POWERSEAL Sealer and Divider

K231327 · Gyrus Acmi, Inc. · GEI · Oct 25, 2023 · General, Plastic Surgery

Device Facts

Record IDK231327
Device NamePOWERSEAL Sealer and Divider
ApplicantGyrus Acmi, Inc.
Product CodeGEI · General, Plastic Surgery
Decision DateOct 25, 2023
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 878.4400
Device ClassClass 2
AttributesTherapeutic

Intended Use

The POWERSEAL Sealer and Divider is a bipolar electrosurgical device intended for use in laparoscopic/minimally invasive or open surgical procedures where ligation of vessels, tissue bundles, and lymphatics is desired. POWERSEAL devices can be used on vessels (arteries and veins, pulmonary arteries, pulmonary veins) up to and including 7 mm, lymphatics, and tissue bundles. POWERSEAL devices are indicated for use in general surgery and such surgical specialties as urologic, colorectal, bariatio, vascular, thoracic, and gynecologic. Procedures may include, but are not limited to. Nissen fundoplication, colectomy, adhesiolysis, sleeve gastrectomy, hysterectomy, oophorectomy. The POWERSEAL Sealer and Divider has not been shown to be effective for tubal sterilization or tubal coagulation for sterilization procedures. Do not use the POWERSEAL devices for these procedures.

Device Story

Bipolar electrosurgical instrument; used in laparoscopic/minimally invasive or open surgery; operated by surgeons. Device features pistol-grip handle, shaft (23, 37, 44 cm), and jaws. Input: manual user control of jaw closure and RF energy delivery. Operation: connects to Olympus electrosurgical generators; delivers bipolar RF energy to seal vessels/tissue; separate manual control activates integral blade to divide tissue along seal line. Output: sealed and divided tissue. Benefits: provides hemostasis and tissue separation; reduces need for multiple instruments. Used in clinical settings; disposable, single-use.

Clinical Evidence

No clinical studies performed. Evidence based on bench testing (vessel burst pressure, electrical safety per IEC 60601-1/2-2, EMC per IEC 60601-1-2) and preclinical animal studies (acute and chronic) evaluating seal performance, thermal safety, and tissue effects.

Technological Characteristics

Bipolar electrosurgical device; pistol-grip handle; 5mm shaft (23, 37, 44 cm); curved/straight jaws. Materials: plastics and stainless steel (ISO 10993 compliant). Energy: RF/HF bipolar. Standards: IEC 60601-1, IEC 60601-1-2, IEC 60601-2-2, ISO 10993, ISO 14971. Sterile, single-use. No software.

Indications for Use

Indicated for ligation of vessels (up to 7mm), lymphatics, and tissue bundles in general, urologic, colorectal, bariatric, vascular, thoracic, and gynecologic surgery. Contraindicated for tubal sterilization or tubal coagulation.

Regulatory Classification

Identification

An electrosurgical cutting and coagulation device and accessories is a device intended to remove tissue and control bleeding by use of high-frequency electrical current.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health and Human Services logo on the left and the FDA text logo on the right. The FDA text logo is in blue and includes the acronym "FDA" followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in a stacked format. October 25, 2023 Gyrus ACMI, Inc. Dolan Mills Program Manager, Regulatory Affairs 800 West Park Drive Westborough, Massachusetts 01581 Re: K231327 Trade/Device Name: POWERSEAL Sealer and Divider Regulation Number: 21 CFR 878.4400 Regulation Name: Electrosurgical Cutting And Coagulation Device And Accessories Regulatory Class: Class II Product Code: GEI Dated: September 21, 2023 Received: September 21, 2023 Dear Dolan Mills: We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download). {1}------------------------------------------------ Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the OS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181). Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4. Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803). please go to https://www.fda.gov/medical-device-safety/medical-device-reporting-mdr-howreport-medical-device-problems. For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and -CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatorythe DICE assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). > Sincerely, Digitally signed by Mark Mark Trumbore -S Trumbore -S 14:51:48-04'00' Mark Trumbore, Ph.D. Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health {2}------------------------------------------------ # Enclosure {3}------------------------------------------------ ## Indications for Use 510(k) Number (if known) K231327 #### Device Name POWERSEAL Sealer and Divider Model numbers: PS-0523CJSA, PS-0537CJSA, PS-0544CISA, PS-0523SJDA, PS-0537SJDA, PS-0544SJDA #### Indications for Use (Describe) The POWERSEAL Sealer and Divider is a bipolar electrosurgical device intended for use in laparoscopic/minimally invasive or open surgical procedures where ligation of vessels, tissue bundles, and lymphatics is desired. POWERSEAL devices can be used on vessels (arteries and veins, pulmonary arteries, pulmonary veins) up to and including 7 mm, lymphatics, and tissue bundles. POWERSEAL devices are indicated for use in general surgery and such surgical specialties as urologic, colorectal, bariatio, vascular, thoracic, and gynecologic. Procedures may include, but are not limited to. Nissen fundoplication, colectomy, adhesiolysis, sleeve gastrectomy, hysterectomy, oophorectomy. The POWERSEAL Sealer and Divider has not been shown to be effective for tubal sterilization or tubal coagulation for sterilization procedures. Do not use the POWERSEAL devices for these procedures. | Type of Use (Select one or both, as applicable) | |-------------------------------------------------| |-------------------------------------------------| X Prescription Use (Part 21 CFR 801 Subpart D) | Over-The-Counter Use (21 CFR 801 Subpart C) #### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {4}------------------------------------------------ ### K231327 510(k) Summary #### General Information Manufacturer: Establishment Registration Number: 510(k) Submitter: Establishment Registration Number: Contact Person: Date Prepared: #### Device Description Classification Name: Regulatory Class Regulation Number Review Panel Product Code Trade / Proprietary Name(s): Generic/Common Name: Model Numbers: Gyrus ACMI, Inc., an Olympus company 9600 Louisiana Ave. North Brooklyn Park, MN 55445 USA Phone: 1-763-416-3000 3011050570 Gyrus ACMI, Inc. 800 West Park Dr. Westborough, MA 01581 3003790304 Dolan Mills Program Manager, Regulatory Affairs May 8, 2023 Electrosurgical cutting and coagulation device and accessories Class 2 21 CFR 878.4400 General & Plastic Surgery GEI POWERSEAL Sealer and Divider Electrosurgical, Cutting & Coagulation PS-0523CJSA, PS-0537CJSA, PS-0544CJSA, PS-0523SJDA, PS-0537SJDA, PS-0544SJDA {5}------------------------------------------------ #### Predicate Device #### POWERSEAL K212643, K203682 The predicate K212643 demonstrated POWERSEAL compatibility with the ESG-410 electrosurgical generator, with no other device changes as compared to its predicate, K203682. K203682 was the initial submission for POWERSEAL and included compatibility with the ESG-400 electrosurgical generator. The predicate has not been subject to a design-related recall. ### Device Description and Technological Characteristics The POWERSEAL 5mm laparoscopic jaw sealer divider is an electrosurgical bipolar device with an integral extending cutting blade. It features a pistol grip handle and is provided in shaft lengths of 23, 37, and 44 cm. The POWERSEAL devices will be provided as sterile, single-use, hand-held bipolar electrosurgical instruments designed for use with Olympus electrosurgical generators to ligate (seal) and divide (cut) vessels, tissue bundles, and lymphatics. The jaws of the POWERSEAL are designed to seal vessels, and grasp and dissect tissue during open and minimally invasive general surgical procedures using radiofrequency (RF) energy, otherwise known as high frequency (HF) energy. A hand-actuated mechanism allows the user to open and close the instrument jaws. When the instrument jaws are correctly placed over the tissue or vessel to be sealed, the user operates a second control to initiate delivery of bipolar energy, which seals the tissue. When the sealing is complete, the user operates a separate control to activate a blade, which divides the tissue along the seal line. The basic fundamental technology, including design, intended use, and principles of operation are the same between the subject and predicate devices. Both the subject and predicate devices connect to an RF energy source The Indications for Use statement is the same as the predicate device. The table below outlines the similarities and differences between the subject and predicate devices. #### Intended Use / Indications The POWERSEAL Sealer and Divider is a bipolar electrosurgical device intended for use in laparoscopic/minimally invasive or open surgical procedures where ligation and division of vessels, tissue bundles, and lymphatics is desired. POWERSEAL devices can be used on vessels (arteries and veins, pulmonary arteries, pulmonary veins) up to and including 7 mm, lymphatics, and tissue bundles. POWERSEAL devices are indicated for use in general surgery and such surgical specialties as urologic, colorectal, bariatric, vascular, thoracic, and gynecologic. Procedures may include, but are not limited to. Nissen fundoplication, colectomy, cholecystectomy, adhesiolysis, sleeve gastrectomy, hysterectomy, oophorectomy. {6}------------------------------------------------ The POWERSEAL Sealer and Divider has not been shown to be effective for tubal sterilization or tubal coagulation for sterilization procedures. Do not use the POWERSEAL devices for these procedures. ## Compliance to Standards The following standards were used during the design and testing of the subject device: | Applied standards | |--------------------------------------| | IEC 60601-1:2005+AMD1:2012+AMD2:2020 | | IEC 60601-1-2 Edition 4.1: 2020 | | IEC 60601-2-2 Ed. 6.0: 2017 | | IEC 62366-1, Edition 1.0, 2015 | | ISO 14971: 2019 | | ISO 11135: 2014+A1:2018 | | ISO 11607-1: 2019 | | ASTM F1980-16: 2021 | | ISO 10993-5: 2009 | | ISO 10993-1:2018 | | ISO 10993-10: 2021 | | ISO 10993-11: 2017 | | ISO 10993-7: 2019 | | ISO 10993-18:2020 | ## Summary of Performance Testing ## Performance Testing Bench For the subject device all data was prepared in accordance with the following FDA guidance documents: "Premarket Notification (510(k)) Submissions for Bipolar Electrosurgical Vessel Sealers for General Surgery," Guidance for Industry and Food and Drug Administration Staff, issued on August 15, 2016; and "Premarket Notification (510(k)) Submissions for Electrosurgical Devices for General Surgery," Guidance for Industry and Food and Drug Administration Staff, issued on March 9, 2020. The guidance documents were followed for all relevant sections. Verification and comparison bench studies were conducted to evaluate the functional performance of the "POWERSEAL" mode. Ex-vivo Vessel Burst Pressure testing was conducted on both the subject and predicate devices to demonstrate vessel sealing performance. System testing demonstrated that the performance requirements defined in the User Requirements Specification and Design Specification were met for the subject devices, and that they exhibit comparable performance characteristics to the predicate device. {7}------------------------------------------------ Bench testing results support the claim of substantial equivalence of the subject device to the predicate device. The following non-clinical and preclinical tests were conducted: - 1) non-clinical (electrical, mechanical, functional) - 2) preclinical (simulated use) evaluation and testing of tissue effects and thermal safety and vessel burst pressure testing and vessel thermal margin Usability and user interface were also assessed according to the risk management plan. Use-related hazardous situations were assessed and risk mitigation measures in terms of usability design for safety were defined. The residual risk was evaluated as acceptable. Risk analysis was carried out in accordance with established internal acceptance criteria based on ISO 14971. Electrical safety and EMC compatibility: Basic safety and performance testing was performed in accordance with IEC 60601-1, IEC 60601-1-2, and IEC 60601-2-2. Mechanical and Functional: Verification and comparison bench studies were conducted to evaluate the mechanical and functional performance as compared to the predicate. Stability: The subject devices have the same packaging and shelf-life as the predicate. Real time age testing will confirm the declared shelf-life. Software: The POWERSEAL Sealer and Divider does not contain software. ### Material The composition of the subject device is nearly identical to the predicate device. Patient contacting materials are primarily plastics and stainless steel. Any differences in materials between the subject and predicate devices were subject to biocompatibility evaluation against ISO 10993. ### Biocompatibility The subject devices are classified in accordance with ISO 10993-1, as an External Communication Device, Tissue/Bone/Dentin, for limited exposure (<24 hours.). In accordance with ISO 10993-1 and the 2016 FDA guidance. Use of International Standard ISO 10993-1, "Biological evaluation of medical devices – Part 1: Evaluation and testing within a risk management process", the subject devices meet all acceptance criteria for the following biocompatibility evaluations: Cytotoxicity, Material Mediated Pyrogen, ISO Acute Systemic Injection Toxicity, ISO Intracutaneous Irritation, and ISO Guinea Pig Maximization Sensitization. ### Clinical and Animal Studies Clinical studies were not necessary for the subject devices. {8}------------------------------------------------ Animal Studies, including Acute and Chronic testing conducted demonstrate substantial equivalence of the subject POWERSEAL Sealer and Divider to the predicate device. | Test | Contents | |----------------------|-------------------------------------------------------------------------------------------------------------------------------------------| | Chronic Animal Study | Chronic animal study was conducted on the subject device to demonstrate seal performance for the indications | | Acute Animal Study | Acute animal study was conducted on both the subject and predicate devices to demonstrate seal performance and safety for the indications | Performance testing demonstrated that the device is as effective, and performs as well as the predicate devices. ## Sterilization Sterilization for the subject device is the exact same as the predicate device. ### Substantial Equivalence In establishing substantial equivalence of the subject POWERSEAL Sealer and Divider to the predicate device, an evaluation of the indications for use, intended use and technological characteristics was conducted. The subject and predicate devices have the same technology, performance, dimensions, and materials. Performance testing confirmed that the subject device is as safe and effective as the predicate device for the proposed indications for use. | Summary of differences and similarities between the subject and predicate devices | | | |-----------------------------------------------------------------------------------|--------------------------------------------------------|---------------------------------------| | Description | Subject Device | Predicate Device<br>(K212643/K203682) | | Intended Use | Exactly the same as the predicate | Same | | Design | Pistol Grip with Shaft and curved jaw and straight jaw | Pistol Grip with Shaft and curved jaw | | Prescription/over- the-counter use | Rx Only | Same | | Size(s) | 5mm diameter in lengths of 23, 37, 44cm | Same | | Output mode and name | Bipolar Sealing = POWERSEAL (Coag) | Same | | Sterile, single use, disposable | Yes | Same | ## Conclusion The performance of the subject device was compared against performance requirements and the predicate listed above. Performance requirements were based on the predicate device and/or its predicate. Testing demonstrated that the performance requirements were met, and that the subject device exhibited comparable performance characteristics to the predicate. Any differences have {9}------------------------------------------------ been validated and demonstrate that the technological differences do not raise different questions of safety and efficacy. In summary, the Gyrus ACMI POWERSEAL Sealer and Divider is substantially equivalent to the predicate device and does not raise different questions of safety and effectiveness.
Innolitics
510(k) Summary
Decision Summary
Classification Order
Enter a record ID and click Load to view the document.
100%