Ranfac Lateral Access Bone Marrow Aspiration Needle (CRVS-BMA-LA)
Applicant
Ranfac Corporation
Product Code
KNW · Gastroenterology, Urology
Decision Date
Feb 21, 2023
Decision
SESE
Submission Type
Special
Regulation
21 CFR 876.1075
Device Class
Class 2
Indications for Use
The Lateral Access Bone Marrow Aspiration Needle is intended for use for aspiration of bone marrow or autologous blood using a standard piston syringe.
Device Story
Manual, sterile, single-use bone marrow aspiration needle; used for harvesting bone marrow or autologous blood. Device comprises access needle (7-gauge outer/11-gauge inner cannula) and 14-gauge aspiration cannula. Access needle features adjustable grip and blunt tip; inner cannula has ground point and side ports. Physician inserts needle into bone (optionally using drillable stylet); rotates grip to withdraw inner cannula, exposing tip to different tract locations for multi-site aspiration without repositioning. Aspiration cannula inserted into access needle; side holes minimize blood contamination. White indicators align side ports. Output is bone marrow/blood collected via standard 10mL luer lock syringe. Used in clinical settings by physicians to obtain tissue specimens.
Clinical Evidence
Bench testing only. Structural integrity testing (resistance to breakage and torque) and simulated use testing were performed to validate design outputs against inputs. Sterilization validation (ethylene oxide, 10^-6 SAL), distribution simulation, and shelf-life studies were conducted.
Indicated for aspiration of bone marrow or autologous blood using a standard piston syringe.
Regulatory Classification
Identification
A gastroenterology-urology biopsy instrument is a device used to remove, by cutting or aspiration, a specimen of tissue for microscopic examination. This generic type of device includes the biopsy punch, gastrointestinal mechanical biopsy instrument, suction biopsy instrument, gastro-urology biopsy needle and needle set, and nonelectric biopsy forceps. This section does not apply to biopsy instruments that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.
{1}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left, there is a symbol representing the Department of Health & Human Services - USA. To the right of this symbol, there is a blue square with the letters "FDA" in white. Next to the blue square, the words "U.S. FOOD & DRUG ADMINISTRATION" are written in blue.
February 21, 2023
Ranfac Corporation Eric Kreuz Vice President/Quality & Regulatory Affairs 30 Doherty Avenue Avon, Massachusetts 02322
## Re: K223612
Trade/Device Name: Ranfac Lateral Access Bone Marrow Aspiration Needle (CRVS-BMA-LA) Regulation Number: 21 CFR 876.1075 Regulation Name: Gastroenterology-Urology Biopsy Instrument Regulatory Class: Class II Product Code: KNW Dated: January 25, 2023 Received: January 25, 2023
## Dear Eric Kreuz:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for
{2}------------------------------------------------
devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Mark Digitally signed by Mark Trumbore -S Trumbore - Date: 2023.02.21 S On behalf of Long Chen. Ph.D. Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
# Indications for Use
Submission Number (if known)
K223612
Device Name
Ranfac Lateral Access Bone Marrow Aspiration Needle (CRVS-BMA-LA)
Indications for Use (Describe)
The Lateral Access Bone Marrow Aspiration Needle is intended for use for aspiration of bone marrow or autologous blood using a standard piston syringe.
Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW."
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for Ranfac. The logo is a shield shape with the left half filled with a gold gradient and the right half is white. Above the shield is the word "Ranfac" in a blue sans-serif font with a blue line above it.
### 510(k) Summary
The following information is provided as required by 21 CFR § 807.87 for Ranfac Lateral Access Bone Marrow Aspiration Needle 510(k) premarket notification. In response to the Safe Medical Devices Act of 1990, the following is a summary of the information upon which the substantial equivalence determination is based.
#### 1. SUBMITTER
| Sponsor: | Ranfac Corp.<br>30 Doherty Ave<br>Avon, MA |
|-----------------|--------------------------------------------------------------------------------------------------------------------------------------|
| Contact Person: | Eric Kreuz<br>Vice President/Regulatory & Quality<br>Ph: 1-508-584-4400, Ext. 137<br>Fax: 1-508-584-8588<br>Email: ekreuz@ranfac.com |
| Date: | December 3, 2022 |
#### 2. DEVICE
| Device Proprietary Name: | Ranfac Lateral Access (LA) Bone Marrow Aspiration<br>Needle |
|----------------------------|-------------------------------------------------------------|
| Common or Usual Name: | Biopsy Needle |
| Regulatory Classification: | Class II |
| Classification Name: | Instrument, biopsy |
| Regulation: | 21 CFR 876.1075 |
| Device Regulation Panel: | Gastroenterology/Urology |
| Device Product Code: | KNW |
#### 3. PREDICATE DEVICE
| Proprietary Name: | Ranfac Marrow Cellution Needle (K150563) |
|----------------------------|------------------------------------------|
| Common Name: | Biopsy Needle |
| Classification Name: | Instrument, Biopsy |
| Regulation: | 21 CFR 876.1075 |
| Product Code: | KNW |
| Regulatory Classification: | Class II |
| Classification Panel: | Gastroenterology/Urology |
{5}------------------------------------------------
Special 510(k) - Ranfac Lateral Access Bone Marrow Aspiration Needle Ranfac Corp. 510(k) Summary - K223612
Image /page/4/Picture/1 description: The image shows the Ranfac logo. The logo is a shield shape that is divided into two halves. The left half of the shield is a gradient of gold, while the right half is white. Above the shield is a horizontal blue line, and above that is the word "Ranfac" in blue text. There is a registered trademark symbol in the bottom right corner of the shield.
#### 4. DEVICE DESCRIPTION
The Ranfac Lateral Access BMA Needle subject of this premarket notification is a manual, sterile disposable needle intended for the purpose of aspirating bone marrow or autologous blood. The device consists of an Access Needle for bone marrow penetration and an Aspirator Cannula through which the aspirate is obtained. The Access Needle is comprised of a 7-gauge outer cannula and an 11-gauge inner cannula with an overall length of 6.5 inches when the inner cannula is fully extended. The outer cannula has an adjustable grip, and a blunt tip that rests on the surface of the bone. The inner cannula has a ground point for bone/bone marrow penetration and side ports for marrow aspiration. When the molded grip of the Access Needle is turned clockwise the inner cannula is withdrawn into the outer cannula exposing the tip to a different location within the needle tract allowing for additional bone marrow aspiration acquisition without having to reposition the needle.
During use, the Access Needle is inserted into bone with the aid of a sharp or drillable stylet. Once the needle is properly positioned and the stylet removed, a 14-gauge Aspiration Cannula is inserted into the Access Needle. The distal tip of the Aspiration Cannula is closed and provided with side holes to allow the physician to aspirate from the sides of the needle (to minimize blood within the aspirate). A white indicator on both the Access Needle and the Aspiration Cannula is provided in order to allow for alignment of the two components' side ports (for aspiration of bone marrow).
An exchangeable Drillable Stylet is provided for use with the device which can mate to a standard surgical drill to aid bone penetration if needed. The device is also packaged with a commercially available, 510k cleared standard 10mL luer lock hypodermic syringe for collection of bone marrow aspirate.
### ட். INDICATIONS FOR USE
The Ranfac Lateral Access BMA Needle is for use for aspiration of bone marrow or autologous blood using a standard piston syringe.
### 6. COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICE
The Ranfac FLEX BMA Needle is substantially equivalent in intended use, principles of operation and fundamental technological characteristics to the legally marketed predicate Marrow Cellution device (K150563). The table below summarizes the similarities and differences in design, materials and dimensions between the subject and predicate device.
{6}------------------------------------------------
Image /page/5/Picture/1 description: The image shows the logo for Ranfac. The logo is a shield shape with the word "Ranfac" in teal at the top right. The shield is divided vertically, with the left side being a textured gold color and the right side being white. There is a registered trademark symbol at the bottom of the shield.
| Table 1. Comparison of the Ranfac LA BMA Needle Technological Characteristics with |
|------------------------------------------------------------------------------------|
| Predicate Ranfac Marrow Cellution Needle |
| | Ranfac LA BMA Needle<br>(This Submission) | Marrow Cellution (K150563) |
|----------------------------------------|------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------|
| Regulation<br>Number | 21 CFR §876.1075 | 21 CFR §876.1075 |
| Intended Use | For harvest of bone marrow tissue | For harvest of bone marrow tissue |
| Indication for Use | For use for aspiration of bone marrow or autologous blood using a standard piston syringe. | For use for aspiration of bone marrow or autologous blood using a standard piston syringe. |
| Performance<br>Characteristics | Needle bores into bone to access marrow cavity | Needle bores into bone to access marrow cavity |
| Overall Product<br>Design | Single-use, sterile disposable needle to acquire tissue specimen. | Single-use, sterile disposable needle to acquire tissue specimen. |
| Mechanics of<br>Operation | Manual instrument | Manual instrument |
| Patient/Tissue<br>Contact<br>Materials | Stainless steel and plastic | Stainless steel and plastic |
| Access Needle<br>Gauge and<br>Length | Adjustable length between 4.8 and 6.5 inches. | Needle provided in lengths of 4 and 6 inches. |
| Needle Cutting<br>Tip Configuration | Stylet has beveled 3-sided trocar tip and needle cannula has 5-sided grind tip for penetration | Stylet has beveled 3-sided trocar tip and needle cannula has 5-sided grind tip for penetration. |
| Aspiration<br>Cannula<br>Configuration | 14-gauge 304 stainless steel hollow cannula with one set of side ports and ABS luer. | 14-gauge 304 stainless steel hollow cannula with one set of side ports and ABS luer. |
| Aspiration<br>Cannula Length | 11.5 inches nominal | 9.2 inches nominal |
| Handle<br>Configuration | T-Shaped Handle Configuration | T-Shaped Handle Configuration |
| Adjustable Depth<br>Guide | Yes | Yes |
| Drillable Stylet | Yes | No |
| 10mL<br>Hypodermic<br>Syringe Provided | Yes | Yes |
| Packaging | Tyvek/Mylar Pouch | Tyvek/Mylar Pouch |
| | Ranfac LA BMA Needle<br>(This Submission) | Marrow Cellution (K150563) |
| Sterilization | Supplied Sterile via Ethylene Oxide<br>validated to $10^{-6}$ Sterility Assurance<br>Level | Supplied Sterile via Ethylene Oxide<br>validated to $10^{-6}$ Sterility<br>Assurance Level |
| Shelf-Life | 6 months | 5 years |
{7}------------------------------------------------
Image /page/6/Picture/1 description: The image shows the Ranfac logo. The logo features the word "Ranfac" in a teal sans-serif font, positioned above a stylized shield. The shield is divided vertically, with the left half being a textured gold color and the right half being white. A thin teal line is above the shield.
Table 1. Comparison of the Ranfac LA BMA Needle Technological Characteristics with Predicate Ranfac Marrow Cellution Needle
### 7. PERFORMANCE TESTING
Structural integrity testing was conducted on the Ranfac LA BMA Needle demonstrating robustness and appropriateness of the design. All samples met or exceeded acceptance criteria. Strength specifications (resistance to breakage and torque) are similar to those of the predicate Ranfac Marrow Cellution Bone Marrow Aspiration Needle. Additionally, simulated use testing was performed to validate that the design output of the Ranfac LA BMA Needle met design input requirements.
The Ranfac LA BMA Needle is provided sterilized with ethylene oxide to an assurance level of 10th in a validated cycle. Distribution simulation and shelf-life studies have been conducted to demonstrate that the device maintains its performance and the packaging will maintain its sterile barrier over the entirety of the intended shelf life.
## Conclusion
Ranfac has demonstrated that the Ranfac LA BMA Needle is substantially equivalent in fundamental design, function, device materials, packaging, sterilization, operating principle, intended use/ indication for use and fundamental technology as the legally marketed predicate device Ranfac Marrow Cellution Needle, which was cleared under Premarket Notification K150563 on May 22, 2015.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.