FebriDx Bacterial / Non-bacterial Point of Care Assay
K230917 · Lumos Diagnostics, Inc. · QXA · Jun 30, 2023 · Microbiology
Device Facts
Record ID
K230917
Device Name
FebriDx Bacterial / Non-bacterial Point of Care Assay
Applicant
Lumos Diagnostics, Inc.
Product Code
QXA · Microbiology
Decision Date
Jun 30, 2023
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 866.3230
Device Class
Class 2
Indications for Use
The FebriDx Bacterial/Non-Bacterial Point-of-Care Assay is a qualitative visually read rapid immunoassay for the detection of human host response proteins, Myxovirus resistance protein A (MxA) and C-reactive protein (CRP) directly from fingerstick blood. FebriDx is indicated for use in patients aged 12-64 who present to urgent care or emergency care settings for evaluation of acute respiratory infection who have had symptoms for less than 7 days and within 3 days of fever onset. FebriDx test results are intended to be used in conjunction with other clinical and diagnostic findings as an aid in the diagnosis of bacterial acute respiratory infection and differentiation from non-bacterial etiology. The assessment of whether a bacterial infection is present should always be based on consideration of all available information, and not based solely on the FebriDx test results. FebriDx test results are not intended to identify a specific pathogen or the severity of infection. For in vitro diagnostic use. For prescription use only.
Device Story
Single-use, integrated lateral flow immunoassay; detects host response proteins MxA and CRP from fingerstick blood. Device includes built-in safety lancet, collection tube, and buffer delivery system. Operator adds blood to device; buffer lyses leukocytes to release intracellular MxA. Analyte-antibody-bead complexes migrate across nitrocellulose membrane; captured by immobilized antibodies. Visual output: black line indicates bacterial infection; red line or no line indicates non-bacterial etiology. Used in urgent/emergency care by professionals. Results aid clinical diagnosis of bacterial vs. non-bacterial respiratory infection; must be used with other clinical findings. Benefits include rapid differentiation of infection etiology to guide management.
Clinical Evidence
Prospective, multi-center U.S. clinical trial (n=520 symptomatic ARI patients). Comparator: Clinical Reference Algorithm (pathogen detection + host immune response). Primary endpoints: PPA 93.2% (84.9-97.0% CI), NPA 88.4% (85.0-91.1% CI). Microbiologically confirmed subgroup: PPA 100%, NPA 91.0%. No significant performance differences across age groups.
Technological Characteristics
Lateral flow immunochromatography; detects MxA and CRP. Single-use disposable cartridge with integrated lancet and buffer delivery. Nitrocellulose membrane with immobilized capture antibodies. Visual readout. No electronic components or software. Sterilization/materials not specified beyond standard IVD requirements.
Indications for Use
Indicated for patients aged 12-64 presenting to urgent or emergency care with acute respiratory infection symptoms <7 days and fever onset within 3 days. Not for use in pediatric patients <12 or elderly patients >64.
Regulatory Classification
Identification
A device to detect and measure non-microbial analytes to aid in the detection and identification of localized human infections is identified as an in vitro diagnostic device intended for the detection and qualitative measurement, quantitative measurement, or both of one or more non-microbial analytes in human clinical specimens to aid in the assessment, identification, or both of a localized microbial infection when used in conjunction with clinical signs and symptoms and other clinical and laboratory findings.
Special Controls
The special controls for this device are:
*Classification.* Class II (special controls). The special controls for this device are:(1) Any sample collection device used must be FDA-cleared, -approved, or -classified as 510(k) exempt (standalone or as part of a test system) for the collection of human specimens; alternatively, the sample collection device must be cleared in a premarket submission as a part of this device.
(2) The labeling required under § 809.10(b) of this chapter must include:
(i) An intended use with a detailed description of what the device detects and measures, the type of results provided to the user, the sample type, whether the measure is qualitative and/or quantitative, the clinical indications for the test use, and the specific population(s) for which the device is intended.
(ii) A detailed description of the performance characteristics of the device for all intended specimen types from the analytical and clinical studies (as applicable) required under paragraphs (b)(3)(ii) and (iii) of this section.
(iii) A detailed explanation of the interpretation of results, including acceptance criteria for evaluating the validity of individual runs (
*e.g.,* assessment of internal and/or external quality controls, as applicable).(iv) The following limiting statements:
(A) A statement that a negative test result does not preclude the possibility of infection;
(B) A statement that the test results should be interpreted in conjunction with other clinical and laboratory data available to the clinician;
(C) A statement that consistent device performance is dependent on adequate specimen collection, transport, storage, and processing. Failure to observe proper procedures in any one of these steps can lead to incorrect results; and
(D) A statement that details any limitations associated with the samples, as appropriate (
*e.g.,* collected on the day of admission to the intensive care unit).(3) Design verification and validation must include the following:
(i) A detailed device description, including as appropriate, all device parts; control elements incorporated into the test procedure; instrument requirements; reagents required but not provided; and the principle of device operation and test methodology, including all preanalytical methods for the processing of specimens and the methodology from obtaining a sample to the result; design of primer/probe sequences; rationale for target analyte selection; and computational path from collected raw data to reported result (
*e.g.,* how collected raw signals are converted into a reported result).(ii) Detailed documentation of analytical studies including analytical sensitivity (Limit of Detection, Limit of Quantitation, and Limit of Blank), inclusivity, cross-reactivity, microbial interference, interfering substances, competitive inhibition, carryover/cross-contamination, specimen stability, within-lab precision, reproducibility, and linearity, as applicable.
(iii) Detailed documentation and results either from a clinical study, that includes prospective (sequentially collected) samples for each intended specimen type that are representative of the intended use populations and, when determined to be acceptable by FDA, additional characterized clinical samples; or, when determined to be acceptable by FDA, an equivalent sample set. The clinical study must compare the device performance to results obtained from an FDA-accepted reference method and/or FDA-accepted comparator method, as appropriate. Documentation from the clinical studies must include the clinical study protocol (
*e.g.,* the predefined statistical analysis plan), clinical study report, testing results, and results of all statistical analyses.(iv) An evaluation of the level of the non-microbial analyte in asymptomatic patients with demographic characteristics (
*e.g.,* age, racial, ethnic, and sex distribution) similar to the intended use population of the device.(v) Documentation of an appropriate end user device training program that will be offered as part of efforts to mitigate the risks of false results, failure to operate the device correctly, and failure to interpret test results correctly.
(vi) An appropriate risk mitigation strategy to ensure that the device does not prevent any other device(s) with which it is indicated for use, including incorporated device(s), from achieving their intended use (
*e.g.,* safety and effectiveness of the functions of the indicated device(s) remain unaffected).(vii) A detailed description of the impact of any software, including software applications and hardware-based devices that incorporate software, on the device's functions.
Predicate Devices
Synovasure Alpha Defensin Lateral Flow Test Kit (DEN180032)
Submission Summary (Full Text)
{1}
FDA
U.S. FOOD & DRUG
ADMINISTRATION
# 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY
ASSAY ONLY
## I Background Information:
A 510(k) Number
K230917
B Applicant
Lumos Diagnostics, Inc.
C Proprietary and Established Names
FebriDx Bacterial / Non-bacterial Point-of-Care Assay
D Regulatory Information
| Product Code(s) | Classification | Regulation Section | Panel |
| --- | --- | --- | --- |
| QXA | Class II | 21 CFR 866.3230 - Device To Detect And Measure Non-Microbial Analytes To Aid In The Detection And Identification Of Localized Human Infections | MI - Microbiology |
## II Submission/Device Overview:
A Purpose for Submission:
To obtain a substantial equivalence determination for the FebriDx Bacterial/Non-bacterial Point-of-Care Assay.
B Measurand:
Myxovirus resistance protein A (MxA) and C-reactive protein (CRP)
C Type of Test:
Lateral flow immunochromatography
Food and Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993-0002
www.fda.gov
{2}
K230917 - Page 2 of 16
## III Intended Use/Indications for Use:
### A Intended Use(s):
See Indications for Use below.
### B Indication(s) for Use:
The FebriDx Bacterial/Non-Bacterial Point-of-Care Assay is a qualitative visually read rapid immunoassay for the detection of human host response proteins, Myxovirus resistance protein A (MxA) and C-reactive protein (CRP) directly from fingerstick blood. FebriDx is indicated for use in patients aged 12-64 who present to urgent care or emergency care settings for evaluation of acute respiratory infection who have had symptoms for less than 7 days and within 3 days of fever onset.
FebriDx test results are intended to be used in conjunction with other clinical and diagnostic findings as an aid in the diagnosis of bacterial acute respiratory infection and differentiation from non-bacterial etiology. The assessment of whether a bacterial infection is present should always be based on consideration of all available information, and not based solely on the FebriDx test results. FebriDx test results are not intended to identify a specific pathogen or the severity of infection.
For in vitro diagnostic use
For prescription use only
### C Special Conditions for Use Statement(s):
Rx - For Prescription Use Only
### D Special Instrument Requirements:
None.
## IV Device/System Characteristics:
### A Device Description:
FebriDx is a 10-minute, point-of-care lateral flow assay that detects Myxovirus resistance protein A (MxA) and C-reactive protein (CRP) in patient fingerstick samples to aid in the evaluation of an infection in patients presenting with acute respiratory symptoms.
The single-use disposable FebriDx test includes a lateral flow test strip, a built-in retractable safety lancet, blood collection and transfer tube, and buffer delivery system.
### B Principle of Operation:
The FebriDx test includes a built-in sample collection and transfer tube detects the presence of MxA and CRP in fingerstick blood specimens using lateral flow technology. A sample of the
{3}
fingerstick blood is added to the lateral flow test device followed by a running buffer that provides sufficient volume to activate the test. The running buffer contains leukocyte membrane lysing agents that release intracellular MxA to allow subsequent detection. The first pad in the device filters out the cellular material as well as intact red-blood cells. The filtered sample then contacts a pad that contains the reagents to adjust pH and reduce potential non-specific binding. Prior to reaching the test strip, free MxA and CRP migrate through a dried formulation of latex beads that have been further conjugated to antibodies specific for binding a particular analyte (MxA or CRP). As the analyte-antibody-bead complex continues to migrate across a porous nitrocellulose membrane, it can interact with one of three capture antibodies that are immobilized on the surface at distinct line positions, including the control line to verify that the sample flowed properly across the device. In the presence of a valid blue control line, the FebriDx test result may be interpreted as one of two different outcomes: the presence of a single black line in the result window indicates a positive bacterial infection result whereas a red line in the result window or the absence of any line in the result window indicates a nonbacterial etiology.
V Substantial Equivalence Information:
A Predicate Device Name(s):
Synovasure Alpha Defensin Lateral Flow Test Kit
B Predicate 510(k) Number(s):
DEN180032
C Comparison with Predicate(s):
| Device & Predicate Device(s): | K230917 | DEN180032 |
| --- | --- | --- |
| Device Trade Name | FebriDx Bacterial/Non-bacterial Point-of-Care Assay | Synovasure Alpha Defensin Lateral Flow Test Kit |
| General Device Characteristic Similarities | | |
| Intended Use/Indications For Use | The FebriDx Bacterial/Non-bacterial Point-of-Care Assay is a qualitative visually read rapid immunoassay for the detection of human host response proteins, Myxovirus resistance protein A (MxA) and C-reactive protein (CRP) directly from fingerstick blood. FebriDx is indicated for use in patients aged 12-64 who present to urgent care or | The Synovasure Alpha Defensin Lateral Flow Test Kit is a qualitative visually read immunochromatographic assay for the detection of human host response proteins, Alpha Defensins 1-3, in the synovial fluid of adults with a total joint replacement who are being evaluated for revision surgery. The Synovasure Alpha Defensin Lateral Flow Test Kit results are intended to be used in conjunction |
K230917 - Page 3 of 16
{4}
K230917 - Page 4 of 16
| | emergency care settings for evaluation of acute respiratory infection who have had symptoms for less than 7 and within 3 days of fever onset.
FebriDx test results are intended to be used in conjunction with other clinical and diagnostic findings as an aid in the diagnosis of bacterial acute respiratory infection and differentiation from non-bacterial etiology. The assessment of whether a bacterial infection is present should always be based on consideration of all available information, and not based solely on the FebriDx test results. FebriDx test results are not intended to identify a specific pathogen or the severity of infection. | with other clinical and diagnostic findings as an aid in the diagnosis of periprosthetic joint infection (PJI). The Synovasure Alpha Defensin Lateral Flow Test Kit is not intended to identify the etiology or severity of a PJI. |
| --- | --- | --- |
| Intended User | Same | Professional Use |
| Test Type | Same | Visually read lateral flow immunochromatographic assay |
| Result Output | Same | Qualitative |
| Analyte | Same | Host biomarkers of infection |
| **General Device Characteristic Differences** | | |
| Measurand | MxA and CRP | Human alpha defensins 1-3 |
| Sample type | Fingerstick blood | Synovial fluid |
| Sample volume | 5 μL | 15 μL |
{5}
VI Standards/Guidance Documents Referenced:
1. ISO 10993-1 Fifth edition 2018-08 Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process
2. ISO 15223-1 Fourth Edition 2021-07 Medical Devices - Symbols To Be Used With Medical Device Labels, Labelling, And Information To Be Supplied - Part 1: General Requirements
3. CLSI EP07, Third Edition, Interference Testing in Clinical Chemistry
4. CLSI EP12-A2, User Protocol for Evaluation of Qualitative Test Performance
5. CLSI EP17-A2, Evaluation of Detection Capability for Clinical Laboratory Measurement Procedures
6. CLSI EP18-A2, Risk Management Techniques to Identify and Control Laboratory Error Sources
7. CLSI EP25-A, Evaluation of Stability of In Vitro Diagnostic Reagents
VII Performance Characteristics (if/when applicable):
A Analytical Performance:
1. Precision/Reproducibility:
A reproducibility study was conducted across three sites with two operators at each site. Spiked whole blood sample panels for the precision and reproducibility studies were prepared using the pre-determined FebriDx C5 and C95 concentrations. A description of each panel member is included in Table 1 below.
Table 1. Reproducibility Study Panel Members
| Panel Member | C-Reactive Protein Concentration | Myxovirus Resistance Protein A Concentration |
| --- | --- | --- |
| P1 | C5 | C95 |
| P2 | C95 | C5 |
| P3 | C95 | C95 |
| P4 | C95 | 120 ng/mL |
| P5 | 150 μg/mL | C95 |
| P6 | 0 | 0 |
A total of 540 determinations were performed by untrained operators at the point of care settings across three different sites over five contiguous days during a two-week period. The study demonstrates overall reproducibility among three lots of material, among three separate sites, and among six separate users. On each day of testing, the Positive and Negative External Controls were run by one individual to ensure the appropriate results were produced. All External Controls tests produced the expected results. Reproducibility study results are summarized below.
K230917 - Page 5 of 16
{6}
Table 2. Reproducibility Study Results for Each Analyte Band – By Site
| Sample | % Agreement with Expected Results | | | | | | | | | |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| | Site 1 | | Site 2 | | Site 3 | | Overall | | | |
| | MxA | CRP | MxA | CRP | MxA | CRP | MxA | 95% CI | CRP | 95% CI |
| P1
MxA C95
CRP C5 | 100%
(30/30) | 80%
(24/30) | 70%
(21/30) | 90%
(27/30) | 93.3%
(28/30) | 83.3%
(25/30) | 87.8%
(79/90) | 79.4-93 | 84.4%
(76/90) | 75.6-90.5 |
| P2
MxA C5
CRP C95 | 100%
(30/30) | 100%
(30/30) | 90%
(27/30) | 100%
(30/30) | 73.3%
(22/30) | 100%
(30/30) | 87.8%
(79/90) | 79.4-93 | 100%
(90/90) | 95.1-100 |
| P3
MxA C95
CRP C95 | 100%
(30/30) | 100%
(30/30) | 90%
(27/30) | 100%
(30/30) | 100%
(30/30) | 100%
(30/30) | 96.7%
(87/90) | 90.7-98.9 | 100%
(90/90) | 95.1-100 |
| P4
MxA High
CRP C95 | 100%
(30/30) | 100%
(30/30) | 100%
(30/30) | 100%
(30/30) | 100%
(30/30) | 100%
(30/30) | 100%
(90/90) | 95.1-100 | 100%
(90/90) | 95.1-100 |
| P5
MxA C95
CRP High | 100%
(30/30) | 100%
(30/30) | 90%
(27/30) | 100%
(30/30) | 93.3%
(28/30) | 100%
(30/30) | 94.4%
(85/90) | 87.7-97.6 | 100%
(90/90) | 95.1-100 |
| P6
MxA Neg
CRP Neg | 100%
(30/30) | 100%
(30/30) | 90%
(27/30) | 100%
(30/30) | 100%
(30/30) | 100%
(30/30) | 100%
(87/90) | 90.7-98.9 | 100%
(90/90) | 95.1-100 |
Lower than expected MxA reactivity was observed for the P1 panel member at site 2. Further analysis of the data indicated that the majority of the discrepant results observed in the reproducibility study were obtained from just one user at site 2 and another user at site 3. Users should ensure they follow the assay instructions to avoid potential erroneous results.
The same results are presented below when interpreting individual cartridge results for each replicate against the expected band reactivities.
Table 3. Reproducibility Study Results Final Interpretation – By Site
| Sample | % Agreement with Expected Results | | | | |
| --- | --- | --- | --- | --- | --- |
| | Site 1 | Site 2 | Site 3 | Overall | 95% CI |
| P1 | 80%
(24/30) | 70%
(21/30) | 76.7%
(23/30) | 75.6% (68/90) | 65.8-83.3 |
| P2 | 100%
(30/30) | 90%
(27/30) | 73.3%
(22/30) | 87.8% (79/90) | 79.4-93.0 |
| P3 | 100%
(30/30) | 90%
(27/30) | 100%
(30/30) | 96.7% (87/90) | 90.7-98.9 |
| P4 | 100%
(30/30) | 100%
(30/30) | 100%
(30/30) | 100% (90/90) | 95.9-100 |
| P5 | 100%
(30/30) | 90%
(27/30) | 93.3%
(28/30) | 94.4%
(85/90) | 87.7-97.6 |
| P6 | 100%
(30/30) | 90%
(27/30) | 100%
(30/30) | 96.7%
(87/90) | 90.7-98.9 |
K230917 - Page 6 of 16
{7}
Additional analyses evaluated potential variability between different FebriDx cartridge lots. The results are summarized in Table 4 below.
Table 4. Reproducibility Study Results Final Interpretation - By Lot
| Sample | % Agreement with Expected Results | | |
| --- | --- | --- | --- |
| | Lot 1 | Lot 2 | Lot 3 |
| P1 | 75% (27/36) | 80.5% (29/36) | 66.7% (12/18) |
| P2 | 83.3% (30/36) | 91.7% (33/36) | 88.9% (16/18) |
| P3 | 100% (36/36) | 100% (36/36) | 83.3% (15/18) |
| P4 | 100% (36/36) | 100% (36/36) | 100% (18/18) |
| P5 | 91.7% (33/36) | 94.4% (34/36) | 100% (18/18) |
| P6 | 100% (36/36) | 100% (36/36) | 83.3% (15/18) |
The reproducibility study results are acceptable and support use of the device in appropriate clinical settings with users trained in the use of the FebriDx test.
# 2. Linearity:
Not applicable
# 3. Analytical Specificity/Interference:
Interfering Substance Study
Two sets of spiked whole blood samples were prepared that separately contained either C95 concentrations of MxA and CRP or C5 concentrations of MxA and CRP. Sample panels were tested by a trained operator prior to using them in the study to confirm the expected reactivity. Both sets of samples were further spiked with the interferents listed in Table 5 below and evaluated on the FebriDx device by three operators in triplicate to assess any potential interference effect.
Table 5. Interference Testing Results
| Test Substance | Concentration | Interference? |
| --- | --- | --- |
| Acetaminophen | 15.6 mg/dL | No |
| Acetylsalicylic acid | 3 mg/dL | No |
| Alcohol | 789 mg/dL | No |
| Azithromycin | 1.11 mg/dL | No |
| Bilirubin (Conjugated) | 40 mg/dL | No |
| Bilirubin (Unconjugated) | 0.4 mg/dL | No |
| Biotin | 3500 ng/mL | No |
| Caffeine | 10 mg/dL | No |
K230917 - Page 7 of 16
{8}
K230917 - Page 8 of 16
| Celecoxib | 0.879 mg/dL | No |
| --- | --- | --- |
| Cetirizine HCL | 0.435 mg/dL | No |
| Dextromethorphan | 1.56 ug/dL | No |
| Doxycycline | 1.8 mg/dL | No |
| Furosemide | 1.59 mg/dL | No |
| HAMA | 8000 ng/mL | No |
| Hemoglobin | 1000 mg/dL | No |
| Ibuprofen | 21.9 mg/dL | No |
| Imipenem | 18 mg/dL | No |
| Levofloxacin | 3.6 mg/dL | No |
| Loratadine | 0.5 mg/dL | No |
| Nicotine | 0.097 mg/dL | No |
| Oxymetazoline HCL | 0.09 mg/dL | No |
| Phenylephrine | 0.003 mg/dL | No |
| Prednisolone | 0.120 mg/dL | No |
| Protein (total) | 9 g/dL | No |
| Rheumatoid Factor | 1000 IU/mL | Yes |
| Salmeterol | 6.03 μg/dL | No |
| Tiotropium | 4.80 ng/dL | No |
| Triglycerides | 1500 mg/dL | No |
| Vancomycin | 12 mg/dL | No |
Initial testing with Rheumatoid Factor at 1000 IU/mL demonstrated interference with CRP band interpretation. Only 5/9 replicates exhibited expected results for samples that contained either C5 or C95 concentrations of CRP in the presence of Rheumatoid Factor. This interferent was diluted and interference testing was repeated until no interference effect was observed. Potential interference with MxA results was also observed at Rheumatoid Factor concentrations below 1000 IU/mL. No interference for either analyte was observed at concentrations below 50 IU/mL.
## Hook Effect Study
To assess whether a hook effect exists for the FebriDx test device, samples containing MxA and CRP at the concentrations described in the table below were evaluated and individual band reactivity for both MxA and CRP was assessed.
Table 6. Hook Effect Study Samples
| Panel Member | MxA (ng/mL) | CRP (μg/mL) |
| --- | --- | --- |
| 1 | 700 | 1000 |
| 2 | 600 | 800 |
| 3 | 500 | 600 |
| 4 | 400 | 500 |
| 5 | 200 | 250 |
| 6 | 100 | 125 |
| 7 | 50 | 62.5 |
| 8 | 0 | 0 |
{9}
All evaluated specimens were reactive for both MxA and CRP. No hook effect was identified for CRP concentrations up to 1000 µg/mL and MxA concentrations up to 700 ng/mL.
4. Assay Reportable Range:
Not applicable.
5. Traceability, Stability, Expected Values (Controls, Calibrators, or Methods):
Controls
The FebriDx test has built-in procedural controls represented by a blue control line.
External controls should be used to demonstrate that the reagents and assay perform properly. External controls (one positive and one negative) are not supplied with the FebriDx test kit (i.e. box of 25 devices) but are available separately from Lumos Diagnostics. External controls should be used, as consistent with good laboratory practices, to verify test performance. The negative external control is composed of buffered solution containing a preservative. The positive external control is composed of buffered stabilizing solution containing preservative and recombinant MxA and CRP proteins near the LoD for each analyte (55.6 ng/mL and 22.9 ug/mL for MxA and CRP, respectively).
Reagent Stability
Protocols for evaluation of real-time stability, transport stability, and in-use stability of the FebriDx test cartridge as well as real-time stability of the external control materials were reviewed and found to be acceptable.
6. Detection Limit:
The limit of detection (LoD) for the individual CRP and MxA bands on the FebriDx test were determined on two lots of FebriDx assay cartridges with two (2) operators. A series of whole blood samples spiked with MxA and CRP analytes at concentrations spanning the assay range were blinded to test operators, tested, and read to determine the LoD concentration (C95) and a high negative concentration (C5). Results for each reagent lot are described separately below in Tables 7-10. In accordance with CLSI EP17-A2, the highest observed value from each individual lot was chosen as the corresponding C5 or C95 value.
Table 7. LoD Study results Lot 1 - MxA
| Sample | [MxA] (ng/mL) | % Positive (n/N) |
| --- | --- | --- |
| 1 | 0 | 0% (0/20) |
| 2 | 10 | 0% (0/20) |
| 3 | 15 | 0% (0/20) |
| 4 | 20 | 15% (3/20) |
| 5 | 25 | 50% (0/20) |
K230917 - Page 9 of 16
{10}
K230917 - Page 10 of 16
| 6 | 30 | 75% (15/20) |
| --- | --- | --- |
| 7 | 35 | 90 (18/20) |
| 8 | 40 | 95% (19/20) |
| 9 | 45 | 100% (20/20) |
| 10 | 60 | 100% (20/20) |
Probit analysis identified C5, C50, and C95 MxA concentrations of 18.71, 26.68, and 38.05, respectively for Lot 1.
Table 8. LoD Study Results for Lot 1 - CRP
| Sample | [CRP] (μg/mL) | % Positive (n/N) |
| --- | --- | --- |
| 1 | 0 | 0% (0/20) |
| 2 | 7 | 0% (0/20) |
| 3 | 9 | 30% (6/20) |
| 4 | 11 | 40% (8/20) |
| 5 | 13 | 50% (10/20) |
| 6 | 15 | 70% (14/20) |
| 7 | 17 | 100% (20/20) |
| 8 | 19 | 100% (20/20) |
| 9 | 25 | 100% (20/20) |
| 10 | 40 | 100% (20/20) |
Probit analysis identified C5, C50, and C95 CRP concentrations of 8.53, 11.89, and 16.56, respectively for Lot 1.
Table 9. LoD Study results Lot 2 - MxA
| Sample | [MxA] (ng/mL) | % Positive (n/N) |
| --- | --- | --- |
| 1 | 0 | 0% (0/20) |
| 2 | 10 | 0% (0/20) |
| 3 | 15 | 0% (0/20) |
| 4 | 20 | 20% (4/20) |
| 5 | 25 | 55% (11/20) |
| 6 | 30 | 60% (12/20) |
| 7 | 35 | 70% (14/20) |
| 8 | 40 | 100% (20/20) |
| 9 | 45 | 100% (20/20) |
| 10 | 60 | 100% (20/20) |
Probit analysis identified C5, C50, and C95 MxA concentrations of 18.78, 26.75, and 38.10, respectively.
{11}
Table 10. LoD Study Results for Lot 2 - CRP
| Sample | [CRP] (μg/mL) | % Positive (n/N) |
| --- | --- | --- |
| 1 | 0 | 0% (0/20) |
| 2 | 7 | 0% (0/20) |
| 3 | 9 | 10% (2/20) |
| 4 | 11 | 45% (9/20) |
| 5 | 13 | 75% (15/20) |
| 6 | 15 | 85% (17/20) |
| 7 | 17 | 100% (20/20) |
| 8 | 19 | 100% (20/20) |
| 9 | 25 | 100% (20/20) |
| 10 | 40 | 100% (20/20) |
Probit analysis identified C5, C50, and C95 CRP concentrations 8.72, 11.64, and 15.54, respectively.
Based on the results of the LoD testing across each test cartridge lot, the overall LoD for MxA was established as 40 ng/mL and the LoD for CRP was established as 16.6 μg/mL.
7. Assay Cut-Off:
Not applicable.
B Comparison Studies:
1. Method Comparison with Predicate Device:
Not applicable.
2. Matrix Comparison:
The FebriDx test is intended to be utilized with freshly collected fingerstick specimens. Due to difficulties obtaining sufficient quantities of this sample matrix, venous whole blood was utilized for the analytical studies described above. To support the use of venous whole blood as a surrogate matrix for fingerstick blood in analytical studies, a study was conducted to demonstrate equivalent LoD for both analytes across both sample matrices of the FebriDx device. Specifically, a matched series of samples was prepared in both venous and fingerstick blood by spiking MxA and CRP to various concentrations and then testing them across 20 replicates of each dilution. The results of the samples for each matrix were fit to a 4-parameter logistic curve and this curve was used to determine the C5 and C95 values for the FebriDx test using either venous blood or fingerstick blood as the sample matrix. Additional confirmatory testing was performed with another 20 replicates at the C5 and C95 concentrations to confirm the values. No significant differences in LoD were observed between fingerstick and venous sample matrices. This data supported the use of venous whole blood matrix in analytical studies and is not intended to support a venous whole blood claim
K230917 - Page 11 of 16
{12}
# C Clinical Studies:
# 1. Clinical Sensitivity:
A prospective multi-center blinded clinical trial (DISRUPT, ClinicalTrials.gov identifier: NCT02018198) was conducted in the United States (U.S.) to evaluate the clinical performance of FebriDx. Patients presenting to a primary care provider, urgent care clinic or Emergency Department with signs/symptoms of acute respiratory infection (ARI) and a recent fever were screened for eligibility. Separately, additional healthy subjects (the asymptomatic cohort) were enrolled in an all corner fashion across the same study sites to evaluate the reference range of the FebriDx device. FebriDx testing for all enrolled patients was conducted by users at the point of care.
A patient's final diagnosis (i.e., clinical truth) was based on results of a clinical reference algorithm in conjunction with adjudication by two independent expert reviewers who had the ability to accept or override the results of the clinical reference algorithm based on the review of available data for each participant. The clinical experts had access to clinical data as well as microbiological results and laboratory tests. Additionally, standard of care tests ordered for the evaluation of acute respiratory infection by treating clinicians were captured, recorded, and made available to the reviewers.
The demographic characteristics of the DISRUPT study populations are described in Table 11 below.
Table 11. Demographic Information of DISRUPT Study Cohorts
| Characteristics | Acute Respiratory Infection Cohort Overall % (Proportion) | Asymptomatic Cohort Overall % (Proportion) |
| --- | --- | --- |
| Sex | | |
| Male | 44.2% (230/520) | 44.7% (76/170) |
| Female | 55.8% (290/520) | 55.3% (94/170) |
| Age | | |
| Median (IQR) | 32.0 (23.0, 48.0) | 38.0 (19.0, 69.0) |
| 1-21 years | 20.8% (108/520) | 31.2% (53/170) |
| 22- 64 years | 71.7% (373/520) | 31.8% (54/170) |
| 65+ | 7.5% (39/520) | 37.1% (63/170) |
| Race | | |
| American Indian | 0.4% (2/520) | 0 |
| Asian | 2.7% (14/520) | 1.2% (2/170) |
| Black | 21.2% (110/520) | 17.1% (29/170) |
| Other | 7.1% (37/520) | 2.4% (4/170) |
| Pacific Islander | 0.2% (1/520) | 0 |
| Unknown | 0 | 0 |
| White | 68.5% (356/520) | 79.4% (135/170) |
| Ethnicity | | |
| Hispanic | 18.7% (97/520) | 12.4% (21/170) |
| Other | 1.3% (6/520) | 0 |
K230917 - Page 12 of 16
{13}
Of the total enrolled participants in the ARI cohort, $14.0\%$ (73/520) were classified as bacterial, and $81.3\%$ (423/520) were considered as a non-bacterial etiology (e.g., either non-infectious or a viral infection) by the comparator clinical reference algorithm that incorporated pathogen detection testing as well as measures of host immune response in conjunction with clinical adjudication by two independent reviewers who were blinded to FebriDx device results. In the event that the two clinical medical experts disagreed, one additional independent expert served as the arbitrator between the two conflicting opinions and a majority opinion, determined by independent review, served to assign a final clinical diagnosis.
A final diagnosis could not be determined by the clinical reference algorithm/adjudication due to insufficient laboratory evidence (e.g., samples lost in transit, thawed in transit, not labelled correctly and thus not tested) for $4.6\%$ (24/520) participants. These subjects were excluded from the performance analysis. The expert panel agreed with classification assigned by the algorithm in $93.3\%$ (463/496) of the adjudicated cases.
Results from the DISRUPT study are summarized in Table 12 and Table 13 below.
Table 12. DISRUPT Study FebriDx Results versus Clinical Comparator
| FebriDx Result | Adjudicated Clinical Reference Algorithm Result | | | |
| --- | --- | --- | --- | --- |
| | Bacterial | Non-bacterial | Unknown | Total |
| Bacterial | 68 | 49 | 7 | 124 |
| Non-bacterial | 5 | 374 | 17 | 396 |
| Invalid | 0 | 0 | 0 | 0 |
| Total | 73 | 423 | 24 | 520 |
K230917 - Page 13 of 16
{14}
Table 13. FebriDx Performance Summary - DISRUPT Study
| | Bacterial Infection |
| --- | --- |
| PPA [95% CI] | 93.2% (68/73) [84.9-97.0%] |
| NPA [95% CI] | 88.4% (374/423) [85.0-91.1%] |
| PPV [95% CI] | 58.1% (68/117) [49.1-66.7%] |
| NPV [95% CI] | 98.7% (374/379) [96.9-99.4%] |
Positive and negative controls were performed every day of enrollment, with every new lot or shipment and with every first-time test-operator. Prior to performing testing on study participants, $100\%$ of positive and negative controls were read as positive and negative, respectively. A small percentage of controls $(0.8\%, 10/1205)$ required repeat testing due to operator error (insufficient buffer delivered caused by weak pressure when pushing the buffer release button and invalid readings).
The DISRUPT study was not adequately powered to evaluate differences in performance among different age groups or between specific races or ethnicities. Results from the FebriDx test should be interpreted with other clinical and diagnostic findings and the presence of a bacterial respiratory infection should not be determined based on FebriDx test results alone.
Table 14. FebriDx Test Performance - DISRUPT Study, Stratified by Race
| Bacterial Infection Performance | Patient Race | | |
| --- | --- | --- | --- |
| | Black | White | Other |
| PPA [95% CI] | 75.0% (6/8) [40.9-92.9%] | 94.8% (55/58) [85.9-98.2%] | 100% (7/7) (64.6-100%) |
| NPA [95% CI] | 89.0% (89/100) [81.4-93.7%] | 87.7% (242/276) [83.3-91.1%] | 91.5% (43/47) [80.1-96.6%] |
Additionally, clinical performance of the FebriDx device has not been established in pediatric patients younger than 12 years old or in elderly patients above 64 years old. The FebriDx test should not be used in these patient populations. These populations may be at increased risk of severe complications from undiagnosed respiratory infections, and differences in the immune response among these age groups could contribute to differences in assay performance.
FebriDx test performance in the DISRUPT study stratified by patient age is summarized in Table 15 below.
K230917 - Page 14 of 16
{15}
Table 15. FebriDx Test Performance – DISRUPT Study, Stratified By Patient Age
| Bacterial Infection Performance | Patient Age Range | | |
| --- | --- | --- | --- |
| | 1-21 years old | 22-64 years old | 65+ years old |
| PPA [95% CI] | 100.0% (13/13)
[77.2-100.0%] | 92.3% (48/52)
[81.8-97.0%] | 87.5% (7/8)
[52.9-97.8%] |
| NPA [95% CI] | 95.7% (88/92)
[89.3-98.3%] | 86.7% (260/300)
[82.4-90.1%] | 83.9% (26/31)
[67.4-92.9%] |
Cumulatively, the data from the DISRUPT study support device performance in patients aged 12-64 years old and a determination of substantial equivalence.
2. Clinical Specificity:
See Clinical Sensitivity section C.1 above.
3. Other Clinical Supportive Data (When 1. and 2. Are Not Applicable):
Not applicable.
D Clinical Cut-Off:
Not applicable.
E Expected Values/Reference Range:
Expected Values
The results from a prospectively enrolled study cohort of 520 individuals with signs and symptoms of Acute Respiratory infection are summarized in Table 16 below.
Table 16. Expected Values of FebriDx Test in DISRUPT study
| FebriDx Result | Adjudicated Clinical Reference Algorithm Result | | | |
| --- | --- | --- | --- | --- |
| | Bacterial | Non-bacterial | Unknown | Total |
| Bacterial | 68 | 49 | 7 | 124 |
| Non-bacterial | 5 | 374 | 17 | 396 |
| Invalid | 0 | 0 | 0 | 0 |
| Total | 73 | 423 | 24 | 520 |
The predictive value of FebriDx test results will depend upon the prevalence of bacterial infections in the population. Users should evaluate FebriDx test results in conjunction with other clinical and diagnostic findings.
K230917 - Page 15 of 16
{16}
Reference Range Study
The clinical study included evaluation of 170 asymptomatic patients who were not suspected of having an infection. The cohort included 44.7% male and 55.3% female participants with a diverse ethnic background. The mean age of participants was 43.5 years and ranged from 3 to 87 years of age. The study enrolled participants in each of three age groups; 31.2% (53/170) of participants were children/adolescents (1 - 21 years); 31.8% (54/170) were adults (22-64 years); and 37.1% (63/170) were elderly (65+ years). Participants presented without signs/symptoms of acute respiratory infection at the time of enrollment. Individual asymptomatic individuals were subjected to FebriDx testing and the same clinical comparator methods described above. Results are summarized in Table 17 below.
Table 17. FebriDx Performance in the Asymptomatic Cohort – DISRUPT Study
| FebriDx Result | Adjudicated Clinical Reference Algorithm Result | | | |
| --- | --- | --- | --- | --- |
| | Bacterial | Non-bacterial | Unknown | Total |
| Bacterial | 1 | 3 | 0 | 4 |
| Non-bacterial | 0 | 157 | 9 | 166 |
| Invalid | 0 | 0 | 0 | 0 |
| Total | 1 | 160 | 9 | 170 |
Clinical Truth could not be determined by the clinical reference algorithm/adjudication due to insufficient laboratory evidence (e.g., samples lost in transit, thawed in transit, not labelled correctly and thus not tested) for 5% (9/170) participants.
VIII Proposed Labeling:
The labeling supports the finding of substantial equivalence for this device.
IX Conclusion:
The submitted information in this premarket notification is complete and supports a substantial equivalence decision.
K230917 - Page 16 of 16
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.