CARTO® 3 EP Navigation System with CARTOSOUND™ 4D Version 7.4
K223733 · Biosense Webster, Inc. · DQK · Feb 16, 2023 · Cardiovascular
Device Facts
Record ID
K223733
Device Name
CARTO® 3 EP Navigation System with CARTOSOUND™ 4D Version 7.4
Applicant
Biosense Webster, Inc.
Product Code
DQK · Cardiovascular
Decision Date
Feb 16, 2023
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.1425
Device Class
Class 2
Indications for Use
The intended use of the CARTO® 3 System is catheter-based cardiac electrophysiological (EP) procedures. The CARTO® 3 System provides information about the electrical activity of the heart and about catheter location during the procedure. The system can be used on patients who are eligible for a conventional electrophysiological procedure. The system has no special contraindications.
Device Story
System performs catheter-based atrial/ventricular mapping; acquires intracardiac ECG signals and catheter location data via magnetic and Advanced Catheter Location (ACL) technologies. Integrates real-time 3D ultrasound data from compatible catheters (e.g., NUVISION NAV) and ultrasound systems (e.g., GE Vivid S70N). Operates in clinical EP labs; used by physicians. Transforms inputs into 3D anatomical and electroanatomical maps; displays 4D volumetric ultrasound reconstructions. Provides real-time visualization of anatomy and catheter position; supports contouring of anatomical structures on ultrasound clips. Output assists physicians in navigating catheters and identifying cardiac electrical activity; facilitates clinical decision-making during ablation or diagnostic procedures. Benefits include enhanced anatomical visualization and procedural guidance.
Clinical Evidence
No human clinical data provided. Evidence consists of bench testing (Proof of Design, functional verification, image quality assessment) and animal testing under simulated clinical conditions. Summative usability testing validated the CARTOSOUND™ 4D module interface. All tests met pre-defined acceptance criteria.
Technological Characteristics
Cardiac mapping system using magnetic and ACL location sensing. Hardware includes PIU, workstation, monitors, signal processing unit, and location pad. Connectivity via Ethernet (LAN) for digital video (H.264) and legacy analog (VGA) support. Software version 7.4. SPU includes ASIC amplifiers and A/D converters. FPGA firmware manages signal calibration. TPI card hardware optimized for noise reduction.
Indications for Use
Indicated for patients eligible for conventional catheter-based cardiac electrophysiological (EP) procedures. No specific age or gender restrictions; no special contraindications.
Regulatory Classification
Identification
A programmable diagnostic computer is a device that can be programmed to compute various physiologic or blood flow parameters based on the output from one or more electrodes, transducers, or measuring devices; this device includes any associated commercially supplied programs.
Predicate Devices
CARTO® 3 EP Navigation System Version 7.2 (K213264)
GE Healthcare's Vivid S70N Ultrasound System (K223832)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
February 16, 2023
Biosense Webster, Inc. Phuong Park Associate Director. Regulatory Affairs 31 Technology Drive, Suite 200 Irvine, California 92618
Re: K223733
Trade/Device Name: CARTO® 3 EP Navigation System with CARTOSOUND 4D version 7.4 Regulation Number: 21 CFR 870.1425 Regulation Name: Programmable Diagnostic Computer Regulatory Class: Class II Product Code: DOK Dated: December 13, 2022 Received: December 13, 2022
Dear Phuong Park:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal
{1}------------------------------------------------
statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Aneesh S. Deoras -S
Aneesh Deoras Assistant Director Division of Cardiac Electrophysiology, Diagnostics and Monitoring Devices Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K223733
Device Name
CARTO® 3 EP Navigation System with CARTOSOUND 4D version 7.4
Indications for Use (Describe)
The intended use of the CARTO® 3 System is catheter-based cardiac electrophysiological (EP) procedures. The CARTO® 3 System provides information about the electrical activity of the heart and about catheter location during the procedure. The system can be used on patients who are eligible for a conventional electrophysiological procedure. The system has no special contraindications.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|---------------------------------------------|
| × Prescription Use (Part 21 CFR 801 Subpart D) | Over-The-Counter Use (21 CFR 801 Subpart C) |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(K) SUMMARY
| Applicant: | Biosense Webster, Inc.<br>31 Technology Drive, Suite 200<br>Irvine, CA 92618, USA<br>Tel.: (800) 729-9010<br>Fax: (909) 839-8500 |
|---------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact Person: | Phuong Park<br>Associate Director, Regulatory Affairs<br>Phone: 949-923-4238<br>Fax: 949-450-6886 |
| Authored by: | Eyal Hendel<br>Regulatory Affairs Manager<br>Biosense Webster (Israel), Ltd.<br>and<br>Phuong Park<br>Associate Director, Regulatory Affairs<br>Biosense Webster, Inc. |
| Date: | February 15, 2023 |
| Device Trade<br>Name: | CARTO® 3 EP Navigation System with CARTOSOUND 4D version 7.4 |
| Device Common<br>Name: | Cardiac Mapping System |
| Manufacturing<br>Number: | FG-5400-00 |
| Device<br>Classification: | Programmable diagnostic computer |
| | Class II, 21 CFR 870.1425 |
| Product Code | DQK |
| Predicate Device: | CARTO® 3 EP Navigation System Version 7.2 510(k)#: K213264 |
{4}------------------------------------------------
| Manufacturing | Biosense Webster (Israel), Ltd. |
|---------------|---------------------------------|
| Facilities: | a Johnson & Johnson Company |
| | 4 Hatnufa Street |
| | Yokneam. ISRAEL 2066717 |
Biosense Webster, Inc. 15715 Arrow Hwy Irwindale, CA 91706 USA
Device Description:
Image /page/4/Picture/3 description: The image shows a medical device on a wheeled stand. The device has a monitor at the top displaying medical images. Below the monitor is a control panel with various buttons and ports. The device is white and has a modern design. The stand has four wheels for easy mobility.
CARTO® The 3 EP Navigation System with CARTOSOUND™ 4D, software V7.4, is a catheter-based atrial and ventricular mapping system designed to acquire and analyze navigation catheter's location and intracardiac ECG signals and use this information to display 3D anatomical and electroanatomical maps of the human heart. The location information needed to create the cardiac maps and the local electrograms are acquired using specialized mapping catheters and reference devices. The CARTO® 3 System uses two distinct types of location technology – magnetic sensor technology and Advanced Catheter Location (ACL) technology.
The CARTO® 3 System with CARTOSOUND™ 4D consists of the following hardware components:
- . Patient Interface Unit (PIU)
- Workstation with Graphic User Interface (GUI)
- . Wide-Screen monitors, keyboard, and mouse
- . Intracardiac In Port
- Intracardiac Out Port ●
- . Power Supply
- Patches Connection Box and Cables (PU)
- Pedals ●
- Location Pad (LP) ●
- Signal Processing Unit (SPU) ●
All hardware components of the CARTO® 3 system with CARTOSOUND™ 4D are the same to those found in the predicate device.
#### The intended use of the CARTO® 3 System is catheter-based cardiac Indications for Use: electrophysiological (EP) procedures. The CARTO® 3 System provides information about the electrical activity of the heart and about catheter location
{5}------------------------------------------------
during the procedure. The system can be used on patients who are eligible for a conventional electrophysiological procedure.
The system has no special contraindications.
The indications for use for the CARTO® 3 System with CARTOSOUND™ 4D are identical to the indications for use of the predicate device, the CARTO® 3 System V7.2.
The modified CARTO® 3 EP Navigation System with CARTOSOUND™ 4D, Technological Characteristics: software V7.4 has the same technological characteristics (i.e., design, material, chemical composition, energy source) as the predicate CARTO® 3 EP Navigation System. Version 7.2 (K213264). A summary of the technological characteristics of the new device compared to the predicate device is as follows:
- . Have identical intended use.
- Use the same fundamental scientific technology.
- . Have the same hardware platform
- . Have identical magnetic and ACL location mapping technology.
- . Have identical magnetic location sensor and ACL location accuracy.
The differences between the predicate device and the modified device are the addition of a new software feature, CARTOSOUND™ 4D, improvements of the legacy CARTOSOUND™ Module and the addition of Letter to File changes as described below.
Note: The CARTOSOUND™ module was cleared with the CARTO® 3 System V1, K090017.
#### Description of Modifications
#### CARTOSOUND™ 4D
The addition of the CARTOSOUND™ 4D Module combines 3D real-time ultrasound image data with CARTO® 3 System Maps. 4D (real-time volumetric imaging with a field of view of up to 90° x 90°) multi-planar, and 2D ultrasound images are acquired with the compatible NUVISION™ NAV Ultrasound Catheter (K223766). The catheter includes an ultrasound transducer and a location sensor. The NUVISION™ Nav catheter has a bifurcated 'tail' originating from the handle. One part terminates in a connector receptacle which mates with a NUVISION Connector Cable to connect the catheter to a compatible ultrasound system. The other part enables connection to the CARTO® 3 Navigation System with CARTOSOUND™ 4D Module via an extension cable. A steering mechanism controls orientation of the image plane by rotating both the NUVISION Nav Catheter's tip and providing variable bi-directional deflection. The real-time ultrasound images are displayed in the CARTO® 3 System and the compatible GE
{6}------------------------------------------------
Healthcare's Vivid S70N Ultrasound System (Vivid S70N, SW version 206.zz, K223832), which is specialized for use with cardiac images.
This module also enables acquisition of clips from the ultrasound system, transferred with a digital communication protocol over an Ethernet connection (LAN) cable. The user can draw contours of the required anatomy on any frame of the ultrasound clip and add this information to a 3D anatomical map reconstruction. The ultrasound clip can be saved with these contours, and the frames containing the contours are marked for easy identification. The ultrasound clip can be also stored to the CARTO® 3 system for future contouring. Contour data contributes anatomical information to the map. Contours and Ultrasound clips are listed in the Map Point List.
The CARTOSOUND™ 4D Module 4D advanced feature presents a 4D rendered reconstruction on the CARTO main map viewer and additional windows in a realtime fashion.
#### CARTOSOUND™ Module Improvements
Improvements were implemented to the user interface of the legacy CARTOSOUND™ module while maintaining the same functionality. The user interface change includes layout and controls for assigning contours to maps and an updated toolbar.
#### Video Over LAN
Improvement was implemented to the CARTO® 3 System, by replacing the Analog interface with a digital communication protocol over an Ethernet connection (LAN) cable with the compatible ultrasound system, by enabling transmitting compressed videos from the GE Healthcare's Vivid S70N Ultrasound System (Vivid S70N, SW version 206.z) to CARTO® 3 System with CARTOSOUND™ 4D using industrial standard compressing protocol H.264 format.
Note: To maintain compatibility with all other Ultrasound Systems, the CARTO® 3 System with CARTOSOUND™ 4D also supports the legacy solution in which data is sent in an analog format (typically VGA) and images are grabbed via a Frame Grabber in the CARTO® 3 System workstation.
#### Letter to File changes
Improvement to OCTARAY DX catheter support through Signal Processing Unit
- Updated coefficients of the Dynamic Threshold (Catheter XML Change) ● to receive the optimal S2 electrode detection in/out of the sheath
Signal Processing Unit (SPU) enhancements
{7}------------------------------------------------
- . Improvement to SPU Build in Test (BIT) mechanism to run the BIT only at SPU startup, to avoid possible delays during the clinical procedure.
- Tissue Proximity Indicator (TPI) Card update: ●
- . HW change: removal of non-functional amplifiers in the TPI card, to decrease the total current (load and self-consumption) of the -5 Volt voltage supply, assembled on the Power Card, in order to reduce the possibility of saturation in extreme conditions.
- . HW change: replacement of 4 resistors in the TPI card to 0 [ohm], to reduce open loop noise due to the removal of the amplifiers. Note: these TPI changes have no impact on the TPI performance.
- Firmware update: ●
- . FPGA Firmware update: updating the calibration multiplier factor with an overflow protected multiplier to address an observed ECG visualization artifact.
Note: The SPU has an analog amplifier (ASIC) followed by an A/D converter and a calibration multiplier. When a full-scale signal is set to the SPU, it may seldom cause an overflow at the multiplier and an ECG visualization artifact
Security Enhancements
- Secure login mode is disabled by default, although selectable by the user ●
- Security and drivers update .
## Support removal
- . The support for the Siemens Sequoia Ultrasound system was removed due to end of life of the ULS system.
The CARTO® 3 EP Navigation System with CARTOSOUND™ 4D underwent Performance Data: verification and validation testing under simulated clinical conditions to verify the new features and to demonstrate with regression testing that the modifications performed did not negatively affect existing features.
## Bench Testing:
Bench testing completed for CARTO® 3 System with CARTOSOUND™ 4D included:
- Proof of Design (POD) - Testing was performed to verify the CARTO® 3 System specifications while connecting it to the NUVISIONT™ NAV Ultrasound catheter and GE Healthcare's Vivid S70N Ultrasound System (Vivid S70N, SW version 206.z). All testing performed were successfully completed and met the acceptance criteria.
{8}------------------------------------------------
- Functional verification Testing was performed to verify the functional ● requirements of CARTO® 3 System with CARTOSOUND™ 4D, including testing of the new CARTOSOUND™ 4D features and CARTOSOUND™ improvements as well as regression testing to verify continued functionality of CARTO® 3 System legacy features and Letter To File modifications. All system features were found to perform according to specifications and met the tests acceptance criteria.
- Image Quality assessment Testing was conducted to evaluate the ULS . image quality as displayed in the monitor of the CARTO® 3 System with CARTOSOUND™ 4D. The test was conducted under simulated clinical workflow and conditions. Assessment was successfully completed and expected results were achieved.
### Animal Testing:
Animal testing was conducted to evaluate the CARTO® 3 System with CARTOSOUND™ 4D functionality under simulated clinical workflow and conditions. All test protocol steps were successfully completed and expected results were achieved.
All testing passed in accordance with appropriate test criteria and standards, and the modified device did not raise new questions of safety or effectiveness.
#### Usability Testing:
Summative Usability testing was conducted to validate that CARTOSOUND™ 4D Module User Interface ease of use. In light of the usability validation study results, it was concluded that the operation of the CARTOSOUND™ 4D Module has been found to be safe and effective for the intended users, uses and use environments.
- The CARTO® 3 EP Navigation System with CARTOSOUND™ 4D software Conclusions: version V7.4 is substantially equivalent to the currently cleared CARTO® 3 EP Navigation System. Version 7.2 based on the completion of verification and validation testing.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.