← Product Code [MAX](/productcode/MAX) · K230942

# SPIRA® Posterior Lumbar Spacers (K230942)

_Camber Spine Technologies · MAX · Jul 11, 2023 · Orthopedic · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K230942

## Device Facts

- **Applicant:** Camber Spine Technologies
- **Product Code:** [MAX](/productcode/MAX.md)
- **Decision Date:** Jul 11, 2023
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 888.3080
- **Device Class:** Class 2
- **Review Panel:** Orthopedic
- **Attributes:** Therapeutic

## Indications for Use

SPIRA® Posterior Lumbar Spacers (SPIRA®-P Posterior Lumbar Spacer, SPIRA®-T Oblique Posterior Lumbar Spacer, and SPIRA®-TA Posterior Lumbar Spacers) are lumbar interbody fusion devices indicated for use at one or more levels of the lumbosacral spine (L1-S1), as an adjunct to fusion in patients with the following indications: degenerative disc disease (DDD), disc herniation with myelopathy and/or radiculopathy), spondylolisthesis, deformity (degenerative scoliosis or kyphosis), spinal stenosis), and failed previous fusion (pseudoarthrosis). DDD is defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic studies. These patients should be skeletally mature and have had at least six (6) months of non-operative treatment. SPIRA® Posterior Lumbar Spacers are intended to be used with additional FDA-cleared supplementary fixation systems. SPIRA® Posterior Lumbar Spacers are intended for use with an autogenous or allogenic bone graft composed of cancellous and/or corticocancellous bone.

## Device Story

SPIRA® Posterior Lumbar Spacers are interbody fusion devices featuring an open matrix design for bone graft packing; used to facilitate fusion in the lumbosacral spine. The device includes a rotating pin for articulation to accommodate specific surgical approaches, including transforaminal insertion. Superior and inferior surfaces are roughened to prevent migration. The device is implanted by a surgeon in an OR setting; it requires use with supplementary fixation systems and autogenous or allogenic bone graft. The device provides structural support to the intervertebral space during the fusion process, potentially alleviating symptoms associated with degenerative disc disease, stenosis, or deformity.

## Clinical Evidence

Bench testing only. Mechanical testing performed per ASTM F2077 (dynamic/static compression shear and axial compression) and ASTM F2267 (subsidence). Results demonstrate performance equivalent to predicate devices.

## Technological Characteristics

Additively manufactured titanium alloy (Ti-6Al-4V) conforming to ASTM F3001-14 (body) and ASTM F136 (internal components). Features open matrix design, rough surface finish, and rotating pin for articulation. Dimensions fall within range of predicate devices. Sterilization method not specified.

## Regulatory Identification

An intervertebral body fusion device is an implanted single or multiple component spinal device made from a variety of materials, including titanium and polymers. The device is inserted into the intervertebral body space of the cervical or lumbosacral spine, and is intended for intervertebral body fusion.

## Special Controls

*Classification.* (1) Class II (special controls) for intervertebral body fusion devices that contain bone grafting material. The special control is the FDA guidance document entitled “Class II Special Controls Guidance Document: Intervertebral Body Fusion Device.” See § 888.1(e) for the availability of this guidance document.(2) Class III (premarket approval) for intervertebral body fusion devices that include any therapeutic biologic (e.g., bone morphogenic protein). Intervertebral body fusion devices that contain any therapeutic biologic require premarket approval.
(c)
*Date premarket approval application (PMA) or notice of product development protocol (PDP) is required.* Devices described in paragraph (b)(2) of this section shall have an approved PMA or a declared completed PDP in effect before being placed in commercial distribution.

## Predicate Devices

- SPIRA®-T Oblique Posterior Lumbar Spacers, SPIRA®-P Posterior Lumbar Spacers ([K210595](/device/K210595.md))
- 4Web Posterior Spine Truss System™ TLIF STS ([K171351](/device/K171351.md), [K143258](/device/K143258.md))
- Globus Medical PATRIOT® Lumbar Spacers – Signature ([K072970](/device/K072970.md), [K122097](/device/K122097.md), [K181357](/device/K181357.md))
- K2M CASCADIA® - TL 3D Interbody System ([K172941](/device/K172941.md))
- Pinnacle InFill (41-TLIF convex, 42-TLIF lordotic oblique, 43-TLIF Contour) ([K150206](/device/K150206.md), [K133721](/device/K133721.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, with the letters "FDA" in a blue square. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

July 11, 2023

Camber Spine Technologies % Christine Scifert Partner MRC Global, LLC 9085 E. Mineral Cir., Suite 110 Centennial. Colorado 80112

Re: K230942

Trade/Device Name: SPIRA® Posterior Lumbar Spacers Regulation Number: 21 CFR 888.3080 Regulation Name: Intervertebral Body Fusion Device Regulatory Class: Class II Product Code: MAX Dated: June 25, 2023 Received: June 26, 2023

Dear Christine Scifert:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal

{1}------------------------------------------------

statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

# Brent Showalter -S

Brent Showalter, Ph.D. Assistant Director DHT6B: Division of Spinal Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

## Indications for Use

510(k) Number (if known) K230942

Device Name SPIRA® Posterior Lumbar Spacers

#### Indications for Use (Describe)

SPIRA® Posterior Lumbar Spacers (SPIRA®-P Posterior Lumbar Spacer, SPIRA®-T Oblique Posterior Lumbar Spacer, and SPIRA®-TA Posterior Lumbar Spacers) are lumbar interbody fusion devices indicated for use at one or more levels of the lumbosacral spine (L1-S1), as an adjunct to fusion in patients with the following indications: degenerative disc disease (DDD), disc hemiation with myelopathy and/or radiculopathy), spondylolisthesis, deformity (degenerative scoliosis or kyphosis), spinal stenosis), and failed previous fusion (pseudoarthrosis). DDD is defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic studies. These patients should be sketally mature and have had at least six (6) months of non-operative treatment. SPIRA® Posterior Lumbar Spacers are intended to be used with additional FDA-cleared supplementary fixation systems. SPIRA® Posterior Lumbar Spacers are intended for use with an autogenous or allogenic bone graft composed of cancellous and/or corticocancellous bone.

| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|

X Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

## CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

## 510(k) Summary SPIRA® Posterior Lumbar Spacers 25 June 2023

| Company:                | Camber Spine Technologies<br>501 Allendale Rd<br>King of Prussia, PA 19406<br>(484) 427-7060 |
|-------------------------|----------------------------------------------------------------------------------------------|
| Company Contact:        | Brooks McAdam<br>VP of Operations<br>(484) 427-7060<br>bmcadam@cambermedtech.com             |
| Official Correspondent: | Christine Scifert – MRC Global, LLC<br>Christine.scifert@askmrcglobal.com<br>901-831-8053    |
| Trade Name:             | SPIRA® Posterior Lumbar Spacers                                                              |
| Common Name:            | Intervertebral Fusion Device With Bone Graft, Lumbar                                         |
| Classification:         | Class II                                                                                     |
| Regulation Number:      | 21 CFR 888.3080 (Intervertebral body fusion device)                                          |
| Panel:                  | Orthopedic                                                                                   |
| Product Code:           | MAX                                                                                          |

#### Device Description:

The Camber Spine Technologies SPIRA® Posterior Lumbar Spacers are lumbar interbody fusion devices that have an open matrix design to permit packing with autogenous graft material to facilitate fusion. The subject submission seeks to expand the indications of the existing SPIRA® Posterior Lumbar Spacers (SPIRA®-P and SPIRA®-T) as well as add the subject SPIRA®-TA components to the Camber Spine Technologies SPIRA®-TA Posterior Lumbar Spacer allows the interbody fusion device to be inserted from a transforaminal approach. The superior and inferior surfaces of the device have a rough surface to help prevent movement of the device while fusion takes place. The device contains a rotating pin that allows for articulation to accommodate the surgical approach. Patients with previous non-fusion spinal surgery at the treated level may be treated.

Camber Spine Technologies SPIRA®-TA device body is additively manufactured from titanium alloy (Ti-6Al-4V) that conforms to ASTM F3001-14. The internal device components (locking cap and pivot pin) are manufactured from titanium alloy (Ti-6Al-4V) per ASTM F136.

{4}------------------------------------------------

#### Indications for Use:

SPIRA® Posterior Lumbar Spacers (SPIRA®-P Posterior Lumbar Spacer, SPIRA®-T Oblique Posterior Lumbar Spacer, and SPIRA®-TA Posterior Lumbar Spacers) are lumbar interbody fusion devices indicated for use at one or more levels of the lumbosacral spine (L1-S1), as an adjunct to fusion in patients with the following indications: degenerative disc disease (DDD), disc herniation with myelopathy and/or radiculopathy), spondylolisthesis, deformity (degenerative scoliosis or kyphosis), spinal stenosis), and failed previous fusion (pseudoarthrosis). DDD is defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic studies. These patients should be skeletally mature and have had at least six (6) months of non-operative treatment. SPIRA® Posterior Lumbar Spacers are intended to be used with additional FDA-cleared supplementary fixation systems. SPIRA® Posterior Lumbar Spacers are intended for use with an autogenous or allogenic bone graft composed of cancellous and/or corticocancellous bone.

#### Substantial Equivalence:

The subject SPIRA® Lumbar Posterior Spacers are substantially equivalent to the following predicate devices:

Primary Predicate:

- . Camber Spine Technologies – SPIRA®-T Oblique Posterior Lumbar Spacers, SPIRA®-P Posterior Lumbar Spacers (K210595)
Secondary Predicate:

- 4Web Posterior Spine Truss System™ TLIF STS (K171351, K143258) ●
- Globus Medical PATRIOT® Lumbar Spacers – Signature (K072970, K122097, K181357)
- K2M CASCADIA® - TL 3D Interbody System (K172941)
- Pinnacle InFill (41-TLIF convex, 42-TLIF lordotic oblique, 43-TLIF Contour) (K150206, K133721)

When comparing the subject SPIRA®-TA and the primary predicate SPIRA®-T interbody fusion device, the only difference is the curved perimeter of the subject device which is attributed to the surgical approach of the device. This design is most similar to the secondary predicate 4Web Posterior Spine Truss System - Curved (K171351, K143258), with the dimensions falling within the range of the secondary predicate Cascadia TL 3D Interbody System (K172941). The materials of the subject device are identical to those of the primary predicate device. The indications for use of the subject device are similar to those of the primary predicate SPIRA®-T device, with the added indications for deformity, similar to the secondary predicate Globus Medical PATRIOT Lumbar Spacers (K072970, K122097, K181357), and the added indication for pseudoarthrosis. Testing shows that the subject SPIRA®-TA performs equivalent to the SPIRA®-T. Therefore, it is concluded that the subject device demonstrates substantial equivalence to the predicate devices and no new issues of safety and effectiveness have been created.

#### Performance Testing:

Mechanical testing, including dynamic compression shear, dynamic axial compression, static compression shear, and static axial compression per ASTM F2077, as well as subsidence per ASTM F2267 have been performed on the subject SPIRA® Posterior Lumbar Spacer interbody device. The results have shown them to be substantially equivalent to the predicate interbody devices.

#### Conclusion:

Based on the test results and the comparison to the predicate devices, the subject device is determined to be substantially equivalent to the predicate devices.

---

**Source:** [https://fda.innolitics.com/device/K230942](https://fda.innolitics.com/device/K230942)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
