Eminent Spine 3D Lumbar Interbody Fusion Systems

K230219 · Eminent Spine, LLC · MAX · May 16, 2023 · Orthopedic

Device Facts

Record IDK230219
Device NameEminent Spine 3D Lumbar Interbody Fusion Systems
ApplicantEminent Spine, LLC
Product CodeMAX · Orthopedic
Decision DateMay 16, 2023
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 888.3080
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The Eminent Spine 3D Lumbar Interbody Fusion Systems (PLIF, TLIF, and ALIF) are indicated for intervertebral body fusion of the lumbar spine, from L2 to S1, in skeletally mature patients who have had six months of non-operative treatment. The device is intended for use at either one level or two contiguous levels for the treatment of degenerative disc disease (DDD) with up to Grade I spondylolisthesis. DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies. The device system is designed for use with supplemental fixation and autograft to facilitate fusion. Hyperlordotic interbody devices (≥20° lordosis) must be used with at least anterior supplemental fixation.

Device Story

Intervertebral body fusion system (PLIF, TLIF, ALIF) designed to replace damaged/unstable lumbar discs; provides mechanical support until biologic fusion occurs. Surgeon-implanted via various surgical approaches; packed with autogenous bone graft. Features serrated superior/inferior surfaces for endplate grip and expulsion resistance. Used in clinical settings for patients with DDD/spondylolisthesis; requires supplemental fixation. Benefits patient by restoring spinal stability and facilitating fusion.

Clinical Evidence

Bench testing only. Mechanical testing included static axial compression, static axial compression-shear, dynamic axial compression, dynamic axial compression-shear (ASTM F2077), subsidence (ASTM F2267), and expulsion testing. Results demonstrate performance equivalent to or better than acceptance criteria.

Technological Characteristics

Material: Ti-6Al-4V per ASTM F3001. Form factor: Hollow rectangular blocks with serrated surfaces for endplate grip. Configurations: PLIF, TLIF, ALIF with varying footprints, heights, and lordotic angles (6°-30°). Energy source: None (mechanical). Connectivity: None. Sterilization: Not specified.

Indications for Use

Indicated for intervertebral body fusion of lumbar spine (L2-S1) in skeletally mature patients with degenerative disc disease (DDD) and up to Grade I spondylolisthesis, following 6 months of failed non-operative treatment. Used at one or two contiguous levels with supplemental fixation and autograft.

Regulatory Classification

Identification

An intervertebral body fusion device is an implanted single or multiple component spinal device made from a variety of materials, including titanium and polymers. The device is inserted into the intervertebral body space of the cervical or lumbosacral spine, and is intended for intervertebral body fusion.

Special Controls

*Classification.* (1) Class II (special controls) for intervertebral body fusion devices that contain bone grafting material. The special control is the FDA guidance document entitled “Class II Special Controls Guidance Document: Intervertebral Body Fusion Device.” See § 888.1(e) for the availability of this guidance document.(2) Class III (premarket approval) for intervertebral body fusion devices that include any therapeutic biologic (e.g., bone morphogenic protein). Intervertebral body fusion devices that contain any therapeutic biologic require premarket approval. (c) *Date premarket approval application (PMA) or notice of product development protocol (PDP) is required.* Devices described in paragraph (b)(2) of this section shall have an approved PMA or a declared completed PDP in effect before being placed in commercial distribution.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is a blue square with the letters "FDA" in white, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue. May 16, 2023 Eminent Spine, LLC % Jennifer Palinchik President Jalex Medical 27865 Clemens Rd. Suite 3 Westlake, Ohio 44145 Re: K230219 Trade/Device Name: Eminent Spine 3D Lumbar Interbody Fusion Systems Regulation Number: 21 CFR 888.3080 Regulation Name: Intervertebral body fusion device Regulatory Class: Class II Product Code: MAX Dated: March 14, 2023 Received: March 14, 2023 Dear Jennifer Palinchik: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {1}------------------------------------------------ Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). > Sincerely. Katherine D. Kavlock -S for Brent Showalter, Ph.D. Assistant Director DHT6B: Division of Spinal Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ # Indications for Use 510(k) Number (if known) K230219 #### Device Name Eminent Spine 3D Lumbar Interbody Fusion Systems #### Indications for Use (Describe) The Eminent Spine 3D Lumbar Interbody Fusion Systems (PLIF, and ALIF) are indicated for intervertebral body fusion of the lumbar spine, from L2 to S1, in skeletally mature patients who have had six months of non-operative treatment. The device is intended for use at either one level or two contiguous levels for the treatment of degenerative disc disease (DDD) with up to Grade I spondylolisthesis. DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies. The device system is designed for use with supplemental fixation and autograft to facilitate fusion. Hyperlordotic interbody devices (≥20° lordosis) must be used with at least anterior supplemental fixation. | Type of Use (Select one or both, as applicable) | | |------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------| | <span style="text-decoration: overline;">\</span> Prescription Use (Part 21 CFR 801 Subpart D) | <span style="text-decoration: ;"> </span> Over-The-Counter Use (21 CFR 801 Subpart C) | CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ Image /page/3/Picture/0 description: The image shows a logo for "Eminent Spine". The logo features a golden sword with a spine-like design in the center, surrounded by a golden, curved shape. The text "EMINENT" is displayed above the word "SPINE" in a smaller font size. # 510(k) Summary | Submitted By: | Eminent Spine, LLC<br>2004 Ventura Dr. Suite #100<br>Plano, TX 75093 | | | |-------------------------------|----------------------------------------------------------------------------------------------------------------------------|--|--| | Date: | 05/15/2023 | | | | Contact Person: | Jennifer Palinchik, President, JALEX Medical | | | | Contact Telephone: | (440) 935-3282 | | | | Contact Fax: | (440) 933-7839 | | | | Device Trade Name: | Eminent Spine 3D Lumbar Interbody Fusion Systems | | | | Common Name: | Intervertebral Body Fusion Device | | | | Device Classification Name: | Intervertebral Body Fusion Device with Bone Graft, Lumbar | | | | Device Classification: | Class II | | | | Reviewing Panel: | Orthopedic | | | | Product Code: | MAX | | | | Primary Predicate Device: | Eminent Spine Interbody Fusion System (K090064) | | | | Additional Predicate Devices: | Eminent Spine 3D Cervical Interbody Fusion System (K212701)<br>CancelleX Porous Titanium Lumbar Interbody Device (K190364) | | | ### Device Description: The Eminent Spine 3D Lumbar Interbody Fusion Systems are intervertebral body fusion systems used in the spine to replace a collapsed, damaged, or unstable disc. The 3D Lumbar Interbody Fusion Systems (PLIF, TLIF, and ALIF) are comprised of various sizes and configurations to accommodate individual patient anatomy. The configurations are designed to provide the surgeon with different surgical approach options, and packed with autogenous bone graft to facilitate fusion. Serrations on the superior and inferior surfaces of each device grip the endplates of the adjacent vertebral bodies to aid in expulsion resistance. The device is intended to provide mechanical support to the implanted level until biologic fusion is achieved. #### Indications for Use: The Eminent Spine 3D Lumbar Interbody Fusion Systems (PLIF, TLIF, and ALIF) are indicated for intervertebral body fusion of the lumbar spine, from L2 to S1, in skeletally mature patients who have had six months of non-operative treatment. The device is intended for use at either one level or two contiguous levels for the treatment of degenerative disc disease (DDD) with up to Grade I spondylolisthesis. DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies. The device system is designed for use with supplemental fixation and autograft to facilitate fusion. Hyperlordotic interbody devices (≥20° lordosis) must be used with at least anterior supplemental fixation. ## Summary of Technological Characteristics: The Eminent Spine 3D Lumbar Interbody Fusion Systems and the predicates have the same intended use and fundamental scientific technology. All devices compare similarly in: {4}------------------------------------------------ Image /page/4/Picture/1 description: The image shows a logo for "Eminent Spine". The logo features a golden spine with a sword running through it, surrounded by a golden arch. The text "EMINENT" is written above the text "SPINE" in a smaller font. - Design features . - Intended use - Materials ● - Dimensions . - Function ● # Table 1: Dimensions and Technological Characteristics Comparison | Item | Eminent Spine 3D<br>Lumbar Interbody<br>Fusion Systems | Eminent Spine<br>Interbody Fusion<br>System (K090064) | Eminent Spine 3D<br>Cervical Interbody<br>Fusion System<br>(K212701) | Comparison | |------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------| | Classification<br>Name | Intervertebral Body<br>Fusion Device | Intervertebral Body<br>Fusion Device | Intervertebral Body<br>Fusion Device | Equivalent | | Regulation | 888.3080 | 888.3080 | 888.3080 | Equivalent | | Product Code | MAX | MAX, MQP, ODP | ODP | Intervertebral<br>body fusion<br>device code<br>equivalent to<br>primary<br>predicate | | Indications for<br>Use | The Eminent Spine 3D<br>Lumbar Interbody<br>Fusion Systems (PLIF,<br>TLIF, and ALIF) are<br>indicated for<br>intervertebral body<br>fusion of the lumbar<br>spine, from L2 to S1, in<br>skeletally mature<br>patients who have had<br>six months of non-<br>operative treatment.<br>The device is intended<br>for use at either one<br>level or two contiguous<br>levels for the treatment<br>of degenerative disc<br>disease (DDD) with up<br>to Grade I<br>spondylolisthesis. DDD<br>is defined as back pain<br>of discogenic origin<br>with degeneration of<br>the disc confirmed by<br>history and<br>radiographic studies.<br>The device system is<br>designed for use with<br>supplemental fixation<br>and autograft to<br>facilitate fusion. | The Eminent Spine<br>Interbody Fusion<br>System (Sidewinder,<br>Python and<br>Cottonmouth) is<br>indicated for<br>intervertebral body<br>fusion of the lumbar<br>spine, from L2 to S1, in<br>skeletally mature<br>patients who have had<br>six months of non-<br>operative treatment.<br>The device is intended<br>for use at either one<br>level or two contiguous<br>levels for the treatment<br>of degenerative disc<br>disease (DDD) with up<br>to Grade I<br>spondylolisthesis. DDD<br>is defined as back pain<br>of discogenic origin<br>with degeneration of<br>the disc confirmed by<br>history and<br>radiographic studies.<br>The device system is<br>designed for use with<br>supplemental fixation | The Eminent Spine 3D<br>Cervical Interbody<br>Fusion System is<br>indicated for<br>intervertebral body<br>fusion in skeletally<br>mature patients with<br>degenerative disc<br>disease (DDD) of the<br>cervical spine at one<br>disc level from the C2-<br>C3 disc to the C7-T1<br>disc. DDD is defined as<br>discogenic pain with<br>degeneration of the disc<br>confirmed by history<br>and radiographic<br>studies. The device<br>system is designed for<br>use with supplemental<br>fixation and autograft to<br>facilitate fusion.<br>Patients should have at<br>least six (6) weeks of<br>non-operative treatment<br>prior to treatment with<br>an intervertebral cage. | Intervertebral<br>body fusion<br>indications<br>equivalent to<br>primary<br>predicate | | | | | | | | Hyperlordotic interbody<br>devices (≥20° lordosis)<br>must be used with at<br>least anterior<br>supplemental fixation. | and autograft to<br>facilitate fusion.<br><br>The Eminent Spine<br>System of implants is<br>indicated for use to<br>replace a vertebral body<br>that has been resected<br>or excised (i.e. partial<br>or total vertebrectomy)<br>due to tumor or<br>trauma/fracture. The<br>implant is intended for<br>use in the<br>thoracolumbar spine<br>(from T1 to L5) and is<br>intended for use with<br>supplemental internal<br>fixation and autograft<br>or allograft bone. These<br>devices are designed to<br>restore the<br>biomechanical integrity<br>of the anterior, middle<br>and posterior spinal<br>column even in the<br>absence of fusion for a<br>prolonged period. | | | | | Description | The Eminent Spine 3D<br>Lumbar Interbody<br>Fusion Systems are<br>intervertebral body<br>fusion systems used in<br>the spine to replace a<br>collapsed, damaged, or<br>unstable disc. The 3D<br>Lumbar Interbody<br>Fusion Systems (PLIF,<br>TLIF, and ALIF) are<br>comprised of various<br>sizes and configurations<br>to accommodate<br>individual patient<br>anatomy. The<br>configurations are<br>designed to provide the<br>surgeon with different<br>surgical approach<br>options, and packed<br>with autogenous bone<br>graft to facilitate fusion. | The Eminent Spine<br>Interbody Fusion<br>System is comprised of<br>various sizes and<br>configuration to<br>accommodate<br>individual patient<br>anatomy. The<br>configurations are<br>designed to provide the<br>surgeon with different<br>surgical approach<br>options. | The Eminent Spine 3D<br>Cervical Interbody<br>Fusion System is<br>comprised of various<br>sizes and configuration<br>to accommodate<br>individual patient<br>anatomy. The device is<br>a hollow rectangular<br>shaped block, which is<br>available in a parallel or<br>lordotic configurations.<br>The device is hollow to<br>allow for placement of<br>bone graft. There are<br>teeth on the superior<br>and inferior surfaces of<br>the device to inhibit<br>movement of the<br>device. | Equivalent to<br>primary<br>predicate | | | | | | | | | surfaces of each device<br>grip the endplates of the<br>adjacent vertebral<br>bodies to aid in<br>expulsion resistance.<br>The device is intended<br>to provide mechanical<br>support to the<br>implanted level until<br>biologic fusion is<br>achieved. | | | | | Footprints | PLIF: 22-31 mm x 9-12<br>mm<br>TLIF: 28-34 mm x 11<br>mm<br>ALIF: 32-42 mm x 21-<br>28 mm | Python (PLIF): 22-31<br>mm x 9-12 mm<br>Sidewinder (TLIF): 28-<br>31 mm x 11 mm<br>Cottonmouth (ALIF):<br>32-42 mm x 21-28 mm | 14x12 mm, 15x13 mm,<br>17x12 mm,<br>17x14 mm, 19x16 mm | Equivalent to<br>primary<br>predicate | | Heights | PLIF: 6-18 mm<br>TLIF: 6-18 mm<br>ALIF: 10-20 mm | Python (PLIF): 6-18<br>mm<br>Sidewinder (TLIF): 6-<br>18 mm<br>Cottonmouth (ALIF):<br>10-20 mm | 5-12 mm | Equivalent to<br>primary<br>predicate | | Lordotic angle | 6°, 12°, 18°, 24°, 30°<br>(depending on model) | 6°, 12°, 18°, 24°, 30°<br>(depending on model) | 0°, 6° | Equivalent to<br>primary<br>predicate | | Material | Ti-6Al-4V per ASTM<br>F3001 | PEEK Optima LT1 and<br>Ti-6Al-4V ELI per<br>ASTM F136 | Ti-6Al-4V per ASTM<br>F3001 | Equivalent to<br>additional<br>predicate | {5}------------------------------------------------ Image /page/5/Picture/0 description: The image shows a logo for "Eminent Spine". The logo features a golden sword with a twisted blade, positioned vertically. The sword is surrounded by a golden, semi-circular frame, and the words "EMINENT SPINE" are written in a serif font below the sword. {6}------------------------------------------------ Image /page/6/Picture/1 description: The image shows a logo for "Eminent Spine". The logo features a golden spine with a sword running through it, surrounded by a golden, curved frame. The text "EMINENT" is written in a serif font above the word "SPINE", which is in a smaller font. ## Mechanical Testing: Substantial equivalence is supported by the results of mechanical testing, including static axial compression, static axial compression-shear, dynamic axial compression, dynamic axial compressionshear per ASTM F2077, subsidence per ASTM F2267, and expulsion. Results support that the subject device performs as well as or better than the chosen acceptance criteria. ## Conclusion: Based on the indications for use, technological characteristics, and comparison with the predicate device, the subject device has demonstrated substantial equivalence.
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...