K231039 · Kuraray Noritake Dental, Inc. · KLE · Dec 13, 2023 · Dental
Device Facts
Record ID
K231039
Device Name
CLEARFIL Universal Bond Quick 2
Applicant
Kuraray Noritake Dental, Inc.
Product Code
KLE · Dental
Decision Date
Dec 13, 2023
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 872.3200
Device Class
Class 2
Indications for Use
The subject device is indicated for the following uses: [1] Direct restorations using light-cured composite resin [2] Sealing of a prepared cavity or abutment tooth as a pretreatment for indirect restorations [3] Treatment of exposed root surfaces [4] Treatment of hypersensitive teeth [5] Intraoral repairs of fractured restorations [6] Post cementation and core build-ups [7] Cementation of indirect restorations
Device Story
CLEARFIL Universal Bond Quick 2 is a light-cure dental bonding agent; used for dentin, enamel, and prosthetic materials. System includes BOND and K-ETCHANT Syringe (35% phosphoric acid, colloidal silica). Operates as self-etching or with K-ETCHANT for selective/total-etching. Dual-curing mechanism activated by CLEARFIL DC Activator; optional for specific Kuraray resin cements. Used in dental clinics by dentists. Applied to tooth structures or restorations to facilitate bonding of composite resins or cements. Benefits include improved restoration retention, reduced hypersensitivity, and sealing of prepared surfaces.
Clinical Evidence
Bench testing only. No clinical data. Performance validated via comparative bench studies against primary predicate, including shear/tensile bond strength to bovine enamel/dentin, dental ceramic, and metal; depth of cure; pH; and fluoride release. All tests met in-house acceptance criteria.
Technological Characteristics
Resin-based dental bonding agent. Components: BOND (light-cure adhesive) and K-ETCHANT Syringe (35% phosphoric acid, colloidal silica). Dual-curing capability via activator. pH 2.4 ± 0.4. Evaluated per ISO 10993 and ISO 7405. No electronic components or software.
Indications for Use
Indicated for patients requiring direct light-cured composite resin restorations, sealing of prepared cavities or abutments for indirect restorations, treatment of exposed root surfaces, treatment of hypersensitive teeth, intraoral repair of fractured restorations, post cementation, core build-ups, and cementation of indirect restorations.
Regulatory Classification
Identification
A resin tooth bonding agent is a device material, such as methylmethacrylate, intended to be painted on the interior of a prepared cavity of a tooth to improve retention of a restoration, such as a filling.
{1}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the FDA logo is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
December 13, 2023
Kuraray Noritake Dental Inc. Yasujiro Ohara General Manager Tokiwabashi Tower, 2-6-4, Otemachi Chiyoda-ku, Tokyo 100-0004 JAPAN
Re: K231039
Trade/Device Name: CLEARFIL Universal Bond Quick 2 Regulation Number: 21 CFR 872.3200 Regulation Name: Resin Tooth Bonding Agent Regulatory Class: Class II Product Code: KLE Dated: November 13, 2023 Received: November 13, 2023
Dear Yasujiro Ohara:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{2}------------------------------------------------
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
{3}------------------------------------------------
Sincerely,
# Michael E. Adjodha -S
Michael E. Adjodha, MChE, RAC, CQIA Assistant Director DHT1B: Division of Dental and ENT Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
# Indications for Use
510(k) Number (if known)
#### Device Name
CLEARFIL Universal Bond Quick 2
Indications for Use (Describe)
- [1] Direct restorations using light-cured composite resin
- [2] Sealing of a prepared cavity or abutment tooth as a pretreatment for indirect restorations
[3] Treatment of exposed root surfaces
[4] Treatment of hypersensitive teeth
[5] Intraoral repairs of fractured restorations
[6] Post cementation and core build-ups
[7] Cementation of indirect restorations
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|--|
|-------------------------------------------------|--|
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{5}------------------------------------------------
Date: 11th December, 2023
# 510(k) Summary
## 5-1. 510(k) owner (submitter)
| 1) Name | Kuraray Noritake Dental Inc. |
|-----------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 2) Address | 1621 Sakazu, Kurashiki, Okayama 710-0801, Japan |
| 3) Contact person | Yasujiro Ohara<br>General Manager<br>Quality Assurance Department |
| 4) Contact person in US | Manabu Suzuki<br>Director<br>Dental Material Division<br>KURARAY AMERICA, INC.<br>32 Old Slip, 7th Floor, New York, NY 10005<br>Tel: (212)-356-3940 (Ext. 115) or (800)-879-1676<br>Fax: (888)-700-5200 |
| 5-2. Name of Device | |
| 1) Trade / Proprietary name | CLEARFIL Universal Bond Quick 2 |
| 2) Classification name | Agent, Tooth Bonding, Resin<br>(21 CFR section: 872.3200. Product code: KLE) |
| 3) Common name | Dental Bonding agent |
{6}------------------------------------------------
| 5-3. Predicate devices | | |
|----------------------------------|-----------------|------------------------------|
| 1) CLEARFIL Universal Bond Quick | 510(k) Number | K161042 |
| (Primary predicate device) | Classification: | Agent, Tooth Bonding, Resin |
| | Product Code: | KLE |
| | 21 CFR Section: | 872.3200 |
| | Applicant: | Kuraray Noritake Dental Inc. |
| 2) BeautiBond Xtreme | 510(k) Number | K213965 |
| (Predicate device) | Classification: | Agent, Tooth bonding, Resin |
| | Product Code: | KLE |
| | 21 CFR Section: | 872.3200 |
| | Applicant: | Shofu Dental Corporation |
| 3) ESTENIA C&B | 510(k) Number | K042929 |
| (Reference device) | Classification: | Material, Tooth shade, Resin |
| | Product Code: | EBF |
| | 21 CFR Section: | 872.3690 |
| | Applicant: | Kuraray Noritake Dental Inc. |
### 5-4. Device Description
CLEARFIL Universal Bond Quick 2 consists of BOND and K-ETCHANT Syringe. BOND is a light-cure bonding agent that allows for the treatment of dentin, enamel, and prosthetic materials. Depending on the indication, BOND is used as self-etching or with K-ETCHANT Syringe for selective enamel etching or total-etching procedures. BOND is intended to be used for both direct and indirect restorations. CLEARFIL DC Activator activates the dual-curing mechanism of BOND; however, the addition of CLEARFIL DC Activator to the adhesive is not required when using BOND with CLEARFIL DC CORE PLUS, self-adhesive resin cements manufactured by Kuraray Noritake Dental Inc. (Kuraray's self-adhesive cements, e.g. PANAVIA SA Cement Universal) or PANAVIA Veneer LC. K-ETCHANT Syringe is an etching gel that consists of 35 % phosphoric acid aqueous solution and colloidal silica.
This is the new registration application for the subject device and there have not been any prior submissions regarding the subject device.
We are seeking clearance for CLEARFIL Universal Bond Quick 2 BOND in this permission. Concerning K-ETCHANT Syringe which is included as components of CLEARFIL Universal Bond Quick 2, the specification of the component was omitted since we already have had 510(k) clearance of K-ETCHANT Syringe (510(k) Number: 133078).
#### 5-5. Statement of Intended Use
The subject device is indicated for the following uses:
[1] Direct restorations using light-cured composite resin
- [2] Sealing of a prepared cavity or abutment tooth as a pretreatment for indirect restorations
- [3] Treatment of exposed root surfaces
- [4] Treatment of hypersensitive teeth
- [5] Intraoral repairs of fractured restorations
- [6] Post cementation and core build-ups
- [7] Cementation of indirect restorations
{7}------------------------------------------------
# 5-6. Substantial Equivalence Discussion
## 1) Intended uses
The intended use of the subject device and the predicate devices are listed on the following table.
| | Trade name | Intended use |
|--------------------------------|---------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Subject<br>device | CLEARFIL<br>Universal Bond<br>Quick 2 | [1] Direct restorations using light-cured composite resin<br>[2] Sealing of a prepared cavity or abutment tooth as a pretreatment for indirect restorations<br>[3] Treatment of exposed root surfaces<br>[4] Treatment of hypersensitive teeth<br>[5] Intraoral repairs of fractured restorations<br>[6] Post cementation and core build-ups<br>[7] Cementation of indirect restorations |
| Primary<br>predicate<br>device | CLEARFIL<br>Universal Bond<br>Quick | [1] Direct restorations using light-cured composite resin<br>[2] Cavity sealing as a pretreatment for indirect restorations<br>[3] Treatment of exposed root surfaces<br>[4] Treatment of hypersensitive teeth<br>[5] Intraoral repairs of fractured restorations<br>[6] Post cementation and core build-ups<br>[7] Cementation of indirect restorations |
| Predicate<br>device | BeautiBond<br>Xtreme | [1] Direct restorations with light-cured composite resin<br>[2] Repair of fractured restorations with light-cured composite resin<br>[3] Post cementation and core build-up<br>[4] Cementation of indirect restorations with light-cured/dual-cured resin cement<br>[5] Treatment of hypersensitive teeth or exposed root surface<br>[6] Sealing of tooth preparation (cavity or abutment) for indirect restorations |
The intended use of the subject device is substantially equivalent to that of CLEARFIL Universal Bond Quick (Primary predicate device) or BeautiBond Xtreme (Predicate device).
Therefore, the intended use of the subject device is substantially equivalent to that of the predicate devices.
- 2) Chemical ingredients/ Safety
All ingredients have been used in the primary predicate device or the reference device as described in "Section 12: Substantial Equivalence Discussion". Regarding these primary predicate device and reference device, there have not been any reported problems or recalls according to the post market adverse event reporting requirements in the US.
We evaluated the subject device referring to Use of International Standard ISO 10993-1, "Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process" Guidance for Industry and Food and Drug Administration Staff", ISO 10993 series and ISO 7405. As a result of the evaluation, it was concluded that the subject device is substantially equivalent in biological safety to the primary predicate device and the reference device.
- 3) Comparison of Technological characteristics/ Effectiveness and Performance
Since there have not been any international standards concerning performance of this type of device, certain tests were performed on this device, in comparison with the primary predicate device and it was confirmed that this device was substantially equivalent to the primary predicate device in terms of the effectiveness and performance.
Shear/tensile bond strength and Fluoride release testing were performed to validate the substantial equivalence of the subject device with the primary predicate device in terms of safety and effectiveness. The test has met all the criteria indicating that the subject device was as effective as the primary predicate device.
{8}------------------------------------------------
| Section | Requirement | CLEARFIL<br>Universal Bond<br>Quick 2<br>(Subject device) | CLEARFIL<br>Universal Bond<br>Quick<br>(Primary predicate<br>device) | BeautiBond<br>Xtreme<br>(Predicate device) |
|--------------------------------------------|---------------------------------------------------------------|-----------------------------------------------------------|----------------------------------------------------------------------|--------------------------------------------|
| Appearance | Yellow or orange transparent liquid<br>No extraneous material | COMPLIES | COMPLIES | |
| Easy of application | Easy to apply | COMPLIES | COMPLIES | |
| Depth of cure | Equal to or greater than 0.5 mm | COMPLIES | COMPLIES | |
| Shear bond strength to<br>bovine enamel | Equal to or greater than 10 MPa | COMPLIES | COMPLIES | COMPLIES |
| Shear bond strength to<br>bovine dentin | Equal to or greater than 10 MPa | COMPLIES | COMPLIES | COMPLIES |
| Tensile bond strength to<br>dental ceramic | Equal to or greater than 10 MPa | COMPLIES | COMPLIES | |
| Tensile bond strength to<br>metal | Equal to or greater than 17 MPa | COMPLIES | COMPLIES | |
| Occlusion dentinal<br>tubules | Dentinal tubules are occluded | COMPLIES | COMPLIES | |
| pH | 2.4 ± 0.4 | COMPLIES | COMPLIES | |
The results of comparative study performed were indicated below.
The results indicate that the subject device and the primary predicate device comply with the requirements of the in-house standard. From the above, it can be said that comparative study of the subject device is substantially equivalent to that of the primary predicate device. It also can be said that shear bond strength to bovine enamel/dentin of the subject device is substantially equivalent to that of the predicate device.
Fluoride release testing was performed to validate the substantial equivalence of the subject device with the primary predicate device.
It was concluded that the subject device is substantially equivalent to the primary predicate device.
#### 5-7. Biocompatibility
The subject device is categorized into the external communicating device (tissue/ bone/dentin) and permanent contact device. There is no new ingredient in the subject device. K-ETCHANT Syringe is categorized into the external communicating device that may contact dentin and whose duration of contact is less than 24 hours.
All ingredients in the subject device have been used in the primary predicate device and the reference device as described in "Section 12: Substantial Equivalence Discussion". Regarding these primary predicate device and reference device, there have not been any reported problems or recalls according to the post market adverse event reporting requirements in the US. Moreover, we already have had 510(k) clearance of K-ETCHANT Syringe (510(k) Number: K133078).
Accordingly, it was considered that the subject device was substantially equivalent in biocompatibility to the primary predicate device and the reference device.
#### 5-8. Conclusion
The comparison for intended uses, chemical ingredients/ safety and performance data shows that the subject device is substantially equivalent to the predicate devices and the reference device. This submission information including the nonclinical testing provided supports that the subject device is as safe and as effective as the primary predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.