← Product Code [KNW](/productcode/KNW) · K230015

# OmniBone™ Bone Marrow Biopsy Kit with Power Driver, OmniBone™ Bone Biopsy Kit with Power Driver (K230015)

_Laurane Medical Sas · KNW · Jun 16, 2023 · Gastroenterology, Urology · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K230015

## Device Facts

- **Applicant:** Laurane Medical Sas
- **Product Code:** [KNW](/productcode/KNW.md)
- **Decision Date:** Jun 16, 2023
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 876.1075
- **Device Class:** Class 2
- **Review Panel:** Gastroenterology, Urology

## Indications for Use

The Omnibone™ Bone Marrow Biopsy Kit is indicated for bone marrow aspiration and biopsy of the anterior or the posterior iliac crest of adult patients. The Omnibone™ Bone Biopsy Kit is indicated for bone biopsy of the vertebral body and bone lesions.

## Device Story

OmniBone™ Powered Biopsy System consists of a rechargeable, handheld, lithium-battery-powered driver and sterile, single-use biopsy kits. System used by clinicians to perform bone marrow aspiration/biopsy or bone lesion biopsy. Driver features 3-position switch (reverse/off/forward) and variable speed trigger; provides rotational kinetic energy to advance needle through cortical/cancellous bone. Clinician attaches needle assembly to driver via coupler and sterile pouch; uses driver to insert needle, then performs biopsy. Output is bone marrow or bone tissue sample for diagnostic evaluation. System aids in efficient needle insertion and sample acquisition compared to manual methods. Reusable driver is disinfected between uses; kits are EtO sterilized.

## Clinical Evidence

Bench testing only. Performance data included battery charging/capacity, tissue sampling, reliability, rotation speed, needle penetration force, coupler reliability, tensile strength, creep testing, and dead space measurements. Biocompatibility per ISO 10993-1, sterilization per ISO 11135:2014, and electrical safety/EMC per IEC 60601-2. All design verification and validation acceptance criteria met.

## Technological Characteristics

Handheld, cordless, rechargeable 4.2V lithium-ion battery-powered driver. Variable speed trigger; 3-position switch (forward/reverse). Sterile, single-use biopsy kits (cannulas, stylets, ejectors). Materials biocompatible per ISO 10993-1. Sterilization: EtO (kits). Electrical safety: IEC 60601-2. No external wired/wireless communication interfaces.

## Regulatory Identification

A gastroenterology-urology biopsy instrument is a device used to remove, by cutting or aspiration, a specimen of tissue for microscopic examination. This generic type of device includes the biopsy punch, gastrointestinal mechanical biopsy instrument, suction biopsy instrument, gastro-urology biopsy needle and needle set, and nonelectric biopsy forceps. This section does not apply to biopsy instruments that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.

## Predicate Devices

- OnControl™ Bone Access and Bone Biopsy System ([K113872](/device/K113872.md))
- OnControl™ Bone Marrow Biopsy System ([K142377](/device/K142377.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logos of the Department of Health and Human Services and the Food and Drug Administration (FDA). The Department of Health and Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.

June 16, 2023

Laurane Medical SAS

Mr. Laurent Fumex Head of Company Chemin de la bayette ZAC de la bayette Le Pradet, 83220 France

Re: K230015

Trade/Device Name: OmniBone™ Bone Marrow Biopsy Kit with Power Driver, OmniBone™ Bone Biopsy Kit with Power Driver

Regulation Number: 21 CFR 876.1075 Regulation Name: Gastroenterology-urology biopsy instrument Regulatory Class: Class II Product Code: KNW Dated: May 4, 2023 Received: May 8, 2023

Dear Mr. Fumex:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of 

{1}------------------------------------------------

Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

Mark
Digitally signed by
Mark Trumbore -S
Trumbore -S Date: 2023.06.16
08:22:14 -04'00'

Mark Trumbore, Ph.D. Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known) K230015

#### Device Name

- Omnibone™ Bone Marrow Biopsy Kit with Power Driver

- Omnibone™ Bone Biopsy Kit with Power Driver

#### Indications for Use (Describe)

The Omnibone™ Bone Marrow Biopsy Kit is indicated for bone marrow aspiration and biopsy of the anterior or the posterior iliac crest of adult patients.

The Omnibone™ Bone Biopsy Kit is indicated for bone biopsy of the vertebral body and bone lesions.

| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|

X Prescription Use (Part 21 CFR 801 Subpart D)

| | Over-The-Counter Use (21 CFR 801 Subpart C)

# CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the logo for Laurane Medical. The logo consists of a green square with a black curved line on the left side, followed by a purple square with a white "L" inside. To the right of the squares is the text "Laurane" in black, with the word "Medical" in a smaller font size below it.

K230015 Laurane Medical SAS Chemin de la Bayette ZAC de la Bayette 83220 Le Pradet France

## 510(k) Summary 21 CFR 807.92

As required by the Safe Medical Devices Act of 1990, coded under section 513, Part (I)(3)(A) of the Food, Drug and Cosmetic Act, a 510(k) summary upon which substantial equivalence determination is based as follows:

# 1. Submitter Information:

| Applicant:       | Laurane Medical SAS<br>Chemin de la Bayette<br>ZAC de la Bayette<br>Le Pradet 83220 - FRANCE                     |
|------------------|------------------------------------------------------------------------------------------------------------------|
| Phone :<br>Fax : | 860-399-9900<br>860-399-9200<br>Contact : Laurent Fumex<br>lfumex@lauranemedical.com: Laurane.medical@wanadoo.fr |

3 January 2023 Date :

#### 2. Subject Device Name :

| Device Trade Name : | OmniBone™ Bone Marrow Biopsy Kit with Power Driver<br>OmniBone™ Bone Biopsy Kit with Power Driver |
|---------------------|---------------------------------------------------------------------------------------------------|
| Common Name:        | Bone Marrow Biopsy Kit<br>Bone Biopsy Kit                                                         |
| Classification:     | Class II<br>KNW (Gastroenterology - urology biopsy instrument)                                    |
| Review Panel:       | Gastroenterology / Urology                                                                        |
| Regulation Number:  | 21 CFR 876.1075                                                                                   |

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image is a logo for Laurane Medical. The logo features a stylized letter "L" in black, with a green square behind it. To the right of the "L" is a purple square with a white "L" inside. The text "Laurane" is written in black, and the word "Medical" is written in a smaller font below it.

### 3. Predicate Device

| Device Trade Name: | OnControl™ Bone Access and Bone Biopsy System by Vidacare®<br>OnControl™ Bone Marrow Biopsy System by Vidacare® |
|--------------------|-----------------------------------------------------------------------------------------------------------------|
| 510(k) Number:     | K113872, K142377                                                                                                |
| Common/Usual Name: | Cement Dispenser Conduit for Vertebroplasty and Bone Biopsy<br>Needle, Bone Marrow Biopsy Needle                |
| Classification:    | Class II<br>KNW (Gastroenterology - Urology biopsy instrument)                                                  |
| Regulation number: | 21 CFR 876.1075                                                                                                 |

### 4. Subject Indications for Use:

The OmniBone™ Bone Biopsy Kit with Power Driver is indicated for bone biopsy of the vertebral body and bone lesions.

The OmniBone™ Bone Marrow Biopsy kit with Power Driver is indicated for bone marrow aspiration and biopsy of the anterior or the posterior iliac crest of adult patients.

#### 5. Subject Device Description

The OmniBone™ Powered Biopsy System consists of a rechargeable Power Driver with Charger and compatible Biopsy Kits with Power Driver Pouch.

The OmniBone™ Bone Biopsy Kit with Power Driver Pouch and the OmniBone™ Bone Marrow Biopsy Kit with Power Driver Pouch are both provided sterile (EtO), are both single-use and are both designed exclusively for use with the OmniBone™ Power Driver. All OmniBone™ Biopsy Kits are available in different needle gauge sizes and lengths.

The OmniBone™ Power Driver with Charger is a handheld, Lith-on battery powered, and rechargeable Power Driver, reusable and supplied non-sterile .

{5}------------------------------------------------

Image /page/5/Picture/0 description: The image is a logo for Laurane Medical. The logo features a stylized letter "L" in black, with a green square behind it. To the right of the green square is a purple square with a white "L" inside. The text "Laurane" is written in black to the right of the squares, and the word "Medical" is written in a smaller font below it.

The OmniBone™ Power Driver has a 3-position switch that controls the motion of the driver: reverse (counter clockwise) – off – forward (clockwise). The OmniBone™ Power Driver has a variable speed trigger that controls the speed of the driver. Applying light pressure on the trigger activates the driver at low speed and increasing the trigger pressure activates higher speeds. Full pressure on the trigger will activate full speed. No pressure on the trigger will stop the motor function. A charging LED indicates the level of charge of the battery of the OmniBone™ Power Driver. An integrated connector allows the OmniBone™ Power Driver to connect to the OmniBone™ Charger.

#### 6. Technological Comparison to Predicate Device

The table below provides a technological comparison between the subject device and the predicate device.

| Characteristic           |                | Predicate<br>OnControl™ Powered Bone Access<br>System<br>OnControl™ Bone Marrow Biopsy System -<br>K142377<br>OnControl™ Bone Access and Bone Biopsy<br>System - K113872                                                                                               | Subject<br>OmniBone™ Powered Biopsy<br>System<br>OmniBone™ Bone Marrow Biopsy kit with<br>Power Driver<br>OmniBone™ Bone Biopsy kit with Power<br>Driver         |
|--------------------------|----------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Regulation Number        |                | 21 CFR 876.1075                                                                                                                                                                                                                                                        | 21 CFR 876.1075                                                                                                                                                  |
| Class                    |                | II                                                                                                                                                                                                                                                                     | II                                                                                                                                                               |
| Primary FDA Product Code |                | KNW - Instrument, Biopsy<br>FCG -<br>Biopsy Needle                                                                                                                                                                                                                     | KNW - Instrument, Biopsy                                                                                                                                         |
| Indications for Use      | Bone<br>Marrow | The OnControl™ Bone Marrow Biopsy<br>System is intended for bone marrow<br>aspiration and biopsy in adult and<br>pediatric patients age 2 and older.                                                                                                                   | The OmniBone™ bone marrow biopsy kit<br>is indicated for bone marrow aspiration<br>and biopsy of the anterior or the posterior<br>iliac crest of adult patients. |
|                          | Bone<br>Lesion | The OnControl™ Bone Access and<br>Bone Biopsy System is intended for use<br>with a standard cement delivery<br>system for the fixation of fractures of<br>the vertebral body using<br>vertebroplasty and/or for bone biopsy<br>of the vertebral body and bone lesions. | The OmniBone™ Bone Biopsy Kit with<br>Power Driver is intended for bone biopsy<br>of the vertebral body and bone lesions.                                        |

{6}------------------------------------------------

Image /page/6/Picture/0 description: The image shows the logo for Laurane Medical. The logo consists of a stylized letter "L" in black, with a green square behind it. To the right of the "L" is the word "Laurane" in black, with the word "Medical" in a smaller font below it. The logo is simple and modern, and the colors are calming and professional.

| K230015 |  |  |
|---------|--|--|
|         |  |  |

| Characteristic                           | Predicate<br>OnControl™ Powered Bone Access<br>System<br>OnControl™ Bone Marrow Biopsy System -<br>K142377<br>OnControl™ Bone Access and Bone Biopsy<br>System - K113872 | Subject<br>OmniBone™ Powered Biopsy<br>System<br>OmniBone™ Bone Marrow Biopsy kit with<br>Power Driver<br>OmniBone™ Bone Biopsy kit with Power Driver             |                                                                                                                                                                      |
|------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Target Population                        | Bone<br>Marrow                                                                                                                                                           | Adult and pediatric patients needing<br>bone marrow aspiration or bone<br>marrow biopsy.                                                                          | Adult patients needing bone marrow<br>aspiration or bone marrow biopsy.                                                                                              |
|                                          | Bone<br>Lesion                                                                                                                                                           | Patients requiring fixation of fractures<br>of the vertebral body or bone biopsy.                                                                                 | Patients requiring bone biopsy.                                                                                                                                      |
| Fundamental Scientific<br>Technology     |                                                                                                                                                                          | Power Driver with Needle<br>Attachments                                                                                                                           | Power Driver with Needle Attachments                                                                                                                                 |
| Power Driver Design                      |                                                                                                                                                                          | Cordless, lithium battery-powered;<br>reusable with disposable sterile sleeve                                                                                     | Cordless, lithium battery-powered;<br>reusable with disposable sterile<br>sleeve                                                                                     |
| Power Driver Energy Delivered            |                                                                                                                                                                          | Non rechargeable, Internal, 18 V<br>Lithium Battery                                                                                                               | Rechargeable, Internal, 4.2 V Lithium<br>Battery                                                                                                                     |
| Power Driver<br>Cybersecurity/Interfaces |                                                                                                                                                                          | Does not contain any external wired<br>and/or wired communication<br>interfaces                                                                                   | Does not contain any external wired<br>and/or wired communication<br>interfaces                                                                                      |
| Charger                                  |                                                                                                                                                                          | No charger                                                                                                                                                        | Charger for power driver with DC power<br>plug                                                                                                                       |
| Needle Design                            |                                                                                                                                                                          | Sterile, single-use, disposable                                                                                                                                   | Sterile, single-use, disposable                                                                                                                                      |
| Operating Principle                      |                                                                                                                                                                          | Needle set attaches to battery<br>powered driver                                                                                                                  | Needle set attaches to battery powered<br>driver                                                                                                                     |
| Mode of Action                           |                                                                                                                                                                          | Single puncture, bone access, and<br>sample                                                                                                                       | Single puncture, bone access, and sample                                                                                                                             |
| Energy Used / Delivered                  |                                                                                                                                                                          | Lithium battery provides rotational<br>kinetic energy to aid the physician in<br>inserting and advancing the needle<br>through cortical and/or cancellous<br>bone | Lithium-on battery provides rotational<br>kinetic energy to aid the physician in<br>inserting and advancing the needle<br>through cortical and/or cancellous<br>bone |
| Imaging Compatibility                    |                                                                                                                                                                          | X-Ray / CT and Ultrasound<br>Compatibility                                                                                                                        | X-Ray / CT and Ultrasound Compatibility                                                                                                                              |

{7}------------------------------------------------

Image /page/7/Picture/0 description: The image shows the logo for Laurane Medical. The logo consists of a stylized "L" shape in black, placed next to a green square with a purple square inside containing the letter "L" in white. To the right of the squares, the word "Laurane" is written in black, with the word "Medical" written in a smaller font size below it.

| K230015 |  |  |
|---------|--|--|
|         |  |  |

| Characteristic                          |             | Predicate<br>OnControl™ Powered Bone Access<br>System<br>OnControl™ Bone Marrow Biopsy System -<br>K142377<br>OnControl™ Bone Access and Bone Biopsy<br>System - K113872                                                                                                                                       | Subject<br>OmniBone™ Powered Biopsy<br>System<br>OmniBone™ Bone Marrow Biopsy Kit with<br>Power Driver<br>OmniBone™ Bone Biopsy Kit with Power<br>Driver                                                                                                                                                                                                              |
|-----------------------------------------|-------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Speed                                   |             | Single-Speed                                                                                                                                                                                                                                                                                                   | Variable                                                                                                                                                                                                                                                                                                                                                              |
| Power Driver Motion                     |             | Forward motion                                                                                                                                                                                                                                                                                                 | Reverse and forward motion                                                                                                                                                                                                                                                                                                                                            |
| Kit<br>Components                       | Bone Marrow | OnControl™ Bone Marrow Biopsy Tray<br>- Alignment Guide<br>- Biopsy Needle<br>- Depth Stop<br>- Ejector Rod<br>- Connector Hub with Sterile Sleeve<br>- Fenestrated Drape                                                                                                                                      | OmniBone™ Bone Marrow Biopsy Kit<br>with Power Driver Pouch<br>● Needle set:<br>Biopsy Cannula and Stylet with depth<br>gauge (Biopsy Needle Assembly)<br>- Ejector Pin<br>- Ejector guide<br>● Pouch set:<br>- Power Driver Pouch<br>- Coupler                                                                                                                       |
|                                         | Bone Lesion | OnControl™ Bone Lesion Biopsy Tray<br>- Transfer Rod<br>- Bone Access Ejector Rod<br>- Bone Access Needle Set<br>- Bone Lesion Biopsy Needle<br>- Bone Lesion Biopsy Ejector<br>Rod<br>- Depth Stop<br>- Manual Handle<br>- Connector Hub with Sterile<br>Sleeve<br>- Fenestrated Drape<br>- Ejector Assistant | OmniBone™ Bone Biopsy Kit with Power<br>Driver Pouch<br>● Needle set:<br>- Bone Access Introducer (Outer cutting<br>Cannula + Drill tip Stylet + depth<br>gauge assembly)<br>- Trephine Biopsy Needle (Trephine<br>Outer Cannula + Stiffener + depth<br>gauge assembly)<br>- Ejector Pin<br>- Handle<br>- Holder<br>● Pouch set:<br>- Power Driver Pouch<br>- Coupler |
| Gauge Sizes                             | Bone Marrow | Biopsy Cannula:11G                                                                                                                                                                                                                                                                                             | Biopsy Cannula: 11G                                                                                                                                                                                                                                                                                                                                                   |
|                                         | Bone Lesion | Access Cannula:<br>10G and 11G<br>Biopsy Cannula:<br>12G and 13G                                                                                                                                                                                                                                               | Introducer Cannula:<br>11G<br>Trephine Outer Cannula:<br>13G                                                                                                                                                                                                                                                                                                          |
| Characteristic                          |             | Predicate<br>OnControl™ Powered Bone Access<br>System<br>OnControl™ Bone Marrow Biopsy System -<br>K142377<br>OnControl™ Bone Access and Bone Biopsy<br>System - K113872                                                                                                                                       | Subject<br>OmniBone™ Powered Biopsy<br>System<br>OmniBone™ Bone Marrow Biopsy Kit with<br>Power Driver<br>OmniBone™ Bone Biopsy Kit with Power Driver                                                                                                                                                                                                                 |
| Overall<br>Needle<br>Assembly<br>Length | Bone Marrow | Biopsy Needle Assembly:<br>102 and 152 mm                                                                                                                                                                                                                                                                      | Biopsy Needle Assembly: 100<br>and 150 mm                                                                                                                                                                                                                                                                                                                             |
|                                         | Bone Lesion | Bone Access Needle Assembly: 62,<br>102 and 152 mm                                                                                                                                                                                                                                                             | Bone Access Introducer Assembly:<br>65,5, 106 and 156 mm                                                                                                                                                                                                                                                                                                              |
|                                         | Bone Lesion | Biopsy Needle Assembly:<br>108, 148, and 198 mm                                                                                                                                                                                                                                                                | Biopsy Trephine Assembly:<br>154,194,5 and 234,5mm                                                                                                                                                                                                                                                                                                                    |
| Biopsy Kits - Sterility                 |             | Single Use, Ethylene Oxide                                                                                                                                                                                                                                                                                     | Single Use, Ethylene Oxide                                                                                                                                                                                                                                                                                                                                            |
| Power Driver - Sterility                |             | Reusable, non-sterile.<br>Used with Single Use Disposable Sterile<br>Sleeve with Connector Hub                                                                                                                                                                                                                 | Reusable, non-sterile.<br>Used with Single Use Disposable<br>Sterile Pouch and Coupler                                                                                                                                                                                                                                                                                |
| Power Driver - Disinfection             |             | Disinfected utilizing an antimicrobial<br>solution                                                                                                                                                                                                                                                             | Disinfected utilizing an antimicrobial<br>solution                                                                                                                                                                                                                                                                                                                    |

{8}------------------------------------------------

Image /page/8/Picture/0 description: The image is a logo for Laurane Médical. The logo features a stylized "L" in black, with a green square behind it. To the right of the "L" is the word "Laurane" in black, with the word "Médical" in a smaller font below it. The "L" is partially contained within a purple square.

The predicate device is the OnControl™ Powered Bone Access System, which is comprised of the OnControl™ Bone Marrow Biopsy System and the OnControl™ Bone Access and Bone Biopsy System. The OnControl™ Powered Bone Access System is referenced as the predicate device for the OmniBone™ Powered Biopsy System as it is the same or similar to the subject device in the following ways:

#### a. Intended Use

#### b. Indications for Use

- The OmniBone™ Bone Biopsy Kit with Power Driver shares the same indications for use as the predicate OnControl™ Bone Access and Bone Biopsy System as they share a subset of the OnControl™ Bone Access and Bone Biopsy System. The subject device and the predicate device are both indicated for the sampling of bone biopsies from the vertebral body and from bone lesions. Indications for vertebroplasty have been omitted.
- The OmniBone™ Bone Marrow Biopsy Kit with Power Driver shares the same indications for use as the predicate OnControl Bone Marrow Biopsy System as they are both indicated for bone marrow aspiration and biopsy in adult patients. Indication for pediatric patients has been omitted.

#### c. Target population

- d. Fundamental Scientific Technology
- e. Performance characteristics

{9}------------------------------------------------

Image /page/9/Picture/0 description: The image is a logo for Laurane Medical. The logo features a stylized letter "L" in black, with a green square behind it. To the right of the "L" is the word "Laurane" in black, with the word "Medical" in a smaller font below it. The logo is simple and modern, and the colors are calming and professional.

- f. Operating Principle, Mechanism of action
- g. Sterility Assurance Level and method of Sterilization

The subject device is different from the predicate device in the following ways:

#### -Enhanced Power Driver Technology

- a. Variable Speed Functionality: The subject device utilizes a variable speed Power Driver. The variable speed functionality was designed to allow the user to control the speed of the needle rotation depending on their needs and the type of procedure being performed.
- b. The subject device's offers 2 possible driving motion: reverse and forward and is controlled by a 3-position switch. The addition of a reverse motion eases the introduction of the access needle through soft tissue and the removal of the needles from the bone.
- Reference LED Indicator that displays Power Level: The subject device utilizes a Reference ﻥ LED indicator that displays 3 level of Power Level.
- d. Rechargeable: the subject device has a DC jack connector and is rechargeable via a dedicated Charger.
- e. Driveshaft: the subject device has no driveshaft extending from the housing but a fully integrated female connector instead facilitating the cleaning of the Power Driver. This difference has no impact on mode of action or on safety.

#### Biopsy Kit Components

- Optimized Biopsy and Introducer Cannula Dimensions and Tip Design: The subject device a. Biopsy and Introducer Cannulas feature enhanced designs for optimal sample acquisitions.
- b. Unlike the connector sleeve of the Predicate, the subject device has a sterile pouch and a sterile coupler that are not factory-assembled together. The pouch and the coupler are assembled during the procedure. This design has no impact on the efficiency of the barrier-like system and shows the same level of performance and safety than the predicate.
- Handle Manual Driver (Bone Biopsy Kit ): The subject device's Manual Driver has a Handle ﻥ and attaches to the needle assembly via an interfacing and locking mechanism.
- d. Easy Connect Release Mechanism: The subject device is designed to release the Needle Assembly from the Power Driver by simply pressing the releasable locking fins of the Luer-Hub of the needles.

#### 7. Performance Data

The OmniBone™ Powered Bone Biopsy System which includes the OmniBone™ Power Driver, the OmniBone™ Charger, the OmniBone™ Bone Biopsy Kit with Power Driver Pouch and the OmniBone™ Bone Marrow Biopsy Kit with Power Driver Pouch was tested to evaluate its performances and support the determination of substantial equivalence.

{10}------------------------------------------------

Image /page/10/Picture/0 description: The image is a logo for Laurane Medical. The logo features a stylized "L" shape in black, placed over a green square. To the right of the green square is a purple square with a white "L" inside. The text "Laurane" is written in black to the right of the squares, and the word "Medical" is written in a smaller font below "Laurane".

The following tests were performed on the subject device:

- Battery charging time -
- Battery capacity -
- -Tissue sampling
- Reliability Testing Power Driver Pouch Sealing -
- Rotation Speed -
- -Needle Penetration Force
- Power Driver Coupler Reliability -
- -Tensile Strength – Needle hub
- Creep Testing -
- Dead Space Measurements -

Additionally, biocompatibility testing was performed in accordance with ISO 10993-1 to demonstrate that the OmniBone™ Powered Biopsy System is biocompatible for its intended use. Sterilization was performed in accordance with ISO 11135:2014 to confirm the Sterility Assurance Level (SAL) of 10-6 for the OmniBone™ Biopsy Kits. The label shelf life is 5 years. Electromagnetic Compatibility and Electrical Safety Testing were performed on the OmniBone™ Power Driver and Charger in accordance with IEC 60601-2. The results from this testing demonstrate that the OmniBone™ Power Driver and Charger meets electrical safety and performance requirements established, and acceptance criteria for all tests were met. The subject device OmniBone™ Powered Biopsy system (OmniBone™ Power Driver, OmniBone™ Charger, OmniBone™ Bone Biopsy Kit with Power Driver Pouch, OmniBone™ Bone Marrow Biopsy Kit with Power Driver Pouch) met all the predetermined acceptance of design verification and validation as specified by applicable standards, guidance, tests protocols and/or customer inputs.

#### 8. Conclusion

The proposed OmniBone™ Powered Biopsy System (OmniBone™ Power Driver, OmniBone™ Charger, OmniBone™ Bone Biopsy Kit with Power Driver Pouch, OmniBone™ Bone Marrow Biopsy Kit with Power Driver Pouch) and the predicate share the same or similar intended use, indications for use, target population, technological characteristics and fundamental scientific technology. Therefore, Laurane Medical SAS has demonstrated that the subject device OmniBone™ Powered Biopsy System (OmniBone™ Power Driver, OmniBone™ Charger, OmniBone™ Bone Biopsy Kit with Power Driver Pouch, OmniBone™ Bone Marrow Biopsy Kit with Power Driver Pouch) is substantially equivalent to the legally marketed predicate device OnControl™ Bone Marrow Biopsy System and OnControl™ Bone Access and Bone Biopsy System, respectively.

---

**Source:** [https://fda.innolitics.com/device/K230015](https://fda.innolitics.com/device/K230015)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
