K230419 · Tampro, Inc. · HEB · Aug 3, 2023 · Obstetrics/Gynecology
Device Facts
Record ID
K230419
Device Name
Sequel Tampon with Plastic Applicator
Applicant
Tampro, Inc.
Product Code
HEB · Obstetrics/Gynecology
Decision Date
Aug 3, 2023
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 884.5470
Device Class
Class 2
Indications for Use
Sequel tampon with plastic applicator is an unscented tampon provided with a plastic applicator used for easing the placement of the tampon correctly into the vagina. The Sequel tampon is inserted into the vagina and used to absorb menstrual or other vaginal discharge.
Device Story
Single-use, non-sterile menstrual tampon with compact plastic applicator; used for absorption of menstrual or vaginal discharge. Device consists of rayon fiber pledget, polyethylene/polyester overwrap, and cotton/polyester withdrawal cord. Applicator is polyethylene. User inserts tampon into vagina via applicator. Device functions as passive absorbent material. No electronic, software, or algorithmic components. Intended for over-the-counter use.
Clinical Evidence
Bench testing only. Biocompatibility testing per ISO 10993-1 (cytotoxicity, sensitization, vaginal irritation, acute systemic toxicity) showed no adverse effects. Performance testing per FDA guidance included Syngyna absorbency (6-9g), string strength, fiber shedding, tampon integrity, and microbial testing (Staphylococcus aureus growth, TSST-1 production, and vaginal microflora impact).
Technological Characteristics
Materials: Rayon fibers (pledget), polyethylene/polyester (overwrap), cotton/polyester (withdrawal cord), polyethylene (applicator). Antiwicking agent: paraffin wax emulsion. Finishing agent: polyglycol ester/fatty alcohol. Design: Compact applicator with plunger; spiral groove tampon. Non-sterile, single-use. No energy source or software.
Indications for Use
Indicated for use by individuals requiring absorption of menstrual or other vaginal discharge. No specific age or gender contraindications provided.
Regulatory Classification
Identification
An unscented menstrual tampon is a device that is a plug made of cellulosic or synthetic material that is inserted into the vagina and used to absorb menstrual or other vaginal discharge. This generic type of device does not include menstrual tampons treated with scent (i.e., fragrance materials) or those with added antimicrobial agents or other drugs.
Predicate Devices
Interlude and private label Unscented 3-piece compact applicator tampon (K173225)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
August 3, 2023
Tampro, Inc. % Chris Staab Vice President of Quality/Regulatory Affairs Regulatory and Technical Associates 30 Neck Road Old Lyme, CT 06371
Re: K230419
> Trade/Device Name: Sequel Tampon with Plastic applicator Regulation Number: 21 CFR§ 884.5470 Regulation Name: Unscented menstrual tampon Regulatory Class: II Product Code: HEB Dated: June 28, 2023 Received: June 30, 2023
Dear Chris Staab:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration. listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies.
{1}------------------------------------------------
You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatoryinformation/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# Monica D. Garcia -S
for
Jason R. Roberts, Ph.D. Assistant Director DHT3B: Division of Reproductive, Gynecology and Urology Devices OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K230419
Device Name Sequel Tampon with Plastic applicator
Indications for Use (Describe)
Sequel tampon with plastic applicator is an unscented tampon provided with a plastic applicator used for easing the placement of the tampon correctly into the vagina. The Sequel tampon is inserted into the vagina and used to absorb menstrual or other vaginal discharge.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|--|
|-------------------------------------------------|--|
Prescription Use (Part 21 CFR 801 Subpart D)
|X | Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
## 510(k) Summary K230419
#### Submitter: Tampro Inc. 1.0
| Name: | Greta Meyer |
|----------|--------------------------------------------------------------|
| Address: | 3749 Buchanan Street #316<br>San Francisco, California 94123 |
| Phone: | 215 260 9081 |
| Email: | Greta@trysequel.com |
Date 510(k) Summary was prepared: July 26, 2023
#### Submission Correspondent: 2.0
| Name: | Chris Staab |
|----------|------------------------------------|
| Address: | 30 Neck Road<br>Old Lyme, CT 06371 |
| Phone: | 860 575 5462 |
| Fax: | 860 434 5892 |
| Email: | RTA1ALI1@aol.com |
#### Name of the Device 3.0
Trade Name: Sequel Tampon with Plastic applicator
Device Name: Sequel Tampon with Plastic applicator
Common Name: Unscented Menstrual Tampon
Device Class: Class II
Regulation Name: Unscented menstrual tampon
Regulation Number: 21 CFR 884.5470
Product Code: HEB (Tampon, Menstrual, Unscented)
### 4.0 Predicate Device Information
The predicate device is Interlude and private label Unscented 3-piece compact applicator tampon (K173225). This predicate device has not been subject to a design-related recall.
{4}------------------------------------------------
#### 5.0 Description of the Device
#### રે ની Device and Description
The Sequel Tampon with Plastic applicator are menstrual tampons used to absorb menstrual fluid. These Tampons are available in Regular absorbency (6-9 grams).
These Tampons are made from rayon fibers and a polyethylene/polyester overwrap in which a cotton/polyester withdrawal cord is inserted. The applicator is a compact style polyethylene applicator with cylindrical barrel with a plunger. The assembled tampon with applicator is wrapped in a printed polypropylene wrapper. The tampon and applicator are single-use and non-sterile.
### 5.2 Device Materials
The pledget absorbing fibers are made of rayon. The non-woven over-wrap is a bicomponent thermoplastic nonwoven. The withdrawal cord is made from polyester and cotton. Finally, the applicator is made from polyethylene.
The antiwicking agent for the withdrawal cord is a widely used paraffin wax emulsion. The tampon cord contains 2,000 micrograms paraffinic additive/tampon.
The finishing agent for the rayon fibers is a polyglycol ester and fatty alcohol mixture, Leomin or glycerol.
### 6.0 Indications for Use
The subject device has the following indications for use:
"Sequel tampon with plastic applicator is an unscented tampon provided with a plastic applicator used for easing the placement of the tampon correctly into the vagina. The Sequel tampon is inserted into the vagina and used to absorb menstrual or other vaginal discharge."
The predicate device indications for use are:
"Interlude and private label Unscented 3-piece applicator tampons are inserted into the vagina and used to absorb menstrual or other vaginal discharge."
The subject and predicate device have the same intended use, to absorb menstrual or other vaginal discharge. There are no intended use concerns.
### 7.0 Comparison of Technological Characteristics
The table below is a comparison of the subject and predicate device technological characteristics.
{5}------------------------------------------------
| Characteristics | Predicate Device | Subject Device |
|--------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------|
| Device Name | Interlude and Private Label Unscented<br>3 Piece Compact Applicator Tampon | Sequel Tampon with Plastic applicator |
| Regulation (Product<br>Code) | 21 CFR 884.5470 (HEB) | 21 CFR 884.5470 (HEB) |
| Manufacturer | Albaad fem | Albaad fem |
| 510(k) Number | K173225 | K230419 |
| 510(k) Owner | Albaad fem | Tampro, Inc |
| Device Design | Compact style applicator with<br>cylindrical barrel with finger grip and<br>two-piece plunger. Digital tampon with<br>round tip and straight grooves. | Compact style applicator with cylindrical<br>barrel and a plunger. Digital tampon with<br>spiral grooves. |
| Absorbency | Regular (6-9 grams)<br>Super (9-12 grams) | Regular (6-9 grams) |
| Pledget<br>Composition | Viscose Rayon | Viscose Rayon |
| Withdrawal cord<br>composition | Polyester and Cotton | Polyester and Cotton |
| Overwrap<br>composition | Polyethylene / Polyethylene<br>Terephthalate Nonwoven | Polyethylene / Polyester |
| Plastic Applicator | High Density Polyethylene, Low<br>Density Polyethylene, Polypropylene | Polyethylene |
| Primary Packaging | Printed Polyethylene | Printed Polypropylene |
The Sequel Tampon with plastic applicator and Interlude and Private Label Unscented3 Piece Compact Applicator Tampon are similar in the pledget composition and components (tampon and applicator). The subject and predicate device differ in the tampon and applicator materials, available absorbencies, as well as the tampon and applicator design. These differences do not raise different questions of safety and effectiveness.
### 8.0 Summary of Performance Testing
The subject device was assessed for biocompatibility in accordance with the 2020 FDA guidance document Use of International Standard ISO 10993-1. "Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process" and ISO 10993-1 (2018). The tampon was assessed for the following endpoints:
{6}------------------------------------------------
- Cytotoxicity ISO 10993-5:2009/(R)2014 ●
- Sensitization (Guinea Pig Maximization Test) – ISO 10993-10:2021
- Vaginal Irritation ISO 10993-23:2021 ●
- Acute Systemic Toxicity - ISO 10993-11:2017
The test results show the tampon is not cytotoxic, non-sensitizing, non-irritating, and does not have acute systemic toxicity. The applicator is the same as the applicator cleared under K020535. Therefore, additional biocompatibility assessments were not completed for the applicator.
The following performance characteristics were assessed in accordance with the 2005 FDA guidance document Menstrual Tampons and Pads: Information for Premarket Notification Submission (510(k)s) - Guidance for Industry and FDA Staff:
- Tampon Absorbency via Syngyna testing, per 21 CFR 801.430 ●
- Assessment of Chemical Residues
- String Strength ●
- . Fiber Shedding
- Tampon Integrity ●
- Microbial Testing to demonstrate the tampon does not: .
- O enhance the growth of Staphylococcus aureus
- increase the production of Toxic Shock Syndrome Toxin-1 (TSST-1) O
- alter the growth of normal vaginal microflora O
### 9. Conclusion
The subject and predicate device have the same intended use. The differences in technological characteristics between the subject and predicate device do not raise different questions of safety and effectiveness. The performance data demonstrate that the Sequel Tampon with Plastic applicator is substantially equivalent to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.