← Product Code [SBF](/productcode/SBF) · K223769

# NextAR™ Spine Platform (K223769)

_Medacta International S.A. · SBF · Jul 21, 2023 · Neurology · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K223769

## Device Facts

- **Applicant:** Medacta International S.A.
- **Product Code:** [SBF](/productcode/SBF.md)
- **Decision Date:** Jul 21, 2023
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 882.4560
- **Device Class:** Class 2
- **Review Panel:** Neurology

## Indications for Use

The NextAR Spine platform is intended as an aid for precisely locating anatomical structures in either open or percutaneous spine procedures. It is indicated for any medical condition in which the use of stereotaxic surgery may be appropriate, when reference to a rigid anatomical structure, such as vertebrae or pelvis, can be identified relative to images of the anatomy. This can include posterior approach spinal procedures, such as: - · Pedicle Screw Placement (Thoracic and Lumbosacral spine) - Iliosacral Screw Placement The NextAR Spine platform is intended to be used in combination with NextAR™Struments and / or Medacta preoperative planning. In the case of pre-operative planning software is used pre-operatively to plan the surgical placement of pedicle screws based upon radiological images of the patient. As an optional display, the NextAR Smart Glasses can be used auxiliary to the NextAR Spine Platform to view stereotaxic information as presented by the NextAR Spine Platform. The NextAR Smart Glasses should not be relied upon solely and should always be used in conjunction with the primary computer display.

## Device Story

NextAR Spine Platform is a CT-based computer-assisted surgical navigation system for open or percutaneous spine procedures. It uses an active infrared camera and active trackers (sensors) mounted on bones to monitor instrument spatial location in real-time. The system registers patient anatomy using either intra-operative 3D C-arm scans or a combination of pre-operative CT and intra-operative 3D C-arm scans. Surgeons use the system to execute surgical plans by visualizing instrument trajectories and screw placement relative to anatomy on a primary monitor; optional AR smart glasses provide auxiliary visualization. The system aids surgeons in precise anatomical localization, potentially improving surgical accuracy. It is intended for use in clinical settings by trained surgeons.

## Clinical Evidence

No clinical trials on human subjects were performed. Evidence consists of benchtop accuracy verification, software verification and validation, performance testing of mechanical and optical properties, and a cadaver validation study to confirm system performance and safety.

## Technological Characteristics

Stereotaxic navigation system utilizing active infrared camera and active trackers. Components include PC-based hardware, optical tracking system, fiducial blocks, and surgical instruments. Registration via 3D C-arm or pre-operative CT/intra-operative 3D C-arm fusion. Connectivity includes primary monitor and optional AR smart glasses. Software-based navigation platform.

## Regulatory Identification

A stereotaxic instrument is a device consisting of a rigid frame with a calibrated guide mechanism for precisely positioning probes or other devices within a patient's brain, spinal cord, or other part of the nervous system.

## Predicate Devices

- NextAR™ Spine Platform ([K210859](/device/K210859.md))

## Reference Devices

- Xvision Spine (XVS) ([K211188](/device/K211188.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is a blue square with the letters "FDA" in white, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue.

July 21, 2023

Medacta International S.A. % Chris Lussier Senior Director, Quality, Regulatory, and Clinical Research Medacta USA 3973 Delp Street Memphis, Tennessee 38118

Re: K223769

Trade/Device Name: NextAR Spine Platform Regulation Number: 21 CFR 882.4560 Regulation Name: Stereotaxic Instrument Regulatory Class: Class II Product Code: OLO, JWH, PBF Dated: June 21, 2023 Received: June 22, 2023

Dear Chris Lussier:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

Shumaya Ali-S

Shumaya Ali, M.P.H. Assistant Director DHT6C: Division of Restorative, Repair and Trauma Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

## Indications for Use

510(k) Number (if known) K223769

Device Name NextAR™ Spine Platform

### Indications for Use (Describe)

The NextAR Spine platform is intended as an aid for precisely locating anatomical structures in either open or percutaneous spine procedures. It is indicated for any medical condition in which the use of stereotaxic surgery may be appropriate, when reference to a rigid anatomical structure, such as vertebrae or pelvis, can be identified relative to images of the anatomy. This can include posterior approach spinal procedures, such as:

- · Pedicle Screw Placement (Thoracic and Lumbosacral spine)
- Iliosacral Screw Placement

The NextAR Spine platform is intended to be used in combination with NextAR™Struments and / or Medacta preoperative planning. In the case of pre-operative planning software is used pre-operatively to plan the surgical placement of pedicle screws based upon radiological images of the patient. As an optional display, the NextAR Smart Glasses can be used auxiliary to the NextAR Spine Platform to view stereotaxic information as presented by the NextAR Spine Platform. The NextAR Smart Glasses should not be relied upon solely and should always be used in conjunction with the primary computer display.

| Type of Use ( <i>Select one or both, as applicable</i> )                                                                                                                                                |                                                                                 |                                                                                |
|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------|--------------------------------------------------------------------------------|
| <table><tr><td><div> <span> </span>     Prescription Use (Part 21 CFR 801 Subpart D)    </div></td><td><div> <span> </span>     Over-The-Counter Use (21 CFR 801 Subpart C)    </div></td></tr></table> | <div> <span> </span>     Prescription Use (Part 21 CFR 801 Subpart D)    </div> | <div> <span> </span>     Over-The-Counter Use (21 CFR 801 Subpart C)    </div> |
| <div> <span> </span>     Prescription Use (Part 21 CFR 801 Subpart D)    </div>                                                                                                                         | <div> <span> </span>     Over-The-Counter Use (21 CFR 801 Subpart C)    </div>  |                                                                                |

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

***DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.***

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

# 510(k) Summary

#### Submitter I.

Medacta International SA Strada Regina 6874 Castel San Pietro (CH) Switzerland Phone (+41) 91 696 60 60 Fax (+41) 91 696 60 66

Contact Person: Stefano Baj, Regulatory Affairs Director, Medacta International SA Applicant Correspondent: Christopher Lussier, Sr. Director, Quality and Regulatory, Medacta USA Date Prepared: July 21, 2023

#### II. Device

| Device Proprietary Name: | NextAR™ Spine Platform  |
|--------------------------|-------------------------|
| Common or Usual Name:    | Navigation System       |
|                          | Total Joint Replacement |
| Classification Name:     | Stereotaxic Instrument  |
| Primary Product Code:    | OLO                     |
|                          | JWH                     |
|                          | PBF                     |
| Regulation Number:       | 21 CFR 882.4560         |
| Device Classification    | II                      |

#### III. Predicate Device

Substantial equivalence is claimed to the following devices:

Primary predicate device:

- NextARTM Spine Platform K210859, Medacta International SA .
Reference devices:

- . Xvision Spine (XVS), K211188, Augmedics Ltd
#### IV. Device Description

The NextARTM Spine Platform is a CT based computer-assisted surgical navigation platform used in either open/mini open or percutaneous spine surgery procedure and includes the following components:

{4}------------------------------------------------

- navigation software which displays information to the surgeon; ●
- Augmented Reality glasses:
- optical tracking system;
- PC based hardware platform; ●
- Fiducial Block;
- Adaptor for sensor:
- Spine attachment instruments ●
- Reusable surgical instruments for spine surgery procedures.

The system operates on the common principle of stereotaxic technology in which markers are mounted on the bones and an infrared camera is used to monitor the spatial location of the instruments. Tracking sensors attached to the bones enable the surgeon to view the position and orientation of the instrumentation relative to the intra-operative data in real-time while performing the surgical procedure. The tracking sensors, the fiducial block and a group of pins and drills are provided sterile.

The NextAR™ Spine Platform aid the surgeon in executing the surgical plan by visualizing all the information in real time in a screen monitor.

The NextAR Spine system is a surgical navigation platform which uses the information of either an intra-operative scan or pre-operative CT in combination with an intra-operative 3D-CArm scan in order to register the spine to navigation elements.

The registration can be performed with one of the following approaches:

- . Direct 3D: based on the use of an intra-operative 3D-CArm scan
- . 3D-3D: based on the use of a pre-operative CT scan and an intra-operative 3D-CArm scan

The system's navigation technology is based on an active infrared camera coupled with an active tracker (Target). These elements allow, by means of the different registration approaches and use of compatible instruments, to accurately prepare trajectories in the vertebrae and/or to implant screws while visualizing information in real time on a screen monitor.

#### V. Indications for Use

The NextAR Spine platform is intended as an aid for precisely locating anatomical structures in either open/mini-open or percutaneous spine procedures. It is indicated for any medical condition in which the use of stereotaxic surgery may be appropriate, when reference to a rigid anatomical structure, such as vertebrae or pelvis, can be identified relative to images of the anatomy. This can include posterior approach spinal procedures, such as:

- · Pedicle Screw Placement (Thoracic and Lumbosacral spine)
- Iliosacral Screw Placement

The NextAR Spine platform is intended to be used in combination with NextAR™ Stereotaxic instruments and / or Medacta preoperative planning. In the case of pre-operative planning, 

{5}------------------------------------------------

surgical planning software is used pre-operatively to plan the surgical placement of pedicle screws based upon radiological images of the patient. As an optional display, the NextAR Smart Glasses can be used auxiliary to the NextAR Spine Platform to view stereotaxic information as presented by the NextAR Spine Platform. The NextAR Smart Glasses should not be relied upon solely and should always be used in conjunction with the primary computer display.

#### VI. Comparison of Technological Characteristics

The NextAR™ Spine System and the predicate NextAR™ Spine Platform, K210859, share the following characteristics:

- principle of operation; ●
- user interface: ●
- power source; ●
- optical tracking system;
- platform: ●
- use of surgical instruments for navigation;
- main system components;
- accuracy;
- computer hardware;
- pre-operative patient anatomy data acquisition; ●
- intra-operative patient anatomy data acquisition. ●

The NextAR™ Spine Platform and the predicate Xvision Spine K211188 share the following characteristics:

- Imaging Modality ●
- Displaying Technology
- Glasses communication; ●
- . Use of surgical instruments for navigation;

## Discussion

The differences between the subject and the primary predicate device are concerning the anatomical views in the AR glasses, glasses communication and the use, as optional, the universal camera adapter.

They do not raise different questions of safety or effectiveness when compared to the predicate device. Both navigation systems utilize stereotaxic technologies.

Minor differences are addressed by performing cadaveric testing/ rational/ test.

#### VII. Performance Data

Testing was conducted according to written protocols with acceptance criteria that were based on standards. The following studies were performed in support of a substantial equivalence determination:

{6}------------------------------------------------

- software verification and validation; ●
- . benchtop accuracy verification;
- performance testing to evaluate mechanical and optical properties; and ●
- . cadaver validation study.

## VIII. Conclusion

The information provided above supports that the NextAR™ Spine Platform is substantially equivalent to the identified predicate device.

Substantial equivalence has been demonstrated through a comparison of intended use, design and technological characteristics, as well as performance evaluations. The NextAR™ Spine Platform can be considered substantially equivalent to the identified predicate device.

---

**Source:** [https://fda.innolitics.com/device/K223769](https://fda.innolitics.com/device/K223769)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
