K230206 · Gravitas Medical, Inc. · KNT · Mar 24, 2023 · Gastroenterology, Urology
Device Facts
Record ID
K230206
Device Name
Entarik Feeding Tube System; Entarik Feeding Tube
Applicant
Gravitas Medical, Inc.
Product Code
KNT · Gastroenterology, Urology
Decision Date
Mar 24, 2023
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 876.5980
Device Class
Class 2
Attributes
AI/ML, Therapeutic
Indications for Use
The Entarik Feeding Tube is intended for the administration of nutrition, fluids and medications by the nasoenteric route for patients who have an intact gastrointestinal tract but are physically unable to manage nutritional intake through normal mastication and deglutition. The Entarik Feeding Tube System is designed to aid, in conjunction with institutional protocols, qualified operators in the placement of the Entarik Enteral Feeding Tube (Entarik FT) into the stomach of patients requiring enteral feeding. The Entarik FT is equipped with sensors designed to provide information about the location of the tube tip relative to the stomach, thus assisting in reducing the incidence of misplacement during first positioning. The Entarik Monitor also monitors the feeding tube position continuously during the course of feeding and automatically and in real-time alerts of tube migration.
Device Story
System comprises nasogastric feeding tube (8Fr/12Fr) with embedded impedance/temperature sensors and portable electronic monitor. Used in clinical settings by qualified operators to guide initial tube placement into stomach and monitor for migration. Sensors transmit real-time impedance/temperature data to monitor; algorithm processes data to determine tube tip location relative to stomach and respiratory tract. Monitor provides visual/auditory alerts for tube migration. Benefits include reduced incidence of misplacement during initial positioning and continuous monitoring during feeding. Healthcare providers use monitor interface to verify placement and respond to migration alerts, facilitating safe enteral nutrition delivery.
Clinical Evidence
Feasibility study in 10 healthy volunteers verified system functionality and safety; 100% success rate in tube placement confirmed by X-ray. No major issues or unexpected adverse events reported. Porcine model study (n=8) validated algorithm's ability to detect respiratory tract placement at various depths.
Technological Characteristics
Nasogastric feeding tube (8Fr, 12Fr) with embedded impedance and temperature sensors. Portable electronic monitor for data processing/display. Biocompatibility per ISO 10993-1. Electrical safety per IEC 60601-1; EMC per IEC 60601-1-2. Software classified as moderate level of concern.
Indications for Use
Indicated for patients with intact gastrointestinal tracts unable to manage nutritional intake via mastication/deglutition requiring nasoenteric administration of nutrition, fluids, and medications.
Regulatory Classification
Identification
A gastrointestinal tube and accessories is a device that consists of flexible or semi-rigid tubing used for instilling fluids into, withdrawing fluids from, splinting, or suppressing bleeding of the alimentary tract. This device may incorporate an integral inflatable balloon for retention or hemostasis. This generic type of device includes the hemostatic bag, irrigation and aspiration catheter (gastric, colonic, etc.), rectal catheter, sterile infant gavage set, gastrointestinal string and tubes to locate internal bleeding, double lumen tube for intestinal decompression or intubation, feeding tube, gastroenterostomy tube, Levine tube, nasogastric tube, single lumen tube with mercury weight balloon for intestinal intubation or decompression, and gastro-urological irrigation tray (for gastrological use).
Special Controls
*Classification.* (1) Class II (special controls). The barium enema retention catheter and tip with or without a bag that is a gastrointestinal tube and accessory or a gastronomy tube holder accessory is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I (general controls) for the dissolvable nasogastric feed tube guide for the nasogastric tube. The class I device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 876.9.
{1}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
March 24, 2023
Gravitas Medical, Inc. Saheel Sutaria Chief Executive Officer 101 Mississippi St San Francisco, CA 94107
Re: K230206
Trade/Device Name: Entarik Feeding Tube System; Entarik Feeding Tube Regulation Number: 21 CFR 876.5980 Regulation Name: Gastrointestinal Tube And Accessories Regulatory Class: Class II Product Code: KNT Dated: January 24, 2023 Received: January 25, 2023
Dear Saheel Sutaria:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{2}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Shanil P. Haugen -S
Shanil P. Haugen, Ph.D. Assistant Director DHT3A: Division of Renal, Gastrointestinal, Obesity and Transplant Devices OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
# Indications for Use
510(k) Number (if known) K230206
Device Name EntarikTM Feeding Tube
Indications for Use (Describe)
The Entarik Feeding Tube is intended for the administration of nutrition, fluids and medications by the nasoenteric route for patients who have an intact gastrointestinal tract but are physically unable to manage nutritional intake through normal mastication and deglutition.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|--|
| | |
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
# Indications for Use
510(k) Number (if known) K230206
Device Name EntarikTM Feeding Tube System
#### Indications for Use (Describe)
The Entarik Feeding Tube System is designed to aid, in conjunction with institutional protocols, qualified operators in the placement of the Entarik Feeding Tube (Entarik FT) into the stomach of patients requiring enteral feeding, The Entarik FT is equipped with sensors designed to provide information about the tube tip relative to the stomach, thus assisting in reducing the incidence of misplacement during first positioning. The Entarik Monitors the feeding tube position continuously during the course of feeding and in real-time alerts of tube migration.
| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|
| <span style="font-family: Arial;"> <svg height="12" width="12"> <rect height="12" style="fill:none;stroke:#000000;stroke-width:1" width="12"></rect> <path d="M2 2 L10 10 M10 2 L2 10" stroke="#000000" stroke-width="1"></path> </svg> Prescription Use (Part 21 CFR 801 Subpart D) </span> |
|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| <span style="font-family: Arial;"> <svg height="12" width="12"> <rect height="12" style="fill:none;stroke:#000000;stroke-width:1" width="12"></rect> </svg> Over-The-Counter Use (21 CFR 801 Subpart C) </span> |
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{5}------------------------------------------------
### 510(k) Summary (21 CFR § 807.92(c))
This summary of 510(k) safety and effectiveness information is submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92.
| 510(k) Number: | K230206 |
|------------------------|-------------------------------------------------------------------------|
| Date Prepared: | January 24, 2023 |
| Applicant Information: | Gravitas Medical, Inc<br>101 Mississippi St.<br>San Francisco, CA 94107 |
| Contact: | Saheel Sutaria |
| Office Number: | 415-926-8616 |
| Email: | ssutaria@gravitasmedinc.com |
#### Device Information:
| Trade/Device Name: | Entarik™ Feeding Tube System<br>Entarik™ Feeding Tube |
|--------------------|-------------------------------------------------------|
| Common Name: | Feeding Tube and Accessories |
| Regulation Number: | 21 CFR 870.5980 |
| Regulation Name: | Gastrointestinal Tube and Accessories |
| Regulatory Class: | Class II |
| Product Code: | KNT |
| Panel: | Gastrointestinal and Urology Devices |
| Predicate Device: | smARTrack™ Feeding Tube System (K142327) |
#### Device Description
The Entarik Feeding Tube System (Entarik System) consists of the Enteral Feeding Tube and the Entarik Monitor. The Entarik Feeding Tube is a nasogastric feeding tube with a single lumen for the administration of nutrition, fluids and medications. The Entarik Feeding Tube includes impedance and temperature sensors embedded within the tube, with wires located in the wall of the feeding tube. The Entarik Monitor is a portable electronic device that serves to measure and record impedance and temperature data from the sensors on the Feeding Tube and provides information to the operator to aid in initial placement and monitoring of feeding tube position.
#### Indications for Use
#### Entarik Feeding Tube
The Entarik Feeding Tube is intended for the administration of nutrition, fluids and medications by the nasoenteric route for patients who have an intact gastrointestinal tract but are physically unable to manage nutritional intake through normal mastication and deglutition.
{6}------------------------------------------------
### Entarik Feeding Tube System
The Entarik Feeding Tube System is designed to aid, in conjunction with institutional protocols, qualified operators in the placement of the Entarik Enteral Feeding Tube (Entarik FT) into the stomach of patients requiring enteral feeding. The Entarik FT is equipped with sensors designed to provide information about the location of the tube tip relative to the stomach, thus assisting in reducing the incidence of misplacement during first positioning. The Entarik Monitor also monitors the feeding tube position continuously during the course of feeding and automatically and in real-time alerts of tube migration.
## Functional and Technological Comparison
The Entarik Feeding Tube System and smARTrack Feeding Tube System are both comprised of an electronic console which holds the system software and serves as the guiding interface to the user during initial tube placement as well as during ongoing use. The consoles both provide an alert when the system senses that the tube has moved out of position during ongoing use. Both the Entarik and smARTrack Feeding Tubes incorporate wired impedance sensors which transmit real-time information to the console regarding the feeding tube location.
There are some technological differences between the Entarik and smARTrack Feeding Tube Systems. The smARTrack device automatically stops feeding using a motorized mechanism in the event of tube dislodgement, whereas the Entarik device does not have that feature. The Entarik Feeding Tube System has an alert to notify the user in event of tube dislodgement. In addition to impedance sensors, the Entarik Feeding Tubes also incorporature sensors for additional information. The number, size, and location of the sensors, and the algorithms to determine tip location, are different between the Entarik and smARTrack systems.
To address the technological differences, the Entarik Feeding Tube System has undergone bench, pre-clinical, and clinical testing to ensure it functions correctly and achieves proper placement as well as correctly identifies tube malposition during the course of use. The performance data is provided in various sections of this submission. The Entarik System has successfully demonstrated its ability to guide the user in correct feeding tube placement. The system also correctly identifies tube movement. In the clinical study which was conducted, tube placement was initially verified with the Entarik System and a confirmatory X-Ray ("gold standard" placement verification) was performed. In all cases the Entarik System successfully guided the user to ensure correct placement was achieved. The available performance data supports the safety and effectiveness profile of the system and negates the technological differences between the Entarik System and the predicate device.
The Entarik feeding tubes are available in 8 and 12 Fr. The smARTrack feeding tube is available in 14Fr only. There are many FDA cleared feeding tubes which are available in various sizes ranging from 6 to 21 Fr. Therefore, this difference in tube diameter does not raise any new safety or effectiveness concerns as both devices are intended to be used in adult patients. Furthermore, biocompatibility tests were also performed to ensure the tubes are safe for use.
## Performance Data
Below is a list of the tests that have been performed and successfully completed for the Entarik Feeding Tube System.
{7}------------------------------------------------
# Biocompatibility:
The biocompatibility evaluation for the Entarik System Feeding Tube was conducted in accordance with the FDA Guidance "Use of ISO 10993 –1, Biological evaluation of medical devices – Part 1: Evaluation and testing within a risk management process" (June 2016) and International Standard ISO 10993-1:2018 "Biological Evaluation of Medical Devices – Part 1: Evaluation and Testing Within a Risk Management Process," as recognized by FDA. The Entarik Enteral Feeding Tube (FT) is considered to fall under the guidelines for a surface contacting device with direct mucosal tissue contact for a prolonged patient contact duration (>24 hours to 30 days) per the ISO/FDA guidelines and an actual patient exposure duration of less than <30 days in the clinical application.
The battery of testing included the following tests:
- -Cytotoxicity Study Using the ISO Elution Method
- ISO Intracutaneous Irritation Test -
- -ISO Maximization Sensitization Test
- -ISO Acute Systemic Toxicity Study
- ISO Sub-acute Systemic Toxicity Study -
- -ISO Intramuscular Implantation Test
- -ISO Material Mediated Pyrogenicity
# Electrical safety and electromagnetic compatibility (EMC):
The system was tested and found to comply with IEC 60601-1:2005/EN 60601-1:2006 Medical electrical equipment - Part 1: General requirements for basic safety and essential performance 3rd Ed and IEC 60601-1-2:2007 Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral standard: Electromagnetic compatibility -Requirements and tests.
## Software Verification and Validation Testing:
Software verification and validation testing were conducted and documentation was provided as recommended by FDA's Guidance for Industry and FDA Staff, "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices", (issued May 11, 2005). The software for this device was considered as a "moderate" level of concern, since a failure or latent flaw in the software could result in minor injury to the patient.
## Bench:
Bench testing performed on the Entarik Feeding Tube included 1) Gastric Soak and Preconditioning, 2) Dimensional. 3) Impedance and Temperature Accuracy and Sensing Rate. 4) Insertion and Withdrawal, 5) Flow and Pressure, 6) Buckle and Kink, 7) Tensile Integrity, 8) Simulated Use, 9) Radiopacity, 10) Product and Label Durability, and 11) Usability. These bench performance studies confirm that the device performed as intended per the product specifications and also demonstrated that the Entarik Feeding Tube System is substantially equivalent to the smARTrack Feeding Tube System predicate device.
## Human Factors:
{8}------------------------------------------------
The Entarik Feeding Tube System was evaluated by fifteen (15) intended users in a simulated clinical setting. This testing was conducted pursuant to the applicable requirements of FDA's Guidance Document titled "Applying Human Factors and Usability Engineering to Medical Devices (issued February 3, 2016)." For this evaluation, the users were trained by Gravitas personnel. Post training, users were provided with the Entarik Feeding Tube System and the proposed Instructions for Use. Tasks were assessed through observation during simulated use and through user knowledge assessments about the device's labeling and procedure steps that were not amendable to simulated use testing. Additionally, the test subjects completed a survey. The human factors assessment demonstrated that the device labeling and training provided to the intended users allowed for the proper use of the Entarik Feeding Tube System in its intended use environment. Further, this assessment confirmed that all identified risks were considered acceptable and that no new risks were identified.
#### Animal:
The Entarik Monitor contains an algorithm to determine if the tip of the feeding tube is in the respiratory tract. A study was conducted to develop and validate the algorithm in a porcine model. After induction of general anesthesia and placement of an endotracheal tube, the feeding tube was intentionally placed in the respiratory tract of eight pigs at various depths of insertion, and the algorithm was successful in detecting respirations in all eight pigs and at all depths of insertion.
#### Clinical:
Clinical studies were conducted to develop an algorithm to guide feeding tube placement into the stomach. A feasibility validation study was conducted in 10 healthy volunteers to assess the functionality and safety of the system. Tube insertions in all subjects were successful with no major issues reported. The entire procedure was well tolerated by the subjects. No unexpected adverse events were reported. Most patients that undergo feeding tube insertion experience some pain and discomfort and therefore none of the feedback that was received from the healthy volunteers was unexpected. It should be noted that the main target population for the system would be unconscious or anesthetized patients for whom minor pain and discomfort would be negligible. Furthermore, according to the patients' subjective feedback, throughout the procedure, no severe pain or discomfort was recorded. The convenience of the feeding tube insertion was assessed via questionnaires completed by the clinical staff. Correct placement of the tube was verified via X-Ray. No incorrect placements occurred while using the feeding tube.
#### Conclusion:
The Entarik Feeding Tube System has a similar intended use as the smARTrack predicate device. The Entarik Feeding Tube System is similar in technological and performance characteristics to the smARTrack predicate. The minor technological differences do not raise any new safety and effectiveness risks or concerns. It can be concluded that the Entarik Feeding Tube System is as safe and effective as the predicate device for the intended use, and is substantially equivalent.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.