EMPOWR Revision VVC+ (Varus / Valgus Constrained), e+ Tibial Insert
Applicant
Encore Medical, L.P Dba Enovis
Product Code
JWH · Orthopedic
Decision Date
Apr 6, 2023
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.3560
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
Joint replacement is indicated for patients suffering from disability due to: - degenerative, post-traumatic or rheumatoid arthritis; - avascular necrosis of the femoral condyle; - post-traumatic loss of joint configuration, particularly when there is patellofemoral erosion, dysfunction or prior patellectomy; - moderate valgus, varus or flexion deformities; - treatment of fractures that are unmanageable using other techniques. This device may also be indicated in the salvage of previously failed surgical attempts. All devices are intended for cemented applications except for the 3DKNEE™ Porous Coated Femur, EMPOWR 3DKNEE™ POROUS Femur, 3DKNEE™ Porous Coated Tibia, and EMPOWR POROUS Tibia which are intended for cementless applications. While knee replacements are not intended to withstand activity levels and loads of normal healthy bone, they are a means of restoring mobility and reducing pain for many patients. The EMPOWR Revision VVC+, e+ Tibial Insert should be considered for use in total knee arthroplasty for patients under the following indications: - absence or loss of both cruciate ligaments - moderate varus-valgus or flexion instability that requires a bearing surface with increased constraint in the clinical judgement of the surgeon - bone loss that requires supplemental fixation in the clinical judgement of the surgeon
Device Story
The EMPOWR Revision VVC+ e+ Tibial Insert is a knee arthroplasty component providing increased femoral/tibial constraint compared to the predicate. It functions as a bearing surface to provide internal/external and varus/valgus rotational stability. The device is implanted by an orthopedic surgeon during total knee arthroplasty. It interfaces with the EMPOWR Revision Femoral implant and a Wrought CoCr reinforcement pin accessory. The insert is manufactured from highly crosslinked UHMWPE with vitamin E. The device restores joint mobility and reduces pain in patients with severe knee instability or bone loss. Clinical decision-making is supported by the surgeon's assessment of the need for increased constraint and supplemental fixation.
Clinical Evidence
No clinical data submitted. Bench testing only, including CAD-based range of motion, femoral-tibial contact area assessment (ASTM F2083-21), intrinsic stability (ASTM F2083-21, ASTM F1223-20), varus/valgus constraint testing (ASTM F2083-21, ASTM F1223-20), and MR conditional labeling (ASTM F2052-21, F2213-17, F2119-07, F2182-10e2, F2503-20).
Technological Characteristics
Material: Highly crosslinked 75kGy UHMWPE with vitamin E (ASTM F2695/F2565). Reinforcement pin: Wrought CoCr (ASTM F1537). Design: 14 sizes, thicknesses 10-25mm. Features: Increased post width and articulating geometry for higher I/E and V/V constraint. Connectivity: None (mechanical implant). Sterilization: Not specified.
Indications for Use
Indicated for patients with disability due to degenerative, post-traumatic, or rheumatoid arthritis; avascular necrosis of femoral condyle; post-traumatic loss of joint configuration; moderate valgus, varus, or flexion deformities; or fractures unmanageable by other techniques. Also indicated for salvage of failed surgical attempts. Specific indications for VVC+ insert include absence/loss of both cruciate ligaments, moderate varus-valgus or flexion instability requiring increased constraint, and bone loss requiring supplemental fixation.
Regulatory Classification
Identification
A knee joint patellofemorotibial polymer/metal/polymer semi-constrained cemented prosthesis is a device intended to be implanted to replace a knee joint. The device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across-the-joint. This generic type of device includes prostheses that have a femoral component made of alloys, such as cobalt-chromium-molybdenum, and a tibial component or components and a retropatellar resurfacing component made of ultra-high molecular weight polyethylene. This generic type of device is limited to those prostheses intended for use with bone cement (§ 888.3027).
Predicate Devices
DJO EMPOWR Revision Varus / Valgus Constrained Tibial Insert (K180930)
Submission Summary (Full Text)
{0}------------------------------------------------
April 6, 2023
Image /page/0/Picture/1 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health and Human Services logo on the left and the FDA logo on the right. The FDA logo is a blue square with the letters "FDA" in white, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue.
Encore Medical, L.P. dba Enovis Trey Thorsen Regulatory Affairs Consultant 9800 Metric Boulevard Austin, Texas 78758
# Re: K230169
Trade/Device Name: EMPOWR Revision VVC+ (Varus / Valgus Constrained), e+ Tibial Insert Regulation Number: 21 CFR 888.3560 Regulation Name: Knee Joint Patellofemorotibial Polymer Semi-Constrained Cemented Prosthesis Regulatory Class: Class II Product Code: JWH, OIY Dated: January 20, 2023 Received: January 20, 2023
Dear Trey Thorsen:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Peter G. Allen -S
Digitally signed by Peter G. Allen -S Date: 2023.04.06 17:15:52 -04'00'
For: [Vacant] Assistant Director DHT6A: Division of Joint Arthroplasty Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K230169
### Device Name
EMPOWR RevisionVVC +(VARUS / VALGUS CONSTRAINED), e + TIBIAL INSERT
Indications for Use (Describe)
Joint replacement is indicated for patients suffering from disability due to:
- · degenerative, post-traumatic or rheumatoid arthritis;
- · avascular necrosis of the femoral condyle;
- · post-traumatic loss of joint configuration, particularly when there is patellofemoral erosion, dysfunction or prior patellectomy;
- · moderate valgus, varus or flexion deformities;
- · treatment of fractures that are unmanageable using other techniques.
This device may also be indicated in the salvage of previously failed surgical attempts. All devices are intended for cemented applications except for the 3DKNEE™ Porous Coated Femur, EMPOWR 3DKNEE™ POROUS Femur, and EMPOWR POROUS Tibia which are intended for cementless applications.
While knee replacements are not intended to withstand activity levels and loads of normal healthy bone, they are a means of restoring mobility and reducing pain for many patients.
The EMPOWR Revision VVC+, e+ Tibial Insert should be considered for use in total knee arthroplasty for patients under the following indications:
· absence or loss of both cruciate ligaments
- · moderate varus-valgus or flexion instability that requires a bearing surface with increased constraint in the clinical judgement of the surgeon
- · bone loss that requires supplemental fixation in the clinical judgement of the surgeon
Type of Use (Select one or both, as applicable)
| For Evaluation (Test & STEP 201 Subject B) | |
|--------------------------------------------|--|
| For Examination (STEP 201 Subject C) | |
× Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW!*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number.
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for Enovis. The word "enovis" is written in a sans-serif font. The "e", "n", and the first half of the "o" are red, while the second half of the "o", "v", "i", and "s" are black. There is a trademark symbol to the right of the "s".
| Page 1 of 3
510(k) Summary
## I. SUBMITTER
Encore Medical, L.P. (dba Enovis) 9800 Metric Blvd. Austin, TX 78758
Contact Person: Trey Thorsen Email: trey.thorsen@enovis.com Phone: 850-450-3932
# II. DEVICE
Name of Device: EMPOWR Revision VVC+ (Varus / Valgus Constrained), e+ Tibial Insert Common or Usual Name: Total Knee Implant Classification Name: Knee joint patellofemorotibial polymer/metal/polymer semi-constrained cemented prosthesis per 21 CFR 888.3560
Regulatory Class: II Primary Product Code: JWH Secondary Product Code: OIY
# III. PREDICATE DEVICE
DJO EMPOWR Revision Varus / Valgus Constrained Tibial Insert, K180930
# IV. DEVICE DESCRIPTION
EMPOWR Revision VVC+ (Varus / Valgus Constrained) e+ Tibial Insert
The EMPOWR Revision VVC+ (Varus / Valgus Constrained) e+ Tibial Insert Implant is a line extension that offers an increased level of femoral / tibial constraint over the EMPOWR VVC Tibial Insert. This tibial insert implant provides additional internal / external and varus / valgus rotational stability.
The sizing architecture for the insert includes 14 sizes (2-11, 4BU, 5BU, 6BD, 7BD) and thicknesses of 10mm, 12mm, 14mm, 16mm, 19mm, 22mm and 25mm with post medial/lateral width of 0.5520 inches for sizes 2-5 and 0.7080 inches for sizes 6 - 11. Along with the post width, the articulating geometry of the proximal side of the tibial insert accommodates an additional amount of internal / external (I/E), varus / valgus (V/V) constraint, and tibial post / femoral cam jump height when interfacing with the EMPOWR Revision Femoral implant. The material (highly crosslinked 75kGy UHMWPE with vitamin E per ASTM F2695/F2565) and distal locking mechanism geometry with the mating tibial baseplate are identical to the predicate.
All tibial insert thickness options are configured to interface with a reinforcement pin accessory made of Wrought CoCr per ASTM F1537. This accessory is configured with a taper interface on the distal end to mate with the EMPOWR UNIVERSAL TIBIAL BASEPLATE, has a diameter and length configured to reinforce the post and thickness for the EMPOWR Revision VVC+ e+ tibial insert, and contains barbs to
{4}------------------------------------------------
| Page 2 of 3
prevent the pin from backout when fully seated within the post of the EMPOWR Revision VVC+, e+ tibial insert.
# V. INDICATIONS FOR USE
Joint replacement is indicated for patients suffering from disability due to:
- . degenerative, post-traumatic or rheumatoid arthritis;
- . avascular necrosis of the femoral condyle;
- . post-traumatic loss of joint configuration, particularly when there is patellofemoral erosion, dysfunction or prior patellectomy;
- moderate valgus, varus or flexion deformities;
- . treatment of fractures that are unmanageable using other techniques.
This device may also be indicated in the salvage of previously failed surgical attempts. All devices are intended for cemented applications except for the 3DKNEE™ Porous Coated Femur, EMPOWR 3DKNEE™ POROUS Femur, 3DKNEE™ Porous Coated Tibia, and EMPOWR POROUS Tibia which are intended for cementless applications.
While knee replacements are not intended to withstand activity levels and loads of normal healthy bone, they are a means of restoring mobility and reducing pain for many patients.
The EMPOWR Revision VVC+, e+ Tibial Insert should be considered for use in total knee arthroplasty for patients under the following indications:
- absence or loss of both cruciate ligaments
- . moderate varus-valgus or flexion instability that requires a bearing surface with increased constraint in the clinical judgement of the surgeon
- . bone loss that requires supplemental fixation in the clinical judgement of the surgeon
# VI. COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICE
The EMPOWR Revision VVC+, e+ Tibial Insert and accessories are a line extension to the predicate EMPOWR Varus / Valgus Constrained Tibial Insert (cleared via K180930).
Comparative testing demonstrates substantial equivalence between the subject and predicate device.
# Biocompatibility testing
The biocompatibility evaluation for the EMPOWR Revision VVC+, e+ Tibial Insert was conducted in accordance with the FDA guidance, Use of International Standard ISO-10993, 'Biological Evaluation of Medical Devices Part 1: Evaluation and Testing, May 1, 1995, and International Standard ISO 10993-1 Biological Evaluation of Medical Devices – Part 1: Evaluation and Testing Within a Risk Management Process, as recognized by FDA.
The EMPOWR Revision VVC+, e+ Tibial Insert is intended for permanent implantation, contacting tissue/bone. Biocompatibility was ensured through the use of qualified materials and contact agents.
{5}------------------------------------------------
Image /page/5/Picture/1 description: The image shows the logo for "enovis". The first two letters, "en", are in red, while the "o" is a black circle with a gradient effect. The letters "vis" are in black, and there is a small trademark symbol in the lower right corner of the logo.
## Performance Testing
The following testing was performed to FDA recognized standards and internal protocols:
- . CAD Based Range of Motion (ROM)
- Femoral – Tibial Contact Area Assessment - ASTM F2083-21
- O Femur - Tibial Intrinsic Stability - ASTM F2083-21, ASTM F1223-20
- . Femur - Tibial Varus / Valgus Constraint Testing- ASTM F2083-21, ASTM F1223-20
- . MR Conditional Labeling - ASTM F2052 -21, ASTM F2213-17, ASTM F2119-07(2013), ASTM F2182-10e2, ASTM F2503-20
# Animal Studies
No animal data submitted.
### Clinical Studies
No clinical data submitted.
### VIII. CONCLUSIONS
All testing and evaluations demonstrate that the device is substantially equivalent to the predicate.
| Page 3 of 3
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.