The KBA3D v2.0.0 is intended for assisting healthcare professionals in viewing and measuring images as well as planning spine surgeries. The device allows surgeons and service providers to perform spine related measurements on images, and to plan surgical procedures. The device also includes tools for measuring anatomical components for design and placement of surgical implants. Clinical judgment and experience are required to properly use the software. The patient population targeted with the use of the KBA3D v2.0.0 software includes patients with mature skeletons requiring imaging measurements and planning of surgical procedures. The Bendini Rod Bending algorithm is intended for patients older than 22 years old.
Device Story
KBA3D v2.0.0 is a SaaS medical image management and processing system for spine surgery planning. It takes two perpendicular standing X-rays (spine/pelvis) as input. The software performs 3D reconstruction of femoral heads, sacral plate, and vertebral bodies; calculates shape/positioning parameters; assesses global balance; simulates surgical effects on spinopelvic alignment; and templates implants (pedicle screws, interbody cages, rods). It integrates with the NuVasive Bendini Web App (via API REST) to retrieve rod bending instructions. Used by orthopedic surgeons, neurosurgeons, and technicians in clinical settings. Output is visualized by the clinician to assist in surgical planning and implant selection. It does not predict surgical outcomes; accuracy is based on theoretical calculations. Benefits include improved visualization of anatomical components and surgical scenarios to aid clinical decision-making.
Clinical Evidence
Bench testing only. Verification included mathematical validation of anatomical parameters against Excel calculations; 3D reconstruction accuracy verified against CT-scan data (mean deviation < 4.3mm); surgical simulation verified against 12 post-op patient cases; and implant positioning algorithms (screws/cages) verified against acceptance criteria (e.g., < 0.5mm/0.5°). Formative and summative usability studies confirmed intuitive use and successful performance of measurements and reconstructions.
Technological Characteristics
SaaS software; PC-compatible; runs on Windows/MAC. Features include 3D reconstruction from 2D X-rays, osteotomy module, and implant templating. Connectivity via API REST for Bendini service. Software development follows IEC 62304, EN ISO 14971, and EN 62366-1. Moderate level of concern.
Indications for Use
Indicated for patients with mature skeletons requiring imaging measurements and surgical planning for spine procedures. The Bendini Rod Bending algorithm is indicated for patients > 22 years old.
Regulatory Classification
Identification
A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.
Special Controls
*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA text logo on the right. The FDA text logo is in blue and includes the letters "FDA" in a square and the words "U.S. Food & Drug Administration".
S.M.A.I.O. % Robert Poggie, Ph.D. President BioVera, Inc. 65 Promenade Saint Louis Notre Dame de L'Ile Perrot, Quebec J7W3J6 CANADA
#### Re: K223841
May 30, 2023
Trade/Device Name: KBA3D v2.0.0 Regulation Number: 21 CFR 892.2050 Regulation Name: Medical image management and processing system Regulatory Class: Class II Product Code: LLZ Dated: May 2, 2023 Received: May 2, 2023
Dear Dr. Poggie:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part
{1}------------------------------------------------
801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4. Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Jessica Lamb
Jessica Lamb, Ph.D. Assistant Director Imaging Software Team DHT8B: Division of Radiological Imaging Devices and Electronic Products OHT8: Office of Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2023 See PRA Statement below.
510(k) Number (if known) K223841
Device Name KBA3D v2.0.0
#### Indications for Use (Describe)
The KBA3D v2.0.0 is intended for assisting healthcare professionals in viewing and measuring images as well as planning spine surgeries. The device allows surgeons and service providers to perform spine related measurements on images, and to plan surgical procedures. The device also includes tools for measuring anatomical components for design and placement of surgical implants. Clinical judgment and experience are required to properly use the software.
The patient population targeted with the use of the KBA3D v2.0.0 software includes patients with mature skeletons requiring imaging measurements and planning of surgical procedures. The Bending algorithm is intended for patients older than 22 years old.
Type of Use (Select one or both, as applicable)X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# K223841
### 510(k) SUMMARY for K223841
In accordance with 21 CFR 807.92 of the Federal Code of Regulations, the following is a summary of safety and effectiveness of S.M.A.I.O's KBA3D V2.0.0 software.
# A. SUBMITTERS INFORMATION Submitter Name:
| Submitter Name: | BioVera, Inc. | | |
|---------------------|-----------------------------------------------------------------------------------|--|--|
| Submitter Address: | 65 Promenade Saint-Louis, Notre-Dame-de-<br>L'Ile-Perrot, Québec, J7W 3J6, CANADA | | |
| Contact Person: | Robert A. Poggie, PhD | | |
| Phone & Fax Number: | 514-901-0796 | | |
| Date of Submission: | May 2, 2023 | | |
# B. DEVICE IDENTIFICATION & MANUFACTURER
| Manufacturer Name: | S.M.A.I.O |
|-----------------------------------------|---------------------------------------------------------------------------|
| Manufacturer Address: | 2, Place Berthe Morisot- Parc Technologique<br>69800 SAINT-PRIEST, FRANCE |
| Registration Number: | 3015383864 |
| Contact Name: | Jean-Charles Roussouly |
| Title: | Operations vice-president |
| Device Trade Name: | KBA3D v2.0.0 |
| Device Common Name: | Image measurement and surgery planning<br>software |
| Classification Name: | Picture archiving and communications<br>system |
| Classification Code: | LLZ |
| Classification Panel: | Radiology Devices |
| Regulation Number: | 21 CFR sections 892.2050 |
| C1. PRIMARY PREDICATE DEVICE<br>K213975 | KEOPS Balance Analyzer 3D (KBA3D) |
| C2. PREDICATE DEVICES | |
| K180091 | UNiD Spine Analyzer, Medicrea International |
| K141669 | Nemaris Surgimap 2.0 |
| C3. REFERENCE DEVICE | |
| K210574 | NuVasive Pulse System |
| | |
{4}------------------------------------------------
# D. DEVICE DESCRIPTION
The subject device KBA3D v2.0.0 is a SaaS software solution developed by S.M.A.I.O for the medical community; it is a second, independent version of the original KBA3D cleared by FDA in K213975. The current version of the KBA3D software is v.2.0.0. The user needs and requirements of the subject device were jointly defined by S.M.A.I.O and NuVasive. The software is intended to view images, perform spine related measurements, and plan surgical procedures such as osteotomies of the spine and templating of implants (screws, cages, rods).
KBA3D V2.0.0 software can be used by health professionals (orthopedic surgeons, radiologists, neurosurgeons) and service providers (imaging technicians, clinical study technicians) who are trained in spine imaging and pathologies. The KBA3D V2.0.0 software is intended for patients requiring measurements and planning of surgical procedures. KBA3D v2.0.0 aims to achieve three objectives:
- 1. From two perpendicular patient's standing x-rays including patient's spine and pelvis from the femoral heads to the cervical levels, provide 3D scaled representation of the femoral heads, sacral plate, and vertebral bodies. Provide related shape and positioning parameters measurements (disc/vertebra/height/angulation), main curvatures description and global balance assessment.
- 2. Simulate potential effects of a spine surgery on spinopelvic alignment and provide related shape and positioning parameters calculation.
- 3. Visualize scaled representation of implant range (pedicle screws, interbody cages, union rods) relative to spinopelvic representation (pre-op versus realigned) to establish possible implant selection scenarios.
The software is not intended to predict the results of surgery as S.M.A.I.O does not provide tools to carry out planning. Therefore, regarding implant sizing, positioning, and correction impacts, accuracy levels provided by S.M.A.I.O are solely based on theoretical calculations that are not correlated to the output of surgery. KBA3D V2.0.0 provides scaled and simplified representations of the screws and cages relative to the patient's spine and pelvis.
Bendini service (part of NuVasive Pulse System, K210574, reference device):
- The first request from the software to the Bendini system enables the software to retrieve the most up-to-date benders/rods configuration file.
- A second request from the software to the Bendini system enables the software to obtain specific bending instructions for the rod.
Note: The Bendini Rod Bending algorithm is intended for patients older than 22 years old. The NuVasive Bendini Web App service is comprised of an API REST web service that facilities outside clients to utilize the NuVasive Bending algorithm. This algorithm is identical to the Bendini algorithm in K210574 for the NuVasive Pulse System and is hosted using the Microsoft Azure App Service. The KEOPS database manages connection to the software, patient data, and storage of data (X-rays and simulations).
{5}------------------------------------------------
# E. INTENDED USE
The KBA3D v2.0.0 is intended for assisting healthcare professionals in viewing and measuring images as well as planning spine surgeries. The device allows surgeons and service providers to perform spine related measurements on images, and to plan surgical procedures. The device also includes tools for measuring anatomical components for design and placement of surgical implants. Clinical judgment and experience are required to properly use the software.
The patient population targeted with the use of the KBA3D v2.0.0 software includes patients with mature skeletons requiring imaging measurements and planning of surgical procedures. The Bendini Rod Bending algorithm is intended for patients older than 22 years old.
# F. TECHNOLOGICAL CHARACTERISTICS AND SUBSTANTIAL EQUIVALENCE
The table below compares the technological characteristics of the KBA3D V2.0.0 to the predicate and reference devices.
| | Subject Device | Predicate Device<br>(Primary) | Predicate<br>Device | Predicate<br>Device | Reference<br>Device |
|--------------------------------------------------------------|-------------------------|-----------------------------------------|------------------------------------|--------------------------------------|-------------------------------------|
| Feature | KBA3D V2.0.0<br>K223841 | KEOPS Balance<br>Analyzer 3D<br>K213975 | Nemaris<br>Surgimap 2.0<br>K141669 | Medicrea<br>UNiD Analyzer<br>K180091 | NuVasive Pulse<br>System<br>K210574 |
| Computer | PC Compatible | PC Compatible | PC Compatible | PC Compatible | PC Compatible |
| Operating System | Windows + MAC | Windows + MAC | Windows + MAC | Windows + MAC | Windows |
| Image Input | Local | Local | Local + PACS<br>connectivity | Local | Local |
| Runs on Server | Yes | Yes | No | Yes | No |
| Osteotomy Module | Yes | Yes | Yes | Yes | No |
| Spine measurements | Yes | Yes | Yes | Yes | Yes |
| Pre-op planning | Yes | Yes | Yes | Yes | No |
| Screws | Yes | No | No | Yes | Yes |
| Cages | Yes | No | Yes | No | No |
| Rods | Yes | Yes | Yes | Yes | Yes (Bendini) |
| Database | No | No | Yes | No | No |
| Case sharing | No | No | Yes | No | No |
| Human intervention for image interpretation and manipulation | Yes | Yes | Yes | Yes | Yes |
# G. PERFORMANCE SUMMARY
The performance characteristics of the KBA3D V2.0.0 were tested and analyzed per the FDA's quidance for Industry and FDA Staff, "Guidance for the Content of Premarket Submissions for Software Contained on Medical Devices" and the FDA quidance document entitled "Content of Premarket Submissions for Management of Cybersecurity in Medical Devices". The pre-clinical tests and results for the subject device included:
- · Anatomical parameters measurements were verified via mathematical calculations. The values were compared to the parameters calculated by the KBA3D V2.0.0 software. The results showed
{6}------------------------------------------------
there was no difference in the calculations by Excel and KBA3D V2.0.0 software.
- · 3D algorithm reconstruction on 2D Images were verified by comparison of Vertebral body dimensions from: (1) 3D reconstructions using the Balance Analyzer and 3D reconstructions from CT-Scan. The results showed that most of the deviations were less than 3mm and that the average was a maximum of 4.3mm (less than the maximum validation criterion of 5mm).
- · Uncertainty testing studies comparing original coordinates and "worst case variation" coordinates demonstrated acceptable variations of values and validated the accuracy of the software. A deviation in distance of less than 2 pixels (px) and a deviation of not more than 2° were considered acceptable.
- Surgical Simulation was verified by applying the post treatment radiographs with the simulation algorithm in comparison with images (postop, 12 patients) by a spine surgeon. The highest value of the SSA error was 3.26% and the set of back types obtained by simulation was identical to the post-op version.
- · Performance of the algorithm to position the screws was verified. A deviation < 10-12 mm was observed for the positioning of the entry point of the screws (less than the acceptance criterion of 0.5 mm) and a deviation < 0.3° was observed for the angulation of each screw in the vertebra ((less than the acceptance criterion of 0.5°)
- · Performance of the algorithm to position the cage was verified. A deviation < 0.4mm offset was observed between the height of the café and the disc space (less than the acceptance criterion of 0.5 mm) and a deviation < 10-12° was observed for the angulation between the two vertebrae and the angulation given by the cage (less than the acceptance criterion of 0.5°)
- · Formative and summative studies demonstrated the safety and usability intuitiveness of the KBA3D V2.0.0 software. Intuitive use was confirmed and the users successfully performed measurements and 3D reconstruction that simulated surgery.
Software validation and performance testing of the KBA3D V2.0.0 consisted of verification and validation activities using the following guidelines and standards throughout the software development process:
| EN ISO 14971:2019 | Application of risk management to medical devices |
|--------------------------------------------|---------------------------------------------------------------------------------------------------------------|
| IEC 62304:2006 | Medical device software - Software life cycle processes |
| EN 62366-1:2015/A1:2020 | Medical Device Part I – Application of the Usability<br>Engineering to Medical Devices |
| IEC 82304-1:2016 | Health Software – Part I General Requirements For<br>Products Safety |
| FDA Guidance for Industry<br>and FDA Staff | Guidance for the content of premarket submissions for<br>software contained in medical devices (May 11, 2005) |
The software for the KBA3D V2.0.0 is of Moderate level of concern.
### H. CONCLUSIONS
The KBA3D v2.0.0 is substantially equivalent to the primary predicate device KEOPS Balance Analyzer 3D (K213975) which we relied upon in determining substantial equivalence for the subject device for similarities in indications for use, design, technological characteristics, performance data, and user interface presented in this 510(k) notification.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.