K223889 · Healthium Medtech Limited · MBI · Mar 22, 2023 · Orthopedic
Device Facts
Record ID
K223889
Device Name
Stativ® Knotted UHMWPE Suture Anchor
Applicant
Healthium Medtech Limited
Product Code
MBI · Orthopedic
Decision Date
Mar 22, 2023
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.3040
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
STATIV® Knotted UHMWPE Suture Anchor with the sutures/tapes are Intended for soft tissue fixation to the bone. These sutures/tapes may be incorporated, as components, into surgeries where constructs including those with allograft or auto graft tissues are used for repair.
Device Story
STATIV® Knotted UHMWPE Suture Anchor is a sterile, single-use fixation device for soft tissue-to-bone repair. It consists of a soft anchor preloaded with non-absorbable UHMWPE sutures or tapes, mounted on an inserter with a handle. Available in self-tap/self-punching variants. During surgery, the anchor is deployed into a pilot hole; upon deployment, the anchor structure bunches into a tri-pod pattern, increasing its size to provide pull-out strength. Used by surgeons in orthopedic procedures (shoulder, knee, hip, foot/ankle, hand/wrist, elbow). The device provides secure mechanical fixation, allowing for tissue healing. Benefits include high pull-out strength and secure construct for various soft tissue repairs.
Clinical Evidence
No clinical data. Substantial equivalence supported by bench performance testing, including pull-out strength, cyclical loading, bacterial endotoxin testing, and biocompatibility (skin sensitization, intracutaneous reactivity, material-mediated pyrogenicity, acute systemic toxicity, in vitro cytotoxicity).
Technological Characteristics
Anchor: UHMWPE loop/suture/tape with polycaprolactone additive. Dispenser: 300 series stainless steel. Handle: ABS. Ring: Silicone rubber. Principle: Mechanical soft tissue-to-bone fixation via bunched anchor deployment. Sterilization: Ethylene Oxide (EO). Single-use.
Indications for Use
Indicated for fixation of soft tissue to bone in shoulder (rotator cuff, SLAP, biceps tenodesis, AC separation, deltoid, capsular reconstruction), foot/ankle (stabilization, Achilles, metatarsal, hallux valgus, tendon transfers, mid-foot), knee (collateral ligaments, patellar tendon, iliotibial band, joint capsule), hand/wrist (scapholunate, collateral ligaments, flexor/extensor tendons, carpal ligaments, CMC arthroplasty), elbow (biceps reattachment, collateral ligaments), and hip (acetabular labral, capsular, gluteal tendon) procedures.
Regulatory Classification
Identification
A smooth or threaded metallic bone fixation fastener is a device intended to be implanted that consists of a stiff wire segment or rod made of alloys, such as cobalt-chromium-molybdenum and stainless steel, and that may be smooth on the outside, fully or partially threaded, straight or U-shaped; and may be either blunt pointed, sharp pointed, or have a formed, slotted head on the end. It may be used for fixation of bone fractures, for bone reconstructions, as a guide pin for insertion of other implants, or it may be implanted through the skin so that a pulling force (traction) may be applied to the skeletal system.
{0}------------------------------------------------
March 22, 2023
Image /page/0/Picture/1 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health and Human Services logo on the left and the FDA logo on the right. The FDA logo is a blue square with the letters "FDA" in white, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue.
Healthium Medtech Limited Pankaj Dawar Deputy General Manager Regulatory Affairs 472-D. 13th Cross, 4th Phase, Peenya Industrial Area Bangalore. Karnataka 560058 India
Re: K223889
Trade/Device Name: Stativ® Knotted UHMWPE Suture Anchor Regulation Number: 21 CFR 888.3040 Regulation Name: Smooth Or Threaded Metallic Bone Fixation Fastener Regulatory Class: Class II Product Code: MBI Dated: December 23, 2022 Received: December 27, 2022
Dear Pankaj Dawar:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part
{1}------------------------------------------------
801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4. Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
# Sara S. Thompson -S
For
Laurence D. Coyne. Ph.D. Director DHT6C: Division of Restorative, Repair and Trauma Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K223889
Device Name STATIV® Knotted UHMWPE Suture Anchor
### Indications for Use (Describe)
The STATIV® Knotted UHMWPE Suture Anchor is indicated for the fixation, by use of sutures, of soft tissue to bone in the following surgical procedures:
1. Shoulder: Rotator Cuff Repair, SLAP Lesion Repair, Biceps Tenodesis, Acromio-Clavicular Separation Repair, Deltoid Repair, Capsular Shift or Capsulolabral Reconstruction.
2. Foot/Ankle: Lateral Stabilization, Medial Stabilization, Achilles Tendon Repair, Metatarsal Ligament Repair, Hallux Valgus reconstruction, digital tendon transfers, Mid-foot reconstruction.
3. Knee: Medial Collateral Ligament Repair, Lateral Collateral Ligament Repair, Patellar Tendon Repair, Posterior Oblique Ligament Repair, Iliotibial Band Tenodesis, Joint Capsule Closure.
4. Hand/Wrist: Scapholunate Ligament Reconstruction of collateral ligaments, Repair of Flexor and Extensor Tendons at the PIP, DIP and MCP joints for all digits, digital tendon transfers, Carpal Ligament Reconstruction and Carpometacarpal joint arthroplasty (basal thumb joint arthroplasty).
5. Elbow: Biceps tendon reattachment, Ulnar or Radial collateral ligament reconstruction.
6. Hip: Acetabular labral repair, Capsular repair, Gluteal Tendon Repair.
| Type of Use (Select one or both, as applicable) | |
|----------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------|
| <span style="font-size: 10pt;"> <span style="font-family: Symbol;">☑</span> </span> Prescription Use (Part 21 CFR 801 Subpart D) | <span style="font-size: 10pt;"> <span style="font-family: Symbol;">☐</span> </span> Over-The-Counter Use (21 CFR 801 Subpart C) |
CONTINUE ON A SEPARATE PAGE IF NEEDED### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for Sironix Arthroscopy Solutions. The word "Sironix" is in large, bold, black letters, with the "O" in "Sironix" being a red circle. Below "Sironix" is the phrase "Arthroscopy Solutions" in smaller, black letters.
Image /page/3/Picture/2 description: The image contains the text "K223889" at the top left. To the right of the text is a logo with an orange circle overlapping a blue circle with a white cross in the middle. Below the text and logo is the word "Healthium" in a blue, sans-serif font.
# 1. 510(k) Summary
#### 1.1. Submitter Information:
| Application Correspondent: | PANKAJ DAWAR<br>Healthium Medtech Limited<br>472-D, 13th Cross, 4th Phase,<br>Peenya Industrial Area,<br>Bangalore, Karnataka, 560058, INDIA. |
|----------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------|
| Phone: | +91-9886529934 |
| E-mail: | pankaj.d@healthiummedtech.com |
| Specification Developer: | Healthium Medtech Limited<br>472-D, 13th Cross, 4th Phase,<br>Peenya Industrial Area,<br>Bangalore, Karnataka, 560058, INDIA. |
| Contact Person: | PANKAJ DAWAR |
| Phone: | + 91-80-41868000 |
| E-mail: | pankaj.d@healthiummedtech.com |
| Date Prepared: | 20-03-2023 |
#### 1.2. Device Identification:
| Device Trade Name: | STATIV® Knotted UHMWPE Suture Anchor |
|----------------------|----------------------------------------------------|
| Device Common Name: | Non-Absorbable Suture Anchor |
| Classification Name: | Smooth or threaded metallic bone fixation fastener |
| Device Class: | Class II |
| Regulation Number: | 21 CFR 888.3040 |
| Product Code: | MBI |
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for Sironix. The word "SIRONIX" is written in a sans-serif font, with the "O" in "SIRONIX" being a red circle. Below the word "SIRONIX" is the phrase "Arthroscopy Solutions" in a smaller font.
Image /page/4/Picture/2 description: The image contains the text "K223889" in a bold, sans-serif font. Below the text is the word "Healthium" in a larger, bold, sans-serif font. To the right of the text is a logo consisting of an orange circle overlapping a blue circle, with a white cross shape where they intersect.
#### 1.3. Predicate Devices:
| Device Name | 510(k) Number |
|------------------------------------------|---------------|
| Arthrex FiberTak Suture Anchor (Primary) | K203268 |
| Fixone All Suture Anchor (Secondary) | K192709 |
#### Device Description 1.4.
The STATIV® Knotted UHMWPE Suture Anchor is a fixation device intended to provide secure fixation of soft tissue to bone and is available in self-tap/self-punching variant also. It consists of a soft Suture Anchor with attached non-absorbable suture(s) to an inserter with handle. The Anchor are available in various sizes, preloaded with suture, tape or suture-tape combinations. This device is provided sterile, for single use only.
# Materials Specifications
- 1. Anchor :
- > Loop: Non-absorbable, UHMWPE USP #3, USP #7
- A Suture : Non-absorbable, UHMWPE USP #2
- A Tape : Non-absorbable UHMWPE 1.5mm Round-Flat-Round
- A Additive: Polycaprolactone (Average Mn 80,000)
- 2. Dispenser : Stainless Steel (SS 300 Series)
- 3. Handle : ABS (Acrylonitrile Butadiene Styrene)
- Ring : Silicone Rubber Ring 4.
#### Intended Use & Indications for Use 1.5.
# Intended Use
STATIV® Knotted UHMWPE Suture Anchor with the sutures/tapes are Intended for soft tissue fixation to the bone. These sutures/tapes may be incorporated, as components, into surgeries where constructs including those with allograft or auto graft tissues are used for repair.
# Indications for Use
The STATIV® Knotted UHMWPE Suture Anchor is indicated for the fixation, by use of sutures, of soft tissue to bone in the following surgical procedures:
- 1. Shoulder: Rotator Cuff Repair, Bankart Repair, SLAP Lesion Repair, Biceps Tenodesis, Acromio-Clavicular Separation Repair, Deltoid Repair, Capsular Shift or Capsulolabral Reconstruction.
- 2. Foot/Ankle: Lateral Stabilization, Medial Stabilization, Achilles Tendon Repair, Metatarsal Ligament Repair, Hallux Valgus reconstruction, digital tendon transfers, Midfoot reconstruction.
| Healthium Medtech Limited | |
|----------------------------------------------------------------------------------------------------|-------------|
| Registered Office: 472/D, 13th Cross, 4th Phase, Peenya Industrial Area, Bengaluru – 560058, India | Page 2 of 7 |
| www.healthiummedtech.com CIN : U03311KA1992PLC013831 | |
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the logo for Sironix Arthroscopy Solutions. The word "Sironix" is written in a bold, sans-serif font, with the "O" in "Sironix" replaced by a red circle. Below the word "Sironix" is the phrase "Arthroscopy Solutions" in a smaller, sans-serif font.
Image /page/5/Picture/2 description: The image contains the text "K223889" at the top, followed by the word "Healthium" in a stylized font. To the right of the text is a logo consisting of a blue circle intersected by a white cross, with an orange circle partially overlapping the blue one. The logo is simple and modern, complementing the text.
- 3. Knee: Medial Collateral Ligament Repair, Lateral Ligament Repair, Patellar Tendon Repair, Posterior Oblique Ligament Repair, Iliotibial Band Tenodesis, Joint Capsule Closure.
- 4. Hand/Wrist: Scapholunate Ligament Reconstruction, Repair/Reconstruction of collateral ligaments, Repair of Flexor and Extensor Tendons at the PIP, DIP and MCP joints for all digits, digital tendon transfers, Carpal Ligament Reconstruction and Carpometacarpal joint arthroplasty (basal thumb joint arthroplasty).
- 5. Elbow: Biceps tendon reattachment, Ulnar or Radial collateral ligament reconstruction.
- 6. Hip: Acetabular labral repair, Capsular repair, Gluteal Tendon Repair.
#### 1.6. Comparison of Technological Characteristics
The fundamental scientific technology, materials of construction and mechanism of operation are similar between the subject device STATIV® Knotted UHMWPE Suture Anchor and the predicate device.
Table 2 summarizes the comparison of technological characteristics between the subject and predicate device.
| S. No | Parameters | Arthrex FiberTak Suture<br>Anchor (K203268) | STATIV® Knotted<br>UHMWPE Suture Anchor<br>(Subject device) | Comments |
|-------|----------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------|
| 1. | Manufacturer | Arthrex Inc. | Healthium Medtech Limited | - |
| 2. | Product Code | MBI | MBI | Similar |
| 3. | Regulation<br>Number | 21 CFR 888.3040 | 21 CFR 888.3040 | Similar |
| 4. | Classification | Class II | Class II | Similar |
| 5. | Intended Use | The Arthrex FiberTak Suture<br>Anchor is intended for<br>fixation of suture (soft tissue)<br>to bone in the shoulder,<br>foot/ankle, knee, hand/wrist,<br>elbow, and hip in the<br>following procedures:<br>Elbow: Biceps Tendon<br>Reattachment, Ulnar or<br>Radial Collateral Ligament<br>Reconstruction<br>Shoulder: Rotator Cuff<br>Repair, Bankart Repair, SLAP<br>Lesion Repair, Biceps<br>Tenodesis, Acromio-<br>Clavicular Separation Repair, | The STATIV® Knotted<br>UHMWPE Suture Anchor is<br>indicated for the fixation, by<br>use of sutures, of soft tissue to<br>bone in the following surgical<br>procedures:<br>1. Shoulder: Rotator Cuff<br>Repair, Bankart Repair,<br>SLAP Lesion Repair,<br>Biceps Tenodesis,<br>Acromio-Clavicular<br>Separation Repair, Deltoid<br>Repair, Capsular Shift or<br>Capsulolabral<br>Reconstruction.<br>2. Foot/Ankle: Lateral<br>Stabilization, Medial | Similar |
# Table 2: Substantial Equivalence Table
| Healthium Medtech Limited | |
|--------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------|
| Registered Office: 472/D, 13th Cross, 4th Phase, Peenya Industrial Area, Bengaluru – 560058, India<br>www.healthiummedtech.com CIN : U03311KA1992PLC013831 | Page 3 of 7 |
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the logo for Sironix. The logo is black, except for the "O" in Sironix, which is red. Below the company name is the text "Arthroscopy Solutions".
Image /page/6/Picture/2 description: The image contains the text "K223889" in a bold, sans-serif font at the top. Below the text is the word "Healthium" in a stylized font, with the "H" being larger than the other letters. To the right of the text is a logo consisting of two overlapping circles, one orange and one blue, with a white cross shape in the center where they overlap.
| S.<br>No | Parameters | Arthrex FiberTak Suture<br>Anchor (K203268) | STATIV® Knotted<br>UHMWPE Suture Anchor<br>(Subject device) | Comments |
|----------|------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------|
| | | Deltoid Repair, Capsular Shift<br>or Capsulolabral<br>Reconstruction<br><b>Hand/Wrist:</b> Scapholunate<br>Ligament Reconstruction,<br>Repair/Reconstruction of<br>collateral ligaments, Repair of<br>Flexor and Extensor Tendons<br>at the PIP, DIP and MCP<br>joints for all digits, digital<br>tendon transfers, Carpal<br>Ligament Reconstruction and<br>Carpometacarpal joint<br>arthroplasty (basal thumb<br>joint arthroplasty)<br><b>Foot/Ankle:</b> Lateral<br>Stabilization, Medial<br>Stabilization, Achilles Tendon<br>Repair, Metatarsal Ligament<br>Repair, Hallux Valgus<br>reconstruction, digital tendon<br>transfers, Mid-foot<br>reconstruction<br><b>Knee:</b> Medial Collateral<br>Ligament Repair, Lateral<br>Collateral Ligament Repair,<br>Patellar Tendon Repair,<br>Posterior Oblique Ligament<br>Repair, Iliotibial Band<br>Tenodesis, Joint Capsule<br>Closure<br><b>Hip:</b> Capsular repair,<br>Acetabular labral repair,<br>Gluteal Tendon Repair. | Stabilization, Achilles<br>Tendon Repair, Metatarsal<br>Ligament Repair, Hallux<br>Valgus reconstruction,<br>digital tendon transfers,<br>Mid-foot reconstruction.<br>3. Knee: Medial Collateral<br>Ligament Repair, Lateral<br>Collateral Ligament<br>Repair, Patellar Tendon<br>Repair, Posterior Oblique<br>Ligament Repair, Iliotibial<br>Band Tenodesis, Joint<br>Capsule Closure.<br>4. Hand/Wrist: Scapholunate<br>Ligament Reconstruction,<br>Repair/Reconstruction of<br>collateral ligaments, Repair<br>of Flexor and Extensor<br>Tendons at the PIP, DIP<br>and MCP joints for all<br>digits, digital tendon<br>transfers, Carpal Ligament<br>Reconstruction and<br>Carpometacarpal joint<br>arthroplasty (basal thumb<br>joint arthroplasty).<br>5. Elbow: Biceps tendon<br>reattachment, Ulnar or<br>Radial collateral ligament<br>reconstruction.<br>6. Hip: Acetabular labral<br>repair, Capsular repair,<br>Gluteal Tendon Repair. | |
| 6. | Anchor Loop<br>Material | Hollow braid of polyester | UHMWPE | SE Analysis 1 |
| 7. | Suture<br>&<br>Tape Material | UHWMPE or a polyblend of<br>UHMWPE and polyester | Suture- UHMWPE USP #2,<br>Tape- UHMWPE 1.5mm flat | Similar |
| 8. | Design | Arthrex Suture Anchor are<br>designed to repair soft tissue<br>to bone through a variety of<br>innovative anchor styles,<br>materials, and suture<br>configurations. As the next | STATIV® Knotted<br>UHMWPE Suture Anchor<br>constructed of strong Ultra<br>High Molecular Weight<br>Polyethylene (UHMWPE)<br>suture material. STATIV® | Similar |
| Healthium Medtech Limited | |
|------------------------------------------------------------------------------------------------------------------|--|
| Registered Office: 472/D, 13th Cross, 4th Phase, Peenya Industrial Area, Bengaluru - 560058, India Page 4 of 7 | |
| www.healthiummedtech.com CIN : U03311KA1992PLC013831 | |
{7}------------------------------------------------
Image /page/7/Picture/0 description: The image shows the logo for Sironix Arthroscopy Solutions. The word "SIRONIX" is in bold, with the "O" in "SIRONIX" being a red circle. Underneath "SIRONIX" is the phrase "Arthroscopy Solutions" in a smaller, less bold font. The logo is simple and professional.
Image /page/7/Picture/2 description: The image contains the text "K223889" in a bold, sans-serif font. Below the text is the word "Healthium" in a stylized, blue font. To the right of the text is a logo consisting of two overlapping circles, one orange and one blue, with a white cross shape in the center where they intersect.
| S.<br>No | Parameters | Arthrex FiberTak Suture<br>Anchor (K203268) | STATIV® Knotted<br>UHMWPE Suture Anchor<br>(Subject device) | Comments |
|----------|-------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------|
| | | evolution of knotless rotator<br>cuff repair, the FiberTak<br>SpeedBridge technique is<br>completed with FiberTak soft<br>Anchor on the medial row and<br>FiberTape® sutures fixated<br>laterally with trusted<br>SwiveLock® Anchor. This<br>knotless repair creates a<br>quick, secure construct in as<br>few as three suture passing<br>steps. | Knotted UHMWPE Suture<br>Anchor is the Suture Anchor<br>with sutures(s)/tape(s)<br>combination. It assured<br>centralized deployment of the<br>anchor into pilot hole. Post<br>deployment, the unique tri-<br>pod bunched up pattern<br>doubles its size, which gives<br>much higher pull out strength. | |
| 9. | Specifications<br>and<br>Dimensions | FiberTak Suture Anchor with #2<br>FiberWire CL (White/Blue) --<br>AR-3600<br>FiberTak Suture Anchor, double<br>loaded with two #2 FiberWire<br>CL (White/Blue, White/Black),-<br>AR-3600-2<br>FiberTak with 1.3 mm<br>SutureTape (White/Blue), - AR-<br>3602<br>FiberTak Suture Anchor, Double<br>Loaded with two 1.3 mm<br>SutureTape (White/Blue,<br>White/Black),- AR-3602-2<br>FiberTak with #2 TigerTail<br>(Blue/Black) -AR-3603<br>FiberTak Suture Anchor, Double<br>Loaded with two #2 TigerTail<br>(Blue/Black and White/Black)-<br>AR-3603-2<br>2.6 FiberTak Soft Anchor,<br>Double Loaded with two #2<br>FiberWire CL (Blue/White,<br>Black/White) -AR-3630-1<br>2.6 FiberTak Soft Anchor, Triple<br>Loaded with three #2 FiberWire<br>CL (Blue/White, Black/White,<br>White)- AR-3631-1<br>2.6 FiberTak Soft Anchor,<br>Double Loaded with two 1.3 mm | 1.5mm Single Suture<br>1.8mm Single Suture<br>1.5mm Single Tape<br>1.8mm Single Tape<br>1.8mm Double Sutures<br>1.8mm Double Tapes<br>2.5 mm Double Sutures (Self-<br>Tap)<br>2.5 mm Double Tapes (Self-Tap)<br>2.5 mm Triple Sutures (Self-Tap)<br>2.9 mm Double Sutures (Self-<br>Tap)<br>2.9 mm Double Tapes (Self-Tap)<br>2.9 mm Triple Sutures (Self-Tap) | SE Analysis 2 |
| Healthium Medtech Limited | |
|--------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------|
| Registered Office: 472/D, 13th Cross, 4th Phase, Peenya Industrial Area, Bengaluru – 560058, India<br>www.healthiummedtech.com CIN : U03311KA1992PLC013831 | Page 5 of 7 |
{8}------------------------------------------------
Image /page/8/Picture/0 description: The image shows the logo for Sironix Arthroscopy Solutions. The word "SIRONIX" is written in a bold, sans-serif font, with the "O" in "SIRONIX" being a red circle. Below "SIRONIX" is the text "Arthroscopy Solutions" in a smaller, sans-serif font.
Image /page/8/Picture/2 description: The image contains the text "K223889" in black font at the top left. Below the text is the word "Healthium" in a blue, stylized font. To the right of the text is a logo consisting of two overlapping circles, one orange and one blue.
| S.<br>No | Parameters | Arthrex FiberTak Suture<br>Anchor (K203268) | STATIV® Knotted<br>UHMWPE Suture Anchor<br>(Subject device) | Comments |
|----------|---------------------|--------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------|
| | | SutureTape (White/Blue and<br>White/Black) - AR-3632 | | |
| | | 2.6 FiberTak Soft Anchor, Triple<br>Loaded with three 1.3 mm<br>SutureTape (White/Blue,<br>White/Black and White)- AR-<br>3633 | | |
| 10. | Single<br>Use/Reuse | Single Use | Single Use | Similar |
| 11. | Shelf Life | 5 years | 5 Years | Similar |
| 12. | Sterilization | Provided in Sterile conditions<br>(EO Sterilization). | Provided in Sterile conditions<br>(EO Sterilization). | Similar |
| 13. | Performance<br>Data | Cyclic pull-out test,<br>Bacterial Endotoxin | Bench Performance:<br>Pull out Strength<br>Cyclical Loading<br>Bacterial Endotoxin | Similar |
| 14. | Safety Data | No Data Available | Biocompatibility:<br>Skin Sensitization,<br>Intracutaneous reactivity,<br>Material Mediated<br>Pyrogenicity,<br>Acute Systemic Toxicity,<br>In vitro cytotoxicity. | - |
# SE Analysis 1:
# Anchor Material:
Difference in Anchor Material does not affect the safety and efficacy of subject device based on the reference comparative study. Predicate device (K203268) has an anchor loop made up of Hollow braid of polyester, whereas the subject device has an anchor made up of UHMWPE.
From the literature study4 it can be inferred that the Polyester suture had lower ultimate load than all groups of sutures used in the study except the suture composed of polyester and UHMWPE (P<.05). Pure UHMWPE suture had higher ultimate failure load than sutures composed of either polyester or polyester plus UHMWPE (P<.05). Predominant failure mode was suture cutting through the meniscus for the groups except for polyester suture which failed by suture rupture.
This literature study shows that UHMWPE material has better ultimate load than the material used in predicate and this change is not raising any questions on the safety and efficacy of the subject device.
| Healthium Medtech Limited | |
|--------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------|
| Registered Office: 472/D, 13th Cross, 4th Phase, Peenya Industrial Area, Bengaluru – 560058, India<br>www.healthiummedtech.com CIN : U03311KA1992PLC013831 | Page 6 of 7 |
{9}------------------------------------------------
Image /page/9/Picture/0 description: The image shows the logo for Sironix Arthroscopy Solutions. The word "SIRONIX" is written in a sans-serif font, with the "O" in "SIRONIX" being a red circle. Below "SIRONIX" is the phrase "Arthroscopy Solutions" in a smaller font.
Image /page/9/Picture/2 description: The image contains the alphanumeric string "K223889" in black font. Below this string is the word "Healthium" in a bold, blue font. To the right of the word "Healthium" is a logo consisting of a blue circle and an orange circle, with a white cross shape overlaying them.
Reference 1: The influence of suture material on the strength of horizontal mattress suture configuration for meniscus repair.
1 https://pubmed.ncbi.nlm.nih.gov/23340094/
# SE Analysis 2:
# Specification and Dimensions:
During the bench testing of subject device, it was observed that the changes in dimensions are not leading to any kind of new risks with the performance of subject device, and the acceptance criteria was met in each of the performance tests as per defined guidelines and applicable standards. By considering the performance data of the subject device, it is evident that the changes in dimensions are not causing and or does not affect the safety and efficacy of the product when comparing it with Predicate device. The different sizes of the subject device are determined through the diameter of the dispenser which is not implantable in nature, however, there is no change in the size of the implant, so the variation in sizes of the dispenser will not lead to any kind of new risks related to safety and performance.
#### 1.7. Summary of Performance Data
The pull-out strength and displacement under cyclic loading of STATIV® Knotted UHMWPE Suture Anchor and Predicate device. The results were reviewed and side by side comparisons were done with the identified predicate device and it demonstrated that there were no significant differences between the STATIV® Knotted UHMWPE Suture Anchor and the predicate device.
#### 1.8. Clinical Testing
Not Applicable
#### 1.9. Conclusion
There are no significant differences between the subject device and the predicate devices that would adversely affect the use of the product. The subject device is substantially equivalent to the predicate device (Arthrex FiberTak Suture Anchor) in design, Intended use, Indications for use, function, Sterilization method, Shelf Life, and operational principles. From the data available we can justify that the STATIV® Knotted UHMWPE Suture Anchor is as safe, and as effective and performs the same indications for use as that of already marketed predicate device identified in 1.3 of 510(k) summary.
| Healthium Medtech Limited | |
|----------------------------------------------------------------------------------------------------|-------------|
| Registered Office: 472/D, 13th Cross, 4th Phase, Peenya Industrial Area, Bengaluru – 560058, India | |
| www.healthiummedtech.com CIN : U03311KA1992PLC013831 | Page 7 of 7 |
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.