K230276 · Techfit Digital Surgery · DZJ · Jun 23, 2023 · Dental
Device Facts
Record ID
K230276
Device Name
TECHFIT DISRP® System
Applicant
Techfit Digital Surgery
Product Code
DZJ · Dental
Decision Date
Jun 23, 2023
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 872.4120
Device Class
Class 2
Indications for Use
TECHFIT Digitally Integrated Surgical Reconstruction Platform (DISRP) System is intended for use as a software system and image segmentation system for the transfer of imaging information from a medical scanner such as a CT based system. The input data file is processed by the DISRP System and the result is an output data file that may then be provided as digital models or used as input to a rapid prototyping portion of the system that produces physical outputs including surgical guides and splints for use in maxillofacial surgery. The DISRP System is also intended as a preoperative software tool for simulating / evaluating surgical treatment options. The DISRP system is compatible with the TECHFIT Patient- Specific Maxillofacial System and the TECHFIT Diagnostic Models and should be used in conjunction with expert clinical judgment.
Device Story
Web-based collaboration platform for orthognathic surgery planning; manages case flow between surgeons, sales reps, and TECHFIT planning staff. Inputs: DICOM data from CT/CBCT scanners; STL surface scan data. Processing: 3D visualization, bone fragment manipulation (translation/rotation), and surgical simulation. Outputs: Digital models (STL) for rapid prototyping of patient-specific surgical guides and splints; osteosynthesis plate lists. Guides are either 3D-printed (Biomed Clear Resin) or machined (CP Titanium Grade 4). Used in clinical settings; operated by surgeons and planning staff. Benefits: Facilitates precise surgical planning, improves transfer of pre-surgical plans to the OR, and assists in correct bone/implant positioning.
Clinical Evidence
Bench testing only. Includes sterilization validation (AAMI/ISO 17665-1, ANSI/AAMI/ISO 14937), dimensional validation, mechanical testing (bending/compressive strength), compatibility testing with maxillofacial systems, and comprehensive biocompatibility testing (ISO 10993-5, -10, -11, -18, -17, -3) for both resin and titanium components. Software verified and validated per IEC 62304.
Technological Characteristics
Web-based software platform; 3D printing (SLA) using Biomed Clear Resin; subtractive machining of CP Titanium Grade 4. Connectivity: Networked/Cloud-based. Sterilization: Steam (non-sterile delivery). Software: Rule-based planning and segmentation algorithms. Dimensions: Patient-specific.
Indications for Use
Indicated for adults undergoing maxillofacial surgery, including Le Fort and Genioplasty procedures, requiring surgical guides and splints for bone positioning and osteotomy guidance.
Regulatory Classification
Identification
A bone cutting instrument and accessories is a metal device intended for use in reconstructive oral surgery to drill or cut into the upper or lower jaw and may be used to prepare bone to insert a wire, pin, or screw. The device includes the manual bone drill and wire driver, powered bone drill, rotary bone cutting handpiece, and AC-powered bone saw.
Predicate Devices
KLS Martin Individual Patient Solutions (IPS) Planning System (K182789)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
June 23, 2023
TECHFIT Digital Surgery Angela María Lema-Pérez Regulatory Affairs Specialist 1511 Aviation Center Pkwy Suite 220H Daytona Beach, Florida 32114
Re: K230276
Trade/Device Name: TECHFIT DISRP® System Regulation Number: 21 CFR 872.4120 Regulation Name: Bone Cutting Instrument and Accessories Regulatory Class: Class II Product Code: DZJ, LLZ Dated: May 30, 2023 Received: May 30, 2023
Dear Angela María Lema-Pérez:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part
{1}------------------------------------------------
801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# Bobak Shirmohammadi -S
For Michael E. Adjodha, M.ChE., CQIA Assistant Director DHT1B: Division of Dental and ENT Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K230276
Device Name TECHFIT DISRP® SYSTEM
### Indications for Use (Describe)
TECHFIT Digitally Integrated Surgical Reconstruction Platform (DISRP) System is intended for use as a software system and image segmentation system for the transfer of imaging information from a medical scanner such as a CT based system. The input data file is processed by the DISRP System and the result is an output data file that may then be provided as digital models or used as input to a rapid prototyping portion of the system that produces physical outputs including surgical guides and splints for use in maxillofacial surgery. The DISRP System is also intended as a preoperative software tool for simulating / evaluating surgical treatment options.
The DISRP system is compatible with the TECHFIT Patient- Specific Maxillofacial System and the TECHFIT Diagnostic Models and should be used in conjunction with expert clinical judgment.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|--|
|-------------------------------------------------|--|
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for Techfit Digital Surgery. The logo features a blue sphere with a gray teardrop shape overlapping it on the left side. To the right of the sphere, the word "TECHFIT" is written in gray, block letters. Below "TECHFIT", the words "Digital Surgery" are written in a smaller, gray font.
#### K230276 - 510(k) summary 1
The 510(k) summary is provided on the following page per 21 CFR 807.92(c).
# 510(k) summary
The following section is included as required by the Safe Medical Devices Act (SMDA) of 1990 and 21CFR 807.92
#### 1.1 Submitter information
Table 1. Submitter information
| Company name | TECHFIT Digital Surgery |
|-----------------------------------|-----------------------------------------|
| Establishment registration number | 3004187715 |
| Street Address | 1511 Aviation Center Pkwy<br>Suite 220H |
| City | Daytona Beach |
| Zip code | FL 32114 |
| Country | United States |
| Phone number | +57 604 322-33-75 Ext. 165 |
| Fax number | +57 604 3383013 |
| Principal contact person | Angela María Lema-Pérez |
| Contact title | Regulatory Affairs Specialist |
| Contact e-mail address | angela.lema@techftids.com |
| Additional contact person | Susana Muñoz-Cuartas |
| Contact title | Regulatory Affairs Specialist |
| Contact e-mail address | susana.munoz@techfitds.com |
#### 1.2 Date Prepared
June 22, 2023.
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for Techfit Digital Surgery. The logo consists of a blue sphere with a white semi-oval shape partially covering it on the left side. To the right of the sphere are the words "TECHFIT" in a simple, sans-serif font, with the words "Digital Surgery" in a smaller font below.
#### 1.3 Submission information
| Table 2. Submission information | |
|------------------------------------------|-----------------------|
| Trade name | TECHFIT DISRP® SYSTEM |
| Common or Usual name | DISRP |
| Classification name | 21 CFR 872.4120 |
| Product code (classification regulation) | DZJ, LLZ |
| Classification Panel | Dental |
| Device class | Class II |
#### 1.4 Predicate device
The primary devices to which substantial equivalence is claimed to:
| Table 3. Primary predicate device | |
|----------------------------------------------------------------------------------|---------------------------------------------------------------|
| Primary predicate: KLS Martin Individual Patient Solutions (IPS) Planning system | |
| Trade or proprietary or model name | KLS Martin Individual Patient Solutions (IPS) Planning System |
| 510(k) number | K182789 |
| Decision date | 2019/03/11 |
| Product code | DZJ, LLZ |
| Manufacturer | KLS-Martin L.P. |
# Table 3. Primary predicate device
#### 1.5 Reference devices
The general information of the reference devices is the following:
| | | | Table 4. Reference Device 1 | | |
|--|--|--|-----------------------------|--|--|
|--|--|--|-----------------------------|--|--|
| Secondary predicate: IPS CaseDesigner 2.0 | |
|-------------------------------------------|----------------------|
| Trade or proprietary or model name | IPS CaseDesigner 2.0 |
| 510(k) number | K200810 |
| Decision date | 2020/10/08 |
| Product code | LLZ |
| Manufacturer | Nobel Biocare AB |
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the logo for Techfit Digital Surgery. The logo consists of a blue sphere with a gray, tear-drop-shaped object overlapping it on the left side. To the right of the sphere are the words "TECHFIT" in gray, block letters, with the words "Digital Surgery" in a smaller font below.
# Table 5. Reference Device 2
| Reference device 1: TECHFIT Patient-Specific Maxillofacial System | |
|-------------------------------------------------------------------|-----------------------------------------------|
| Trade or proprietary or model name | TECHFIT Patient-Specific Maxillofacial System |
| 510(k) number | K203282 |
| Decision date | 2021/05/19 |
| Product code | JEY |
| Manufacturer | TECHFIT Digital Surgery INC. |
# Table 6. Reference Device 3
| Reference device 2: AFFINITY Proximal Tibia System | |
|----------------------------------------------------|-----------------------------------|
| Trade or proprietary or model name | AFFINITY Proximal Tibia System |
| 510(k) number | K220199 |
| Decision date | 2022/03/21 |
| Product code | HRS, HWC |
| Manufacturer | Industrias Medicas Sampedro S.A.S |
#### 1.6 Device Information
Table 7 presents the device information of the subject device and predicate devices.
| | Subject device:<br>TECHFIT<br>SYSTEM<br>DISRP® | Primary predicate device:<br>KLS Martin Individual<br>Patient Solutions (IPS)<br>Planning System<br>(K182789) |
|----------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------|
| Product code | DZJ, LLZ | DZJ, LLZ |
| Classification | 21 CFR 872.4120, Class II<br>21 CFR 892.2050, Class II | 21 CFR 872.4120,<br>Class II<br>21 CFR 892.2050,<br>Class II |
| Intended Use | TECHFIT Digitally Integrated<br>Surgical Reconstruction<br>Platform (DISRP) System is<br>intended for use as a software<br>system and image<br>segmentation system for the<br>transfer of imaging<br>information from a medical | The KLS Martin<br>Individual Patient<br>Solutions (IPS)<br>Planning System is<br>intended for use as a<br>software system and<br>image segmentation |
# Table 7. Device information
510(k) Premarket Notification
TECHFIT DIGITAL SURGERY
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the logo for Techfit Digital Surgery. The logo features a blue and gray sphere on the left, followed by the word "TECHFIT" in gray, sans-serif font. Below "TECHFIT" is the phrase "Digital Surgery" in a smaller, gray, sans-serif font. The logo is clean and modern, and the colors are professional and calming.
#### 1.6.1 Device Description
# The TECHFIT DISRP SYSTEM is composed by Orthognathic Surgical Guides and the Digitally Integrated Surgical Reconstruction Platform (DISRP).
# Digitally Integrated Surgical Reconstruction Platform (DISRP)
The Digitally Integrated Surgical Reconstruction Platform (DISRP) is a web-based collaboration software for digital surgery case flow management and Orthognathic surgery planning that reflects the production process and allows for the interaction of multiple users: surgeons, sales representatives, and the TECHFIT case planning staff (case planning assistant, case planners, and operations director), using multiple devices. It allows easy collaboration in the planning process. Being web-based allows immediate and convenient sharing without the installation or maintenance of the application at the user's end.
# Orthognathic Surgical Guides
| 510(k) Premarket Notification | TECHFIT DIGITAL SURGERY |
|-------------------------------|-------------------------|
|-------------------------------|-------------------------|
{7}------------------------------------------------
Image /page/7/Picture/1 description: The image shows the logo for Techfit Digital Surgery. The logo consists of a blue sphere with a smaller, gray sphere partially overlapping it on the left side. To the right of the spheres, the word "TECHFIT" is written in gray capital letters, with the words "Digital Surgery" written in a smaller font size below it.
Orthognathic Surgical Guides are Patient-Specific single use devices that are designed to assist the surgeon in transferring the pre-surgical plan to the operation room. Surgical Guides are intended for Orthognathic surgeries in adults and have drilling holes and slots for making drillings and osteotomies, as well as they guide the correct positioning of bones and implants.
Orthognathic Surgical Guided into two types: Resin Orthognathic Surgical Guides and Machined Orthognathic Surgical Guides.
- Resin Orthognathic Surgical Guides ●
Resin Orthognathic Surgical Guides include surgical guides and splints.
Surgical Guides consist of the Le Fort and Genioplasty surgical quides, which are composed of a body that is manufactured by TECHFIT Digital Surgery and produced by rapid prototyping with the Form 3B printer and Biomed Clear Resin from Formlabs, Somerville, United States.
In surgery Surgical Guides must be used with Metal Sleeves. There are three types of metal sleeves: slot metal sleeve, drill metal sleeve and screw metal sleeve. The slot, drill and screw metal sleeves are manually fitted by the healthcare professional during surgery into the slots, drilling holes and fixation holes of the surgical guide. The metal sleeves are produced from commercially pure titanium grade 4 through machining and are manufactured equivalent to the TECHFIT Patient-Specific Maxillofacial System (K203282) and AFFINITY Proximal Tibia System (K220199). In addition, the Metal Sleeves are single-use and patient-specific accessories.
The Splints are optional quides used in orthognathic surgery to quide to the correct teeth positioning and to validate the patient's final occlusion. The Splints are manufactured by TECHFIT Digital Surgery and produced by rapid prototyping using the Form 3B printer and Biomed Clear Resin from Formlabs, Somerville, United States.
- Machined Orthognathic Surgical Guides .
Machined Orthognathic Surgical Guides consist of the Le Fort and Genioplasty surgical guides that are manufactured from the same material (commercially pure titanium grade 4) and in a manufacturing process equivalent to the TECHFIT Patient-Specific Maxillofacial System (K203282) and AFFINITY Proximal Tibia System (K220199) manufactured through machining.
Machined Orthognathic Surgical Guides and Metal Sleeves are manufactured with the same material and same manufacturing process.
# 1.6.2 Indications for use
TECHFIT Digitally Integrated Surgical Reconstruction Platform (DISRP) System is intended for use as a software system and image segmentation system for the transfer of imaging information from a medical scanner such as a CT based system. The input data file is processed by the DISRP System and the result is an output data file that may then be provided as digital models or used as input to a rapid prototyping portion of the system that produces physical outputs including
{8}------------------------------------------------
Image /page/8/Picture/0 description: The image shows the logo for Techfit Digital Surgery. The logo features a blue sphere with a gray teardrop shape overlapping it on the left side. To the right of the sphere, the word "TECHFIT" is written in gray capital letters, with the words "Digital Surgery" written in a smaller font size below it.
surgical guides and splints for use in maxillofacial surgery. The DISRP System is also intended as a pre-operative software tool for simulating / evaluating surgical treatment options.
The DISRP system is compatible with the TECHFIT Patient- Specific Maxillofacial System and the TECHFIT Diagnostic Models and should be used in conjunction with expert clinical judgment.
#### 2 Comparison to Predicate Devices
Both subject device and primary predicate have the same indications for use: orthognathic surgery, including Le Fort and Genioplasty Surgical Guides and splints, reconstructive surgery, and trauma of the midface and maxillofacial skeleton. A device comparison table is more detailed in section 12 and Table 8.
{9}------------------------------------------------
Image /page/9/Picture/0 description: The image shows the logo for Techfit Digital Surgery. The logo consists of a blue sphere with a gray semi-circle overlapping it on the left side. To the right of the sphere, the word "TECHFIT" is written in gray, all-caps letters. Below "TECHFIT", the words "Digital Surgery" are written in a smaller, gray font.
# Table 8. Comparison to predicate device - an overview.
| Criteria | Subject device | Primary predicate | Reference device 1 | Reference device 2 | Reference device<br>3 |
|------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 510K number | TECHFIT DISRP®<br>SYSTEM | KLS Martin Individual<br>Patient Solutions (IPS)<br>Planning System | IPS CaseDesigner<br>2.0 | TECHFIT Patient-<br>Specific<br>Maxillofacial<br>system | AFFINITY<br>Proximal Tibia<br>System |
| 510K number | K230276 | K182789 | K200810 | K203282 | K220199 |
| Indications for<br>Use | TECHFIT Digitally<br>Integrated Surgical<br>Reconstruction Platform<br>(DISRP) System is<br>intended for use as a<br>software system and<br>image segmentation<br>system for the transfer of<br>imaging information from<br>a medical scanner such<br>as a CT based system.<br>The input data file is<br>processed by the DISRP<br>System and the result is<br>an output data file that<br>may then be provided as<br>digital models or used as<br>input to a rapid<br>prototyping portion of the<br>system that produces<br>physical outputs including<br>surgical guides and splints<br>for use in maxillofacial<br>surgery. The DISRP<br>System is also intended<br>as a pre-operative<br>software tool for<br>simulating / evaluating<br>surgical treatment options. | The KLS Martin<br>Individual Patient<br>Solutions (IPS) Planning<br>System is intended for<br>use as a software system<br>and image segmentation<br>system for the transfer of<br>imaging information from<br>a medical scanner such<br>as a CT based system.<br>The input data file is<br>processed by the IPS<br>Planning System, and the<br>result is an output data<br>file that may then be<br>provided as digital<br>models or used as input<br>to a rapid prototyping<br>portion of the system that<br>produces physical<br>outputs including<br>anatomical models,<br>guides, splints, and case<br>reports for use in<br>maxillofacial surgery. The<br>IPS Planning System is<br>also intended as a pre-<br>operative software tool | IPS CaseDesigner is<br>indicated for use as a<br>software and image<br>segmentation system<br>for the transfer of<br>imaging information<br>from a scanner such<br>as a CT scanner. It is<br>also indicated to<br>support the diagnostic<br>and treatment<br>planning process of<br>craniomaxillofacial<br>procedures. IPS<br>CaseDesigner<br>facilitates the service<br>offering of<br>individualized surgical<br>aids. | TECHFIT Patient-<br>Specific Maxillofacial<br>System is intended<br>for use in the<br>stabilization, fixation,<br>and reconstruction of<br>the<br>maxillofacial/midface<br>and mandibular<br>skeletal regions. | AFFINITY Proximal<br>Tibia System is<br>intended to treat<br>fractures,<br>nonunions,<br>malunions of the<br>proximal tibia<br>including simple,<br>comminuted,<br>lateral wedge<br>depression, medial<br>wedge, bicondylar<br>combination of<br>lateral wedge and<br>depression,<br>periprosthetic, and<br>fractures with<br>associated shaft<br>fractures.<br>-Simple<br>metaphyseal<br>fractures<br>(Classification AO<br>41-A2)<br>-Multifragmentary<br>metaphyseal<br>fractures<br>(Classification AO |
{10}------------------------------------------------
Image /page/10/Picture/0 description: The image shows the logo for Techfit Digital Surgery. The logo consists of a blue sphere with a gray crescent shape partially covering it on the left side. To the right of the sphere is the word "TECHFIT" in a simple, sans-serif font, with the words "Digital Surgery" in a smaller font below it.
{11}------------------------------------------------
Image /page/11/Picture/0 description: The image shows the logo for Techfit Digital Surgery. The logo consists of a blue sphere with a white teardrop shape overlapping it on the left side. To the right of the sphere is the word "TECHFIT" in a simple, sans-serif font, and below that is the phrase "Digital Surgery" in a smaller font.
| The DISRP system is<br>compatible with the<br>TECHFIT Patient- Specific<br>Maxillofacial System and<br>the TECHFIT Diagnostic<br>Models and should be<br>used in conjunction with<br>expert clinical judgment | surgical treatment<br>options. | | | -Simple bicondylar<br>fractures<br>(Classification<br>AO41-C1, 41-C2)<br><br>-Multifragmentary<br>bicondylar<br>fractures<br>(Classification AO<br>41-C3)<br><br>-Simple joint<br>simple<br>metaphyseal<br>fractures<br>(Classification AO<br>41-C1)<br><br>-Diaphisary<br>fractures<br>(Classification AO<br>42A and 42B) | |
|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------|---------------------------------------------|-------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------|
| Clinical Use | Orthognathic treatment | Maxillofacial and<br>orthognathic treatment | Cranio-maxillofacial<br>and orthognathic<br>treatment | Maxillofacial/<br>Midface & Mandible | Orthopedic<br>Proximal tibial<br>condyles |
| Product Code | DZJ, LLZ | DZJ, LLZ | LLZ | JEY | HRS, HWC |
{12}------------------------------------------------
Image /page/12/Picture/0 description: The image shows the logo for Techfit Digital Surgery. The logo features a blue sphere with a gray semi-circle overlapping it on the left side. To the right of the sphere, the word "TECHFIT" is written in gray, block letters. Below "TECHFIT" the words "Digital Surgery" are written in a smaller font.
| Classification | 21 CFR 872.4120, Class II<br>21 CFR 892.2050, Class II | 21 CFR 872.4120, Class II<br>21 CFR 892.2050, Class II | 21 CFR 892.2050,<br>Class II | 21 CFR 872.4760,<br>Class II | 21 CFR 888.3030,<br>Class II |
|-----------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------|------------------------------|
| Target<br>Population | Adults | Pediatric & Adults | Adults | Adults | Adults |
| Image Import | DICOM data format from<br>CT/CBCT scanner<br>3D models and surface<br>scan data in generic open<br>file format (STL) | DICOM data format from<br>CT scanner 3D models<br>and surface scan data in<br>generic open file format<br>(STL). | DICOM data format<br>from CT/CBCT<br>scanner<br>3D models and<br>surface scan data in<br>generic open file<br>format (STL) | DICOM data format<br>from CT/CBCT<br>scanner<br>3D models and<br>surface scan data in<br>generic open file<br>format (STL) | No |
| Software<br>output | Encrypted STL file with<br>design of the final patient<br>specific device such as<br>surgical guides, splints<br>and case reports | N.A. | Encrypted STL file<br>with design of the final<br>splint for centralized<br>manufacturing at KLS<br>Martin | N.A | N.A |
| | 3D Visualization | 2D and 3D Visualization | 2D and 3D<br>Visualization | N.A | N.A |
| | Distance and angle<br>measurements | Distance and angle<br>measurements | Distance and angle<br>measurements | N.A | N.A |
| Software<br>functions | Bone fragments can be<br>moved in the 3D space<br>(translations and<br>rotations) in order to plan<br>the ideal post-operative<br>position of each fragment.<br>Supported osteotomies:<br>Le Fort I, Chin, Splint | Bone fragments can be<br>moved in the 3D space<br>(translations and<br>rotations) in order to plan<br>the ideal post-operative<br>position of each<br>fragment.<br>Supported osteotomies:<br>Le Fort I, Sagittal, Splint,<br>Ramus, Chin | Bone fragments can<br>be moved in the 3D<br>space (translations<br>and rotations) in order<br>to plan the ideal post-<br>operative position of<br>each fragment.<br>Supported<br>osteotomies:<br>Le Fort I, Sagittal,<br>Splint, Ramus, Chin,<br>Segmental (Split, Y-<br>cut, H-cut) | N.A | N.A |
| | 3D Surgical Model<br>creation from CBCT/CT<br>scan data and STL file | 3D Surgical Model<br>creation from<br>CT scan data | 3D Surgical Model<br>creation from<br>CBCT/CT scan data<br>and STL file | N.A | N.A |
| Software<br>functions | 3D photo mapping | Soft tissue simulation | Soft tissue simulation,<br>3D photo mapping | N.A | N.A |
{13}------------------------------------------------
Image /page/13/Picture/0 description: The image shows the logo for Techfit Digital Surgery. The logo features a blue sphere with a gray teardrop shape overlapping it on the left side. To the right of the sphere are the words "TECHFIT" in a simple, sans-serif font, with the words "Digital Surgery" in a smaller font below.
{14}------------------------------------------------
Image /page/14/Picture/0 description: The image contains the logo for Techfit Digital Surgery. The logo consists of a blue sphere with a gray teardrop shape overlapping it on the left side. To the right of the sphere and teardrop is the word "TECHFIT" in gray, block letters, with the words "Digital Surgery" in a smaller font below it.
| Based on the created<br>orthognathic plan the user<br>can export surgical guides<br>and splints order files.<br>These files can be used to<br>calculate and produce<br>orthognathic surgical<br>guides and splints in the<br>production backend | Based on the created<br>orthognathic plan the<br>user can export surgical<br>splint order files. These<br>files can be used to<br>calculate and produce<br>orthognathic splints in the<br>production backend | Based on the created<br>orthognathic plan the<br>user can export<br>surgical splint order<br>files. These files can<br>be used to calculate<br>and produce<br>orthognathic splints in<br>the production<br>backend | N.A | N.A |
|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----|-----|
| DISRP: the user can<br>create a list of required<br>osteosynthesis plates<br>based on a created<br>surgical plan. | N.A. | Osteosynthesis Plates<br>& IPS Gate: the user<br>can create a list of<br>required<br>osteosynthesis plates<br>based on a created<br>surgical plan. | N.A | N.A |
{15}------------------------------------------------
Image /page/15/Picture/0 description: The image shows the logo for Techfit Digital Surgery. The logo features a blue sphere with a gray teardrop shape overlapping it on the left side. To the right of the sphere, the word "TECHFIT" is written in gray, uppercase letters. Below "TECHFIT", the words "Digital Surgery" are written in a smaller, gray font.
{16}------------------------------------------------
Image /page/16/Picture/0 description: The image shows the logo for Techfit Digital Surgery. The logo consists of a blue sphere with a gray teardrop shape partially overlapping it on the left side. To the right of the sphere are the words "TECHFIT" in a simple, sans-serif font, with the words "Digital Surgery" in a smaller font below.
| | - New algorithm for<br>occlusion alignment<br>- New algorithm for virtual<br>occlusion | N.A. | New algorithm<br>for occlusion<br>alignment - New<br>algorithm for<br>virtual<br>occlusion | N.A | N.A |
|-------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------|-------------|-------------|
| Operating<br>system<br>requirements | Operating system:<br>Windows 64-bit (Windows<br>7 and/or later)<br>Mac OS (Yosemite or<br>later) | Windows 64-bit<br>(Windows 7, 8,<br>10)<br>Mac OS X (Yosemite, El<br>Capitan) | Windows 64-bit<br>(Windows 7, 10)<br>Mac OS Catalina,<br>Mojave, High Sierra | N.A | N.A |
| Material | Resin Orthognathic<br>Surgical guides: Biomed<br>Clear Resin.<br>Machined Orthognathic<br>Surgical Guides: CP<br>Titanium.<br>Metal sleeves: CP<br>Titanium. | Anatomical Models:<br>Epoxy/Resin, Acrylic<br>Cutting/Marking Guides:<br>Polyamide, Titanium Alloy<br>(Ti-6Al-4V), CP Titanium<br>Splints: methacrylate | Not available | CP Titanium | CP Titanium |
{17}------------------------------------------------
Image /page/17/Picture/0 description: The image shows the logo for Techfit Digital Surgery. The logo features a blue sphere with a gray crescent shape partially overlapping it on the left side. To the right of the sphere and crescent is the word "TECHFIT" in a simple, sans-serif font, with the words "Digital Surgery" in a smaller font below it.
| Manufacturing<br>Method | Resin: 3D printing -<br>Stereolithography (SLA).<br>CP Titanium: Traditional<br>(Subtractive) | Epoxy/Resin, Acrylic:<br>Stereolithography (SLA)<br>CP Titanium: Traditional<br>(Subtractive)<br>Ti-6Al-4V: 3D (Additive;<br>Selective Laser Melting)<br>Polyamide: 3D (Additive;<br>Selective Laser Sintering) | Not available | Traditional<br>(Subtractive) | Traditional<br>(Subtractive) |
|----------------------------------------|-----------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------|-----------------------------------------------------------------------------------------------|----------------------------------------------------------|
| Sterilization | Non-sterile (Steam) | Non-sterile (Steam) | Not available | Non-sterile (Steam) | Non-sterile<br>(Steam) |
| Manufacturer | TECHFIT Digital Surgery | KLS Martin | Nobel Biocare | TECHFIT Digital<br>Surgery | Industrias Médicas<br>Sampedro |
| Patient-<br>specific<br>configuration? | Yes. Devices are<br>manufactured patient-<br>specific, based on a CT<br>scan of the patient | Yes. Devices are<br>manufactured patient-<br>specific, based on a CT<br>scan of the patient | Yes | Yes. Devices are<br>manufactured<br>patient-specific,<br>based on a CT scan<br>of the patient | No. The plates are<br>standard<br>manufactured<br>plates |
{18}------------------------------------------------
Image /page/18/Picture/0 description: The image shows the logo for Techfit Digital Surgery. The logo consists of a blue sphere with a gray teardrop shape overlapping it on the left side. To the right of the sphere is the word "TECHFIT" in gray, all-caps lettering. Below "TECHFIT" are the words "Digital Surgery" in a smaller, gray font.
#### 3 Performance data
The following non-clinical testing was conducted as a basis for the determination of substantial equivalence:
| Name | Test method | Conclusion |
|-------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Safety testing | | |
| Validating sterilization<br>methods | AAMI/ISO 17665-1:2006/(R)2013<br>Sterilization of health care products<br>- Moist heat - Part 1 Requirements<br>for the development, validation and<br>routine control of a sterilization<br>process for medical devices.<br>ANSI/AAMI/ISO<br>14937:2009/(R)2013 Sterilization of<br>Healthcare Products -General<br>Requirements for Characterization<br>of a Sterilizing Agent and the<br>Development, Validation and<br>Routine Control of a Sterilization<br>Process for Medical Devices. | The results of the steam<br>sterilization validation<br>show that TECHFIT<br>Orthognathic Surgical<br>Guides sterilized to a<br>SAL of 10-6 using the<br>recommended steam<br>sterilization instructions |
| Performance testing | | |
| Dimensional validation | Scanning of orthognathic surgical<br>guides and then comparing them<br>versus the original files, before and<br>after sterilization. | DISRP and<br>Orthognathic surgical<br>guides maintain the<br>proportions and<br>dimensions of the<br>original digital design<br>after manufacturing and<br>sterilization. |
| Mechanical testing | Bending and compressive strength,<br>after sterilization. Evaluating if the<br>orthognathic surgical guides can<br>withstand the maximum bite force,<br>force required for drilling and handle<br>force required for a general surgery. | Orthognathic Surgical<br>Guides withstand bit<br>force and the force<br>applied by the surgeon<br>during handling. |
| Compatibility testing | Verifying compatibility between the<br>Maxillofacial System and<br>orthognathic surgical guides | Orthognathic Surgical<br>Guides are compatible<br>with the Maxillofacial<br>System and Metal<br>Sleeves. |
| Software testing | Software verification and validation<br>activities | DISRP software was<br>verified and validated as<br>per IEC 62304 |
### Table 9. Performance data
TECHFIT DIGITAL SURGERY
{19}------------------------------------------------
Image /page/19/Picture/0 description: The image shows the logo for Techfit Digital Surgery. The logo features a blue sphere with a gray semi-circle overlapping it on the left side. To the right of the sphere, the word "TECHFIT" is written in gray, all-caps letters. Below "TECHFIT", the words "Digital Surgery" are written in a smaller, gray font.
- Biocompatibility testing .
Resin Orthognathic Surgical Guides are manufactured with Biomed Clear (Formlabs). The following are the tests summarizes the biocompatibility testing:
| | Table 10. Overview of Biomed Clear biocompatibility testing | | | | |
|--|-------------------------------------------------------------|--|--|--|--|
|--|-------------------------------------------------------------|--|--|--|--|
| Test / Assessment Description | Test Report Conclusion |
|------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------|
| Cytotoxicity:<br>ISO 10993-5 | No cytotoxic effect properties in<br>compliance with requirements of<br>the ISO 10993-5 guidelines |
| Sensitization:<br>ISO 10993-10 | No sensitizing properties in<br>compliance with requirements of<br>the ISO 10993-10 guidelines |
| Irritation:<br>ISO 10993-10 | No irritant properties in compliance<br>with requirements of the ISO<br>10993-10 guidelines |
| Acute systemic toxicity:<br>ISO 10993-11 and USP <88> | No evidence of systemic toxicity<br>test passed and is considered<br>negative based on standards set by<br>ISO 10993-11 |
| Pyrogenicity:<br>ISO 10993-11 | No pyrogenic properties in<br>compliance with requirements of<br>the ISO 10993-11 |
| Chemical Characterization and risk<br>assessment:<br>ISO 10993-18 and ISO 10993-17 | Non-toxic chemical characterization<br>as per ISO 10993-18 and ISO<br>10993-17 standards |
An overview of TECHFIT's biocompatibility testing is shown below:
# Table 11. Overview of TECHFIT biocompatibility testing for Resin Orthognathic Surgical Guides
| Test / Description | Assessment | Test Report Conclusion |
|--------------------------------|------------|----------------------------------------------------------------------------------------------|
| Cytotoxicity:<br>ISO 10993-5 | | No cytotoxic effect properties in compliance with requirements of the ISO 10993-5 guidelines |
| Sensitization:<br>ISO 10993-10 | | No sensitizing properties in compliance with requirements of the ISO 10993-10 guidelines |
| Irritation:<br>ISO 10993-23 | | No irritant properties in compliance with requirements of the ISO 10993-10 guidelines |
{20}------------------------------------------------
Image /page/20/Picture/0 description: The image shows the Techfit Digital Surgery logo. The logo features a blue sphere with a gray teardrop shape overlapping it on the left side. To the right of the sphere, the word "TECHFIT" is written in gray, with the words "Digital Surgery" written in a smaller font size below it.
Machined Orthognathic Surgical Guides and Metal Sleeves are manufactured with commercially pure Titanium grade 4. The following are the tests summarizes the biocompatibility testing:
### Table 12. Overview of TECHFIT biocompatibility testing for Machined Orthognathic Surgical Guides and Metal Sleeves.
| Test / Assessment Description | Test Report Conclusion |
|------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------|
| Cytotoxicity:<br>ISO 10993-5 | No cytotoxic effect properties in<br>compliance with requirements of<br>the ISO 10993-5 guidelines |
| Sensitization:<br>ISO 10993-10 | No sensitizing properties in<br>compliance with requirements of<br>the ISO 10993-10 guidelines |
| Irritation:<br>ISO 10993-10 | No irritant properties in compliance<br>with requirements of the ISO<br>10993-10 guidelines |
| Acute systemic toxicity:<br>ISO 10993-11 | No evidence of systemic toxicity<br>test passed and is considered<br>negative based on standards set by<br>ISO 10993-11 |
| Pyrogenicity:<br>ISO 10993-11 | No pyrogenic properties in<br>compliance with requirements of<br>the ISO 10993-11 |
| Genotoxicity:<br>ISO 10993-3 | No genotoxicity properties in<br>compliance with requirements of<br>the ISO 10993-3 |
| Chemical Characterization and risk<br>assessment:<br>ISO 10993-18 and ISO 10993-17 | Non-toxic chemical characterization<br>as per ISO 10993-18 and ISO<br>10993-17 standards |
#### 4 Conclusion
Non-clinical tests demonstrate that the DISRP System is substantially equivalent to the predicate devices.
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.