Proseal Enamel Coating Resin is a light curing material for sealing the enamel pits and fissures of teeth.
Device Story
Proseal Enamel Coating Resin is a light-curing dental sealant composed of methacrylate resin, photo initiator system, and inorganic fillers. It is supplied in syringes or bottles for professional use by dental clinicians. The material is applied to the enamel pits and fissures of teeth; it features a color-change technology, appearing pink upon application to aid in placement accuracy and transitioning to an opaque off-white color upon light-curing. This process provides a physical barrier to prevent dental caries. The device is used in a clinical dental setting to improve oral health outcomes by protecting susceptible tooth surfaces.
Clinical Evidence
No clinical data. Substantial equivalence is supported by in-vitro bench testing, including depth of cure, flexural strength, shear bonding strength, compressive strength, water sorption, solubility, film thickness, and elastic modulus, all meeting ISO 6874:2015 criteria. Biocompatibility testing (cytotoxicity, sensitization, irritation, systemic toxicity, subchronic toxicity, genotoxicity) was performed per ISO 10993.
Technological Characteristics
Flowable paste composed of methacrylate resin, photo initiator, and inorganic fillers. Light-cured. Complies with ISO 6874:2015 (Class 2 sealant), ISO 7405, and ISO 10993-1. Delivered via syringe or bottle.
Indications for Use
Indicated for sealing enamel pits and fissures of teeth in patients requiring dental sealant treatment.
Regulatory Classification
Identification
A pit and fissure sealant and conditioner is a device composed of resin, such as polymethylmethacrylate, intended for use primarily in young children to seal pit and fissure depressions (faults in the enamel) in the biting surfaces of teeth to prevent cavities.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
July 25, 2023
Rizhao HuGe Biomaterials Company, Ltd. Maggie Zheng Regulatory Affairs Manager No.2 North Zhaoyang Road, District of Donggang Rizhao City, Shandong Province 276800 CHINA
Re: K230455
Trade/Device Name: Enamel Coating Resin Regulation Number: 21 CFR 872.3765 Regulation Name: Pit And Fissure Sealant And Conditioner Regulatory Class: Class II Product Code: EBC Dated: April 3, 2023 Received: April 7, 2023
Dear Maggie Zheng:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Michael E. Adjodha -S
Michael E. Adjodha, M.ChE. Assistant Director DHT1B: Division of Dental and ENT Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
#### Indications for Use
510(k) Number (if known) K230455
#### Device Name Proseal Enamel Coating Resin
Indications for Use (Describe)
Proseal Enamel Coating Resin is a light curing material for sealing the enamel pits and fissures of teeth.
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
## K230455
# 005 510 (k) Summary
This summary of 510(k) for the subjective device equivalence information is being submitted in accordance with the requirements of 21 C.F.R. 807.92.
#### 1. Date Summary Prepared: July 21, 2023
#### 2. Submitter Information:
| Owner's Name | Rizhao HuGe Biomaterials Company, Ltd. |
|----------------|-------------------------------------------------------------------------------------------------|
| Address | No.2 North Zhaoyang Road, District of Donggang, Rizhao City,<br>Shandong Province, China 276800 |
| Telephone | 0086 633 2277268 |
| Contact Person | Ms. Maggie Zheng |
| Contact Title | Regulatory Affairs Manager |
| E-mail | zhengxy@hugedent.com |
#### 3. Device Name
Trade name: Proseal Enamel Coating Resin Common name: Dental sealant, pit and fissure sealant Classification name: Sealant, Pit And Fissure, And Conditioner Regulatory Class: II Product Code: EBC
#### 4. Predicate Device Information
| Owner/Operator | Device Trade Name | 510 (k) No. | Product Code | Predicate |
|----------------|-------------------------------|-------------|--------------|-----------|
| 3M COMPANY | 3MTM ESPETM ClinproTM Sealant | K992326 | EBC | Primary |
No reference devices were used in this submission.
#### 5. Description of Device
Proseal Enamel Coating Resin is a light-curing fissure sealant and is mainly composed of methacrylate resin, photo initiator system and inorganic fillers. It is packaged in syringes or
{4}------------------------------------------------
bottles.
### 6. Indications for Use
Proseal Enamel Coating Resin is a light curing material for sealing the enamel pits and fissures of teeth.
### 7. Summary of Physical Properties Tests
Proseal Enamel Coating Resin is classified as Class 2 sealant in accordance with ISO 6874: 2015. The physical properties were determined and tested according to ISO 6874: 2015 and FDA guidance "Guidance for Industry and FDA Staff Dental Composite Resin Devices -Premarket Notification [510(k)]Submissions Document issued on: October 26, 2005".
In-vitro bench tests were performed on the Proseal Enamel Coating Resin. The test results indicated that the Proseal Enamel Coating Resin meets the pass/fail criteria and supports substantial equivalence when compared to the predicate device on physical properties. The main test items inleude depth of cure, flexural strength, shear bonding strength, compressive strength, water sorption, solubility, film thickness and elastic modulus.
The risk analysis of Proseal Enamel Coating Resin was conducted according to ISO 14971 and the residual risk is considered acceptable.
### 8. Technological Characteristics
The following table shows the significant technological characteristics and indicates the similarities and differences with the predicate device:
| | Technological<br>Characteristics | Subject device | Primary predicate device |
|--|---------------------------------------|------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------|
| | | Proseal Enamel Coating Resin(K230455) | 3MTM ESPETM ClinproTM Sealant(K992326) |
| | Physical Form | Flowable Paste | Flowable Paste |
| | Indications of Use | Proseal Enamel Coating Resin is a light<br>curing material for sealing the enamel pits<br>and fissures of teeth. | 3MTM ESPETM ClinproTM Sealant is<br>indicated for:<br>Pit and fissure sealant. |
| | Prescription/over-<br>the-counter use | Prescription | Prescription |
| | Color-change<br>technology | The color-change technology is a unique<br>color-change feature. The product (Type | The color-change technology is a unique<br>color-change feature. The product is pink |
{5}------------------------------------------------
| Technological<br>Characteristics | Subject device | Primary predicate device |
|----------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | Proseal Enamel Coating Resin(K230455) | 3M™ ESPET™ Clinpro™ Sealant(K992326) |
| | F) is pink when applied to the tooth<br>surface, and changes to an opaque<br>off-white color when exposed to light. The<br>pink color aids the dental professional in<br>the accuracy and amount of material<br>placed during the sealant procedure. When<br>light-cured, the pink sealant will transform<br>to an opaque off-white color. The change<br>of color from pink to opaque off-white is<br>not a cure indicator. | when applied to the tooth surface, and<br>changes to an opaque off-white color when<br>exposed to light. The pink color aids the<br>dental professional in the accuracy and<br>amount of material placed during the sealant<br>procedure. When light-cured, the pink<br>sealant will transform to an opaque off-white<br>color. The change of color from pink to<br>opaque off-white is not a cure indicator. |
| Curing method | Light cure | Light cure |
| Delivery form | Syringe or bottle | Syringe or bottle |
| Physical Properties | The subject device and the predicate device have substantially equivalent physical<br>properties as they all meet the criteria per ISO 6874 and FDA Staff Dental Composite<br>Resin Devices -Premarket Notification [510(k)] Submissions Document issued on:<br>October 26, 2005”. | |
| FDA-Recognized<br>Standards | ISO 6874; ISO 7405; ISO 10993-1 | ISO 6874; ISO 7405; ISO 10993-1 |
All compositions of the subject device are based upon industry well-known chemistry. The technological characteristics of the subject device are very similar to those of the predicate device, 3M™ ESPE™ Clinpro™ Sealant (K992326). The subject device is a similar product, manufactured with similar materials and used in the same way by the same types of users and patient populations. The subject device and primary predicate device have minor difference in Composition of Materials. However, the difference does not affect the intended use or substantial equivalence, both products are intended for pit and fissure sealant. Besides, other comparison items such as physical form, curing method, delivery form and physical properties, etc. are the same or very similar. And both products are supplied for prescription use.
#### 9. Summary of Biocompatibility
The subject device, Proseal Enamel Coating Resin, is substantially equivalent to the predicate device.
Biocompatibility tests were performed fully following the ISO 10993 standards. The test items include Cytotoxicity, Sensitization, Irritation, Systemic Toxicity, Subchronic Toxicity and
{6}------------------------------------------------
Genotoxicity.
## 10. Clinical Performance Data
Not applicable. Clinical performance testing has not been performed on the subject device.
#### 11. Conclusions
Based on the indications for use, technological characteristics, performance testing and comparison to predicate device, the subject device is substantially equivalent to the predicate device described herein.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.