DynaFlex ® Clear Bracket & Buttons is indicated for alignment of teeth during orthodontic treatment.
Device Story
DynaFlex Clear Brackets & Buttons are ceramic orthodontic appliances used to align teeth. Dental professionals bond these brackets to natural teeth using orthodontic adhesive. Once bonded, the clinician inserts orthodontic wires into the bracket slots; the system exerts pressure on the teeth to move them to a desired position. The brackets feature a self-ligating door with a nickel-titanium spring mechanism. The device is intended for single-use and is provided non-sterile. By applying controlled forces via archwires, the device facilitates progressive tooth movement, correcting malocclusion and improving dental alignment. The use of nickel-titanium springs provides the necessary mechanical force for the self-ligating function, similar to predicate systems using stainless steel springs.
Clinical Evidence
Bench testing only. Testing included bond strength evaluation to ensure the bracket remains affixed to the tooth under orthodontic forces. Biocompatibility testing was conducted in accordance with ISO 10993-1. Manufacturing process validation confirmed that the finished device meets technical specifications. No clinical data was provided.
Technological Characteristics
Materials: Polycrystalline (translucent) alumina bracket body, nickel-titanium alloy spring mechanism. Design: Self-ligating ceramic bracket with mechanical lock base. Dimensions: In/out range 0.027"-0.039". Connectivity: None (mechanical). Sterilization: Non-sterile. Single-use.
Indications for Use
Indicated for alignment of teeth during orthodontic treatment in patients requiring correction of misalignment and malocclusion.
Regulatory Classification
Identification
An orthodontic plastic bracket is a plastic device intended to be bonded to a tooth to apply pressure to a tooth from a flexible orthodontic wire to alter its position.
{1}------------------------------------------------
Image /page/0/Picture/0 description: The image contains two logos. On the left is the logo for the Department of Health & Human Services - USA, featuring a stylized human figure. To the right is the logo for the U.S. Food & Drug Administration (FDA), with the FDA acronym in a blue square and the full name "U.S. FOOD & DRUG ADMINISTRATION" written in blue text.
June 5, 2023
Dyna Flex Matthew Malabey Director of Operations 8050 Hawk Ridge Trail Lake Saint Louis, Missouri 63367
Re: K230225
Trade/Device Name: DynaFlex Clear Brackets & Buttons Regulation Number: 21 CFR 872.5470 Regulation Name: Orthodontic Plastic Bracket Regulatory Class: Class II Product Code: NJM Dated: February 21, 2023 Received: February 21, 2023
Dear Matthew Malabey:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{2}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
# Michael E. Adjodha -S
Michael E. Adjodha, M.ChE. Assistant Director DHT1B: Division of Dental and ENT Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
#### Indications for Use
510(k) Number (if known)
#### K230225
Device Name DynaFlex Clear Brackets & Buttons
Indications for Use (Describe)
DynaFlex ® Clear Bracket & Buttons is indicated for alignment of teeth during orthodontic treatment.
Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995,
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
Image /page/3/Picture/0 description: The image displays the logo for DynaFlex, a company specializing in orthodontic solutions. The logo features a stylized graphic to the left, followed by the company name "DynaFlex" in a bold, red font, with a registered trademark symbol. Below the company name, the tagline "Your Total Orthodontic Solution!" is written in a smaller, dark teal font, with a service mark symbol.
## 510(k) Summary (K230225)
## Submitter:
DynaFlex 8050 Hawk Ridge Trail. Lake St. Louis, MO 63367
## Contact:
Matthew Malabey Director of Operations DynaFlex 314-426-4020- Phone 314-429-7575-Fax
Date Summary Prepared: June 1st, 2023
## Device Name:
- Trade Name DynaFlex Clear Brackets & Buttons .
- Classification name Orthodontic Plastic Bracket O
- Regulation Medical Specialty Dental 0
- o Review Panel - Dental
- o Product Code - NJM
- . Regulation Number 21 CFR& 872.5470
- o Device Class - 2
## Devices for Which Substantial Equivalence is Claimed:
- 0 Primary Predicate: Ortho Organizers (Henry Schein) .- Carriere SLX 3D Clear (K173440)
{5}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for DynaFlex, a company that provides orthodontic solutions. The logo consists of a stylized red square with a white abstract design on the left, followed by the company name "DynaFlex" in bold, red letters. Below the company name is the tagline "Your Total Orthodontic Solution!" in a smaller, black font.
#### Device Descrintion:
DynaFlex ® Clear Brackets & Buttons is a series of clear, lightweight, ceramic brackets, and buttons. DynaFlex ® Clear Brackets & Buttons are used in the alignment of teeth through orthodontic treatment of misalignment and malocclusion in patients by moving teeth progressively to a final, treated state.
DynaFlex ® Clear Brackets & Buttons are designed to move teeth to improve their alignment. Ceramic brackets are bonded to natural teeth by dental professionals to connect with orthodontic wires to cause tooth movement to a more preferred position. Ceramic brackets are intended for use in affixed to a tooth so that pressure can be exerted on the teeth. Bonding supplies are used to bond the bracket on to a tooth. Ceramic brackets are bonded to natural teeth by dental professionals to connect with orthodontic wires to cause tooth movement to a more preferred position.
DynaFlex ® Clear Brackets & Buttons may be adjusted by the dentist. The appliance is provided as a non-sterile device and is single use. DynaFlex® Clear Brackets & Buttons is for prescription only.
#### Indications for Use:
DynaFlex ® Clear Bracket & Buttons is indicated for alignment of teeth during orthodontic treatment.
#### Summary of Technological Characteristics:
A dental health care professional (e.g. Orthodontist or dentist), prescribes the Clear Brackets & Buttons based on an characteristics assessment of the patient's teeth. determines a course of treatment with the system. DynaFlex and Buttons all have a bracket body made of ceramic. DynaFlex brackets are self-ligating ceramic door with a nickeltitanium spring mechanism. Orthodontic brackets are affixed to teeth using an orthodontic adhesive. Pressure is exerted on a tooth when the brackets are used in combination with archwires and/or other intraoral modules. The pressure causes tooth movement.
## Performance Characteristics:
Bench testing of the clear brackets & buttons was performed. Specifically, bond strength testing was performed to determine the amount of force required to separate the bonded bracket from the tooth. The minimum specification for bond strength was met by all samples tested.
## Non-Clinical Performance Testing
As part of demonstrating substantial equivalence to the predicate devices that are subject to this 510/k) submission. DynaFlex completed a number of non-clinical performance tests. The DynaFlex Clear Bracket & Buttons meet all the requirements for overall design,
{6}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the logo for DynaFlex, a company that provides orthodontic solutions. The logo features a stylized graphic of a bird in flight, followed by the company name in bold, red letters. Below the company name is the tagline "Your Total Orthodontic Solution!" in a smaller, black font.
biocompatibility, and performance results confirming that the design output meets the design inputs and specifications for the device.
The DynaFlex Clear brackets & buttons meet all the testing in accordance with internal requirements, national standards, and international standards to support substantial equivalence of the subject device.
Manufacturing Process Flow validation was performed to ensure that the finished device matches the specification of the technical requirements. The output and specifications were tested and compared. DynaFlex Clear Bracket & Buttons met the specifications of this testing.
DynaFlex has completed biocompatibility testing per ISO 10993-1 and its applicable parts, as appropriate for the Clear Bracket & Buttons contact and duration. The mechanical properties of the clear brackets and buttons have been demonstrated by the manufacturer as appropriate for use with clear brackets.
#### Substantial Equivalence:
Chevel Creating Comments Company Contraction Come Con-
DynaFlex® Clear brackets & buttons is substantially equivalent to the Carriere SLX 3D (Clear Kl 7 3 440). DynaFlex Clear Bracket & Buttons is used in a manner similar to the Carriere SLX 3D Clear (K173440) system (Primary Predicate) marketed by Ortho Organizers (Henry Schein). The proposed and predicate devices are of similar design; made from similar materials and are fabricated using similar manufacturing methods that are common to the dental device industry.
Furthermore, DynaFlex® Clear Bracket & Buttons have similar intended uses as the predicate devices. Both the predicate and the subject device serve as an oral appliance constructed of same polymer components, which are intended to move teeth.
There are no substantial technical or functional differences between the predicated orthodontic ceramic bracket and the predicate device in terms of design, function. safety and intended use. Both are ceramic.
However, the self-ligating brackets made of polycrystalline alumina in the proposed device are made
with nickel-titanium springs, unlike the stainless-steel springs of the Predicate Device (K173440). Nickel titanium springs in the proposed device can be considered substantially equivalent to the predicate device, as this material presents no additional risks - technologically, and clinically. Therefore, there are no substantial differences between the subject device and predicate device in terms of the spring used.
The properties and characteristics of the predicate devices are compared to DynaFlex● Clear Bracket & Buttons in the Table 5-1.
{7}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the logo for DynaFlex, a company that provides orthodontic solutions. The logo consists of a red square with a white abstract design on the left, followed by the word "DynaFlex" in red, with a registered trademark symbol. Below the company name is the text "Your Total Orthodontic Solution!" in a smaller, dark green font, with a service mark symbol.
| Table 5-1<br>Substantial Equivalence Comparison | | | |
|-------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------|
| Manufacturer: | DynaFlex® | Ortho Organizer (Henry Schein Co, Predicate) | Comparison |
| Trade Name: | DynaFlex® Clear<br>Brackets & Buttons | Carriere SLX 3D Clear | N/A |
| Product Images: | Image: DynaFlex Clear Brackets & Buttons | Image: Carriere SLX 3D Clear | N/A |
| Product Code: | NJM | NJM | Same |
| Medical Specialty | Dental | Dental | Same |
| Regulation Number: | 872.5470 | 872.5470 | Same |
| Regulation Name: | Orthodontic plastic bracket | Orthodontic plastic bracket | Same |
| 510(k): | K230225 | K173440 | N/A |
| Indications for Use: | DynaFlex® Clear Bracket &<br>Buttons is indicated for<br>alignment of teeth during<br>orthodontic treatment. | The Carriere SLX 3D Clear<br>orthodontic ceramic bracket<br>system is intended to aid in<br>the movement of teeth<br>during<br>orthodontic treatment | Similar |
| Design | Archwire slot, tiewings for<br>ligature and identification<br>marks for placement Hooks<br>for ligation, for additional<br>tooth movement Molded<br>ceramic body with rounded<br>corners and e Slot to hold<br>orthodontic wires edges | Archwire slot, tiewings for<br>ligature and identification<br>marks for placement Hooks<br>for ligation, for additional<br>tooth movement Molded<br>ceramic body with rounded<br>corners and e Slot to hold<br>orthodontic wires edges | Same |
| Base: | Mechanical Lock Base | Mechanical Lock Base | Same |
| In/Out | .027" - .039" | .025" - .046" | Within range of<br>predicate |
| Torque | -12° through + 12° | -17° through +17° | Within range of<br>predicate |
| Angulation | Up to + 9° | Up to + 9° | Within range of<br>predicate |
| Rotation | Up to + 5° | Up to + 12° | Within range of<br>predicate |
| Mode of Use | An archwire (provided by<br>clinician and worn by<br>patient)<br>is inserted into the device<br>and provides the light<br>orthodontic forces required<br>to move teeth per the dental<br>professional's technique<br>and<br>treatment goals. | An archwire (provided by<br>clinician and worn by<br>patient)<br>is inserted into the device<br>and provides the light<br>orthodontic forces required<br>to move teeth per the dental<br>professional's technique<br>and<br>treatment goals. | Same |
| Application | Bonded with Orthodontic<br>Adhesive | Bonded with Orthodontic<br>Adhesive | Same |
| Minimum Bond<br>Strength (N/MPa) | 80/6.0 | 103.7/7.63 | Within range of<br>predicate |
| Supplied Sterile: | No | No | Same |
| | | | |
| Single Use: | Yes | Yes | Same |
| Target Users | Dental Professionals trained in orthodontics | Dental Professionals trained in orthodontics | Same |
| Physical Properties | Polycrystalline Alumina /<br>Nickle-Titanium Alloy | Polycrystalline Alumina /<br>Stainless Steel | Similar |
| Biocompatibility | Passed ISO 10993-1 and<br>series | Passed ISO 10993-1 and<br>series | Same |
| Material | Polycrystalline (translucent)<br>alumina, | Polycrystalline (translucent)<br>alumina | Same |
{8}------------------------------------------------
Image /page/7/Picture/0 description: The image shows the logo for "DynaFlex", a company that provides orthodontic solutions. The logo features a stylized graphic to the left of the company name, both in red. Below the company name is the tagline "Your Total Orthodontic Solution!" in a smaller, black font.
#### Substantial Equivalence Conclusion
By definition, a device is substantially equivalent to a predicate device when the device has the same intended use and the same technological characteristics as the previously cleared predicate device. Based on the comparison and analysis above, the DynaFlex Clear Brackets & Buttons is determined to be substantially equivalent to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.