← Product Code [FED](/productcode/FED) · K230801

# Pathfinder Endoscope Overtube with Balloon Device (K230801)

_Neptune Medical, Inc. · FED · Jul 28, 2023 · Gastroenterology, Urology · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K230801

## Device Facts

- **Applicant:** Neptune Medical, Inc.
- **Product Code:** [FED](/productcode/FED.md)
- **Decision Date:** Jul 28, 2023
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 876.1500
- **Device Class:** Class 2
- **Review Panel:** Gastroenterology, Urology

## Indications for Use

The Pathfinder Endoscope Overtube with Balloon is intended to be used with an endoscope to facilitate intubation, change of endoscopes, and treatment in the gastrointestinal (GI) tract in adult patients (22 years of age and older).

## Device Story

Hollow overtube with distal balloon; used with flexible GI endoscope. Overtube provides stabilization via balloon contact with lumen walls. Vacuum-assisted rigidization mechanism allows device to transition between flexible and rigid states; rigid state maintains shape for endoscope advancement/withdrawal. Irrigation line provides lubrication; vacuum line connects to external source for rigidization; balloon line connects to external controller for inflation/deflation. Used in endoscopy suite by physicians. Output is physical stabilization of endoscope within GI tract; facilitates navigation and treatment. Benefits include improved endoscope positioning and reduced anatomical disturbance.

## Clinical Evidence

No clinical data. Substantial equivalence supported by bench testing, including simulated use, lubricity, insufflation, insertion/removal, navigation, rigidization/de-rigidization, endoscope compatibility, balloon reliability, balloon pressure stability, balloon inflation/deflation, and balloon dimensions. Biocompatibility testing performed per ISO 10993-1.

## Technological Characteristics

Vacuum-assisted rigidizing overtube with distal balloon. Materials: biocompatible polymers (ISO 10993-1 compliant). Dimensions: 85cm and 110cm lengths, 14mm nominal inner diameter, 40mm nominal balloon outer diameter. Energy: external vacuum source and external balloon controller. Connectivity: Luer lock fittings for irrigation and inflation. Sterilization: Ethylene Oxide (EO). Single-use.

## Regulatory Identification

An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.

## Special Controls

*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.

In combination with the general controls of the FD&C Act, the integrated operating table-electromechanical surgical system is subject to the following special controls:

1. (1) Premarket clinical performance testing, or a combination of premarket clinical performance testing and postmarket surveillance (in accordance with special control (2)), must include the following:
   1. (i) Objective performance measures (e.g., rate and number of conversions to other surgical modalities, rate of device related adverse events (including tissue injury, hematoma, and increased blood loss), and their severity, cause, and outcomes) must be reported with relevant descriptive comparator performance measures.
   2. (ii) The data must demonstrate the performance of the device for providing accurate and precise control of attached surgical instruments in range of clinical conditions relevant to the device's intended use.
   3. (iii) The test dataset must include data collected from a patient population representative of the intended patient population under anticipated conditions of use.
2. (2) Data obtained from postmarket surveillance must demonstrate, in consideration of the premarket data obtained in accordance with special control (1), that the device performs in accordance with special control (1), unless FDA determines, based on the totality of the premarket data, that data from postmarket surveillance is not required to demonstrate that the device performs as intended. Such postmarket surveillance must be conducted per a protocol determined appropriate by FDA to demonstrate that the device performs as intended (in consideration of the premarket data obtained in accordance with special control (1)), and must include initiation, enrollment, and reporting requirements to ensure timely periodic updates to FDA on post-market surveillance progress and outcomes.
3. (3) Animal performance testing must evaluate the extent of port site trauma due to repositioning of table during surgical procedures when utilizing robotic minimally invasive and laparoscopic approaches
4. (4) The device manufacturer must develop, and update as necessary, a device-specific use training program that ensures proper device setup/use/shutdown, accurate control of instruments to perform the intended surgical procedures, troubleshooting and handling during unexpected events or emergencies, and safe practices to mitigate use error.
5. (5) The device manufacturer may only distribute the device to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program.
(6) Human factors assessment must demonstrate that the user can correctly use the device system across all intended use environments with the provided instructions and training materials, including patient access during normal operating conditions and emergency situations, and effects arising from the integrated nature of the operating table and robotic surgical arms.
(7) Labeling must include:
(i) A detailed summary of clinical performance testing conducted with the device, including study population, results, adverse events, and comparisons to any comparator groups identified;
(ii) A statement in the labeling that the safety and effectiveness for the representative specific procedures was based on evaluation of the device as a surgical tool and did not include evaluation of outcomes related to the treatment of the patient's underlying disease or condition, unless FDA determines that it can be removed or modified based on clinical performance data submitted to FDA;
(iii) Identification of compatible devices;
(iv) The list of surgical procedures for which the device has been determined to be safe with clinical justification;
(v) Reprocessing instructions for reusable components;
(vi) A shelf life for any sterile components;
(vii) A description of the device-specific use training program;
(viii) A statement that the device is only for distribution to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program; and
(ix) A summary of any completed postmarket surveillance data collected as required by special control (2), including updated labeling to accurately reflect outcomes observed in postmarket surveillance.
(8) Non-clinical performance testing must demonstrate that the device performs as intended under anticipated conditions of use and must include:
(i) Device motion accuracy and repeatability;
(ii) System testing;
(iii) Instrument reliability;
(iv) Crosstalk;
(v) Table motion control;
(vi) Thermal effects on tissue;
(vii) User-device interface performance;
(viii) Workspace access testing; and
(ix) Performance testing with compatible devices.
(9) Software verification, validation, and hazard analysis must be performed.
(10) Electromagnetic compatibility and electrical, thermal, and mechanical safety testing must be performed.
(11) Performance data must demonstrate the sterility of all patient-contacting device components.
(12) Performance data must support the shelf life of the device components provided sterile by demonstrating continued sterility and package integrity over the labeled shelf life.
(13) Performance data must validate the reprocessing instructions for the reusable components of the device.
(14) Performance data must demonstrate that all patient-contacting components of the device are biocompatible.
(15) Performance data must demonstrate that all patient-contacting components of the device are non-pyrogenic.
(16) The device manufacturer must submit a report to the FDA annually on the anniversary of initial marketing authorization for the device, until such time as FDA may terminate such reporting, which comprises the following information:
(i) Cumulative summary, by year, of complaints and adverse events since date of initial marketing authorization; and
(ii) Identification and rationale for changes made to the device, labeling, or device specific use training program, which did not require submission of a premarket notification during the reporting period.

## Predicate Devices

- Pathfinder Endoscope Overtube ([K211301](/device/K211301.md))

## Reference Devices

- Dilumen Endolumenal Interventional Platform ([K210851](/device/K210851.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logos of the U.S. Department of Health & Human Services and the U.S. Food & Drug Administration (FDA). The Department of Health & Human Services logo is on the left and features a stylized human figure. The FDA logo is on the right and includes the FDA acronym in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.

July 28, 2023

Neptune Medical, Inc. % Eric Bannon Regulatory Consultant Alvamed. Inc. 935 Great Plain Ave, Suite 166 Needham, MA 02492

Re: K230801

Trade/Device Name: Pathfinder Endoscope Overtube with Balloon Device Regulation Number: 21 CFR 876.1500 Regulation Name: Endoscope And Accessories Regulatory Class: Class II Product Code: FED Dated: June 28, 2023 Received: June 29, 2023

Dear Eric Bannon:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's

{1}------------------------------------------------

requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

# Shanil P. Haugen -S

Shanil P. Haugen, Ph.D. Assistant Director DHT3A: Division of Renal, Gastrointestinal, Obesity and Transplant Devices OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

## Indications for Use

510(k) Number (if known)

K230801

Device Name

Pathfinder Endoscope Overtube with Balloon

Indications for Use (Describe)

The Pathfinder Endoscope Overtube with Balloon is intended to be used with an endoscope to facilitate intubation, change of endoscopes, and treatment in the gastrointestinal (GI) tract in adult patients (22 years of age and older).

| Type of Use (Select one or both, as applicable) |  |
|-------------------------------------------------|--|
|-------------------------------------------------|--|

X Prescription Use (Part 21 CFR 801 Subpart D)

| Over-The-Counter Use (21 CFR 801 Subpart C)

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

1-1

## 1.0 510(K) SUMMARY

1.1 Name and Address of Owner/Submitter

Neptune Medical, Inc. 1828 El Camino Real, Suite 508 Burlingame, CA 94010 USA Phone: +1 (650) 307-2176 Alex Tilson, CEO alex@neptunemedical.com

1.2 Correspondent/Primary Contact Person

Eric Bannon Vice President of Regulatory and Clinical Affairs AlvaMed, Inc. ebannon@alvamed.com Phone: +1 (781) 710-8243 Fax: +1 (617) 249-0955

### 1.3 Submission Information

| Date Summary Prepared:      | March 22, 2023                                        |
|-----------------------------|-------------------------------------------------------|
| Name of Device:             | Pathfinder Endoscope Overtube with Balloon            |
| Device Classification Name: | Endoscope and Accessories                             |
| Common or Usual Name:       | Endoscope Accessory Overtube                          |
| Classification:             | Class II                                              |
| Product Code:               | FED (21 CFR 876.1500)                                 |
| Predicate Device:           | Pathfinder Endoscope Overtube (K211301)               |
| Reference Device:           | Dilumen Endolumenal Interventional Platform (K210851) |

1.4 Predicate/Reference Device

|                     | 510(k) Number | Code | Device Name                                          | Submitter Name           |
|---------------------|---------------|------|------------------------------------------------------|--------------------------|
| Predicate<br>Device | K211301       | FED  | Pathfinder<br>Endoscope Overtube                     | Neptune Medical,<br>Inc. |
| Reference<br>Device | K210851       | FDF  | Dilumen<br>Endolumenal<br>Interventional<br>Platform | Lumendi                  |

{4}------------------------------------------------

Image /page/4/Picture/1 description: The image contains the logo for Neptune Medical. The word "Neptune" is written in blue, and the word "Medical" is written in red below it. To the right of the text is a circular graphic that depicts a planet with rings in space.

Traditional 510(k): Pathfinder Endoscope Page: Overtube with Balloon Device

### 1.5 Device Description

The Pathfinder Endoscope Overtube with Balloon device consists of a hollow tube with a balloon at the distal end for use over a flexible gastrointestinal endoscope. The purpose of the balloon is to contact the lumen walls to provide stabilization to the distal end of the overtube within the GI tract. The endoscope is inserted through the proximal end of the device and comes out the distal end. The free space between the overtube and the endoscope is lubricated with water through the irrigation line by connecting to the irrigation/water Luer (female Luer lock fitting) and injecting water. The vacuum line is connected to free space within the device and is completely contained. A source of vacuum must be connected to the barb fitting per the diagram below for the overtube to transition between the flexible and rigid conditions. The balloon line is connected to an extrusion within the device that creates an air pathway from handle to the balloon. An inflation device must be connected to the Luer (male Luer lock fitting) to inflate and/or deflate the balloon. The handle rotator has two positions. The first position connects the device to atmosphere (vent) to stay in the flexible condition. The second position connects the device to a source of vacuum to transition to the rigid condition. When transitioned to the rigid condition, the device maintains its shape at the time of rigidization, allowing the endoscope to advance or withdraw relative to the overtube with minimal disturbance to surrounding anatomy. Balloon inflation and deflation are actuated and controlled by an existing balloon controller unit within the endoscopy suite. The device must be connected to the balloon controller to transition into the inflated and deflated condition.

#### 1.6 Comparison of Subject Predicate, and Reference Device

|                                | Subject Device:                                                                                                                                                                                                                                                    | Predicate Device:                                                                                                                                                                                                                                 | Reference Device:                                                                                                                                                                                                                                                                                                                                                                                                                                                         |
|--------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
|                                | Pathfinder Endoscope<br>Overtube with Balloon                                                                                                                                                                                                                      | Pathfinder Endoscope<br>Overtube                                                                                                                                                                                                                  | Dilumen Endoluminal<br>Interventional Platform                                                                                                                                                                                                                                                                                                                                                                                                                            |
| 510(k)<br>Number               | TBD                                                                                                                                                                                                                                                                | K211301                                                                                                                                                                                                                                           | K210851                                                                                                                                                                                                                                                                                                                                                                                                                                                                   |
| 510(k)<br>Submitter/<br>Holder | Same as predicate                                                                                                                                                                                                                                                  | Neptune Medical, Inc.                                                                                                                                                                                                                             | Lumendi                                                                                                                                                                                                                                                                                                                                                                                                                                                                   |
| Regulation<br>Code             | Same as predicate                                                                                                                                                                                                                                                  | FED                                                                                                                                                                                                                                               | FDF                                                                                                                                                                                                                                                                                                                                                                                                                                                                       |
| Regulation<br>Number           | Same as predicate/<br>reference                                                                                                                                                                                                                                    | 876.1500                                                                                                                                                                                                                                          | 876.1500                                                                                                                                                                                                                                                                                                                                                                                                                                                                  |
| Common Name                    | Same as predicate                                                                                                                                                                                                                                                  | Endoscope Access Overtube, Gastroenterology-urology                                                                                                                                                                                               | Colonoscope and accessories, flexible/rigid                                                                                                                                                                                                                                                                                                                                                                                                                               |
| Regulation Description         | Same as predicate/ reference                                                                                                                                                                                                                                       | Endoscope and accessories                                                                                                                                                                                                                         | Endoscope and accessories                                                                                                                                                                                                                                                                                                                                                                                                                                                 |
| Device Classification          | Same as predicate/ reference                                                                                                                                                                                                                                       | Class II                                                                                                                                                                                                                                          | Class II                                                                                                                                                                                                                                                                                                                                                                                                                                                                  |
| Indications<br>for Use         | The Pathfinder Endoscope<br>Overtube with Balloon is<br>intended to be used with<br>an endoscope to facilitate<br>intubation, change of<br>endoscopes, and<br>treatment in the<br>gastrointestinal (GI) tract<br>in adult patients (22 years<br>of age and older). | The Pathfinder Endoscope<br>Overtube is intended to be<br>used with an endoscope<br>to facilitate intubation,<br>change of endoscopes,<br>and treatment in the<br>gastrointestinal (GI) tract<br>in adult patients (22 years<br>of age and older) | The Lumendi DiLumen is<br>an accessory to an<br>endoscope. The DiLumen<br>dual balloon accessory is<br>intended for use with any<br>standard endoscope that<br>has a distal tip outer<br>diameter of 12.5 – 14.3<br>mm. The device is<br>indicated to ensure<br>complete positioning of an<br>endoscope during<br>navigation in the large<br>intestine, while assisting<br>with optical visualization,<br>diagnosis, tissue<br>manipulation, and<br>endoscopic treatment. |
| Intended Population            | Same as predicate                                                                                                                                                                                                                                                  | Adult patients undergoing<br>gastrointestinal endoscopy                                                                                                                                                                                           | Not Available                                                                                                                                                                                                                                                                                                                                                                                                                                                             |
| Sterilization<br>Method        | Same as predicate                                                                                                                                                                                                                                                  | Ethylene Oxide (EO)<br>Sterilization                                                                                                                                                                                                              | Non-Sterile                                                                                                                                                                                                                                                                                                                                                                                                                                                               |
| Single Use                     | Yes                                                                                                                                                                                                                                                                | Yes                                                                                                                                                                                                                                               | Yes                                                                                                                                                                                                                                                                                                                                                                                                                                                                       |
| Device<br>Technology           | Vacuum-assisted<br>rigidizing overtube with<br>balloon for endoscopic<br>procedures in the GI<br>Tract.                                                                                                                                                            | Vacuum-assisted<br>rigidizing overtube for<br>endoscopic procedures in<br>the GI Tract.                                                                                                                                                           | The DiLumen utilizes two<br>balloons to position and<br>stabilize the endoscope<br>within a patient's large<br>intestine. After the<br>DiLumen is installed over<br>the endoscope, the<br>endoscope and DiLumen<br>are navigated to the target                                                                                                                                                                                                                            |

Table 1.6: Comparison of Predicate/Reference Device to Subject Device

{5}------------------------------------------------

K230801 Page 3 of 5

Image /page/5/Picture/1 description: The image shows the logo for Neptune Medical. The word "Neptune" is written in blue, with the word "Medical" written in red underneath. To the right of the text is an image of a planet with rings around it, set against a background of stars.

#### Traditional 510(k): Pathfinder Endoscope Page: Overtube with Balloon Device

{6}------------------------------------------------

| Image: Neptune Medical Logo | Traditional 510(k): Pathfinder Endoscope<br>Overtube with Balloon Device | Page: | 1-4 |
|-----------------------------|--------------------------------------------------------------------------|-------|-----|
|-----------------------------|--------------------------------------------------------------------------|-------|-----|

|            |                                                                                                                                             |                                                                                                               | zone with the balloons<br>deflated.                                                                                      |
|------------|---------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------|
| Dimensions | Two Sizes, with total<br>length of 85cm and<br>110cm. With nominal inner<br>diameter of 14mm.<br>Nominal balloon outer<br>diameter of 40mm. | Nine sizes, with total<br>length from 65 cm to 145<br>cm. With nominal inner<br>diameter of 11mm to 16<br>mm. | Fore and aft balloon<br>outer diameter: 60mm<br>Fore inner diameter:<br>17.6mm Working length:<br>168 cm, 130 cm, 103 cm |

- 1.7 Environment of Use, and Contraindications
The intended environment of use for the Pathfinder Endoscope Overtube with Balloon Device is the Gastrointestinal tract.

The contraindications are patients with esophageal bleeding, lesion(s), and/or laceration; esophageal strictures and/or varices; laryngeal perforation; trauma to teeth, gums, and/or pharynx; aspiration pneumonia; or any other condition that may preclude endoscopy.

#### 1.8 Performance Data: Summary of Functional and Performance Testing

The following Functional and Performance Bench has been completed:

- · Simulated use testing
- · Lubricity
- Insufflation
- Insertion/Removal
- Navigation
- Rigidization/De-Rigidization
- Endoscope Compatibility
- Balloon Reliability
- · Balloon Pressure Stability
- Balloon Inflation/Deflation
- Balloon dimensions
- 1.9 Performance Data: Summary of Biocompatibility Testing

The biocompatibility evaluation for the Pathfinder Endoscope Overtube with Balloon device was conducted in accordance with the guidance document "Use of International Standard ISO 10993-1, 'Biological Evaluation of Medical Devices Part 1: Evaluation and Testing Within a Risk Management

{7}------------------------------------------------

Process," June 16, 2016, and International Standard ISO 10993-1 "Biological Evaluation of Medical Devices - Part 1: Evaluation and Testing Within a Risk Management Process," as recognized by FDA and other applicable standards. The testing included the following:

- · Cytotoxicity
- · Sensitization
- Irritation
- · Systemic Toxicity
- · Material Mediated Pyrogenicity

The Pathfinder Endoscope Overtube with Balloon Device is considered to contact breached or compromised surfaces for a duration of less than 24 hours.

- 1.10 Performance Data: Summary of Clinical Testing
No clinical testing was applicable to this submission.

- 1.11 Conclusions
Like the predicate originally cleared by the FDA, the modified Pathfinder Endoscope Overtube with Balloon Device has been shown to be safe through performance testing. Design changes made to the device do not raise any questions of safety or effectiveness. Verification and validation data support substantial equivalence of the modified Pathfinder Endoscope Overtube with Balloon Device to the legally marketed predicate device.

---

**Source:** [https://fda.innolitics.com/device/K230801](https://fda.innolitics.com/device/K230801)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
