← Product Code [IPF](/productcode/IPF) · K223802

# accufit (K223802)

_Mettler Electronics Corporation · IPF · Oct 6, 2023 · Physical Medicine · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K223802

## Device Facts

- **Applicant:** Mettler Electronics Corporation
- **Product Code:** [IPF](/productcode/IPF.md)
- **Decision Date:** Oct 6, 2023
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 890.5850
- **Device Class:** Class 2
- **Review Panel:** Physical Medicine
- **Attributes:** Therapeutic

## Indications for Use

Relaxation of muscle spasms Prevention or retardation of disuse atrophy Increase local blood circulation Muscle re-education Maintaining or increasing range of motion Immediate post-surgical stimulation of calf muscles to prevent venous thrombosis

## Device Story

The accufit is a noninvasive powered muscle stimulator used in outpatient clinical settings under clinician supervision. It delivers Biphasic and Interferential (4P) electrical waveforms to muscles via IntelliSTIM electrodes and IntelliGEL hydrogel adhesive strips. The device features four preprogrammed treatment regimens designed to mimic physical exercises, with electrode placement guided by the device's graphical user interface (GUI). An intelligent real-time impedance feedback system monitors the interface between the electrodes and the patient's skin. The clinician selects treatment duration (15, 30, or 45 minutes) and intensity. The device output is intended to stimulate muscles for re-education, spasm relaxation, and circulation improvement. It includes a patient-operated interrupt switch for safety. The system is designed to provide therapeutic benefits through controlled electrical stimulation, aiding in muscle maintenance and post-surgical recovery.

## Clinical Evidence

Bench testing only. No clinical or animal data was required. Performance was validated through electrical safety (IEC 60601-1), electromagnetic compatibility (IEC 60601-1-2), and software verification/validation testing. Biocompatibility of the hydrogel was confirmed per ISO 10993-5 and ISO 10993-10.

## Technological Characteristics

Powered muscle stimulator; ABS plastic housing; 100-240VAC power source. Delivers Biphasic and Interferential (4P) waveforms. Features real-time impedance feedback. Connectivity: Standalone device with GUI. Software: Moderate level of concern. Complies with IEC 60601-1, IEC 60601-1-2, IEC 60601-2-10, and ISO 14971. Electrodes use copper-coated carbon film and hydrogel adhesive.

## Regulatory Identification

A powered muscle stimulator is an electrically powered device intended for medical purposes that repeatedly contracts muscles by passing electrical currents through electrodes contacting the affected body area.

## Predicate Devices

- Mettler Sys*Stim ME 240 ([K113017](/device/K113017.md))
- Self-adhesive Electrode ([K222252](/device/K222252.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services seal on the left and the FDA acronym along with the full name of the agency on the right. The FDA part of the logo is in blue, with the acronym in a square and the full name written out to the right of the square.

October 6, 2023

Mettler Electronics Corporation An Le QS/RA Manager 1333 South Claudina Street Anaheim. California 92805

Re: K223802

Trade/Device Name: accufit Regulation Number: 21 CFR 890.5850 Regulation Name: Powered Muscle Stimulator Regulatory Class: Class II Product Code: IPF Dated: October 5, 2023 Received: October 5, 2023

Dear An Le:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

{1}------------------------------------------------

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

# Jitendra V. Virani -S

CDR Jitendra Virani, MS, MBA Assistant Director DHT5B: Division of Neuromodulation and Rehabilitation Devices OHT5: Office of Neurological and Physical Medicine Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

{2}------------------------------------------------

## Indications for Use

510(k) Number (if known) K223802

Device Name accufit

Indications for Use (Describe)

- Relaxation of muscle spasms
- · Prevention or retardation of disuse atrophy
- Increase local blood circulation
- Muscle re-education
- · Maintaining or increasing range of motion
- · Immediate post-surgical stimulation of calf muscles to prevent venous thrombosis

Type of Use (Select one or both, as applicable)

| X Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

FORM FDA 3881 (6/20)

{3}------------------------------------------------

## Mettler Electronics, Corporation Traditional 510k Submission accufit Powered Muscle Stimulator

21 CFR §890.5850

510(k) Summary – accufit (510k number K223802)

### Mettler Electronics Corporation

accufit 510k Summary

#### l. Submitter

Mettler Electronics, Corporation 1333 South Claudina Street Anaheim, CA USA 92805

Contact Person Timothy Duggins Phone: 602-200-4981 Email: timd@horizonphoenix.biz Date Revised: March 27, 2023

#### II. Device

| Trade Name:        | accufit                   |
|--------------------|---------------------------|
| Usual Name:        | powered muscle stimulator |
| Regulation Number: | 21 CFR 890.5850           |
| Regulation Name:   | Powered Muscle Stimulator |
| Regulatory Class:  | Class II                  |
| Product Code:      | IPF                       |

#### lll. Predicate Device

Premarket Notification:

| Primary Predicate Device (for the accufit): |                                  |
|---------------------------------------------|----------------------------------|
| Trade Name:                                 | Mettler Sys*Stim ME 240, K113017 |
| Common or Usual Name:                       | powered muscle stimulator        |
| Regulation Number:                          | 21 CFR 890.5850                  |
| Regulation Name:                            | Powered Muscle Stimulator        |
| Regulatory Class:                           | Class II                         |
| Product Code:                               | IPF, GZJ, ILY, LIH               |

Secondary Predicate Device: Self-adhesive Electrode, K222252 for the IntelliSTIM and IntelliGEL

| Trade Name:             | Self-adhesive Electrode                                        |
|-------------------------|----------------------------------------------------------------|
| Common or Usual Name:   | Cutaneous electrode                                            |
| Regulation Number:      | 21 CFR 882.1320                                                |
| Regulation Name:        | cutaneous electrode                                            |
| Regulatory Class:       | Class II                                                       |
| Product Code:           | GXY                                                            |
| Premarket Notification: | Shenzhen Roundwhale Technology Co., Ltd, K222252, (01/19/2023) |

Mettler Electronics K113017 (06/22/2012)

{4}------------------------------------------------

#### IV. Device Description

accufit (subject device) when using the Biphasic/Interferential (4P) waveforms is a noninvasive powered muscle stimulator used to provide an electrical stimulation of muscles for muscle re-education. The accufit is used on an outpatient basis under the supervision of a clinician.

The accufit is the result of a close collaboration between Lutronic Corporation and Mettler Electronics, Corporation. Mettler is the manufacturer of both the subject and predicate device. Using the Mettler Sys*Stim ME240 (K113017) as a starting point, the accufit utilizes two of the original and unchanged waveforms for muscle reeducation.

There are four preprogrammed treatment regimens that enable users to mimic common physical exercises. Each treatment regimen has recommended treatment electrode placement depicted in both the accufit operator's manual and on the accufit graphic user interface (GU). The IntelliSTIM treatment electrodes have been designed to support the accufit and are controlled by an intelligent real-time impedance feedback Mettler System. The IntelliSTIM are held in place with hydrogel adhesive strips and Velcro body wraps. The user may select the treatment duration of either 15, 30 or 45 minutes. The size, type of electrode and technical/performance characteristics.

## V. Indications for Use

As the accufit (subject device) and the Mettler Sys* Stim ME240 (predicate device K113017) have nearly identical technical and performance specifications. The two devices share the same indications for use considering the accufit uses only 2 of the Mettler Sys*Stim ME240 9 outputs/waveforms as indicated:

| Waveform                      | Indication for Use                                                                                                                                                                                                                                      |
|-------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Biphasic, Interferential (4P) | Relaxation of muscle spasms Prevention or retardation of disuse atrophy Increase local blood circulation Muscle re-education Maintaining or increasing range of motion Immediate post-surgical stimulation of calf muscles to prevent venous thrombosis |

## VI. Comparison of the technology characteristics with the predicate device

Mettler Electronics, Corporation believes that the accufit described herein and for use under the conditions of the proposed labeling is substantially equivalent to the Mettler Sys*Stim ME240, predicate device (K113017). Owing to the close collaboration between Lutronic Corporation and Mettler Electronics the subject and predicate device have nearly identical Technical Characteristics.

{5}------------------------------------------------

21 CFR §890.5850 Utilizing FDA's Guidance for Industry and FDA Staff "Format for traditional and Abbreviated 510(k)s", the accufit (subject device) is substantially equivalent to the Mettler Sys*Stim ME 240 (predicate device K113017).

## Indications for Use

The accufit (subject device) and the Mettler Sys*Stim ME240 (predicate device K113017) have the following identical indications for use considering the two common waveforms (Biphasic and Interferential (4P) bolded).

| Waveform                                                                                       | accufit                                                                                                                                                                                                                                                                         | ME240 (K113017)                                                                                                                                                                                                                                                                   | Comment         |
|------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------|
| Biphasic<br>Interferential (4P)                                                                | Relaxation of muscle spasms, Prevention<br>or retardation of disuse atrophy, Increase<br>local blood circulation, Muscle re-<br>education, Maintaining or increasing range<br>of motion, Immediate post-surgical<br>stimulation of calf muscles to prevent<br>venous thrombosis | Relaxation of muscle spasms,<br>Prevention or retardation of disuse<br>atrophy, Increase local blood<br>circulation, Muscle re-education,<br>Maintaining or increasing range of<br>motion, Immediate post-surgical<br>stimulation of calf muscles to<br>prevent venous thrombosis | Identical       |
| Microcurrent,<br>Interferential (4P),<br>Premodulated (2P3),<br>Biphasic and TENS<br>waveforms | Not applicable.                                                                                                                                                                                                                                                                 | Symptomatic relief and<br>management of chronic, intractable<br>pain, Post-traumatic acute pain,<br>Post-surgical acute pain                                                                                                                                                      | See Note below. |
| DC (Direct Current)<br>Mode                                                                    | Not applicable.                                                                                                                                                                                                                                                                 | Relaxation of muscle spasm                                                                                                                                                                                                                                                        | See Note below. |
| Laser and cluster<br>applicators of the<br>Sys*Stim 240emit<br>infrared energy                 | Not applicable.                                                                                                                                                                                                                                                                 | Temporary increase in blood<br>circulation, Temporary relief of<br>minor muscles and joint aches,<br>pains and stiffness, relaxation of<br>muscles, temporary relief of muscle<br>spasms, temporary relief of minor<br>pain and stiffness associated with<br>arthritis.           | See Note below. |

NOTE: The ME 240 (predicate device K113017) has other modes (Microcurrent, Premodulated (2P3), TENS, DC (direct current), and laser/cluster applicators which are NOT present in the acufit (subject device). There is no claim of substantial equivalency with these modes and their indications for use.

{6}------------------------------------------------

### Powered Muscle Stimulator

#### 21 CFR §890.5850

| Characteristic                                                            | accufit510k -K223802<br>(subject device)                                                                   | Mettler Sys*Stim ME240,<br>K113017 (predicate device)                                                                                                                         | 21 CFR §890.5850<br>Substantial equivalence discussion |
|---------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------|
| Manufacturer                                                              | Mettler Electronics Corporation                                                                            | Mettler Electronics<br>Corporation                                                                                                                                            | Identical                                              |
| Classification                                                            | Class II- 890-5850 - IPF                                                                                   | Class II- 890-5850 - IPF, GZJ,<br>ILY, LIH                                                                                                                                    | Identical                                              |
| Power Source                                                              | 100-240VAC, 50/60Hz, 1.0A<br>(Fuse: 250V / 6.3A)                                                           | 100-240VAC, 50/60Hz, 1.0A<br>(Fuse: 250V / 6.3A)                                                                                                                              | Identical                                              |
| Therapeutic Modality                                                      | Electrical muscle stimulator                                                                               | Electrical muscle stimulator                                                                                                                                                  | Identical                                              |
| Treatment Output Mode                                                     | Electrode                                                                                                  | Electrode                                                                                                                                                                     | Identical                                              |
| Method of Line Current Isolation                                          | Double Insulated Wire<br>Non- Conductive Enclosure                                                         | Double Insulated Wire<br>Non- Conductive Enclosure                                                                                                                            | Identical                                              |
| Measured Patient Leakage:                                                 |                                                                                                            |                                                                                                                                                                               |                                                        |
| Normal Condition (μ A)                                                    | Less than 100µA                                                                                            | Less than 100µA                                                                                                                                                               | Identical                                              |
| Single Fault Condition (µA)                                               | Less than 500µA                                                                                            | Less than 500µA                                                                                                                                                               | Identical                                              |
| Number of Output Modes                                                    | 2                                                                                                          | 9                                                                                                                                                                             | See Note #7.                                           |
| Number of Output Channels                                                 | 2                                                                                                          | 2                                                                                                                                                                             | Identical                                              |
| Regulated Current or Voltage?                                             | Current = 100mA maximum                                                                                    | Current = 100mA maximum                                                                                                                                                       | Identical                                              |
| Synchronous or Alternating                                                | Synchronous Channel 1&2                                                                                    | Synchronous Channel 1&2                                                                                                                                                       | Identical                                              |
| Constant Current                                                          | NA                                                                                                         | Optional                                                                                                                                                                      | See Note #5.                                           |
| Constant Voltage                                                          | NA                                                                                                         | Optional                                                                                                                                                                      | See Note #5.                                           |
| Method of Channel Isolation                                               | Line current isolation                                                                                     | Line current isolation                                                                                                                                                        | Identical                                              |
| Pulse Intensities                                                         | Adjustable, 0-100 mA peak into 500ohm<br>load each Electrode channel                                       | Not known                                                                                                                                                                     | See Note #6.                                           |
| Software/Firmware/Microprocessor<br>Control?                              | Yes                                                                                                        | Yes                                                                                                                                                                           | Substantially Equivalent see Note<br>#1                |
| Automatic Overload Trip?                                                  | Yes                                                                                                        | Yes                                                                                                                                                                           | Identical                                              |
| Automatic No Load Trip?                                                   | Yes                                                                                                        | Yes                                                                                                                                                                           | Identical                                              |
| Automatic Shut Off?                                                       | Yes                                                                                                        | Yes                                                                                                                                                                           | Identical                                              |
| Patient Override Control?                                                 | Patient Interrupt (Stop) Switch                                                                            | Patient interrupt Switch                                                                                                                                                      | Identical                                              |
| Indicator Display:                                                        | Yes                                                                                                        | Yes                                                                                                                                                                           | Substantially Equivalent see Note<br>#1                |
| On/ Off Status?                                                           | Yes                                                                                                        | Yes                                                                                                                                                                           | Substantially Equivalent see Note<br>#1                |
| Low Battery?                                                              | NA                                                                                                         | Yes                                                                                                                                                                           | NA See Note #3.                                        |
| Voltage/ Current Level?                                                   | Yes                                                                                                        | Yes                                                                                                                                                                           | Substantially Equivalent see Note<br>#1                |
| Timer range                                                               | 15, 30, 45 mins                                                                                            | 0 - 60 minutes                                                                                                                                                                | Substantially Equivalent see Note<br>#2                |
| Compliance: Voluntary Standards                                           | IEC 60601-1; 60601-1-2; 60601-2-10, ISO<br>14971, UL 60601, CSA C22.2 No 606.1,<br>MDD 93/42/EEC, Annex II | IEC 60601-1; 60601-1-2;<br>60601-2-10, ISO 14971, UL<br>60601, CSA C22.2 No 606.1,<br>MDD 93/42/EEC, Annex II                                                                 | Identical                                              |
| Compliance: 21 CFR 898                                                    | Yes                                                                                                        | Yes                                                                                                                                                                           | Yes                                                    |
| Weight                                                                    | 27 kg                                                                                                      | 4.5 lbs.,<br>5.5 lbs. with battery                                                                                                                                            | See Notes #3 and #4.                                   |
| Dimensions (mm) W x L x H                                                 | 452 x 582 x 1101                                                                                           | 8 inches (H) x 8 inches<br>(W) x 13 inches (L)                                                                                                                                | See Note #4                                            |
| Housing, Materials and Construction                                       | ABS plastic                                                                                                | ABS plastic                                                                                                                                                                   | Identical                                              |
| Powered Muscle Stimulator Device                                          | accufit 510k - K223802                                                                                     | Mettler SysStim ME240 -<br>K113017 (predicate device)                                                                                                                         | Substantial<br>Equivalence Discussion                  |
|                                                                           | (subject device)                                                                                           |                                                                                                                                                                               |                                                        |
| Waveforms                                                                 | 2 (Biphasic and Interferential (4P))                                                                       | Medium Frequency (Russian),<br>Biphasic, High Volt Pulsed<br>Current (HVPC), Interferential<br>(4P), Premodulated (2P),<br>Microcurrent, TENS and DC<br>(Direct Current) Mode | Different<br>See Note #2                               |
| Shape                                                                     | Square (Biphasic)<br>Sinusoidal (Interferential (4P))                                                      | Square (Biphasic) Sinusoidal<br>(Interferential (4P))                                                                                                                         | Identical                                              |
| Maximum Output Voltage (Biphasic)<br>(Volts, peak to peak)                | 104 @ 500Ω; 197 @ 2kΩ; 178 @ 10kΩ                                                                          | 104 @ 500Ω; 197 @ 2kΩ; 178 @<br>10kΩ                                                                                                                                          | Identical                                              |
| Maximum Output Voltage (Interferential<br>(4P)) (Volts, peak to peak)     | 106 @ 500Ω; 144 @ 2kΩ; 122 @ 10kΩ                                                                          | 106 @ 500Ω; 144 @ 2kΩ; 122 @<br>10kΩ                                                                                                                                          | Identical                                              |
| Maximum Output Current (Biphasic)<br>(mA, ±20%)                           | 106 @ 500Ω; 49 @ 2kΩ; 9 @ 10kΩ                                                                             | 106 @ 500Ω; 49 @ 2kΩ; 9 @<br>10kΩ                                                                                                                                             | Identical                                              |
| Maximum Output Current<br>(Interferential (4P)) (mA, ±20%)                | 106 @ 500Ω; 35 @ 2kΩ; 6.3 @ 10kΩ                                                                           | 106 @ 500Ω; 35 @ 2kΩ; 6.3 @<br>10kΩ                                                                                                                                           | Identical                                              |
| Pulse Width (Biphasic)                                                    | 40-800 µSec                                                                                                | 40-800 µSec                                                                                                                                                                   | Identical                                              |
| Pulse Width (Interferential (4P))                                         | 200-400 µSec                                                                                               | 200-400 µSec                                                                                                                                                                  | Identical                                              |
| Frequency (Biphasic)                                                      | 1-200 Hz                                                                                                   | 1-200 Hz                                                                                                                                                                      | Identical                                              |
| Frequency (Interferential (4P))                                           | 2.5 kHz, 4 kHz, 5 kHz                                                                                      | 2.5 kHz, 4 kHz, 5 kHz                                                                                                                                                         | Identical                                              |
| For Interferential (4P) modes only: Best<br>Frequency (Hz)                | 1-250 Hz                                                                                                   | 1-250 Hz                                                                                                                                                                      | Identical                                              |
| For Multiphasic waveforms only:<br>Symmetrical phases?                    | Yes                                                                                                        | Yes                                                                                                                                                                           | Identical                                              |
| Phase duration (Biphasic) - (μSec)                                        | ~35 µSec                                                                                                   | Not known                                                                                                                                                                     | See Note #1.                                           |
| Phase duration (Interferential (4P)) -<br>(μSec)                          | 100, 125, 200 µSec                                                                                         | 100, 125, 200 µSec                                                                                                                                                            | Identical                                              |
| Net Charge (μC micro Coulombs per<br>pulse). If zero, method of achieving | Net Charge = 0 µC (Biphasic &<br>Interferential (4P) pulses are<br>symmetrical about OV)                   | Net Charge = 0 µC<br>(Biphasic &<br>Interferential (4P) pulses are<br>symmetrical about 0V)                                                                                   | Identical                                              |
| Duration of Primary (depolarizing) Phase                                  | ~35 µSec                                                                                                   | Not known                                                                                                                                                                     | See Note #1.                                           |

NOTE 1: In each device the software operates the GUI. In the case of the accufit the software only activates the 2 waveforms described, the Biphasic and the Interferential (4P). The difference does not raise any new safety or effectiveness questions and each device has identical safety and performance characteristics.

Note 2: The accufit uses identical levels of intensity however for a shorter duration. The difference does not raise any new safety or effectiveness questions and each device has identical safety and performance characteristics.

{7}------------------------------------------------

## Mettler Electronics, Corporation Traditional 510k Submission accufit Powered Muscle Stimulator

#### 21 CFR §890.5850

Note 3: The accufit does not have a battery. The difference does not raise any new safety or effectiveness questions and each device has identical safety and performance characteristics.

Note 4: The accufit is enclosed in an upright, wheeled configuration which is merely a branding initiative. The difference does not raise any new safety or effectiveness questions and each device has identical safety and performance characteristics.

Note 5: The accufit does have these options. The difference does not raise any new safety or effectiveness questions and each device has identical safety and performance characteristics.

Note 6: The accufit (subject device) and ME 240 (predicate device K113017) have identical electronic characteristics, therefore the difference in pulse intensities (if any) do not raise any new safety or effectiveness questions and each device has identical safety and performance characteristics.

Note 7: The accufit (subject device) and ME240 (predicate device K113017) generate identical (40) waveforms. The ME240 (predicate device K113017) generates six additional waveforms and one additional mode which are not found in the accufit (subject device). Substantially equivalency claim is limited only to the Biphasic and Interferential (4P) waveforms. The difference does not raise any new safety or effectiveness questions and each device has identical safety and performance characteristics.

#### VII. Comparison of the output specifications with the predicate device (K113017)

{8}------------------------------------------------

#### Powered Muscle Stimulator

#### 21 CFR §890.5850

| Pulse Duration                                                                                       | 360 µSec                                   | Not known           | See Note #1. |
|------------------------------------------------------------------------------------------------------|--------------------------------------------|---------------------|--------------|
| Maximum Phase Charge(µC) (Biphasic)                                                                  | 42.4 µC, 500 Ω                             | 42.4 µC, 500 Ω      | Identical    |
| Maximum Phase Charge (Interferential<br>(4P))                                                        | 14.9 µC, 500 Ω                             | 14.9 µC, 500 Ω      | Identical    |
| Maximum Power Density (Biphasic) (using<br>smallest electrode conductive surface<br>area)            | 0.043 W/cm², 500 Ω                         | 0.043 W/cm², 500 Ω  | Identical    |
| Maximum Power Density (Interferential<br>(4P)) (using smallest electrode conductive<br>surface area) | 0.176 W/cm², 500 Ω                         | 0.176 W/cm², 500 Ω  | Identical    |
| Maximum Current Density (Biphasic)                                                                   | 5.13 mA/cm², 500 Ω                         | 5.13 mA/cm², 500 Ω  | Identical    |
| Maximum Current Density (Interferential<br>(4P))                                                     | 5.196 mA/cm², 500 Ω                        | 5.196 mA/cm², 500 Ω | Identical    |
| Burst Mode (pulse trains)                                                                            | NA (no bursts)                             | NA (no bursts)      | Identical    |
| ON Time (seconds)                                                                                    | Biphasic 1 – 240<br>Interferential (4P) NA | Not known           | See Note #1. |
| OFF Time (seconds)                                                                                   | Biphasic 1 – 240<br>Interferential (4P) NA | Not known           | See Note #1. |

Note 1: The accufit (subject device) and ME 240 (predicate device K113017) have identical electronic characteristics, therefore the difference in these characteristics (if any) do not raise any new safety or effectiveness questions and each device has identical safety and performance characteristics.

Note 2: The accufit (subject device) and ME240 (predicate device K113017) generate identical (4P) waveforms. The ME240 (predicate device K113017) generates six additional waveforms and one additional mode which are not found in the accufit (subject device). Substantially equivalency claim is limited only to the Biphasic and Interferential (40) waveforms. The difference does not raise any new safety or effectiveness questions and each device has identical safety and performance characteristics.

The IntelliSTIM and IntelliGEL are substantially equivalent to the Self Adhesive Electrode predicate device (K222252) as shown in the following table:

| Characteristic<br>See Note 1                    | IntelliSTIM<br>IntelliGEL<br>K223802<br>Subject Device                                                                                                                                                                                                                                          | Self-adhesive Electrode<br>Predicate Device<br>K222252                                                                                                                                                                                                                                                                                                                                                                                                                              | Discussion                                 |
|-------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------|
| Regulation                                      | 21 CFR 882.1320                                                                                                                                                                                                                                                                                 | 21 CFR 882.1320                                                                                                                                                                                                                                                                                                                                                                                                                                                                     | Identical                                  |
| Product Code                                    | GXY                                                                                                                                                                                                                                                                                             | GXY                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 | Identical                                  |
| Manufacturer                                    | Mettler Electronics<br>Corporation                                                                                                                                                                                                                                                              | Shenzhan Roundwhale<br>Technology Co. Limited                                                                                                                                                                                                                                                                                                                                                                                                                                       | NA                                         |
| Classification Name                             | Cutaneous electrode                                                                                                                                                                                                                                                                             | Cutaneous Electrode                                                                                                                                                                                                                                                                                                                                                                                                                                                                 | Identical                                  |
| Model Name                                      | IntelliSTIM (electrodes)<br>IntelliGEL (hydrogel<br>adhesive)                                                                                                                                                                                                                                   | Self-adhesive Electrode<br>with lead wire                                                                                                                                                                                                                                                                                                                                                                                                                                           | See Note #1.                               |
| Common Name                                     | Reusable Neurostimulation<br>Electrode                                                                                                                                                                                                                                                          | Reusable Neurostimulation<br>Electrode                                                                                                                                                                                                                                                                                                                                                                                                                                              |                                            |
| OTC/Prescription                                | Prescription                                                                                                                                                                                                                                                                                    | OTC and Prescription                                                                                                                                                                                                                                                                                                                                                                                                                                                                | See Note #5.                               |
| Target Population                               | General (Adult)                                                                                                                                                                                                                                                                                 | General (Adult)                                                                                                                                                                                                                                                                                                                                                                                                                                                                     | Identical                                  |
| Electrical Connection                           | Lead wire                                                                                                                                                                                                                                                                                       | Lead wire, snap button,<br>Magnetic Button                                                                                                                                                                                                                                                                                                                                                                                                                                          | Identical<br>See Note #2.                  |
| Size and Shape                                  | Rectangular 60*70mm                                                                                                                                                                                                                                                                             | Round 50±0.5cm),<br>Rectangular (50*<br>30±0.5mm), Square<br>50*50±0.5mm, Square<br>40*40±0.5mm, Oval<br>40*80±0.5mm                                                                                                                                                                                                                                                                                                                                                                | Substantially Equivalent<br>See Note #2.   |
| Protective Cover/Liner                          | ABS                                                                                                                                                                                                                                                                                             | PET                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 | Substantially Equivalent<br>See Note #3.   |
| Conductive Component                            | Copper coated carbon film                                                                                                                                                                                                                                                                       | Silver coated carbon film                                                                                                                                                                                                                                                                                                                                                                                                                                                           | Substantially Equivalent<br>See Note #4.   |
| Color                                           | Red and Black                                                                                                                                                                                                                                                                                   | White                                                                                                                                                                                                                                                                                                                                                                                                                                                                               | See Note #5.                               |
| Lead Wire Connector                             | .080.0.1" female socket or<br>pin connector                                                                                                                                                                                                                                                     | .080/0.1" female socket or<br>pin connector                                                                                                                                                                                                                                                                                                                                                                                                                                         | Identical                                  |
| Design Features                                 | (1) Protective film lining of<br>the applied part consisting<br>of Polyethylene<br>terephthalate and Silicon;<br>(2) the hydrogel with silver<br>chloride substrate and (3) a<br>liner protecting the<br>hydrogel and the electrode<br>consisting of Polyethylene<br>terephthalate and Silicon. | Six layers: 1.Insulation<br>backing material:<br>Fabric/Foam/ Tan fabric<br>2.Conductive film:<br>Aluminum foil film /Carbon<br>film/Carbon film coated<br>with silver/ 3.Conductive<br>hydrogel 4.Connection<br>5.double sides adhesive<br>tape 6.Release liner                                                                                                                                                                                                                    | See Note #1.                               |
| Reusable                                        | Reusable                                                                                                                                                                                                                                                                                        | Reusable                                                                                                                                                                                                                                                                                                                                                                                                                                                                            | See Note #1.                               |
| Single Patient Use                              | Yes                                                                                                                                                                                                                                                                                             | Yes                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 | Identical                                  |
| Biocompatibility                                | ISO 10993 compliant                                                                                                                                                                                                                                                                             | ISO 10993 compliant                                                                                                                                                                                                                                                                                                                                                                                                                                                                 | Identical                                  |
| Sterile                                         | Nonsterile only                                                                                                                                                                                                                                                                                 | Nonsterile only                                                                                                                                                                                                                                                                                                                                                                                                                                                                     | Identical                                  |
| Adhesive Type                                   | Self-adhesive<br>(biocompatible self-<br>adhesive conductive<br>hydrogel)                                                                                                                                                                                                                       | Self-adhesive                                                                                                                                                                                                                                                                                                                                                                                                                                                                       | Identical See Note #1.                     |
| Electrical Safety                               | ANSI AAMI ES60601-<br>1:2005/(R)2012 & A1:2012,<br>C1:2009/(R)2012 &<br>A2:2010/(R)2012 (Cons. Text)<br>[Incl. AMD2:2021] , Medical<br>electrical equipment - Part 1:<br>General requirements for basic<br>safety and essential<br>performance                                                  | ANSI AAMI ES60601-<br>1:2005/(R)2012 & A1:2012,<br>C1:2009/(R)2012 &<br>A2:2010/(R)2012 (Cons. Text)<br>[Incl. AMD2:2021] , Medical<br>electrical equipment - Part 1:<br>General requirements for<br>basic safety and essential<br>performance                                                                                                                                                                                                                                      | Identical See Note 5.                      |
| emc/emi                                         | ANSI AAMI ES60601-<br>1:2005/(R)2012 & A1:2012,<br>C1:2009/(R)2012 &<br>A2:2010/(R)2012 (Cons. Text)<br>[Incl. AMD2:2021], Cl.8.5.2.3                                                                                                                                                           | ANSI AAMI ES60601-<br>1:2005/(R)2012 & A1:2012,<br>C1:2009/(R)2012 &<br>A2:2010/(R)2012 (Cons. Text)<br>[Incl. AMD2:2021], Cl.8.5.2.3                                                                                                                                                                                                                                                                                                                                               | Identical                                  |
| Biocompatibility                                | Complies with ISO 10933                                                                                                                                                                                                                                                                         | Complies with ISO 10933                                                                                                                                                                                                                                                                                                                                                                                                                                                             | Identical                                  |
| Force required to remove<br>wire from electrode | More than six pounds of force                                                                                                                                                                                                                                                                   | More than six pounds of force                                                                                                                                                                                                                                                                                                                                                                                                                                                       | Identical                                  |
| AC Impedance                                    | <200Ω                                                                                                                                                                                                                                                                                           | <300Ω                                                                                                                                                                                                                                                                                                                                                                                                                                                                               | Substantially<br>Equivalent - See Note 2.  |
| Adhesive strength                               | >185g/in                                                                                                                                                                                                                                                                                        | ≥7N                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 | See Note 6.                                |
| Indications for use                             | The IntelliSTIM and IntelliGel<br>when used according to the<br>accufit operator's manual<br>forms a conductive adhesive<br>interface between the accufit<br>and the patient's skin.                                                                                                            | Self-Adhesive Electrode are<br>intended for use as a<br>reusable, conductive adhesive<br>interface between the<br>patient's skin and the<br>marketed electrical<br>stimulators (i.e. TENS<br>(Transcutaneous Electrical<br>Nerve Stimulation), EMS<br>(Electrical Muscular<br>Stimulation), IF or IFC<br>(Interferential) or PGF (Pulsed<br>Galvanic Stimulation) for<br>transmitting electrical<br>current. The electrode is for<br>OTC (Over-The-Counter) or<br>Prescription use. | Substantially Equivalent -<br>See Note #1. |

{9}------------------------------------------------

## Mettler Electronics, Corporation Traditional 510k Submission accufit Powered Muscle Stimulator

## 21 CFR §890.5850

{10}------------------------------------------------

Powered Muscle Stimulator

Note 1 – In each device the electrode floats above the patient through the use of a hydrogel pad. The accufit IntelliStim electrode and IntelliGel self-adhesive hydrogel pad forms an identical combination to the predicate device (K22252) however the IntelliStim electrode being separate is reusable. The IntelliGel pad is disposable. The Self Adhesive Electrodes [Predicate Device, K222252] is a substantially equivalent combination of electrode and hydrogel pad however as the electrode and hydrogel pad are a single unit, therefore entire accessory is single use. The IntelliGEL are for use only with the accufit and no claims are made for use on any other device. The difference does not raise any new safety or effectiveness questions and each device has identical biocompatible safety and performance characteristics

Note 2 – The electrode's size is directly related to the area being treated. Its AC impedance, size and shape evenly distributes the electrical stimulation preventing burns or similar injuries. The IntelliSTIM electrode was purpose-developed for use the accufit and for no other device, there is no snap button or magnetic button required. There is no substantial equivalence claimed for any size of the Predicate's (K22252) Square 50*50±0.5mm electrode. The difference does not raise any new safety or effectiveness questions and each device has identical safety and performance characteristics.

Note 3 – Both materials are thermoplastics with identical performance characteristics. The difference does not raise any new safety or effectiveness questions and each device has identical safety and performance characteristics.

### 21 CFR §890.5850

{11}------------------------------------------------

## Powered Muscle Stimulator

#### 21 CFR §890.5850

Note 4 - Both copper and silver are highly conductive materials. Copper is used in the IntelliSTIM because the metal is stronger and therefore more suited for a reusable accessory. Both the subject and predicate device (K222252) were tested according to the ISO 10993 series of Standard and were found to have zero toxic, irritating or sensitization risks. The difference does not raise any new safety or effectiveness questions and each device has identical safety and performance characteristics.

Note 5 – The IntelliSTIM electrode set is designed to be used with the accufit and no other device. The Self-adhesive Electrode (K22252)is designed to be used with multiple devices. The IntelliSTIM electrodes was included in the accufit's electrical safety and emc/emi testing; for example IEC 60601 clause 8.5.2.3, the combination accufit and IntelliSTIM electrode was found to be in compliance with all pertinent requirements. The difference does not raise any new safety or effectiveness questions and each device has identical safety and performance characteristics.

Note 6 - 185g/in was determined using a different methodology then that used in the predicate device (K22252). The adhesive strength is stronger in the subject device owing to having two elements (IntelliSTIM and IntelliGEL) instead of one and it was designed for exclusive use with the accufit. Design did not have to consider the variety of devices that the electrode/adhesive might be used on. The difference does not raise any new safety or effectiveness questions and each device has identical safety and performance characteristics.

### VIII. Indications for Use Statement Comparison

The accufit (subject device) uses only two of the Mettler Sys*Stim ME240's waveforms/outputs:

| Waveform                        | accufit  | ME240 (K113017) |
|---------------------------------|----------|-----------------|
| Medium Frequency (Russian)      | Not used | ×               |
| Biphasic                        | ×        | ×               |
| High Volt Pulsed Current (HVPC) | Not used | ×               |
| Interferential (4P)             | ×        | ×               |
| Premodulated (2P3) waveforms    | Not used | ×               |
| Microcurrent                    | Not used | ×               |
| Interferential (4P3)            | Not used | ×               |
| Premodulated (2P)               | Not used | ×               |
| DC (Direct Current)             | Not used | ×               |
| Laser/cluster infrared          | Not used | ×               |

{12}------------------------------------------------

## 21 CFR §890.5850

#### IX. Performance Data

### Biocompatibility

## Testing

The accufit and the IntelliSTIM treatment electrodes have no meaningful biocompatibility risk or hazard as there is no patient contact.

The IntelliGel hydrogel adhesive strips biocompatibility evaluation was conducted in accordance with the FDA's Guidance for Industry and FDA Staff "Use of International Standard ISO 10993-1, "Biological Evaluation of Medical Devices Part 1: Evaluation and Testing within a Risk Management Process" using:

- a. ISO 10993-5, Biological evaluation of medical devices Part 5: Tests for in vitro cytotoxicity,
- b. ISO 10993-10, Biological evaluation of medical devices Part 10: Tests for Irritation and Sensitization,

No appreciable risk or hazard was detected.

## Sterilization and Shelf-Life

The accufit, IntelliSTIM and IntelliGel are not provided sterile and does not need to be sterilized by the end user. The accufit and IntelliSTIM electrodes are cleaned with a soft cloth usually moistened with isopropyl alcohol or a gentle household cleaner. The IntelliSTIM treatment electrodes must have the hydrogel removed from the electrode using a lint free cloth soaked with 90% isopropyl alcohol. The IntelliGel expected service life is 2 years and has been established by accelerated shelf life testing.

## Electrical Safety and Electromagnetic Compatibility (EMC)

Electrical safety and EMC testing were conducted on the accufit and IntelliSTIM treatment electrodes (including the hospital grade power cord and patient stop switch) by a duly accredited third party test laboratory whose scope includes electromedical equipment. In every instance the accufit and the IntelliSTIM treatment electrodes were found to be in compliance with:

- . IEC 60601-1, an International Standard on Medical electrical equipment general requirements for safety (FDA Recognition List Number 19-4);
- . IEC 60601-1-2, an International Standard on Medical electrical equipment, electromagnetic compatibility (FDA Recognition List Number 19-8);
- . IEC 60601-1-6, an International Standard on Medical electrical equipment – Part 1-6: General requirements for basic safety and essential performance – Collateral standard: Usability (FDA Recognition List Number 5-132), and

{13}------------------------------------------------

## Powered Muscle Stimulator

### 21 CFR §890.5850

- IEC 60601-2-10 an International Standard on Medical electrical equipment Part 2-10: . Particular requirements for the basic safety and essential performance of nerve and muscle stimulators (FDA Recognition List Number 17-16).
## Software Verification and Validation Testing

Software verification and validation testing were conducted, and documentation was provided as recommended by FDA's Guidance for Industry and FDA Staff, "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices" and ISO 62304 Medical device software — Software life cycle processes. The software for the accufit was considered as a "moderate" level of concern (FDA) and as a Class B (ISO 62304) since a failure of the software could result in minor injury to a patient or to a user of the device.

## Performance Testing

No animal or clinical testing was required to demonstrate the accufit is safe and will perform as intended. Bench Testing was performed to ensure that the accufit performs as intended and documentation was provided. Testing included normal, single fault and multiple fault scenarios using the requirements of FDA recognized Standards for electrical safety, electromagnetic disturbances and powered muscle stimulators.

#### X. Conclusions

The accufit [subject device] and the ME240 (primary predicate device, K113017) emanate Biphasic/Interferential (4P) waveforms. The predicate device (K113107)has other modes and indications, but no claim of substantial equivalency is made for these. When used according to the accufit operator's manual, with only those two waveforms, the accufit has the identical intended use, identical indications for use, and the identical fundamental scientific technology and performance characteristics as the predicate device (K113017). The IntelliSTIM and IntelliGel (subject devices) have identical intended use, identical fundamental scientific technology and performance characteristics with the secondary predicate device the Selfadhesive electrode (K22252). The differences between the subject and primary (K113017) and secondary predicate device (K22252) are minor are NOT the result of any short coming or adverse event. Therefore, the accufit is as safe, as effective, and performs as well as the legally marketed predicate devices (K113017) and (K222252).

---

**Source:** [https://fda.innolitics.com/device/K223802](https://fda.innolitics.com/device/K223802)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
