← Product Code [GCJ](/productcode/GCJ) · K231003

# Laparoscope (21033FA, 21033FC, 21033WA, 21033WC, 20533FA, 20533FC, 20533WA, 20533WC) (K231003)

_Guangdong Optomedic Technologies, Inc. · GCJ · Jul 6, 2023 · Gastroenterology, Urology · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K231003

## Device Facts

- **Applicant:** Guangdong Optomedic Technologies, Inc.
- **Product Code:** [GCJ](/productcode/GCJ.md)
- **Decision Date:** Jul 6, 2023
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 876.1500
- **Device Class:** Class 2
- **Review Panel:** Gastroenterology, Urology
- **Attributes:** Therapeutic

## Indications for Use

The OptoMedic Laparoscopes are intended to be used for endoscopy and endoscopic surgery within the thoracic and peritoneal cavities including the female reproductive organs. The OptoMedic Laparoscopes are intended to be used by trained healthcare professional in diagnostic and therapeutic procedures.

## Device Story

OptoMedic Laparoscopes are rigid endoscopes used for visualization during endoscopic surgery. Device consists of slender, tubular optical instrument containing series of lenses; external light source illuminates tissue via light guide cable; reflected light collected/transmitted through laparoscope to compatible camera head; image signal processed and displayed on monitor. Used in thoracic and peritoneal cavities, including female reproductive organs. Operated by trained healthcare professionals in clinical settings. Provides visualization to assist surgeons in diagnostic and therapeutic procedures; benefits patient through minimally invasive access.

## Clinical Evidence

Bench testing only. No clinical data provided. Performance verified through testing against standards: ANSI/AAMI ES 60601-1, IEC 60601-2-18, ISO 8600 series (optical resolution, field of view, insertion width), and ISO 10993 series (biocompatibility: cytotoxicity, sensitization, systemic toxicity, irritation). Reprocessing validated per AAMI TIR 30 and AAMI TIR 12.

## Technological Characteristics

Rigid endoscope; optical lens system; external light source; reusable. Available in 5.6mm and 10.1mm insertion widths; 310mm and 330mm working lengths; 0° and 30° directions of view. Sterilization via steam or low-temperature plasma. Complies with ISO 8600, IEC 60601-2-18, and ISO 10993 standards.

## Regulatory Identification

An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.

## Special Controls

*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.

In combination with the general controls of the FD&C Act, the integrated operating table-electromechanical surgical system is subject to the following special controls:

1. (1) Premarket clinical performance testing, or a combination of premarket clinical performance testing and postmarket surveillance (in accordance with special control (2)), must include the following:
   1. (i) Objective performance measures (e.g., rate and number of conversions to other surgical modalities, rate of device related adverse events (including tissue injury, hematoma, and increased blood loss), and their severity, cause, and outcomes) must be reported with relevant descriptive comparator performance measures.
   2. (ii) The data must demonstrate the performance of the device for providing accurate and precise control of attached surgical instruments in range of clinical conditions relevant to the device's intended use.
   3. (iii) The test dataset must include data collected from a patient population representative of the intended patient population under anticipated conditions of use.
2. (2) Data obtained from postmarket surveillance must demonstrate, in consideration of the premarket data obtained in accordance with special control (1), that the device performs in accordance with special control (1), unless FDA determines, based on the totality of the premarket data, that data from postmarket surveillance is not required to demonstrate that the device performs as intended. Such postmarket surveillance must be conducted per a protocol determined appropriate by FDA to demonstrate that the device performs as intended (in consideration of the premarket data obtained in accordance with special control (1)), and must include initiation, enrollment, and reporting requirements to ensure timely periodic updates to FDA on post-market surveillance progress and outcomes.
3. (3) Animal performance testing must evaluate the extent of port site trauma due to repositioning of table during surgical procedures when utilizing robotic minimally invasive and laparoscopic approaches
4. (4) The device manufacturer must develop, and update as necessary, a device-specific use training program that ensures proper device setup/use/shutdown, accurate control of instruments to perform the intended surgical procedures, troubleshooting and handling during unexpected events or emergencies, and safe practices to mitigate use error.
5. (5) The device manufacturer may only distribute the device to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program.
(6) Human factors assessment must demonstrate that the user can correctly use the device system across all intended use environments with the provided instructions and training materials, including patient access during normal operating conditions and emergency situations, and effects arising from the integrated nature of the operating table and robotic surgical arms.
(7) Labeling must include:
(i) A detailed summary of clinical performance testing conducted with the device, including study population, results, adverse events, and comparisons to any comparator groups identified;
(ii) A statement in the labeling that the safety and effectiveness for the representative specific procedures was based on evaluation of the device as a surgical tool and did not include evaluation of outcomes related to the treatment of the patient's underlying disease or condition, unless FDA determines that it can be removed or modified based on clinical performance data submitted to FDA;
(iii) Identification of compatible devices;
(iv) The list of surgical procedures for which the device has been determined to be safe with clinical justification;
(v) Reprocessing instructions for reusable components;
(vi) A shelf life for any sterile components;
(vii) A description of the device-specific use training program;
(viii) A statement that the device is only for distribution to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program; and
(ix) A summary of any completed postmarket surveillance data collected as required by special control (2), including updated labeling to accurately reflect outcomes observed in postmarket surveillance.
(8) Non-clinical performance testing must demonstrate that the device performs as intended under anticipated conditions of use and must include:
(i) Device motion accuracy and repeatability;
(ii) System testing;
(iii) Instrument reliability;
(iv) Crosstalk;
(v) Table motion control;
(vi) Thermal effects on tissue;
(vii) User-device interface performance;
(viii) Workspace access testing; and
(ix) Performance testing with compatible devices.
(9) Software verification, validation, and hazard analysis must be performed.
(10) Electromagnetic compatibility and electrical, thermal, and mechanical safety testing must be performed.
(11) Performance data must demonstrate the sterility of all patient-contacting device components.
(12) Performance data must support the shelf life of the device components provided sterile by demonstrating continued sterility and package integrity over the labeled shelf life.
(13) Performance data must validate the reprocessing instructions for the reusable components of the device.
(14) Performance data must demonstrate that all patient-contacting components of the device are biocompatible.
(15) Performance data must demonstrate that all patient-contacting components of the device are non-pyrogenic.
(16) The device manufacturer must submit a report to the FDA annually on the anniversary of initial marketing authorization for the device, until such time as FDA may terminate such reporting, which comprises the following information:
(i) Cumulative summary, by year, of complaints and adverse events since date of initial marketing authorization; and
(ii) Identification and rationale for changes made to the device, labeling, or device specific use training program, which did not require submission of a premarket notification during the reporting period.

## Predicate Devices

- HOPKINS Telescopes ([K223923](/device/K223923.md))
- O-Mec Laparoscopes 690 Series ([K201151](/device/K201151.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo features the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.

July 6, 2023

Guangdong OptoMedic Technologies, Inc. Weijuan Guo Regulatory Affairs Engineer Suite 503, Building A, Golden Valley Intellicreation Community, No. 2 Yonganbei Street Foshan, 528200 China

Re: K231003

Trade/Device Name: Laparoscope (21033FA, 21033FC, 21033WA, 21033WC, 20533FA, 20533FC, 20533WA, 20533WC) Regulation Number: 21 CFR 876.1500 Regulation Name: Endoscope And Accessories Regulatory Class: Class II Product Code: GCJ Dated: June 9, 2023 Received: June 9, 2023

Dear Weijuan Guo:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

Mark
Digitally signed by
Mark Trumbore -S
Trumbore -S Date: 2023.07.06
10:45:39-04'00'

Mark Trumbore, Ph.D. Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

#### K231003

DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

## Indications for Use

Page 1 of 1 Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2023 See PRA Statement below.

Submission Number (if known)

K231003

Device Name

Laparoscope (21033FA, 21033FC, 21033WA, 21033WC, 20533FA, 20533FC, 20533WA, 20533WC)

Indications for Use (Describe)

The OptoMedic Laparoscopes are intended to be used for endoscopy and endoscopic surgery within the thoracic and peritoneal cavities including the female reproductive organs. The OptoMedic Laparoscopes are intended to be used by trained healthcare professional in diagnostic and therapeutic procedures.

Type of Use (Select one or both, as applicable)

> Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/1 description: The image shows the logo for OptoMedic. The logo consists of the word "OptoMedic" in a bold, sans-serif font. The "Opto" portion of the word is in a lighter shade of green, while the "Medic" portion is in black. To the left of the word "OptoMedic" is a green circle that is not fully closed.

Traditional 510(k) Premarket Notification K231003 Laparoscope

# 510(k) Summary

Prepared in accordance with the requirements of 21 CFR Part 807.92

Date Prepared: March 31, 2023

## I. General Information

| 510(k) Submitter/Owner: | Guangdong OptoMedic Technologies, Inc.                          |
|-------------------------|-----------------------------------------------------------------|
|                         | Suite 503, Building A, Golden Valley Intellicreation Community, |
|                         | No. 2 Yonganbei Street, Daxu, Guicheng, Nanhai, Foshan,         |
|                         | Guangdong, 528200, P.R. China                                   |
|                         | Establishment Registration Number: Not yet registered           |
| Contact Person:         | Jane Guo                                                        |
|                         | Regulatory Affairs Engineer                                     |
|                         | Tel: +86 (757) 8670 2920                                        |
|                         | Email: guoweijuan@optomedic.com                                 |

#### II. Device Identification

| Device Trade Name:    | Laparoscope                         |
|-----------------------|-------------------------------------|
| Common or Usual Name: | Laparoscope                         |
| Model:                | 21033FA, 21033FC, 21033WA, 21033WC, |
|                       | 20533WA, 20533WC, 20533FA, 20533FC  |
| Regulation Name:      | Endoscopes and accessories          |
| Regulation Number:    | 21 CFR 876.1500                     |
| Regulatory Class:     | Class II                            |
| Product Code:         | GCJ                                 |
|                       |                                     |

## III. Predicate Device

| Primary Predicate Device   | 510(k) Number: | K223923                       |
|----------------------------|----------------|-------------------------------|
|                            | Product Name:  | HOPKINS Telescopes            |
| Secondary Predicate Device | 510(k) Number: | K201151                       |
|                            | Product Name:  | O-Mec Laparoscopes 690 Series |

{4}------------------------------------------------

Image /page/4/Picture/1 description: The image shows the logo for OptoMedic. The logo consists of a green and yellow circular design on the left, followed by the word "OptoMedic" in bold, black letters. The circular design appears to be an abstract representation of an eye or lens.

Page 2 of 5

## IV. Device Description

The OptoMedic laparoscopes are rigid endoscopes that are used for endoscopy and endoscopic surgery within the thoracic and peritoneal cavities including the female reproductive organs. An endoscope is a slender, tubular optical instrument used as a viewing system for examining an inner part of the body. The inside of the endoscope contains a series of lenses that transmit the endoscopic image, which is illuminated by an external light source.

The Laparoscope has 8 models which are available in two insertion widths (5.6 mm and 10.1mm), two working lengths (320mm and 310mm) and two different directions of view (0° and 30°). The specifications of the proposed laparoscopes are listed in Table 1.

| No. | Model   | Direction of View | Maximum Insertion Portion Width | Working Length |
|-----|---------|-------------------|---------------------------------|----------------|
| 1   | 21033FA | 0°                | 10.1mm                          | 330mm          |
| 2   | 21033FC | 30°               | 10.1mm                          | 330mm          |
| 3   | 21033WA | 0°                | 10.1mm                          | 330mm          |
| 4   | 21033WC | 30°               | 10.1mm                          | 330mm          |
| 5   | 20533FA | 0°                | 5.6mm                           | 310mm          |
| 6   | 20533FC | 30°               | 5.6mm                           | 310mm          |
| 7   | 20533WA | 0°                | 5.6mm                           | 310mm          |
| 8   | 20533WC | 30°               | 5.6mm                           | 310mm          |

|  | Table 1 Model Specifications |
|--|------------------------------|
|--|------------------------------|

## V. Indications for Use

The OptoMedic Laparoscopes are intended to be used for endoscopy and endoscopic surgery within the thoracic and peritoneal cavities including the female reproductive organs. The OptoMedic Laparoscopes are intended to be used by trained healthcare professional in diagnostic and therapeutic procedures.

{5}------------------------------------------------

Image /page/5/Picture/1 description: The image shows the logo for OptoMedic. The logo consists of a green and yellow circular graphic on the left, followed by the word "OptoMedic" in black font. The font is sans-serif and appears to be bolded.

.

## VI. Comparison of Technological Characteristics with The Predicate Device

| Description                           | Subject Device (K231003)                                                                                                                                                                                                                                                                                                                                                                                 | Primary Predicate Device (K223923)                                                                                                                                                                                                                                                                                                                                                                                                    | Secondary Predicate Device (K201151)                                                                                                                                                                                                                                                                                                                                                                                   |
|---------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Regulation Number                     | 21 CFR 876.1500                                                                                                                                                                                                                                                                                                                                                                                          | 21 CFR 876.1500                                                                                                                                                                                                                                                                                                                                                                                                                       | 21 CFR 876.1500                                                                                                                                                                                                                                                                                                                                                                                                        |
| Product Code                          | GCJ, HET, NMH                                                                                                                                                                                                                                                                                                                                                                                            | GCJ                                                                                                                                                                                                                                                                                                                                                                                                                                   | GCJ                                                                                                                                                                                                                                                                                                                                                                                                                    |
| Device class                          | Class II                                                                                                                                                                                                                                                                                                                                                                                                 | Class II                                                                                                                                                                                                                                                                                                                                                                                                                              | Class II                                                                                                                                                                                                                                                                                                                                                                                                               |
| Indication for use                    | The OptoMedic Laparoscopes<br>are intended to be used for<br>endoscopy and<br>endoscopic surgery within the<br>thoracic and peritoneal cavities<br>including the female<br>reproductive organs.<br>The OptoMedic Laparoscopes<br>are intended to be used by<br>trained healthcare professional<br>in diagnostic and therapeutic<br>procedures.                                                           | For telescopes with<br>diameter ranging from<br>3mm-5mm:<br>The HOPKINS Telescopes<br>are intended to provide<br>visualization during<br>laparoscopy, thoracoscopy<br>and general surgery in<br>adults and pediatrics.<br>For telescopes with<br>diameter ranging from<br>5.5mm- 11mm:<br>The HOPKINS Telescopes<br>are intended to provide<br>visualization during<br>laparoscopy, thoracoscopy<br>and general surgery in<br>adults. | The O-Mec Laparoscopes<br>690 Series (Models, 90-<br>331000H, 690-331030H,<br>690-300500H, 690-<br>300530H) are intended to<br>be used by surgeons in<br>diagnostic and therapeutic<br>procedures. Laparoscopic<br>minimally invasive<br>procedures are performed<br>in the abdominal cavity by<br>means of small skin<br>punctures that allow the<br>insertion of the laparoscope<br>and laparoscopic<br>instruments. |
| Prescription/<br>Over-the-counter use | Prescription                                                                                                                                                                                                                                                                                                                                                                                             | Prescription                                                                                                                                                                                                                                                                                                                                                                                                                          | Prescription                                                                                                                                                                                                                                                                                                                                                                                                           |
| Single use /Reusable                  | Reusable                                                                                                                                                                                                                                                                                                                                                                                                 | Reusable                                                                                                                                                                                                                                                                                                                                                                                                                              | Reusable                                                                                                                                                                                                                                                                                                                                                                                                               |
| Principle of<br>operation             | The illumination light enters the<br>laparoscope from the light guide<br>cable, and exits from objective<br>lens to irradiate the tissue. The<br>light reflected by the tissue is<br>collected and transmitted by the<br>laparoscope compatible with a<br>camera head. The image<br>information is converted into an<br>image signal through the camera<br>head and finally displayed on<br>the monitor. | The HOPKINS Telescopes<br>are rigid telescopes that<br>utilize the rod lens<br>technology. At the distal<br>end of the telescope's shaft<br>is the lens and the other<br>end of the shaft is attached<br>to the eyepiece.<br>Throughout the central<br>lumen of the HOPKINS<br>Telescopes, optical glass<br>rods are used to transmit<br>and magnify<br>the image received from                                                       | The illumination light<br>enters the laparoscope from<br>the light guide cable, and<br>exits from objective lens to<br>irradiate the tissue. The<br>light reflected by the tissue<br>is collected and transmitted<br>by the laparoscope<br>compatible with a camera<br>head. The image<br>information is converted<br>into an image signal<br>through the camera head<br>and finally displayed on the                  |
|                                       |                                                                                                                                                                                                                                                                                                                                                                                                          |                                                                                                                                                                                                                                                                                                                                                                                                                                       |                                                                                                                                                                                                                                                                                                                                                                                                                        |
|                                       |                                                                                                                                                                                                                                                                                                                                                                                                          | the lens.                                                                                                                                                                                                                                                                                                                                                                                                                             | monitor.                                                                                                                                                                                                                                                                                                                                                                                                               |
| Physical Characteristics              |                                                                                                                                                                                                                                                                                                                                                                                                          |                                                                                                                                                                                                                                                                                                                                                                                                                                       |                                                                                                                                                                                                                                                                                                                                                                                                                        |
| Endoscope Type                        | Rigid endoscope                                                                                                                                                                                                                                                                                                                                                                                          | Rigid endoscope                                                                                                                                                                                                                                                                                                                                                                                                                       | Rigid endoscope                                                                                                                                                                                                                                                                                                                                                                                                        |
| Working Length                        | 310mm, 330mm                                                                                                                                                                                                                                                                                                                                                                                             | 18cm- 50cm                                                                                                                                                                                                                                                                                                                                                                                                                            | 331mm, 334mm, 301mm, 303mm                                                                                                                                                                                                                                                                                                                                                                                             |
| Maximum insertion<br>portion width    | 5.6mm, 10.1mm                                                                                                                                                                                                                                                                                                                                                                                            | 3mm- 11mm                                                                                                                                                                                                                                                                                                                                                                                                                             | 10.1mm, 5.5mm                                                                                                                                                                                                                                                                                                                                                                                                          |
| Optical Characteristics               |                                                                                                                                                                                                                                                                                                                                                                                                          |                                                                                                                                                                                                                                                                                                                                                                                                                                       |                                                                                                                                                                                                                                                                                                                                                                                                                        |
| Direction of View                     | 0°, 30°                                                                                                                                                                                                                                                                                                                                                                                                  | 0°, 6°, 30°, 45°                                                                                                                                                                                                                                                                                                                                                                                                                      | 0°, 30°                                                                                                                                                                                                                                                                                                                                                                                                                |
| Field of View                         | 68°, 75°                                                                                                                                                                                                                                                                                                                                                                                                 | 55°- 80°                                                                                                                                                                                                                                                                                                                                                                                                                              | 75°                                                                                                                                                                                                                                                                                                                                                                                                                    |
| Light source                          | External                                                                                                                                                                                                                                                                                                                                                                                                 | External                                                                                                                                                                                                                                                                                                                                                                                                                              | External                                                                                                                                                                                                                                                                                                                                                                                                               |
| Reprocessing Methods                  |                                                                                                                                                                                                                                                                                                                                                                                                          |                                                                                                                                                                                                                                                                                                                                                                                                                                       |                                                                                                                                                                                                                                                                                                                                                                                                                        |
| Cleaning                              | Manual, Automatic                                                                                                                                                                                                                                                                                                                                                                                        | Manual, Automatic                                                                                                                                                                                                                                                                                                                                                                                                                     | Manual                                                                                                                                                                                                                                                                                                                                                                                                                 |
| Sterilization                         | Steam sterilization<br>/Low temperature plasma<br>sterilization                                                                                                                                                                                                                                                                                                                                          | Steam (prevacuum),<br>STERRAD 100S,<br>STERRAD NX,<br>STERRAD 100NX,<br>STERIS VPRO1, VPRO 1 Plus                                                                                                                                                                                                                                                                                                                                     | Autoclavable                                                                                                                                                                                                                                                                                                                                                                                                           |

Table 2 General Comparison

.

{6}------------------------------------------------

Image /page/6/Picture/1 description: The image shows the logo for OptoMedic. The logo consists of a green and yellow circular graphic on the left, followed by the word "OptoMedic" in bold, black letters. The circular graphic appears to be a stylized representation of an eye or lens.

Traditional 510(k) Premarket Notification K231003 Laparoscope

The differences technological characteristics do not raise different questions of safety and effectiveness.

## VII. Performance data

Non clinical tests were conducted to verify that the proposed device met all design specifications as is Substantially Equivalent (SE) to the predicate device. The test results demonstrated that the proposed device complies with the following standards:

- ANSI/AAMI ES 60601-1: 2005+A2 (R2012) +A1 Medical Electrical Equipment- Part 1: . General requirements for basic safety and essential performance.
- IEC 60601-2-18:2009 Medical electrical equipment Part 2-18: Particular requirements for • the basic safety and essential performance of endoscopic equipment
- ISO 8600-1-2015 General requirements ●
- ISO 8600-3-2019 Determination of field of view and direction of view of endoscopes with . optics
- ISO 8600-4-2014 Determination of maximum width of insertion portion .
- ISO 8600-5-2020 Determination of optical resolution of rigid endoscopes with optics .
- ISO 10993-5:2009 Biological evaluation of medical devices Part 5: Tests for in vitro . cytotoxicity
- ISO 10993-10:2021 Biological evaluation of medical devices Part 10: Tests for skin ● sensitization
- ISO 10993-11:2017 Biological evaluation of medical devices Part 11: Tests for systemic . toxicity
- ISO 10993-23:2021Biological evaluation of medical devices Part 23: Tests for irritation .

{7}------------------------------------------------

#### K231003

Image /page/7/Picture/1 description: The image shows the logo for OptoMedic. The logo consists of a green and yellow circular graphic to the left of the company name. The company name is written in a bold, sans-serif font, with "Opto" in black and "Medic" in black.

Traditional 510(k) Premarket Notification K231003 Laparoscope

- AAMI TIR 30:2016 A compendium of processes, materials, test methods, and acceptance . criteria for cleaning reusable medical devices
- AAMI TIR 12:2020 Designing, testing, and labeling reusable medical devices for reprocessing . in health care facilities: A guide for medical device manufacturers

Performance testing were also conducted on the subject device and demonstrate that the proposed system performs according to specifications and functions as intended.

#### VIII. Conclusions

The performance testing summarized above supports a substantial equivalence determination. The performance testing demonstrate that the subject device is as safe and as effective as the legally marketed predicate device.

---

**Source:** [https://fda.innolitics.com/device/K231003](https://fda.innolitics.com/device/K231003)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
