K223838 · Amo Manufacturing USA, LLC · OOE · Apr 12, 2023 · Ophthalmic
Device Facts
Record ID
K223838
Device Name
Catalys™ Precision Laser System
Applicant
Amo Manufacturing USA, LLC
Product Code
OOE · Ophthalmic
Decision Date
Apr 12, 2023
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 886.4390
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Catalys™ Precision Laser System is indicated for use in patients undergoing cataract surgery for removal of the crystalline lens. Intended uses in cataract surgery include anterior capsulotomy, phacofragmentation, and the creation of single plane and multi-plane arc cuts/incisions in the cornea, each of which may be performed either individually or consecutively during the same procedure.
Device Story
Ophthalmic surgical laser system used in hospitals, ASCs, and surgeon offices. Employs femtosecond laser technology with integrated Spectral Domain Optical Coherence Tomography (OCT) for real-time imaging. System uses common optical path for OCT and treatment laser, ensuring co-registered imaging and treatment beams. Surgeon operates system via graphical user interface to define incisions based on OCT-visualized ocular anatomy (anterior chamber, lens, cornea). Laser pulses create localized plasma and cavitation to perform capsulotomy, lens fragmentation, and corneal incisions. Software updates include advanced astigmatism management, Stevens nomogram, Toric Mark Axis Calculator, and expanded pre-op data import capabilities (e.g., IOLMaster 700). System facilitates efficient lens removal and precise corneal incisions, potentially improving surgical outcomes.
Clinical Evidence
No clinical data. Substantial equivalence supported by bench testing, software verification/validation, and regression testing.
Technological Characteristics
Class 4 femtosecond laser (1030 nm wavelength); Spectral Domain OCT (820-930 nm source); pulse duration <600 fs; max pulse energy 10 µJ; repetition rate 9-120 kHz. Includes Liquid Optics™ Interface (single-use, sterile). Software-controlled via microprocessor. Connectivity includes pre-op exam import. Complies with 21 CFR 1040.10/1040.11 and ISO 15004-2:2007.
Indications for Use
Indicated for patients undergoing cataract surgery for removal of the crystalline lens, including anterior capsulotomy, phacofragmentation, and corneal arc cuts/incisions.
Regulatory Classification
Identification
An ophthalmic laser is an AC-powered device intended to coagulate or cut tissue of the eye, orbit, or surrounding skin by a laser beam.
{1}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the seal of the Department of Health & Human Services. To the right of the seal is the FDA logo, which consists of the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG" in blue, and the word "ADMINISTRATION" in a smaller font size below.
April 12, 2023
AMO Manufacturing USA, LLC Laarni Ricafort Manager, Regulatory Affairs 510 Cottonwood Drive Milpitas, CA 95035
Re: K223838
Trade/Device Name: Catalys™ Precision Laser System Regulation Number: 21 CFR 886.4390 Regulation Name: Ophthalmic Laser Regulatory Class: Class II Product Code: OOE Dated: March 10, 2023 Received: March 14, 2023
## Dear Laarni Ricafort:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling
{2}------------------------------------------------
(21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safetyreporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act): 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-devicereporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
J Angelo Green -S
Image /page/1/Picture/6 description: The image contains the letters '-S' on the left side of the image. To the right of the letters is the acronym 'FDA' in a light blue color. The letters are large and take up a significant portion of the image.
for Tieuvi Nguyen, Ph.D. Director DHT1A: Division of Ophthalmic Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
## Indications for Use
510(k) Number (if known) K223838
Device Name CATALYSTM Precision Laser System
#### Indications for Use (Describe)
The CATALYS™ Precision Laser System is indicated for use in patients undergoing cataract surgery for removal of the crystalline lens. Intended uses in cataract surgery include anterior capsulotomy, phacofragmentation, and the creation of single plane and multi-plane arc cuts/ incisions in the cornea, each of which may be performed either individually or consecutively during the same procedure.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------------------------------------------|------------------------------------------------------------------------------------|
| <span style="font-size:100%;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) | <span style="font-size:100%;">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
# 510(k) SUMMARY
The following 510(k) Summary is being submitted in accordance with the requirements of 21 CFR 807.92
| 510(k) Summary: | K223838 |
|-----------------|---------|
|-----------------|---------|
[807.92(a)(1)] Submitter Information
| Sponsor/Submitter: | AMO Manufacturing USA, LLC<br>510 Cottonwood Drive<br>Milpitas, CA 95035 USA<br>Phone 408-273-5166<br>Fax:408 273-5966 |
|------------------------|----------------------------------------------------------------------------------------------------------------------------|
| Contact Person: | Laarni Ricafort<br>Manager, Regulatory Affairs<br>Email: lricafor@its.jnj.com<br>Phone: 408-273-4016<br>Fax: :408 273-5966 |
| Date Summary Prepared: | April 10, 2023 |
[807.92(a)(2)] Name of Device
| Device Trade Name: | CATALYST™ Precision Laser System |
|------------------------|----------------------------------|
| Common Name: | Opthalmic laser |
| Device Classification: | Class II |
| Regulation Number: | 21 CFR 886.4390 |
| Classification Name: | Opthalmic Femtosecond Laser |
| Product Code: | OOE |
[807.92(a)(3)] Legally Marketed Devices
| Predicate Device: | CATALYST™ Precision Laser System<br>K200056; cleared May 18, 2020 |
|-------------------|-------------------------------------------------------------------|
|-------------------|-------------------------------------------------------------------|
{5}------------------------------------------------
### [807.92(a)(4)] Device Description
Catalys™ Precision Laser System ophthalmic surgical laser system Device Description: used in healthcare facilities such as hospitals. Ambulatory Surgerv Centers (ASCs) and surgeon office settings. The System is an electromedical device which contains software. System components include a single-use Liquid Optics™ Interface and optional Mobile Patient Bed. The Catalys™ Precision Laser System (also referred to as the Catalys™ System or System) is an ophthalmic surgical laser system indicated for use in cataract surgery to create a precise anterior capsulotomy and/or to effect lens fragmentation, thus facilitating efficient lens removal. The System also creates single plane and multi-plane arc cuts/incisions in the cornea, each of which may be performed either individually or consecutively during the same procedure. The System employs femtosecond ("FS") laser technology with integrated Optical Coherence Tomography ("OCT"), all of which are controlled and monitored by dedicated electronics. The System utilizes a common optical path for the OCT and femtosecond treatment laser (including the threedimensional scanner and Liquid Optics™ [patient] Interface). As such, the beams are intrinsically co-registered and provide for precise overlap between imaging and treatment beams. Ocular surfaces recognized by the system software include anatomy within the anterior chamber, such as the anterior and posterior corneal surfaces and the anterior and posterior surfaces of the crystalline lens. Detailed axial or sagittal cross-sectional views are available via OCT, to demarcate proposed incisions versus adjacent ocular structures (for example, iris, pupil and limbus). The Catalys™ Precision Laser System laser classification per 21 CFR 1040.10 and 1040.11 is Class 4.
### [807.92(a)(5)] Intended Use
The Catalys™ Precision Laser System is indicated for use in patients undergoing cataract surgery for removal of the crystalline lens. Intended uses in cataract surgery include anterior capsulotomy, phacofragmentation, and the creation of single plane and multi-plane arc cuts/incisions in the cornea, each of which may be performed either individually or consecutively during the same procedure.
Indications for Use:
{6}------------------------------------------------
## The Catalys™ Precision Laser System is indicated for use in patients undergoing cataract surgery for removal of the crystalline lens. Intended uses in cataract surgery include anterior Difference in capsulotomy, phacofragmentation, and the creation of single Indications from plane and multi-plane arc cuts/incisions in the cornea, each of Predicate Device which may be performed either individually or consecutively during the same procedure. The subject device and the predicate device have the same indications for use. [807.92(a)(6)] Technical Characteristics The subject Catalys Precision Laser System is unchanged in regard to its technological characteristics, indications for use, and intended uses. The software revisions for the subject device include updates to the graphical user interface and host to improve the efficiency of the workflow as well as the addition of software modules to support the import of patient exams and iris registration. The revisions in the modified device are constrained to only software changes, of which most are related to advanced astigmatism management and additional updates to currently implemented features. Technological Software verification and validation testing in addition to bench Characteristics: testing was performed to verify the ability of the software to meet its intended use and to ensure that no adverse effects have been introduced due to the software change. This testing included subsystem level verification and regression testing as well as system validation using the latest software, Mobile Patient Bed, and Liquid Optics Interface. The following table provides a comparison of the predicate device and subject device for the purpose of demonstrating substantial equivalence to the predicate device for its indication for use, intended use, technological characteristics and added new features.
#### Similarities and Differences Between Cleared Predicate Device and Subject Device
| Attribute | Predicate Device<br>Catalys Precision Laser<br>System | Subject Device<br>Catalys Precision Laser<br>System |
|--------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------|
| 510(k) Number | K200056 | TBD |
| Regulation Number | 886.4390 | Same |
| Attribute | Predicate Device<br>Catalys Precision Laser<br>System | Subject Device<br>Catalys Precision Laser<br>System |
| Regulation Name | Ophthalmic Laser | Same |
| Regulatory Class | Class II | Same |
| Product Code | OOE | Same |
| Indications for Use | The Catalys™ Precision Laser<br>System is indicated for use in<br>patients undergoing cataract<br>surgery for removal of the<br>crystalline lens. Intended uses in<br>cataract surgery include anterior<br>capsulotomy,<br>phacofragmentation, and the<br>creation of single plane and<br>multi-plane arc cuts/incisions in<br>the cornea, each of which may<br>be performed either individually<br>or consecutively during the<br>same procedure. | Same |
| System Type | Ophthalmic Femtosecond Laser<br>with Spectral Domain OCT | Same |
| Laser Mechanism of Action | Plasma, Cavitation | Same |
| Treatment Laser Wavelength<br>(nm) | $1030 \pm 5$ | Same |
| Output Power, Max | Per ISO 15004-2:2007 limits | Same |
| Maximum Pulse Energy (µJ) | 10 | Same |
| Repetition Rate (kHz) | 9-120 | Same |
| Pulse Duration (fs) | < 600 | Same |
| Spot Size; diameter (µm) | 5 | Same |
| System controls | Microprocessor with Graphical<br>User Interface | Same |
| Patient Contact Interface | Suction-ring type interface<br>devices (marketed as Liquid<br>Optics™ Interface) Sterile and<br>Single-use | Same |
| Attribute | Predicate Device<br>Catalys Precision Laser<br>System | Subject Device<br>Catalys Precision Laser<br>System |
| LOI Suction Ring Seal<br>Diameters (mm) | LOI<br>External (mm): 21.6<br>Internal (mm): 14.1 | Same |
| | LOI-12<br>External (mm):19<br>Internal (mm):12 | Same |
| | 0180-1401<br>External (mm): 21.6<br>Internal (mm): 14.1 | Same |
| | 0180-1201<br>External (mm): 19<br>Internal (mm): 12 | Same |
| OCT Axial Resolution (µm) | 30 | Same |
| OCT transverse Resolution<br>(µm) | 15 | Same |
| Scan speed (A-scans/sec) | 1000 | Same |
| A-scan depth (nm) | 2 | Same |
| Optical Source | 820-930 | Same |
| Optical Power | ANSI Class 1 < 3.48mW at<br>cornea | Same |
| Iris Imaging | Live iris view | Same |
| Trajectory Timing<br>Synchronization | FPGA coordinates from one<br>non-reentrant VI | Same |
| Communication Method for<br>Watchdog for Host PC with<br>the Mobile Patient Bed<br>Pairing | Direct FPGA Interface | Same |
| Attribute | Predicate Device<br>Catalys Precision Laser<br>System | Subject Device<br>Catalys Precision Laser<br>System |
| Software Features | Built-in Nomogram | Addition of Stevens nomogram |
| | Toric Alignment Marks | Toric Mark Axis Calculator |
| | Lens Fragmentation Alignment | Additional Nomogram<br>customizations |
| | Pre-op Manual Entry | |
| | Pre-op Import (from Cassini<br>pre-op device) | Pre-op Import (from IOLMaster<br>700 pre-op device) |
| | Iris Registration (built by<br>Cassini Technologies) | Iris Registration (built by<br>Johnson and Johnson) |
{7}------------------------------------------------
{8}------------------------------------------------
{9}------------------------------------------------
# [807.92(b)(1)] Determination of Substantial Equivalence
| | The cOS 7.0 software within the subject device was subjected to<br>hardware and software bench tests, in conjunction with simulated<br>use testing. |
|-----------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | Software-specific bench testing of the CatalysTM System was<br>conducted to demonstrate the System's ability to meet all<br>intended design specifications related to the software design |
| Non-Clinical<br>Performance Data: | changes. |
| | Bench testing of the predicate device is directly applicable to the<br>subject device as there are no significant changes to the subject<br>device other than the design changes resident in the software. |
| | Bench testing, when coupled with software regression testing,<br>verification and validation testing presented for the subject<br>device, including regression testing provides reasonable<br>assurance that the System remains safe and effective for its<br>intended use and furthermore, that it is substantially equivalent to<br>the identified predicate device |
{10}------------------------------------------------
| Clinical Performance<br>Data: | Clinical Data was deemed not necessary for the Catalys™<br>Precision Laser System. The performance data demonstrated that<br>the device performs as intended.<br>The subject device does not include any changes to the<br>indications for use or intended use of the predicate device. It does<br>not introduce any new harms or unacceptable risks, and therefore<br>does not require clinical testing to assess safety and performance<br>or to demonstrate equivalence. |
|----------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| [807.92(b)(3)] Conclusion | |
| Conclusions from Non-<br>Clinical and Clinical<br>Tests: | The subject Catalys Precision Laser System is substantially<br>equivalent to the currently cleared Catalys Precision Laser<br>System. The subject Catalys Precision Laser System is<br>substantially equivalent to the predicate in terms of indications<br>for use, technological characteristics and fundamental scientific<br>technology. The mechanism of laser cutting is the same for both<br>systems, in that the ultra-short laser pulses create a highly<br>localized plasma and subsequent cavitation event that, when<br>controlled by a computerized scanning system, direct the laser<br>beam through a three-dimensional pattern to produce a precise<br>capsulotomy, fragment the crystalline lens and create arc<br>cuts/incisions in the cornea. |
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.