K223884 · Elidah, Inc. · QAJ · Feb 24, 2023 · Gastroenterology, Urology
Device Facts
Record ID
K223884
Device Name
ELITONE Urge Urinary Incontinence Device
Applicant
Elidah, Inc.
Product Code
QAJ · Gastroenterology, Urology
Decision Date
Feb 24, 2023
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 876.5330
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
ELITONE URGE is a non-implanted muscle stimulator designed to treat urge urinary incontinence in women. It applies stimulation to the pelvic floor muscles and surrounding tissues.
Device Story
ELITONE Urge is a wearable, battery-powered muscle stimulator for treating urge urinary incontinence in women. The device consists of a control unit (55mm x 45mm x 10mm) and a disposable, flexible cutaneous electrode applied to the perineal region. The electrode uses mildly adhesive, electrically conductive hydrogel to transfer stimulation to pelvic floor muscles. The user operates the device via two pushbuttons to adjust intensity (0-35 levels, 0-15mA RMS) and control operation (start/pause/stop). The device delivers a symmetric, amplitude-modulated, 10 Hz sinusoidal waveform in 20-minute sessions, recommended 4-5 times per week. The 10 Hz frequency is intended to calm detrusor muscle activity. The device is used at home by the patient. The healthcare provider prescribes the device; the patient self-administers treatment. The output helps reduce urgency symptoms and urge leaks, potentially improving patient quality of life.
Clinical Evidence
Clinical study was a single-arm, prospective study (n=34) of women with urge urinary incontinence. Subjects self-administered treatment for 6 weeks. Primary endpoint: percent change in urge leaks per day. Results: 76% of participants were responders (≥50% reduction in urge leaks). Secondary endpoints: 76% responder rate for urgency episodes, 91% for I-QoL, and 32% for pad use. No adverse events resulting in injury or requiring medical treatment were reported. Supporting evidence included a clinical literature review on perineal-applied EMS for female urinary incontinence.
Technological Characteristics
Non-implanted electrical muscle stimulator. Materials: ABS (controller housing), polyethylene, hydrogel, PET (electrode). Energy: Battery-powered. Waveform: Symmetric, amplitude-modulated, 10 Hz sinusoidal. Dimensions: 55mm x 45mm x 10mm. Connectivity: Standalone. Sterilization: Non-sterile. Software: Embedded, moderate level of concern. Standards: IEC 60601-1, IEC 60601-2-10, IEC 60601-1-2, ISO 10993-1.
Indications for Use
Indicated for women with urge urinary incontinence. Designed to treat urge urinary incontinence by applying electrical stimulation to pelvic floor muscles and surrounding tissues.
Regulatory Classification
Identification
A transcutaneous electrical continence device consists of cutaneous electrodes that are used to apply external stimulation to reduce urinary incontinence.
Special Controls
In combination with the general controls of the FD&C Act, the transcutaneous electrical continence device is subject to the following special controls:
*Classification.* Class II (special controls). The special controls for this device are:(1) Non-clinical performance testing must characterize the electrical stimulation, including the following: Waveforms, output modes, maximum output voltage, maximum output current, pulse duration, frequency, net charge per pulse, maximum phase charge at 500 ohms, maximum current density, maximum average current, and maximum average power density.
(2) The patient-contacting materials must be demonstrated to be biocompatible.
(3) Performance data must demonstrate the electromagnetic compatibility (EMC), electrical safety, thermal safety, and mechanical safety of the device.
(4) Software verification, validation, and hazard analysis must be performed.
(5) Labeling must include the following:
(i) Instructions for use, including specific instructions regarding the proper placement of electrodes;
(ii) A summary of electrical stimulation parameters; and
(iii) Cleaning instructions and reuse information.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo features the letters 'FDA' in a blue square, followed by the words 'U.S. FOOD & DRUG ADMINISTRATION' in blue text.
February 24, 2023
Elidah, Inc. Gloria Kolb CEO 810 Main St., Suite C Monroe, CT 06468
Re: K223884
Trade/Device Name: ELITONE Urge Urinary Incontinence Device Regulation Number: 21 CFR§ 876.5330 Regulation Name: Transcutaneous Electrical Continence Device Regulatory Class: II Product Code: OAJ Dated: December 27, 2022 Received: December 27, 2022
Dear Gloria Kolb:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies.
U.S. Food & Drug Administration 10903 New Hampshire Avenue Silver Spring, MD 20993 www.fda.gov
{1}------------------------------------------------
You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatoryinformation/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Negeen Haghighi -S
for
Jessica K. Nguyen, Ph.D. Assistant Director DHT3B: Division of Reproductive. Gynecology and Urology Devices OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K223884
Device Name
ELITONE Urge Urinary Incontinence Device
#### Indications for Use (Describe)
ELITONE Urge is a non-implanted muscle stimulator designed to treat urge urinary incontinence in women. It applies stimulation to the pelvic floor muscles and surrounding tissues.
| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|
| <span></span> Prescription Use (Part 21 CFR 801 Subpart D) | <span></span> Over-The-Counter Use (21 CFR 801 Subpart C) |
|------------------------------------------------------------|-----------------------------------------------------------|
|------------------------------------------------------------|-----------------------------------------------------------|
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary
| Sponsor: | Elidah, Inc.<br>810 Main St, Suite C.<br>Monroe, CT 06468<br>Contact: Gloria Kolb<br>Phone: 978-435-4324<br>Date Prepared: 2/20/23 |
|---------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Device Information: | Trade Name: ELITONE Urge Urinary Incontinence Device<br>Common Name: Non-implanted electrical continence device<br>Classification Name: Transcutaneous electrical continence device<br>Classification Product Code: QAJ<br>Regulation Number: 876.5330<br>Device Classification: Class II |
| Predicate Device: | ELITONE (K183585), manufactured by Elidah, Inc.<br>The predicate device has not been subject to a design-related recall. |
# Device Description:
The ELITONE Urge Urinary Incontinence Device (hereafter ELITONE URGE or ELITONE UUI) provides electrical muscle stimulation (commonly called EMS or E-stim) to treat urge urinary incontinence. Stimulation is delivered to the pelvic floor muscles and surrounding structures through a disposable cutaneous electrode applied to the perineal region. One end of the thin, flexible electrode is positioned proximate the pubis and the other end is positioned proximate the ischial tuberosity. Mildly adhesive and electrically conductive hydrogel portions of each electrode to the perineal tissues and allow comfortable transfer of electrical stimulation through the pelvic floor. The stimulation is controlled by a battery powered, wearable control unit. The housing of the control unit is approximately 55mm x 45mm x 10mm. It includes two pushbuttons that allow the user to increment the voltage (i.e., intensity), and to start, pause and stop device operation. The stimulation utilizes a symmetric, amplitude-modulated, sinusoidal waveform. The output intensity operates over a range of levels (0-35) that approximately correspond to a 0-15mA RMS current. The stimulation frequency is 10 Hz, a frequency typically used to calm detrusor muscle activity, decreasing symptoms of urgency type incontinence. Each treatment session lasts 20 minutes, after which the device turns off automatically. Women are recommended to use the device 4-5 times per week.
# Intended Use:
ELITONE URGE is a non-implanted muscle stimulator designed to treat urge urinary incontinence in women. It applies stimulation to the pelvic floor muscles and surrounding tissues.
#### Comparison to the Predicate Device:
The table below provides comparisons of the proposed ELITONE UUI device to the predicate ELITONE device (hereafter ELITONE SUI for clarity). The difference between ELITONE UUI and the predicate is the
{4}------------------------------------------------
composition of the output signal. Each stimulation cycle of the predicate's output combines a 50Hz signal (understood to treat stress urinary incontinence) and a 10Hz signal (understood to treat urge urinary incontinence), whereas the proposed ELITONE UUI output is solely 10Hz. This difference is achieved through a modification to the embedded software. Product labeling has been changed to reflect the new Indication for Use. All other elements of the two products are identical. Differences in technological characteristics do not raise different questions of safety and effectiveness.
| Technological<br>Characteristic | ELITONE UUI Device<br>(New Device) | ELITONE SUI Device<br>(K183585) | Comparison and Impact<br>on Safety/Performance |
|---------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------|
| Indication for Use | non-implanted muscle<br>stimulator designed to treat<br>urge urinary incontinence in<br>women. It applies stimulation<br>to the pelvic floor muscles<br>and surrounding tissues. | non-implanted muscle<br>stimulator designed to treat<br>stress urinary incontinence<br>in women. It applies<br>stimulation to the pelvic floor<br>muscles and surrounding<br>tissues. | Substantially Equivalent.<br>Both have an intended<br>use for the treatment of<br>urinary incontinence in<br>women. |
| Therapeutic modality | Electrical muscle stimulation | Electrical muscle stimulation | Identical |
| Rx or OTX | Rx and OTC | Rx and OTC | Identical |
| Targeted tissue | Pelvic floor muscles and<br>surrounding structures | Pelvic floor muscles and<br>surrounding structures | Identical |
| Anatomic site of<br>stimulation application | Perineal region | Perineal region | Identical |
| Device Materials | Controller Housing - ABS<br>Electrode - Polyethylene,<br>Hydrogel, PET | Controller Housing - ABS<br>Electrode - Polyethylene,<br>Hydrogel, PET | Identical |
| Sterility Status | Non-sterile | Non-sterile | Identical |
| Shelf-Life | 2-Year shelf life on GelPads | 2-Year shelf life on GelPads | Identical |
| Number of output<br>modes | 1 | 1 | Identical |
| Number of output<br>channels | 1 | 1 | Identical |
| Controls | 2 buttons provide all<br>functionality | 2 buttons provide all<br>functionality | Identical |
| Compliance with<br>voluntary standards | IEC 60601-1<br>IEC 60601-2-10<br>IEC 60606-1-2 | IEC 60601-1<br>IEC 60601-2-10<br>IEC 60606-1-2 | Identical |
| Waveform | Modulated, symmetric<br>(250µs/phase). | Modulated, symmetric<br>(250µs/phase). | Identical |
| Stimulation Frequency | 10 Hz<br>Amplitude modulated at<br>2000Hz | 50 Hz, 10 Hz<br>Amplitude modulated at<br>2000Hz | Substantially equivalent:<br>Both utilize frequencies<br>known to treat UUI (10-15<br>Hz). |
| Time (Cycle<br>Duration?) | 6s at 10 Hz<br>6s no stimulation | 4s at 50 Hz<br>2s at 10 Hz<br>6s no stimulation | Substantially equivalent:<br>Both utilize contract/relax<br>sequencing consistent<br>with known treatments for<br>incontinence. |
| Max output voltage<br>(500Ω) | Vrms = 8V<br>Vpp = 43V | Vrms = 15V at 50Hz, 8V at<br>10Hz<br>Vpp = 43V | Identical to 10Hz portion<br>of predicate |
| Technological<br>Characteristic | ELITONE UUI Device<br>(New Device) | ELITONE SUI Device<br>(K183585) | Comparison and Impact<br>on Safety/Performance |
| Max output current<br>(500Ω) | Ipp — 97mA<br>Irms -11.4mA | Ipp — 97mA<br>Irms -28.2mA at 50Hz,<br>11.4mA at 10 Hz | Identical to 10Hz portion<br>of predicate |
| Max current density | RMS: 0.4 mA/cm²<br>Peak: 3.5 mA/cm² | RMS: 1.0 mA/cm² at 50Hz,<br>0.4mA/cm² at 10Hz<br>Peak: 3.5 mA/cm² | Identical to 10Hz portion<br>of predicate |
| Maximum Power<br>Density (W/cm²) | 0.0032 W/cm² | 0.015 W/cm² at 50Hz,<br>0.0032 W/cm² at 10Hz | Identical to 10Hz portion<br>of predicate |
| Maximum Phase<br>Charge (µC @<br>500Ohms) | 9.09 µC | 9.09 µC | Identical |
{5}------------------------------------------------
# Testing Summary:
The following performance testing was provided in support a substantial equivalence determination.
- Software Verification and Validation Testing - The software for the device was considered to be a "moderate" level of concern. Verification and validation testing was conducted in accordance with the FDA guidance document "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices" issued May 11, 2005.
- . Electrical Safety and Electromagnetic Compatibility (EMC) Testing - Testing was completed by external labs per IEC 60601-1 and IEC 60601-2-10 standards for safety and IEC 60601-1-2 standard for EMC.
- Other Performance Testing - Performance of the electrode component was assessed by measuring its impedance, its adhesiveness and its current dispersion. Comparisons were made to legally marketed EMS electrodes.
- . Biocompatibility Assessment – A biocompatibility evaluation was completed in accordance with the ISO 10993-1 standard. The electrode is considered tissue contacting for a duration of less than 24 hours. The materials, manufacturing processes and use application are the same as those of legally marketed electrodes. Accordingly, no new biocompatibility testing was needed to demonstrate safety and efficacy.
- . Clinical Literature Evaluation – A literature review was conducted to demonstrate the safety and efficacy of perineal-applied EMS in the treatment of female urinary incontinence. Articles were reviewed to assess their relevance to the performance and safety of the device.
- Clinical Testing – Clinical testing was conducted on women with urge urinary incontinence in the US. The study design was a single-arm study with the patient serving as own control (i.e., comparison to baseline). Subjects self-administered treatment for 6-weeks. The pre-specified primary efficacy endpoint was the percent change in urge leaks per day, determined by comparing 7-day baseline data with data from the 6th week of treatment. Responders were defined as achieving ≥50% reduction in urge leaks. Pre-specified secondary efficacy endpoints included: changes in urgency episodes (i.e., combined number of urge leaks and urgency events that did not result in a leak) with responders defined as achieving ≥50% reduction, changes in pad use with responders defined as achieving ≥50% reduction, and changes in I-QoL score with responders defined as achieving ≥2.5 points change. Subjects with urge incontinence symptoms predominant to stress incontinence symptoms (if present) were randomly assigned to treat with one of two
{6}------------------------------------------------
devices, one of which was the subject ELITONE UUI device. 15 subjects were included in analysis of the ELITONE UUI device. Clinical testing was later expanded (under a second IRB-approved protocol) to include a cohort of women with urge urinary incontinence without stress incontinence symptoms. These subjects self-administered treatment for 6-weeks with the ELITONE UUI device according to the same protocol as the first cohort. Pre-specified primary and secondary endpoints aligned with those from the first cohort of subjects. 19 subjects were included in the analysis of this second cohort. The protocol detailing the expanded data collection specifies intent to pool data from the two testing periods. Accordingly, 34 subjects were included in the final analysis. For the primary efficacy endpoint, 76% of participants responded to the treatment. Responder rates for the secondary endpoints were: 76% for urgency episodes, 91% for I-QoL, and 32% for pad use. The primary safety endpoint was defined as adverse events that resulted in an injury or required medical treatment. None were reported. The self-administered, at home treatment limited the opportunity for direct oversight by study personnel, making it relatively easy for subjects to enroll and then subsequently opt-out of participation before initiating treatment. In the first cohort, 26 subjects were shipped the ELITONE UUI device (i.e., enrollment), 15 returned the study device demonstrating at least one treatment session (i.e., treatment), and 15 were found to have completed the study protocol (i.e., analysis). Enrollment, treatment, and analysis values for the second cohort were 30, 23, and 19 respectively.
# Risk Management Summary:
ELITONE UUI has been designed according to Elidah's internal procedures with traceability between the design inputs, design outputs, verification and validation activities. ELITONE UUI has been evaluated for risks according to Elidah's internal procedures based on ISO 14971. The risks associated with ELITONE UUI were individually and collectively assessed, and the risk/benefit analysis was acceptable.
# Conclusion:
The non-clinical data demonstrate that ELITONE UUI is as safe as the predicate. The software verification and validation demonstrate that ELITONE UUI performs as intended in the specified use conditions. The clinical literature evaluation supports the use of transcutaneous electrical stimulation as an effective treatment of urge urinary incontinence in women. The clinical study data demonstrates that ELITONE UUI is as safe and effective as the predicate device. Collectively, performance testing supports a Substantial Equivalence determination.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.