The NEBA® Compact EEG 2R Mobile Headset) is intended to capture, amplify, and wirelessly transmit electrical brain activity for review by a trained medical professional using the CEEG2R recording software. The NEBA Headset comprises an electrode positioning system for placing single-use disposable electrodes on the head (Vermed® A10005, ANSI/AAMI EC12 compliant silver/silver-chloride (Ag/AgCl) electrodes). Contained within the headset is a wireless EEG amplifier module. The NEBA Headset and its associated software do not provide any diagnostic conclusion or automated alerts of any clinical event about a patient's condition. The NEBA Headset transmits electrophysiological signals from the electrodes to a CEEG 2R IEC/UL 60950-1 safety compliant computer running the NEBA CEEG 2R recording software. The system supports six electrode locations (CZ, left ocular, right ocular, left ear, right ear, and ground) on four flat flexible leads and one plastic tab (the latter to support the CZ electrode). The CEEG2 Headset is intended for use in clinical settings in individuals six years of age and older. (Rx only).
Device Story
Battery-powered wireless EEG headset; captures electrical brain activity via 6 disposable Ag/AgCl electrodes; uses Bluetooth LE to transmit signals to a computer running CEEG2R recording software. Headset features flexible platform and adjustable arms for 10-20 International Electrode Placement System positioning. Used in clinical settings by trained medical professionals; provides raw EEG data for review. No diagnostic or automated alert capabilities. Benefits include simplified electrode placement and wireless signal transmission for professional monitoring. Device is reusable; electrodes are single-use.
Clinical Evidence
No clinical data. Substantial equivalence demonstrated via bench testing, including electrical safety (IEC 60601-1), electromagnetic compatibility (IEC 60601-1-2), EEG-specific safety (IEC 60601-2-26), wireless emissions/exposure (47 CFR Part 15), and biocompatibility (ISO 10993-5).
Technological Characteristics
Battery-powered (3.7V Li-ion) wireless EEG headset; Qi-compliant wireless charging. Materials: polyurethane and polyester. Sensing: 6 passive Ag/AgCl electrodes (ANSI/AAMI EC12). Connectivity: Bluetooth LE. Software: Integrated firmware for signal transmission; CEEG2R recording software for monitoring. Biocompatibility per ISO 10993-1/5/10.
Indications for Use
Indicated for individuals 6 years of age and older in clinical settings to capture, amplify, and wirelessly transmit electrical brain activity for review by a trained medical professional. Not indicated for diagnostic conclusions or automated clinical alerts.
Regulatory Classification
Identification
An electroencephalograph is a device used to measure and record the electrical activity of the patient's brain obtained by placing two or more electrodes on the head.
{1}------------------------------------------------
March 2, 2023
Image /page/0/Picture/1 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left side of the logo is the Department of Health & Human Services logo. To the right of that is a blue square with the letters FDA in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
Neba Health, LLC Howard Merry President 2052 Gordon Highway, Ste B. Augusta. Georgia 30909
Re: K223628
Trade/Device Name: NEBA® Compact EEG2R Mobile Headset Regulation Number: 21 CFR 882.1400 Regulation Name: Electroencephalograph Regulatory Class: Class II Product Code: OMC, GXY Dated: October 26, 2022 Received: December 5, 2022
Dear Howard Merry:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part
{2}------------------------------------------------
801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4. Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# Patrick Antkowiak -S
for Jay Gupta Assistant Director DHT5A: Division of Neurosurgical, Neurointerventional and Neurodiagnostic Devices OHT5: Office of Neurological and Physical Medicine Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
# Indications for Use
510(k) Number (if known) K223628
Device Name NEBA® Compact EEG2R Mobile Headset
#### Indications for Use (Describe)
The NEBA® Compact EEG 2R Mobile Headset) is intended to capture, amplify, and wirelessly transmit electrical brain activity for review by a trained medical professional using the CEEG2R recording software. The NEBA Headset comprises an electrode positioning system for placing single-use disposable electrodes on the head (Vermed® A10005, ANSI/AAMI EC12 compliant silver/silver-chloride (Ag/AgCl) electrodes).
Contained within the headset is a wireless EEG amplifier module.
The NEBA Headset and its associated software do not provide any diagnostic conclusion or automated alerts of any clinical event about a patient's condition.
The NEBA Headset transmits electrophysiological signals from the electrodes to a CEEG 2R IEC/UL 60950-1 safety compliant computer running the NEBA CEEG 2R recording software.
The system supports six electrode locations (CZ, left ocular, left ear, right ear, and ground) on four flat flexible leads and one plastic tab (the latter to support the CZ electrode).
The CEEG2 Headset is intended for use in clinical settings in individuals six years of age and older.
| Type of Use (Select one or both, as applicable) | |
|----------------------------------------------------------------------------------|--------------------------------------------------------------------------------|
| <div> <span> Prescription Use (Part 21 CFR 801 Subpart D) </span> </div> | <div> <span> Over-The-Counter Use (21 CFR 801 Subpart C) </span> </div> |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
# K223628 – Traditional 510(k) Summary
| Submitter: | NEBA Health, LLC<br>2052B Gordon Highway<br>Augusta, GA 30909<br>NEBA Health, LLC.<br>Phone: (706) 736-5864<br>Fax: (706) 650-2160 |
|--------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact person: | E. Howard Merry,<br>President NEBA Health<br>hmerry@NEBAHealth.net |
| Date summary prepared: | February 22, 2023 |
| Subject device | |
| Device trade name; proprietary name: | NEBA® Compact EEG2R Mobile Headset (CEEG2R) |
| Common name: | Electroencephalogram (EEG) |
| Product Code; Classification: | OMC (Primary); Reduced montage<br>electroencephalograph 21 CFR 882.1400;<br>Electroencephalograph, Class II.<br>GXY (Secondary); Cutaneous Electrode<br>21 CFR 882.1320; Class II; Cutaneous Electrode |
| Review Panel: | Neurology; Diagnostic Devices |
#### Predicate and reference devices
| Predicate Type | 510K Number | Device | Manufacturer |
|---------------------|-------------|---------------------|--------------|
| Primary Predicate | K203827 | REMI | Epitel, Inc. |
| Secondary Predicate | K183529 | AE-120A EEG Headset | Nihon Kohden |
{5}------------------------------------------------
#### Subject device description
The NEBA® Compact EEG2R Mobile Headset is a battery powered (3.7v Lithium-lon battery wirelessly charged using a Qi-compliant receiver) wireless (Bluetooth LE) EEG headset which facilitates the placement of EEG electrodes. The NEBA® Compact EEG2R Mobile Headset (CEG2R) has both hardware and software components. Hardware comprises an EEG electrode system that serves to conduct EEG potentials from the human scalp for transfer to a built-in wireless EEG amplifier. The software provides EEG amplifier hardware control, recording, storage, and user interfaces for waveform monitoring, patient information entry and storage, and accessing stored EEG data.
The CEEG2R Headset has two primary purposes:
- . To aid in EEG electrode positioning on the head such that electrodes are positioned in intended locations accurately and reliably per the standard 10-20 International Electrode Placement System. The default leads configuration includes CZ, left ocular (OC-L), right ocular (OC-R), left ear (E-L), right ear (E-R), and Ground (GND) positions;
- . To transmit electrophysiological signals from positioned electrodes to an EEG recording and monitoring device via an internal EEG amplifier with wireless transfer communication.
The CEEG2R Headset is constructed using biocompatible patient contact surfaces, the major components formed of polyurethane and polyester. The headset is held in place using a flexible platform at the top of the head and adjustable arms that terminate at lateral supports at the sides of the head.
The headset is designed for use with accessory disposable silver/silver-chloride (Ag/AgCl) electrodes (K781430).
An integral, counter posing pressure tab accepts a disposable electrode for placement at the top medial portion of the head (at location CZ of the 10-20 International Electrode Placement System). Connection components facilitate the quick insertion and removal of the electrodes and flexible leads. Electrodes for use on the scalp and hair bearing scalp region use a low-viscosity integrated wet-gel conductive medium embedded in the electrodes along with integrated low-tack adhesive optimized for hair compatibility and system stability.
The system is stabilized on the head through use of low tack adhesive-lined electrodes, conformable lateral support linings, spring hinges, and a flexible headband platform.
The system interfaces with the CEG2R recording and monitoring software for signal acquisition, signal measurement, and electrode impedance measurement by way of a built-in wireless amplifier. The battery, wireless battery charger, wireless communication transmitter and amplifier printed circuit board (PCB) are located atop a midline-located headband platform of the headset.
#### Subject device Intended Use
The NEBA® Compact EEG 2R Mobile Headset) is intended to capture, amplify, and wirelessly transmit electrical brain activity for review by a trained medical professional using the CEEG2R recording software. The NEBA Headset comprises an electrode positioning system for placing single-use disposable electrodes on the head (Vermed® A10005, ANSI/AAMI EC12 compliant silver/silver-chloride
{6}------------------------------------------------
(Ag/AgCl) electrodes cleared in K781430). Contained within the headset is a wireless EEG amplifier module. The NEBA Headset and its associated software do not provide any diagnostic conclusion or automated alerts of any clinical event about a patient's condition. The NEBA Headset transmits electrophysiological signals from the electrodes to a CEEG 2R IEC/UL 60950-1 safety compliant computer running the NEBA CEEG 2R recording software. The system supports six electrode locations (CZ, left ocular, right ocular, left ear, right ear, and ground) on four flat flexible leads and one plastic tab (the latter to support the CZ electrode). The CEEG2 Headset is intended for use in clinical settings in individuals six years of age and older. (Rx only).
### Comparison of predicate devices to subject device
The NEBA® Compact EEG2R Mobile Headset (CEEG2R) is substantially equivalent to the REMI device when combined with the Nihon Kohden AE-120A EEG headset. NEBA shares the same technological characteristics as the predicate devices, but there are some differences. Both battery-powered predicates are reduced montage EEG devices that wirelessly transmit EEG data to an accessory device running software.
The primary predicate is the REMI device. The NEBA® Compact EEG2R Mobile Headset (CEG2R) is designed to amplify, capture, and wirelessly transmit EEG data from the CZ channel to an accessory software intended for monitoring and reviewing EEG data. Similarly, the REM device is called out as the primary predicate as it is intended to wirelessly transmit a single channel of electrical activity for subsequent review. Both are intended for professional prescription use only and for patients 6 years and older.
Neither the REMI device nor the NEBA® Compact EEG2R Mobile Headset are indicated for making any diagnosis or diagnostic recommendations and both are intended only to facilitate the monitoring and review of physiological signals.
The secondary predicate, the Nihon Kohden AE-120A, is called out to achieve the NEBA® Compact EEG2R Mobile Headset positionality. Both the Nihon Kohden AE-120A and the NEBA® Compact EEG2R Mobile Headset use mechanical headsets to aid in EEG electrode positioning to approximate the 10-20 International Electrode Placement System.
The predicate systems and the NEBA® Compact EEG2R Mobile Headset were evaluated for safety and effectiveness in quality systems per FDA guidance and using compliance standards. The predicate systems and the NEBA® Compact EEG2R Mobile Headset were evaluated according to ISO 10993-1 biological evaluation and associated FDA guidance.
Given the discussion above and the substantial equivalence table below, NEBA Health asserts that the differences between the NEBA® Compact EEG2R Mobile Headset and the predicate devices do not raise different questions of safety and effectiveness based on the performance testing used to address the technological characteristics.
NEBA Health asserts the NEBA Health device (CEEG2R) is substantially equivalent to the combined predicate devices.
{7}------------------------------------------------
Table 5a compares the technical characteristics of the subject device to the primary and secondary predicate devices.
#### Table 5a – Technical Characteristics Comparison
| Attribute | REMI | NKC | NEBA CEEG 2R<br>mobile Headset | Discussion |
|----------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| classification<br>regulation | K203827 | K183529 | K223628 | Same. |
| | Class II per 21 CFR<br>882.1400<br>Electroencephalograph (Head<br>Set) | Class II per 21 CFR<br>882.1400<br>Electroencephalograph (Head<br>Set) Class II per 21 CFR<br>882.1320 (for<br>electrodes within<br>headset) | Class II per 21 CFR<br>882.1400<br>Electroencephalograph (Head<br>Set) Class II per 21 CFR<br>882.1320 (for<br>electrodes within<br>headset) | |
| Product code | OMC / GXY | OMC/GXY | OMC/GXY | Same. |
| IFU | the REMI Platform is<br>intended to be used in<br>healthcare settings<br>where<br>near real-time and/or<br>remote EEG is<br>warranted. REMI<br>consists of epilog<br>disposable Sensors a<br>single use, single<br>patient, disposable,<br>wearable sensor<br>intended to amplify,<br>capture, and<br>wirelessly transmit a<br>single<br>channel of electrical<br>activity of the brain<br>for up to 48 hours.<br>The REMI-Mobile<br>software and REMI-<br>Tablet are intended to<br>receive and transmit<br>data from four Epilog<br>Sensors to secure<br>cloud storage for<br>subsequent viewing<br>and reviewing of EEG<br>on third-party<br>software. REMI does<br>not make any | The AE-120A EEG<br>Head Set is intended<br>to amplify, capture,<br>and wirelessly<br>transmit electrical<br>activity of the brain<br>for review by a<br>trained medical<br>professional using the<br>previously cleared and<br>validated Nihon<br>Kohden<br>electroencephalograph systems (EEG-1200A<br>series and EEG-9100)<br>to assist in the<br>diagnosis of<br>neurological<br>disorders. The AE-<br>120A EEG Head Set<br>and its associated EEG<br>Software do not<br>provide any diagnostic<br>conclusion or<br>automated alerts of<br>an adverse clinical<br>event about a<br>patient's condition.<br>The device is intended<br>for use by trained<br>medical professionals | The NEBA® Compact<br>EEG 2R Mobile<br>Headset (NEBA<br>Headset) is intended<br>to capture, amplify,<br>and wirelessly<br>transmit electrical<br>brain activity for<br>review by a trained<br>medical professional<br>using the CEEG2R<br>recording software.<br>The NEBA Headset<br>comprises an<br>electrode positioning<br>system for placing<br>single-use disposable<br>electrodes on the<br>head (Vermed®<br>A10005, ANSI/AAMI<br>EC12 compliant<br>silver/silver-chloride<br>(Ag/AgCl) electrodes).<br>Contained within the<br>headset is a wireless<br>EEG amplifier module.<br><br>The NEBA Headset<br>and its associated<br>software do not<br>provide any diagnostic | The NEBA Headset<br>and the predicate<br>devices are all<br>reduced montage<br>EEGs with the<br>intended use to<br>amplify, record and<br>wirelessly transmit<br>EEG data to<br>accessories running<br>software for<br>monitoring and<br>reviewing EEG data.<br>All 3 devices are for<br>prescription use only<br>by professionals. All<br>devices have similar<br>intended use<br>environments. The<br>primary predicate, the<br>REMI device and the<br>NEBA Headset have<br>the same population<br>age range. The Nihon<br>Kohden device and<br>the NEBA headset<br>have in common an<br>EEG positioning<br>system. |
| Attribute | REMI | NKC | NEBA CEEG 2R mobile Headset | Discussion |
| | | | | |
| | K203827 | K183529 | K223628 | |
| | diagnosis or<br>recommendations<br>and is intended only<br>as a physiological<br>signal monitor. Epilog<br>Sensors are intended<br>for use by trained<br>medical professionals<br>in a professional<br>healthcare facility<br>environment. Epilog<br>Sensors are intended<br>for use with adult and<br>pediatric patients<br>(6+). (Rx only). | in a medical facility<br>such as a physician's<br>office, laboratory, or<br>clinic. The device is<br>intended for use on<br>adults (ages 18 and<br>above). (Rx Only). | conclusion or<br>automated alerts of a<br>clinical event about a<br>patient's condition.<br><br>The NEBA Headset<br>transmits<br>electrophysiological<br>signals from the<br>electrodes to a CEEG<br>2R IEC/UL 60950-1<br>safety compliant<br>computer running the<br>NEBA CEEG 2R<br>recording software.<br><br>The system supports<br>six electrode locations<br>(CZ, left ocular, right<br>ocular, left ear, right<br>ear, and ground) on<br>four flat flexible leads<br>and one plastic tab<br>(the latter to support<br>the CZ electrode).<br><br>The CEEG2 Headset is<br>intended for use in<br>clinical settings in<br>individuals six years of<br>age and older. (Rx<br>only). | |
| Physiological signal<br>acquired | EEG | EEG | EEG | Same. |
| Type of patient<br>contact | Contacts patient scalp | contacts patient scalp | Contacts patient scalp | Same. |
| electrodes | 2 passive gold<br>electrodes using a<br>conductive hydrogel | 10 passive Ag/AgCl<br>electrodes | 6 passive Ag/AgCl<br>electrodes | The electrodes for the<br>Nihon Kohden device<br>and the NEBA Headset<br>are the same. The<br>disposable electrode<br>for the NEBA Headset<br>is 510(k) approved<br>and ANSI/AAMI EC12<br>compliant and is<br>manufactured by a |
| Attribute | REMI | NKC | NEBA CEEG 2R<br>mobile Headset | Discussion |
| | K203827 | K183529 | K223628 | NEBA approved<br>vendor. |
| type of use | Epilog Disposable is<br>single use, non-sterile,<br>disposable | Electrodes: Single use,<br>non-sterile,<br>disposable. EEG<br>Device: AE-120A EEG<br>Head set is reusable<br>and<br>nonpatient<br>contacting. Belts are<br>patient contacting<br>and reusable. | Electrodes: Single use,<br>non-sterile,<br>disposable. The NEBA<br>Headset is reusable<br>and comprises both<br>patient and non-<br>patient contacting<br>parts. The headset is<br>reusable and designed<br>for cleaning between<br>uses. | The intended use of<br>the Nihon Kohden<br>device and the NEBA<br>device are similar and<br>differences do not<br>raise questions of<br>safety and<br>effectiveness. |
| Channel # | 0-10 | 8 | 1 | Same. The NEBA<br>Headset total EEG<br>channel number is<br>within the range<br>established by the<br>predicates. |
| Montage | 10/20 system - Epilog-<br>Disposable can be<br>placed anywhere in<br>the 10/20 system<br>where each channel<br>represents a bipolar<br>derivation<br>approximation of the<br>10/20 system | 10/20 System - AE-<br>120A EEG Head Set<br>approximates the<br>10/20 montage but<br>may deviate slightly<br>depending<br>on the patient's head<br>shape. | 10/20 System - the<br>NEBA Headset is<br>intended to facilitate<br>the placement of<br>electrodes per the<br>10/20 montage | Same. |
| Electrical safety and<br>EMC | IEC 60601-1,<br>IEC 60601-1-2,<br>IEC 60601-2-26 | IEC 60601-1,<br>IEC 60601-1-2,<br>IEC 60601-2-26 | IEC 60601-1,<br>IEC 60601-1-2,<br>IEC 60601-2-26 | Same. |
| Input dynamic range | 1 mVp-p | 1 mVp-p | ±187.5 mV peak-to-<br>peak input | The difference<br>between the<br>predicate devices and<br>the CEEG2R device for<br>the Input Dynamic<br>Range is negligible<br>when compared to<br>EEG signals in the low<br>micro-volt range and<br>does not affect safety<br>and effectiveness. |
| Attribute | REMI | NKC | NEBA CEEG 2R<br>mobile Headset | Discussion |
| | K203827 | K183529 | K223628 | |
| input noise | 5 µVp-p | 5 µVp-p or less (0.53<br>to 60<br>Hz) | ± 4 µV maximum<br>peak-peak 0.5 Hz<br>through 50 Hz a noise<br>limit | Same. |
| data transfer | Bluetooth 2.4 GHz | Bluetooth 2.4 GHz | Bluetooth LE 2.4 GHz | Same. |
| power source | CR2016 primary<br>lithium (not<br>rechar…
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.