Medical Video Endoscope System

K230475 · Huger Medical Instrument Co., Ltd. · FGB · Nov 3, 2023 · Gastroenterology, Urology

Device Facts

Record IDK230475
Device NameMedical Video Endoscope System
ApplicantHuger Medical Instrument Co., Ltd.
Product CodeFGB · Gastroenterology, Urology
Decision DateNov 3, 2023
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 876.1500
Device ClassClass 2
AttributesTherapeutic

Indications for Use

Medical Video Endoscope System has been designed to be used with endo-therapy accessories such as a biopsy forcens and other ancillary equipment for endoscopic surgery within urinary tract and interior of the kidney. Medical Video Processor is intended to be used in conjunction with the endoscopes to enhance and display the image of the field of view of the endoscope on the monitor, and also provides functional supply for the endoscope. Single-use Video Ureterorenoscope is intended to connect with Medical Video Processor to provide video imaging.

Device Story

System comprises single-use flexible video ureterorenoscope and reusable medical video processor; used in hospital settings by physicians. Endoscope captures visual data via distal CMOS sensor; transmits video signal to processor. Processor enhances image; displays output on external monitor; provides power/functional supply to endoscope. Enables visualization of urinary tract/kidney interior for diagnostic and surgical procedures; facilitates use of ancillary instruments (e.g., biopsy forceps) via single instrument channel. Benefits include real-time visualization for minimally invasive urological interventions.

Clinical Evidence

No clinical data. Bench testing only, including biocompatibility (cytotoxicity, sensitization, irritation), optical performance (resolution, depth of field, color reproduction), physical/functional testing (leakage, use-life), and electrical safety/EMC compliance per IEC 60601 series.

Technological Characteristics

System includes single-use flexible ureterorenoscope (TPU, PC, Pebax, Barium sulfate) and reusable video processor. CMOS sensor imaging; LED illumination. Connectivity via DVI output. Sterilization of endoscope via Ethylene Oxide (SAL 10^-6). Complies with ISO 8600 series, ISO 10993, IEC 60601-1, IEC 60601-1-2, and IEC 60601-2-18.

Indications for Use

Indicated for endoscopic surgery and diagnosis within the urinary tract and interior of the kidney in patients requiring such procedures. Used with endo-therapy accessories (e.g., biopsy forceps).

Regulatory Classification

Identification

An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.

Special Controls

*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image contains the logos of the Department of Health & Human Services and the Food and Drug Administration (FDA). The Department of Health & Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text. November 3, 2023 HUGER Medical Instrument Co., Ltd. % Diana Hong General Manager Mid-Link Consulting Co., Ltd P.O. Box 120-119 Shanghai, 200120 China Re: K230475 > Trade/Device Name: Medical Video Endoscope System Regulation Number: 21 CFR& 876.1500 Regulation Name: Endoscope and Accessories Regulatory Class: II Product Code: FGB Dated: September 26, 2023 Received: September 27, 2023 # Dear Diana Hong: We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {1}------------------------------------------------ Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download). Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181). Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. {2}------------------------------------------------ Page 3 For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, # Mark J. Antonino -S Mark J. Antonino, M.S. Assistant Director DHT3B: Division of Reproductive, Gynecology and Urology Devices OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {3}------------------------------------------------ # Indications for Use 510(k) Number (if known) K230475 Device Name Medical Video Endoscope System #### Indications for Use (Describe) Medical Video Endoscope System has been designed to be used with endo-therapy accessories such as a biopsy forcens and other ancillary equipment for endoscopic surgery within urinary tract and interior of the kidney. Medical Video Processor is intended to be used in conjunction with the endoscopes to enhance and display the image of the field of view of the endoscope on the monitor, and also provides functional supply for the endoscope. Single-use Video Ureterorenoscope is intended to connect with Medical Video Processor to provide video imaging. | Type of Use (Select one or both, as applicable) | | |-------------------------------------------------|--| |-------------------------------------------------|--| X Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C) ### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Druq Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2023 See PRA Statement below. {4}------------------------------------------------ # Tab #6 510(k) Summary This 510(k) Summary is being submitted in accordance with requirements of Title 21, CFR Section 807.92. The assigned 510(k) Number: - 1. Date of Preparation: 02/16/2023 - 2. Sponsor Identification ## HUGER Medical Instrument Co., Ltd. Building 26A, No. 3825, Xinzhuan Highway, Dongjing Town, Songjiang District, 201619, Shanghai, China. Establishment Registration Number: Not yet registered. Contact Person: Lei Lu Position: Compliance Manager Tel: +86-021-67626235 Fax: +86-021-67691721 Email: lulei@huger.cn - 3. Designated Submission Correspondent Ms. Diana Hong (Primary Contact Person) Ms. Tingting Su (Alternative Contact Person) ## Mid-Link Consulting Co., Ltd. P.O. Box 120-119, Shanghai, 200120, China Tel: +86-21-22815850, Fax: 360-925-3199 Email: info@mid-link.net {5}------------------------------------------------ #### 4. Identification of Proposed Device Trade Name: Medical Video Endoscope System; Common Name: Flexible Urology Endoscope System Regulation Name: Endoscope and Accessories Classification: II Product Code: FGB Regulation Number: 21CFR 876.1500 Review Panel: Gastroenterology/Urology Indications for Use: Medical Video Endoscope System has been designed to be used with endo-therapy accessories such as a biopsy forceps and other ancillary equipment for endoscopic surgery within urinary tract and interior of the kidney. Medical Video Processor is intended to be used in conjunction with the endoscopes to enhance and display the image of the field of view of the endoscope on the monitor, and also provides functional supply for the endoscope. Single-use Video Ureterorenoscope is intended to connect with Medical Video Processor to provide video imaging. Device Description: Medical Video Endoscope System is consist of a Single-use Video Ureterorenoscope and a Medical Video Processor. The proposed device has been designed to be used with endo-therapy accessories such as a biopsy forceps and other ancillary equipment for endoscopy and endoscopic surgery within urinary tract and interior of the kidney. | System | Component | Model | |-----------------------------|-----------------------------------------|---------| | Medical<br>Endoscope System | Video<br>Single-use<br>Ureterorenoscope | SVU-220 | | | Medical Video Processor | VEP-220 | | | | | Table 1 Product Model | | |--|--|--|-----------------------|--| |--|--|--|-----------------------|--| The Single-use Video Ureterorenoscope is a single use device. The Single-use Video Ureterorenoscope is consist of distal end, bending section, insertion portion, operational portion and connection portion. The Single-use Video Ureterorenoscope is a single-channel endoscope. Only one instrument channel is in the distal end of the endoscope. {6}------------------------------------------------ The Single-use Urology Videoscope is sterilized by Ethylene Oxide Gas to achieve a SAL of 106 and supplied sterility maintenance which could maintain the sterility of the device during the shelf life of three vears. The Medical Video Processor is a reusable device. The Medical Video Processor is composed of a power circuit, an image processing circuit and software. - న్. Identification of Predicate Device 510(k) Number: K172098 Product Name: Medical Video Endoscope system - 6. Non-Clinical Test Conclusion Non clinical tests were conducted to verify that the proposed device met all design specifications as was Substantially Equivalent (SE) to the predicate device. The test results demonstrated that the proposed device complies with the following standards: - > ISO 10993-7:2008 Biological evaluation of medical devices - Part 7: Ethylene oxide sterilization residuals - A ASTM F88/F88M-15 Standard test method for seal strength of flexible barrier materials - > ASTM F1929-15 Standard Test Method for Detecting Seal Leaks in Porous Medical Packaging by Dve Penetration - > ISO 8600-1:2015 Endoscopes-Medical endoscopes and endotherapy devices - Part 1: General requirements - A ISO 8600-3:1997/Amd1:2003 Optics and optical instruments -Medical endoscopes and endoscopic accessories - Part 3: Determination of field of view and direction of view of endoscopes with optics - > ISO 8600-4:2014 Endoscopes - Medical endoscopes and certain accessories - Part 4: Determination of maximum width of insertion portion - > ISO 10993-5:2009 Biological evaluation of medical devices - Part 5: Tests for in vitro cytotoxicity - > ISO 10993-10:2010 Biological evaluation of medical devices - Part 10: Tests for irritation and skin sensitization - > IEC 60601-1-2005+CORR.1:2006+CORR.2:2007+AM1:2012, Medical Electrical Equipment Part 1: General requirements for basic safety and essential performance, including the US National Differences - A IEC 60601-1-2:2014 Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance- Collateral standard: Electromagnetic disturbances - Requirements and tests - > IEC 60601-2-18:2009 Medical electrical equipment - Part 2-18: Particular requirements for the basic safety and essential performance of endoscopic equipment {7}------------------------------------------------ - > ASTM D4169-16 Standard Practice for Performance Testing of Shipping Containers and Systems - A ISO 14971:2019 Medical Devices - Application of risk management to medical devices - > IEC 62471:2006 Photobiological safetv of lamps and lamp systems #### Biocompatibilitv testing The contact level of the proposed device is mucosal membrane, and the contact duration is limited contact (<24 hours). The proposed endoscope was evaluated for the following tests. The results of the biocompatibility testing showed that there are no negative impacts from the materials that are used in the proposed device. - > Cytotoxicity, - A Sensitization, - > Intracutaneous irritation, #### Software verification and validation testing Software verification and validation testing were conducted, and documentation was provided as recommended by FDA's Guidance for Industry and FDA Staff. "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices." #### Bench performance testing #### Optical performance testing - > Photobiological safety test according to IEC 62471: 2006; Color reproduction, Resolution, Depth of Field of view, Noise, Geometric distortion, A Intensity uniformity, Image frame frequency and system delay testing compared with the predicate device (K172098). The test results show that the performance of the proposed device is similar to the predicate device. #### Physical/functional performance testing > Irrigation valve leakage test was performed on the proposed device. #### Endoscope and video processor use-life testing The optical performance comparison test was conducted on the un-aged Single-use Video Ureterorenoscope and aged Single-use Video Ureterorenoscope. The test results demonstrate that the optical performance of the aged endoscope is similar as those of the un-aged endoscope. A use-life verification was conducted to determine the image system's use life. After accelerating aging and running, the qualitative was conducted on the video processor. The test results demonstrate that the performance of the proposed system doesn't reduced after accelerating aging and the use-life statement of six years are accepted. In addition, the quantitative optical performance comparison testing was conducted on the new image processor and the image processor after accelerating aged and {8}------------------------------------------------ running. The test results demonstrate that the optical performance of the image process after accelerating aged and running is similar as those of the new one. - 7. Clinical Test Conclusion No clinical study is included in this submission. - Substantially Equivalent (SE) Comparison 8. | ITEM | Proposed Device | Predicate Device<br>K172098 | Remark | |--------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------| | Product Code | FGB | FGB | Same | | Regulation No. | 876.1500 | 876.1500 | Same | | Class | II | II | Same | | Indication for Use | Medical Video Endoscope<br>System has been designed to be used with endo-therapy<br>accessories such as a biopsy<br>forceps and other ancillary<br>equipment for endoscopy and<br>endoscopic surgery within urinary<br>tract and interior of the kidney.<br><br>Medical Video Processor is<br>intended to be used in<br>conjunction with the endoscopes<br>to enhance and display the image<br>of the field of view of the<br>endoscope on the monitor, and<br>also provides functional supply<br>for the endoscope.<br><br>Single-use Video<br>Ureterorenoscope is intended to<br>connect with Medical Video<br>Processor to provide video imaging. | The instrument has been<br>designed to be used with<br>endo-therapy accessories such<br>as a biopsy forceps and other<br>ancillary equipment for<br>endoscopy and endoscopic<br>surgery within urinary tract and<br>interior of the kidney. | Similar | | Single use/ Reuse | Endoscope: Single use | Endoscope: Single use | Same | Table 2 SE Comparison {9}------------------------------------------------ | | Video Processor: Reuse | Image Processor: Reuse | | |------------------------------------------------|-----------------------------------------------------------------|-------------------------------------------------------------------------------|-----------| | Sterile | Yes for disposable endoscope | Yes for disposable endoscope | Same | | Anatomical Site | Urinary tract and interior of the kidney | Urinary tract and interior of the kidney | Same | | Where used | Hospital | Hospital | Same | | Main Configuration | Single-use Video<br>Ureterorenoscope<br>Medical Video Processor | Uscope (single-use endoscope)<br>Eview (touch PC with data processing center) | Similar | | Label/Labeling | Conform with 21CFR Part 801 | Conform with 21CFR Part 801 | Same | | <b>Single-use video Ureterorenoscope</b> | | | | | Scope type | Flexible | Flexible | Same | | Field of view | 120° | 120° | Same | | Direction of view | 0° | 0° | Same | | Depth of field | 2-100mm | 3-50mm | Different | | Sensor type | CMOS | CMOS | Same | | Max. outer<br>diameter of<br>insertion section | 2.9mm | 3.2mm | Similar | | Up/down deflection | Up:270° Down: 270° | Up:270° Down: 270° | Same | | Work length | 700mm | 650mm | Similar | | Minimum<br>instrument channel<br>width | 1.2mm | 1.0mm | Similar | | Product<br>Performance | Comply with ISO 8600-1, ISO 8600-3, ISO 8600-4 and ISO 8600-7 | Comply with ISO 8600-1 | Different | | Patient-contact component and material | | | | | Bending section | Thermoplastic Polyurethane (TPU) | Sheath SUS 304 and PEBAX | | | Insertion section | Thermoplastic Polyurethane (TPU) | Controllable portion SUS 304, covered with silicone rubber | | | Distal End Section | Polycarbonate (PC) | Distal tip ABS | Different | | CMOS Front End | Glass | Working channel Fluorinated ethylene propylene | | | Instrument channel | Pebax and Barium sulfate | / | | | Tuohy Borst Valve | Polycarbonate (PC) and | | | | | | | | | Assembly | Polyethylene copolymer | | | | UV Glue | Epoxy resin | | | | | Cytotoxicity: ISO 10993-5 | Cytotoxicity: ISO 10993-5 | | | Biocompatibility | Sensitization: ISO 10993-10 | Sensitization: ISO 10993-10 | Same | | | Irritation: ISO 10993-10 | Irritation: ISO 10993-10 | | | Sterilization | | | | | Method | EO sterilization | EO sterilization | Same | | SAL | 10-6 | 10-6 | Same | | Medical Video Processor | | | | | Illumination source | LED | LED | Same | | Power supply | AC: 100-240V±10% 50/60 Hz | AC: 100-240V 50/60 Hz | Same | | Dimension | 320(W)×96(H)×370(D) mm | 315(W)×187(H)×308 (D) mm | Different | | Weight | 9Kg | 4.3 Kg | Different | | Input power | 50VA | Unknown | Different | | Video signal output | DVI | HDMI | Different | | Auto white balance | Automatically adjusted | Automatically adjusted | Same | | Communication<br>with endoscope | Provided | Provided | Same | | Electrical Safety | Comply with IEC 60601-1 | Comply with IEC 60601-1 | Same | | EMC | Comply with IEC 60601-1-2 | Comply with IEC 60601-1-2 | Same | | Particular<br>requirements | Comply with IEC 60601-2-18 | Comply with IEC 60601-2-18 | Same | {10}------------------------------------------------ Similar Analysis 1-Indication for Use The indication for use of the proposed device is different from that of the predicate device in expression. The indications for use of the proposed device are provided in the form of system's indications for use, video processor's indications for use and disposable endoscope's indications for use. The predicate device only provides the whole system's indications for use. The expression of the indications for use of the proposed device provides the indications for use of video processor and disposable endoscope. Many cleared endoscope systems adopted this expression form of indications for use. Both devices are used for endoscopic diagnosis and therapies within the urinary system. Based on above analysis, the indication for use of the proposed device and the predicate device is only different in expression. Therefore, the difference will not affect the new safety and effectiveness of the proposed device. #### Similar Analysis 2- Main Configuration The main configuration of the proposed device is different with the predicate device. However, it's just the different name. The Single-use Video Ureterorenoscope of the proposed device and the Uscope of the predicate device have the same function and the Medical Video Processor of the proposed device and the view of the predicate device have function. Therefore, the {11}------------------------------------------------ difference will not affect the safety and effectiveness of the proposed device. #### Different Analysis 3 - Depth of Field The proposed device has a wider range of depth of field than the predicate device to give physicians more options for diagnosis and treatment based on the patient's condition. Therefore, the difference will not affect the safety and effectiveness of the proposed device. #### Similar Analysis 4- Max. outer diameter of insertion section The max. outer diameter of insertion of the proposed device is similar to that of the predicate device. The out diameter of the proposed device is 2.9mm, and the out diameter of the predicate device is 3.2mm. The proposed device was complied with the requirements of ISO 8600 series standards. Therefore, the difference will not affect the safety and effectiveness of the proposed device. #### Similar Analysis 5- Work length The work length of the proposed device is similar as that of the predicate device. The work length of the proposed device is 700mm, and the work length of the predicate device is 650mm. The proposed device was complied with the requirements of ISO 8600 series standards. Therefore, this difference does not raise new problems on safety and effectiveness of proposed device. #### Similar Analysis 6- Min. inner diameter instrument channel The minimum instrument channel width of the proposed device is similar to that of the predicate device. The instrument channel inner diameter of the proposed device is 1.2mm, and the instrument channel inner diameter of the predicate device is 1.0mm. The device was tested per ISO 8600 standard, which can prove that the performance of the proposed endoscope is comply with ISO 8600 standards. Endoscopes with different sizes can be applied to different size of surgical instruments. This difference on minimum instrument channel width between the proposed device and predicate devices does not affect the safety and effectiveness of the proposed device. #### Different Analysis 7- Product Performance The performance tests of the proposed device follow more standards than those of the predicate device, ISO 8600-3, ISO 8600-4 and ISO 8600-7. And the performance test result of the proposed device meets the requirements of ISO 8600-1, ISO 8600-4 and ISO 8600-7. Therefore, the difference will not affect the safety and effectiveness of the proposed device. #### Different Analysis 8- Patient-contact component and material The patient-conduct component and material of the proposed device is different with the predicate device. However, the biocompatibility tests were conducted on the proposed device and the test result showed that the proposed device does not raise the adverse effect on the material. Therefore, the difference will not affect the safety and effectiveness of the proposed device. {12}------------------------------------------------ ### Different Analysis 9-Dimension and weight The dimension and weight for the proposed image processor is different from predicate image processor. However, the dimension and weight difference is just in physical specification and this difference will not raise new issues in safety and effectiveness. #### Different Analysis 10- Input power The input power for the predicate image processor is unknown. However, the input power of proposed image processor complies with IEC 60601-1 standard, the difference in input power is just the difference in device design. Therefore, this difference on input power is considered not affect the safety and effectiveness of the proposed device. #### Different Analysis 11-Video Signal Output The types of video signal output are different between proposed image processor and predicate image processor. The proposed device has DVI interface, the predicate device has HDMI interface. The image quality of DVI output and HDMI output of the proposed device and predicative device have been tested in the quantitative image quality testing. The test results demonstrate that there is no difference in image quality between the two products. Therefore, the difference on video signal output will not affect the safety and effectiveness of the proposed device. #### 9. Substantially Equivalent (SE) Conclusion The proposed device is as safe, as effective, and performs as well as or better than the legally marketed predicate device cleared under K172098.
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...