← Product Code [HRS](/submissions/OR/subpart-d%E2%80%94prosthetic-devices/HRS) · K230141

# LOQTEQ® VA Elbow Plates 2.7/3.5 System (K230141)

_Aap Implantate AG · HRS · Mar 17, 2023 · Orthopedic · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/HRS/K230141

## Device Facts

- **Applicant:** Aap Implantate AG
- **Product Code:** [HRS](/submissions/OR/subpart-d%E2%80%94prosthetic-devices/HRS.md)
- **Decision Date:** Mar 17, 2023
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 888.3030
- **Device Class:** Class 2
- **Review Panel:** Orthopedic
- **Attributes:** Therapeutic

## Indications for Use

The LOQTEQ® VA Elbow Plates 2.7/3.5 includes the LOQTEQ® VA Distal Medial Humerus Plate 2.7/3.5, LOQTEQ® VA Distal Dorsolateral Humerus Plate 2.7/3.5, LOQTEQ® VA Distal Lateral Humerus Plate 2.7/3.5 and LOOTEO® VA Olecranon Plate 2.7/3.5. LOQTEQ® VA Distal Medial Humerus Plate 2.7/3.5, LOQTEQ® VA Distal Dorsolateral Humerus Plate 2.7/3.5, LOQTEQ® VA Distal Lateral Humerus Plate 2.7/3.5 are indicated for: Intra-articular fractures of the distal humerus Supracondylar fractures of the distal humerus Osteotomies, and non-unions of the distal humerus LOQTEQ® VA Olecranon Plate 2.7/3.5 is indicated for: Fixation of fractures, osteotomies and non-unions of the olecranon, particularly in osteopenic bone

## Device Story

Trauma implants consisting of metallic bone plates and screws; implanted by surgeons to achieve internal fixation of bone fragments following fractures or joint reconstruction. Plates utilize fixateur interne principle; angular stable plate-screw connection achieved by screwing threaded screw heads into threaded plate holes. Conventional screws allow off-axis angulation up to 15° within plate holes. Shape adapted to anatomy of distal humerus and olecranon. Implants removed after fracture healing. Used in clinical/surgical settings.

## Clinical Evidence

No clinical data submitted. Substantial equivalence supported by non-clinical bench testing, including mechanical testing against predicate K120717 and biocompatibility validation. Sterilization validated per ANSI AAMI ISO 17665-1.

## Technological Characteristics

Metallic bone fixation plates and screws. Features variable angle locking mechanism (angular stable connection). Compatible with locking and non-locking screws. Sterilization via moist heat per ANSI AAMI ISO 17665-1.

## Regulatory Identification

Single/multiple component metallic bone fixation appliances and accessories are devices intended to be implanted consisting of one or more metallic components and their metallic fasteners. The devices contain a plate, a nail/plate combination, or a blade/plate combination that are made of alloys, such as cobalt-chromium-molybdenum, stainless steel, and titanium, that are intended to be held in position with fasteners, such as screws and nails, or bolts, nuts, and washers. These devices are used for fixation of fractures of the proximal or distal end of long bones, such as intracapsular, intertrochanteric, intercervical, supracondylar, or condylar fractures of the femur; for fusion of a joint; or for surgical procedures that involve cutting a bone. The devices may be implanted or attached through the skin so that a pulling force (traction) may be applied to the skeletal system.

## Predicate Devices

- aap LOQTEQ® Elbow System ([K132787](/device/K132787.md))
- aap LOQTEQ® Olecranon Plate ([K140607](/device/K140607.md))
- aap LOQTEQ® Distal Lateral Humerus Plate 2.7/3.5 ([K161696](/device/K161696.md))
- Synthes Variable Angle LCP Elbow System ([K120717](/device/K120717.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

March 17, 2023

Image /page/0/Picture/1 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health and Human Services logo on the left and the FDA logo on the right. The FDA logo features the letters 'FDA' in a blue square, followed by the words 'U.S. FOOD & DRUG ADMINISTRATION' in blue text.

aap Implantate AG % Melissa Burbage Senior Regulatory Specialist PaxMed International. LLC 12264 El Camino Real, Suite 400 San Diego, California 92130

Re: K230141

Trade/Device Name: LOQTEQ® VA Elbow Plates 2.7/3.5 System Regulation Number: 21 CFR 888.3030 Regulation Name: Single/Multiple Component Metallic Bone Fixation Appliances And Accessories Regulatory Class: Class II Product Code: HRS Dated: November 18, 2022 Received: January 18, 2023

Dear Melissa Burbage:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal

{1}------------------------------------------------

statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Image /page/1/Picture/5 description: The image shows the name "Shumaya Ali-S" in a large, clear font. The text is horizontally oriented and appears to be the main subject of the image. The background is plain and does not distract from the text.

Shumaya Ali, MPH Assistant Director DHT6C: Division of Restorative, Repair and Trauma Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known) K230141

Device Name

LOQTEO® VA Elbow Plates 2.7/3.5

Indications for Use (Describe)

The LOQTEQ® VA Elbow Plates 2.7/3.5 includes the LOQTEQ® VA Distal Medial Humerus Plate 2.7/3.5, LOQTEO® VA Distal Dorsolateral Humerus Plate 2.7/3.5, LOQTEQ® VA Distal Lateral Humerus Plate 2.7/3.5 and LOQTEQ® VA Olecranon Plate 2.7/3.5.

LOQTEQ® VA Distal Medial Humerus Plate 2.7/3.5, LOQTEQ® VA Distal Dorsolateral Humerus Plate 2.7/3.5, LOQTEQ® VA Distal Lateral Humerus Plate 2.7/3.5 are indicated for:

- Intra-articular fractures of the distal humerus
· Supracondylar fractures of the distal humerus

· Osteotomies, and non-unions of the distal humerus

LOQTEQ® VA Olecranon Plate 2.7/3.5 is indicated for:

· Fixation of fractures, osteotomies and non-unions of the olecranon, particularly in osteopenic bone

Type of Use (Select one or both, as applicable)

| > Prescription Use (Part 21 CFR 801 Subpart D)

_ Over-The-Counter Use (21 CFR 801 Subpart C)

### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

{3}------------------------------------------------

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

FORM FDA 3881 (6/20)

Page 1 of 1

PSC Publishing Services (301) 443-6740 EF

{4}------------------------------------------------

# 510(k) Summary

# K230141

# LOQTEQ® VA Elbow Plates 2.7/3.5

# aap Implantate AG

March 17, 2022

#### ADMINISTRATIVE INFORMATION

| Manufacturer Name         | aap Implantate AG                               |                                        |
|---------------------------|-------------------------------------------------|----------------------------------------|
|                           | Lorenzweg 5                                     |                                        |
|                           | 12099 Berlin Germany                            |                                        |
|                           | Telephone:                                      | +49 30 750 19 129                      |
|                           | Fax:                                            | +49 30 750 19 111                      |
| Official Contact          | Agnieszka Mierzejewska, Chief Operating Officer |                                        |
| Representative/Consultant | Melissa Burbage                                 |                                        |
|                           | Kevin A. Thomas, PhD; Floyd G. Larson, MS, MBA  |                                        |
|                           | PaxMed International, LLC                       |                                        |
|                           | 12264 El Camino Real, Suite 400                 |                                        |
|                           | San Diego, CA 92130                             |                                        |
|                           | Telephone:                                      | +1 858-792-1235                        |
|                           | Fax:                                            | +1 858-792-1236                        |
|                           | Email:                                          | mburbage@paxmed.com                    |
|                           |                                                 | kthomas@paxmed.com; flarson@paxmed.com |

#### DEVICE NAME AND CLASSIFICATION

| Trade/Device Name    | LOQTEQ® VA Elbow Plates 2.7/3.5                                                                                                                                         |
|----------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Common Name          | Plate, Fixation, Bone                                                                                                                                                   |
| Regulation Number    | 21 CFR 888.3030                                                                                                                                                         |
| Regulation Name      | Single/multiple component metallic bone fixation appliances and accessories                                                                                             |
| Regulatory Class     | Class II                                                                                                                                                                |
| Product Code         | HRS                                                                                                                                                                     |
| Classification Panel | Orthopedic                                                                                                                                                              |
| Reviewing Division   | Office of Orthopedic Devices (OHT6), Division of Restorative, Repair and Trauma Devices (DHT6C) Stereotaxic, Bone Growth Stimulators and Fracture Fixation Devices Team |

### PREDICATE DEVICE INFORMATION

Primary Predicate Device K132787, aap LOQTEQ® Elbow System, aap Implantate AG

Additional Predicate Devices K140607, aap LOQTEQ® Olecranon Plate, aap Implantate AG K161696, aap LOQTEQ® Distal Lateral Humerus Plate 2.7/3.5, aap Implantate AG K120717, Synthes Variable Angle LCP Elbow System, Synthes

{5}------------------------------------------------

# INDICATIONS FOR USE STATEMENT

The LOQTEQ® VA Elbow Plates 2.7/3.5 includes the LOQTEQ® VA Distal Medial Humerus Plate 2.7/3.5, LOQTEQ® VA Distal Dorsolateral Humerus Plate 2.7/3.5, LOQTEQ® VA Distal Lateral Humerus Plate 2.7/3.5 and LOOTEO® VA Olecranon Plate 2.7/3.5.

LOQTEQ® VA Distal Medial Humerus Plate 2.7/3.5, LOQTEQ® VA Distal Dorsolateral Humerus Plate 2.7/3.5, LOQTEQ® VA Distal Lateral Humerus Plate 2.7/3.5 are indicated for:

• Intra-articular fractures of the distal humerus

· Supracondylar fractures of the distal humerus

· Osteotomies, and non-unions of the distal humerus

LOQTEO® VA Olecranon Plate 2.7/3.5 is indicated for:

· Fixation of fractures, osteotomies and non-unions of the olecranon, particularly in osteopenic bone

### SUBJECT DEVICE DESCRIPTION

The components of the LOOTEO VA Elbow Plates 2.7/3.5 are trauma implants, bone plates and screws, which are implanted into the human body to achieve an internal fixation of bone fragments, typically after fractures or other bone injuries (e.g., joint reconstruction). LOQTEQ plates function according to the fixateur interne principle, i.e., they incorporate an angular stable plate-screw connection as a result of screwing threaded screw heads into threaded plate holes. Conventional bone screws can be used in all plate holes and allow for off-axis angulation up to 15° inside threaded plate holes. After fracture healing, the implants (plates and screws) are to be removed.

The shape of the bone plates is adapted to the anatomy of and the biomechanical stress on the distal humerus and the olecranon (ulna), respectively.

The LOQTEQ VA Elbow Plates 2.7/3.5 consists of the LOQTEQ VA Distal Dorsolateral Humerus Plate, LOQTEQ VA Distal Lateral Humerus Plate, LOQTEQ VA Distal Medial Humerus Plate and the LOQTEQ VA Olecranon Plate 2.7/3.5.

#### PERFORMANCE DATA

Non-clinical testing data submitted, referenced, or relied upon to demonstrate substantial equivalence to from predicate devices K132787, K140607, and K161696 included biocompatibility and validation of the recommended end-user moist heat sterilization cycle according to ANSI AAMI ISO 17665-1 Sterilization of health care products – Moist heat – Part 1: Requirements for the development, validation and routine control of a sterilization process for medical devices. Mechanical testing against the predicate device K120717 was completed. Clinical data were not submitted in this premarket notification.

### EQUIVALENCE TO MARKETED DEVICES

The subject device, predicate devices K132787, K140607, K161696, and K120717 have the same technological characteristics and use the same operating principles for bone fixation. The plate

{6}------------------------------------------------

design and dimensions are the same as predicate devices K132787, K140607, and K161696. Slight modifications have been made to the plates to accommodate the variable angle feature that is similar to the predicate device K120717.

The subject devices are compatible with aap locking and non-locking screws previously cleared with predicate devices K132787, K140607, and K161696.

### CONCLUSION

The subject devices, the primary predicate device, and the additional predicate device have the same intended use and similar technological characteristics. They encompass a similar range of physical dimensions, are manufactured from the same materials, and are to be sterilized using identical methods. The data included in this submission demonstrate substantial equivalence to the predicate devices listed above.

---

**Source:** [https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/HRS/K230141](https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/HRS/K230141)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
