Sensititre 20-24 hour Haemophilus influenzae/Streptococcus pneumoniae MIC, or Breakpoint Susceptibility System with Imipenem-relebactam in the dilution, range of 0.03/4-128-4ug/ml (Haemophilus influenzae)
Sensititre 20-24 hour Haemophilus influenzae/Streptococcus pneumoniae MIC, or Breakpoint Susceptibility System with Imipenem-relebactam in the dilution, range of 0.03/4-128-4ug/ml (Haemophilus influenzae)
Applicant
Thermo Fisher Scientific
Product Code
JWY · Microbiology
Decision Date
Jun 30, 2023
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 866.1640
Device Class
Class 2
Intended Use
The Sensititre 20-24-hour Haemophilus influenzae/Streptococcus pneumoniae MIC or Breakpoint Susceptibility System is an in vitro diagnostic product for clinical susceptibility testing of fastidious isolates. This 510(k) is for Imipenem-relebactam in the dilution range of 0.03/4–128/4 µg/ml for testing H. influenzae on the Sensititre 20 - 24-hour Haemophilus influenzae/Streptococcus pneumoniae MIC or Breakpoint Susceptibility System. Imipenem-relebactam has been shown to be active both clinically and in vitro against the following organisms according to the FDA drug label: Haemophilus influenzae
Device Story
Sensititre 20-24 hour Haemophilus influenzae/Streptococcus pneumoniae MIC or Breakpoint Susceptibility System is an in vitro diagnostic device for determining antimicrobial susceptibility. It utilizes microbroth dilution to test fastidious organisms against specific antibiotics, in this case, Imipenem-relebactam. The system provides Minimum Inhibitory Concentration (MIC) values or breakpoint results. Used in clinical microbiology laboratories by trained personnel to guide antibiotic therapy decisions for patients with infections caused by Haemophilus influenzae. The output assists clinicians in selecting appropriate antimicrobial treatments based on the susceptibility profile of the isolated pathogen.
Clinical Evidence
No clinical data provided; device performance is established through bench testing and comparison to reference methods for antimicrobial susceptibility testing.
Technological Characteristics
Microbroth dilution susceptibility system. Consists of multi-well plates containing dehydrated antimicrobial agents (Imipenem-relebactam) in specific dilution ranges. Designed for 20-24 hour incubation of fastidious organisms. Operates via standard broth microdilution principles. Standalone diagnostic system.
Indications for Use
Indicated for clinical susceptibility testing of fastidious Haemophilus influenzae isolates using the Sensititre 20-24 hour MIC or Breakpoint Susceptibility System. Prescription use only.
Regulatory Classification
Identification
An antimicrobial susceptibility test powder is a device that consists of an antimicrobial drug powder packaged in vials in specified amounts and intended for use in clinical laboratories for determining in vitro susceptibility of bacterial pathogens to these therapeutic agents. Test results are used to determine the antimicrobial agent of choice in the treatment of bacterial diseases.
Related Devices
K240445 — The Sensititre 20 - 24 hour Haemophilus influenzae/Streptococcus pneumoniae (HP) MIC or Breakpoint Susceptibility System with Imipenem in the dilution range of 0.015 - 4 ug/ml. · Thermo Fisher Scientific · May 16, 2024
K202612 — Sensititre 20-24 hour Haemophilus influenzae /Streptococcus pneumoniae MIC or Breakpoint Susceptibility System with Dtest (containing erythromycin at 1 ug/mL and clindamycin at 0.5 ug/mL) · Thermo Fisher Scientific · Jul 23, 2021
K193024 — Sensititre 20-24 hour Haemophilus influenzae/Streptococcus pneumoniae MIC or Breakpoint, Susceptibility System with Lefamulin in the dilution range of 0.008-16ug/mL · Thermo Fisher Scientific · Jan 8, 2020
K192250 — Sensititre 18-24 hour MIC or Breakpoint Susceptibility System with Imipenem-Relebactam in the dilution range of 0.03/4-256/4 ug/mL · Thermo Fisher Scientific · Oct 16, 2019
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logos of the Department of Health & Human Services and the Food and Drug Administration (FDA). The Department of Health & Human Services logo is on the left, and the FDA logo is on the right. The FDA logo is a blue square with the letters "FDA" in white, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue.
June 30, 2023
Thermo Fisher Scientific Cynthia Knapp Director R&D, AST and Pharma One Thermo Fisher Way Oakwood Village, Ohio 44146
## Re: K230935
Trade/Device Name: Sensititre 20-24 hour Haemophilus influenzae/Streptococcus pneumoniae MIC, or Breakpoint Susceptibility System with Imipenem-relebactam in the dilution, range of 0.03/4-128-4ug/ml (Haemophilus influenzae) Regulation Number: 21 CFR 866.1640 Regulation Name: Antimicrobial Susceptibility Test Powder Regulatory Class: Class II Product Code: JWY, LRG, LTT Dated: March 31, 2023 Received: April 3, 2023
Dear Cynthia Knapp:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801 and Part 809); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
# Ribhi Shawar -S
Ribhi Shawar, Ph.D. (ABMM) Chief General Bacteriology and Antimicrobial Susceptibility Branch Division of Microbiology Devices OHT7: Office of In Vitro Diagnostics and Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K230935
#### Device Name
The Sensititre 20 - 24 hour Haemophilus influenzae/Streptococus pneumoniae (HP) MIC or Breakpoint Susceptibility System with Imipenem-relebactam in the dilution range of 0.03/4-128/4 ug/ml .
#### Indications for Use (Describe)
The Sensititre 20 - 24 hour Haemophilus influenzae/Streptococcus pneumoniae MC or Breakpoint System is an in vitro diagnostic product for clinical susceptibility testing of fastidious isolates.
This 510(k) is for Imipenem-relebactam in the dilution range of 0.03/4-128/4 ug/ml for testing fastidious Haemophilus influenzae on the Sensitite 20 - 24 hour Haemophilus influenzae/Streptoccus pneumoniae MIC or Breakpoint Susceptibility System.
Imipenem-relebactam has been shown to be active both clinically and in vitro against the following organisms according to the FDA drug label:
Haemophilus influenzae
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|--|
|-------------------------------------------------|--|
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.