aScope™ 5 Broncho is intended for endoscopic procedures and examination within the airways and tracheobronchial tree. aScope™ 5 Broncho is intended to provide visualization via a compatible Ambu displaying unit, and to allow passing of endotherapy instruments via its working channel. The aBox™ 2 is intended to display live imaging data from compatible Ambu visualization devices.
Device Story
System comprises sterile, single-use flexible bronchoscope (aScope 5) and reusable digital monitor (aBox 2). Bronchoscope features distal tip camera module with dual LED illumination; insertion portion enters airway via mouth, nose, or artificial airway. Working channel allows endotherapy instrument passage and fluid instillation; suction system removes blood/mucus. aBox 2 processes and displays live video from endoscope on 12.8" LCD screen; supports snapshot/video recording and external monitor connection. Used in clinical settings by physicians for airway visualization and intervention. Output provides real-time visual feedback to clinician, facilitating diagnostic and therapeutic airway procedures.
Clinical Evidence
Bench testing only. Evaluated per ISO 8600 series (general requirements, field/direction of view, insertion width). Verification tests included bending, suction, duration of use, and optical performance (resolution, color, distortion). Safety validated via IEC 60601-1/2-18 (electrical), IEC 60601-1-2 (EMC), IEC 62471 (photobiological), and ISO 10993-1 (biocompatibility). Sterilization validated per ISO 11135.
Technological Characteristics
Single-use flexible bronchoscope; reusable digital monitor. Features: maneuverable tip, distal LED illumination, rotary function, suction system. Sterilization: Ethylene Oxide. Connectivity: integrated LCD monitor with external monitor output. Software: video processing, brightness/contrast/zoom adjustment, recording/storage.
Indications for Use
Indicated for endoscopic procedures and examination of the airways and tracheobronchial tree in patients requiring visualization and/or use of endotherapy instruments.
Regulatory Classification
Identification
A bronchoscope (flexible or rigid) and accessories is a tubular endoscopic device with any of a group of accessory devices which attach to the bronchoscope and is intended to examine or treat the larynx and tracheobronchial tree. It is typically used with a fiberoptic light source and carrier to provide illumination. The device is made of materials such as stainless steel or flexible plastic. This generic type of device includes the rigid ventilating bronchoscope, rigid nonventilating bronchoscope, nonrigid bronchoscope, laryngeal-bronchial telescope, flexible foreign body claw, bronchoscope tubing, flexible biopsy forceps, rigid biopsy curette, flexible biopsy brush, rigid biopsy forceps, flexible biopsy curette, and rigid bronchoscope aspirating tube, but excludes the fiberoptic light source and carrier.
Ambu® aScope™ 4 Broncho Slim and Regular (K173727)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains two logos. On the left is the Department of Health & Human Services logo. On the right is the FDA logo, which includes the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
August 10, 2023
Ambu A/S % Sanjay Parikh Senior Director, QA/RA Ambu Inc. 6721 Columbia Gateway Drive, Suite 200 Columbia, Maryland 21046
Re: K230428
Trade/Device Name: Ambu® aScope™ 5 Broncho 2.7/1.2, Ambu® aScope™ 5 Broncho 4.2/2.2, Ambu® aBox™ 2 Regulation Number: 21 CFR 874.4680 Regulation Name: Bronchoscope (Flexible Or Rigid) And Accessories Regulatory Class: Class II Product Code: EOO Dated: July 10, 2023 Received: July 11, 2023
Dear Sanjay Parikh:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal
{1}------------------------------------------------
statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Image /page/1/Picture/5 description: The image shows the name "Joyce C. Lin -S" in a large, clear font. The text is horizontally aligned and appears to be a name or title. The background is plain, ensuring the text is easily readable.
for Shu-Chen Peng, Ph.D. Assistant Director DHT1B: Division of Dental and ENT Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number: K230428
Device Name Ambu® aScope™ 5 Broncho 4.2/2.2 Ambu® aScope™ 5 Broncho 2.7/1.2 Ambu® aBox™ 2
Indications for Use (Describe)
aScope™ 5 Broncho is intended for endoscopic procedures and examination within the airways and tracheobronchial tree. aScope™ 5 Broncho is intended to provide visualization via a compatible Ambu displaying unit, and to allow passing of endotherapy instruments via its working channel.
The aBox™ 2 is intended to display live imaging data from compatible Ambu visualization devices.
| Type of Use (Select one or both, as applicable) | |
|----------------------------------------------------------------------------------|--------------------------------------------------------------------------------|
| <div> <span> Prescription Use (Part 21 CFR 801 Subpart D) </span> </div> | <div> <span> Over-The-Counter Use (21 CFR 801 Subpart C) </span> </div> |
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the word "Ambu" in a bold, sans-serif font. The color of the text is a vibrant pink or red. The letters are closely spaced, creating a solid block of text.
# 510(k) Summary – Ambu® aScope™ 5 Broncho System
This 510(k) summary has been prepared in accordance with 21 CFR 807.87(h) and the content and format of the 510(k) summary has been prepared in accordance with 21 CFR 807.92.
| Submitter | Ambu A/S<br>Baltorpbakken 13<br>2750 Ballerup<br>Denmark<br>Tel.: +45 7225 2000<br>Fax.: +45 7225 2050 | |
|-----------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------|
| Contact Person | Name: Karina Matthiesen<br>Job Title: Senior Regulatory Affairs Professional<br>Address: Ambu A/S, Baltorpbakken 13, 2750 Ballerup, Denmark<br>Email: kama@ambu.com<br>Tel.: +45 7225 2000 | |
| Date Summary<br>Prepared | August 10, 2023 | |
| Device Trade<br>Name | Ambu® aScope™ 5 Broncho 2.7/1.2<br>Ambu® aScope™ 5 Broncho 4.2/2.2<br>Ambu® aBox™ 2 | |
| Device Common<br>Name | Endoscopy system | |
| Device<br>Classification | Bronchoscope (flexible or rigid) and accessories<br>Product Codes: EOQ<br>21 CFR 874.4680<br>Class II | |
| Legally<br>Marketed<br>Devices to which<br>the Device is<br>Substantially<br>Equivalent | Ambu® aScope™ 5 Broncho: | Ambu® aBox™ 2: |
| | Predicate Device A (K220606): | Predicate Device A (K220606): |
| | Ambu® aScope™ 5 Broncho HD 5.0/2.2, Ambu® aScope™ 5 Broncho HD 5.6/2.8 | Ambu® aBox™ 2 |
| | Reference Device B (K173727): | |
| | Ambu® aScope™ 4 Broncho Slim and Regular | |
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the word "Ambu" in a bold, sans-serif font. The color of the text is a vibrant pink or red. The letters are closely spaced, creating a compact and easily readable logo.
#### Description of The Ambu® aScope™ 5 Broncho System is a combination of a displaying the Device unit, the Ambu® aBox™ 2, and a compatible endoscope, Ambu® aScope™ 5 Broncho.
Ambu® aScope™ 5 Broncho is a sterile, single-use flexible bronchoscope designed to conduct endoscopic procedures and examination of the airways and tracheobronchial tree. The endoscope is available in two size confiqurations: Ambu® aScope™ 5 Broncho 4.2/2.2 and Ambu® aScope™ 5 Broncho 2.7/1.2. Apart from the size, the endoscopes share a similar design. The insertion portion is inserted into the patient airway through the mouth, nose, endotracheal tube, tracheostomy tube, etc. There is a working channel system within the endoscope for use with endotherapy instruments or instillation of fluids. An introducer (luer lock adaptor), which is supplied together with the endoscopes, can be attached to the working channel port during use. Suctioning of blood, saliva, and mucus from airway is possible through the suction system.
Ambu® aScope™ 5 Broncho features an integrated camera module with built-in dual LED illumination. The image module provides a cropped 400x400 pixels signal from the 160 Kpixel sensor.
The Ambu® aBox™ 2, also referred to as displaying unit, is a non-sterile digital monitor intended to display live imaging data from Ambu visualization devices. The product consists of a base unit with a 12.8″ LCD screen mounted on the top. The device is powered by an integrated power supply and comes with country specific power cables.
The Ambu® aBox™ 2 has the following physical and performance characteristics:
- Displays the image from Ambu® aScope™ 5 Broncho endoscope . on the screen
- . Can record snapshots or video of image from Ambu® aScope™ 5 Broncho endoscope
- Can connect to an external monitor .
- . Is a reusable device
Intended aScope 5 Broncho is intended for endoscopic procedures and examination Use/Indications within the airways and tracheobronchial tree. for Use
aScope 5 Broncho is intended to provide visualization via a compatible Ambu displaying unit, and to allow passing of endotherapy instruments via its working channel.
The aBox™ 2 is intended to display live imaging data from compatible Ambu visualization devices.
{5}------------------------------------------------
Image /page/5/Picture/1 description: The image shows the word "Ambu" in a bold, sans-serif font. The letters are a bright pink color. The word is centered and takes up most of the frame.
| Summary of the<br>technological<br>characteristics<br>in comparison to<br>the predicate<br>devices | Ambu® aScope™ 5 Broncho and the predicate endoscope, Ambu®<br>aScope™ 5 Broncho HD, are both sterile, single-use flexible<br>bronchoscopes. Furthermore, they share the following technological<br>characteristics:<br>Both devices have a maneuverable tip controlled by the user. Both devices have a camera and LED light source located in the<br>distal tip. Both devices have a rotary function enabling the orientation of<br>camera and working channel in relation to the bronchoscope<br>handle to be altered. Both devices have two endoscope buttons to communicate with<br>the displaying unit. Both devices have a suction system activated by a suction button. Both devices are sterilized by Ethylene Oxide. Both devices are compatible with the displaying unit Ambu®<br>aBox™ 2. |
|----------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | Ambu® aScope™ 5 Broncho has the same handle as the predicate device,<br>but the insertion portion differs in the following ways:<br>The outer diameter of the insertion portion is smaller than for the<br>predicate device. The working channel is smaller than for the predicate device. Due to the smaller endoscope tip, the camera technology differs<br>from the predicate e.g., lower resolution. |
| | The Ambu® aBox™ 2 and the predicate displaying unit share the following<br>technological characteristics:<br>Both are video processors displaying live video-imaging data of<br>the connected visualization device to a monitor. Both provide video output formats, recording and data storage<br>and data transport functions. Both share technical functionalities such as brightness control,<br>image contrast and sharpness adjustment and zoom function. Both are portable and have an integrated monitor in addition to<br>the possibility to connect to an external monitor. The differences between the aBox 2 and the predicate device are as<br>follows:<br>The predicate device aBox 2 is not compatible with the applicant<br>device Ambu aScope 5 Broncho |
| Performance<br>Data - Bench | Performance requirements were evaluated in accordance with the ISO<br>8600 series:<br>ISO 8600-1 Endoscopes - Medical endoscopes and endotherapy<br>devices - Part 1: General requirements ISO 8600-3 Endoscopes - Medical endoscopes and endotherapy<br>devices - Part 3: Determination of field of view and direction of<br>view of endoscopes with optics |
{6}------------------------------------------------
Image /page/6/Picture/1 description: The image shows the word "Ambu" in a bold, sans-serif font. The color of the text is a bright pink or red. The letters are closely spaced together, creating a compact and solid appearance.
- ISO 8600-4 Endoscopes Medical endoscopes and endotherapy . devices - Part 4: Determination of maximum width of insertion portion
The following tests were performed to verify/validate the design and evaluate the performance of the Ambu® aScope™ 5 Broncho System:
- . Verification tests including:
- o Insertion portion dimensions
- Bending performance o
- o Suction performance
- Duration of use o
- . Optical performance tests including:
- O Field of view
- Direction of view o
- Depth of field* o
- O Resolution*
- Color performance* O
- o Image intensity uniformity*
- Geometric distortion* O
- Photobiological safety according to IEC 62471 ●
- Transportation study ●
- Sterilization validation according to ISO 11135
- Stability study to document shelf life
- Performance tests o
- Sterile packaging integrity o
- Biocompatibility according to ISO 10993-1 including tests for: ●
- Cytotoxicity (ISO 10993-5) O
- O Irritation (ISO 10993-23)
- Sensitization (ISO 10993-10) o
- Electrical Safety and performance according to IEC 60601-1 and . IEC 60601-2-18
- . Electromagnetic Compatibility according to IEC 60601-1-2
The camera technology of Ambu® aScope™ 5 Broncho differs from that the predicate device, however, the different technological of characteristics do not raise different questions of safety or effectiveness. The safety and effectiveness of the applicant device was ensured by comparing the results of the optical performance tests marked with * to reference device, Ambu® aScope™ 4 Broncho, thereby the demonstrating substantial equivalence.
Overall, the Ambu® aScope™ 5 Broncho System performed as expected and met the test specifications set.
{7}------------------------------------------------
Image /page/7/Picture/1 description: The image shows the word "Ambu" in a bold, sans-serif font. The letters are all capitalized and colored in a bright pink. The word is displayed horizontally and appears to be a logo or brand name.
### Conclusion The Ambu® aScope™ 5 Broncho System, consisting of Ambu® aScope™ 5 Broncho and Ambu® aBox™ 2, has the same intended use/indications for use and similar technological characteristics and principles of operation as the predicate device.
The minor technological differences between Ambu® aScope™ 5 Broncho System and its predicate device do not raise any different questions regarding safety or effectiveness.
Therefore, it is concluded that the Ambu® aScope™ 5 Broncho System is substantially equivalent to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.