OptiMesh Multiplanar Expandable Interbody Fusion System
K230927 · Spineology, Inc. · OQB · Nov 1, 2023 · Orthopedic
Device Facts
Record ID
K230927
Device Name
OptiMesh Multiplanar Expandable Interbody Fusion System
Applicant
Spineology, Inc.
Product Code
OQB · Orthopedic
Decision Date
Nov 1, 2023
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.3085
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The OptiMesh Multiplanar Expandable Interbody Fusion System is indicated for use as an adjunct to fusion in an intervertebral body fusion at one level in the lumbar spine from L2 to S1 in skeletally mature patients with degenerative disc disease (DDD) with up to Grade I spondylolisthesis at the involved level. DDD is defined as discogenic back pain with degeneration of the disc confirmed by patient history, physical examination, and radios. Eligible patients shall have undergone six (6) months of conservative (non-operative) care. The OptiMesh device with compatible allograft, or a bone void filler as cleared by FDA for use in intervertebral body fusion to facilitate fusion, is intended for use with supplemental posterior fixation systems intended for use in the lumbar spine.
Device Story
OptiMesh is a non-rigid, expandable intervertebral body graft containment device; designed to contain bone graft within an internal cavity. AFT Allograft Filler Tube (composed of human demineralized bone matrix, nondemineralized bone, and sodium hyaluronate) is provided pre-loaded in a delivery tube for use as a bone void filler. Device is inserted into the intervertebral disc space during spinal fusion surgery. Operated by surgeons in a clinical/OR setting. The system facilitates fusion by containing graft material; used alongside supplemental posterior fixation. Benefits include stabilization of the intervertebral space and support for bone fusion in patients with DDD.
Clinical Evidence
No new clinical trials were conducted. Performance was supported by a critical engineering and clinical comparison to a bone graft used in a prospective, multi-center, non-randomized, FDA/IRB-approved performance goal clinical investigation. Bench testing included mechanical ASTM testing and comparison to ISO 23089-2 data, biocompatibility, pyrogenicity, and risk assessment.
Technological Characteristics
Non-rigid, expandable intervertebral graft containment device. AFT filler composed of human demineralized bone matrix, human nondemineralized bone, and sodium hyaluronate. Mechanical performance evaluated per ASTM standards and ISO 23089-2. Device is provided sterile for single-patient use.
Indications for Use
Indicated for skeletally mature patients with degenerative disc disease (DDD) and up to Grade I spondylolisthesis at one lumbar level (L2-S1) who have failed 6 months of conservative care. Used as an adjunct to intervertebral body fusion with supplemental posterior fixation.
Regulatory Classification
Identification
An intervertebral body graft containment device is a non-rigid, implanted spinal device that is designed to contain bone graft within its internal cavity. The device is inserted into the intervertebral body space of the spine and is intended as an adjunct to intervertebral body fusion.
Special Controls
*Classification.* Class II (special controls). The special controls for this device are:(1) Clinical performance testing must include an assessment of any adverse events observed during clinical use, as well as intervertebral body fusion, and compare this to a clinically acceptable fusion rate.
(2) Non-clinical performance testing must demonstrate the mechanical function and durability of the implant, as well as the ability of the device to be inserted, deployed, and filled with bone graft consistently.
(3) Device must be demonstrated to be biocompatible.
(4) Validation testing must demonstrate the cleanliness and sterility of, or the ability to clean and sterilize, the device components, and device-specific instruments.
(5) Design characteristics of the device, including engineering schematics, must ensure that the geometry and material composition are consistent with the intended use.
(6) Labeling must bear all information required for the safe and effective use of the device, specifically including the following:
(i) A clear description of the technological features of the device including identification of device materials, compatible components in the fusion construct, and the principles of device operation;
(ii) Intended use and indications for use, including levels of fixation;
(iii) Identification of magnetic resonance (MR) compatibility status;
(iv) Cleaning and sterilization instructions for devices and instruments that are provided nonsterile to the end user; and
(v) Detailed instructions of each surgical step, including device removal.
Predicate Devices
Thoracolumbar Interbody Systems and Attrax Putty (K203714)
{0}------------------------------------------------
November 1, 2023
Image /page/0/Picture/1 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA acronym followed by the full name of the agency on the right. The FDA part of the logo is in blue, with the acronym in a larger font size than the rest of the name. The logo is simple and professional, reflecting the agency's role in protecting public health.
Spineology Inc. Andrew Adams Group Director of Regulatory & Ouality Affairs 7800 3rd Street North Suite 600 Saint Paul, Minnesota 55128
Re: K230927
Trade/Device Name: OptiMesh Multiplanar Expandable Interbody Fusion System Regulation Number: 21 CFR 888.3085 Regulation Name: Intervertebral body graft containment device Regulatory Class: Class II Product Code: OQB, MBP Dated: March 31, 2023 Received: April 3, 2023
Dear Mr. Adams:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"
{1}------------------------------------------------
## K230927 - Andrew Adams
(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100. Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
> Sincerely, Jesse Muir -S Date: 2023.11.01 14:56:03 Jesse Muir, Ph.D. Assistant Director DHT6C: Division of Restorative, Repair and Trauma Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
# 510(k) Number (if known)
K230927
## Device Name
OptiMesh Multiplanar Expandable Interbody Fusion System
## Indications for Use (Describe)
The OptiMesh Multiplanar Expandable Interbody Fusion System is indicated for use as an adjunct to fusion in an intervertebral body fusion at one level in the lumbar spine from L2 to S1 in skeletally mature patients with degenerative disc disease (DDD) with up to Grade I spondylolisthesis at the involved level. DDD is defined as discogenic back pain with degeneration of the disc confirmed by patient history, physical examination, and radios. Eligible patients shall have undergone six (6) months of conservative (non-operative) care. The OptiMesh device with compatible allograft, or a bone void filler as cleared by FDA for use in intervertebral body fusion to facilitate fusion, is intended for use with supplemental posterior fixation systems intended for use in the lumbar spine.
| Type of Use (Select one or both, as applicable) | |
|----------------------------------------------------------------------------------|--------------------------------------------------------------------------------|
| <div> <span> Prescription Use (Part 21 GER 801 Subpart D) </span> </div> | <div> <span> Over-The-Counter Use (21 CER 801 Subpart C) </span> </div> |
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# Indications for Use
510(k) Number (if known)
K230927
### Device Name
AFT Allograft Filler Tube
## Indications for Use (Describe)
AFT is intended for use as a bone void filler in the extremities, spine, intervertebral disc space, and pelvis for voids or gaps that are not intrinsic to the stability of the bony structure. AFT is indicated for use in the treatment of osseous defects caused by surgery or traumatic injury. When used in intervertebral body fusion procedures, AFT must be used on its own with an intervertebral body graff containment device cleared by FDA for use with a bone void filler.
AFT is intended for single patient use only.
Type of Use (Select one or both, as applicable):
| <span style="font-family: Arial;"> <svg height="12" width="12"> <rect fill="none" height="12" stroke="black" width="12" x="0" y="0"></rect> <path d="M2,2 L10,10 M2,10 L10,2" stroke="black"></path> </svg> Prescription Use (Part 21 CFR 801 Subpart D)</span> |
|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| <span style="font-family: Arial;"> <svg height="12" width="12"> <rect fill="none" height="12" stroke="black" width="12" x="0" y="0"></rect> </svg> Over-The-Counter Use (21 CFR 801 Subpart C)</span> |
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
# 510(k) Summary
| Spineology | 510(k) Summary |
|--------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------|
| Date Prepared: | October 25, 2023 |
| Submitter: | Spineology Inc.<br>7800 3rd Street North<br>Suite 600<br>Saint Paul, MN 55128<br>Establishment Registration Number: 2135156 |
| Contact Person: | Andrew Adams<br>Group Director of Regulatory & Quality Affairs<br>Phone: 651.256.8500<br>Fax: 651.256.8505<br>Email: aadams@spineology.com |
| Device Name and Classification | |
| Trade Name: | OptiMesh Multiplanar Expandable Interbody Fusion System |
| Classification Name: | Intervertebral Body Graft Containment Device |
| Product Codes: | OQB |
| Regulatory Class: | Class II |
| Regulation Number: | 21 CFR 888.3085 |
| Panel: | Orthopedic |
| Trade Name: | AFT Allograft Filler Tube |
| Classification Name: | Resorbable Calcium Salt Bone Void Filler Device |
| Product Codes: | MBP |
| Regulatory Class: | Class II |
| Regulation Number: | 21 CFR 888.3045 |
| Panel: | Orthopedic |
| Predicate Device | |
| Primary: | K203714 Thoracolumbar Interbody Systems and<br>Attrax Putty |
| Reference: | DEN200010 Spineology Interbody Fusion System |
| Reference: | K060161 AFT Allograft Filler Tube |
{5}------------------------------------------------
### 1. Purpose
The purpose of this premarket notification is to obtain FDA clearance for the addition of AFT Allograft Filler Tube as a compatible allograft for use with the OptiMesh Multiplanar Interbody Fusion System (formerly branded as Spineology Interbody Fusion System).
### 2. Device Description
The OptiMesh Multiplanar Expandable Interbody Fusion System is an intervertebral body graft containment device that is a non-rigid, implanted spinal device that is designed to contain bone graft within its internal cavity. The device is inserted into the intervertebral body space of the spine and is intended as an adjunct to intervertebral body fusion.
AFT Allograft Filler Tube is composed of human demineralized bone matrix, human nondemineralized bone, and sodium hyaluronate. All components of AFT are resorbable. AFT is aseptically processed and provided pre-loaded into a disposable delivery tube.
### Indications for Use 3.
# OptiMesh Multiplanar Expandable Interbody Fusion System
The OptiMesh Multiplanar Expandable Interbody Fusion System is indicated for use as an adjunct to fusion in an intervertebral body fusion at one level in the lumbar spine from L2 to S1 in skeletally mature patients with degenerative disc disease (DDD) with up to Grade I spondylolisthesis at the involved level. DDD is defined as discogenic back pain with degeneration of the disc confirmed by patient history, physical examination, and radiographic studies. Eligible patients shall have undergone six (6) months of conservative (non-operative) care. The OptiMesh device with compatible allograft, or a bone void filler as cleared by FDA for use in intervertebral body fusion to facilitate fusion, is intended for use with supplemental posterior fixation systems intended for use in the lumbar spine.
# AFT Allograft Filler Tube
AFT is intended for use as a bone void filler in the extremities, spine, intervertebral disc space, and pelvis for voids or gaps that are not intrinsic to the stability of the bony structure. AFT is indicated for use in the treatment of osseous defects caused by surgery or traumatic injury. When used in intervertebral body fusion procedures, AFT must be used on its own with an intervertebral body graft containment device cleared by FDA for use with a bone void filler.
AFT is intended for single patient use only.
### 4. Technological Characteristics
When compared to the predicate device, the OptiMesh Multiplanar Expandable Interbody Fusion System with AFT has the same intended use and the same or similar technological characteristics, including:
- Indications for Use ●
- Function / Performance ●
- Control Mechanism ●
- . Energy Type
- Principle of Operation ●
- . Use with Compatible Fill Material
- Use with Supplemental Fixation Systems .
{6}------------------------------------------------
Different technological characteristics exist, including:
- Materials of Construction ●
- Structure of Construction
- Mechanism of Mechanical Support ●
The subject OptiMesh interbody device and AFT bone void filler device are identical in technological characteristics to the reference devices. The reference devices support the comparison of the subject device to the primary predicate device, particularly that the differences in technological characteristics do not raise different questions of safety and effectiveness.
### ട. Non-Clinical Testing
Non-clinical testing was conducted to support AFT as a compatible allograft for use with the OptiMesh device confirming function and performance.
- . Benchtop mechanical ASTM testing and comparison confirmed that the OptiMesh device with AFT performs as intended in comparison to the reference device and ISO 23089-2 mechanical performance data.
- A critical comparison between AFT bone void filler and the reference bone graft and its ● impact on bone biology was conducted and assessed.
- Biocompatibility, pyrogenicity / endotoxin monitoring, sterilization, packaging, and shelf-● life were compared and assessed.
- . A risk assessment was performed and confirmed that the OptiMesh device with AFT does not alter the risk profile for the device or present new issues of safety or effectiveness when compared to the predicate device.
### 6. Clinical Performance Data
Clinical data has been utilized to demonstrate the performance and safety of the addition of AFT as a compatible allograft for use with the OptiMesh Multiplanar Interbody Fusion System. The clinical data includes evaluation through a critical engineering and clinical comparison to a bone graft that was used in a prospective, multi-center, non-randomized, FDA and IRB approved performance goal clinical investigation. This evaluation supports the performance of the subject device for its intended use.
### 7. Conclusion
Based on the intended use, technological characteristics, comparison to the predicate device, and support from the reference devices, the subject addition of AFT Allograft Filler Tube as a compatible allograft for use with the OptiMesh Multiplanar Expandable Interbody Fusion System has been shown to be substantially equivalent to the legally marketed predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.