The Auryon™ Atherectomy System and Auryon Atherectomy Catheters with aspiration are indicated for use as atherectorny devices for arterial stenoses, including in-stent restenosis (ISR), and to aspirate thrombus adjacent to stenoses in native and stented infra-inguinal arteries. The Auryon™ Atherectomy System and Auryon Atherectomy Catheters without aspiration are indicated for use in the treatment, including atherectomy, of infra-inguinal stenoses and occlusions.
Device Story
Auryon Atherectomy System consists of single-use catheter and laser console; catheter utilizes array of optic fibers surrounded by circumferential blunt blade at distal tip. System transmits laser energy (355 nm wavelength) at pre-set fluence levels (50 and 60 mJ/mm²) to occluded/narrowed artery to perform atherectomy. Operates over 300cm 0.014" guidewire. Larger catheters (2.0mm, 2.35mm) include aspiration feature for debris/thrombus removal; 2.35mm model includes off-center feature for debulking beyond catheter diameter. Used in clinical setting by physicians; provides debulking of arterial lesions. Modification adds platinum/iridium marker band to 0.9mm catheter for enhanced radiographic visualization. System benefits patients by restoring arterial patency through mechanical and laser-assisted tissue removal.
Clinical Evidence
Bench testing only. Performance testing included catheter shaft dimensions, guard tube length, trackability, leakage, pull testing, optical fiber exposure, optical functionality, torque testing, and hydrophilic coating evaluation. All tests met established acceptance criteria.
Indicated for patients with arterial stenoses, including in-stent restenosis (ISR), and thrombus in native and stented infra-inguinal arteries. Also indicated for treatment of infra-inguinal stenoses and occlusions.
Regulatory Classification
Identification
An intraluminal artery stripper is a device used to perform an endarterectomy (removal of plaque deposits from arterisclerotic arteries.)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food & Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
June 9, 2023
Eximo Medical Ltd. % James Welsh Director, Regulatory Affairs AngioDynamics 603 Oueensbury Ave Queensbury, New York 12804
Re: K230709
Trade/Device Name: Auryon Atherectomy System Regulation Number: 21 CFR 870.4875 Regulation Name: Intraluminal Artery Stripper Regulatory Class: Class II Product Code: MCW Dated: March 14, 2023 Received: March 14, 2023
Dear James Welsh:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Ariel G. Ashshakoor -S
Digitally signed by Ariel G. Ash-shakoor -S Date: 2023.06.09 16:02:44 -04'00'
For
Gregory O'Connell Assistant Director DHT2C: Division of Coronary and Peripheral Intervention Devices OHT2: Office of Cardiovascular Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K230709
Device Name Auryon™ Atherectomy System
#### Indications for Use (Describe)
The Auryon™ Atherectomy System and Auryon Atherectomy Catheters with aspiration are indicated for use as atherectorny devices for arterial stenoses, including in-stent restenosis (ISR), and to aspirate thrombus adjacent to stenoses in native and stented infra-inguinal arteries.
The Auryon™ Atherectomy System and Auryon Atherectomy Catheters without aspiration are indicated for use in the treatment, including atherectomy, of infra-inguinal stenoses and occlusions.
| Type of Use (Select one or both, as applicable) | |
|------------------------------------------------------------------------------------------------|---------------------------------------------|
| <span style="text-decoration: overline;">X</span> Prescription Use (Part 21 CFR 801 Subpart D) | Over-The-Counter Use (21 CFR 801 Subpart C) |
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Eximo Medical Ltd. Auryon Atherectomy System, Special 510(k)
## 510(k) Summary for the Eximo Medical Ltd. Auryon Atherectomy System
Date Prepared: April 13, 2023
## Sponsor
Eximo Medical Ltd Pekeris St 3 Rehovot, Israel 7670203
# Contact
Yoel Zabar Sr. Director, R&D & Site Manager Eximo Medical Ltd. Telephone: +972-8-6307639 Email: yoelz@eximomedical.com
## Subject Device
- Trade Name: Common Name: Regulation Number: Regulation Name: Regulatory Class: Product Code: Classification Panel:
#### Predicate Device
- 510(k) Reference Trade Name: Common Name: Regulation Number: Regulation Name: Regulatory Class: Product Code: Classification Panel:
Auryon Atherectomy System Peripheral Atherectomy Catheter 21CFR870.4875 Intraluminal Artery Stripper Class 2 MCW Cardiovascular Devices
#### K220116
Auryon Atherectomy Catheters Peripheral Atherectomy Catheter 21CFR870.4875 Intraluminal Artery Stripper Class 2 MCW Cardiovascular Devices
## Purpose
The purpose of this Special 510(k) is to introduce into commercial distribution a slight modification of the Auryon Atherectomy Catheters previously cleared under predicate 510(k) K220116; specifically, to modify the design of the existing 0.9 mm catheter to add a platinum/iridium marker band near the tip of the catheter for enhanced radiographic image contrast.
{4}------------------------------------------------
## Device Description
The Auryon™ Atherectomy System consists of two sub-units: 1) a single use catheter ("Auryon catheter"); and 2) a laser console. The Auryon catheter is a single use catheter that is made of an array of optic fibers surrounded by a circumferential blunt blade at its distal tip. The Auryon™ catheter is connected to the laser system via its connector and transmits energy at pre-set fluence levels of 50 and 60 ml/mm² to the occluded or narrowed artery. The Auryon™ Atherectomy System must work over a commercially available 300cm 0.014" guide wire that crosses the lesion intra-luminally. The catheters are available in four sizes (0.9mm, 1.5mm, 2.0mm and 2.35mm), with and without hydrophilic coating.
For the small size catheters (i.e., 0.9mm and 1.5mm), there is a designated lumen tube for a guidewire at the center of the inner blunt blade. The 0.9mm and 1.5mm catheters do not have an aspiration feature and have not been tested in ISR lesions.
The larger Auryon catheters (i.e., 2.0mm and 2.35mm) have an eccentric guidewire lumen, and include additional features consisting of an aspiration feature (both catheters) and an "off-center" feature 2.35mm only). The aspiration feature is intended for debris and thrombus collection and removal from the vessel during the atherectomy procedure. These devices are also indicated for treatment of In-Stent Restenosis (ISR) lesions.
The "off-center" feature is included in the 2.35 mm catheter only and is designed to facilitate debulking of lesions in blood vessels beyond the catheter's diameter.
The modification associated with this submission is to add a platinum/iridium marker band within the wall of the 0.9 mm coated and uncoated catheters adjacent to the existing stainless steel inner blade. This marker band will enhance the radiographic contrast of the device.
There are no changes to the Auryon Laser Console unit associated with this submission.
## Indications for Use/Intended Use
The Auryon™ Atherectomy System and Auryon Atherectomy Catheters with aspiration are indicated for use as atherectomy devices for arterial stenoses, including in-stent restenosis (ISR), and to aspirate thrombus adjacent to stenoses in native and stented infra-inguinal arteries.
The Auryon™ Atherectomy System and Auryon Atherectomy Catheters without aspiration are indicated for use in the treatment, including atherectomy, of infra-inguinal stenoses and occlusions.
#### Comparison of Similarities and Differences in Technological Characteristics and Performance
As detailed below, the proposed Auryon Atherectomy System is Substantially Equivalent to the predicate device Auryon Atherectomy Catheters.
| Device Comparison | Subject Device: Auryon<br>Atherectomy System | Predicate Device: Auryon<br>Atherectomy Catheters (K220116) |
|--------------------|----------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------|
| Indication for Use | The Auryon™ Atherectomy System<br>and Auryon Atherectomy Catheters<br>with aspiration are indicated for use<br>as atherectomy devices for arterial | The Auryon™ Atherectomy System<br>and Auryon Atherectomy Catheters<br>with aspiration are indicated for use<br>as atherectomy devices for arterial |
{5}------------------------------------------------
| | stenoses, including in-stent<br>restenosis (ISR), and to aspirate<br>thrombus adjacent to stenoses in<br>native and stented infra-inguinal<br>arteries.<br>The Auryon™ Atherectomy System<br>and Auryon Atherectomy Catheters<br>without aspiration are indicated for<br>use in the treatment, including<br>atherectomy, of infra-inguinal<br>stenoses and occlusions. | stenoses, including in-stent<br>restenosis (ISR), and to aspirate<br>thrombus adjacent to stenoses in<br>native and stented infra-inguinal<br>arteries.<br>The Auryon™ Atherectomy System<br>and Auryon Atherectomy Catheters<br>without aspiration are indicated for<br>use in the treatment, including<br>atherectomy, of infra-inguinal<br>stenoses and occlusions. |
|----------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Regulation Number | 21 CFR §870.4875 | 21 CFR §870.4875 |
| Regulatory Class | Class II | Class II |
| Product Code | MCW | MCW |
| Active Medium | Nd:YAG | Nd:YAG |
| Laser Wavelength | 355 nm | 355 nm |
| Laser Fluence levels | 50 and 60 mJ/mm² | 50 and 60 mJ/mm² |
| Pulse Rate | 40 Hz | 40 Hz |
| Pulse Duration | 10-25 ns | 10-25 ns |
| Maximum output | 33.5 mJ | 33.5 mJ |
| Catheter sizes | 0.9mm, 1.5 mm, 2.0mm, 2.35mm<br>(only size 0.9 impacted by change) | 0.9mm, 1.5 mm, 2.0mm, 2.35mm |
| Platinum/Iridium marker<br>band | Included in 0.9 mm catheter only | None |
| Catheter Sterilization<br>Method | Ethylene Oxide | Ethylene Oxide |
## Comparison of Performance Data
The modified 0.9 mm Auryon Atheter was tested using the same methods and acceptance criteria as was done in the predicate device 510(k). The specific tests are listed below
#### Summary of Performance Testing
Catheter shaft ID, OD, and working length Catheter guard tube length Catheter trackability in simulated anatomical shape Freedom from leakage during liquid infusion Pull testing of all joints Freedom from exposed optical fibers Optical Functionality test Catheter torque test
{6}------------------------------------------------
#### Summary of Performance Testing
Evaluation of the hydrophilic coating of the catheters
## Substantial Equivalence
Assessment of the similarities and differences of the proposed Auryon Atherectomy System and the predicate Auryon Atherectomy Catheters concludes that the devices are substantially equivalent to one another; specifically:
- . The proposed and predicate device have the identical ProCode, Regulation Number, Regulation Name, and Regulatory Class;
- The proposed and predicate device have identical Indications for Use;
- . The proposed and predicate devices incorporate the identical operating principle, mechanism of action, and are intended for the same patient populations; and,
- With the exception of the added Pt/Ir component, the proposed and predicate employ an identical overall design, materials of manufacture, performance testing, sizes, and configurations.
The sum of these evaluations and determinations lead Eximo Medical Ltd. to conclude that substantial equivalence has been demonstrated, and that the existing data and additional testing have confirmed that there are no new questions of safety or effectiveness.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.