Natural Cycles

K231274 · Natural Cycles Nordic AB · PYT · Aug 24, 2023 · Obstetrics/Gynecology

Device Facts

Record IDK231274
Device NameNatural Cycles
ApplicantNatural Cycles Nordic AB
Product CodePYT · Obstetrics/Gynecology
Decision DateAug 24, 2023
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 884.5370
Device ClassClass 2
AttributesSoftware as a Medical Device

Indications for Use

Natural Cycles is a stand-alone software application, intended for women 18 years and older, to monitor their fertility. Natural Cycles can be used for preventing a pregnancy (contraception) or planning a pregnancy (conception).

Device Story

Natural Cycles is a standalone web/mobile software application for fertility monitoring. Users input daily basal body temperature (BBT) via manual entry, Oura Ring, or Apple Watch; menstrual cycle dates; and optional ovulation/pregnancy test results. A proprietary algorithm processes these inputs to determine daily fertility status. The device operates in three modes: NC° Birth Control (identifies fertile/non-fertile days to guide barrier method use or abstinence), NC° Plan Pregnancy, and NC° Follow Pregnancy. Used in home environments by patients. Output provides fertility status (fertile/non-fertile), ovulation date, and educational information. Helps users make informed decisions regarding pregnancy prevention or planning.

Clinical Evidence

Clinical study compared Apple Watch temperature data to manual oral thermometer (two-decimal) inputs in 272 users across 1918 cycles (505 complete). Ovulation identification was validated against 153 positive luteinizing hormone (LH) tests. Results showed the algorithm successfully identified ovulation using Apple Watch data. The Apple Watch input resulted in 0.6 fewer 'green' (non-fertile) days in the luteal phase compared to oral thermometer, with no increased risk of unintended pregnancy.

Technological Characteristics

Standalone software application; mobile/web-based. Inputs: manual BBT, Oura Ring, or Apple Watch. Algorithm: proprietary fertility status determination. Connectivity: smartphone/cloud-based data transfer. Software level of concern: major.

Indications for Use

Indicated for women 18+ to monitor fertility for contraception or conception. Not for women with medical conditions where pregnancy poses significant risk. Caution advised for those who used hormonal birth control within 60 days.

Regulatory Classification

Identification

A software application for contraception is a device that provides user-specific fertility information for preventing a pregnancy. This device includes an algorithm that performs analysis of patient-specific data (e.g., temperature, menstrual cycle dates) to distinguish between fertile and non-fertile days, then provides patient-specific recommendations related to contraception.

Special Controls

1. Clinical performance testing must demonstrate the contraceptive effectiveness of the software in the intended use population. 2. Human factors performance evaluation must be provided to demonstrate that the intended users can self-identify that they are in the intended use population and can correctly use the application, based solely on reading the directions for use for contraception. 3. Software verification, validation, and hazard analysis must be performed. Documentation must include the following: a. A cybersecurity vulnerability and management process to assure software functionality; and b. A description of the technical parameters of the software, including the algorithm used to determine fertility status and alerts for user inputs outside of expected ranges. 4. Labeling must include: The following warnings and precautions: a. i. A statement that no contraceptive method is 100% effective. ii. A statement that another form of contraception (or abstinence) must be used on days specified by the application. iii. Statements of any factors that may affect the accuracy of the contraceptive information. iv. A warning that the application cannot protect against sexually transmitted infections. b. Hardware platform and operating system requirements. Instructions identifying and explaining how to use the software application, including C. required user inputs and how to interpret the application outputs. d. A summary of the clinical validation study and results, including effectiveness of the application as a stand-alone contraceptive and how this effectiveness compares to other forms of legally marketed contraceptives.

*Classification.* Class II (special controls). The special controls for this device are:(1) Clinical performance testing must demonstrate the contraceptive effectiveness of the software in the intended use population. (2) Human factors performance evaluation must be provided to demonstrate that the intended users can self-identify that they are in the intended use population and can correctly use the application, based solely on reading the directions for use for contraception. (3) Software verification, validation, and hazard analysis must be performed. Documentation must include the following: (i) A cybersecurity vulnerability and management process to assure software functionality; and (ii) A description of the technical parameters of the software, including the algorithm used to determine fertility status and alerts for user inputs outside of expected ranges. (4) Labeling must include: (i) The following warnings and precautions: (A) A statement that no contraceptive method is 100% effective. (B) A statement that another form of contraception (or abstinence) must be used on days specified by the application. (C) Statements of any factors that may affect the accuracy of the contraceptive information. (D) A warning that the application cannot protect against sexually transmitted infections. (ii) Hardware platform and operating system requirements. (iii) Instructions identifying and explaining how to use the software application, including required user inputs and how to interpret the application outputs. (iv) A summary of the clinical validation study and results, including effectiveness of the application as a stand-alone contraceptive and how this effectiveness compares to other forms of legally marketed contraceptives.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue. August 24, 2023 Natural Cycles Nordic AB % Shelia Hemeon-Heyer, MS, JD, FRAPS President Heyer Regulatory Solutions, LLC 125 Cherry Lane Amherst, MA 01002 Re: K231274 Trade/Device Name: Natural Cycles Regulation Number: 21 CFR§ 884.5370 Regulation Name: Software Application for Contraception Regulatory Class: II Product Code: PYT Dated: July 27, 2023 Received: July 27, 2023 Dear Shelia Hemeon-Heyer: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. {1}------------------------------------------------ You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatoryinformation/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, # Michael T. Bailey -S For Monica D. Garcia, Ph. D. Assistant Director DHT3B: Division of Reproductive, Gynecology and Urology Devices OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ ## Indications for Use 510(k) Number (if known) K231274 Device Name Natural Cycles Indications for Use (Describe) Natural Cycles is a stand-alone software application, intended for women 18 years and older, to monitor their fertility. Natural Cycles can be used for preventing a pregnancy (contraception) or planning a pregnancy (conception). Prescription Use (Part 21 CFR 801 Subpart D) X Over-The-Counter Use (21 CFR 801 Subpart C) ### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. ### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ # 510(k) Summary K231274 - Natural Cycles - A. Submitter: Natural Cycles Nordic AB St Eriksqatan 63B 112 34 Stockholm, Sweden Contact: Raoul Scherwitzl, PhD, CEO and Co-Founder Phone: +46 707174866 E-mail: Raoul.scherwitzl@naturalcycles.com - B. Correspondent: Hever Requlatory Solutions LLC 125 Cherry Lane Amherst, MA 01002 Contact: Sheila Hemeon-Heyer E-mail: Sheila@heyer-requlatory.com - C. Date Prepared: August 23, 2023 #### Device Name and Classification Information: D. | Trade Name: | Natural Cycles | |--------------------|-------------------------------------------------------------------------| | Common Name: | Software application for contraception | | Regulation Name: | Software application for contraception | | Regulation Number: | 21 CFR 884.5370 | | Regulatory Class: | II | | Product Code: | PYT (Device, fertility diagnostic, contraceptive, software application) | #### E. Predicate Device(s): K202897 Natural Cycles The predicate device has not been subject to a design-related recall. #### F. Device Description: Natural Cycles is an over-the-counter web and mobile-based standalone software application that monitors a woman's menstrual cycle using information entered by the user and informs the user about her past, current, and future fertility status. The following information is used by the Natural Cycles software: - daily body temperature measurements - information about the user's menstruation cycle (i.e., start date, number of days) - optional ovulation or pregnancy test results ● A proprietary algorithm evaluates the data and returns the user's fertility status. Natural Cycles is available in three modes: Contraception (NC° Birth Control), Conception (NC° Plan Pregnancy), and Pregnancy (NCº Follow Pregnancy). For NCº Birth Control mode, the {4}------------------------------------------------ device provides predictions of "not fertile," shown as green days, and "use protection," shown as red days, that allow the user to determine the days on which her risk of conception is highest, and then make choices about either abstaining from sex or using a barrier method of contraception to prevent pregnancy. In addition to measuring daily basal body temperature with an oral thermometer with two decimal points, the predicate submission cleared the Oura Ring for automatic temperature input to the Natural Cycles algorithm, and the current submission expands to allow the device to utilize automatic temperature inputs from the Apple Watch. Natural Cycles can be used by women 18 years and older. Women who have been on hormonal birth control within 60 days prior to using Natural Cycles have a higher risk of becoming pregnant when compared to women who have not been on hormonal birth control within the 12 months prior to using the device may not be right for women who have a medical condition where pregnancy would be associated with a significant risk to the mother or the fetus. #### G. Indications for Use Statement Natural Cycles is a stand-alone software application, intended for women 18 years and older, to monitor their fertility. Natural Cycles can be used for preventing a pregnancy (contraception) or planning a pregnancy (conception). #### Comparison with Predicate Device H. A detailed comparison of the intended use and technological features of the subject and predicate device are described in the table below: | Parameter | Subject Device<br>Natural Cycles<br>K231274 | Predicate Device<br>Natural Cycles<br>K202897 | Comparison | |----------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------| | Indications for<br>Use Statement | Natural Cycles is a stand-<br>alone software application,<br>intended for women 18 years<br>and older, to monitor their<br>fertility.<br>Natural Cycles can be used<br>for preventing a pregnancy<br>(contraception) or planning a<br>pregnancy (conception). | Natural Cycles is a stand-<br>alone software application,<br>intended for women 18 years<br>and older, to monitor their<br>fertility.<br>Natural Cycles can be used<br>for preventing a pregnancy<br>(contraception) or planning<br>a pregnancy (conception). | Identical | | Use<br>Environment | App is downloaded to<br>user's smartphone and<br>used in the home<br>environment | App is downloaded to<br>user's smartphone and<br>used in the home<br>environment | Identical | | Parameter | Subject Device<br>Natural Cycles<br>K231274 | Predicate Device<br>Natural Cycles<br>K202897 | Comparison | | Input<br>Information | • Manual input of two-<br>decimal daily basal body<br>temperature (BBT)<br>measurements, or<br>automatic input from the<br>Oura ring or Apple Watch.<br>• Manual input of<br>information about the<br>user's menstruation<br>cycle, i.e., start date,<br>number of days.<br>• Optional manual input of<br>ovulation or pregnancy<br>test results. | • Manual input of two-<br>decimal daily basal<br>body temperature<br>(BBT) measurements<br>or automatic input from<br>the Oura ring.<br>• Manual input of<br>information about the<br>user's menstruation<br>cycle, i.e., start date,<br>number of days.<br>• Optional manual input of<br>ovulation or pregnancy<br>test results. | Expanded<br>option for a<br>new source of<br>daily basal<br>body<br>temperature<br>information. | | Output<br>Information | • NC° Birth Control: For<br>each day, whether the<br>woman is fertile (red) or<br>non- fertile (green), with<br>descriptive texts.<br>• NC° Plan Pregnancy:<br>Fertility status results are<br>displayed as a scale for<br>fertile days, and green for<br>non-fertile days, together<br>with description texts.<br>• NC° Follow<br>Pregnancy: Provides<br>educational<br>information about the<br>progress of the<br>pregnancy.<br>• All uses in Prevent<br>mode or Plan mode<br>receive the ovulation<br>date for the month and<br>daily statement of<br>fertility status. Historic<br>data is available for all | • Prevent mode: For each<br>day, whether the woman<br>is fertile (red) or non-<br>fertile (green), with<br>descriptive texts.<br>• Plan mode: Fertility<br>status results are<br>displayed as a scale for<br>fertile days, and green<br>for non-fertile days,<br>together with<br>description texts.<br>• Pregnancy mode:<br>Provides educational<br>information about the<br>progress of the<br>pregnancy.<br>• All uses in Prevent<br>mode or Plan mode<br>receive the ovulation<br>date for the month and<br>daily statement of<br>fertility status. Historic<br>data is available for all | Identical; titles<br>for the modes<br>have been<br>changed | {5}------------------------------------------------ The subject and predicate device have identical indications for use statements and have the same intended use - to predict fertile and non-fertile days for use in providing patient-specific recommendations related to contraception. The subject and predicate device have different technological characteristics (i.e., temperature data input sources). This difference does not raise different questions of safety and effectiveness as compared to the predicate device. {6}------------------------------------------------ #### l. Summary of Non-Clinical Performance Testing - . Software documentation provided in accordance with the 2005 FDA guidance document Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices to support device software with a major level of concern. - . Cybersecurity information provided in accordance with the 2014 FDA guidance document Content of Premarket Submissions for Management of Cybersecurity in Medical Devices. - . Human factors testing in DEN170052 was used to support that device users could safely use the device. #### Summary of Clinical Performance Testing J. The purpose of this 510(k) is to add the option for the user to obtain the daily basal body temperature information used by the Natural Cycles fertility algorithm from the Apple Watch. There have been no changes to how the Natural Cycles algorithm determines the daily fertility status. A clinical study was conducted to compare the daily temperatures generated by a wearable temperature monitor, the Apple Watch, to a two decimal place oral thermometer for use by the Natural Cycles algorithm. The study included data of de-identified information from 272 users after all applied criteria were met. A subset of the data, 104 users, was assessed for the ovulation key performance indicator (KPI), which required a positive result from a luteinizing hormone (LH) test. The mean age of the study participants was 30.5 ± 5.1 years, and the mean menstrual cycle length for complete cycles during the study was 28.8 ± 5.5 days. The majority (58%) of the women were located in the European Union and the United Kingdom, with the remainder located in the US (42%). The women wore the Apple Watch nightly, with the user's temperature information being transferred to the subject device each morning and recorded their temperature each morning as taken using a two-decimal place oral thermometer. Data were collected over a total of 1918 menstrual cycles, 505 of which were considered complete cycles and met the applied criteria, per the study definition. The participants were also asked to record the results of luteinizing hormone (LH) tests each cycle. At least one positive LH test was submitted in 153 of the complete study cycles. The results of the clinical study demonstrated that when temperature was inputted from either the Apple Watch or the two-decimal place oral thermometer, the Natural Cycles algorithm was able to identify that ovulation had occurred, which was confirmed in the study by comparison to the 153 positive LH test results. Compared to the two-decimal place oral thermometer, the Natural Cycles algorithm provided 0.6 fewer green days (not fertile) in the luteal phase of the menstrual cycle when the input temperature was from the Apple Watch without increasing the risk of unintended pregnancy. This is in contrast with the predicate device, which provided an additional 1.6 green days when the input temperature was from the Oura Ring; however, this difference does not impact the safety and effectiveness of Natural Cycles for its intended use. {7}------------------------------------------------ #### K. Conclusion A comparison of intended use and technological characteristics combined with performance data demonstrates that Natural Cycles is as safe and effective as the predicate device and supports a determination of substantial equivalence.
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...