K231254 · Actuated Medical, Inc. · FDF · Nov 3, 2023 · Gastroenterology, Urology
Device Facts
Record ID
K231254
Device Name
GripTract-GI Endoscopic Tissue Manipulator
Applicant
Actuated Medical, Inc.
Product Code
FDF · Gastroenterology, Urology
Decision Date
Nov 3, 2023
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 876.1500
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The GripTract-Gl Endoscopic Tissue Manipulator (GripTract) is an accessory intended to assist in positioning the distal end of an endoscope from the mucosal surface and assist with optical visualization, diagnosis, and endoscopic treatment. GripTract is indicated for use in the large intestine with any standard endoscope that has a distal tip outer diameter of 11.5 - 12.0 mm and working length of 168 - 170 cm.
Device Story
GripTract-GI is a single-use, non-sterile endoscopic accessory for large intestine procedures. It consists of a handpiece attaching to the endoscope control body and a distal end cap with two integrated fingers. The physician operates two control knobs on the handpiece to rotate, extend, or retract the fingers independently of the endoscope. This allows for direct tissue lifting, retraction, and guidance of working tools to improve visualization, diagnosis, and treatment. The device is used in clinical surgical settings by trained medical specialists. It provides mechanical tissue manipulation to facilitate endoscopic tasks. Benefits include improved endoscope positioning and enhanced visualization compared to endoscope-only procedures.
Clinical Evidence
No human clinical data. Evidence includes bench performance testing (verification of specifications, distribution per ASTM D4169-16, accelerated shelf-life, reliability, and bench safety), human factors/usability testing (n=15 users for setup and use), biocompatibility testing (per ISO 10993 series), and in vivo porcine testing. Porcine study compared GripTract to endoscope-only procedures, showing equivalent visual and histological mucosal damage with no device-attributable adverse events at acute or chronic time points.
Technological Characteristics
Non-sterile, single-use accessory. Materials: polyurethane, polycarbonate (End Cap); PTFE (Sheath); ETFE-coated stainless steel (Fingers); stainless steel (Torque Tubes). Mechanical actuation via manual control knobs and torque tubes. Compatible with 11.5-12.0 mm OD endoscopes. Biocompatibility per ISO 10993-1. Distribution testing per ASTM D4169-16.
Indications for Use
Indicated for use in the large intestine to assist in positioning the distal end of an endoscope from the mucosal surface and to assist with optical visualization, diagnosis, and endoscopic treatment. Compatible with standard endoscopes having a distal tip outer diameter of 11.5-12.0 mm and a working length of 168-170 cm.
Regulatory Classification
Identification
An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.
Special Controls
*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.
In combination with the general controls of the FD&C Act, the integrated operating table-electromechanical surgical system is subject to the following special controls:
1. (1) Premarket clinical performance testing, or a combination of premarket clinical performance testing and postmarket surveillance (in accordance with special control (2)), must include the following:
1. (i) Objective performance measures (e.g., rate and number of conversions to other surgical modalities, rate of device related adverse events (including tissue injury, hematoma, and increased blood loss), and their severity, cause, and outcomes) must be reported with relevant descriptive comparator performance measures.
2. (ii) The data must demonstrate the performance of the device for providing accurate and precise control of attached surgical instruments in range of clinical conditions relevant to the device's intended use.
3. (iii) The test dataset must include data collected from a patient population representative of the intended patient population under anticipated conditions of use.
2. (2) Data obtained from postmarket surveillance must demonstrate, in consideration of the premarket data obtained in accordance with special control (1), that the device performs in accordance with special control (1), unless FDA determines, based on the totality of the premarket data, that data from postmarket surveillance is not required to demonstrate that the device performs as intended. Such postmarket surveillance must be conducted per a protocol determined appropriate by FDA to demonstrate that the device performs as intended (in consideration of the premarket data obtained in accordance with special control (1)), and must include initiation, enrollment, and reporting requirements to ensure timely periodic updates to FDA on post-market surveillance progress and outcomes.
3. (3) Animal performance testing must evaluate the extent of port site trauma due to repositioning of table during surgical procedures when utilizing robotic minimally invasive and laparoscopic approaches
4. (4) The device manufacturer must develop, and update as necessary, a device-specific use training program that ensures proper device setup/use/shutdown, accurate control of instruments to perform the intended surgical procedures, troubleshooting and handling during unexpected events or emergencies, and safe practices to mitigate use error.
5. (5) The device manufacturer may only distribute the device to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program.
(6) Human factors assessment must demonstrate that the user can correctly use the device system across all intended use environments with the provided instructions and training materials, including patient access during normal operating conditions and emergency situations, and effects arising from the integrated nature of the operating table and robotic surgical arms.
(7) Labeling must include:
(i) A detailed summary of clinical performance testing conducted with the device, including study population, results, adverse events, and comparisons to any comparator groups identified;
(ii) A statement in the labeling that the safety and effectiveness for the representative specific procedures was based on evaluation of the device as a surgical tool and did not include evaluation of outcomes related to the treatment of the patient's underlying disease or condition, unless FDA determines that it can be removed or modified based on clinical performance data submitted to FDA;
(iii) Identification of compatible devices;
(iv) The list of surgical procedures for which the device has been determined to be safe with clinical justification;
(v) Reprocessing instructions for reusable components;
(vi) A shelf life for any sterile components;
(vii) A description of the device-specific use training program;
(viii) A statement that the device is only for distribution to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program; and
(ix) A summary of any completed postmarket surveillance data collected as required by special control (2), including updated labeling to accurately reflect outcomes observed in postmarket surveillance.
(8) Non-clinical performance testing must demonstrate that the device performs as intended under anticipated conditions of use and must include:
(i) Device motion accuracy and repeatability;
(ii) System testing;
(iii) Instrument reliability;
(iv) Crosstalk;
(v) Table motion control;
(vi) Thermal effects on tissue;
(vii) User-device interface performance;
(viii) Workspace access testing; and
(ix) Performance testing with compatible devices.
(9) Software verification, validation, and hazard analysis must be performed.
(10) Electromagnetic compatibility and electrical, thermal, and mechanical safety testing must be performed.
(11) Performance data must demonstrate the sterility of all patient-contacting device components.
(12) Performance data must support the shelf life of the device components provided sterile by demonstrating continued sterility and package integrity over the labeled shelf life.
(13) Performance data must validate the reprocessing instructions for the reusable components of the device.
(14) Performance data must demonstrate that all patient-contacting components of the device are biocompatible.
(15) Performance data must demonstrate that all patient-contacting components of the device are non-pyrogenic.
(16) The device manufacturer must submit a report to the FDA annually on the anniversary of initial marketing authorization for the device, until such time as FDA may terminate such reporting, which comprises the following information:
(i) Cumulative summary, by year, of complaints and adverse events since date of initial marketing authorization; and
(ii) Identification and rationale for changes made to the device, labeling, or device specific use training program, which did not require submission of a premarket notification during the reporting period.
{1}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA acronym with the full name of the agency on the right. The FDA part of the logo is in blue, with the acronym in a square and the full name written out to the right of the square.
November 3, 2023
Actuated Medical, Inc. Douglas Dillon Director, Ouality Assurance & Regulatory Affairs 320 Rolling Ridge Drive Bellefonte, Pennsylvania 16823
Re: K231254
Trade/Device Name: GripTract-GI Endoscopic Tissue Manipulator Regulation Number: 21 CFR 876.1500 Regulation Name: Endoscope And Accessories Regulatory Class: Class II Product Code: FDF Dated: October 6, 2023 Received: October 6, 2023
Dear Douglas Dillon:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"
{2}------------------------------------------------
(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the OS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
{3}------------------------------------------------
Sincerely,
Shanil P. Haugen -S
Shanil P. Haugen, Ph.D. Assistant Director DHT3A: Division of Renal, Gastrointestinal, Obesity and Transplant Devices OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
# Indications for Use
510(k) Number (if known) K231254
Device Name
Griptract-GI Endoscopic Tissue Manipulator
#### Indications for Use (Describe)
The GripTract-Gl Endoscopic Tissue Manipulator (GripTract) is an accessory intended to assist in positioning the distal end of an endoscope from the mucosal surface and assist with optical visualization, diagnosis, and endoscopic treatment.
GripTract is indicated for use in the large intestine with any standard endoscope that has a distal tip outer diameter of 11.5 - 12.0 mm and working length of 168 - 170 cm.
| Type of Use (Select one or both, as applicable) | |
|--------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------|
| <div> <span> <span style="font-size:16px;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) </span> </div> | <div> <span> □ Over-The-Counter Use (21 CFR 801 Subpart C) </span> </div> |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{5}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for Actuated Medical. The logo consists of a brown plus sign with a curved line through it, followed by the word "ACTUATED" in gray and "MEDICAL" in orange. Below the logo is the tagline "Innovative motion + Positive outcomes" in gray.
310 Rolling Ridge Drive Bellefonte, PA 16823 + p (814) 355-0003
+ f (814) 355-1532
ActuatedMedical.com
#### Certified ...
- + ISO 13485:2016
- + Medical Device Single Audit Program (MDSAP)
- + Women's Business Enterprise (WBE)
- + Women-Owned Small Business (WOSB)
## K231254 510(k) SUMMARY
## Applicant Information
Date Prepared: October 6, 2023
Name and Address: Actuated Medical, Inc. 320 Rolling Ridge Drive Bellefonte, PA 16823 Ph: (814) 355-0003 Fx: (814) 355-1523
Contact Person: Douglas R. Dillon Director, Quality Assurance and Regulatory Affairs Ph: (814) 355-0003 x107 Fx: (814) 355-1523 Email: Douglas.Dillon@actuatedmedical.com
## Device Information
| Trade Name: | GripTract-GI™ Endoscopic Tissue Manipulator |
|----------------------|---------------------------------------------|
| Common Name: | GripTract-GI™ Endoscopic Tissue Manipulator |
| Classification: | 21 CFR §876.1500 |
| Classification Name: | Endoscope and accessories |
| Product Code: | FDF |
Image /page/4/Picture/16 description: The image is a close-up of an orange background with a pattern of lighter orange squares and wavy lines. The squares are arranged in a staggered pattern, creating a sense of depth. The wavy lines add a dynamic element to the composition. The overall effect is a textured and visually interesting design.
K231254 Page 1 of 11
{6}------------------------------------------------
Image /page/5/Picture/1 description: The image is a close-up of an orange background with a pattern of squares and wavy lines. The squares are arranged in a checkerboard pattern, with alternating light and dark orange squares. The wavy lines are also orange and run diagonally across the image. The overall effect is a bright and cheerful design.
| Type | 510(k)<br>Number | Trade Name | Submitter |
|---------------------|------------------|----------------------------------------------------------------|---------------------------|
| Predicate<br>Device | K210851 | DiLumen<br>Endolumenal<br>Interventional<br>Platform (DiLumen) | Lumendi Ltd. |
| Reference<br>Device | K984358 | D-201-12704<br>Disposable Distal<br>Attachment | Olympus Americas,<br>Inc. |
### Predicate & Reference Devices
# Device Description
GripTract is a single-use, non-sterile, endoscope accessory intended to ensure complete positioning of an endoscope and assist with optical visualization, diagnosis, and endoscopic treatment. It is indicated for use in the large intestine with any standard endoscope that has a distal tip outer diameter of 11.5 - 12.0 mm and a working length of 168 - 170 cm.
The GripTract Handpiece attaches to the endoscope control body just below the working channel. The soft End Cap with two integrated Fingers for tissue manipulation is placed on the distal end of the endoscope. Two Control Knobs in the Handpiece each operate a corresponding Finger, permitting the User to rotate and extend/retract the Fingers for tissue manipulation and visualization independent of the endoscope's movement or the presence of other tools in the endoscope's working channel.
## Indications for Use
The GripTract-Gl Endoscopic Tissue Manipulator (GripTract) is an accessory intended to assist in positioning the distal end of an endoscope from the mucosal surface and assist with optical visualization, diagnosis, and endoscopic treatment.
GripTract is indicated for use in the large intestine with any standard endoscope that has a distal tip outer diameter of 11.5 - 12.0 mm and working length of 168 - 170 cm.
{7}------------------------------------------------
Image /page/6/Picture/1 description: The image is an abstract design featuring a combination of geometric and organic elements. The background is a gradient of orange hues, transitioning from a solid, darker shade on the left to a lighter, patterned area on the right. The pattern consists of alternating squares in two different shades of orange, creating a checkerboard effect. A wavy line in a similar orange shade meanders across the patterned area, adding a fluid, organic contrast to the geometric shapes.
K231254 Page 3 of 11
## Technological Characteristics
The table below provides a comparison of the technological characteristics of the two devices. Many items, including intended use, user population, and anatomical area of use are identical. Key differences include method of manipulating tissue, method of attaching controls to endoscope, and materials of construction; none of which raise different questions of safety and efficacy as compared to the Predicate Device.
| Category | Predicate<br>DiLumen | Proposed<br>GripTract | Comparison |
|---------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Intended / Indications<br>for use | The Lumendi<br>DiLumen is an<br>accessory to an<br>endoscope. The<br>DiLumen dual<br>balloon accessory<br>is intended for<br>use with any<br>standard<br>endoscope that<br>has a distal tip<br>outer diameter of<br>12.5 - 14.3 mm.<br>The device is<br>indicated to<br>ensure complete<br>positioning of an<br>endoscope during<br>navigation in the<br>large intestine,<br>while assisting<br>with optical<br>visualization,<br>diagnosis, tissue<br>manipulation, and<br>endoscopic<br>treatment. | GripTract is an<br>accessory<br>intended to assist<br>in positioning the<br>distal end of an<br>endoscope from<br>the mucosal<br>surface and assist<br>with optical<br>visualization,<br>diagnosis, and<br>endoscopic<br>treatment. The<br>device is<br>indicated for use<br>in the large<br>intestine with any<br>standard<br>endoscope that<br>has a distal tip<br>outer diameter of<br>11.5 - 12.0 mm<br>and a working<br>length of 168-170<br>cm. | Same - intended<br>use is the same.<br>Indications differ<br>only by the<br>specific<br>endoscope<br>dimensions.<br>GripTract is<br>designed for<br>smaller diameter<br>endoscopes with<br>the same length<br>as one specific<br>model of the<br>Predicate. |
| Category | Predicate<br>DiLumen | Proposed<br>GripTract | Comparison |
| Compatible<br>Endoscope OD | Standard<br>endoscope with<br>diameter 12.5-<br>14.3 mm | Standard<br>endoscope with<br>diameter 11.5-<br>12.0 mm | Different -<br>GripTract<br>designed for<br>endoscopes with<br>smaller OD. |
| Working Length | 168 cm (different<br>models for<br>different length<br>endoscopes) | 168-170 cm | Same |
| Sterility | DiLumen is a<br>non-sterile device<br>that does not<br>require cleaning<br>or sterilization<br>prior to use. | GripTract is a<br>non-sterile device<br>that does not<br>require cleaning<br>or sterilization<br>prior to use. | Same |
| Single Use/reusable | Single Use | Single Use | Same |
| Shelf Life | 24 months | 24 months | Same |
| Distal End Attachment<br>Method | Close-fitting<br>sleeve fits<br>securely over the<br>distal end of the<br>endoscope. | Low durometer<br>End Cap fits<br>securely onto the<br>distal end of the<br>endoscope. | Same |
| Proximal Control<br>Attachment Method | Endoscope is<br>placed through<br>the Device Lock<br>then a radial knob<br>is twisted to<br>tighten the Device<br>Lock securing the<br>DiLumen controls<br>in the same plane<br>as the endoscope<br>controls. | O-Rings wrap<br>around<br>endoscope<br>control body<br>securing the<br>GripTract controls<br>in the same plane<br>as the endoscope<br>controls. | Different - both<br>attachment<br>methods securely<br>hold the device<br>onto the<br>endoscope control<br>body and place<br>controls in the<br>same location. |
| Category | Predicate<br>DiLumen | Proposed<br>GripTract | Comparison |
| Operating Principle | DiLumen is<br>secured to the<br>endoscope at<br>both the distal<br>and proximal<br>ends. The User<br>manipulates the<br>DiLumen controls<br>with the same<br>hand as the<br>endoscope<br>controls. The<br>Fore Balloon can<br>be inflated,<br>extended, and<br>retracted to<br>improve<br>endoscopic<br>visualization and<br>treatment. | GripTract is<br>secured to the<br>endoscope at<br>both the distal<br>and proximal end.<br>The User<br>manipulates the<br>GripTract controls<br>with the same<br>hand as the<br>endoscope<br>controls. The<br>Fingers can be<br>rotated, extended,<br>and retracted to<br>improve<br>endoscopic<br>visualization and<br>treatment. | Same |
{8}------------------------------------------------
Image /page/7/Picture/1 description: The image is an abstract design featuring shades of orange. The left side of the image is a solid, lighter shade of orange, while the right side transitions into a pattern of alternating darker and lighter orange squares, creating a checkerboard effect. Overlaying the checkerboard pattern are wavy lines in a lighter shade of orange, adding a dynamic element to the design.
{9}------------------------------------------------
Image /page/8/Picture/1 description: The image is an abstract design featuring a combination of geometric shapes and wavy lines. The color palette is dominated by shades of orange, with a lighter orange used for the background and a darker orange for the shapes and lines. The geometric shapes appear to be squares or rectangles arranged in a grid-like pattern, while the wavy lines meander across the image, adding a sense of movement and fluidity to the design.
# K231254
Page 5 of 11
{10}------------------------------------------------
Image /page/9/Picture/1 description: The image is an abstract design featuring shades of orange. The background consists of a pattern of squares in alternating light and dark orange hues, creating a checkerboard effect. Overlaying this pattern are wavy lines in a lighter shade of orange, adding a dynamic element to the composition.
# K231254
Page 6 of 11
| Category | Predicate<br>DiLumen | Proposed<br>GripTract | Comparison |
|---------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Tissue Manipulation | Tissue is<br>manipulated<br>using the Controls<br>to inflate and<br>deflate the Fore<br>and Aft Balloons,<br>which maintain<br>position of or in<br>tissue. The Fore<br>Balloon with<br>Suture Loops can<br>also be extended<br>and retracted to<br>enhance<br>endoscopic<br>visualization and<br>treatment.<br>Attaching tissue<br>to the Suture<br>Loops allows<br>lifting and<br>retraction of<br>tissue. | Tissue is<br>manipulated<br>using the Controls<br>to extend, retract,<br>and rotate the<br>coated Fingers to<br>enhance<br>endoscopic<br>visualization and<br>treatment. The<br>Fingers can be<br>used to directly lift<br>and retract tissue,<br>and to help guide<br>working tools. | Different - tissue<br>manipulation tools<br>are different, but<br>both physically<br>manipulate/retract<br>tissue as needed<br>in order to<br>facilitate<br>visualization and<br>diagnosis /<br>treatment in the<br>gastrointestinal<br>tract. |
| Accessories | DiLumen can be<br>used with various<br>additional<br>endoscope<br>accessories but is<br>not supplied with<br>any designated<br>accessories. | GripTract can be<br>used with various<br>additional<br>endoscope<br>accessories but is<br>not supplied with<br>any designated<br>accessories. | Same |
{11}------------------------------------------------
Image /page/10/Picture/1 description: The image is an abstract design featuring a combination of geometric and organic elements. The background consists of alternating squares of two shades of orange, creating a checkerboard pattern. Overlaying this pattern are wavy lines in a lighter shade of orange, adding a sense of movement and fluidity to the composition. The overall effect is a visually engaging and modern design.
# K231254
Page 7 of 11
| Category | Predicate<br>DiLumen | Proposed<br>GripTract | Comparison |
|-----------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------|
| Biocompatibility<br>Testing | Biocompatibility<br>tested per ISO<br>10993-1:<br>Cytoxicity,<br>Sensitization,<br>Irritation,<br>Systemic Toxicity,<br>Pyrogenicity, and<br>EO Residuals<br>(Note: Predicate<br>is NOT sterile) | Biocompatibility<br>tested per ISO<br>10993-1:<br>Cytotoxicity,<br>Sensitization,<br>Irritation,<br>Systemic Toxicity,<br>and<br>Pyrogenicity<br>(Note: Proposed<br>is NOT sterile) | Different -<br>additional test<br>performed by<br>Predicate is not<br>relevant to non-<br>sterile devices. |
| Disposal | Dispose of the<br>DiLumen in<br>accordance with<br>accepted medical<br>practice and local,<br>state, and federal<br>regulations for a<br>single use<br>medical device<br>disposal. | Dispose of the<br>GripTract in<br>accordance with<br>accepted medical<br>practice and local,<br>state, and federal<br>regulations for a<br>single use<br>medical device<br>disposal. | Same |
| Use Environment | Clinical surgical<br>settings (e.g.,<br>hospital,<br>outpatient care<br>facility) where<br>endoscopic<br>procedures are<br>performed. | Clinical surgical<br>settings (e.g.,<br>hospital,<br>outpatient care<br>facility) where<br>endoscopic<br>procedures are<br>performed. | Same |
{12}------------------------------------------------
Image /page/11/Picture/1 description: The image is an abstract design featuring shades of orange. The left side of the image is a solid, lighter shade of orange, while the right side transitions into a pattern of alternating darker and lighter orange squares, creating a checkerboard effect. A wavy line in a lighter shade of orange runs across the checkerboard pattern, adding a dynamic element to the design.
### K231254 Page 8 of 11
| Category | Predicate<br>DiLumen | Proposed<br>GripTract | Comparison |
|------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| User Population | Medical<br>specialists who<br>have proper<br>training and<br>competencies in<br>endoscopic<br>procedures and<br>endoscopic<br>equipment. | Medical<br>specialists who<br>have proper<br>training and<br>competencies in<br>endoscopic<br>procedures and<br>endoscopic<br>equipment. | Same |
| Anatomic Area of Use | Natural orifice for<br>access to colon. | Natural orifice for<br>access to colon. | Same |
| Materials of<br>Construction | Sleeve:<br>Pellathane<br>(polyurethane)<br>with 2.5%<br>Glycolube<br>Balloon:<br>polyurethane<br>Inflation<br>Tube/Push<br>Rods: Isoplast<br>(polyurethane),<br>Pebax (polyether<br>block amide),<br>Pellathane<br>(polyurethane) | End Cap:<br>polyurethane,<br>polycarbonate<br>Sheath: PTFE<br>(polytetrafluoroeth<br>ylene)<br>Fingers: ETFE<br>(ethylene<br>tetrafluoroethylen<br>e) coated<br>stainless steel<br>Torque Tubes:<br>stainless steel | Different –<br>GripTract uses<br>different<br>biocompatible<br>polymeric<br>materials and also<br>uses stainless<br>steel to ensure<br>effective force<br>transmission. |
# Non-Clinical Performance Data
The methods and performance data for evaluating the technological differences and questions of safety and effectiveness include bench performance testing, usability testing, biocompatibility testing, and in vivo porcine testing. This testing is summarized in the table below.
Bench Performance Testing: Verification of Product Specification testing confirmed that GripTract met all product specifications and acceptance criteria in ambient, high temperature/humidity, and low temperature environmental conditions. Distribution testing conforming to ASTM D4169-
{13}------------------------------------------------
Image /page/12/Picture/1 description: The image is an abstract design featuring a pattern of orange squares and a wavy line. The squares are arranged in a checkerboard pattern, with alternating shades of orange. The wavy line runs diagonally across the image, adding a sense of movement and dynamism to the design. The overall effect is visually appealing and modern.
K231254 Page 9 of 11
16 (FDA Recognition Number 14-499) confirmed that GripTract met all product specifications and acceptance criteria after being exposed to simulated shipping and transportation conditions. Accelerated Shelf-Life testing confirmed that GripTract met all product specifications and acceptance criteria after exposure to conditions simulating a shelf-life of two years. Reliability testing confirmed that GripTract met all product specifications and acceptance criteria following worst-case simulated use. Bench Safety testing confirmed that GripTract is sufficiently safe during benchtop evaluations that approximate worst-case scenarios and interactions that may occur between biological tissue and the GripTract Fingers. A comparison of the endoscope viewing area between GripTract and the Reference Device confirmed that GripTract does not block visualization relative to a standard distal end cap.
Usability Testing: GripTract was tested with two different user populations based on their interactions with the device, and the tasks and responsibilities they perform. Specifically, there is a user population for setting up the device and another for actually using the device clinically with the Patient. Human factors validation testing of both user populations and tasks confirmed GripTract can safely and effectively be set-up and used for its intended use in the expected use environment without encountering serious use errors or other problems.
Biocompatibility Testing: GripTract is categorized as a surface device contacting breached or compromised surfaces for a limited duration. Cytotoxicity, sensitization, irritation, acute systemic toxicity, and material mediated pyroqenicity testing was conducted on GripTract in its final finished form. All tests confirmed the suitability of GripTract.
Porcine Testing: Worst-case procedures completed with GripTract exhibited equivalent visual and histological mucosal damage compared to endoscope-only (control) procedures and there were no device-attributable adverse events at either acute or chronic time points. This demonstrates that GripTract is safe and effective for its intended use.
{14}------------------------------------------------
Image /page/13/Picture/1 description: The image is a background with a pattern of orange squares and squiggly lines. The squares are arranged in a checkerboard pattern, with alternating light and dark orange squares. The squiggly lines are a lighter shade of orange and run diagonally across the squares. The background is simple and abstract, and the colors are warm and inviting.
K231254
Page 10 of 11
| Summary of non-clinical performance testing. | | | |
|----------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------|----------------------|
| Test<br>Performed | Test Description | Standards<br>Organization<br>and<br>Designation | Results |
| Verification of<br>Product<br>Specification | Verification of Product Specifications<br>after exposure to ambient, high temp,<br>and low temperature environmental<br>conditions. | NA | Pass |
| Distribution | Confirmation of Product Specifications<br>following exposure to simulated<br>distribution stress and conditions. | ASTM<br>D4169-16 | Pass |
| Accelerated<br>Shelf Life | Confirmation of Product Specifications<br>following exposure to accelerated<br>conditions simulating a shelf-life of<br>two years. | NA | Pass |
| Shelf Life | Confirmation of Product Specifications<br>following exposure to real-time<br>ambient conditions for two years. | NA | Test is on-<br>going |
| Reliability | Confirmation of Product Specifications<br>following worst-case simulated use. | NA | Pass |
| Bench Safety | Assessed safety of worst-case<br>interactions between biological tissue<br>and GripTract Fingers. | NA | Pass |
| Usability<br>(Set-Up) | A total of 15 Users assessed whether<br>the intended user population can<br>successfully set up GripTract without<br>serious use errors, failure to perform<br>critical tasks, or encountering<br>problems that affect the intended use<br>and expected use environment. | NA | Pass |
| Usability<br>(Use) | A total of 15 Users assessed whether<br>the intended user population can<br>successfully use GripTract without<br>serious use errors, failure to perform<br>critical tasks, or encountering<br>problems that affect the intended use<br>and expected use environment. | NA | Pass |
| Biocompatibility | Final, finished devices tested for<br>cytotoxicity, sensitization, irritation,<br>acute systemic toxicity, and material<br>mediated pyrogenicity. | ISO 10993-12<br>ISO 10993-5<br>ISO 10993-10<br>ISO 10993-23<br>ISO 10993-11 | Pass |
| Porcine Testing | In vivo porcine testing GripTract<br>against standard of care in worst-case<br>procedure. Acute and chronic time<br>points examined for differences in<br>visual and histological mucosal<br>damage as well as presence of<br>device-attributable adverse events. | NA | Pass |
{15}------------------------------------------------
Image /page/14/Picture/1 description: The image is an abstract design featuring a combination of solid and patterned areas. The left side of the image is a solid orange color, which gradually transitions into a checkered pattern of alternating light and dark orange squares. Overlaying the checkered pattern are wavy lines in a lighter shade of orange, adding a dynamic element to the design.
K231254 Page 11 of 11
# Conclusions
After evaluating GripTract for its intended use, then identifying, evaluating, and mitigating the risks associated with both use and foreseeable misuse, it is concluded that GripTract is substantially equivalent to DiLumen when used as indicated to assist in positioning the distal end of an endoscope from the mucosal surface and assist with optical visualization, diagnosis, and endoscopic treatment.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.