Perforated Baskets

K231291 · Richard Wolf Medical Instruments Corporation · KCT · Jun 3, 2024 · General Hospital

Device Facts

Record IDK231291
Device NamePerforated Baskets
ApplicantRichard Wolf Medical Instruments Corporation
Product CodeKCT · General Hospital
Decision DateJun 3, 2024
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 880.6850
Device ClassClass 2

Indications for Use

The products are used to store medical devices before, during, and after sterilization. The products are intended to enable sterilization of the enclosed medical devices but are not intended to maintain sterility on their own; instead, they are used in conjunction with an FDA-approved sterilization wrap that is legally available on the market.

Device Story

Perforated Baskets consist of metal base and lid with locking mechanism; optional instrument carriers/adapters for multi-level storage; small parts basket included. Used in clinical settings by professionals to house medical devices during steam sterilization. Input: medical devices placed inside basket. Operation: perforated structure allows steam penetration during pre-vacuum steam sterilization (132°C, 3 min exposure, 20 min dry time). Output: sterilized medical devices contained within basket. Device does not maintain sterility post-sterilization; must be used with FDA-cleared sterilization wrap. Benefits: facilitates organized, effective steam sterilization of surgical instruments; supports maximum load of 4.5 kg.

Clinical Evidence

Bench testing only. Cleaning validation performed per AAMI TIR 30:2011 and ANSI/AAMI ST98:2022; results showed no visible soil, protein < 6.4 µg/cm², and hemoglobin < 2.2 µg/cm². Steam sterilization validation performed per EN ISO 14937, Annex D, achieving Sterility Assurance Level (SAL) > 10⁻⁶.

Technological Characteristics

Metal perforated baskets with locking lid mechanism. Reusable. Non-sterile delivery. Steam sterilization (pre-vacuum, 132°C). Maximum load capacity 4.5 kg. No software. No direct/indirect patient contact.

Indications for Use

Indicated for professional use to store medical devices during sterilization processes. No specific patient population or medical indications. Contraindicated for use without an FDA-approved sterilization wrap.

Regulatory Classification

Identification

A sterilization wrap (pack, sterilization wrapper, bag, or accessories, is a device intended to be used to enclose another medical device that is to be sterilized by a health care provider. It is intended to allow sterilization of the enclosed medical device and also to maintain sterility of the enclosed device until used.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue. June 3, 2024 Richard Wolf Medical Instruments Corporation Mike McAndrew US Head of Regulatory - OA/OC 353 Corporate Woods Parkway Vernon Hills, Illinois 60061 Re: K231291 Trade/Device Name: Perforated Baskets Regulation Number: 21 CFR 880.6850 Regulation Name: Sterilization Wrap Regulatory Class: Class II Product Code: KCT Dated: April 30, 2024 Received: May 1, 2024 Dear Mike McAndrew: We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" {1}------------------------------------------------ (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download). Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review. the OS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181). Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely. Christopher K. Dugard -S Christopher K. Dugard, M.S. Assistant Director DHT4B: Division of Infection Control and Plastic and Reconstructive Surgery Devices OHT4: Office of Surgical {2}------------------------------------------------ and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {3}------------------------------------------------ # Indications for Use 510(k) Number (if known) K231291 Device Name Perforated Baskets Indications for Use (Describe) The products are used to store medical devices before, during, and after sterilization. The products are intended to enable sterilization of the enclosed medical devices but are not intended to maintain sterility on their own; instead, they are used in conjunction with an FDA-approved sterilization wrap that is legally available on the market. The products have no medical indications. Sterilization cycle times are as follow: Exposure time: 3 min Temperature: 132 ℃ Drying time: 20 min The maximum validated load weight for all devices listed below is 4.5 kg (approx. 9.92 lbs). All measurements are in millimeter and are coded as follows [WidthxHeightxLength]. UNIVERSAL PERFORATED BASKET 1 (33001); 732x117x282) UNIVERSAL PERFORATED BASKET 2 (33002); (743x200x282) UNIVERSAL PERFORATED BASKET 3 (33003); (538x117x250) UNIVERSAL PERFORATED BASKET 4 (33004); (538x117x250) UNIVERSAL PERFORATED BASKET 5 (33005); (549x200x250) PERFORATED BASKET FOR SMALL PARTS (33006); (128x40x121) PERF. BASKET – KNEE ARTHROSC. BASIC SET (33007); (549x150x250) PERF. BASKET - KNEE ARTHROSC. ACL/PCL 1 (33008); (549x150x250) PERF. BASKET - KNEE ARTHROSC. ACL/PCL 2 (33009); (549x150x250) PERF. BASKET FOR KNEE ARTHROSC. PREP TAB (33010); (549x150x250) PERF. BASKET FOR VERTEBRIS LUMBAR (33011); (549x150x250) PERF. BASKET FOR SHOULDER ARTHROSCOPY (33012); (538x117x250) PERF. BASKET FOR VERTEBRIS FORAMINOTOMY (33013); (525x65x255) PERF. BASKET FOR VERTEBRIS STENOSIS (33015); (549x150x250) PERFORATED BASKET FOR CAMERA HEAD (38002); (448x73x200) PERFORATED BASKET, UPPER URINARY TRACT 1 (38003); (635x75x132) PERF. BASKET FOR CYSTOSCOPE/TELESCOPE (38004); (467x75x132) PERFORATED BASKET FOR DISCOSCOPE (38005); (467x75x132) PERFORATED BASKET FOR HYSTEROSCOPE 1 (38006); (467x75x132) PERF. BASKET FOR VIDEO MEDIASTINOSCOPE (38007); (256x85x250) PERFORATED BASKET FOR TELESCOPE (38008); (467x75x132) PERFORATED BASKET FOR HYSTEROSCOPE 2 (38009); (467x90x132) PERFORATED BASKET FOR STEREO TELESCOPE (38010); (477x139x196) PERFORATED BASKET FOR NEPHROSCOPE (38011); (467x75x172) PERFORATED BASKET FOR MORCESCOPE (38012); (467x75x172) PERFORATED BASKET FOR CYSTO-URETHROSCOPE (38013); (451x64x80) PERFORATED BASKET FOR 3D TELESCOPE (38014); (583x80x200) PERF. BASKET FOR STANDARD TELESCOPE 1 (38015); (297x54x59) PERF. BASKET FOR STANDARD TELESCOPE 2 (38016); (481x54x59) PERF. BASKET FOR STANDARD TELESCOPE 3 (38017); (619x54x59) PERFORATED BASKET FOR PAIN THERAPY (33014); (525x95x255) {4}------------------------------------------------ #### Type of Use (Select one or both, as applicable) X Prescription Use (Part 21 CFR 801 Subpart D) #### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {5}------------------------------------------------ Image /page/5/Picture/1 description: The image is a logo for Richard Wolf. The text "RICHARD WOLF" is displayed in a bold, sans-serif font. The word "RICHARD" is in black, while the word "WOLF" is in red. To the right of the text is a square with rounded corners, containing a stylized image of a wolf with a nail in its mouth. The wolf is drawn in black lines, and the nail is pointing downwards. # K231291 510(k) # Summary # Submitter Richard Wolf Medical Instruments Corporation 353 Corporate Woods Parkway Vernon Hills, IL 60061 Phone:(847) 913 1113 Contact Person, Title: Mike McAndrew, US Head of Regulatory -QA/QC Date Prepared:05/30/2024 # Legal Manufacturer Richard Wolf GmbH Pforzheimer Straße 32 75438 Knittlingen, Germany ## 1. Devices | Device Trade Name | Device classification name | Regulation<br>Number and<br>Name | Product<br>Code | Device<br>Class | Review<br>Panel | |-----------------------|---------------------------------------------------------------------------|----------------------------------|-----------------|-----------------|---------------------| | PERFORATED<br>BASKETS | Sterilization Wrap Containers,<br>Trays, Cassettes & Other<br>Accessories | 880.6850<br>General Hospital | KCT | II | General<br>Hospital | {6}------------------------------------------------ Image /page/6/Picture/1 description: The image contains the words "RICHARD WOLF" in a stylized font. The word "RICHARD" is in black, while the word "WOLF" is in red. To the right of the words is a logo of a wolf's head with a medical instrument going through it, all enclosed in a rounded square. ### 2. Predicate Device Name of Predicate Device: KARL STORZ Metal Sterilization Trays 510(k) Number: K203198 Regulatory Class: II Product Code: KCT Manufacturer: Karl Storz Endoscopy America Inc # 3. Device Description # 3.1. Subject Device Identification The PERFORATED BASKETS all have the same basic configuration, a metal basket base and a basket lid, with optional product specific instrument carriers (for multiple levels) or adapters and an additional basket for small parts. The base part and the lid of the PERFORATED BASKET are connected by a locking mechanism, which assures a tight connection of both parts. ### 3.2. Subject device characteristics #### User PERFORATED BASKETS are used for professional use only. ### Delivered sterile / non-sterile PERFORATED BASKETS are delivered in a non-sterile state. ### Single use / reusable PERFORATED BASKETS are reusable. ### Software PERFORATED BASKETS do not have software. #### Materials with patient contact PERFORATED BASKETS do not have direct or indirect contact with the patient. ### 4. Indications for Use The products are used to store medical devices before, during, and after sterilization. The products are intended to enable sterilization of the enclosed medical devices but are not intended to maintain sterility on their own; instead, they are used in conjunction with an FDA-approved sterilization wrap that is legally available on the market. The products have no medical indications. Sterilization cycle times are as follow: Exposure time: 3 min Temperature: 132 °C Drying time: 20 min {7}------------------------------------------------ Image /page/7/Picture/1 description: The image shows the logo for Richard Wolf. The words "RICHARD" and "WOLF" are stacked on top of each other, with "RICHARD" in a smaller font size than "WOLF". The word "WOLF" is in a bold, red font. To the right of the text is a square with rounded corners, inside of which is a stylized drawing of a wolf's head and a nail. The maximum validated load weight for all devices listed below is 4.5 kg (approx. 9.92 lbs). All measurements are in millimeter and are coded as follows [WidthxHeightxLength]. UNIVERSAL PERFORATED BASKET 1 (33001); 732x117x282) UNIVERSAL PERFORATED BASKET 2 (33002); (743x200x282) UNIVERSAL PERFORATED BASKET 3 (33003); (538x117x250) UNIVERSAL PERFORATED BASKET 4 (33004); (538x117x250) UNIVERSAL PERFORATED BASKET 5 (33005); (549x200x250) PERFORATED BASKET FOR SMALL PARTS (33006); (128x40x121) PERF. BASKET - KNEE ARTHROSC. BASIC SET (33007); (549x150x250) PERF. BASKET – KNEE ARTHROSC. ACL/PCL 1 (33008); (549x150x250) PERF. BASKET – KNEE ARTHROSC. ACL/PCL 2 (33009); (549x150x250) PERF. BASKET FOR KNEE ARTHROSC. PREP TAB (33010); (549x150x250) PERF. BASKET FOR VERTEBRIS LUMBAR (33011); (549x150x250) PERF. BASKET FOR SHOULDER ARTHROSCOPY (33012); (538x117x250) PERF. BASKET FOR VERTEBRIS FORAMINOTOMY (33013); (525x65x255) PERF. BASKET FOR VERTEBRIS STENOSIS (33015); (549x150x250) PERFORATED BASKET FOR CAMERA HEAD (38002); (448x73x200) PERFORATED BASKET, UPPER URINARY TRACT 1 (38003); (635x75x132) PERF. BASKET FOR CYSTOSCOPE/TELESCOPE (38004); (467x75x132) PERFORATED BASKET FOR DISCOSCOPE (38005): (467x75x132) PERFORATED BASKET FOR HYSTEROSCOPE 1 (38006); (467x75x132) PERF. BASKET FOR VIDEO MEDIASTINOSCOPE (38007); (256x85x250) PERFORATED BASKET FOR TELESCOPE (38008); (467x75x132) PERFORATED BASKET FOR HYSTEROSCOPE 2 (38009); (467x90x132) PERFORATED BASKET FOR STEREO TELESCOPE (38010); (477x139x196) PERFORATED BASKET FOR NEPHROSCOPE (38011); (467x75x172) PERFORATED BASKET FOR MORCESCOPE (38012); (467x75x172) PERFORATED BASKET FOR CYSTO-URETHROSCOPE (38013); (451x64x80) PERFORATED BASKET FOR 3D TELESCOPE (38014); (583x80x200) PERF. BASKET FOR STANDARD TELESCOPE 1 (38015); (297x54x59) PERF. BASKET FOR STANDARD TELESCOPE 2 (38016); (481x54x59) PERF. BASKET FOR STANDARD TELESCOPE 3 (38017); (619x54x59) PERFORATED BASKET FOR PAIN THERAPY (33014); (525x95x255) {8}------------------------------------------------ Image /page/8/Picture/1 description: The image is a logo for "Richard Wolf". The words "Richard Wolf" are stacked on top of each other, with "Richard" in a smaller, thinner font than "Wolf". To the right of the text is a square with rounded corners, inside of which is a stylized drawing of a wolf's head with a nail being hammered into the wolf's head. # 5. Comparison of Technological Characteristics with the Predicate Device The PERFORATED BASKETS are comparable to KARL STORZ Metal Sterilization Trays. Sterilization occurs through the perforated structure in the tray base and lid. After sterilization, the FDA cleared sterilization wrap maintains sterility. All of these characteristics are the same as the predicate device. The devices and the predicate have essentially the same indications for use and perform in a similar manner. | Item | Subject devices | Predicate devices | Explanation of<br>differences | |-----------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------| | 510(k) number | K231291 | K203198 | N/A | | Manufacturer | Richard Wolf GmbH | Karl Storz Endoscopy<br>America Inc | N/A | | Product Code | KCT | KCT | None. | | Regulation Number | 880.6850 | 880.6850 | None | | Indications for Use | Perforated baskets<br>The products are used to store<br>medical devices before,<br>during, and after sterilization.<br>The products are intended to<br>enable sterilization of the<br>enclosed medical devices but<br>are not intended to maintain<br>sterility on their own; instead,<br>they are used in conjunction<br>with an FDA-approved<br>sterilization wrap that is legally<br>available on the market.<br>The products have no medical<br>indications.<br>Sterilization cycle times are as<br>follows:<br>Exposure time: 3 min<br>Temperature: 132 °C<br>Drying time: 20 min<br>The maximum validated load<br>weight for all devices listed<br>below is 4.5 kg (approx. 9.92<br>lbs).<br>All measurements are in<br>millimeter and are coded as<br>follows [WidthxHeightxLength].<br>Authors note: At this position<br>corresponding devices and<br>their dimensions are listed in<br>every Instruction for use. | The KARL STORZ Metal<br>Sterilization Trays are<br>intended only for use to<br>encase and protect specific<br>KARL STORZ reusable<br>medical devices for<br>sterilization in prevacuum<br>steam sterilization cycles<br>132°C for four minutes with<br>a 20 minute dry time).<br>When used in conjunction<br>with an FDA cleared<br>sterilization wrap, sterility of<br>the enclosed medical<br>device is maintained until<br>used. | Similar. | | Prescription/<br>over-<br>the-counter-use | Prescription Use | Over the Counter | Different | | Cleaning<br>and<br>Sterilization method<br>for the basket | Devices must be reprocessed<br>before the first usage. | Information not available | Different | | Item | Subject devices | Predicate devices | Explanation of<br>differences | | Validated<br>Sterilization Cycle | Pre-Vacuum<br>Exposure time: 3 min,<br>Temperature: 132°C | Pre-Vacuum<br>Exposure time: 4min,<br>Temperature: 132°C | Similar | | Load condition | Maximum weight and loading<br>conditions are either limited by<br>the place holders for the<br>intended devices if the<br>submitted can hold more than<br>one device, the loading<br>template for specific<br>instruments sets or the weight<br>of the individual medical<br>device if the submitted device<br>is only intended to reprocess<br>one item. | Maximum weight and<br>loading conditions are<br>either limited by the place<br>holders for the intended<br>devices if the predicate<br>device can hold more than<br>one device, the loading<br>template for specific<br>instruments sets or the<br>weight of the individual<br>medical device if the<br>predicate device is only<br>intended to reprocess one<br>item. | Different | | Maximum validated<br>load weight | Up to 4.5 kg (approx. 9.92 lbs) | 5 lbs | Different | {9}------------------------------------------------ # K231291 510(k) Summary Image /page/9/Picture/1 description: The image contains the words "RICHARD WOLF" in a stylized font. The word "RICHARD" is in a smaller, sans-serif font above the word "WOLF", which is in a larger, bolder, red font. To the right of the words is a logo featuring a stylized drawing of a wolf's head with a vertical line through it, enclosed in a rounded square. ### 6. Performed evaluation of impact of differences The following testing was performed in order to support device performance. For each device it was evaluated whether specific testing was necessary. The following subsections present an overview including the decision whether specific testing was deemed necessary or not. #### 6.1. Cleaning and Sterilization The devices are delivered in a non-sterile status. All devices need to be processed before each usage. Thus, the cleaning process was validated in compliance with the relevant standards and guidelines and the device was found to be suitable for the described procedures. A sterilization validation for the device was also conducted to validate their performance. #### 6.2. Performance Testing Bench The performance of the devices in regards to their capability to be used for steam sterilization was tested. This sterilization process was validated in compliance with the relevant standards and quidelines and the device was found to be suitable for the described procedure. | Test | Test Method (Standard) | Acceptance Criteria | Results | |------------------------|---------------------------------------------|-----------------------------------------------|-----------| | Manual<br>Cleaning | AAMI TIR 30: 2011,<br>ANSI/ AAMI ST98: 2022 | 1: Visible soil: not visible | 1: passed | | | | 2: Protein/cm2 [µg/cm2]:<br>< 6.4 µg/cm2 | 2: passed | | | | 3: Hemoglobin [µg/cm2]:<br>< 2.2 µg/cm2 | 3: passed | | Steam<br>Sterilization | EN ISO 14937, Annex D | Sterilization assurance<br>level (SAL) > 10-6 | passed | {10}------------------------------------------------ Image /page/10/Picture/1 description: The image is a logo for Richard Wolf. The words "RICHARD WOLF" are stacked on top of each other, with "RICHARD" in black and "WOLF" in red. To the right of the text is a white square with rounded corners, inside of which is a black line drawing of a wolf's head and a medical instrument. # 7. Conclusion The conclusions drawn from the nonclinical testing demonstrate the device in 510(k) K231291, the PERFORATED BASKETS, is as safe, as effective and performs as well, or better, than the legally market predicate device, cleared via 510(k) K203198.
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...