Retrospective patient MR perfusion and diffusion imaging data
Retrospective clinical patient data was used to conduct comparative performance testing of the Virtual Expert module, specifically for volumetric and spatial agreement analysis, and to support secondary clinical application assessments.
Patients undergoing MR perfusion and diffusion imaging for stroke assessment
Predicate device (CMN Capillary Function K202793)
Volumetric and spatial agreement per-patient and per-lesion
Indications for Use
Cercare Medical Neurosuite and associated modules, including the Capillary Function module, is an image processing software package to be used by trained professionals, including physicians and medical technicians. The software package runs on standard off-the-shelf computer or a virtual platform, such as VMware, and can be used to perform image viewing, processing, and analysis of images. Data and images are acquired through DICOM (Digital Imaging and Communications in Medicine) compliant imaging devices Cercare Medical Neurosuite provides viewing capabilities, whereas the Capillary Function module provides analysis capabilities for functional and dynamic imaging datasets acquired with MRI including a Diffusion Weighted MRI (DWI) Module and a Dynamic Analysis Module (dynamic contrast-enhanced imaging data for MRI). The Capillary Function module is used for visualization and analysis of dynamic imaging data, showing properties of changes in contrast over time. This functionality includes calculation of parameters related to tissue flow (perfusion) and tissue blood volume. In addition, the Capillary Function module's DWI technology is used to visualize local water diffusion properties from the analysis of diffusion-weighted MRI data. The Virtual Expert module allows the calculation of regions of interest and the visual inspection of time attenuation curves. One clinical application is to visualize the apparent blood perfusion and to calculate ADC threshold volume, Tmax threshold volume, and Mismatch Ratio in the brain tissue affected by acute stroke. Areas of decreased perfusion appear as areas of changed signal intensity: - · Lower signal intensity for CBF and CBV · Higher signal intensity for TTP, MTT, and Tmax
Device Story
Software-only device for neuroimaging analysis; runs on standard PC or virtual platform (VMware). Inputs: DICOM-compliant MRI datasets (DWI and dynamic contrast-enhanced). Processing: Capillary Function module calculates hemodynamic markers (CBF, CBV, MTT, Tmax, etc.) and diffusion properties; Virtual Expert module performs automatic ROI delineation, thresholding, and volumetric calculations (e.g., Mismatch Ratio). Output: Parameter maps and volumetric data for clinician review. Used in clinical settings by physicians/technicians as decision support for stroke assessment. Benefits: Provides quantitative hemodynamic and diffusion information to assist in evaluating brain tissue affected by stroke or tumors; aids clinical decision-making by visualizing perfusion deficits and calculating mismatch volumes.
Clinical Evidence
No clinical trials; bench testing only. Performance validated using simulated digital phantoms with known hemodynamic parameters and retrospective clinical MRI data. Metrics included absolute bias, correlation coefficients, and multi-scale structural similarity index. Virtual Expert module validated via volumetric and spatial agreement analysis compared to predicate and reference devices. All established acceptance criteria met.
Technological Characteristics
Software-only; DICOM-compliant. Runs on off-the-shelf PC/VMware. Supports secure VPN/SSH networking. Implements SVD algorithm for perfusion markers. Conforms to ISO 13485, ISO 14971, IEC 62304, IEC 82304-1, IEC 62366-1, ISO 15223-1, ISO 20417, and NEMA PS 3.1-3.20.
Indications for Use
Indicated for trained professionals, including physicians and medical technicians, to perform image viewing, processing, and analysis of DICOM-compliant MRI datasets for patients, including those with acute stroke, to visualize tissue perfusion, diffusion, and hemodynamic status.
Regulatory Classification
Identification
A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.
Special Controls
*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logos of the Department of Health & Human Services and the Food and Drug Administration (FDA). The Department of Health & Human Services logo is on the left, and the FDA logo is on the right. The FDA logo is a blue square with the letters "FDA" in white, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue.
Cercare Medical A/S % Jennifer Willner President JW Regulatory Consulting LLC 406 Wacouta Street, Suite 417 SAINT PAUL MN 55101
July 14, 2023
Re: K230016
Trade/Device Name: Cercare Medical Neurosuite (CMN) Capillary Function with Virtual Expert Regulation Number: 21 CFR 892.2050 Regulation Name: Medical image management and processing system Regulatory Class: Class II Product Code: LLZ Dated: June 6, 2023 Received: June 14, 2023
Dear Jennifer Willner:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements,
{1}------------------------------------------------
including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4. Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Jessica Lamb
Jessica S. Lamb, Ph.D. Assistant Director Imaging Software Team DHT8B: Division of Radiological Imaging Devices and Electronic Products OHT8: Office of Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K230016
#### Device Name
Cercare Medical Neurosuite (CMN) Capillary Function with Virtual Expert
#### Indications for Use (Describe)
Cercare Medical Neurosute and associated modules, including the Capillary Function module, is an image processing software package to be used by trained professionals, including physicians and medical technicians.
The software package runs on standard off-the-shelf computer or a virtual platform, such as VMware, and can be used to perform image viewing, processing, and analysis of images. Data and images are acquired through DICOM (Digital Imaging and Communications in Medicine) compliant imaging devices
Cercare Medical Neurosuite provides viewing capabilities, whereas the Capillary Function module provides analysis capabilities for functional and dynamic imaging datasets acquired with MRI including a Diffusion Weighted MRI (DWI) Module and a Dynamic Analysis Module (dynamic contrast-enhanced imaging data for MRI).
The Capillary Function module is used for visualization and analysis of dynamic imaging data, showing properties of changes in contrast over time. This functionality includes calculation of parameters related to tissue flow (perfusion) and tissue blood volume. In addition, the Capillary Function module's DWI technology is used to visualize local water diffusion properties from the analysis of diffusion-weighted MRI data.
The Virtual Expert module allows the calculation of regions of interest and the visual inspection of time attenuation curves. One clinical application is to visualize the apparent blood perfusion and to calculate ADC threshold volume, Tmax threshold volume, and Mismatch Ratio in the brain tissue affected by acute stroke.
Areas of decreased perfusion appear as areas of changed signal intensity:
- · Lower signal intensity for CBF and CBV
· Higher signal intensity for TTP, MTT, and Tmax
| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
| Prescription Use (Part 21 CFR 801 Subpart D) |
| Over-The-Counter Use (21 CFR 801 Subpart C) |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the text "CMN K230016 510(k) Submission". The text is in a simple, sans-serif font and is left-aligned. The text is likely a title or heading for a document or presentation related to a 510(k) submission, which is a type of premarket submission made to the FDA to demonstrate that a medical device is safe and effective.
Image /page/3/Picture/1 description: The image contains the logo for Cercare Medical. The logo consists of a cluster of blue circles with one red circle at the top left, followed by the text "Cercare Medical" in a light blue sans-serif font. The text is positioned to the right of the circle cluster.
## 510(K) SUMMARY
| 510(k) Number: | K230016 |
|----------------|---------------|
| Date Prepared: | July 12, 2023 |
### Table 1: Submitter Information
| Manufacturer: Cercare Medical A/S | Manufacturer's Contact Person: |
|-----------------------------------|--------------------------------|
| Inge Lehmanns Gade 10 | Mikkel Bo Hansen |
| DK-8000 Aarhus | Chief Scientific Officer |
| Denmark | Phone: +45 22890125 |
| US FDA ERN: 3019844085 | Email: mbh@cercare-medical.com |
#### Table 2: Device Information
| Trade Name | Cercare Medical Neurosuite (CMN) Capillary Function<br>with Virtual Expert for MRI |
|----------------------------|------------------------------------------------------------------------------------|
| Common Name | CMN Capillary Function with Virtual Expert |
| Classification Name | System, Image Processing, Radiological |
| Regulation | 21 CFR 892.2050 |
| Product Code | LLZ |
| Regulatory Classification: | Class II |
| Device Panel: | Radiology |
CMN Capillary Function with Virtual Expert MRI is substantially equivalent to the CMN Capillary Function Predicate Device (Table 3) and its modifications are similar to those provided in the iSchemaView RAPID Reference Device (Table 4). Neither of these have been subject to a design-related recall.
### Table 3: Predicate Device
| Predicate Device | Manufacturer | FDA 510(k) |
|------------------------|---------------------|------------|
| CMN Capillary Function | Cercare Medical A/S | K202793 |
#### Table 4: Reference Device
| Predicate Device | Manufacturer | FDA 510(k) |
|------------------|--------------|------------|
| RAPID | iSchemaView | K182130 |
### Device Description
Cercare Medical Neurosuite is a software-only device designed to streamline medical image processing by providing for the visualization and study of medical images. CMN can be installed on a customer PC or it can be accessed remotely using remote desktop technologies. CMN provides viewing, quantification, analysis and reporting capabilities. CMN is not intended as a dedicated PACS system for long term persistent storage of patient data.
CMN is software that is intended for use by trained professionals, including physicians and medical technicians. The software provides cerebral image processing capabilities. CMN is
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image contains the logo for Cercare Medical. The logo consists of a cluster of small turquoise circles with one larger red circle at the top left, followed by the text "Cercare Medical" in a turquoise sans-serif font. The text is positioned to the right of the circle cluster.
intended to be used as decision support software only and the clinician continues to provide all treatment decisions.
The software is intended to visualize and study neuroimaging by image viewing and registration of medical images. CMN works with MRI (Magnetic Resonance Image) technology.
CMN accepts and produces data sets in the DICOM format. DICOM is a standard format for storing and transmitting medical image data in vendor neutral format and is managed by the DICOM Standards Committee.
CMN is a platform that allows for the addition of certain modules for further analysis. One of these modules included in this submission is Capillary Function.
## CMN Capillary Function
Capillary Function, when activated in the installed Cercare Medical Neurosuite, provides further functionalities for reading, writing, visualizing and studying medical images.
Capillary Function provides perfusion post-processing technologies, where dynamically acquired perfusion MRI series can be processed to yield information relevant for assessment of the hemodynamic status of a patient.
Capillary Function generates hemodynamic markers, which can be used for management of diseases with possibly compromised hemodynamic function, such as ischemic stroke and tumors.
The generated output maps can be viewed by standard DICOM image viewers. In addition. Capillary Function includes the possibility for post-processing diffusion-weighted imaging (DWI) MRI data. Post-processing of DWI data results in maps reflective of local water diffusion properties. The post-processed DWI-derived maps can be viewed in standard DICOM image viewers. Capillary Function thus works with MRI technology.
## CMN Virtual Expert
Virtual Expert, when activated in the installed CMN Capillary Function, provides further functionalities for reading, writing, visualizing, and studying medical images.
Virtual Expert provides automatic delineation of regions of interest (ROI) relevant for stroke patient assessment based on perfusion and diffusion image output generated by the Capillary Function module. Specifically, diffusion MRI images are used to generate threshold masks of perceived core lesions, whereas perfusion MRI images are used to generate threshold masks of perceived perfusion restriction. Virtual Expert thus works with MRI technology.
The generated masks can be combined into a mismatch region of interest.
Volumetric calculations and ratios can be calculated from the computed regions of interest.
## Indications for Use
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the logo for Cercare Medical. The logo consists of a cluster of small teal circles with one larger red circle at the top left, followed by the text "Cercare Medical" in teal. The text is in a clean, sans-serif font and is aligned horizontally with the circle cluster.
Cercare Medical Neurosuite and associated modules, including the Capillary Function module, is an image processing software package to be used by trained professionals, including physicians and medical technicians.
The software package runs on standard off-the-shelf computer or a virtual platform, such as VMware, and can be used to perform image viewing, processing, and analysis of images. Data and images are acquired through DICOM (Digital Imaging and Communications in Medicine) compliant imaging devices
Cercare Medical Neurosuite provides viewing capabilities, whereas the Capillary Function module provides analysis capabilities for functional and dynamic imaging datasets acquired with MRI including a Diffusion Weighted MRI (DWI) Module and a Dynamic Analysis Module (dynamic contrast-enhanced imaging data for MRI).
The Capillary Function module is used for visualization and analysis of dynamic imaging data, showing properties of changes in contrast over time. This functionality includes calculation of parameters related to tissue flow (perfusion) and tissue blood volume. In addition, the Capillary Function module's DWI technology is used to visualize local water diffusion properties from the analysis of diffusion-weighted MRI data.
The Virtual Expert module allows the calculation of mirrored regions of interest and the visual inspection of time attenuation curves. One clinical application is to visualize the apparent blood perfusion and to calculate Hypoperfused Area and Mismatch Ratio in the brain tissue affected by acute stroke.
Areas of decreased perfusion appear as areas of changed signal intensity:
- Lower signal intensity for CBF and CBV
- Higher signal intensity for TTP, TTD, MTT, and Tmax ●
## Technological Characteristics
CMN with Capillary Function provides the following functions:
- processes DICOM images from multiple sources to provide visualization of changes of ● tissue perfusion, diffusion and change
- receives DICOM images from external DICOM image providers (modalities (MRI ● Scanners), PACS and Workstations) and sends DICOM images to external image consumers
- storage of status and results, and references therein, in a searchable database
CMN Capillary Function with Virtual Expert is a DICOM-compliant PACS software that provides comprehensive functionality to transfer, process and display modality specific imaging data. CMN runs on standard off-the-shelf computer and networking hardware and is entirely independent from CT, MRI or PACS platforms. It supports secure VPN networking or
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image contains the logo for Cercare Medical. The logo consists of a cluster of small turquoise circles with one larger red circle at the top left. To the right of the circles is the text "Cercare Medical" in a turquoise sans-serif font. The logo is simple and modern, with a focus on the company name.
encapsulated Secure Shell (SSH), and seamlessly integrates into an existing radiological data network.
The primary users of CMN software are medical imaging professionals who analyze tissue using CT or MRI images. The images generated by CMN provide additional diagnostic information, which is derived from the temporal/diffusion/density features of the native MRI image.
## Differences in Technical Characteristics/Performance with Respect to the Predicate Device
CMN Capillary Function with Virtual Expert is identical to the Predicate CMN Capillary Function (K202793) with the exception of the following features:
- . Output from the so-called SVD algorithm for computing perfusion derived markers has been exposed to the end-user and entails the common perfusion markers CBF, CBV, MTT, and Tmax, which to clearly distinguish these from the similarly named markers of the predicate device are called, respectively, CBF-basic, CBV-basic, MTT-basic, and Tmax-basic.
- Image thresholding features have been implemented via the Virtual Expert module, which provide thresholding of images and subsequent volumetric calculations relevant to ischemic stroke patient assessment.
The Subject and Predicate devices are based on the same technological elements of viewing, processing and analyzing DICOM image data to assist the clinician during diagnostic procedures. These differences do not impact the intended use or raise new questions with the safety and performance of the device.
## Performance Standards
CMN Capillary Function with Virtual Expert has been developed in conformance with the following standards and FDA guidance, as applicable:
- ISO 13485:2016, Quality management systems Requirements for regulatory purposes ●
- ISO 14971:2019, Medical devices Application of risk management to medical ● devices
- IEC 62304:2006/Amd 1:2015, Medical device software Software lifecycle processes ●
- IEC 82304-1:2016, Health Software Part 1: General Requirements for Product Safety
- IEC 62366-1:2015, Medical devices Part 1: Application of usability engineering to . medical devices
- ISO 15223-1:2021, Medical devices – Symbols to be used with medical device labels, labeling and information to be supplied
- ISO 20417:2021, Medical devices Information to be supplied by the manufacturer ●
- NEMA PS 3.1-3.20:2022 Digital Imaging and Communications in Medicine (DICOM) ●
{7}------------------------------------------------
Image /page/7/Picture/1 description: The image contains the logo for Cercare Medical. The logo consists of a cluster of small turquoise circles with one larger red circle at the top left, followed by the text "Cercare Medical" in a turquoise sans-serif font. The text is positioned to the right of the circle cluster.
- . Content of Premarket Submissions for Management of Cybersecurity in Medical Devices, October 2018
- Computer-Assisted Detection Devices Applied to Radiology Images and Radiology . Device Data - Premarket Notification [510(k)] Submissions, July 2012
## Performance Data
CMN Capillary Function with Virtual Expert complies with DICOM (Digital Imaging and Communications in Medicine), developed by the American College of Radiology and the National Electrical Manufacturers Association - NEMA PS 3.1-3.20.
Cercare conducted extensive performance validation testing and software verification and validation testing of the CMN Capillary Function with Virtual Expert. This performance validation testing demonstrated that the CMN system provides accurate representation of key processing parameters under a range of clinically relevant parameters associated with the intended use of the software. Software performance, validation and verification testing demonstrated that the CMN system met all design requirements and specifications.
Comparative performance testing was performed for the Subject Device with respect to the Predicate Device and Reference Device through bench testing including simulated digital phantoms and retrospective clinical data.
For the Capillary Function module, the retrospective clinical data was primarily used for visual inspection due to the relative nature of most perfusion image biomarkers, while rigorous testing was conducted through digital phantoms where the true parameter combinations were known. The simulated phantom data included variations of hemodynamic parameter combinations, while simultaneously simulating various experimental conditions such as patient motion, signal-tonoise ratios, and diffusion gradient schemes. Following the process undertaken with FDA on the Predicate Device (K202793), only human-like phantom testing and analyses were conducted since this approach allows for more realistic experimental settings (motion correction, noise, etc) and performance quantified through comparison of absolute bias, correlation coefficients, and multi-scale structural similarity index obtained for both devices. The established acceptance criteria were reached in all tests conducted.
For the Virtual Expert module, comparison testing of retrospective patient MR perfusion and diffusion imaging consisted of primary analyses of volumetric and spatial agreement per-patient and per-lesion. Secondary clinical application assessments (high-level context of the devices not associated with clinical claims) were then conducted in terms of image-driven decision to treat analysis through the so-called DEFUSE3 criteria. The Virtual Expert comparison testing was modeled by the comparison testing conducted for the Predicate Device (K202213) and in the associated published technical comparison study (Bathla et al, J. Neurointerventional Surg., 12:1028-1032 (2020)).
Together with software verification and validation, the performance validation demonstrated that CMN Capillary Function with Virtual Expert satisfies all design requirements and device specifications and is substantially equivalent to the Predicate Device and Reference Device.
{8}------------------------------------------------
Image /page/8/Picture/0 description: The image shows the logo for Cercare Medical. The logo consists of a cluster of blue circles with one red circle, followed by the text "Cercare Medical" in a light blue sans-serif font. The circles are arranged in a pattern, with the red circle standing out from the rest.
## Substantial Equivalence
In comparison with the Predicate CMN Capillary Function device (K202793) and Reference RAPID device (K182130), CMN Capillary Function with Virtual Expert has the same intended use and similar indications, technological characteristics, and principles of operation as described in the comparison table below.
{9}------------------------------------------------
Image /page/9/Picture/0 description: The image shows the logo for Cercare Medical. The logo consists of a cluster of small blue circles with one red circle at the top left, followed by the text "Cercare Medical" in a light blue sans-serif font. The logo is simple and modern, with a focus on the company name.
| Description | Subject Device | Predicate Device CMN<br>Capillary Function (K202793) | Reference Device IschemaView<br>Rapid (K182130) |
|--------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Product<br>Name | CMN Capillary Function with Virtual<br>Expert | CMN Capillary Function | IschemaView Rapid |
| Product Code<br>/ Regulation | LLZ / 21CFR/892.2050 | LLZ / 21CFR/892.2050 | LLZ / 21CFR/892.2050 |
| Indications<br>for Use | CMN and associated modules, including<br>the Capillary Function module, is an image<br>processing software package to be used by<br>trained professionals, including physicians<br>and medical technicians.<br>The software package runs on standard off-<br>the-shelf computer or a virtual platform,<br>such as VMware, and can be used to<br>perform image viewing, processing and<br>analysis of images. Data and images are<br>acquired through DICOM compliant<br>imaging devices.<br>CMN provides viewing capabilities,<br>whereas the Capillary Function module<br>provides analysis capabilities for<br>functional and dynamic imaging datasets<br>acquired with MRI including a Diffusion<br>Weighted MRI (DWI) Module and a<br>Dynamic Analysis Module (dynamic<br>contrast-enhanced imaging data for MRI).<br>The Capillary Function module is used for<br>visualization and analysis of dynamic<br>imaging data, showing properties of | CMN and associated modules,<br>including the Capillary Function<br>module, is an image processing<br>software package to be used by<br>trained professionals, including<br>physicians and medical technicians.<br>The software package runs on<br>standard off-the-shelf computer or a<br>virtual platform, such as VMware,<br>and can be used to perform image<br>viewing, processing and analysis of<br>images. Data and images are<br>acquired through DICOM compliant<br>imaging devices.<br>CMN provides viewing capabilities,<br>whereas the Capillary Function<br>module provides analysis capabilities<br>for functional and dynamic imaging<br>datasets acquired with MRI including<br>a Diffusion Weighted MRI (DWI)<br>Module and a Dynamic Analysis | iSchemaView RAPID is an image<br>processing software package to be<br>used by trained professionals,<br>including but not limited to physicians<br>and medical technicians.<br>The software runs on a standard off-<br>the-shelf computer or a virtual<br>platform, such as VMware, and can be<br>used to perform image viewing,<br>processing and analysis of images.<br>Data and images are acquired through<br>DICOM compliant imaging devices.<br>The iSchemaView RAPID provides<br>both viewing and analysis capabilities<br>for functional and dynamic imaging<br>datasets acquired with CT Perfusion<br>(CT-P), CT Angiography (CTA), and<br>MRI including a Diffusion Weighted<br>MRI (DWI) Module and a Dynamic<br>Analysis Module (dynamic contrast- enhanced imaging data for MRI and<br>CT). |
| Description | Subject Device | Predicate Device CMN<br>Capillary Function (K202793) | Reference Device IschemaView<br>Rapid (K182130) |
| | changes in contrast over time. This<br>functionality includes calculation of<br>parameters related to tissue flow<br>(perfusion) and tissue blood volume. In<br>addition, the Capillary Function module's<br>DWI technology is used to visualize local<br>water diffusion properties from the<br>analysis of diffusion-weighted MRI data.<br>The Virtual Expert module allows the<br>calculation of mirrored regions of interest<br>and the visual inspection of time<br>attenuation curves. One clinical application<br>is to visualize the apparent blood perfusion<br>and to calculate Hypoperfused Area and<br>Mismatch Ratio in the brain tissue affected<br>by acute stroke.<br>Areas of decreased perfusion appear as<br>areas of changed signal intensity:<br>Lower signal intensity for CBF and CBV Higher signal intensity for TTP, TTD, MTT, and TMax | Module (dynamic contrast-enhanced<br>imaging data for MRI).<br>The Capillary Function module is<br>used for visualization and analysis of<br>dynamic imaging data, showing<br>properties of changes in contrast over<br>time. This functionality includes<br>calculation of parameters related to<br>tissue flow (perfusion) and tissue<br>blood volume. In addition, the<br>Capillary Function module's DWI<br>technology is used to visualize local<br>water diffusion properties from the<br>analysis of diffusion-weighted MRI<br>data. | The DWI Module is used to visualize<br>local water diffusion properties from<br>the analysis of diffusion - weighted<br>MRI data.<br>The Dynamic Analysis Module is used<br>for visualization and analysis of<br>dynamic imaging data, showing<br>properties of changes in contrast over<br>time. This functionality includes<br>calculation of parameters related to<br>tissue flow (perfusion) and tissue<br>blood volume.<br>RAPID CT-Perfusion and RAPID<br>MR-Perfusion can be used by<br>physicians to aid in the selection of<br>acute stroke patients (with known<br>occlusion of the intracranial internal<br>carotid artery or proximal middle<br>cerebral artery) for endovascular<br>thrombectomy.<br>Instructions for use of contrast agents<br>for this indication can be found in<br>Appendix A of the User's Manual.<br>Additional information for safe and<br>effective drug use is available in<br>product specific iodinated CT and<br>gadolinium-based MR contrast drug<br>labeling. |
| Description | Subject Device | Predicate Device CMN Capillary Function (K202793) | Reference Device IschemaView Rapid (K182130) |
| Basic PACS Functions | View, process and analyze medical images. Communication of results through service class user protocols. | Yes | In addition to the RAPID imaging criteria, patients must meet the clinical requirements for thrombectomy, as assessed by the physician, and have none of the following contraindications or exclusions.<br><br>Contraindications/Exclusions:<br>Bolus Quality: absent or inadequate bolus. Patient Motion: excessive motion leading to artifacts that make the scan technically inadequate. Presence of Hemorrhage. |
| PACS Functionality | | | |
| Computer Platform | Standard off-the-shelf PC workstation / server<br><br>Virtual platform such as VMware | Yes | Yes |
| DICOM Compliance | Yes | Yes | Yes |
| Functional Overview | CMN is software package that provides for the visualization and study of changes of tissue in digital images captured by MRI. CMN provides viewing and quantification | Same | Same |
| Description | Subject Device | Predicate Device CMN Capillary Function (K202793) | Reference Device IschemaView Rapid (K182130) |
| Data Acquisition | Acquires medical image data from DICOM compliant imaging devices and modalities | Same | Same |
| Data / Image Types | MRI via DICOM format | Same | Same |
| <b>Acquisition and Modalities Features</b> | | | |
| MRI | Diffusion Weighted Image (DWI) | Yes | Yes |
| | Dynamic Analysis tissue flow (perfusion) and tissue blood volume | Yes | Yes |
| <b>Computed Parameter Maps</b> | | | |
| Diffusion MRI | Isotropic DWI (isoDWI) | Yes…
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.