Dentis I-FIX Abutment

K230203 · Dentis Co., Ltd. · NHA · May 3, 2023 · Dental

Device Facts

Record IDK230203
Device NameDentis I-FIX Abutment
ApplicantDentis Co., Ltd.
Product CodeNHA · Dental
Decision DateMay 3, 2023
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 872.3630
Device ClassClass 2

Intended Use

Dentis I-FIX Abutment is intended for use as an aid in prosthetic rehabilitation.

Device Story

Dentis I-FIX Abutment system comprises I-FIX Angled Type Fixture Healing Abutment, I-FIX Angled Type Fixture Healing Abutment Screw, and I-FIX Abutment Screw. Device functions as prosthetic rehabilitation aid; healing abutments protect fixture inner holes and adjust height during healing; abutment screws connect abutments to fixtures. Used in dental clinics by dentists/specialists. Healing abutment components provided sterile; abutment screws provided non-sterile for end-user sterilization. Output is physical support for dental prosthetics, facilitating restoration of masticatory function. Benefits include protection of implant site during healing and secure attachment of prosthetic components.

Clinical Evidence

No clinical data. Substantial equivalence established via bench testing, including sterilization validation (ISO 11137, ISO 17665), shelf-life testing (ASTM F1980), and biocompatibility assessment (ISO 10993-1). MR environment safety evaluated via scientific rationale and literature review.

Technological Characteristics

Materials: Pure Titanium Grade 4 (ASTM F67) for healing abutment body; Ti-6Al-4V ELI (ASTM F136) for screws. Dimensions: 4.03mm diameter for healing abutment; 2.15mm diameter for screws. Sterilization: Gamma radiation for healing abutment set; end-user sterilization for abutment screw. Connectivity: None (mechanical).

Indications for Use

Indicated for use as an aid in prosthetic rehabilitation in patients requiring dental implant abutments.

Regulatory Classification

Identification

An endosseous dental implant abutment is a premanufactured prosthetic component directly connected to the endosseous dental implant and is intended for use as an aid in prosthetic rehabilitation.

Special Controls

*Classification.* Class II (special controls). The guidance document entitled “Class II Special Controls Guidance Document: Root-Form Endosseous Dental Implants and Endosseous Dental Implant Abutments” will serve as the special control. (See § 872.1(e) for the availability of this guidance document.)

Predicate Devices

Reference Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is a blue square with the letters "FDA" in white, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue. Dentis Co., Ltd. % April Lee Consultant Withus Group Inc. 106 Superior Irvine, California 92620 Re: K230203 Trade/Device Name: Dentis I-FIX Abutment Regulation Number: 21 CFR 872.3630 Regulation Name: Endosseous Dental Implant Abutment Regulatory Class: Class II Product Code: NHA Dated: March 27, 2023 Received: March 27, 2023 Dear April Lee: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {1}------------------------------------------------ Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531- 542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely. # Andrew I. Steen -S Andrew I. Steen Assistant Director DHT1B: Division of Dental and ENT Devices OHT1: Office of Ophthalmic. Anesthesia. Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ # Indications for Use 510(k) Number (if known) K230203 Device Name Dentis I-FIX Abutment Indications for Use (Describe) Dentis I-FIX Abutment is intended for use as an aid in prosthetic rehabilitation. Type of Use (Select one or both, as applicable) X Prescription Use (Part 21 CFR 801 Subpart D) | Over-The-Counter Use (21 CFR 801 Subpart C) #### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ ### 510(k) Summary #### Submitter Dentis Co., Ltd. EunJin, Shin 99, Seongseoseo-ro, Dalseo-gu Daegu, 42718 Korea Email: ej0616@dentis.co.kr Tel. +82-53-589-3646 Fax. +82-53-289-7922 #### Device Information - 510(k) number: K230203 - Trade Name: Dentis I-FIX Abutment ● - Common Name: Dental Abutment System - Classification Name: Endosseous dental implant abutment ● - Product Code: NHA - Panel: Dental - Regulation Number: 872.3630 ● - Device Class: Class II ● - Date prepared: 05/03/2023 ● #### Predicate Devices: The subject device is substantially equivalent to the following predicate devices: #### Primary Predicate - . K210826, Healing Abutment, Cover Screw manufactured by MegaGen Implant Co., Ltd. Reference devices - K083586, I-FIX SYSTEM manufactured by Dentis Co., Ltd. ● - K111364, HAPTITE COATING IMPLANT SYSTEM by Dentis Co., Ltd. - K171027, Dentis Dental Implant System manufactured by Dentis Co., Ltd. ● #### Indication for Use: Dentis I-FIX Abutment is intended for use as an aid in prosthetic rehabilitation. #### Device Description: Dentis I-FIX Abutment consists of I-FIX Angled Type Fixture Healing Abutment, I-FIX Angled Type Fixture Healing Abutment Screw, and I-FIX Abutment Screw. I-FIX Angled Type Fixture Healing Abutment and I-FIX Angled Type Fixture Healing Abutment Screw are compatible with following Implant Systems: | Proprietary Name | I-FIX System (Angled Type Fixture) | |--------------------------------|------------------------------------| | Compatible Implants (K number) | K083586 | | Implant diameter size | Ø 2.0, 2.5, 3.0 mm | | Implant length | 10~16 mm | I-FIX Abutment Screw is only used with I-FIX Cemented Abutment and I-FIX Free Abutment cleared in K083586. #### Official Correspondent Withus Group Inc. April Lee 106 Superior, Irvine, CA 92620 USA Email: withus6664@gmail.com Phone: 1-909-274-9971 Fax: 1-909-460-8122 {4}------------------------------------------------ The dimensions of abutments are as following: | No. | Device Name | Dimension (mm) | Angulation | |-----|-----------------------------------------------------|--------------------------------------|------------| | 1 | I-FIX Angled Type Fixture<br>Healing Abutment | (D) 4.03 X (L) 3.2, 4, 5 and 7 | 0° | | 2 | I-FIX Angled Type Fixture<br>Healing Abutment Screw | (D) 2.15 X (L) 4.0, 4.8, 6.1 and 8.0 | 0° | | 3 | I-FIX Abutment Screw | (D) 2.15 X (L) 4.7 | 0° | Tolerance of dimension for Abutments shall be within ± 1% range. The Abutments have below featured: | Name | Uses | Surface | Materials | |-----------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------|---------|------------------| | I-FIX Angled Type Fixture Healing<br>Abutment | The healing Abutment is used for<br>protecting inner hole of fixture and | Non | Titanium Grade 4 | | I-FIX Angled Type Fixture Healing<br>Abutment Screw | adjusting the appropriate height during the<br>healing period | Non | Ti-6Al-4V ELI | | I-FIX Abutment Screw | I-FIX Abutment screw is used with the<br>previously cleared devices, K083586 such<br>as I-FIX Cemented and Free Abutment for<br>connecting with fixture. | Non | Ti-6Al-4V ELI | I-Fix Angled Type Fixture Healing Abutment and I-FIX Angled Type Fixture Healing Abutment Screw are packaged as a set and provided sterilized. I-FIX Abutment Screw is provided non-sterilized. ### Materials - I-FIX Angled Type Fixture Healing Abutment is fabricated from pure titanium (Conforming to . ASTM Standard F67) - I-FIX Angled Type Fixture Healing Abutment Screw and I-FIX Abutment Screw are fabricated . from Ti-6Al-4V ELI (Conforming to ASTM Standard F136) {5}------------------------------------------------ # Summaries of Technological Characteristics & Substantial Equivalence Discussion | | | | I-FIX Angled Type Fixture Healing Abutment & I-FIX Angled Type Fixture Healing Abutment Screw | | | |---------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--|--| | | Subject Device | Primary Predicate | Reference Device | | | | K number | K230203 | K210826 | K171027 | | | | Manufacturer | Dentis Co., Ltd. | MegaGen Implant Co., Ltd. | Dentis Co., Ltd. | | | | Device Name | Dentis I-FIX Abutment | Healing Abutment, Cover Screw | Dentis Dental Implant System | | | | Model Name | I-FIX Angled Type Fixture Healing Abutment & I-FIX Angled Type Fixture Healing Abutment Screw | Healing Abutment & Scan Healing Abutment Screw | s-Clean Healing Abutment | | | | Design | Image: I-FIX Angled Type Fixture Healing Abutment & I-FIX Angled Type Fixture Healing Abutment Screw | Image: Healing Abutment & Scan Healing Abutment Screw | Image: s-Clean Healing Abutment | | | | Diameter | Ø 4.03 mm | Ø 4.0 ~ 10.0 mm | Ø 4.0, 4.5, 4.8, 5.0, 5.5, 6.0, 6.5, 7.0 and 7.5 mm | | | | Cuff Length | 0.8 mm | 2.0~7.0 mm | 1.0, 1.5, 2.0, 2.5, 3.0, 3.5, 4.0, 4.5, 5.0, 5.5 mm | | | | Male Screw x Pitch | M1.6 X 0.35P | M1.6 X 0.35P<br>M1.8 X 0.35P<br>M2.0 X 0.35P | M1.6 X 0.35P | | | | Coating | Non | Non<br>Screw: Anodizing | Non | | | | Material | Body: CP Titanium Gr4 (ASTM F67)<br>Screw: Ti-6Al-4V ELI (ASTM F136) | Ti-6Al-4V ELI (ASTM F136) | Titanium Gr4 (ASTM F67) | | | | Indications for Use | Dentis I-FIX Abutment is intended for use as an aid in prosthetic rehabilitation. | MegaGen Prosthetics are intended for use as an aid in prosthetic rehabilitation. | Dentis Dental Implant System is indicated for use in partially or fully edentulous mandibles and maxillae, in support of single or multiple unit restorations including cemented retained, screw retained, or overdenture restorations. Dentis Dental Implant System serves as temporary support for provisional prosthetic device during the healing phase of permanent endosseous dental implants. This is an implant abutment support prosthesis such as the artificial teeth used temporarily for the restoration of masticatory function of patients. | | | | Sterilization | Gamma sterile | Gamma sterile | Gamma sterile | | | | Comparison | When comparing the subject device to the primary predicate, the indication for use, use with screw, Male Screw X Pitch, function, and coating are the same. The difference between two devices are material and dimensions. To support the difference of material, K171027 was added. The dimensions of subject device are slightly smaller than the identified predicate devices; however, this difference is not an important factor to device performance. Therefore, subject device and predicate device are substantially equivalent. | | | | | # I-FIX Angled Type Fixture Healing Abutment & I-FIX Angled Type Fixture Healing Abutment Screw {6}------------------------------------------------ #### I-FIX Abutment Screw | | Subject Device | Reference Device | |--------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------| | K number | K230203 | K083586 | | Manufacturer | Dentis Co., Ltd. | Dentis Co., Ltd. | | Model Name | I-FIX Abutment Screw | s-Clean Abutment Screw | | Design | Image: I-FIX Abutment Screw | Image: s-Clean Abutment Screw | | Head Diameter | Ø 2.15 mm | Ø 2.2 mm | | Length | 4.7mm | 4.05mm | | Male Screw x Pitch | M1.6 X 0.35P | M1.6 X 0.35P | | Coating | Non | Non | | Material | Ti-6Al-4V ELI (ASTM F136) | Ti-6Al-4V ELI (ASTM F136) | | Sterilization | End User Sterilization | End User Sterilization | | Comparison | Subject Device and Reference Device have the same indications for use, function material,<br>coating, and sterilization method. The Head diameter is different; however, this difference<br>is not an important factor to device performance. Therefore, subject device and predicate<br>device are substantially equivalent. | | # Non-Clinical Test Data Below tests were performed for predicate devices and leveraged for the subject device: - Sterilization validation test on healing abutment made of Titanium Gr4 according to ISO . 11137-1,2,3 referenced in K210134 - Sterilization validation test on healing abutment made of Ti-6A1-4A1 ELI according to ISO ● 11137-1,2,3 - End User Sterilization Validation Test Report on Abutments according to ANSI/AAMI ST79, ● ISO 17665-1,-2, ISO 11737-1,-2, and ISO 11138-1 referenced in K111364 - Shelf Life Test on Healing Abutment according to ASTM F1980 referenced in K171027 - Biocompatibility testing according to ISO 10993-1:2009 for Pure Titanium Grade 4 or Ti-6Al-● 4V ELI referenced in K210134 and K16124 The results of the above tests have met the criteria of the standards and demonstrated the substantial equivalence with the predicate device. Non-clinical test data was conducted in accordance with FDA Guidance "Class II Special Controls Guidance Document: Root-form Endosseous Dental Implants and Endosseous Dental Implant Abutments". Clinical testing was not necessary to establish substantial equivalency of the device. # MR Environment Condition Non-clinical worst-case MRI review was performed to evaluate the metallic Dentis I-FIX Abutment in the MRI environment using scientific rationale and published literature (e.g., Woods, Terry O., Jana G. Delfino, and Sunder Rajan. 'Assessment of Magnetically Induced Displacement Force and Torque on Metal Alloys Used in Medical Devices.' Journal of Testing and Evaluation 49.2 (2019): 783-795), based on the entire system including all variations (all compatible implant bodied, dental abutments, and fixation screws) and material composition. Rationale addressed parameters per the FDA Guidance "Testing and Labeling Medical Devices for Safety in the Magnetic Resonance (MR) Environment," including magnetically induced displacement force and torque. {7}------------------------------------------------ # Conclusion Dentis I-FIX Abutment constitutes a substantially equivalent medical device, meeting all the declared requirements of its intended use. This system has the same intended use and fundamental scientific technology as its predicate devices. Therefore, Dentis I-FIX Abutment and its predicates are substantially equivalent.