← Product Code [LZA](/productcode/LZA) · K230542

# Disposable Nitrile Gloves (ABC-DG01) (K230542)

_Huizhou Anboson Technology Co., Ltd. · LZA · May 31, 2023 · General Hospital · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K230542

## Device Facts

- **Applicant:** Huizhou Anboson Technology Co., Ltd.
- **Product Code:** [LZA](/productcode/LZA.md)
- **Decision Date:** May 31, 2023
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 880.6250
- **Device Class:** Class 1
- **Review Panel:** General Hospital

## Indications for Use

The Disposable Nitrile Glove is a disposable device intended for medical purposes that is worn on the examiner's hand to prevent contamination between patient and examiner.

## Device Story

Disposable nitrile examination gloves; ambidextrous; blue or black color; powder-free. Worn by clinicians during patient examinations to provide a barrier against contamination. Manufactured per ASTM D6319-19. Non-sterile, single-use. Protects both patient and examiner from cross-contamination.

## Clinical Evidence

No clinical data. Bench testing only. Performance verified against ASTM D6319-19 (dimensions, physical properties), ASTM D6124-06 (residual powder), and ASTM D5151-19 (freedom from holes). Biocompatibility testing performed per ISO 10993-5, 10993-10, 10993-11, and 10993-23.

## Technological Characteristics

Material: Nitrile. Design: Ambidextrous, powder-free, blue or black. Standards: ASTM D6319-19, ASTM D6124-06, ASTM D5151-19. Non-sterile. Single-use.

## Regulatory Identification

A non-powdered patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. A non-powdered patient examination glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.

## Predicate Devices

- Disposable Powder Free Nitrile Examination Glove, White/Blue/ Black/ Pink Color ([K171422](/device/K171422.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health and Human Services logo on the left and the FDA logo on the right. The FDA logo features the letters "FDA" in a blue square, followed by the words "U.S. Food & Drug Administration" in blue text.

May 31, 2023

Huizhou Anboson Technology Co., Ltd. % Ivv Wang Consultant Shanghai Sungo Management Consulting Co. Ltd. 14th Floor, 1500# Century Avenue Shanghai. 200122 China

Re: K230542

Trade/Device Name: Disposable Nitrile Gloves (ABC-DG01) Regulation Number: 21 CFR 880.6250 Regulation Name: Non-Powdered Patient Examination Glove Regulatory Class: Class I, reserved Product Code: LZA Dated: May 5, 2023 Received: May 5, 2023

Dear Ivy Wang:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

# Bifeng Qian -S

Bifeng Qian, M.D., Ph.D. Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known) K230542

Device Name Disposable Nitrile Gloves (ABC-DG01 )

Indications for Use (Describe)

The Disposable Nitrile Glove is a disposable device intended for medical purposes that is worn on the examiner's hand to prevent contamination between patient and examiner.

| Type of Use (Select one or both, as applicable) |  |
|-------------------------------------------------|--|
|-------------------------------------------------|--|

Prescription Use (Part 21 CFR 801 Subpart D)

X Over-The-Counter Use (21 CFR 801 Subpart C)

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

# 510(K) Summary

(As requirement by 21 CFR 807.92)

## K230542

Date prepared: 2023-05-05

#### A. Applicant:

Name: Huizhou Anboson Technology Co., Ltd.

Address: 2F, No.7, Hechang East Fourth Road, Zhongkai High-tech Zone, Huizhou City, Guangdong, 516001 China

Contact: Ms. Xuanli Weng Title: Manager Tel: +86-752-2607756

Submission Correspondent: Primary contact: Ms. Ivy Wang Shanghai SUNGO Management Consulting Co., Ltd. Room 1401, Dongfang Building, 1500# Century Ave., Shanghai 200122, China Tel: +86-21-58817802 Email: haiyu.wang@sungoglobal.com Secondary contact: Mr. Raymond Luo Room 1401, Dongfang Building, 1500# Century Ave., Shanghai 200122, China Tel: +86-21-68828050 Email: fda.sungo@gmail.com

#### B. Device:

Trade Name: Disposable Nitrile Gloves Common Name: Nitrile Patient Examination Gloves (Powder Free) Model: ABC-DG01

Regulatory Information Classification Name: Non-powdered patient examination glove Classification: Class I Product code: LZA Regulation Number: 21 CFR 880.6250 Review Panel: General Hospital

#### C. Predicate device:

K171422 Disposable Powder Free Nitrile Examination Glove, White/Blue/ Black/ Pink Color Ever Global (Vietnam) Enterprise Corp

#### D. Indications for use of the device:

{4}------------------------------------------------

The Disposable Nitrile Gloves is a disposable device intended for medical purposes that is worn on the examiner's hand to prevent contamination between patient and examiner.

#### E. Device Description:

The Disposable Nitrile Gloves are non-sterile, single use only, disposable examination gloves intended for medical purposes to be worn by examiners to prevent contamination between the patient and the examiner. The gloves are blue or black color, powder free, nitrile ambidextrous gloves. The gloves are offered in large (L) size, packed in a paper box.

The gloves are designed and manufactured in accordance with the ASTM D6319-19 standard.

| Table 1 General Comparison of Proposed and Predicate Devices |                                                                                                                                                                               |                                                                                                                                                                                                                                       |        |
|--------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------|
| Device                                                       | Proposed Device                                                                                                                                                               | Predicate Device                                                                                                                                                                                                                      | Result |
| 510K #                                                       | K230542                                                                                                                                                                       | K171422                                                                                                                                                                                                                               | -      |
| Product Name                                                 | Disposable Nitrile Gloves                                                                                                                                                     | Disposable Powder Free Nitrile Examination Glove, White/Blue/ Black/ Pink Color                                                                                                                                                       | -      |
| Product Code                                                 | LZA                                                                                                                                                                           | LZA                                                                                                                                                                                                                                   | Same   |
| Classification                                               | Class I                                                                                                                                                                       | Class I                                                                                                                                                                                                                               | Same   |
| Regulation Number                                            | 21 CFR 880.6250                                                                                                                                                               | 21 CFR 880.6250                                                                                                                                                                                                                       | Same   |
| Indications for use                                          | The Disposable Nitrile Gloves is a disposable device intended for medical purposes that is worn on the examiner's hand to prevent contamination between patient and examiner. | The Disposable Powder Free Nitrile Examination Glove, White/ Blue/ Black/ Pink Color is a disposable device intended for medical purposes that is worn on the examiner's hands to prevent contamination between patient and examiner. | Same   |
| Powder free                                                  | Yes                                                                                                                                                                           | Yes                                                                                                                                                                                                                                   | Same   |
| Design feature                                               | Ambidextrous                                                                                                                                                                  | Ambidextrous                                                                                                                                                                                                                          | Same   |
| Material                                                     | Nitrile                                                                                                                                                                       | Nitrile                                                                                                                                                                                                                               | Same   |
| OTC use                                                      | Yes                                                                                                                                                                           | Yes                                                                                                                                                                                                                                   | Same   |
| Sterility                                                    | Non-sterile                                                                                                                                                                   | Non-sterile                                                                                                                                                                                                                           | Same   |
| Use                                                          | Singe use                                                                                                                                                                     | Single use                                                                                                                                                                                                                            | Same   |
| Label                                                        | Single-use, indication, powder free, device color, device name, glove size and quantity, Disposable Nitrile Glove, Non-Sterile                                                | Single-use, indication, powder free, device color, device name, glove size and quantity, Disposable Powder Free Nitrile Examination Glove, Non-                                                                                       | Same   |

Table 1 General Comparison of Proposed and Predicate Devices

F. Summary of Technological Characteristics

Table 2 Device dimension comparison

| Predicate<br>device | Designation | Size |   |   |   |    | Tolerance |
|---------------------|-------------|------|---|---|---|----|-----------|
|                     |             | XS   | S | M | L | XL |           |

Sterile

{5}------------------------------------------------

| (K171422)                     | Length, mm       | 230           | 230 | 230 | 230 | 230       | min      |
|-------------------------------|------------------|---------------|-----|-----|-----|-----------|----------|
|                               | Width, mm        | 75            | 85  | 90  | 105 | 115       | $\pm$ 5  |
|                               |                  | Thickness, mm |     |     |     |           |          |
|                               | Finger           | 0.05          |     |     |     |           | min      |
|                               | Palm             | 0.05          |     |     |     |           | min      |
|                               |                  | Size          |     |     |     |           |          |
|                               | Designation<br>L |               |     |     |     | Tolerance |          |
| Proposed<br>device<br>K230542 | Length, mm       | 230           |     |     |     |           | min      |
|                               | Width, mm        | 110           |     |     |     |           | $\pm$ 10 |
|                               | Thickness, mm    |               |     |     |     |           |          |
|                               | Finger           | 0.05          |     |     |     |           | min      |
|                               | Palm             | 0.05          |     |     |     |           | min      |
| Result                        | Similar          |               |     |     |     |           |          |

Analysis: The physical dimensions are different with that of the predicate, but they all meet the requirements of ASTM D6319-19, so the differences do not raise any new safety or performance questions.

The proposed device has less size than that of the predicate device, while this difference will not affect the safety or performance of the proposed device.

| Item                   |                        |                     | Proposed<br>device (K230542)                                                             | Predicate<br>device (K171422)                                                      | Result                   |         |
|------------------------|------------------------|---------------------|------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------|--------------------------|---------|
| Colorant               |                        |                     |                                                                                          | Blue/ Black                                                                        | White/ Blue/ Black/ Pink | Similar |
| Physical<br>properties | Before aging           | Tensile strength    | 14MPa, min                                                                               | 14MPa, min                                                                         | Same                     |         |
|                        |                        | Ultimate elongation | 500%, min                                                                                | 500%, min                                                                          | Same                     |         |
|                        | After aging            | Tensile strength    | 14MPa, min                                                                               | 14MPa, min                                                                         | Same                     |         |
|                        |                        | Ultimate elongation | 400%, min                                                                                | 400%, min                                                                          | Same                     |         |
|                        | Comply with ASTM D6319 |                     |                                                                                          | Comply with ASTM<br>D6319                                                          | Same                     |         |
| Freedom from holes     |                        |                     | Be free from<br>holes when<br>tested in<br>accordance with<br>ASTM D5151<br>G-1, AQL 2.5 | Be free from holes when<br>tested in accordance<br>with ASTM D5151<br>G-1, AQL 2.5 | Same                     |         |
| Residual Powder        |                        |                     | Meet the<br>requirements of<br>ASTM D6124                                                | Meet the requirements<br>of ASTM D6124                                             | Same                     |         |

Table 3 Performance comparison

Analysis: The subject device is available in blue or black color; however, the predicate is available in multiple colors (white, blue, black, pink). Biocompatibility testing was successfully completed for the subject device, demonstrating that any color differences do not affect the safety of the proposed device.

Table 4 Biocompatibility comparison

{6}------------------------------------------------

| Item             | Proposed<br>device (K230542)               | Predicate device<br>(K171422)                                                                                             | Result                      |           |
|------------------|--------------------------------------------|---------------------------------------------------------------------------------------------------------------------------|-----------------------------|-----------|
| Material         | Nitrile                                    | Nitrile                                                                                                                   | Same                        |           |
| Biocompatibility | Irritation<br>ISO 10993-23                 | Under the conditions of the<br>study, not an<br>irritant.                                                                 | Comply with ISO<br>10993-10 | Same      |
|                  | Sensitization<br>ISO 10993-10              | Under the<br>conditions of the<br>study, not a<br>sensitizer.                                                             |                             |           |
|                  | In Vitro<br>Cytotoxicity<br>ISO 10993-5    | Under the<br>conditions of the<br>study, the test<br>article extract<br>showed<br>potential toxicity.                     | Not available               | Different |
|                  | Acute Systemic<br>Toxicity<br>ISO 10993-11 | Under the<br>conditions of the<br>study, the device<br>extract does not<br>induce acute<br>systemic toxicity<br>response. | Not available               | Different |

Analysis: In vitro cytotoxicity and acute systemic toxicity information for the predicate device is not publicly available. This does not raise different safety or performance questions since the subject device has acceptable biocompatibility per the biocompatibility endpoint assessment.

## G. Summary of Non-Clinical Testing

#### A Biocompatibility

The following tests for the subject device were conducted the biocompatibility of Disposable Nitrile Gloves:

- ISO 10993-5: In Vitro Cytotoxicity
- ISO 10993-23: Primary Skin Irritation
- ISO 10993-10: Dermal Sensitization
- · ISO 10993-11: Acute Systemic Toxicity

#### A Performance Testing

Performance testing of the proposed device was conducted as per ASTM D6319-19 Standard Specification for Nitrile Examination Gloves for Medical Application.

To summarize, the performance testing of the subject device was conducted to adequately demonstrate the effectiveness of the device in accordance with the relevant test methods cited below:

{7}------------------------------------------------

- ASTM D6319-19 Standard Specification for Nitrile Examination Gloves for Medical Application
- ASTM D6124-06 (Reapproved 2017) Standard Test Method for Residual Powder on Medical Gloves
- ASTM D5151-19 Standard Test Method for Detection of Holes in Medical Gloves

| Test Method                                                    | Purpose                                                                                                                | Acceptance Criteria                                                                                                                                      | Results (Blue & Black)                                                                                                                                                  |
|----------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Dimensions<br>(width)<br>(thickness)<br>ASTM<br>D6319-19       | The purpose of the test is to<br>evaluate the physical<br>dimension of the glove                                       | Width 110 ±10mm<br>Length 230mm min<br><br>Palm - 0.05mm min.<br>Finger-0.05mm min.                                                                      | Pass<br>106mm min width<br>250mm min length<br><br>Pass<br>Palm – 0.06mm min.<br>Finger-0.12mm min                                                                      |
| Physical<br>properties<br>ASTM<br>D6319-19                     | The purpose of the test is to<br>evaluate the tensile<br>strength and ultimate<br>elongation before and after<br>aging | Tensile Strength:<br>Before Aging ≥ 14 MPa,<br>min.<br>After Aging ≥ 14 MPa,<br>min.<br>Elongation:<br>Before Aging 500%, min.<br>After Aging 400%, min. | Pass<br>Tensile Strength:<br>Before Aging 19.07MPa,<br>min.<br>After Aging 18.11 MPa,<br>min.<br>Elongation:<br>Before Aging 586.47%, min.<br>After Aging 469.20%, min. |
| Freedom<br>from holes<br>ASTM D5151-<br>19                     | The purpose of the test is to<br>detect holes in the gloves                                                            | No leakage at sampling<br>level of G-1, AQL 2.5                                                                                                          | Pass<br>No leakage, 125 of 125<br>passed of each color                                                                                                                  |
| Residual<br>Powder<br>ASTM<br>D6124-06<br>(Reapproved<br>2017) | The purpose of the test is to<br>detect the powder residue<br>in the glove                                             | <2mg per glove                                                                                                                                           | Pass<br>average 0.07 mg per glove                                                                                                                                       |

## H. Clinical Test Conclusion

No clinical study is included in this submission.

#### l. Conclusion

The conclusion drawn from the nonclinical tests demonstrates that the subject device in 510(K) submission, the Disposable Nitrile Gloves are as safe, as effective, and performs as well as or better than the legally marketed predicate device cleared under K171422.

---

**Source:** [https://fda.innolitics.com/device/K230542](https://fda.innolitics.com/device/K230542)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
