The VYVO Life Watch is a small, wrist-worn device indicated for use in measuring and displaying functional oxygen saturation of arterial hemoglobin (%SpO2) and pulse rate. It is intended for spot-checking of adult patients in hospitals, clinics, long-term care, and home use.
Device Story
The Vyvo Life Watch is a wrist-worn, non-sterile, single-patient use pulse oximeter. It utilizes reflectance pulse oximetry to measure SpO2 and pulse rate. The device features a sensor with two LEDs (660 nm red, 940 nm infrared) and a photodiode. The user initiates measurement by pressing a finger onto the sensor located on the side of the watch. The device processes light absorption changes in the arteriolar bed to calculate physiological parameters, which are displayed on an integrated OLED screen. It is powered by a rechargeable lithium-ion polymer battery and includes a micro-USB connector for charging. The device is intended for spot-checking by adult patients or healthcare providers in clinical and home environments. It does not provide life-supporting or life-sustaining functions. The output allows for immediate assessment of oxygenation and pulse rate, aiding in clinical decision-making for spot-check monitoring.
Clinical Evidence
Clinical validation conducted per ISO 80601-2-61:2017 on 12 healthy adult subjects (6 male, 6 female, ages 19-42) with varying Fitzpatrick skin types. Results showed equivalence to simultaneous measurements from the predicate device. No adverse events were reported.
Indicated for spot-checking functional oxygen saturation (%SpO2) and pulse rate in adult patients in hospitals, clinics, long-term care, and home settings.
Regulatory Classification
Identification
An oximeter is a device used to transmit radiation at a known wavelength(s) through blood and to measure the blood oxygen saturation based on the amount of reflected or scattered radiation. It may be used alone or in conjunction with a fiberoptic oximeter catheter.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: a symbol on the left and the FDA name and title on the right. The symbol on the left is a stylized representation of a human figure, while the text on the right reads "FDA U.S. FOOD & DRUG ADMINISTRATION" in blue letters.
Vyvo Technology Corp.(VT) Aileen Fu Official Correspondent Shenzhen Global Medical Technology Services Co., Ltd Room 1702 17th Floor, Shenzhen Taifeng Building, B0. 86, Qianjin 1st Road, Bao An Shenzhen, Guangdong 8100 China
Re: K231288
March 4, 2024
Trade/Device Name: Vyvo Regulation Number: 21 CFR 870.2700 Regulation Name: Oximeter Regulatory Class: Class II Product Code: DOA Dated: December 15, 2023 Received: February 2, 2024
Dear Aileen Fu:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"
{1}------------------------------------------------
(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review. the OS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic.
{2}------------------------------------------------
See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# Bradley Q. Quinn -S
Bradley Quinn Assistant Director DHT1C: Division of Sleep Disordered Breathing, Respiratory and Anesthesia Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
# Indications for Use
510(k) Number (if known) K231288
Device Name VYVO Life Watch
#### Indications for Use (Describe)
The VYVO Life Watch is a small, wrist-worn device indicated for use in measuring and displaying functional oxygen saturation of arterial hemoglobin (%SpO2) and pulse rate. It is intended for spot-checking of adult patients in hospitals, clinics, long-term care, and home use.
| Type of Use (Select one or both, as applicable) | |
|----------------------------------------------------------------------------------|---------------------------------------------------------------------------------|
| <div> <span> Prescription Use (Part 21 CFR 801 Subpart D) </span> </div> | <div> <span> Over-The-Counter Use (21 CFR 801 Subpart C) </span> </div> |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
# 510(k) Summary
This summary of 510(k) information is submitted as specified in SMDA and 21 CFR §807.92.
| Submission Date | December 15, 2023 |
|--------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Manufacturer<br>information | Submitter's Name: Vyvo Technology Corp.(VT)<br>Address: 123 NW 23 RD Street City Miami, FL<br>Contact person: Alfonso Cioffi<br>TEL: +1 2016470536<br>E-Mail: a.cioffi@helohealth.com |
| Submission<br>Correspondent | Contact person: Fu Ailing<br>Tel: +86-13538216349<br>Email: Aileenfu@coctd.com<br>Company: Shenzhen Global Medical Technology Services Co., Ltd<br>Address: Room 1702, Shenxintaifeng Building, No. 86, Qianjin 1st<br>Road, Bao'an District, Shenzhen, 518101, Guangdong, China |
| Establishment<br>registration number | NA |
# 1 Administrative Information
## 2 Proposed Device Information
| Common name of the device | WL Watch Oximeter |
|----------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Trade name of the device | VYVO Life Watch |
| Type/Model of the device | WL |
| Classification information | Regulation Medical Specialty: Cardiovascular<br>510(k) Review Panel: Anesthesiology<br>Classification name: Oximeter<br>Regulation Number: 21 CFR 870.2700<br>Device Class: II<br>Product Code: DQA |
| Type of 510(k) submission | Traditional |
# 3 Predicate and Reference Devices
Predicate Device:
| Sponsor: | Bio-Beat Technologies Ltd. |
|----------------|----------------------------|
| Device: | BB-613 Watch Oximeter |
| 510(K) Number: | K181006 |
#### Reference Device:
| Sponsor: | Beijing Choice Electronic Technology Company, Limited |
|----------------|-------------------------------------------------------|
| Device: | Wrist Pulse Oximeter |
| 510(K) Number: | K122046 |
{5}------------------------------------------------
VOL 005:001 510(k) Summary Version: 2.0
# Product: Vyvo Life Watch
#### 4 Device Description
The Life Watch, Model WL is a small, wrist-worn device indicated for use in measuring and displaying functional oxygen saturation of arterial hemoglobin (% SpO2) and pulse rate (PR). The expected measuring location of the device is the finger, which actively presses onto the sensor on the side of the watch for measurement. The Life Watch is a watch-like device with a reflectance pulse oximetry sensor located at the side of the reflectance pulse oximetry sensor includes two light emitting diodes (LEDs) of red wavelength (660 nm) and infrared wavelength (940 nm), and one photodiode light detector placed next to each other. Light beams are emitted from LEDs through the skin to the arteriolar bed of the tissue. Changes in light absorption during the pulsing cycle are measured by the photodiode light detector as scattered lights are reflected back from the pulsating arteriolar bed. The functional oxygen saturation of arterial hemoglobin (% SpO2) and pulse rate are measured by the well-established non-invasive pulse oximetry technology where the red and infrared light is absorbed in different amounts depending on the oxygenation of the blood during the arterial pulsing. The maximum optical output power is less than 2 mW. The Life Watch is a single-patient use, non-sterile pulse oximeter. It is available in one configuration as a standalone device with a wrist pulse oximeter and a detachable watchband for wearing the pulse oximeter on the wrist.
The Life Watch is a compact and light weight device which consists of a reflectance pulse oximetry sensor, a color graphic OLED display, a lithium ion polymer rechargeable battery, an analog and digital unit, a microprocessor and an operating software. The functions of the life watch include the following: (1) Measurement and display of the functional oxygen saturation of arterial hemoglobin (% SpO2) and pulse rate; (2) Spot-checking of specific physiological parameters of adult patients; (3) Easy to read OLED graphic display; (4) Peripheral micro-USB connector used as the battery charging base; (5) Built-in rechargeable lithium ion polymer battery (3.8V, 250mAh) (6) Low battery power indicator in OLED graphic display.
The device consists of the watch case, band, charging case, USB cable.
The proposed device is not used for life-supporting or life-sustaining, and not for implant. The device does not contain drug or biological products.
# 5 Intended Use/ Indications for Use
The VYVO Life Watch is a small, wrist-worn device indicated for use in measuring and displaying functional oxygen saturation of arterial hemoglobin (%SpO2) and pulse rate. It is intended for spot-checking of adult patients in hospitals, clinics, long-term care, and home use.
# 6 Comparison
{6}------------------------------------------------
Version: 2.0
| Table 1 Substantial Equivalence Comparison | | | | | |
|------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Item | Subject Device<br>(K231288) | Primary Predicate Device<br>(K181006) | Reference Device<br>(K122046) | Explanation of<br>Differences Between<br>Predicate and<br>Subject Devices | |
| Device Name | VYVO Life Watch | BB-613 Watch Oximeter | Wrist Pulse Oximeter | N/A | |
| Device Model | WL | BB-613 | MD300W4 | N/A | |
| Manufacturer | Vyvo Technology Co., Ltd. | Bio-Beat Technologies Ltd. | Beijing Choice Electronic<br>Technology Company,<br>Limited | N/A | |
| Product Code | DQA | DQA | DQA | Same | |
| Regulation Number | 21 CFR870.2700 | 21 CFR870.2700 | 21 CFR870.2700 | Same | |
| Indication for Use | The VYVO Life Watch is a<br>small, wrist-worn device<br>indicated for use in<br>measuring and displaying<br>functional oxygen<br>saturation of arterial<br>hemoglobin (%SpO2) and<br>pulse rate. It is intended for<br>spot-checking of adult<br>patients in hospitals,<br>clinics, long-term care, and<br>home use. | The BB-613 Watch<br>Oximeter is a small, wrist-<br>worn device indicated for<br>use in measuring and<br>displaying functional<br>oxygen saturation of<br>arterial hemoglobin<br>(%SpO2) and pulse rate. It<br>is intended for spot-<br>checking of adult patients<br>in hospitals, clinics, long-<br>term care, and home use. | The MD300W4 wrist pulse<br>oximeter is a portable non-<br>invasive device intended<br>for spot checking, data<br>collection and recording of<br>arterial oxygen saturation<br>(SpO2) and pulse rate of<br>adult and pediatric patient<br>at home and hospital<br>(including clinical use In<br>internist/surgery,<br>Anesthesia, intensive care<br>and etc). | Same | |
| Intended Population | Adult | Adult | Adult, pediatric | Same | |
| Composition | A main unit | A main unit | A main unit, SpO2 sensor | Same | |
| Principle of<br>Operation | Pulse reflectance<br>technology, Two LED<br>(red + IR) and photo<br>diode absorbs<br>reflected light | Pulse reflectance<br>technology, Four LED<br>(red + IR) and photo<br>diode absorbs<br>reflected light | Pulse reflectance<br>technology, Two<br>(red + IR) and photo<br>diode absorbs<br>reflected light | Same as the<br>reference device<br>K122046. | |
| Intended Application Site | Finger | Wrist | Finger | Same as the<br>reference device<br>K122046 | |
| | | | | | |
| Measurement type | Spot | Spot | Spot | Same | |
| Emitted light peak wavelength | 940nm (IR), 660nm<br>(Red) | 880nm (IR), 650nm<br>(Red) | 940nm (IR), 660nm<br>(Red) | Same as the<br>reference device<br>K122046. | |
| Display Parameter | SpO2, pulse rate | SpO2, pulse rate | SpO2, pulse rate | Same | |
| Sensor | MAX30101 | Unspecified | M-50G | Due to different<br>design scheme, the<br>difference does not<br>raise any new issue of<br>substantial<br>equivalence according<br>to performance test<br>reports. | |
| SpO2 module | Integrated into main unit | Integrated into main unit | M-50G, Insertion into main<br>unit as an accessory | Same | |
| SpO2 Display Range | 70%-100% | 70%-100% | 0%-100% | Same | |
| SpO2 Measuring Range | 70% to 100% | 70% to 100% | 70% to 100% | Same | |
| SpO2 Measuring Accuracy | ±3% for 70% ~ 80%;<br>±3% for 80% ~ 90%;<br>±3% for 90% ~ 100%; | 70%-80%±3%<br>80%-90%±3%<br>90%-100%±3% | 70%-100%±3%<br>90%-100%±2%<br>80%-90%±3%<br>70%-80%±3%<br>>70% unspecified | Same | |
| SpO2 Resolution | 1% | 70%-100%, ±3% or ±3<br>digits<br>Display: 2 characters | 1% | Same as the<br>reference device<br>K122046. | |
| Pulse Rate Measuring Range | 40 to 200 bpm | 40 to 240 bpm | 30 to 235 bpm | Difference but is<br>within the PR<br>measuring range of<br>the predicate device<br>or the reference<br>device. | |
| Pulse Rate Measuring Accuracy | ±2bpm or ±2% (whichever<br>is greater) | ±3bpm | ±2bpm or 2% ( the greater) | Same as the<br>reference device<br>K122046. | |
| PR Resolution | 1bpm | ±3% or ±3 digits | 1bpm | Same as the | |
| | | | Display: 3 characters | | reference device<br>K122046. |
| Patient-Contacting Components/<br>Materials | | Case/Polycarbonate;<br>Photodiode window;<br>Watchband/silicone | Case/Polycarbonate;<br>Photodiode window;<br>Watchband/Silicone | Belt/Medical silica gel and<br>Nylon | Same |
| Contact Duration | | Prolonged contact duration<br>>24h to 30d | Prolonged contact duration<br>>24h to 30d | Prolonged contact duration<br>>24h to 30d | Same |
| Application Method | | User wears the<br>device as a watch<br>and powers it on | User wears the<br>device as a watch<br>and powers it on | User wears the device on<br>finger | Same |
| Sterility | | Supplied and used<br>non-sterile | Supplied and used<br>non-sterile | Supplied and used<br>non-sterile | Same |
| Display | | OLED | LCD on device | OLED | Same as the<br>reference device<br>K122046. |
| Data storage | | No | No | 72-Hour | Same |
| Transmission mode | | No | Bluetooth | USB or GPRS | N/A |
| Working Time | | About 5-7 days of<br>continuous use | 5 days of continuous use | Work for 10h continuously | Same or better |
| Power supply | | 3.8V<br>Lithium-ion rechargeable<br>battery | 5V<br>Lithium-ion rechargeable<br>battery | 4.2V<br>Lithium-ion rechargeable<br>battery | Due to different<br>design scheme, the<br>difference does not<br>raise any new issue of<br>substantial<br>equivalence according<br>to test reports on<br>battery and device. |
| Type of protection against Electrical<br>shock | | Type BF | Type BF | Type BF | Same |
| Operating<br>Environment | Temperature | 0°C ~35°C | 4°C to 39°C<br>(39°F to 103°F) | 5°C~ 40°C | The slightly different<br>operating temperature<br>does not raise any<br>new issue of<br>substantial<br>equivalence according |
| | | | | | |
| | Relative<br>Humidity | 10%~95%, non-condensing | Up to 95%, non-<br>condensing | ≤80%, non-condensing | to IEC 60601-1 test<br>report.<br>Difference but is<br>within the range of the<br>predicate device. |
| | Atmospheric<br>pressure | 70kPa~106kPa | 700 to 1060hPa<br>(70kPa~106kPa) | 86kPa~106kPa | Same |
| Storage and<br>Transportation<br>Environment | Temperature | 5°C ~28°C | -20°C to 70°C<br>(-4°F to 158° F) | -20℃~55℃ | Difference but is<br>within the range of the<br>predicate device or<br>the reference device. |
| | Relative<br>Humidity | 20%~75%, non-condensing | Up to 95%, non-<br>condensing | ≤93%, non-condensation | Difference but is<br>within the range of the<br>predicate device or<br>the reference device. |
| | Atmospheric<br>pressure | 70kPa~106kPa | 465 hpa to 1060hPa<br>(46.5kPa~106kPa) | 50kPa~106kPa | Difference but is<br>within the range of the<br>predicate device or<br>the reference device. |
| Enclosure | | 49x37x137mm | 48X38X16mm | 133mmx63mmx33mm | Difference but does<br>not raise any new<br>issue of substantial<br>equivalence |
| SpO2 Sensor | | Built in main unit | Built in main unit | Accessory of main unit | Same |
| EMC and<br>Electrical Safety | Electrical Safety | Compliance with IEC<br>60601-1 | Compliance with IEC<br>60601-1 | Compliance with IEC<br>60601-1 | Same |
| | | Compliance with IEC<br>60601-1-11 | Compliance with IEC<br>60601-1-11…
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.