K230304 · Puyang Linshi Medical Supplies Co., Ltd. · KGO · Aug 9, 2023 · General, Plastic Surgery
Device Facts
Record ID
K230304
Device Name
Polyisoprene Surgical Gloves
Applicant
Puyang Linshi Medical Supplies Co., Ltd.
Product Code
KGO · General, Plastic Surgery
Decision Date
Aug 9, 2023
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4460
Device Class
Class 1
Indications for Use
The Polyisoprene Surgical Gloves are sterile and single use device intended to be worn on the hands of operating room personnel to protect a surgical wound from contamination.
Device Story
Polyisoprene surgical gloves; sterile, single-use, powder-free, hand-specific, beaded cuff design. Used by operating room personnel to provide a protective barrier between the wearer's hands and a surgical wound. Device acts as a physical barrier to prevent contamination. No electronic, software, or algorithmic components. Device is a passive medical supply.
Clinical Evidence
No clinical data. Bench testing only, including biocompatibility (ISO 10993-5, -7, -10, -11, -23), residual powder (ASTM D6124), hole detection (ASTM D5151), protein content (ASTM D5712), and physical properties/dimensions (ASTM D3577-19).
Technological Characteristics
Material: Synthetic polyisoprene rubber. Design: Sterile, powder-free, hand-specific, beaded cuff. Dimensions and physical properties meet ASTM D3577-19. Sterilization: Ethylene Oxide (EO) per ISO 11135:2014. SAL 10^-6.
Indications for Use
Indicated for use by operating room personnel to protect surgical wounds from contamination. Suitable for sterile, single-use applications.
Regulatory Classification
Identification
A non-powdered surgeon's glove is a device intended to be worn on the hands of operating room personnel to protect a surgical wound from contamination. A non-powdered surgeon's glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
August 9, 2023
Puyang Linshi Medical Supplies Co., Ltd. % Boyle Wang General Manager Shanghai Truthful Information Technology Co., Ltd. Room 608, No.738, Shangcheng Rd., Pudong Shanghai, Shanghai 200120 China
Re: K230304
Trade/Device Name: Polyisoprene Surgical Gloves Regulation Number: 21 CFR 878.4460 Regulation Name: Non-Powdered Surgeon's Glove Regulatory Class: Class I, reserved Product Code: KGO Dated: July 8, 2023 Received: July 11, 2023
Dear Boyle Wang:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal
{1}------------------------------------------------
statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# Allan Guan -S
For Bifeng Qian, M.D., Ph.D. Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K230304
Device Name Polyisoprene Surgical Gloves
Indications for Use (Describe)
The Polyisoprene Surgical Gloves are sterile and single use device intended to be worn on the hands of operating room personnel to protect a surgical wound from contamination.
Type of Use (Select one or both, as applicable)Prescription Use (Part 21 CFR 801 Subpart D)
X Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary (K230304)
This summary of 510(k) safety and effectiveness information is being submitted in accordance with 21 CFR 807.92.
# 1.0 submitter's information
Name: Puyang Linshi Medical Supplies Co., Ltd. Address: East of Panjin Road and North of Fumin Road in Puyang County, Puyang City, Henan Province 457001, China. Phone Number: +86-19839327898 Contact: Catherine Liu Date of Preparation: 08.01.2023
# Designated Submission Correspondent
Mr. Boyle Wang Shanghai Truthful Information Technology Co., Ltd. Tel: +86-21-50313932 Email: Info@truthful.com.cn
# 2.0 Device information
Trade name: Polyisoprene Surgical Gloves Common name: Surgeon's Gloves Classification name: Surgeon's Gloves Model(s): 6.5, 7, 7.5, 8
# 3.0 Classification
Production code: KGO Regulation number: 21CFR878.4460 Classification: Class I Panel: General Hospital
# 4.0 Predicate device information
Better Care Plastic Technology Co., Ltd. Manufacturer: Sterile Polyisoprene Powder Free Surgical Gloves Device: 510(k) number: K171047
# 5.0 Indications for use
{4}------------------------------------------------
The Polyisoprene Surgical Gloves are sterile and single use device intended to be worn on the hands of operating room personnel to protect a surgical wound from contamination.
### 6.0 Device description
The proposed device is Polyisoprene Surgical Gloves, sterile and disposable devices. The proposed devices are made of polyisoprene. The proposed device is white. The design of proposed device is addressing the standards as ASTM D6124, ASTM D5151, and ASTM D3577.
# 7.0 Summary comparing technological characteristics with predicate device
| Item | Proposed device | Predicated device | Remark |
|-----------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------|
| 510(k) number | K230304 | K171047 | |
| Product Code | KGO | KGO | Same |
| Regulation No. | 21CFR878.4460 | 21CFR878.4460 | Same |
| Class | I | I | Same |
| Indications<br>for Use | The Polyisoprene Surgical Gloves are sterile and single use device intended to<br>be worn on the hands of<br>operating room personnel to<br>protect a surgical wound<br>from contamination. | This surgeon's glove is a sterile and single use device intended to<br>be worn on the hands of<br>operating room personnel to<br>protect a surgical wound from<br>contamination. | Same |
| Prescription or Over<br>The Counter Use | Over-The-Counter-Use | Over-The-Counter-Use | Same |
| Materials | Synthetic polyisoprene rubber | Synthetic polyisoprene rubber | Same |
| Design | Single use<br>Sterile<br>Powder-free<br>Hand Specific<br>Beaded cuff | Single use<br>Sterile<br>Powder-free<br>Hand Specific<br>Beaded cuff | Same |
| Color | White | Clear | Difference 1 |
| Dimensions and<br>physical properties | Meets ASTM D3577- 2019 | Meets ASTM D3577-09 (2015) | Difference 2 |
| Sterilization<br>method | EO Sterilization | Radiation | Difference 3 |
| Sterility Assurance | 10-6 SAI | 10-6 SAI | Same |
#### Table1-General Comparison
{5}------------------------------------------------
| Level (SAL) | | | |
|----------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------|--------------|
| Freedom from<br>holes | Meets ASTM D3577- 2019 Clause<br>8.3 (ASTM D5151-19)<br>Inspection level/AQL: GI/AQL 1.5 | Meets ASTM D3577- 09(2015)<br>Inspection level/AQL: GI/AQL 1.5 | Difference 2 |
| Powder-Free | Meets ASTM D 6124-06 The<br>averaged residual powder content<br>for the glove during process<br>validation is 0.1mg per glove | Meets ASTM D 6124-06 The<br>averaged residual powder<br>content for the glove<br>during process validation is<br>0.16mg per glove | Similar |
| Primary Skin<br>Irritation<br>ISO 10993- 10:2010 | Under the conditions of the<br>study (per ISO 10993-23), the<br>device is not an irritant | Under the conditions of the<br>study (per ISO 10993-10), the<br>device is not an irritant | Difference 4 |
| Dermal<br>Sensitization -<br>ISO 10993-10:2010 | Under the conditions of the<br>study (per ISO 10993-10), not<br>a sensitizer | Under the conditions of the<br>study (per ISO 10993-10), not<br>a sensitizer | Same |
| Acute Systemic<br>Toxicity - ISO<br>10993-11: 2006 | Under the conditions of the study,<br>there was no mortality or evidence<br>of Acute systemic<br>toxicity | Under the conditions of the study,<br>there was no mortality or<br>evidence of Acute systemic<br>toxicity | Same |
| Shelf Life | 2 years | 3 years | Difference 5 |
#### Analysis:
Difference 1: The colors of proposed device are different with those of the predicate, but they all meet the requirements of ASTM D3577-19, so the differences do not raise any new safety or performance questions.
Difference 2: The proposed device and the predicate device meet different requirements ASTM D3577-19 and ASTM D3577-09 as the ASTM D3577-19 is the only Consensus Standard for Rubber Surgical Gloves Recognized by FDA, but the differences do not raise any new safety or performance questions.
Difference 3: The proposed device and the predicate device sterilized by different methods EO Sterilization and Radiation, but the EO validation of proposed device was implemented based on ISO 11135:2014, including sterilizer installation, OQ, PPQ and MPQ of sterilization. And parameters in Sterilization Validation Report can prove these products being SAL of 10-6.
Difference 4: New standard ISO10993-23:2021 replaces the Skin Irritation test in ISO10993-10:2010. Difference 5: The shelf life of proposed device is verified by Product Performance Test Report after 2 Years Accelerated Aging.
{6}------------------------------------------------
# 8.0 Summary of Non-Clinical Performance Testing
The following performance data has been provided to demonstrate that the subject device meets the acceptance criteria in the standard.
| No. | Name of the Test<br>Methodology / Standard | Purpose | Acceptance Criteria | Results |
|-----|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 1 | ISO 10993-10:2010<br>Biological Evaluation Of<br>Medical Devices - Part<br>10: Tests For Skin<br>Sensitization. | This part of ISO<br>10993 assesses<br>possible contact<br>hazards released<br>from medical devices, | Skin Sensitization Test:<br>provided<br>grades less than 1,<br>otherwise sensitization. | All grades are 0.<br><br>All animals were survived, and no abnormal signs were observed during the study. |
| 2 | ISO 10993-23:2021<br>Biological evaluation of<br>medical devices - Part<br>23: Tests for irritation | which may produce<br>skin and mucosal<br>irritation, eye irritation<br>or skin sensitization. | Skin Irritation Test:<br>If the primary irritation<br>index is 0-0,4, the<br>response category is<br>Negligible.<br>0,5-1,9 means slight<br>2-4,9 means moderate<br>5-8 means severe | The primary irritation index is 0.<br><br>The response of the proposed device was categorized as negligible under the test condition |
| 3 | ISO 10993-5:2009<br>Biological Evaluation Of<br>Medical Devices - Part<br>5: Tests For In Vitro<br>Cytotoxicity | This part of ISO<br>10993 describes test<br>methods to assess<br>the in vitro cytotoxicity<br>of medical devices. | The viab.% of the 100%<br>extract of the test article<br>is the final result, and if<br>viability is reduced to<br><70% of the blank, it<br>has cytotoxic potential. | Viab.% of 100% test article<br>extract is 6.1%<br><br>It means the proposed<br>device have potential toxicity<br>to L-929 in the MTT method |
| 4 | ISO 10993-11: 2017<br>Biological evaluation of<br>medical devices — Part<br>11: Tests for systemic<br>toxicity | To evaluate the<br>potential for medical<br>device materials to<br>cause adverse<br>systemic reactions. | Within the monitoring<br>period (72 h), if the<br>toxicosis response of<br>testing group is not<br>greater than that of<br>control group, the<br>testing sample is<br>regarded as acceptable. | There was no evidence of<br>systemic toxicity from the<br>extract. |
| 5 | ISO 10993-7 standards<br>for EO/ECH residual<br>testing | This part of ISO<br>10993 specifies<br>allowable limits for<br>residual ethylene<br>oxide (EO) | Limit<br>(< 24 h) EO 4 mg<br>ECH 9 mg<br>Prolonged<br>(> 24 h < 30 d) EO 60<br>mg/30 d<br>ECH 60 mg/30 d | EO residue: ≤ 4mg in<br>the first 24h; ≤ 60mg in<br>the first 30d; EO<br>residue shall also ≤<br>10ug/g;<br>ECH residue: ≤ 9mg in<br>the first 24h; ≤ 60mg in<br>the first 30d. |
| 6 | ASTM D6124-06<br>(Reapproved 2017),<br>Standard Test Method<br>for Residual Powder on<br>Medical Gloves | This standard is<br>designed to<br>determine the amount<br>of residual powder (or<br>filter-retained mass)<br>found<br>on medical gloves | powder residue limit of<br>2.0 mg | 0.1mg/glove |
| 7 | ASTM<br>D5151-19, Standard Test<br>Method for Detection of<br>Holes in Medical Gloves.<br>12/23/2019 | This test method<br>covers the detection<br>of holes in<br>medical gloves. | Samples number: 200<br>gloves<br>AQL: 1.5 (ISO 2859)<br>Criterion ≤7 gloves for<br>water leakage | 0 glove water leakage found |
| 8 | ASTM D5712-15<br>Standard Test Method<br>for Analysis of Aqueous<br>Extractable Protein in<br>Natural Rubber in Latex,<br>Natural Rubber, and<br>Elastomeric Products<br>Using the Modified<br>Lowry Method | This test method, for<br>the determination of<br>protein levels<br>in latex, is primarily<br>intended to test NR,<br>latex, and elastomeric<br>materials for residual<br>protein content. | Have a recommended<br>aqueous soluble protein<br>content limit of 200<br>ug/dm² | 52ug/dm² |
| 9 | ASTM<br>D3577-19, Standard<br>Specification for Rubber<br>Surgical Gloves | The specification is<br>intended as a<br>reference to the<br>performance and<br>safety of rubber<br>surgical gloves. The<br>safe and proper use<br>of rubber surgical<br>gloves is beyond the<br>scope of this<br>specification. | Dimensions:<br>-6.5: width 83±6mm<br>Length ≥265 mm<br>-7: width 89±6mm<br>Length ≥265 mm<br>-7.5: width 95±6mm<br>Length ≥265 mm<br>-8: width 102±6mm<br>Length ≥265 mm<br>Thickness:<br>-Finger ≥0.10 mm<br>-Palm ≥0.10 mm<br>-Cuff ≥0.10 mm<br><br>Physical properties:<br>● Before aging<br>● Tensile strength<br>≥17MPa<br>● Ultimate<br>● Elongation ≥<br>650%<br>● Stress at 500%<br>Elongation ≤ 7.0<br>MPa<br>● After Accelerated<br>Aging<br>● Tensile strength<br>≥12MPa<br>● Ultimate Elongation<br>≥ 490% | Dimensions:<br>6.5: width: 83-84 mm<br>Length 280-289 mm<br>Thickness:<br>Finger 0.230-0.260 mm<br>Palm 0.188-0.207 mm<br>Cuff 0.137-0.151 mm<br>7: width 91-93 mm<br>Length 271-278 mm<br>Thickness:<br>Finger 0.211-0.241 mm<br>Palm 0.181-0.193 mm<br>Cuff 0.133-0.144 mm<br>7.5: width 96-98 mm<br>Length 273-280 mm<br>Thickness:<br>Finger 0.218-0.237mm<br>Palm 0.180-0.192 mm<br>Cuff 0.136-0.146 mm<br>8: width 103-105 mm<br>Length 268-283 mm<br>Thickness:<br>Finger 0.221-0.272 mm<br>Palm 0.189-0.212 mm<br>Cuff 0.143-0.153 mm<br><br>Physical properties: Before<br>aging<br>-Tensile strength 25.2-30.9<br>MPa<br>-Ultimate Elongation 721% -<br>777%<br>-Stress at 500% Elongation<br>6.5-7.0 MPa<br>After Accelerated Aging<br>-Tensile strength 23.0-29.0<br>MPa<br>-Ultimate Elongation 680% -<br>716% |
# Table 2 Summary of Non-Clinical Performance Testing
{7}------------------------------------------------
{8}------------------------------------------------
{9}------------------------------------------------
# 9. Summarv of Clinical Performance Test
No clinical study is included in this submission.
# 10.0 Conclusion
The conclusions drawn from the nonclinical tests demonstrate that the proposed device is as safe, as effective, and performs as well as or better than the legally marketed predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.