Enhanced AURORA™ Medical Diode System, and related accessories
K230076 · Psoria-Shield, Inc. · FTC · Mar 8, 2023 · General, Plastic Surgery
Device Facts
Record ID
K230076
Device Name
Enhanced AURORA™ Medical Diode System, and related accessories
Applicant
Psoria-Shield, Inc.
Product Code
FTC · General, Plastic Surgery
Decision Date
Mar 8, 2023
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4630
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Enhanced AURORA™ Medical Diode system and related accessories is indicated for use in targeted PUVA photochemistry and UVB phototherapy for the treatment of skin conditions including psoriasis, vitiligo, atopic dermatitis, eczema, and seborrheic dermatitis. In addition, the system UVB channel is indicated for the treatment of leukoderma.
Device Story
Enhanced AURORA™ Medical Diode system provides targeted UVA or UVB phototherapy for dermatologic conditions. System comprises base station, touchscreen interface, and two wavelength-specific handpieces (UVA or UVB). Clinician selects treatment type; activates energy via handpiece trigger. Device delivers light to affected skin area; includes USB-connected thermal printer for treatment parameter verification. Used in clinical settings by healthcare providers. Benefits include targeted light delivery to specific skin lesions, minimizing exposure to healthy tissue. Software version 1.2.4 manages printer communication and treatment area display.
Clinical Evidence
Bench testing only. Verification confirmed UVA and UVB energy dosage output is within 10% of user-selected power settings using a Coherent PowerMax PC thermopile sensor array.
Technological Characteristics
Diode-based medical device; UVA (350-395nm) and UVB (300-320nm) light emission. Powered by 110V wall outlet. Features touchscreen, dual handpieces, and USB printer connectivity. Safety features include key lock and emergency stop button. Dimensions: 19" x 14" x 8"; Weight: 20 lbs. Software version 1.2.4.
Indications for Use
Indicated for patients with psoriasis, vitiligo, atopic dermatitis, eczema, seborrheic dermatitis, or leukoderma requiring targeted PUVA photochemistry or UVB phototherapy.
Regulatory Classification
Identification
An ultraviolet lamp for dermatologic disorders is a device (including a fixture) intended to provide ultraviolet radiation of the body to photoactivate a drug in the treatment of a dermatologic disorder if the labeling of the drug intended for use with the device bears adequate directions for the device's use with that drug.
PhotoMedex, Inc. XTRAC Ultra2 Excimer Laser System, Model AL10000 (K073659)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logos of the Department of Health and Human Services and the Food and Drug Administration (FDA). The Department of Health and Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
March 8, 2023
Psoria-Shield, Inc. John Yorke General Manager 409 Mandeville St Utica, New York 13502
Re: K230076
Trade/Device Name: Enhanced AURORA™ system, and related accessories Regulation Number: 21 CFR 878.4630 Regulation Name: Ultraviolet Lamp For Dermatologic Disorders Regulatory Class: Class II Product Code: FTC, GEX Dated: January 9, 2023 Received: January 10, 2023
Dear John Yorke:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Jianting Wang -S
Jianting Wang Acting Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K230076
Device Name
Enhanced AURORA TM Medical Diode system, and related accessories
Indications for Use (Describe)
The Enhanced AURORA™ Medical Diode system and related accessories is indicated for use in targeted PUVA photochemistry and UVB phototherapy for the treatment of skin conditions including psoriasis, vitiligo, atopic dermatius, eczema, and seborrheic dermatitis. In addition, the system UVB channel is indicated for the treatment of leukoderma.
| Type of Use (Select one or both, as applicable) | |
|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| <span style="font-family: Arial, sans-serif;"> <span style="font-size: 11pt;"> <span style="color: black;"> <span style="font-weight: normal;"> <span style="font-style: normal;"> <span style="text-decoration: none;"> </span> </span> </span> </span> </span> </span> <span style="font-family: Wingdings;"> <span style="font-size: 11pt;"> <span style="color: black;"> <span style="font-weight: normal;"> <span style="font-style: normal;"> <span style="text-decoration: none;"> </span> </span> </span> </span> </span> </span> Prescription Use (Part 21 CFR 801 Subpart D) | <span style="font-family: Arial, sans-serif;"> <span style="font-size: 11pt;"> <span style="color: black;"> <span style="font-weight: normal;"> <span style="font-style: normal;"> <span style="text-decoration: none;"> </span> </span> </span> </span> </span> </span> <span style="font-family: Wingdings;"> <span style="font-size: 11pt;"> <span style="color: black;"> <span style="font-weight: normal;"> <span style="font-style: normal;"> <span style="text-decoration: none;"> </span> </span> </span> </span> </span> </span> Over-The-Counter Use (21 CFR 801 Subpart C) |
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
#### 510(k) SUMMARY K230076
Psoria-Shield Inc's Enhanced AURORA™Medical Diode System, and related accessories
The following information is provided as required by 21 C.F.R. § 807.87 (2003) for Psoria-Shield Inc's AURORA system Traditional 510(k) premarket notification:
Document Control Clerk,
Pursuant to 21 CFR 807.87, this is to notify you of the intention of Psoria-Shield, Inc. to introduce the Enhanced AURORA™ Medical Diode system and related accessories into commerce.
### GENERAL INFORMATION:
| Premarket Application | |
|-----------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------|
| Submitter: | Psoria-Shield, Inc.<br>409 Mandeville St<br>Utica, NY 13502, USA<br>Phone: 315 864 3465<br>Email: jy@psoriashield.com |
| Contact: | John Yorke<br>General Manager<br>Psoria-Shield, Inc.<br>409 Mandeville St<br>Utica, NY 13502, USA<br>Phone: 315 864 3465<br>Email: jy@psoriashield.com |
| Manufacturer: | Psoria-Shield, Inc.<br>409 Mandeville St<br>Utica, NY 13502, USA<br>Phone: 315 864 3465<br>Email: jy@psoriashield.com |
| Date Prepared: | January 9, 2023 |
| Classification Name: | Ultraviolet Lamp For Dermatologic Disorders<br>Laser surgical instrument for use in general and plastic surgery and in<br>dermatology |
| CFR Citation Number: | 21 CFR 878.4630<br>21 CFR 878.4810 |
| Product Code: | FTC, GEX |
{4}------------------------------------------------
| Classification: | Class II |
|------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Review Panel: | General & Plastic Surgery |
| Trade Name: | Enhanced AURORA™ Medical Diode system and related<br>accessories |
| Common Name: | The Enhanced AURORA™ Medical Diode system and related<br>accessories is a Diode medical device for dermatological disorders. |
| Generic Name: | AURORA™ system, and related accessories |
| Establishment<br>Registration<br>Number: | 3008872264 |
| Predicate Devices: | The Psoria-Shield, Inc. AURORA™ (K192411 July 2, 2020)<br>The PhotoMedex, Inc. XTRAC Ultra2 Excimer Laser System, Model<br>AL10000 (K073659 January 2, 2008). |
| Purpose of this<br>Submission: | Psoria-Shield is seeking to introduce the Enhanced AURORA™<br>Medical Diode system and related accessories into commerce<br>claiming substantial equivalence to the Psoria-Shield AURORA™<br>(K192411) and the PhotoMedex XTRAC Ultra2 Excimer Laser<br>System, Model AL10000 (K073659). |
| Indications for Use: | The Enhanced AURORA™ Medical Diode system and related<br>accessories is indicated for use in targeted PUVA photochemistry<br>and UVB phototherapy for the treatment of skin conditions including<br>psoriasis, vitiligo, atopic dermatitis, eczema, and seborrheic<br>dermatitis. In addition, the system UVB channel is indicated for the<br>treatment of leukoderma. |
| Device Description: | The Enhanced AURORA™ Medical Diode system and related<br>accessories is designed to allow a clinician to easily apply UVA or<br>UVB light to the patient's affected skin. The clinician selects the<br>desired treatment type (UVA or UVB light) and applies the light<br>through the appropriate handpiece by pointing the exit aperture<br>towards the affected skin area. |
| | The device is comprised of four main components: |
| | - A base station<br>- A touchscreen where the user can enter system commands<br>- Two handpieces connected to the base by power cables. One<br>- handpiece contains diode emitters that, when activated, emit UVA<br>light. The other handpiece emits UVB light. Either handpiece will only<br>operate if the appropriate wavelength is selected, so a user cannot |
{5}------------------------------------------------
| - | accidentally activate the incorrect handpiece. Activation of energy<br>occurs on depression of a trigger in the handpiece.<br>A printer accessory that can print out verification of treatment<br>parameters for hard copy records.<br>The technical specifications for the AURORA are as follows: | | |
|----------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------|
| | ●<br>●<br>●<br>●<br>●<br>●<br>● | Weight:<br>Size:<br>Min/Max power output (UVA):<br>Min/Max power output (UVB):<br>Power Increase Increments<br>Size of beam:<br>Dosage Accuracy: | 20 lbs<br>19" (W) x 14" (L) x 8" (H)<br>100 mJ/cm² / 3400 J/cm²<br>100 mJ/cm² / 3400 mJ/cm²<br>20mJ/100mJ<br>2.88 cm²<br>10% |
| Technical Characteristics: | The Enhanced AURORA™ Medical Diode system and related<br>accessories has the exact technological characteristics as the<br>previously cleared AURORA™ (K192411 July 2,2020). Additionally,<br>Psoria Shield is claiming substantial equivalence to the PhotoMedex<br>XTRACT (K073659 January 2,2008).<br><br>The Enhanced AURORA™ Medical Diode system and related<br>accessories emits either UVA or UVB light, depending on which has<br>been selected by the clinician. The power output is the same between<br>UVA or UVB. The treatment area is the same between UVA or UVB.<br>Treatment instructions are the same between both the UVA and UVB.<br>Thus, the modified AURORA is the same device as the cleared<br>AURORA™ (K192411 July 2,2020) except for the modifications listed<br>in table below, and for which Psoria-Shield is seeking clearance: | | |
| Modification | Detailed description of modification |
|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| A modification was<br>implemented to improve<br>the communication<br>between the Enhanced<br>AURORA and the<br>thermal printer. Minor<br>changes were made to<br>the code of the<br>Enhanced AURORA.<br>These modified<br>changes are<br>incorporated in the new<br>version of the code<br>1.2.4. | The update consisted of updating the USB Printer to improve the<br>communication of the USB thermal printer address. The update<br>was done to ensure that the firmware/software connection was<br>fixed to allow the Enhanced AURORA™ Medical Diode system to<br>properly function for the user. This was done to allow the clinician<br>to utilize the thermal printer function while additionally preventing<br>them from overriding the Patient ID function while utilizing the<br>device. A software change to the code was made in order for the<br>clinician to read results before printing and to include an added<br>treatment area to the results screen on the Enhanced AURORA ™<br>Medical Diode system. |
{6}------------------------------------------------
| Performance Data: | Psoria-Shield has performed bench testing consistent with design<br>verification activities required under 21 CFR 820 to confirm the<br>design modifications have no effect on the safety or effectiveness of<br>the modified device. This was confirmed through verifying that UVA<br>and UVB energy dosage output are within 10% of user-selected<br>power settings.<br><br>Testing was performed using a production-equivalent Enhanced<br>AURORA™ Medical Diode system and related accessories, with<br>energy dosage measured using a Coherent PowerMax PC thermopile<br>sensor array. Testing was performed to confirm the joule output of the<br>device in a cm² area in the center of both the thermopile and using<br>the center of the chipset arrays for the most accurate reading<br>possible. Screenshots of the AURORA device were taken at the<br>same time to demonstrate the dosage that the device was calibrated<br>to deliver in the pulse. |
|--------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Substantial Equivalence: | The Enhanced AURORA™ Medical Diode system and related<br>accessories has the exact same use as the previously cleared<br>AURORA™ (K192411 July 2, 2020). In addition, the Enhanced<br>AURORA™ Medical Diode system and related accessories has the<br>exact indications and principles of operation as the predicate devices,<br>the exact technological characteristics to the AURORA™ (K192411<br>July 2,2020) and very similar technical characteristics to the<br>PhotoMedex XTRACT (K073659 January 2, 2008).<br><br>Although there are minor differences between the Enhanced<br>AURORA™ Medical Diode system and related accessories and its<br>predicate devices. Note that the indications for use (IFU's) are not<br>being changed, modified, or altered in any fashion from the predicate<br>devices. Thus, the Enhanced AURORA™ Medical Diode system and<br>related accessories is substantially equivalent to the predicate<br>devices. |
| Conclusion: | In summary, the company's Enhanced AURORA™ Medical Diode<br>system and related accessories has the exact same use as the<br>previously cleared AURORA™ (K192411 July 2,2020). In addition,<br>the Enhanced AURORA™ Medical Diode system and related<br>accessories has the exact indications and principles of operation as<br>the predicate devices, and the exact technological characteristics to<br>the AURORA™ (K192411 July 2,2020) and very similar technical<br>characteristics to the PhotoMedex XTRACT (K073659 January 2,<br>2008). Although there are minor differences between the Enhanced<br>AURORA™ Medical Diode system and related accessories and its<br>predicate devices as summarized in table below. Note that the<br>indications for use (IFU's) are not being changed, modified, or altered<br>in any fashion from the predicate devices. Thus, the Enhanced |
{7}------------------------------------------------
AURORA™ Medical Diode system and related accessories is substantially equivalent to the predicate devices.
| | Psoria Shield<br>Enhanced<br>AURORA™ Medical<br>Diode system and<br>related accessories<br>(K230076) | Psoria Shield<br>AURORA™<br>(K192411) | PhotoMedex<br>XTRAC Ultra²<br>Excimer Laser<br>System, Model<br>AL10000<br>(K073659) |
|---------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Intended Use | The Enhanced<br>AURORA™ and<br>related accessories<br>is intended to be<br>used for the<br>treatment of<br>dermatologic<br>disorders. | The AURORA is<br>intended to be used<br>for the treatment of<br>dermatologic<br>disorders. | The intended use is<br>targeted UVB<br>phototherapy for<br>treatment of the<br>skin conditions<br>including<br>psoriasis, vitiligo,<br>atopic dermatitis,<br>and leukoderma. |
| Indications for Use | The Enhanced<br>AURORA™ Medical<br>Diode system and<br>related accessories<br>is indicated for use<br>in targeted PUVA<br>photochemistry and<br>UVB phototherapy<br>for the treatment of<br>skin conditions<br>including psoriasis,<br>vitiligo, atopic<br>dermatitis, eczema,<br>and seborrheic<br>dermatitis. In<br>addition, the system<br>UVB channel is<br>indicated for the<br>treatment of<br>leukoderma. | The AURORA<br>system is indicated<br>for use in targeted<br>PUVA<br>photochemistry and<br>UVB phototherapy<br>for the treatment of<br>skin conditions<br>including psoriasis,<br>vitiligo, atopic<br>dermatitis (eczema),<br>and seborrheic<br>dermatitis. In<br>addition, the system<br>UVB channel is<br>indicated for the<br>treatment of<br>leukoderma. | UVB phototherapy<br>for psoriasis,<br>vitiligo, atopic<br>dermatitis, and<br>leukoderma. |
| Wavelength | UVA:350-395nm<br>UVB:300-320nm | UVA: 350-395nm<br>UVB:320-320nm | 308nm |
## Substantial Equivalence Table Comparing Subject and Predicate Device
{8}------------------------------------------------
| Maximum Fluence<br>to Tissue: | Up to 3,400mJ/cm² | Up to 3,400mJ/cm² | 2-3.8 mJ/cm2 |
|-------------------------------|--------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------|
| Pulse Duration: | C/W | C/W | Quasi continuous<br>output |
| Beam Delivery: | Flexible Fiber Optic<br>and Therapeutic<br>Handpieces | Flexible Fiber Optic<br>and Therapeutic<br>Handpieces | Flexible Fiber Optic<br>and Therapeutic<br>Handpieces |
| Aiming Beam | YES | YES | YES |
| Power Source | 110V wall power | 110V wall power | 110V wall power |
| Dimensions | H: 19 | H:19 | H: 34 |
| | D: 14 | D:14 | D: 39 |
| | W: 8 | W: 8 | W:18 |
| Safety Features | Key lock to prevent<br>accidental energy<br>emission and<br>Emergency stop<br>button | Key lock to prevent<br>accidental energy<br>emission and<br>Emergency stop<br>button | Key lock to prevent<br>accidental energy<br>emission and<br>emergency stop<br>button |
| Weight | 20lbs | 20lbs | 220lbs |
| Accessories | USB-connected<br>printer | USB-connected<br>printer | NA |
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.