The Gamma4 System is intended to achieve functionally stable osteosyntheses and stabilization of bones and bone fragments.
Device Story
Gamma4 System is a hip fracture nailing system consisting of sterile implants (nails, lag screws, end caps) and non-sterile instruments (targeting devices). Used by orthopedic surgeons in clinical settings to achieve stable osteosynthesis. Implants are inserted into the femur to stabilize fractures or correct deformities. System includes intermediate nails, RC lag screws, and targeting sleeves. Device provides mechanical stabilization of bone fragments to facilitate healing. No software or AI components.
Clinical Evidence
No clinical data. Substantial equivalence supported by bench testing, including targeting accuracy, bending fatigue strength (ASTM F1264), construct fatigue strength, cut-out testing, torsional stiffness (ASTM F1264), and MR evaluation.
Technological Characteristics
Implants: Titanium alloy (Ti6Al4V ELI) per ASTM F136. Instruments: Stainless steel (AISI 431), carbon fiber reinforced PEEK, and unreinforced PEEK. System includes intramedullary nails, lag screws, end caps, and targeting devices. Mechanical fixation principle. No software or electronic components.
Indications for Use
Indicated for treatment of stable and unstable fractures, stabilization of bones, and correction of bone deformities in intracapsular, trochanteric, subtrochanteric, and shaft regions of the femur (including osteoporotic and osteopenic bone). RC Lag Screw is also indicated for rotationally unstable fractures.
Regulatory Classification
Identification
An intramedullary fixation rod is a device intended to be implanted that consists of a rod made of alloys such as cobalt-chromium-molybdenum and stainless steel. It is inserted into the medullary (bone marrow) canal of long bones for the fixation of fractures.
T2 Alpha Femur Antegrade GT/PF Nailing System (K203819)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. Underneath the square are the words "U.S. FOOD & DRUG ADMINISTRATION" in blue.
June 9, 2023
Stryker GmbH Cindy Leon Sr. Staff Specialist, Regulatory 325 Corporate Drive Mahwah, New Jersey 07430
Re: K230512
Trade/Device Name: Gamma4 System Regulation Number: 21 CFR 888.3020 Regulation Name: Intramedullary fixation rod Regulatory Class: Class II Product Code: HSB Dated: May 25, 2023 Received: May 25, 2023
Dear Cindy Leon:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal
{1}------------------------------------------------
statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
| Sincerely, | |
|------------|-------------------------------------------|
| Farzana | Digitally signed by<br>Farzana Sharmin -S |
| Sharmin -S | Date: 2023.06.09<br>16:04:36 -04'00' |
Farzana Sharmin, Ph.D. Acting Assistant Director DHT6A: Division of Joint Arthroplasty Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known)
K230512
Device Name Gamma4 System
Indications for Use (Describe)
The Gamma4 System is indicated for the treatment of stable and unstable fractures as well as for stabilization of bones and correction of bone deformities in the intracapsular, trochanteric and shaft regions of the femur (including osteoporotic and osteopenic bone). The RC Lag Screw is also indicated for rotationally unstable fractures.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|---------------------------------------------|
| Prescription Use (Part 21 CFR 801 Subpart D) | Over-The-Counter Use (21 CFR 801 Subpart C) |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
| Section 9: 510(k) Summary | |
|---------------------------|-----------------------------------------------------|
| I. SUBMITTER | |
| Sponsor: | Stryker GmbH |
| | Bohnackerweg 1 |
| | 2545 Selzach / Switzerland |
| Contact Person: | Cindy Leon |
| | Sr. Staff Specialist, Regulatory Affairs |
| | 325 Corporate Drive |
| | Mahwah, NJ 07430 |
| | Phone: 201-831-5374 |
| | Fax: 201-831-6020 |
| Date Prepared: | February 24, 2023 |
| II. DEVICE | |
| Name of Device: | Gamma4 System |
| Common Name: | Gamma4 System |
| | Rod, fixation, intramedullary and accessories |
| Regulation Number / Name: | Gamma4 System |
| | 21CFR 888.3020 (Intramedullary fixation rod) |
| Product Code: | Gamma4 System |
| | HSB (Rod, fixation, intramedullary and accessories) |
| Regulatory Class: | Class II |
{4}------------------------------------------------
Traditional 510(k) Premarket Notification
### III, PREDICATE DEVICE
Primary Predicate: Gamma3 System (K213328)
Additional Predicates:
T2 Alpha Femur Antegrade GT/PF Nailing System (K203819)
## IV. DEVICE DESCRIPTION
This Traditional 510(k) submission is being supplied to the U.S. FDA to gain clearance to market new devices into the Gamma4 System, add new indications for use for the RC Lag Screws and update the current surgical technique to include an optional section to use bone void fillers.
The Gamma4 System was previously cleared in K222309, is a hip fracture nailing system and includes sterile implants (Trochanteric and Long Nails in various diameters and sizes, Lag Screws, and End Caps) as well as non-sterile instruments (targeting devices). The subject devices (Intermediate Nails and RC Lag Screw) are also sterile implants in various diameters and sizes as well as a non-sterile Intermediate Targeting Sleeve. All nails and screws are made of titanium alloy (Ti6Al4V ELI) as per ASTM F136 and the targeting device is manufactured from stainless steel per AISI 431, carbon fiber reinforced PEEK and PEEK unreinforced.
#### V. INTENDED USE
#### Gamma4 System
The introduction of the Intermediate Nails, RC Lag Screw and Intermediate Targeting Sleeve does not alter the intended use of the subject system previously cleared in K222309. The intended use is provided below:
The Gamma4 System is intended to achieve functionally stable osteosyntheses and stabilization of bones and bone fragments.
{5}------------------------------------------------
## VI. INDICATIONS FOR USE
#### Gamma4 System
The Gamma4 System is indicated for the treatment of stable and unstable fractures as well as for stabilization of bones and correction of bone deformities in the intracapsular, trochanteric, subtrochanteric and shaft regions of the femur (including osteoporotic and osteopenic bone).
The RC Lag Screw is also indicated for rotationally unstable fractures.
#### Summary of Technologies
A comparison of the systems demonstrated that the subject Gamma4 System is substantially equivalent to the following predicates in regards to intended use, material, design, and operational principles:
- Gamma3 System (K213328) .
- . T2 Alpha Femur Antegrade GT/PF Nailing System (K203819)
#### Non-Clinical Testing
The following performance assessment was made in support of substantial equivalence:
The following tests were performed:
- . Targeting accuracy testing
- Bending Fatigue Strength per ASTM F1264 .
The following performance assessment was conducted:
- Construct Fatigue Strength Testing ●
- . Cut-out testing
- . Torsional Stiffness per ASTM F1264
- MR evaluation ●
Testing demonstrated that the Gamma4 System is equivalent in mechanical performance to the Gamma3 System primary predicate device (K213328) and T2 Alpha Femur Antegrade GT/PF Nailing System (K203819) additional predicate device.
{6}------------------------------------------------
# CLINICAL TESTING
Clinical testing was not required for this submission.
## CONCLUSION
The subject Gamma4 System is substantially equivalent to the previously cleared Gamma3 System (K213328) primary predicate.
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.