The mask is intended to provide an interface for CPAP or bi-level devices.
Device Story
Whitsundays Mask System is a non-invasive interface for delivering CPAP or bi-level positive airway pressure to patients with Obstructive Sleep Apnea or requiring ventilatory support. Device consists of nasal cushion, conduit frame, elbow, and headgear. It connects to PAP flow generators via 22mm conical connector (ISO 5356-1). Used in home (single-patient) or hospital/institutional (multi-patient) settings. Airflow is delivered to the patient's upper airway via a seal under the nose. Healthcare providers select appropriate cushion/frame sizes for patient fit. Device facilitates therapy compliance by providing a stable, comfortable interface. SLM variant is validated for multi-patient reprocessing using high-level thermal disinfection.
Clinical Evidence
Bench testing only. No clinical data. Performance verified against ISO 17510:2015 (CO2 rebreathing, exhaust flow, resistance to flow). Mechanical integrity, pressure accuracy, and environmental testing (cleaning, storage, free fall) performed. Biocompatibility evaluated per ISO 10993-1 and ISO 18562-1. Reprocessing validation (cleaning/disinfection efficacy) performed per ISO 17664.
Technological Characteristics
Materials: silicone, nylon elastane, polycarbonate, polybutylene terephthalate, polyurethane, nylon elastane polyurethane laminate. Interface: nasal cushion with tubular conduit frame. Connection: 22mm conical connector (ISO 5356-1). Operating pressure: 4-25 cmH2O. Non-sterile. Reprocessing: High-level thermal disinfection for SLM variant. Complies with ISO 17510, ISO 18562, ISO 10993, and ISO 17664 standards.
Indications for Use
Indicated for patients weighing >66 lb (30 kg) prescribed non-invasive CPAP or bi-level PAP therapy. Whitsundays mask is for single-patient home use; Whitsundays SLM is for multi-patient hospital/institutional use and requires reprocessing between patients.
Regulatory Classification
Identification
A noncontinuous ventilator (intermittent positive pressure breathing-IPPB) is a device intended to deliver intermittently an aerosol to a patient's lungs or to assist a patient's breathing.
{1}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
July 6, 2023
ResMed Pty Ltd (Registration Number: 3004604967) % Sheila Bruschi Senior Director, Regulatory Affairs Resmed Corp (Registration Number: 3007573469) 9001 Spectrum Center Boulevard San Diego, California 92123
Re: K223747
Trade/Device Name: Whitsundays Mask System Regulation Number: 21 CFR 868.5905 Regulation Name: Noncontinuous Ventilator (IPPB) Regulatory Class: Class II Product Code: BZD Dated: December 14, 2022 Received: June 5, 2023
Dear Sheila Bruschi:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal
{2}------------------------------------------------
statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# Rachana Visaria -S
Rachana Visaria, Ph.D. Assistant Director DHT1C: Division of Sleep Disordered Breathing, Respiratory and Anesthesia Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
### Indications for Use
510(k) Number (if known) K223747
Device Name Whitsundays Mask System
Indications for Use (Describe)
The Whitsundays Mask System has two product variants:
• Whitsundays mask
· Whitsundays SLM (Sleep Lab Mask)
Both masks are intended for patients weighing more than 66 lb (30 kg), who have been prescribed non-invasive CPAP or bi-level positive airway pressure (PAP) therapy. The Whitsundays mask is intended for single-patient reuse in the home environment and the Whitsundays SLM is intended for multi-patient reuse in the hospital/institutional environment. The Sleep Lab Mask (SLM) is the only variant that is validated and intended for multi-patient reprocessing and must be reprocessed if reused between patients.
Type of Use (Select one or both, as applicable)X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the ResMed logo. The logo consists of a curved line of dots that transitions from blue to red. Below the dots is the word "ResMed" in a bold, sans-serif font.
# 510(k) Summary
[As required by 21 CFR 807.92(c)]
| Date Prepared: | July 6, 2023 |
|--------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Company Name/Owner: | ResMed Pty Ltd<br>1 Elizabeth Macarthur Drive<br>Bella Vista, NSW, 2153 Australia |
| Prepared and Submitted by: | Ms. Shu-Ying Huang<br>Regulatory Affairs Manager<br>Tel: +65 65727105<br>Shuying.huang@resmed.com.sg |
| Official Contact: | Ms. Sheila Bruschi<br>Senior Director, Regulatory Affairs<br>ResMed Corp.<br>9001 Spectrum Center Blvd<br>San Diego CA 92123 USA<br>Tel +1 858 922 1803<br>sheila.bruschi@resmed.com |
| Device Trade Name: | Whitsundays Mask System |
| Device Common Name: | Vented Nasal Mask |
| Classification and<br>Classification Name: | 21 CFR 868.5905, 73 BZD (Class II)<br>Accessory to Noncontinuous Ventilator (IPPB) |
| Product Code: | BZD |
| Predicate Device: | Scone mask (K180497) |
{5}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for ResMed. The logo consists of a curved line of dots that transition in color from blue to red. Below the dots is the word "ResMed" in a bold, sans-serif font.
#### Device Description:
The Whitsundays Mask System is an accessory to deliver airflow and positive air pressure generated by positive airway pressure (PAP) devices such as CPAP or bi-level flow generator systems to the patient's airway, for the treatment of Obstructive Sleep Apnea (OSA) or ventilatory support (non-life support). It delivers the treatment pressure from the source to the patient's upper airway by providing an air seal between the PAP device and the bottom of the patient's nose.
The Whitsundays mask system has two product variants:
- Whitsundays mask This is the home use variant that is intended for single patient re-use.
- . Whitsundays SLM (Sleep Lab Mask) This is the SLM variant that is intended for multi-patient re-use and must be reprocessed if reused between patients.
The Whitsundays Mask System is made up of 4 main component assemblies: nasal cushion, conduit frame, elbow, and headgear. The nasal cushion and conduit frame are available in various sizes to allow for adequate mask fit in the intended patient population.
The Whitsundays Mask System is a prescription device supplied non- sterile.
#### Indications for Use:
The Whitsundays Mask System has two product variants:
- Whitsundays mask
- . Whitsundays SLM (Sleep Lab Mask)
Both masks are intended for patients weighing more than 66 lb (30 kg), who have been prescribed noninvasive CPAP or bi-level positive airway pressure (PAP) therapy. The Whitsundays mask is intended for single-patient reuse in the home environment and the Whitsundays SLM is intended for multi-patient reuse in the hospital/institutional environment. The Sleep Lab Mask (SLM) is the only variant that is validated and intended for multi-patient reprocessing and must be reprocessed if reused between patients.
{6}------------------------------------------------
Image /page/5/Picture/0 description: The image contains the logo for ResMed. The logo features a curved line of dots that transition in color from blue to red. Below the dots is the word "ResMed" in a bold, sans-serif font.
## Comparison Table:
| Design parameter or<br>feature | Predicate device:<br>Error! Reference source not found.<br>Error! Reference source not found. | Subject device:<br>Error! Reference source not found. | Comments | |
|-------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------|
| Indications for Use | The Scone mask is intended to be<br>used by patients weighing more<br>than 66 lb (30 kg) who have been<br>prescribed non-invasive positive<br>airway pressure (PAP) therapy such<br>as CPAP or bi-level therapy. The<br>mask is intended for single patient<br>re-use in the home and multi-<br>patient re-use in the<br>hospital/institutional environment. | The Whitsundays mask system has<br>two product variants:<br>• Whitsundays mask<br>• Whitsundays SLM (Sleep Lab<br>Mask)<br>Both masks are intended for<br>patients weighing more than 66 lb<br>(30 kg), who have been prescribed<br>non-invasive CPAP or bi-level<br>positive airway pressure (PAP)<br>therapy. The Whitsundays mask is<br>intended for single-patient reuse in<br>the home environment and the<br>Whitsundays SLM is intended for<br>multi-patient reuse in the<br>hospital/institutional environment.<br>The Sleep Lab Mask (SLM) is the<br>only variant that is validated and<br>intended for multi-patient<br>reprocessing and must be<br>reprocessed if reused between<br>patients. | Equivalent | |
| Intended Use | The mask is intended to provide an<br>interface for CPAP or bi-level<br>devices. | The mask is intended to provide an<br>interface for CPAP or bi-level<br>devices. | Identical | |
| FDA Product Code | BZD | BZD | Identical | |
| Patient population | Patients weighing more than 66lb<br>(30kg) for whom positive airway<br>pressure therapy has been<br>prescribed. | Patients weighing more than 66lb<br>(30kg) for whom positive airway<br>pressure therapy has been<br>prescribed. | Identical | |
| Design parameter or<br>feature | Predicate device:<br>Error! Reference source not found.<br>Error! Reference source not found. | Subject device:<br>Error! Reference source not found. | Comments | |
| Environment of Use | Home or hospital/institutional<br>environment. | Home or hospital/institutional<br>environment. | Identical | |
| Reprocessing claims | Single patient re-use or multi-<br>patient re-use. | Single patient re-use or multi-<br>patient re-use. | Identical | |
| Sterility state as provided | Non-sterile | Non-sterile | Identical | |
| Validated reprocessing<br>methods | High-Level Thermal disinfection | High-Level Thermal disinfection | Identical | |
| Materials | Materials include silicone, polycarbonate, polybutylene<br>terephthalate and nylon elastane polyurethane laminate. | Whitsundays mask system has new materials for the nasal cushion and<br>for the sleeve on the the conduit frame.<br>Device materials include silicone, nylon elastane, polycarbonate, polybutylene terephthalate, polyurethane and nylon elastane polyurethane laminate | Equivalent. All of the materials used<br>in the subject device have been<br>included in previously cleared<br>submissions or have undergone<br>biological evaluation as per ISO<br>10993-1 and ISO 18562-1. | |
| Cushion type | Cradles patient's nose and seals<br>under the nose | Cradles patient's nose and seals<br>under the nose | Identical. The cushion is designed to<br>seal in the same way. | |
| Frame type | Tubular conduit | Tubular conduit | Identical. | |
| PAP tubing connection<br>point | 22mm conical connector as per ISO<br>5356-1 over the head connection | 22mm conical connector as per ISO<br>5356-1 over the head connection | Identical. | |
| Exhaust port location | Elbow and cushion | Elbow and cushion | Identical. | |
| Sizes | Cushion available in four sizes<br>Frame available in two sizes<br>Headgear available in one size | Cushion available three sizes<br>Frame available in two sizes<br>Headgear available in one size | Equivalent | |
| Mask exhaust<br>flow (Nominal)<br>ISO 17510:2015<br>Annex B | Pressure<br>(cm<br>H2O) | Flow<br>(L/min)<br>'Pillows' curve | Flow<br>(L/min)<br>'Pillows' curve | |
| | 4 | 20 | 20 | Identical |
| | 9 | 31 | 31 | |
| | 15 | 41 | 41 | |
| | 20 | 49 | 49 | |
| | 25 | 55 | 55 | |
{7}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the logo for ResMed. The logo features a curved line made of dots that transition in color from blue to red. Below the line of dots, the text "ResMed" is written in a bold, sans-serif font.
{8}------------------------------------------------
Image /page/7/Picture/0 description: The image shows the logo for ResMed. The logo consists of the word "ResMed" in a bold, sans-serif font, with the "R" slightly larger than the other letters. Above the wordmark is a curved line made up of a series of dots that transition in color from blue on the left to red on the right. The dots are arranged to create a wave-like pattern.
| Design parameter or<br>feature | Predicate device:<br>Error! Reference source not found.<br>Error! Reference source not found. | Subject device:<br>Error! Reference source not found. | Comments | | | | | | | | | | |
|-----------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------|---------------------------------------------------------|-----------|-------------------------------------------------------------------------------------------------------------------------|-----------------|------------|------------|-----|-----|-------------------------------------------------|--|-----|------------------------------------------------------------------------------------------------------------|
| CO2 re-breathing<br>performance (<20%), ISO<br>17510:2015<br>Annex F | Complies with ISO 17510:2015 CO2<br>requirements (<20%) | Complies with ISO 17510:2015 CO2<br>requirements (<20%) | Identical | | | | | | | | | | |
| Physical Dead space<br>(mL) | Frame size | | SML | STD | Cushion<br>Size | Frame size | | SML | STD | Equivalent. This is a record-only<br>parameter. | | | |
| | Cushion<br>Size | S | 109 | 116 | | SW | 116 | 124 | | | | | |
| | | SW | 105 | 112 | | M | 119 | 127 | | | | | |
| | | M | 113 | 120 | | L | 123 | 130 | | | | | |
| W | | 112 | 119 | | | | | | | | | | |
| Resistance to flow (Pressure<br>drop across mask in cmH2O)<br>ISO 17510:2015<br>Annex C | @50 L/min | | 0.5 | @100 L/min | | 2.3 | @50 L/min | | 0.4 | @100 L/min | | 1.6 | Resistance to flow values is reported<br>in the instructions for use in<br>accordance with ISO 17510:2015. |
| Flow generator setting on<br>compatible ResMed CPAP<br>and Bi-level flow<br>generators. | "Pillows" | | | "Pillows" | | | Identical | | | | | | |
| Operating pressure range<br>(cmH2O) | 4 - 25 | | | 4 - 25 | | | Identical | | | | | | |
| Sound | Sound power level: 21 dBA<br>Sound pressure level: 14 dBA | | | Sound power level: 24 dBA<br>Sound pressure level: 17 dBA | | | Equivalent | | | | | | |
| Operating and storage<br>temperature | Operating temperature: 5°C to 40°C<br>Storage temperature: -20°C to<br>+60°C | | | Operating temperature: 5°C to 40°C<br>Storage temperature: -20°C to<br>+60°C | | | Identical | | | | | | |
| Use life | Scone mask: Visual inspection per<br>instructions for use<br>Scone SLM: 30 validated<br>reprocessing cycles | | | Whitsundays mask: Visual<br>inspection per instructions for use<br>Whitsundays SLM: 30 validated<br>reprocessing cycles | | | Identical | | | | | | |
{9}------------------------------------------------
Image /page/8/Picture/0 description: The image shows the ResMed logo. The logo features a curved line of dots that transition in color from blue to red. Below the dots, the word "ResMed" is written in a bold, italicized font. The logo is simple and modern, with a focus on the company name and a visual element that suggests movement or flow.
#### Non-Clinical Data Submitted:
Non-clinical verification and validation testing completed for the new device demonstrated that Whitsundays Mask System met all intended performance requirements. These included:
Applicable performance and safety tests in accordance with ISO 17510:2015 Medical devices – Sleep apnoea breathing therapy - Masks and application accessories:
- CO2 rebreathing
- Mask exhaust flow .
- Resistance to flow
Other bench tests:
- Physical dead space
- . Pressure accuracy and pressure swing performance
- . Mechanical Integrity of the mask system before and after the following environmental tests:
- Home cleaning
- Transportation and Storage
- . Operation environment
- Free fall and sit test
- Cleaning and Reprocessing
Biocompatibility evaluation was conducted in accordance with ISO 18562-1 and ISO 10993-1. This evaluation was conducted on components that had patient exposure classifications of long-term external communicating device (tissue) and /or long-term skin contact.
Validation of reprocessing claims (in accordance with ISO 17664-1 and ISO 17664-2) included a combination of cleaning efficacy, disinfection efficacy, residual toxicity, and mechanical integrity testing.
Whitsundays mask system was designed and tested in accordance with the applicable requirements in relevant FDA consensus standards including:
| Standards | Title |
|-------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------|
| ISO 17510:2015 | Medical devices -- Sleep apnoea breathing therapy -- Masks and application accessories |
| ISO 18562-1:2017 | Biocompatibility evaluation of breathing gas pathways in healthcare applications -- Part 1: Evaluation and testing within a risk management process |
| ISO 18562-2:2017 | Biocompatibility evaluation of breathing gas pathways in healthcare applications – Part 2: Tests for emissions of particulate matter |
| ISO 18562-3:2017 | Biocompatibility evaluation of breathing gas pathways in healthcare applications – Part 3: Tests for emissions of volatile organic compounds (VOCs) |
| ISO 18562-4:2017 | Biocompatibility evaluation of breathing gas pathways in healthcare applications – Part 4: Tests for leachables in condensate |
| ISO 10993-1:2018 | Biological evaluation of medical devices -- Part 1: Evaluation and testing within a risk management process |
| ISO 10993-5:2009 | Biological evaluation of medical devices – Part 5: Test for in vitro cytotoxicity |
| ISO 10993-10:2021 | Biological evaluation of medical devices – Part 10: Tests for skin sensitization |
| ISO 10993-17:2002 | Biological evaluation of medical devices – Part 17: Establishment of allowable limits for leachable substances |
| ISO 10993-18:2021 | Biological evaluation of medical devices – Part 18: Chemical characterization of materials |
{10}------------------------------------------------
Image /page/9/Picture/0 description: The image shows the logo for ResMed. The logo consists of a curved line made up of a series of dots that transition in color from blue to purple to red. Below the line of dots is the word "ResMed" in a bold, sans-serif font.
| ISO 17664-1:2021 | Processing of health care products - Information to be provided by the medical device<br>manufacturer for the processing of medical devices - Part 1: Critical and semi-critical<br>medical devices |
|------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| ISO17664-2:2021 | Processing of health care products - Information to be provided by the medical device<br>manufacturer for the processing of medical devices - Part 2: Non-critical medical<br>devices. |
| ISO 5356-1:2015 | Anaesthetic and respiratory equipment — Conical connectors — Part 1: Cones and<br>sockets |
| ISO 15223-1:2021 | Medical devices - Symbols to be used with information to be supplied by the<br>manufacturer - Part 1: General requirements |
| ISO 20417:2021 | Medical devices — Information to be supplied by the manufacturer |
### Substantial Equivalence Conclusion:
The Whitsundays mask system is substantially equivalent to the predicate Scone mask (K180497):
- It has equivalent intended use .
- It has similar technological characteristics
- It has similar performance characteristics ●
- . The differences do not raise any new questions of safety or effectiveness
- It is as safe and as effective as the predicate device
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.