K230142 · Biomed Ent, Inc. · LYA · Mar 22, 2023 · Ear, Nose, Throat
Device Facts
Record ID
K230142
Device Name
Epi-Stop Nasal Gel/epistaxis pack
Applicant
Biomed Ent, Inc.
Product Code
LYA · Ear, Nose, Throat
Decision Date
Mar 22, 2023
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 874.4780
Device Class
Class 1
Attributes
Therapeutic
Indications for Use
Epi-Stop™ Nasal Gel/Epistaxis Pack is a sterile, single use in patients suffering from anterior epistaxis to: - Help control minimal bleeding following trauma by barrier function, blood absorption and platelet aggregation; - Act as an adjunct to aid in the natural healing process. Epi-Stop™ Nasal Gel/Epistaxis Pack is indicated for use as a nasal packing to treat anterior epistaxis. Epi-Stop™ Nasal Gel/Epistaxis Pack is intended for use under the direction of a licensed healthcare provider.
Device Story
Epi-Stop Nasal Gel/Epistaxis Pack is a sterile, single-use, injectable hemostatic gel; composed of crosslinked porcine-derived gelatin and hyaluronic acid. Used in anterior nasal cavity to treat epistaxis; applied by medical professional or trained personnel via syringe and applicator. Gel creates moist environment; provides barrier function; absorbs blood; facilitates platelet aggregation. Remains in situ for at least 72 hours; removed via gentle irrigation with water. Device provides tamponade effect; adapts to implant area due to viscoelasticity. Benefits patient by stopping nosebleeds and supporting healing. Used in clinical settings under direction of licensed healthcare provider.
Clinical Evidence
Bench testing and in vivo animal study. Bench tests compared absorption, occlusion, dynamic viscosity, extrusion force, and pH against predicate. Both devices showed comparable water absorption and occlusion (no leakage at 60 mins). In vivo animal study confirmed both devices remained at injection site for 72 hours and successfully stopped anterior mild bleeding within the same acceptance criteria range.
Technological Characteristics
Injectable nasal gel; materials: porcine-derived gelatin, hyaluronic acid, phosphate buffer solution. Supplied sterile (steam sterilization, SAL 10^-6). Single-use. Delivered via pre-filled syringe with applicator. Biocompatibility per ISO 10993-1, 10993-5, 10993-10, 10993-11, and 10993-23.
Indications for Use
Indicated for patients suffering from anterior epistaxis to control minimal bleeding following trauma and aid in natural healing. Contraindicated in patients with known hypersensitivity/allergy to porcine-derived gelatin or its derivatives.
Regulatory Classification
Identification
An intranasal splint is intended to minimize bleeding and edema and to prevent adhesions between the septum and the nasal cavity. It is placed in the nasal cavity after surgery or trauma. The intranasal splint is constructed from plastic, silicone, or absorbent material.
{1}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which consists of the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
March 22, 2023
BioMed ENT, Inc. % Marcela Garcia Principal Investigator BioMed Elements B.V. Kerkenbos 1077-V Nijmegen, Gelderland 6546 BB Netherlands
Re: K230142
Trade/Device Name: Epi-Stop Nasal Gel/Epistaxis Pack Regulation Number: 21 CFR 874.4780 Regulation Name: Intranasal Splint Regulatory Class: Class I Product Code: LYA Dated: January 18, 2023 Received: January 18, 2023
Dear Marcela Garcia:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal
{2}------------------------------------------------
statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Joyce C. Lin -S
for Shu-Chen Peng, Ph.D. Assistant Director DHT1B: Division of Dental and ENT Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
## Indications for Use
510(k) Number (if known) K230142
#### Device Name Epi-Stop™ Nasal Gel/Epistaxis Pack
#### Indications for Use (Describe)
Epi-Stop™ Nasal Gel/Epistaxis Pack is a sterile, single use in patients suffering from anterior epistaxis to:
- Help control minimal bleeding following trauma by barrier function, blood absorption and platelet aggregation;
- Act as an adjunct to aid in the natural healing process.
Epi-Stop™ Nasal Gel/Epistaxis Pack is indicated for use as a nasal packing to treat anterior epistaxis.
Epi-Stop™ Nasal Gel/Epistaxis Pack is intended for use under the direction of a licensed healthcare provider.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|--|
|-------------------------------------------------|--|
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
Image /page/3/Picture/0 description: The image contains the logo for BioMed ENT. The words "BioMed" are in orange, and the word "ENT" is in gray. To the right of the words is a graphic of seven hexagons arranged in a triangular shape. The hexagons are colored in shades of orange and blue.
# 510(K) SUMMARY
In accordance with the Food and Drug Administration Rule to implement provisions of the Safe Medical Devices Act of 1990 and in conformance with 21 CFR 807.92, this information serves as a Summary of Safety and Effectiveness for the use of Epi-Stop™ Nasal Gel/Epistaxis Pack.
| Applicant: | BioMed ENT, Inc.<br>18911 Hardy Oak Blvd,<br>San Antonio, TX 78258,<br>United States,<br>210-846-0692 |
|------------------------|-------------------------------------------------------------------------------------------------------|
| Trade name: | Epi-Stop™ Nasal Gel/Epistaxis Pack |
| Common name: | Nasal Packing |
| Classification name: | Intranasal Splint |
| Number: | 21 CFR 874.4780 |
| Device Classification: | Class I |
| Device Product Code: | LYA |
| Device Panel: | Ear, Nose and Throat |
| Predicate Device: | Chitogel® Endoscopic Sinus Surgery Kit (K172179) |
| Contact Person: | Tonny Voermans<br>President<br>T: +31(0) 24 711 4079<br>E: t.voermans@biomed-ent.com |
| Date: | 2023 March 21 |
## Device Description
Epi-Stop™ Nasal Gel/Epistaxis Pack is a single use, hemostatic, and injectable gel, made of crosslinked gelatin and hyaluronic acid, indicated for the treatment of anterior epistaxis. Epi-Stop™ Nasal Gel/Epistaxis Pack creates a moist environment for healing and the ability of absorbing and stopping nose bleeding within minutes. Due to its viscoelasticity and low extrusion force, Epi-Stop™ Nasal Gel/Epistaxis Pack can be injected into the anterior nasal cavity and the gel must be left in the nose for at least 72 hours. Afterwards, the gel can be
{5}------------------------------------------------
Image /page/4/Picture/0 description: The image contains the logo for BioMed ENT. The words "BioMed" are in orange, and the word "ENT" is in gray. To the right of the words is a graphic of seven hexagons arranged in a triangular shape. The hexagons are colored in shades of orange and blue.
removed by using safe water for gentle irrigation (either sterile or boiled water) or using safe lukewarm water.
Principle of operation. In Figure 1 the commercial presentation of Epi-Stop™ Nasal Gel/Epistaxis Pack is shown and it consists of a cardboard box (secondary package) as an outer package (a). Inside the box, a leaflet with the instructions for use (IFU) can be found, together with a sterile sealed-pouch (primary package) containing a capped syringe and an applicator (b). The sterile pouch can be peeled-open, and the capped syringe and applicator can be taken out of the primary package (c). The gel is contained and sealed in this syringe and should only be injected through the applicator by a medical professional or trained personnel. To properly use Epi-Stop™ Nasal Gel/Epistaxis Pack, a medical professional or trained personnel should twist-and-remove the cap tip from the syringe (d) and attach the applicator instead (e). The applicator can then be introduced in the anterior nasal cavity against the injury site, and the gel can be injected in one smooth notion in compliance with the IFU.
Image /page/4/Figure/4 description: The image shows five different views of a medical product called Epi-Stop Nasal Gel. Image a) shows the product's packaging, which is a box with the product name and the manufacturer's logo. Images b) through e) show the product itself, which is a syringe filled with a gel-like substance. The syringe is shown in different stages of use, with the gel being dispensed from the syringe in images d) and e).
Figure 1. Visual representation of Epi-Stop™ Nasal Gel/Epistaxis Pack and its key components.
Mechanism of Action. Briefly, to successfully stop anterior mild bleeding, Epi-Stop™ Nasal Gel/Epistaxis Pack relies primarily on an optimal occlusion, followed by good absorption to retain blood, and lastly on an optimal viscosity designed to facilitate the subject device's ability to properly adapt to the implant area once it has been injected in the anterior nasal cavity.
Device components. A brief summary in Table 1 of the product components with the duration and type of contact.
| together with duration and type of contact for each of the items listed. | | | |
|--------------------------------------------------------------------------|---------------------|-----------------------------------|------------|
| | Device components | Type of contact | Duration |
| 1 | Cardboard box | Not a patient-contacting material | N/A |
| 2 | IFU | Not a patient-contacting material | N/A |
| 3 | Pouch | Not a patient-contacting material | N/A |
| 4 | Cap tip | Not a patient-contacting material | N/A |
| 5 | Syringe | patient-contacting material. | < 1 minute |
| 6 | Applicator1 | patient-contacting material. | < 1 minute |
| 7 | Epi-Stop™ Nasal Gel | patient-contacting material. | 72 hours |
Table 1. This table contains a list of all product components of Epi-Stop™ Nasal Gel/Epistaxis Pack,
{6}------------------------------------------------
Image /page/5/Picture/0 description: The image contains the logo for BioMed ENT. The words "BioMed" are in orange, and the word "ENT" is in gray. To the right of the words is a design made up of seven hexagons in three different colors. The top hexagon is orange, the two hexagons below it are light orange, the two hexagons below that are light blue, and the two hexagons at the bottom are dark blue.
### Indications for use
Epi-Stop™ Nasal Gel/Epistaxis Pack is a sterile, single use device indicated for use in patients suffering from anterior epistaxis to:
- -Help control minimal bleeding following trauma by barrier function, blood absorption and platelet aggregation;
- Act as an adjunct to aid in the natural healing process. -
Epi-Stop™ Nasal Gel/Epistaxis Pack is indicated for use as a nasal packing to treat anterior epistaxis.
Epi-Stop™ Nasal Gel/Epistaxis Pack is intended for use under the direction of a licensed healthcare provider.
### Contraindications:
This product is contraindicated for use in patients with a known hypersensitivity/allergy to porcine-derived gelatin or its derivatives.
### Possible Adverse Reactions:
Possible adverse effects associated with product use can include Allergic reactions, Scarring/synechia and Toxic syndrome (rare). For more information, please read Warnings and Precautions section in the Instructions For Use.
{7}------------------------------------------------
Image /page/6/Picture/0 description: The image contains the logo for Biomed ENT. The words "BioMed" are in orange, and the word "ENT" is in gray. To the right of the words is a design made up of seven hexagons. Three of the hexagons are orange, two are light blue, and two are dark blue.
# Comparison of Technological Characteristics with the Predicate Device.
Table 2. Comparison of predicate device with subject device Epi-Stop™ Nasal Gel/Epistaxis Pack
| | Subject device | Predicate device | |
|-------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--|
| | K230142 | K172179 | |
| | Epi-StopTM Nasal Gel/Epistaxis<br>Pack | Chitogel® Endoscopic Sinus<br>Surgery Kit | |
| | BioMed ENT, Inc. | Chitogel Ltd | |
| Product Codes | LYA | LYA, | |
| Regulations | 874.4780 | 874.4780, | |
| Device<br>Classification | Class I | Class I | |
| | Indications for use: | Indications for use: | |
| Indications for<br>use | Epi-StopTM Nasal Gel/Epistaxis<br>Pack is a sterile, single use<br>device indicated for use in<br>patients suffering from anterior<br>epistaxis to:<br>- Help control minimal bleeding<br>following trauma by barrier<br>function, blood absorption and<br>platelet aggregation;<br>- Act as an adjunct to aid in the<br>natural healing process. | Chitogel® Endoscopic Sinus<br>Surgery Kit is indicated for use in<br>patients undergoing nasal/sinus<br>surgery as a space occupying<br>packing to:<br>- Separate tissue or structures<br>compromised by surgical trauma.<br>- Separate and prevent adhesions<br>between mucosal surfaces in the<br>nasal cavity.<br>- Control minimal bleeding | |
| | Epi-StopTM Nasal Gel/Epistaxis<br>Pack is indicated for use as a<br>nasal packing to treat anterior<br>epistaxis. | following surgery or trauma by<br>tamponade effect, blood<br>absorption and platelet<br>aggregation.<br>- Act as an adjunct to aid in the | |
| | Epi-StopTM Nasal Gel/Epistaxis<br>Pack is intended for use under<br>the direction of a licensed<br>healthcare provider. | natural healing process.<br>The Chitogel® Endoscopic Sinus<br>Surgery Kit is indicated for use as<br>a nasal packing to treat epistaxis. | |
| Technological<br>Characteristics | Injectable Nasal Gel | Injectable Nasal Gel | |
| | Subject device | Predicate device | |
| | Epi-Stop™ Nasal Gel/Epistaxis<br>Pack is a single use, injectable<br>gel, supplied sterile.<br>Epi-Stop™ Nasal Gel/Epistaxis | Chitogel® Endoscopic Sinus<br>Surgery Kit is a single use,<br>injectable gel, supplied sterile.<br>In Chitogel® Endoscopic Sinus | |
| How supplied | Pack consists of a pre-filled<br>syringe containing the gel, and an<br>attachment/applicator to help<br>deliver the gel in the<br>Kiesselbach's plexus. | Surgery Kit all components of the<br>gel need to be mixed on-site by<br>the physician, using the syringe<br>and the attachment provided. The<br>pre-measured components mix to<br>form the biodegradable gel. | |
| Materials | -Porcine-derived gelatin (derived<br>from collagen) and<br>-Hyaluronic acid and | - Chitosan and<br>- Dextran Aldehyde Powder and<br>- Phosphate Buffer Solution | |
| | -Phosphate Buffer solution | | |
| | Method of Action. | Method of Action. | |
| | Gel injected in the anterior nasal<br>cavity. | Gel injected in the sinus cavity | |
| | Removal. | Removal. | |
| Method of | Epi-Stop™ Nasal | Chitogel® Endoscopic Sinus | |
| Action/Removal | Gel/Epistaxis Pack can be | Surgery Kit is eliminated via<br>gentle irrigation using saline or | |
| | removed by using safe water<br>for gentle irrigation (either | water. | |
| | sterile or boiled water) or using | | |
| | safe lukewarm water. | | |
| Sterilization | Supplied sterile - Steam<br>sterilization | Supplied sterile - ETO & Gamma | |
| Sterilization<br>10-6 SAL<br>Assurance<br>Level | | 10-6 SAL | |
| Single Use | Yes | Yes | |
| Shelf life | 18 months | 3 months | |
| Biocompatibility | Complies with ISO 10993-1 | Complies with ISO 10993-1 | |
{8}------------------------------------------------
Image /page/7/Picture/0 description: The image contains the logo for BioMed ENT. The words "BioMed" are in orange, and the letters "ENT" are in gray. To the right of the text is a pyramid shape made of six hexagons in shades of orange and blue.
{9}------------------------------------------------
Image /page/8/Picture/0 description: The image contains the logo for BioMed ENT. The words "BioMed" are in orange, and the word "ENT" is in gray. To the right of the words is a design made up of seven hexagons in different shades of orange and blue.
## Biocompatibility testing
The Epi-Stop™ Nasal Gel/Epistaxis Pack consists of well-known and commonly used materials. All materials are purchased from reputable vendors, and Epi-Stop™ Nasal Gel/Epistaxis Pack has been evaluated under biocompatibility studies, and have successfully passed all of the following:
Cytotoxicity. A Cytotoxicity study was conducted to analyze if Epi-Stop™ Nasal Gel/Epistaxis Pack could cause cell damage or cell death. Thus, the cytotoxic effect of Epi-Stop™ Nasal Gel/Epistaxis Pack was investigated through a Cytotoxic-Direct Contact test in compliance with ISO 10993-5:2009 and the standards of Good Laboratory Practice. According to the results and acceptance criteria stated in ISO 10993-5:2009, the test item Epi-Stop™ Nasal Gel/Epistaxis Pack must be considered NOT CYTOTOXIC.
Sensitization. A sensitization study was conducted to investigate if Epi-Stop™ Nasal Gel/Epistaxis Pack could induce an abnormally sensitive reaction on skin. Thus, Epi-Stop™ Nasal Gel/Epistaxis Pack was submitted to a biological evaluation in order to identify its potential sensitizing effects; this was done through a Delayed Hypersensitivity test (GPMT test) in compliance with ISO 10993-10:2010 and the standards of Good Laboratory Practice. To summarize, on the basis of the results, interpreted according to ISO 10993-10:2010, the test item Epi-Stop™ Nasal Gel/Epistaxis Pack must be considered NOT SENSITIZING.
In Vivo Acute Skin Irritation. An in vivo acute skin irritation test was conducted to investigate if Epi-Stop™ Nasal Gel/Epistaxis Pack could cause an irritant effect on skin. Thus, a biological evaluation was performed on Epi-Stop™ Nasal Gel/Epistaxis Pack via the following test: Skin Irritation in compliance with ISO 10993-23:2021 and the standards of Good Laboratory Practice. To summarize, on the basis of the results, interpreted according to ISO 10993-23:2021, the test item Epi-Stop™ Nasal Gel/Epistaxis Pack must be considered Negligibly irritant for the skin.
Acute Systemic Toxicity. An acute systemic toxicity test was conducted to investigate if Epi-Stop™ Nasal Gel/Epistaxis Pack could induce any adverse or toxic effect following it's placement. Thus, a biological evaluation was conducted on Epi-Stop™ Nasal Gel/Epistaxis Pack via a Systemic Toxicity test in compliance with ISO 10993-11:2017 and the standards of Good Laboratory Practice. In summary, and based on the results, which were interpreted in accordance with ISO 10993-11:2017, the test sample Epi-Stop™ Nasal Gel/Epistaxis Pack does not cause toxic symptoms and satisfies with the requirements of the test.
{10}------------------------------------------------
Image /page/9/Picture/0 description: The image shows the logo for BioMed ENT. The words "BioMed" are in orange, and the word "ENT" is in gray. To the right of the words is a design made up of six hexagons in different shades of orange and blue.
| Test | Applicable standards | Status |
|----------------------------------|--------------------------------------------------|-----------------|
| Cytotoxic-Direct Contact<br>test | ISO 10993-5:2009<br>Good Laboratory<br>Practice | Not cytotoxic |
| Sensitization | ISO 10993-10:2010<br>Good Laboratory<br>Practice | Not sensitizing |
| In Vivo Acute Skin<br>Irritation | ISO 10993-23:2021<br>Good Laboratory<br>Practice | Not irritant |
| Acute Systemic Toxicity | ISO 10993-11:2017<br>Good Laboratory<br>Practice | Not toxic |
All Biocompatibility studies mentioned above were conducted in compliance with the following standards of Good Laboratory Practice:
- OECD Series on Principles of Good Laboratory Practice and ● Compliance Monitoring - OECD principles of Good Laboratory Practice (as revised in 1997) - Environment Directorate - Organization for Economic Co- Operation and Development, Paris 1998.
- Legislative decree n. 50 of March the 2ª, 2007. Enforcement of ● Community Directives 2004/9/CE and 2004/10/CE, concerning the inspection and verification of Good Laboratory Practice and the drawing of the legislative, regulatory and administrative dispositions relative to the application of Good Laboratory Practice rules, to the control of their application on the assays performed on the chemical substances (GU n.86 of April the 13th, 2007).
United States Food and Drug Administration, Title 21 Code of . Federal Regulations Part 58, Federal Register 22 December 1978, and subsequent amendments.
GLP Certification n. 2019/25 released by the Italian Ministry of Health . on September 12th 2019 authorizing Eurofins Biolab S.r.l. to perform analyses in compliance with the principles of good laboratory practices.
## Performance testing
Performance of Epi-Stop™ Nasal Gel/Epistaxis Pack was analyzed and evaluated together with the predicate device Chitogel® Endoscopic Sinus Surgery Kit. In summary, absorption, occlusion, dynamic viscosity, extrusion force, and pH were the five tests conducted.
## Occlusion.
Occlusion of the injury site is an important property responsible for the sealing between the wound's edges and the gel surface. A good occlusion will allow the hemostat to stop any leakage or outflow of blood outside the perimeter of the wound. The occlusion property of Epi-Stop™ Nasal Gel/Epistaxis Pack's was
{11}------------------------------------------------
Image /page/10/Picture/0 description: The image contains the logo for BioMed ENT. The words "BioMed" are in orange, and the word "ENT" is in gray. To the right of the words is a design made up of seven hexagons in a honeycomb pattern. The hexagons are in shades of orange and blue.
tested in comparison with Chitogel® Endoscopic Sinus Surgery Kit. Briefly, in summary to the occlusion results, there was no leakage of water seen in any of the devices tested for a period of 60 minutes. Both, the predicate and subject device succeeded in sealing the open surface of the glass flask, preventing the water from leaking outside the flask containers.
### Absorption.
The absorption property of an hemostat is relevant in allowing the retention of blood while the body carries on, independently, with its coagulation process. To test the retention that the Gelatin-HA meshwork can provide to Epi-Stop™ Nasal Gel/Epistaxis Pack an absorption test was conducted in comparison with Chitogel® Endoscopic Sinus Surgery Kit. In summary to the absorption results, it was concluded that both, the predicate device and subject device performed highly similar regarding the absorption property. After three hours of being submerged in water, both devices showed a comparable absorbance of water, based on their initial weight.
### Dynamic Viscosity.
The internal resistance of Epi-Stop™ Nasal Gel/Epistaxis Pack is an important characteristic as it reflects upon the force or pressure required to spread the gel. Once adhered to the edges of the wound, Epi-Stop™ Nasal Gel/Epistaxis Pack will adapt to the surrounding areas to render the device more comfortable upon placement. For this, the internal resistance of Epi-Stop™ Nasal Gel/Epistaxis Pack was measured and compared to Chitogel® Endoscopic Sinus Surgery Kit. In summary, both Epi-Stop™ Nasal Gel/Epistaxis Pack and Chitogel® Endoscopic Sinus Surgery Kit showed viscosities within the acceptable range for a hemostat. However, although the viscosity of Epi-Stop™ Nasal Gel/Epistaxis Pack is lower than that of Chitogel® Endoscopic Sinus Surgery Kit, it will not compromise its intended use, but rather only improves the handling characteristics with respect to injectability and adaptability to the surrounding implant area. The different viscosity levels seen in the subject device (when compared to the predicate) are primarily intended to support the handling properties of the device. This in mind, to successfully stop anterior minimal bleeding, Epi-Stop™ Nasal Gel/Epistaxis Pack relies primarily on an optimal occlusion, followed by good absorption to retain blood, and lastly on an optimal viscosity designed to facilitate the subject device's ability to properly adapt to the implant area once it has been injected in the anterior nasal cavity.
## Extrusion Force.
The applied force needed in order to eject the gel from the syringe was also analyzed in both devices. It was expected that both gels should be extruded from its own syringe by using similar extrusion force. Both devices scored within the acceptance range.
{12}------------------------------------------------
Image /page/11/Picture/0 description: The image contains the logo for BioMed ENT. The words "BioMed" are in orange, and the word "ENT" is in gray. To the right of the words is a design made up of seven hexagons in a triangular shape. The hexagons are in shades of orange and blue.
To help establish the stability of Epi-Stop™ Nasal Gel/Epistaxis Pack over the entirety of its shelf life, the pH has been measured so far at 0, 3.5, and 18 months real-time. In summary, the pH of Epi-Stop™ Nasal Gel/Epistaxis Pack remained stable for the entirety of the proposed shelf-life, 18 months.
### In-vivo study.
To verify that Epi-Stop™ Nasal Gel/Epistaxis Pack stays at the initial site of placement (bleeding site) for a period of three (3) days, and to test the hemostatic capabilities of the device, an in vivo animal study was conducted. Chitogel® Endoscopic Sinus Surgery Kit was included in this study as it is the predicate device that Epi-Stop™ is claiming substantial equivalence to.
In conclusion, both Epi-Stop™ Nasal Gel/Epistaxis Pack and Chitogel® Endoscopic Sinus Surgery Kit were able to stop anterior mild bleeding within the timeframe indicated. Although the predicate was slightly faster, both devices performed within the same acceptance criteria range. Together with this, according to the observations from the study, both gels also remained at the injection site and no dripping was observed during 24, 48, and 72 hours of monitoring.
# Conclusion
In conclusion, Chitogel® Endoscopic Sinus Surgery Kit and Epi-Stop™ Nasal Gel/Epistaxis Pack are considered substantially equivalent since both devices are single use, and injectable gels that are provided sterile for the treatment of anterior epistaxis. These devices are also made of biopolymers recognized for their mucoadhesive properties (i.e. chitosan or gelatin). Both devices are indicated for use as a nasal packing to treat anterior epistaxis, and supplied sterile with a SAL of 10-6. Both, the predicate and subject device were tested through biocompatibility analyses and both, Chitogel® Endoscopic Sinus Surgery Kit and Epi-Stop™ Nasal Gel/Epistaxis Pack, passed Cytotoxicity, Sensitization, and Acute Systemic Toxicity. In addition to this, Epi-Stop™ Nasal Gel/Epistaxis Pack also passed in vivo skin irritation test.
The differences between the predicate and subject device are not significant enough to compromise the safety and effective performance of Epi-Stop™ Nasal Gel/Epistaxis Pack since it shares the same characteristics as Chitogel® Endoscopic Sinus Surgery Kit in order to effectively treat anterior epistaxis. Epi-Stop™ Nasal Gel/Epistaxis Pack also shares almost identical indications for use, highly similar technology (e.g., both are single use gels delivered in syringes), and mode of removal with regards to the predicate device discussed above.
To finalize, performance between devices was compared by testing properties such as absorption, occlusion, dynamic viscosity, extrusion force, and pH. Hence, Chitogel® Endoscopic Sinus Surgery Kit and Epi-Stop™ Nasal
{13}------------------------------------------------
Image /page/12/Picture/0 description: The image contains the logo for Biomed ENT. The words "BioMed" are in orange, and the word "ENT" is in gray. To the right of the words is a design made up of seven hexagons in different shades of orange and blue.
Gel/Epistaxis Pack were analyzed and both devices exhibited values within acceptable ranges. Moreover, the in vivo study performed in both, the subject and predicate also showed that both devices remain at the injection site and are capable of stopping anterior bleeding in less than a minute. Based on these data, it was concluded that Epi-Stop™ Nasal Gel/Epistaxis Pack is as safe and effective as Chitogel® Endoscopic Sinus Surgery Kit when it comes to the treatment of anterior epistaxis.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.