K230968 · ShenB Co., Ltd. · GEI · Jul 6, 2023 · General, Plastic Surgery
Device Facts
Record ID
K230968
Device Name
VYBE RF Electrosurgical System
Applicant
ShenB Co., Ltd.
Product Code
GEI · General, Plastic Surgery
Decision Date
Jul 6, 2023
Decision
SESE
Submission Type
Special
Regulation
21 CFR 878.4400
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The VYBE RF Electrosurgical System is intended for use in dermatologic and general surgical procedures for electrocoagulation and hemostasis and the percutaneous treatment of facial wrinkles for use with Fitzpatrick Skin Type I to Skin Type V when using the 1MHz setting. The 2MHz setting has not been evaluated for use in the percutaneous treatment of facial wrinkles and is intended for use in dermatologic and general procedures for electrocoagulation and hemostasis.
Device Story
VYBE RF Electrosurgical System delivers bipolar fractional radiofrequency (RF) energy to target tissue to induce electrothermal coagulation. System components include main console with LCD touch screen, RF generator, power supply, handpiece, foot switch, and sterile disposable 36-needle electrode cartridge (6x6 array). Operated by clinicians in dermatologic/surgical settings. RF energy passes through skin via needle electrodes; creates minor dermal damage for coagulation/hemostasis. Output controlled via console; provides thermal effect on tissue. Benefits include controlled coagulation for surgical procedures and wrinkle treatment. Device is identical to predicate VIVACE; packaging change to blister pack with Tyvek.
Clinical Evidence
Bench testing only. Evaluations included electrical safety (IEC 60601 series), biocompatibility (ISO 10993 series), sterility (ISO 11135, 11138, 11737, 10993-7), software verification/validation, and thermal effects testing per FDA guidance.
Technological Characteristics
Bipolar fractional RF energy source; 1MHz and 2MHz frequencies. 36-needle electrode cartridge (6x6 array, 0.3mm diameter, 0.5-3.5mm depth). Power: 59W max (180 VA). Dimensions: 380x340x1200mm. Components: LCD touch screen, RF generator, SMPS, handpiece, foot switch. Sterile single-use disposable tips. Electrical safety per IEC 60601-1/2/6; biocompatibility per ISO 10993; sterility per ISO 11135/11138/11737.
Indications for Use
Indicated for dermatologic and general surgical electrocoagulation and hemostasis; percutaneous treatment of facial wrinkles in Fitzpatrick Skin Types I-V (1MHz setting only).
Regulatory Classification
Identification
An electrosurgical cutting and coagulation device and accessories is a device intended to remove tissue and control bleeding by use of high-frequency electrical current.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health and Human Services logo on the left and the FDA logo on the right. The FDA logo features the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
July 6, 2023
ShenB Co Ltd % Connie Hoy Consultant Hoy and Associates 1830 Bonnie Way Sacramento, California 95825
Re: K230968
Trade/Device Name: VYBE RF Electrosurgical System Regulation Number: 21 CFR 878.4400 Regulation Name: Electrosurgical cutting and coagulation device and accessories Regulatory Class: Class II Product Code: GEI Dated: June 12, 2023 Received: June 12, 2023
Dear Connie Hoy:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal
{1}------------------------------------------------
statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
| Mark<br>Trumbore -S | Digitally signed by<br>Mark Trumbore -S<br>Date: 2023.07.06<br>14:32:19 -04'00' |
|---------------------|---------------------------------------------------------------------------------|
|---------------------|---------------------------------------------------------------------------------|
Mark Trumbore, Ph.D. Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K230968
Device Name VYBE RF Electrosurgical System
Indications for Use (Describe)
The VYBE RF Electrosurgical System is intended for use in dermatologic and general surgical procedures for electrocoagulation and hemostasis and the percutaneous treatment of facial wrinkles for use with Fitzpatrick Skin Type I to Skin Type V when using the 1MHz setting. The 2MHz setting has not been evaluated for use in the percutaneous treatment of facial wrinkles and is intended for use in dermatologic and general procedures for electrocoagulation and hemostasis.
Type of Use (Select one or both, as applicable)
| <span style="font-size:10pt"><b></b></span> | <span style="font-size:10pt"><b></b></span> |
|--------------------------------------------------------------------------------------|----------------------------------------------------------------------|
| <input checked="true" type="checkbox"/> Prescription Use (Part 21 CFR 801 Subpart D) | <input type="checkbox"/> Over-The-Counter Use (21 CFR 801 Subpart C) |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
#### K230968
# 510(k) Summary VYBE RF
This 510(k) Summary is in accordance with the requirements of the Safe Medical Device Act (SMDA) of 1990. The content of this 510(k) summary is provided in conformance with 21 CFR Part 807.92.
# A. SPONSOR INFORMATION
| Name: | ShenB Co., Ltd. |
|----------|--------------------------------------------------------------------------------|
| Address: | Shenb Bldg 148, Seongsui-ro<br>Seongdong-Gu, Seoul, 04796<br>Republic of Korea |
| Phone: | 82-2-466-0010 |
| Fax: | 82-2-466-5473 |
#### Official Correspondent:
| Name: | Connie Hoy |
|----------|----------------------------------------------------------------------------------------|
| Address: | Hoy & Associates Regulatory Consulting, LLC<br>1830 Bonnie Way<br>Sacramento, CA 95825 |
| Phone: | 530-908-4903 |
| Email: | conniehoy@hoyregulatory.com |
Establishment Reg. No.: 3010226575
Date Prepared: June 12, 2023
{4}------------------------------------------------
### B. DEVICE NAME
| Trade Name: | VYBE RF Electrosurgical System |
|------------------------|----------------------------------------------------------------|
| Common Name: | Electrosurgical System and Accessories |
| Classification Name: | Electrosurgical cutting and coagulation device and accessories |
| Classification Number: | 21 CFR 878.4400 (Class II) |
| Product Code: | OUH |
| Classification Panel: | General and Plastic Surgery |
### C. PREDICATE DEVICE
The VYBE RF Electrosurgical System is substantially equivalent to the VIVACE Electrosurgical System (K193070). Both the subject device (VYBE RF) and the predicate device (VIVACE) are manufactured by ShenB Co., Ltd.
## D. INDICATIONS FOR USE
The VYBE RF Electrosurgical System is intended for use in dermatologic and general surgical procedures for electrocoagulation and hemostasis and the percutaneous treatment of facial wrinkles for use with Fitzpatrick Skin Type I to Skin Type V when using the 1MHz setting. The 2MHz setting has not been evaluated for use in the percutaneous treatment of facial wrinkles and is intended for use in dermatologic and general surgical procedures for electrocoagulation and hemostasis.
# E. DEVICE DESCRIPTION
The VYBE RF Electrosurgical System is comprised of the following components:
- . The system main body, consisting of:
- o LCD touch screen control panel
- High-frequency generating output section (main P.C.B board or RF Generator) O
- Power supply component Switch Mode Power Supply (SMPS) O
- The accessories to the device include:
- Handpiece with Handpiece Connector O
- Disposable micro-needle cartridge (electrode) for insertion into Handpiece o
- Foot switch O
{5}------------------------------------------------
Radio frequency current (RF energy) is delivered from the RF Generator, through the handpiece and electrode tip into the target tissue. As RF energy passes through the skin, it generates an electro thermal reaction, which is capable of coagulating (causing minor dermal damage) the tissue. The bi-polar RF energy is delivered between independent adjacent electrode pairs (total 36 needle electrode, 6 x 6 array insertion). The RF generator and hand piece are not disposable. Each disposable micro-needle cartridge (electrode) is supplied sterile and is for single patient use only and cannot be re-sterilized.
## F. TECHNOLOGICAL SPECIFICATIONS
The VYBE RF Electrosurgical System is substantially equivalent to the noted predicate devices, with respect to technological characteristics, such as intended use, principles of operation, target population, and energy source.
| Electrical voltage and frequency | AC 120V 50/60 Hz | |
|----------------------------------|---------------------------------|--------------------------------|
| Power conception | 180 VA | |
| Maximum Power | 59W 10% (load resistance 500Ω) | |
| Frequency | 1 MHz, 2 MHz | |
| Microneedle<br>Cartridge | Electrode | 36 each, 6 x 6 array |
| | Exposed length | 0.5 ~ 3.5mm (0.1mm increments) |
| | Outer diameter | Ø 0.3mm |
| Dimensions | (W)380mm x (L)340mm x (H)1200mm | |
# G. PERFORMANCE DATA
To establish safety and efficacy of the VYBE RF Electrosurgical System, the following evaluations were completed following standards and FDA guidance documents. All acceptance criteria were met:
- . Electrical Testing per:
- o IEC 60601-1:2005/(R)2012, A1:2012
- o IEC 60601-1-2:2014, Ed. 4.0
- o IEC 60101-2-2:2017, Ed. 6.0
- o IEC 60601-1-6:2010 +A1 2013
- o IEC 62366:2007/AMD 1:2014
- 0 Biocompatibility Testing per:
- o ISO 10993-1:2009
- o ISO 10993-5:2009
- o ISO 10993-10:2010
{6}------------------------------------------------
- Sterility Testing:
- o The micro-needle cartridge is supplied sterility conforms to a Sterility Assurance Level (SAL) of 10-6:
- ISO 11135:2014
- ISO 11138-1:2017
- ISO 11138-2:2017
- ISO 11737-1:2018
- ISO 11737-2:2019
- ISO 10993-7:2008
- Software Verification and Validation Testing:
- o Software verification and validation testing were conducted as recommended by FDA's Guidance for Industry and FDA Staff, "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices", issued May 11, 2005.
- Thermal Testing: .
- o Testing of the thermal effects on tissue was completed in accordance with FDA's Guidance for Industry and FDA Staff "Premarket Notification (510(k)) Submissions for Electrosurgical Devices for General Surgery", issued March 9, 2020.
| Device name | VYBE RF (Subject Device) | PREDICATE VIVACE (K193070) | Comparison |
|---------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------|
| Manufacturer | ShenB Co., Ltd. | ShenB Co., Ltd. | SAME |
| Classification #<br>& Product<br>Code | 878.4400<br>OUH, GEI | 878.4400<br>OUH | SAME |
| Indication for<br>Use | The VYBE RF Electrosurgical System is<br>intended for use in dermatologic and<br>general surgical procedures for<br>electrocoagulation and hemostasis and<br>the percutaneous treatment facial<br>wrinkles for use with Fitzpatrick Skin<br>Type I to Skin Type V when using the<br>1MHz setting. The 2MHz setting has<br>not been evaluated for use in the<br>percutaneous treatment of facial<br>wrinkles and is intended for use in<br>dermatologic and general surgical<br>procedures for electrocoagulation and<br>hemostasis. | The VIVACE Electrosurgical System is<br>intended for use in dermatologic and<br>general surgical procedures for<br>electrocoagulation and hemostasis and<br>the percutaneous treatment facial<br>wrinkles for use with Fitzpatrick Skin<br>Type I to Skin Type V when using the<br>1MHz setting. The 2MHz setting has<br>not been evaluated for use in the<br>percutaneous treatment of facial<br>wrinkles and is intended for use in<br>dermatologic and general surgical<br>procedures for electrocoagulation and<br>hemostasis. | SAME |
#### H. Comparison to the Predicate Device
{7}------------------------------------------------
#### 510(k) Summary
#### VYBE RF
| Device name | VYBE RF (Subject Device) | PREDICATE VIVACE (K193070) | Comparison |
|---------------------|------------------------------------------------|------------------------------------------------|------------|
| Source of<br>Energy | Bipolar Fractional RF | Bipolar Fractional RF | SAME |
| Delivery system | Bipolar Handpiece + Micro needle<br>electrodes | Bipolar Handpiece + Micro needle<br>electrodes | SAME |
| RF Frequency | 1MHz (±10%) | 1MHz (±10%) | SAME |
| | 2MHz (±10%) | 2MHz (±10%) | SAME |
| Max power | 1MHz: 36 W (Load resistance 500Ω) | 1MHz: 36 W (Load resistance 500Ω) | SAME |
#### I. CONCLUSION
The subject VYBE RF Electrosurgical System is identical to the Vivace Electrosurgical device. The only difference is the packaging for the sterile disposable tip has been changed to a blister pak with Tyvek. This change does not raise any new concerns for the safety or effectiveness of the device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.