K230510 · Cynosure, LLC · GAT · Jul 6, 2023 · General, Plastic Surgery
Device Facts
Record ID
K230510
Device Name
MyEllevate
Applicant
Cynosure, LLC
Product Code
GAT · General, Plastic Surgery
Decision Date
Jul 6, 2023
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.5000
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The MyEllevate device is indicated for soft tissue approximation and elevation of sub dermis and underlying muscle.
Device Story
MyEllevate is a surgical suture system for soft tissue approximation and elevation of sub dermis and underlying muscle. The system includes a synthetic non-absorbable braided polyester suture and manual surgical instruments: skin marking tape, lancet, tunneling rod, two stainless steel Suturods, and a light handle with an LED source. Each Suturod features an internal fiber optic light pipe that connects to the light handle to illuminate the distal tip. This allows the physician to visually verify the location and depth of the Suturod as it is passed through the subcutaneous space. All components are provided sterile for single-use. The device is used in a clinical setting by a physician to assist in the accurate placement of sutures. The illumination feature provides real-time visual feedback during the tunneling process, aiding the surgeon in precise tissue manipulation, which can improve procedural accuracy and patient outcomes.
Clinical Evidence
No clinical data submitted. Substantial equivalence is based on non-clinical bench testing, including electrical safety (IEC 60601-1), electromagnetic compatibility (IEC 60601-1-2), usability (IEC 60601-1-6), and photobiological safety (IEC 62471), alongside comparative performance data for the suture material.
Technological Characteristics
System includes non-absorbable braided polyester suture and manual instruments (stainless steel Suturods, lancet, tunneling rod). Features fiber optic light pipe for distal tip illumination via LED light handle. Electrical safety standards: IEC 60601-1, IEC 60601-1-2, IEC 60601-1-6, IEC 62471. Biocompatibility per ISO 10993-1. Sterilization: Ethylene oxide. Single-use.
Indications for Use
Indicated for soft tissue approximation and elevation of sub dermis and underlying muscle in patients requiring such procedures.
Regulatory Classification
Identification
Nonabsorbable poly(ethylene terephthalate) surgical suture is a multifilament, nonabsorbable, sterile, flexible thread prepared from fibers of high molecular weight, long-chain, linear polyesters having recurrent aromatic rings as an integral component and is indicated for use in soft tissue approximation. The poly(ethylene terephthalate) surgical suture meets U.S.P. requirements as described in the U.S.P. Monograph for Nonabsorbable Surgical Sutures; it may be provided uncoated or coated; and it may be undyed or dyed with an appropriate FDA listed color additive. Also, the suture may be provided with or without a standard needle attached.
Special Controls
*Classification.* Class II (special controls). The special control for this device is FDA's “Class II Special Controls Guidance Document: Surgical Sutures; Guidance for Industry and FDA.” See § 878.1(e) for the availability of this guidance document.
{1}------------------------------------------------
Build Correspondence
Image /page/0/Picture/2 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: a symbol on the left and the FDA name on the right. The symbol on the left is a stylized image of a human figure, while the FDA name on the right is written in blue letters.
July 6, 2023
Cynosure, LLC Cynthia Aguirre Regulatory Affairs Specialist II 5 Carlisle Road Westford, Massachusetts 01886
Re: K230510
Trade/Device Name: MyEllevate Regulation Number: 21 CFR 878.5000 Regulation Name: Nonabsorbable Poly(Ethylene Terephthalate) Surgical Suture Regulatory Class: Class II Product Code: GAT Dated: April 14, 2023 Received: April 17, 2023
Dear Cynthia Aguirre:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part
{2}------------------------------------------------
801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE(@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Alexander Nquyen -S
Digitally signed by Alexander Nquyen -S Date: 2023.07.06 16:23:26 -04'00'
for Cynthia J. Chang, Ph.D. Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
## Indications for Use
510(k) Number (if known) K230510
Device Name MyEllevate
Indications for Use (Describe)
MyEllevate is indicated for use in soft tissue approximation of sub dermis and underlying muscle
Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)
_ Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
### Section 5- 510(K) Summary
{5}------------------------------------------------
Section 5
510(k) Summary
#### 510(k) Summary for Cynosure MyEllevate
A summary of 510(k) safety and effectiveness information in accordance with the requirements of 21 CFR 807.92.
| 807.92(a)(1) Submitter Information | |
|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------|
| Applicant | Cynosure, LLC |
| Address | 5 Carlisle Road, Westford MA, 01886 |
| Phone Number | 480-401-9884 |
| Fax Number | N/A |
| Establishment Registration Number | 1222993 |
| Contact Person | Cynthia Aguirre |
| Preparation Date | July 6, 2023 |
| 807.92(a)(2) Name of Device | |
| Trade or Proprietary Name | MyEllevate |
| Common or Usual Name | Nonabsorbable Polyester Suture |
| Classification Name | Nonabsorbable poly(ethylene terephthalate) surgical suture |
| Classification Panel | General & Plastic Surgery |
| Regulation | 21 CFR, §878.5000 |
| Regulatory Class | II |
| Product Code(s) | GAT |
| 807.92 (a)(3) Legally marketed device(s) to which equivalence is claimed | |
| Predicate Devices | Implicitguide™ Surgical Suture System (K091061) TEVDEK ® Polyester Suture (K1001434, K021019) |
| 807.92(a)(4) Device Description | |
| The MyEllevate device consists of a Surgical Suture System consisting of synthetic non-absorbable, braided, polyester surgical suture, and a set of instruments designed to efficiently and accurately place the suture for the purpose of performing soft tissue approximation. MyEllevate contains a Skin Marking Tape, a Lancet, a Tunneling Rod, two stainless steel Suturods, and a Light Handle which contains an LED light source. Each Suturod™ has an internal fiber optic "light pipe" which, when connected to the Ligh Handle, illuminates the distal tip of the Suturod™ and permits visual verification of the location and depth of the Suturod™ as it is passed through the subcutaneous space. All components of MyEllevate are provided sterile and are intended for single use only. While the nonabsorbable suture is a Class II device as describes by 21CFR, §878.5000, the | |
| remaining instruments are Class I devices exempt from 510(k) requirements. The Lancet, Clearing Device, and Tunneling Rod are manual surgical devices described by 21 CFR §878.4800. The Skin Marking Pen is described in 21 CFR §878.4660, and the fiber optic "light pipe" and Light Handle to illuminate surgical sites, is described in 21 CFR, §876.4530. | |
| <b>807.92(a)(5) Intended Use of the Device</b> | |
| The MyEllevate device is indicated for soft tissue approximation and elevation of sub dermis and underlying muscle. | |
| <b>807.92(b)(1) Non-clinical tests submitted</b> | |
| The following non- clinical tests have been included in this 510(k) submission in determination of substantial equivalence between the test device and the referenced predicates. | |
| <b>Electromagnetic Compatibility and Electrical Safety</b><br>Performance testing was conducted for MyEllevate to demonstrate that the device satisfies electrical safety requirements of the following standards: | |
| • | IEC 60601-1:2005/AMD1L2012 & AMD2:2020 Medical Electrical Equipment Part 1:<br>General requirements for basic safety and essential performance |
| • | IEC 60601-1-2: Ed 4.1 Medical Electrical Equipment Part 1-2: General |
{6}------------------------------------------------
Section 5
- IEC 60601 60601-1-2: Ed 4.1 Medical Electrical Equipment Part 1-2: General requirements for basic safety and essential performance – Collateral standard: Electromagnetic Compatibility - Requirements and Tests
- IEC 60601-1-6:2005/AMD1:2012 & AMD2:2020Medical Electrical Equipment Part ● 1-6: General requirements for basic safety and essential performance - Collateral standard: Usability
IEC 62471: 2008 Photobiological safety of lamps and lamp system
# 807.92(b)(2) Clinical tests submitted – N/A – No clinical tests submitted 807.92(b)(3) Conclusions drawn from clinical and non-clinical tests submitted
The results of our comparisons demonstrated the substantial equivalence of the subject device to the identified predicate devices.
{7}------------------------------------------------
Cynosure MyEllevate 510(k)
K230510
Page 4 of 5 Section 5
Cynosure, LLC
510(k) Summary
| Device | MyEllevate<br>(K230510) | Implicitguide™ Surgical Suture<br>System<br>(K091061) | TEVDEK® II Suture<br>(K001434, K021019) |
|-------------------------------------|------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------|
| Indication for Use | The MyEllevate device is<br>indicated for soft tissue<br>approximation and elevation of<br>sub dermis and underlying muscle. | The Implicitguide™ Surgical<br>Suture System is indicated for use<br>in soft tissue approximation and<br>elevation of sub dermis and<br>underlying muscle | Soft tissue approximation |
| Suture<br>Composition | Poly (Ethylene terephthalate) | Poly (Ethylene terephthalate) | Poly (Ethylene terephthalate) |
| Suture<br>Construction | Braided | Braided | Braided |
| Suture Color | Undyed (Natural)<br><br>Dyed (green) | Undyed (Natural)<br><br>Dyed (green) | Undyed (Natural)<br><br>Dyed (green) |
| Suture Size | USP 4-0 | USP 4-0 | USP 4-0 |
| USP Official<br>Monograph | Conforms | Conforms | Conforms |
| <861> Suture<br>Diameter | Conforms | Conforms | Conforms |
| <881> Tensile<br>Strength | Conforms | Conforms | Conforms |
| Attached Needles | Yes | Yes | Yes |
| Needle<br>Composition | Stainless Steel | Stainless Steel | Stainless Steel |
| <871> Needle<br>attachment- Sutures | Conforms | Conforms | Conforms |
{8}------------------------------------------------
| Manufacture | Cynosure, LLC. | ImplicitCare, LLC. | Teleflex |
|---------------------------|-------------------------------------------------------|-------------------------------------------------------|-------------------------------------------------------|
| How Supplied | Sterile; single- use | Sterile; single- use | Sterile; single- use |
| Sterilization<br>Method | Ethylene oxide | Ethylene oxide | Ethylene oxide |
| Biocompatibility | Meets ISO 10993-1 | Meets ISO 10993-1 | Meets ISO 10993-1 |
| Additional<br>instruments | Manual surgical instruments per<br>§878.4800 supplied | Manual surgical instruments per<br>§878.4800 supplied | Manual surgical instruments per<br>§878.4800 supplied |
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.