← Product Code [FDF](/productcode/FDF) · K230588

# Aer-O-Scope Colonoscope System (K230588)

_Gi View , Ltd. · FDF · Aug 17, 2023 · Gastroenterology, Urology · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K230588

## Device Facts

- **Applicant:** Gi View , Ltd.
- **Product Code:** [FDF](/productcode/FDF.md)
- **Decision Date:** Aug 17, 2023
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 876.1500
- **Device Class:** Class 2
- **Review Panel:** Gastroenterology, Urology
- **Attributes:** Therapeutic

## Indications for Use

The Aer-O-Scope Colonoscope System is intended to provide visualization (via a video monitor) and diagnostic/ therapeutic access to the adult lower gastrointestinal tract, (including but not limited to, the anus, rectum, sigmoid colon, transverse colon, cecum and ileocecal valve) for endoscopy. The Aer-O-Scope Disposable colonoscope is a single use disposable endoscope and cannot be reprocessed.

## Device Story

Aer-O-Scope 3 Colonoscope System is a flexible, operator-controlled endoscope for lower GI tract visualization and therapeutic access. System consists of a Video Controller and a single-use, disposable colonoscope. Disposable colonoscope features an optical imaging head with CMOS sensor and LED lighting at the distal tip, providing a 200° panoramic field of view. Channels are provided for irrigation, insufflation, lens wash, therapeutic access, and suction. Operator controls deflection via angulation knobs on the handle and manages system functions via a touchpad on the Video Controller. Used in physician clinics, ambulatory surgical centers, and hospitals. Provides real-time video to a monitor for clinician assessment. Facilitates colonic intubation via hydrophilic coating and variable stiffness insertion tube. Benefits include single-use disposability, eliminating reprocessing needs, and wide-angle visualization for screening, diagnostic, and surveillance colonoscopy.

## Clinical Evidence

Clinical investigation performed to demonstrate safety and effectiveness of cecal intubation and overall system performance. Study confirmed successful cecal intubation in the intended adult population and verified therapeutic access to the lower GI tract for screening, diagnostic, and surveillance colonoscopy. Bench testing included packaging, shelf-life, biocompatibility (ISO 10993), EMC, electrical safety, and optical performance (photobiological safety, resolution, geometric distortion). Pre-clinical in vivo swine studies also conducted.

## Technological Characteristics

Flexible endoscope with CMOS image sensor and LED lighting. Insertion tube constructed of polyurethane with biocompatible hydrophilic coating. Features include irrigation, insufflation, lens wash, and suction channels. Distal tip deflection via manual angulation knobs. Variable stiffness insertion tube. Compliant with ISO 10993 for biocompatibility and voluntary ASCA Pilot for electrical/EMC safety. Single-use, non-reprocessable.

## Regulatory Identification

An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.

## Special Controls

*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.

## Predicate Devices

- Evis Exera II 180 System ([K100584](/device/K100584.md))

## Reference Devices

- Aer-O-Scope Colonoscope System ([K161791](/device/K161791.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food & Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

August 17, 2023

GI View Ltd. % Kristin Davenport Partner Covington & Burling LLP One City Center, 850 Tenth St, NW Washington, DC, District of Columbia 20001-4956

Re: K230588

Trade/Device Name: Aer-O-Scope Colonoscope System Regulation Number: 21 CFR 876.1500 Regulation Name: Endoscope And Accessories Regulatory Class: Class II Product Code: FDF Dated: July 17, 2023 Received: July 18, 2023

Dear Kristin Davenport:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal

{1}------------------------------------------------

statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

# Shanil P. Haugen -S

Shanil P. Haugen, Ph.D. Assistant Director DHT3A: Division of Renal, Gastrointestinal, Obesity and Transplant Devices OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

## Indications for Use

510(k) Number (if known) K230588

#### Device Name Aer-O-Scope Colonoscope System

#### Indications for Use (Describe)

The Aer-O-Scope Colonoscope System is intended to provide visualization (via a video monitor) and diagnostic/ therapeutic access to the adult lower gastrointestinal tract, (including but not limited to, the anus, rectum, sigmoid colon, transverse colon, cecum and ileocecal valve) for endoscopy. The Aer-O-Scope Disposable colonoscope is a single use disposable endoscope and cannot be reprocessed.

| Type of Use (Select one or both, as applicable)                                                                                                                         |                                                                                                                                                                        |
|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| <div style="display:flex; align-items:center;"> <span> <span style="font-size: 20px;">☑</span> </span> <span>Prescription Use (Part 21 CFR 801 Subpart D)</span> </div> | <div style="display:flex; align-items:center;"> <span> <span style="font-size: 20px;">☐</span> </span> <span>Over-The-Counter Use (21 CFR 801 Subpart C)</span> </div> |

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

## 510(K) SUMMARY

| Owner:                                 | GI View Ltd.                                                                                                                |
|----------------------------------------|-----------------------------------------------------------------------------------------------------------------------------|
| Address:                               | 5 Shoham St.<br>5251001 Ramat Gan<br>Israel                                                                                 |
| Contact:                               | Sharon Goldfarb<br>VP Regulatory and Clinical<br>Phone: +972-54-6454034<br>Fax: +972-3-761-7974<br>Email: Sharon@giview.com |
| Summary Date:                          | March 1, 2023                                                                                                               |
| Device Name, 21 CFR 807.92(a)(2)       |                                                                                                                             |
| Trade Name:                            | Aer-O-Scope Colonoscope System¹                                                                                             |
| Common Name:                           | Colonoscope                                                                                                                 |
| Classification Name:                   | Endoscope and accessories, 21 CFR § 876.1500                                                                                |
| Product Code:                          | FDF                                                                                                                         |
| Predicate Device, 21 CFR 807.92(a)(3): |                                                                                                                             |
| Trade Name:                            | Evis Exera II 180 System                                                                                                    |
| 510(k) Number:                         | K100584                                                                                                                     |
| Common Name:                           | Endoscopic Video Imaging System                                                                                             |
| Classification Name:                   | Endoscope and accessories, 21 CFR 876.1500.                                                                                 |

Product Code: NWB, FDF, FDS

Reference Device, 21 CFR 807.92(a)(3):

<sup>1</sup> The Aer-O-Scope Colonoscope System subject to the 510(k) Premarket Notification and the Aer-O-Scope Reference device share the same name and hereafter will be distinguished by referring to as Aer-O-Scope 2 (the Reference K161791) and Aer-O-Scope 3 (the device subject to this 510(k) submission).

{4}------------------------------------------------

Trade Name: Aer-O-Scope Colonoscope System (2)

510(k) Number: K161791

Common Name: Colonoscope

Classification Name: Endoscope and accessories, 21 CFR 876.1500

Product Code: FDF

#### Device Description, 21 CFR 807.92(a)(4)

The Aer-O-Scope 3 Colonoscope System is a flexible, operator-controlled, colonoscope that provides visualization and therapeutic access to the colon. It is comprised of two major components, the Video Controller and the Aer-O-Scope 3 Disposable Colonoscope. The Video Controller contains all components and subsystems required for operation and control of the Aer-O-Scope 3 Colonoscope System. The Aer-O-Scope 3 Disposable Colonoscope includes channels for irrigation, insufflation, lens wash, therapeutic access, and suction. The disposable colonoscope includes an optical imaging head with a CMOS sensor and lenses for visualization with a 200° panoramic omni field of view. All commands are controlled by the operator and regulated through the Video Controller.

The Aer-O-Scope 3 Disposable Colonoscope is a single use device and cannot be reprocessed. The main material that comes in contact with the patient is polyurethane that is coated with a biocompatible hydrophilic coating. The Aer-O-Scope 3 Disposable Colonoscope is biocompatible according to the ISO 10993 harmonized and FDA consensus standard.

## Predicate Comparison, 21 CFR 807.92(a)(6)

The main technological and operational features of the Aer-O-Scope 3 (other than the visualization system) are the same as those of the Evis Exera II 180 System (Evis Exera II Video System Center Olympus CV-180 ("CV-180") & Evis Exera II Colonovideoscope Olympus CF type H180AL ("CF H180AL")). The only differences between the Aer-O-Scope 3 and the predicate relate to visualization (including a CMOS chip and LED lighting) and disposability and the hydrophilic coating, features that are also present in the Aer-O-Scope Colonoscope System 2 (reference device). GI View submitted data from the Aer-O-Scope 2 Colonoscope System for the visualization, hydrophilic coating & disposability features, as well as performance testing to show that the technological differences between the Aer-O-Scope 3 and the predicate do not affect the substantial equivalence determination.

The Aer-O-Scope 3 Colonoscope System has a subset of the indicated uses and associated technological and operational characteristics of the predicate Olympus Evis Exera II 180 System (the indication for the Olympus device also includes surgery). Both devices are indicated for adult populations who may undergo colonoscopy in physicians' clinics, ambulatory surqical centers, and hospital settings.

{5}------------------------------------------------

While operationally the same as the Olympus predicate device, the technology differs in that the Aer-O-Scope 3 uses a CMOS chip and LED lighting and the Olympus predicate uses a CCD chip and fiber optic lighting.

Like the predicate, the Aer-O-Scope 3 Disposable Colonoscope is a long flexible endoscope with optics at the deflectable tip of the device and openings for irrigation, insufflation, and suction. The Aer-O-Scope 3's Video Controller supplies and controls the Aer-O-Scope Disposable Colonoscope according to user command inputs via a touchpad (on the Video Controller) or buttons on the colonoscope handle, similar to the Olympus predicate. The distal tip deflection of both the device subject to this Premarket Notification and the predicate are controlled by angulation knobs on the scope handles. The Aer-O-Scope 3 utilizes a CMOS image sensor and LEDs incorporated directly into the distal tip of the Aer-O-Scope Disposable Colonoscope; whereas the CCD image sensor and Xenon light source of the Olympus predicate are external, and signals are carried over fiber optics. Like the predicate, the Aer-O-Scope 3 Colonoscope System is equipped with insufflation, iens wash and suction.

Hydrophilic coating and predefined variable stiffness along the insertion tube are incorporated into the Aer-O-Scope 3 to facilitate colonic intubation. The Olympus CF H180AL Colonoscope has an operator controlled variable stiffness mechanism to facilitate colonic intubation. While the mechanisms for colonic intubation facilitation are different, both methods have shown to be successful for streamlining colonic intubation. These differences do not affect the safety and effectiveness of the Aer-O-Scope 3 compared to the predicate.

{6}------------------------------------------------

#### Indications for Use Comparison:

| Aer-O-Scope 3<br>Colonoscope System                                                                                                                                                                                                                                                                                                                                                                                                                                        | Evis Exera II 180<br>System<br>(K100584)                                                                                                                                                                                                                                                                                                                                                                                                                                                       | (Reference) Aer-O-<br>Scope 2 Colonoscope<br>System                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       |
|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Indications for Use<br>Statement                                                                                                                                                                                                                                                                                                                                                                                                                                           | Indications for Use<br>Statement                                                                                                                                                                                                                                                                                                                                                                                                                                                               | (K161791)<br>Indications for Use<br>Statement                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             |
| The Aer-O-Scope (3)<br>Colonoscope System is<br>intended to provide<br>visualization (via a<br>video monitor) &<br>diagnostic / therapeutic<br>access to the adult<br>lower gastrointestinal<br>tract, (including but not<br>limited to, the anus,<br>rectum, sigmoid colon,<br>transverse colon,<br>cecum and ileocecal<br>valve) for endoscopy.<br>The Aer-O-Scope<br>Disposable<br>colonoscope is a single<br>use disposable<br>endoscope and cannot<br>be reprocessed. | EVIS EXERA II<br>COLONOVIDEOSCOPE<br>OLYMPUS CF TYPE<br>H180AL,<br>These instruments have<br>been designed to be<br>used with an Olympus<br>video system<br>center, light source,<br>documentation<br>equipment, monitor,<br>Endotherapy accessories<br>(such as a biopsy<br>forceps), and other<br>ancillary equipment for<br>endoscopy and<br>endoscopic<br>surgery within the<br>lower digestive tract<br>(including the anus,<br>rectum, sigmoid colon,<br>colon,<br>and ileocecal valve). | The Aer-O-Scope (2)<br>Colonoscope System is<br>intended to provide<br>panoramic (360°)<br>visualization (via a<br>video<br>monitor) & diagnostic /<br>therapeutic access to<br>the adult lower<br>gastrointestinal tract,<br>(Including but not<br>limited to, the anus,<br>rectum, sigmoid colon,<br>transverse colon,<br>cecum and ileocecal<br>valve) for endoscopy.<br>The Aer-O-Scope<br>Disposable Scanner<br>(Colonoscope<br>component of the Aer-O-<br>Scope<br>Colonoscope System) is<br>a single use disposable<br>device. An Aer-O-Scope<br>Scanner cannot be<br>reprocessed. |

#### Performance Data - Bench, 21 CFR 807.92(b)(1):

GI View Ltd. conducted packaging and shelf-life testing which met all the criteria.

Biocompatibility tests were performed and met all the required criteria.

EMC and Electrical safety were tested (voluntary ASCA Pilot) and found compliant with the applicable standards.

Bench tests to measure the safety and effectiveness of device components were performed.

{7}------------------------------------------------

Bench tests to measure optical parameters and safety such as photobiological safety, resolution and geometric distortion were performed.

In all instances the Aer-O-Scope 3 Colonoscope System functioned as intended and met the individual test specifications and/or FDA consensus standards.

Pre-clinical in vivo tests in swine were performed on the final device to demonstrate overall safety and efficacy.

## Performance Data - Clinical, 21 CFR 807.92(b)(2);

Clinical data support the determination of substantial equivalence. A clinical investigation was performed to demonstrate safety and effectiveness of cecal intubation and the performance of the Aer-O-Scope Colonoscope System 3. The data demonstrated that the Aer-O-Scope 3 Colonoscope System (with a wider field of view (200°) compared to the predicate) successfully allows cecal intubation in the intended use population and provides therapeutic access to the lower gastrointestinal tract for screening, diagnostic and surveillance endoscopy (colonoscopy) as the predicate devices.

## Summary, 21 CFR 807.92(b)(3):

The bench tests included studies related to forces, biocompatibility, optical performance and constructive and electrical safety. The tests demonstrated that the Aer-O-Scope 3 Colonoscope System performed as expected and met the specifications/FDA consensus standards requirements. As such, the bench tests demonstrated the Aer-O-Scope 3 performed as well as the predicate devices. The data also demonstrated that the Aer-O-Scope 3 Colonoscope System successfully provided a wider field of view (200°) as well as therapeutic access to the lower gastrointestinal tract for screening, diagnostic and surveillance endoscopy as other colonoscopes.

---

**Source:** [https://fda.innolitics.com/device/K230588](https://fda.innolitics.com/device/K230588)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
