← Product Code [HEB](/productcode/HEB) · K230419

# Sequel Tampon with Plastic Applicator (K230419)

_Tampro, Inc. · HEB · Aug 3, 2023 · Obstetrics/Gynecology · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K230419

## Device Facts

- **Applicant:** Tampro, Inc.
- **Product Code:** [HEB](/productcode/HEB.md)
- **Decision Date:** Aug 3, 2023
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 884.5470
- **Device Class:** Class 2
- **Review Panel:** Obstetrics/Gynecology

## Indications for Use

Sequel tampon with plastic applicator is an unscented tampon provided with a plastic applicator used for easing the placement of the tampon correctly into the vagina. The Sequel tampon is inserted into the vagina and used to absorb menstrual or other vaginal discharge.

## Device Story

Single-use, non-sterile menstrual tampon with compact plastic applicator; used for absorption of menstrual or vaginal discharge. Device consists of rayon fiber pledget, polyethylene/polyester overwrap, and cotton/polyester withdrawal cord. Applicator is polyethylene. User inserts tampon into vagina via applicator. Device functions as passive absorbent material. No electronic, software, or algorithmic components. Intended for over-the-counter use.

## Clinical Evidence

Bench testing only. Biocompatibility testing per ISO 10993-1 (cytotoxicity, sensitization, vaginal irritation, acute systemic toxicity) showed no adverse effects. Performance testing per FDA guidance included Syngyna absorbency (6-9g), string strength, fiber shedding, tampon integrity, and microbial testing (Staphylococcus aureus growth, TSST-1 production, and vaginal microflora impact).

## Technological Characteristics

Materials: Rayon fibers (pledget), polyethylene/polyester (overwrap), cotton/polyester (withdrawal cord), polyethylene (applicator). Antiwicking agent: paraffin wax emulsion. Finishing agent: polyglycol ester/fatty alcohol. Design: Compact applicator with plunger; spiral groove tampon. Non-sterile, single-use. No energy source or software.

## Regulatory Identification

An unscented menstrual tampon is a device that is a plug made of cellulosic or synthetic material that is inserted into the vagina and used to absorb menstrual or other vaginal discharge. This generic type of device does not include menstrual tampons treated with scent (i.e., fragrance materials) or those with added antimicrobial agents or other drugs.

## Predicate Devices

- Interlude and private label Unscented 3-piece compact applicator tampon ([K173225](/device/K173225.md))

## Reference Devices

- [K020535](/device/K020535.md)

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

August 3, 2023

Tampro, Inc. % Chris Staab Vice President of Quality/Regulatory Affairs Regulatory and Technical Associates 30 Neck Road Old Lyme, CT 06371

Re: K230419

> Trade/Device Name: Sequel Tampon with Plastic applicator Regulation Number: 21 CFR§ 884.5470 Regulation Name: Unscented menstrual tampon Regulatory Class: II Product Code: HEB Dated: June 28, 2023 Received: June 30, 2023

Dear Chris Staab:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration. listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies.

{1}------------------------------------------------

You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatoryinformation/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

# Monica D. Garcia -S

for

Jason R. Roberts, Ph.D. Assistant Director DHT3B: Division of Reproductive, Gynecology and Urology Devices OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

## Indications for Use

510(k) Number (if known) K230419

Device Name Sequel Tampon with Plastic applicator

Indications for Use (Describe)

Sequel tampon with plastic applicator is an unscented tampon provided with a plastic applicator used for easing the placement of the tampon correctly into the vagina. The Sequel tampon is inserted into the vagina and used to absorb menstrual or other vaginal discharge.

| Type of Use (Select one or both, as applicable) |  |
|-------------------------------------------------|--|
|-------------------------------------------------|--|

Prescription Use (Part 21 CFR 801 Subpart D)

|X | Over-The-Counter Use (21 CFR 801 Subpart C)

### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

## 510(k) Summary K230419

#### Submitter: Tampro Inc. 1.0

| Name:    | Greta Meyer                                                  |
|----------|--------------------------------------------------------------|
| Address: | 3749 Buchanan Street #316<br>San Francisco, California 94123 |
| Phone:   | 215 260 9081                                                 |
| Email:   | Greta@trysequel.com                                          |

Date 510(k) Summary was prepared: July 26, 2023

#### Submission Correspondent: 2.0

| Name:    | Chris Staab                        |
|----------|------------------------------------|
| Address: | 30 Neck Road<br>Old Lyme, CT 06371 |
| Phone:   | 860 575 5462                       |
| Fax:     | 860 434 5892                       |
| Email:   | RTA1ALI1@aol.com                   |

#### Name of the Device 3.0

Trade Name: Sequel Tampon with Plastic applicator

Device Name: Sequel Tampon with Plastic applicator

Common Name: Unscented Menstrual Tampon

Device Class: Class II

Regulation Name: Unscented menstrual tampon

Regulation Number: 21 CFR 884.5470

Product Code: HEB (Tampon, Menstrual, Unscented)

### 4.0 Predicate Device Information

The predicate device is Interlude and private label Unscented 3-piece compact applicator tampon (K173225). This predicate device has not been subject to a design-related recall.

{4}------------------------------------------------

#### 5.0 Description of the Device

#### રે ની Device and Description

The Sequel Tampon with Plastic applicator are menstrual tampons used to absorb menstrual fluid. These Tampons are available in Regular absorbency (6-9 grams).

These Tampons are made from rayon fibers and a polyethylene/polyester overwrap in which a cotton/polyester withdrawal cord is inserted. The applicator is a compact style polyethylene applicator with cylindrical barrel with a plunger. The assembled tampon with applicator is wrapped in a printed polypropylene wrapper. The tampon and applicator are single-use and non-sterile.

### 5.2 Device Materials

The pledget absorbing fibers are made of rayon. The non-woven over-wrap is a bicomponent thermoplastic nonwoven. The withdrawal cord is made from polyester and cotton. Finally, the applicator is made from polyethylene.

The antiwicking agent for the withdrawal cord is a widely used paraffin wax emulsion. The tampon cord contains 2,000 micrograms paraffinic additive/tampon.

The finishing agent for the rayon fibers is a polyglycol ester and fatty alcohol mixture, Leomin or glycerol.

### 6.0 Indications for Use

The subject device has the following indications for use:

"Sequel tampon with plastic applicator is an unscented tampon provided with a plastic applicator used for easing the placement of the tampon correctly into the vagina. The Sequel tampon is inserted into the vagina and used to absorb menstrual or other vaginal discharge."

The predicate device indications for use are:

"Interlude and private label Unscented 3-piece applicator tampons are inserted into the vagina and used to absorb menstrual or other vaginal discharge."

The subject and predicate device have the same intended use, to absorb menstrual or other vaginal discharge. There are no intended use concerns.

### 7.0 Comparison of Technological Characteristics

The table below is a comparison of the subject and predicate device technological characteristics.

{5}------------------------------------------------

| Characteristics                | Predicate Device                                                                                                                                      | Subject Device                                                                                            |
|--------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------|
| Device Name                    | Interlude and Private Label Unscented<br>3 Piece Compact Applicator Tampon                                                                            | Sequel Tampon with Plastic applicator                                                                     |
| Regulation (Product<br>Code)   | 21 CFR 884.5470 (HEB)                                                                                                                                 | 21 CFR 884.5470 (HEB)                                                                                     |
| Manufacturer                   | Albaad fem                                                                                                                                            | Albaad fem                                                                                                |
| 510(k) Number                  | K173225                                                                                                                                               | K230419                                                                                                   |
| 510(k) Owner                   | Albaad fem                                                                                                                                            | Tampro, Inc                                                                                               |
| Device Design                  | Compact style applicator with<br>cylindrical barrel with finger grip and<br>two-piece plunger. Digital tampon with<br>round tip and straight grooves. | Compact style applicator with cylindrical<br>barrel and a plunger. Digital tampon with<br>spiral grooves. |
| Absorbency                     | Regular (6-9 grams)<br>Super (9-12 grams)                                                                                                             | Regular (6-9 grams)                                                                                       |
| Pledget<br>Composition         | Viscose Rayon                                                                                                                                         | Viscose Rayon                                                                                             |
| Withdrawal cord<br>composition | Polyester and Cotton                                                                                                                                  | Polyester and Cotton                                                                                      |
| Overwrap<br>composition        | Polyethylene / Polyethylene<br>Terephthalate Nonwoven                                                                                                 | Polyethylene / Polyester                                                                                  |
| Plastic Applicator             | High Density Polyethylene, Low<br>Density Polyethylene, Polypropylene                                                                                 | Polyethylene                                                                                              |
| Primary Packaging              | Printed Polyethylene                                                                                                                                  | Printed Polypropylene                                                                                     |

The Sequel Tampon with plastic applicator and Interlude and Private Label Unscented3 Piece Compact Applicator Tampon are similar in the pledget composition and components (tampon and applicator). The subject and predicate device differ in the tampon and applicator materials, available absorbencies, as well as the tampon and applicator design. These differences do not raise different questions of safety and effectiveness.

### 8.0 Summary of Performance Testing

The subject device was assessed for biocompatibility in accordance with the 2020 FDA guidance document Use of International Standard ISO 10993-1. "Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process" and ISO 10993-1 (2018). The tampon was assessed for the following endpoints:

{6}------------------------------------------------

- Cytotoxicity ISO 10993-5:2009/(R)2014 ●
- Sensitization (Guinea Pig Maximization Test) – ISO 10993-10:2021
- Vaginal Irritation ISO 10993-23:2021 ●
- Acute Systemic Toxicity - ISO 10993-11:2017

The test results show the tampon is not cytotoxic, non-sensitizing, non-irritating, and does not have acute systemic toxicity. The applicator is the same as the applicator cleared under K020535. Therefore, additional biocompatibility assessments were not completed for the applicator.

The following performance characteristics were assessed in accordance with the 2005 FDA guidance document Menstrual Tampons and Pads: Information for Premarket Notification Submission (510(k)s) - Guidance for Industry and FDA Staff:

- Tampon Absorbency via Syngyna testing, per 21 CFR 801.430 ●
- Assessment of Chemical Residues
- String Strength ●
- . Fiber Shedding
- Tampon Integrity ●
- Microbial Testing to demonstrate the tampon does not: .
	- O enhance the growth of Staphylococcus aureus
	- increase the production of Toxic Shock Syndrome Toxin-1 (TSST-1) O
	- alter the growth of normal vaginal microflora O

### 9. Conclusion

The subject and predicate device have the same intended use. The differences in technological characteristics between the subject and predicate device do not raise different questions of safety and effectiveness. The performance data demonstrate that the Sequel Tampon with Plastic applicator is substantially equivalent to the predicate device.

---

**Source:** [https://fda.innolitics.com/device/K230419](https://fda.innolitics.com/device/K230419)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
