Klockner Abutments are intended for use in dental implants to provide a support structure so that the edentulous or partially edentulous patients will regain their masticatory and aesthetic functions. All digitally designed abutments for use with Klockner TiBase Abutments are intended to be sent to a SOADCO S.L. validated milling center for manufacture.
Device Story
Klockner Abutments provide support structures for dental implant prosthetics; includes healing abutments, temporary abutments, straight abutments, overdenture attachment systems, and TiBase abutments. Healing abutments connect to implants during healing; temporary abutments retain temporary cemented prostheses; straight abutments support single/multiple-unit cement-retained restorations. K-LOCK overdenture system uses snap-fit retention inserts and housings to attach prostheses to implants, accommodating up to 40° divergence. TiBase abutments serve as two-piece bases for patient-specific CAD/CAM zirconia superstructures. Devices are used in dental clinics by clinicians; digitally designed components are sent to validated milling centers for manufacture. Output is a stable prosthetic support structure; benefits include restored masticatory and aesthetic function. Materials include titanium alloy (ASTM F136) and zirconia (ISO 13356), with TiCN or ZrN coatings on specific models.
Clinical Evidence
No clinical data. Substantial equivalence supported by non-clinical bench testing: sterilization validation (ISO 17665), biocompatibility (ISO 10993-5, -10, -12), mechanical dynamic loading (ISO 14801), retention testing, coating characterization, and MRI safety assessment.
Indicated for edentulous or partially edentulous patients requiring dental implant support for prosthetic restoration (crowns, bridges, overdentures).
Regulatory Classification
Identification
An endosseous dental implant abutment is a premanufactured prosthetic component directly connected to the endosseous dental implant and is intended for use as an aid in prosthetic rehabilitation.
Special Controls
*Classification.* Class II (special controls). The guidance document entitled “Class II Special Controls Guidance Document: Root-Form Endosseous Dental Implants and Endosseous Dental Implant Abutments” will serve as the special control. (See § 872.1(e) for the availability of this guidance document.)
{1}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
SOADCO S.L. % Melissa Burbage Senior Regulatory Specialist PaxMed International. LLC 12264 El Camino Real, Suite 400 San Diego, California 92130
Re: K230103
Trade/Device Name: Klockner Abutments Regulation Number: 21 CFR 872.3630 Regulation Name: Endosseous dental implant abutment Regulatory Class: Class II Product Code: NHA Dated: January 27, 2023 Received: January 27, 2023
Dear Melissa Burbage:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{2}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
# Andrew I. Steen -S
Andrew I. Steen Assistant Director DHT1B: Division of Dental and ENT Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
# Indications for Use
510(k) Number (if known)
K230103
Device Name Klockner Abutments
Indications for Use (Describe)
Klockner Abutments are intended for use in dental implants to provide a support structure so that the edentulous or partially edentulous patients will regain their masticatory and aesthetic functions.
All digitally designed abutments for use with Klockner TiBase Abutments are intended to be sent to a SOADCO S.L. validated milling center for manufacture.
Type of Use (Select one or both, as applicable)
| <span style="font-family: Arial, sans-serif;"> <input checked="checked" type="checkbox"/> Prescription Use (Part 21 CFR 801 Subpart D) </span> |
|----------------------------------------------------------------------------------------------------------------------------------------------------|
| <span style="font-family: Arial, sans-serif;"> <input type="checkbox"/> Over-The-Counter Use (21 CFR 801 Subpart C) </span> |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
#### 510(k) Summary
# K230103
# Klockner Abutments
# SOADCO, S.L.
# April 24, 2023
#### ADMINISTRATIVE INFORMATION
| Manufacturer Name | SOADCO S.L.<br>Av. del Pessebre, 76-82<br>AD700 Escaldes - Engordany, Andorra<br>Telephone: +37 (6) 800-590 |
|---------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Official Contact | Ms. Mercedes Roldán, General Manager |
| Representative/Consultant | Melissa Burbage<br>Kevin A. Thomas, PhD; Floyd G. Larson, MS, MBA<br>PaxMed International, LLC<br>12264 El Camino Real, Suite 400<br>San Diego, CA 92130<br>Telephone: +1 858-792-1235<br>Fax: +1 858-792-1236<br>Email: mburbage@paxmed.com<br>kthomas@paxmed.com; flarson@paxmed.com |
#### DEVICE NAME AND CLASSIFICATION
| Trade/Device Name | Klockner Abutments |
|----------------------|------------------------------------|
| Common Name | Dental implant abutment |
| Classification Name | Endosseous dental implant abutment |
| Regulation Number | 21 CFR 872.3630 |
| Regulation Name | Endosseous dental implant abutment |
| Regulatory Class | Class II |
| Product Code | NHA |
| Classification Panel | Dental Products Panel |
| Reviewing Branch | Dental Devices Branch |
#### PREDICATE DEVICE INFORMATION
#### Primary Predicate Device
K151194, Klockner Dental Implant Abutments (II), SOADCO, S.L.
#### Reference Devices
K170022, Klockner Vega TiBase for CEREC®, SOADCO, S.L. K222288, DESS Dental Smart Solutions, Terrats Medical SL K191123, Multi-unit Abutments, Medentika GmbH K220612, PrimeLOC Attachment System, Innovative Product Brands, Inc.
{5}------------------------------------------------
Klockner Abutments are intended for use in dental implants to provide a support structure so that the edentulous or partially edentulous patients will regain their masticatory and aesthetic functions.
All digitally designed abutments for use with Klockner TiBase Abutments are intended to be sent to a SOADCO S.L. validated milling center for manufacture.
# SUBJECT DEVICE DESCRIPTION
The purpose of this submission is to obtain marketing clearance for abutments that are compatible with components from the Klockner Dental Implant Systems cleared in K153098, K082200, K080224. and K010132. The abutments consist of healing abutments, temporary abutments, straight abutments, and an overdenture attachment system that consists of abutments, denture housings, and retention inserts, and TiBase abutments with screws.
The subject device Vega Mimetic® Healing Abutments are designed to be connected to the implant during the healing period between implant placement and final abutment placement. Restorations are not placed on the healing components. The healing components are compatible with the Klockner Vega Implant System. They are offered in coronal diameter of 3.6 mm and with gingival heights of 2.0, 3.0, and 5.0 mm. The healing abutments attach to the dental implant with an integrated apical threaded portion. They are manufactured from titanium alloy and are color anodized to aid in identification.
The subject device Vega Mimetic® Temporary Abutments are straight, prepable abutments designed to retain a temporary cemented prosthesis. They are provided straight only and not intended to provide any angulation correction. They are compatible with the Klockner Vega Implant System. The abutment has a hex connection and attaches to the implant with a separate screw, previously cleared under K153098. The abutments are offered in a prosthetic platform diameter of 3.6 mm and a 7.0 mm post height, with gingival heights of 1.0, 2.0, and 3.0 mm. They are manufactured from titanium alloy and are anodized on a portion of the surface for identification purposes.
The subject device Vega Straight Permanent Abutments are designed for single-unit and multipleunit cement-retained restorations. The straight abutments are compatible with the Klockner Vega Implant System. They are available in an engaging and a non-engaging connection. The engaging design has a hex connection and attaches to the implant with a separate screw. The non-engaging design has a threaded apical portion. They are offered in a prosthetic platform diameter of 4.2 mm and gingival heights of 1.0, 2.0, and 3.0 mm, with no angulation. They are manufactured from titanium alloy and anodized on a portion of the surface for identification purposes.
The subject device K-LOCK Overdenture Attachment System is compatible with the Klockner Vega and Essential Cone Implant System and consists of abutments, retention inserts, and denture housings for attachment of prostheses to endosseous dental implants. All K-LOCK abutments are made of titanium alloy and have the same coronal ridge retention design that attaches to the overdenture component by an interference (snap) fit. The threaded apical end of the abutment connects to the implant and is specific to each compatible implant system and diameter. The
{6}------------------------------------------------
510(k) Summary Page 3 of 8
retention connection is designed to accommodate a path of insertion on implants that are divergent up to 40° (with 20°on each side). K-LOCK abutments are provided with a titanium carbonitride (TiCN) coating and are available in cuff (gingival) heights up to 6 mm, with a minimum implant interface diameter of 3.0 mm. A thin titanium carbonitride (TiCN) coating is applied to the abutments in the area that contacts gingival tissue and in the area that contacts the retention inserts.
Subject device TiBase Abutments are two-piece abutments composed of a bottom titanium component and are designed for fabrication of a patient-specific CAD/CAM zirconia superstructure on which a crown may be placed. The patient-specific CAD/CAM zirconia superstructure is the top-half of the two-piece abutment. They also may be used for support of a crown directly on the abutment. There are two types of TiBase Abutments; TiBases that attach directly to the implant, and TiBases that attach to previously-cleared abutments from the sponsor (similar to a prosthetic coping).
All patient-specific custom abutment fabrication for TiBase Abutments is by prescription on the order of the clinician. Zirconia superstructures for use with the subject device TiBase Abutments will be made at a SOADCO validated milling center under FDA quality system regulations and the material will conform to ISO 13356 Implants for surgery – Ceramic materials based on yttria-stabilized tetragonal zirconia (Y-TZP). The cement recommended in labeling for bonding of superstructures is Panavia F 5.0 cement by Kuraray Noritake Dental, cleared under K183537.
TiBase Abutments that attached directly to the implant are available in an engaging design and a non-engaging design. TiBase Abutments are made of titanium alloy. The TiBase Abutments are provided in four different designs: MedproTiBase, Mimetic TiBase, EC Essential TiBase, and SK2 Medpro TiBase. The EC Essential TiBase is a line extension from previously cleared TiBases for Cerec (K170022). The MedproTiBase abutments are anodized gold. The Mimetic TiBase abutments are coated with a thin layer of zirconium nitride (ZrN).
The design parameters for the CAD/CAM zirconia superstructure to be used on MedproTiBase Abutments are as follows:
Minimum wall thickness – 0.5 mm Minimum post height - 4.0 mm1 Minimum gingival height - 0.5 mm Maximum gingival height – 6.0 mm All zirconia superstructures are for straight abutments only 1 For devices 84 01 36 must be a minimum post height of 5.5 mm
The design parameters for the CAD/CAM zirconia superstructure to be used on Mimetic TiBase Abutments are as follows:
Minimum wall thickness - 0.35 mm Minimum post height - 4.2 mm Minimum gingival height - 0.5 mm Maximum gingival height - 6.0 mm All zirconia superstructures are for straight abutments only
{7}------------------------------------------------
510(k) Summary Page 4 of 8
The design parameters for the Sirona Dental CAD/CAM system to be used on EC Essential TiBase Abutments for Cerec® are as follows:
Minimum wall thickness - 0.8 mm Minimum post height - 5.2 mm Minimum gingival height - 0.5 mm Maximum gingival height – 6.0 mm Maximum angulation of the final abutment - 20°
TiBase Abutments that attached directly to previously-cleared angled abutments (like a coping) from SOADCO S.L. are available in an engaging design and a non-engaging design. These TiBase Abutments have no angulation, but some may be placed on abutments that are angled up to 30°. TiBase Abutments are made of titanium alloy and anodized gold. The TiBase Abutments are provided in three different designs: Permanent Vega Medpro, Essential Cone 25° Medpro, and Essential Cone Multi-Cone Medpro. The differences between these designs are primarily related to the compatible implant system or related to connection to a straight or angled abutment.
The design parameters for the CAD/CAM zirconia superstructure to be used on Medpro TiBase to Abutments are as follows:
Minimum wall thickness - 0.5 mm Minimum post height - 4.0 mm1 Minimum gingival height - 0.5 mm Maximum gingival height - 6.0 mm Not intended to provide angulation correction 1 For devices 84 01 26 and 84 01 27 must be a minimum post height of 5.5 mm
### PERFORMANCE DATA
Non-clinical testing data submitted to demonstrate substantial equivalence included:
- Sterilization validation according to ISO 17665-1 Sterilization of health care products Moist ● heat – Part 1: Requirements for the development, validation, and routine control of a sterilization process for medical devices and ISO 17665-2 Sterilization of health care products — Moist heat — Part 2: Guidance on the application of ANSI/AAMI/ISO 17665-1 was leveraged from previously cleared submissions K151194 and K170022.
- Biocompatibility testing for Klockner TiBase, straight abutments, healing abutments, temporary abutments, and denture housing subject devices leveraged from K170022 and K151194.
- Biocompatibility testing for TiCN coated K-Lock Abutments and ZrN coated Mimetic . Abutments according to ISO 10993-5 Biological evaluation of medical devices - Part 5: Tests for in vitro cytotoxicity, ISO 10993-10 Biological evaluation of medical devices – Part 10: Tests for skin sensitization for irritation and sensitization, and ISO 10993-12 Biological evaluation of medical devices – Part 12: Sample preparation and reference materials.
- Mechanical testing according to ISO 14801 Dentistry Implants Dynamic loading test ● for endosseous dental implants.
- . Retention testing was conducted to ensure that retention is maintained throughout the expected use of the product.
- The coating characterization using the methods outlined in FDA Guidance Class II Special ● Controls Guidance Document: Root-form Endosseous Dental Implants and Endosseous
{8}------------------------------------------------
Dental Implant Abutments, issued May 12, 2002.
- Non-clinical worst-case MRI review to evaluate the metallic devices in the MRI . environment using scientific rationale and published literature (e.g., Woods, Terry O., Jana G. Delfino, and Sunder Rajan. "Assessment of Magnetically Induced Displacement Force and Torque on Metal Alloys Used in Medical Devices." Journal of Testing and Evaluation 49.2 (2019): 783-795), based on the subject device components and material composition. Rationale addressed parameters per the FDA guidance "Testing and Labeling Medical Devices for Safety in the Magnetic Resonance (MR) Environment," including magnetically induced displacement force and torque.
# EQUIVALENCE TO MARKETED DEVICES
The Indications for Use Statement (IFUS) for the subject device is substantially equivalent to that of the primary predicate K151194 and reference device K170022. There is difference in terminology used, however the intended use is the same. All abutments are intended to be used with dental implants for support for prosthetic restoration for edentulous or partially edentulous patients. The digitally designed abutments are to be sent to a validated milling center for manufacture.
The subject device consists of various types of abutments; healing abutments, temporary abutments, straight abutments, overdenture attachment system, and TiBase abutments.
The subject device healing abutments, temporary abutments, and straight abutments are identical in design, materials, and technological characteristics to similar abutments cleared in the primary predicate K151194.
The subject device overdenture attachment system consists of abutments, retention inserts and denture housings for resilient attachment of prostheses to endosseous implants. The subject device abutment is substantially equivalent to the abutment of the reference device K220612 in its design, materials, and technological characteristics. It is similar to the reference device K220612 in coronal geometry in that the mechanism for overdenture retention is the same. The ability of the attachment system to accommodate abutment divergence is the same as that of the reference device K220612. The diameters and cuff heights offered are similar to those of the reference device K220612. The subject device is manufactured from titanium alloy and has the external surfaces above the implant/abutment interface modified by application of titanium carbonitride (TiCN) coating, similar to the titanium nitride (Ti\) coating of the reference device K220612. As with the reference device K220612, the coating is non-porous, the surface roughness of the machined abutment surface is maintained, and the coating is not applied to enhance tissue attachment to the device.
The subject device TiBase (to implant) abutment is substantially equivalent to the predicate device K170022 and the reference device K222288 in terms of design, materials, and technological characteristics. The subject device Mimetic TiBase abutments have external surfaces above the implant/abutment interface modified by application of zirconium nitride (Zr\) coating, similar to the zirconium nitride (ZrN) coating of the reference device K222288. The superstructure design parameters, including angulation, are similar to those of reference device K22288.
{9}------------------------------------------------
510(k) Summary Page 6 of 8
The subject device TiBase to abutment (coping) is substantially equivalent to the reference device K191123 in terms of design, materials, and technological characteristics. The TiBase to abutment does not allow for any additional angulation but can be used on abutments with angulation up to 30°.
The subject device abutments are made of titanium alloy conforming to ASTM F136 Standard Specification for Wrought Titanium-4 Vanadium ELI (Extra Low Interstitial) Allov for Surgical Implant Applications (UNS R56401). The titanium alloy subject device components are manufactured from identical materials, in the identical facilities using the identical manufacturing processes as used for SOADCO S.L. products cleared previously in K151194 and K170022.
All zirconia superstructures for use with the subject device TiBase Abutments will be made at a SOADCO S.L. validated milling center under FDA quality system regulations. The zirconia is inCoris ZI, cleared under K123664, and conforming to ISO 13356. Implants for surgerv -Ceramic materials based on vttria-stabilized tetragonal zirconia (Y-TZP). The cement recommended in labeling for bonding of superstructures is Panavia F 5.0 cement by Kuraray Noritake Dental, cleared under K183537. The zirconia and cement are the same material that was used in previously cleared products K170022.
Provided at the end of this summary is a table comparing the Indications for Use Statements (IFUS) and the technological characteristics of the subject device, the primary predicate device, and the reference devices.
#### CONCLUSION
The subject device, the primary predicate device, and reference devices have the same intended use and similar technological characteristics. They encompass a similar range of physical dimensions, are manufactured from the same materials, and are to be sterilized using identical methods. The data included in this submission demonstrate substantial equivalence to the predicate devices listed above.
{10}------------------------------------------------
| | Subject Device | Primary Predicate Device | Reference Device | Reference Device | Reference Device | Reference Device |
|------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | Klockner Abutments<br>SOADCO S.L. | K151194<br>Klockner Dental Implant<br>Abutment (II)<br>SOADCO S.L. | K170022<br>Klockner Vega TiBase for<br>CEREC®<br>SOADCO S.L. | K222288<br>DESS Dental Smart<br>Solutions<br>Terrats Medical SL | K191123<br>Multi-unit Abutments<br>Medentika GmbH | K220612<br>PrimeLOC Attachment<br>System<br>Innovative Product<br>Brands, Inc. |
| | Klockner Abutments are<br>intended for use in dental<br>implants to provide a support<br>structure so that the<br>edentulous or partially<br>edentulous patients will<br>regain their masticatory and<br>aesthetic functions.<br>All digitally designed<br>abutments for use with<br>Klockner TiBase Abutments<br>are intended to be sent to a<br>SOADCO S.L. validated<br>milling center for<br>manufacture. | The Klockner Dental Implant<br>Abutments are intended to be<br>placed into dental implants to<br>provide support for<br>prosthetic reconstructions<br>such as crowns, bridges.<br>The Klockner Dental Implant<br>Abutments include protective<br>cap, temporary abutment,<br>angled abutments, straight<br>abutments, cast abutments.<br>All abutments are intended to<br>be used with the Klockner<br>Dental Implant Systems,<br>models: Essential EC,<br>Essential EC 1.5. | The Klockner Vega TiBase<br>for CEREC® abutments are<br>titanium alloy abutments<br>placed onto Klockner Vega<br>Implants to provide support<br>for customized prosthetic<br>restorations. The Klockner<br>Vega TiBase for CEREC®<br>abutments are indicated for<br>screw-retained single tooth<br>or cement-retained single<br>tooth and bridge restorations.<br>All digitally designed<br>copings and/or crowns for<br>use with the Klockner Vega<br>TiBase for CEREC®<br>abutments are to be designed<br>and milled using Sirona<br>CEREC Premium SW 4.2<br>software and manufactured<br>using a Sirona CEREC or<br>inLab MC X or MC XL<br>milling unit. | DESS Dental Smart<br>Solutions abutments are<br>intended to be used in<br>conjunction with endosseous<br>dental implants in the<br>maxillary or mandibular arch<br>to provide support for<br>prosthetic restorations.<br>All digitally designed custom<br>abutments for use with DESS<br>Bases or Blanks are to be<br>sent to a Terrats Medical<br>validated milling center for<br>manufacture.<br>For complete Indications for<br>Use statement on OEM<br>Compatibility see 510(k)<br>Summary for K222288 in<br>Section 12 | Multi-unit abutments are<br>indicated for use with dental<br>implants as a support for<br>multi-unit screw retained<br>bridges and bars in the<br>maxilla or mandible of a<br>partially or fully edentulous<br>patient.<br>Multi-unit Abutments are<br>used for the restoration of the<br>following dental implant<br>systems:<br>For complete Indications for<br>Use statement on OEM<br>Compatibility see 510(k)<br>Summary for K191123 in<br>Section 12. | The PrimeLOC Attachment<br>System is designed to<br>facilitate patient removal of a<br>dental prosthesis for use with<br>full arch overdentures or<br>partial dentures retained in<br>whole or in part by<br>endosseous implants in the<br>mandible or maxilla. |
| Reason for<br>Predicate Device | Not applicable | Healing abutment, temporary<br>abutment, straight abutment | Ti Base, validated milling<br>center | TiBase to implant, 30<br>degrees | TiBase to abutment (coping) | Attachment system design |
| Product Codes | NHA | NHA | NHA | NHA | NHA | NHA |
| Designs/Features | | | | | | |
| Abutment Design | Healing, Temporary,<br>Straight, Attachment System,<br>TiBase, TiBase to abutment<br>(coping) | Conventional one-piece<br>abutments | Titanium Base Abutments | Healing, Temporary,<br>Straight, TiBase to implant | TiBase to abutment (coping) | Overdenture Attachment |
| Prosthesis<br>Attachment | Cement-retained<br>Screw-retained | Cement-retained<br>Screw-retained | Cement-retained<br>Screw-retained | Cement-retained<br>Screw-retained | Cement-retained<br>Screw-retained | Internal Thread |
| Restoration | Single-unit, Multi-unit | Single-unit, Multi-unit | Single-unit, Multi-unit | Single-unit, Multi-unit | Multi-unit | Multi-unit |
| | Subject Device | Primary Predicate Device | Reference Device | Reference Device | Reference Device | Reference Device |
| | Klockner Abutments<br>SOADCO S.L. | K151194<br>Klockner Dental Implant<br>Abutment (II)<br>SOADCO S.L. | K170022<br>Klockner Vega TiBase for<br>CERECⓇ<br>SOADCO S.L. | K222288<br>DESS Dental Smart<br>Solutions<br>Terrats Medical SL | K191123<br>Multi-unit Abutments<br>Medentika GmbH | K220612<br>PrimeLOC Attachment<br>System<br>Innovative Product<br>Brands, Inc. |
| Abutment-Implant<br>Platform Diameter<br>(mm) | 3.0 - 4.8 | 3.0 - 4.8 | 3.0 - 4.8 | 2.52 - 6.5 | 3.3 - 5.0 | 3.5 - 4.1 |
| Prosthetic Platform<br>Diameter (mm) | 3.5 - 6.6 | 3.5 - 6.6 | 4.5 - 7 | 4.5 - 6.5 | 3.3 - 6.5 | 3.5 |
| Angulation | Up to 30° | 0°, 17°, 30° | Up to 20° | Up to 30° | 0°, 17°, 30° | 0° |
| Abutment-Implant<br>Interface | Internal, External | Internal | Internal | Internal, External | Internal | Internal |
| Overdenture<br>Divergence<br>Allowance | 20°/40° | n/a | n/a | n/a | n/a | 20°/40° |
| Overdenture<br>Attachment Type | Retention Insert retained in<br>Denture Attachment Housing | n/a | n/a | n/a | n/a | Retention Insert retained in<br>Denture Attachment Housing |
| Materials | | | | | | |
| Abutment Materials | Titanium alloy, ASTM F136<br>Zirconia, ISO 13356<br>(copings) | Titanium alloy, ASTM F136 | Titanium alloy, ASTM F136<br>Compatible with the Sirona<br>Dental System inCoris ZI<br>Meso L (K100152 /<br>K111421) | Titanium alloy, ASTM F136<br>Cobalt-chromium alloy,<br>ASTM F1537<br>Zirconia, ISO 13356<br>(copings) | Ti 6Al-4V ELI | Ti 6Al-4V ELI |
| Surface Coating<br>(overdenture &<br>Mimetic TiBase) | Titanium Carbonitride,<br>Zirconium nitride…
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.