The ClearTip is intended for Ultrasonically Guided Fine Needle Aspiration, (FNA) of submucosal and extraluminal lesions of the Tracheobronchial Tree and Gastrointestinal Tract, (e.g., lymph nodes, abnormal tissue in the mediastinum).
Device Story
ClearTip is a manually operated endoscopic biopsy instrument; used to obtain tissue specimens from submucosal and extraluminal lesions in the gastrointestinal tract and tracheobronchial tree. Device consists of handle, insertion part, syringe, stopcock, and connector. Operated by physician via endoscope working channel; needle is advanced to target lesion under ultrasound guidance; aspiration is performed manually using syringe. Output is a tissue sample for diagnostic analysis. Single-patient use; sterile; disposable. Benefits include minimally invasive tissue acquisition for pathological evaluation.
Clinical Evidence
No clinical data. Substantial equivalence supported by non-clinical bench testing including sterilization validation, shelf-life, appearance, dimensions, operability, elasticity, bending strength, pull-out, tensile force, and biocompatibility testing.
Technological Characteristics
Manual biopsy needle; Nitinol needle and stylet; PEEK and PTFE sheath. Compatible with Olympus and Pentax endoscopes. Minimum accessory channel diameter 2.0 mm. Ethylene oxide (EO) sterilization. Single-use. No software or electronic components.
Indications for Use
Indicated for patients requiring ultrasonically guided fine needle aspiration of submucosal and extraluminal lesions within the tracheobronchial tree and gastrointestinal tract, including lymph nodes and mediastinal tissue.
Regulatory Classification
Identification
A gastroenterology-urology biopsy instrument is a device used to remove, by cutting or aspiration, a specimen of tissue for microscopic examination. This generic type of device includes the biopsy punch, gastrointestinal mechanical biopsy instrument, suction biopsy instrument, gastro-urology biopsy needle and needle set, and nonelectric biopsy forceps. This section does not apply to biopsy instruments that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.
Predicate Devices
SonoTip Pro and Pro Flex EBUS-TBNA Needle System (K133763)
{1}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health and Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
June 30, 2023
Finemedix Co., Ltd. % Priscilla Chung Regulatory Affairs Consultant LK Consulting Group USA. Inc. 18881 Von Karman Ave STE 160 Irvine, CA 92612
K231267 Re:
Trade/Device Name: ClearTip Regulation Number: 21 CFR 876.1075 Regulation Name: Gastroenterology-urology biopsy instrument Regulatory Class: Class II Product Code: FCG Dated: April 28, 2023 Received: May 2, 2023
Dear Priscilla Chung:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any
{2}------------------------------------------------
Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safetyreporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medicaldevice-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advicecomprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# Shanil P. Haugen -S
Shanil P. Haugen, Ph.D. Assistant Director DHT3A: Division of Renal, Gastrointestinal, Obesity and Transplant Devices OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
## Indications for Use
510(k) Number (if known) K231267
Device Name ClearTip
Indications for Use (Describe)
The Clear Tip is intended for Ultrasonically Guided Fine Needle Aspiration, (FNA) of submucosal and extraluminal lesions of the Tracheobronchial Tree and Gastrointestinal Tract, (e.g., lymph nodes, abnormal tissue in the mediastinum).
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
K231267 Page 1 of 4
# 510(K) SUMMARY
(K231267)
June 8, 2023
## 1. Submitted by:
FINEMEDIX Co., Ltd. 140-9, Yuram-ro, Dong-gu, Daegu, 41059 Republic of Korea Tel : 82-53-741-8388, Fax : 82-53-741-8168
#### 2. US Agent/ Official Correspondent:
Priscilla Chung
LK Consulting Group USA, Inc. 18881 Von Karman Ave STE 160 Irvine, CA 92612 Fax: 714.409.3357 Tel: 714.202.5789 Email: juhee.c@LKconsultingGroup.com
#### 3. Device Name:
| - Trade Name | : ClearTip |
|-----------------------|----------------------------|
| - Classification | : Class II |
| - Classification Name | : Biopsy Needle |
| - Product Code | : FCG |
| - Regulation Number | : 876.1075 |
| - Review Panel | : Gastroenterology/Urology |
#### 4. Predicate Device:
- Primary Predicate Device: SonoTip Pro and Pro Flex EBUS-TBNA Needle System (K133763) by Medi-Globe Corporation
- Reference Predicate Device: Clear-Tip EUS-FNA (K180363) by FINEMEDIX Co., Ltd.
{5}------------------------------------------------
#### 5 Device Description:
The ClearTip is a manually operated endoscopic instrument intended to obtain tissue specimens of gastrointestinal tract (=digestive tract) and tracheobronchial tree.
The subject device mainly consists of a handle unit with an insertion part, a syringe, a stopcock, and a connector. This device passes through a working channel of an endoscope, and the average contact time with the submucosal or external lesions of a gastrointestinal tract or a tracheobronchial tree is less than 1 hour. This device is supplied sterile for single-patient use and shall be not reused or resterilized.
## 6 Indications for Use Statement
The ClearTip is intended for Ultrasonically Guided Fine Needle Aspiration, (FNA) of submucosal and extraluminal lesions of the Tracheobronchial Tree and Gastrointestinal Tract, (e.g., lymph nodes, abnormal tissue in the mediastinum).
#### Substantial Equivalence Discussion: 7
#### 7.1.Comparison Chart
| | Subject Device | Primary Predicate | Reference Device |
|-----------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Manufacturer | FINEMEDIX Co., Ltd. | Medi-Globe Corporation | FINEMEDIX Co., Ltd. |
| Device Name | ClearTip | SonoTip Pro and Pro Flex<br>EBUS-TBNA Needle<br>System | Clear-Tip EUS-FNA |
| 510(k) Number | K231267 | K133763 | K180363 |
| Device<br>Classification Name | Gastroenterology-urology<br>Biopsy instrument | Gastroenterology-<br>Urology Biopsy<br>Instrument | Gastroenterology-urology<br>Biopsy instrument |
| Product Code | FCG | FCG | FCG |
| Regulation Number | 21 CFR 876.1075 | 21 CFR 876.1075 | 21 CFR 876.01075 |
| Indications for Use | The ClearTip is intended<br>for Ultrasonically Guided<br>Fine Needle Aspiration,<br>(FNA) of submucosal and<br>extraluminal lesions of<br>the Tracheobronchial<br>Tree and Gastrointestinal<br>Tract, (e.g., lymph nodes,<br>abnormal tissue in the<br>mediastinum). | The SonoTip Pro and Pro<br>Flex - TBNA Needle<br>System is intended for<br>Ultrasonically Guided<br>Fine Needle Aspiration,<br>(FNA) of submucosal<br>and extraluminal lesions<br>of the Tracheobronchial<br>Tree and Gastrointestinal<br>Tract, (e.g., lymph nodes,<br>abnormal tissue in the<br>mediastinum). | The Clear-Tip EUS-FNA<br>is used with an ultrasound<br>endoscope for fine needle<br>biopsy of submucosal<br>lesions, mediastinal<br>masses, lymph nodes and<br>intraperitoneal masses<br>within or adjacent to the<br>gastrointestinal tract. |
| Needle Gauge | 22 Gauge | 22 Gauge | 19,22,25 Gauge |
| Needle Material | Nitinol | Stainless Steel / Nitinol | Stainless Steel |
| Needle length<br>Adjustment range | 0~5 cm | 0~4 cm | 0~8cm |
| Stylet Material | Nitinol | Nitinol | Nitinol |
{6}------------------------------------------------
| Sheath length | 738.5 mm | 730 mm | 1,507 mm |
|----------------------------------------------------|---------------------------------------------------|---------------------------------------------------|---------------------------------------------------|
| Sheath Material | PEEK, PTFE | PVDF | PEEK |
| Sheath length<br>Adjustment range | 0~3cm | 0~4cm | 0~5cm |
| Accessory Channel<br>Diameter | Minimum Accessory<br>Channel Diameter<br>(2.0 mm) | Minimum Accessory<br>Channel Diameter<br>(2.0 mm) | Minimum Accessory<br>Channel Diameter<br>(2.0 mm) |
| Endoscope<br>Compatibility<br>(Use with connector) | 1) Olympus, 2) Pentax | 1) Olympus, 2) Pentax | 1) Olympus |
| Use with a Syringe<br>and stopcock | Yes | Yes | Yes |
| Principle of<br>Operation | Manual (sampling using<br>aspiration) | Manual (sampling using<br>aspiration) | Manual (sampling using<br>aspiration) |
| Shelf life | 3 Years | 5 Years | 3 Years |
| Sterility | Ethylene oxide(EO) | Ethylene oxide(EO) | Ethylene oxide(EO) |
| Single use | Yes | Yes | Yes |
#### 7.2. Substantial Equivalence Discussion
The subject device is substantially equivalent to the primary predicate devices in the indications for use, needle gauge, needle material, accessory channel diameter, endoscope compatibility, use with a syringe and stopcock, principle of operation, sterility, and single use.
The differences between the subject device and the primary predicate device are as below:
-Needle Length Adjustment Range: The maximum needle length of the subject device is 5 cm, and the predicate device is 4 cm. The difference of 1 cm between the two devices does not affect safety and effectiveness during biopsy. It is just an option for the operator to adjust the length of the needle in consideration of the surgical area (position and depth of the tissue to be biopsied).
-Sheath Length: The sheath length of the subject device is 738.5 mm, and the predicate device is 730 mm. The difference of 8.5 mm does not affect safety and effectiveness during biopsy since the physician can control the length as needed.
-Sheath Material: The raw material of the sheath of the subject device is PEEK & PTFE, and the predicate device is PVDF. We performed biocompatibility testing on the subject device, and based on the test result, we conclude that the subject device is biocompatible and substantially equivalent to the predicate devices.
-Shelf Life: The shelf life of the subject device is 3 years, and the shelf life of the predicate device is 5 years. We performed shelf-life testing, and based on the test result, we conclude that the subject device has 3-year shelf life.
{7}------------------------------------------------
We are presenting Clear-Tip EUS-FNA (K180363) made by our company as a reference device. It shares similarities with the subject device in terms of indications for use statement, needle gauge, stylet material, accessory channel diameter, endoscope compatibility, use with a syringe and stopcock, principle of operation, sterility, and single use.
The differences between the subject device and the reference predicate device are as follows:
- Claim: The reference device does not include a claim on the Tracheobronchial Tree.
- Sheath Length: The sheath of the subject device is shorter than that of the reference device. The shortened length of the subject device is for user convenience, especially when using the device on the Tracheobronchial Tree.
- This reference device is presented to leverage some of the test items, such as EO sterilization validation, Real-Time Shelf-Life, and Shipping Validation, for the subject device. We have also referenced the test parameters, test method, and P/F criteria from this 510k.
#### 8. Non-clinical Tests
The following tests were performed on the subject device and the test results support that the subject device is substantially equivalent to the predicate devices.
- Sterilization Validation Test ●
- Shelf-Life Test ●
- Appearance
- Dimensions ●
- Operability ●
- Elasticity ●
- Bending Strength ●
- Pull-out ●
- Tensile Force ●
- Biocompatibility Tests ●
#### 9. Conclusion:
Based on the information provided herein and the test results, we conclude that the subject device is substantially equivalent to its predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.