K223611 · Alphatec Spine, Inc. · MAX · Mar 29, 2023 · Orthopedic
Device Facts
Record ID
K223611
Device Name
Calibrate LTX Interbody System
Applicant
Alphatec Spine, Inc.
Product Code
MAX · Orthopedic
Decision Date
Mar 29, 2023
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.3080
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Calibrate LTX Interbody System is indicated for spinal fusion procedures from T1 to S1 in skeletally mature patients for the treatment of a symptomative disc disease (DDD), degenerative spondylolisthesis, spinal stenosis, and or thoracic disc herniation (myelopathy with or without axial pain) at one or two adjacent levels. DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies. Additionally, the Calibrate LTX Interbody System can be used as an adjunct to fusion in patients diagnosed with multilevel degenerative scoliosis and sagittal deformity. The Calibrate LTX Interbody System is intended for use on patients who have had at least six months of non-operative treatment. It is intended to be used with autograft and/or allogenic bone graft comprised of cortical, cancellous, and/or corticocancellous bone, and/or demineralized allograft bone marrow aspirate and supplemental fixation systems that are cleared by FDA for use in the thoracic and lumbar spine. Calibrate LTX spacers expanded greater than 20° must be used with the provided bone screws in addition to supplemental fixation. Calibrate LTX spacers without integrated fixation features may be used with AMP-X System as integrated fixation in addition to supplemental fixation.
Device Story
Lordotic expandable thoracolumbar intervertebral body fusion system; inserted via lateral or anterolateral approach. Device consists of titanium alloy spacers, inserters, trials, and instruments. Spacers feature anti-migration teeth and grit-blast surfaces; provide lordotic expansion, sagittal alignment restoration, and indirect decompression. Includes AMP-X anti-migration plate system with fixation plates and locking screws. Used in clinical settings by surgeons; requires supplemental fixation. Spacers expanded >20° require integrated bone screws. Benefits include stabilization of spinal segments and facilitation of fusion in patients with degenerative conditions or deformities.
Clinical Evidence
No clinical data. Substantial equivalence supported by non-clinical bench testing including dynamic axial compression, dynamic compression-shear, and static screw push-out testing per ASTM F2077.
Technological Characteristics
Materials: Titanium alloy (Ti-6Al-4V ELI) per ASTM F136. Design: Lordotic expandable interbody spacers with anti-migration teeth and grit-blast endplates. Connectivity: None. Energy: None (mechanical). Sterilization: Not specified.
Indications for Use
Indicated for skeletally mature patients requiring spinal fusion (T1-S1) for symptomatic degenerative disc disease, degenerative spondylolisthesis, spinal stenosis, thoracic disc herniation, multilevel degenerative scoliosis, or sagittal deformity at 1-2 levels, following 6 months of failed non-operative treatment.
Regulatory Classification
Identification
An intervertebral body fusion device is an implanted single or multiple component spinal device made from a variety of materials, including titanium and polymers. The device is inserted into the intervertebral body space of the cervical or lumbosacral spine, and is intended for intervertebral body fusion.
Special Controls
*Classification.* (1) Class II (special controls) for intervertebral body fusion devices that contain bone grafting material. The special control is the FDA guidance document entitled “Class II Special Controls Guidance Document: Intervertebral Body Fusion Device.” See § 888.1(e) for the availability of this guidance document.(2) Class III (premarket approval) for intervertebral body fusion devices that include any therapeutic biologic (e.g., bone morphogenic protein). Intervertebral body fusion devices that contain any therapeutic biologic require premarket approval.
(c)
*Date premarket approval application (PMA) or notice of product development protocol (PDP) is required.* Devices described in paragraph (b)(2) of this section shall have an approved PMA or a declared completed PDP in effect before being placed in commercial distribution.
Predicate Devices
IdentiTi and Transcend Interbody Systems (K222028)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
March 29, 2023
Alphatec Spine, Inc. Neha Mohindroo Regulatory Affairs Associate 1950 Camino Vida Roble Carlsbad, California 92008
Re: K223611
Trade/Device Name: Calibrate LTX Interbody System Regulation Number: 21 CFR 888.3080 Regulation Name: Intervertebral Body Fusion Device Regulatory Class: Class II Product Code: MAX, PHM, OVD Dated: December 22, 2022 Received: December 23, 2022
Dear Neha Mohindroo:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal
{1}------------------------------------------------
statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Katherine D. Kavlock -S
for
Brent Showalter, Ph.D. Assistant Director DHT6B: Division of Spinal Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K223611
Device Name Calibrate LTX Interbody System
### Indications for Use (Describe)
The Calibrate LTX Interbody System is indicated for spinal fusion procedures from T1 to S1 in skeletally mature patients for the treatment of a symptomative disc disease (DDD), degenerative spondylolisthesis, spinal stenosis, and or thoracic disc herniation (myelopathy with or without axial pain) at one or two adjacent levels. DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies.
Additionally, the Calibrate LTX Interbody System can be used as an adjunct to fusion in patients diagnosed with multilevel degenerative scoliosis and sagittal deformity.
The Calibrate LTX Interbody System is intended for use on patients who have had at least six months of non-operative treatment. It is intended to be used with autograft and/or allogenic bone graft comprised of cortical, cancellous, and/or corticocancellous bone, and/or demineralized allograft bone marrow aspirate and supplemental fixation systems that are cleared by FDA for use in the thoracic and lumbar spine.
Calibrate LTX spacers expanded greater than 20° must be used with the provided bone screws in addition to supplemental fixation. Calibrate LTX spacers without integrated fixation features may be used with AMP-X System as integrated fixation in addition to supplemental fixation.
| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|
| <div> <span> <span style="font-size: 16px;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) </span> </div> | <div> <span> □ Over-The-Counter Use (21 CFR 801 Subpart C) </span> </div> |
|--------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------|
|--------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------|
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for "atec". The "a" is green, while the "tec" is black. There is a trademark symbol to the right of the "c". There is a black line underneath the logo.
This 510(k) summary of safety and effectiveness is being submitted in accordance with the requirements of 21 CFR 807.92.
| I. | SUBMITTER: | Alphatec Spine, Inc.<br>1950 Camino Vida Roble<br>Carlsbad, CA 92008<br>Phone: (760) 431-9286 |
|-----|------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | Contact Person: | Neha Mohindroo<br>Associate, Regulatory Affairs<br>Contact Phone: (760) 356-6596 |
| | Date Summary Prepared: | February 28, 2023 |
| II. | DEVICE | |
| | Name of Device:<br>Common or Usual Name:<br>Classification Name: | Calibrate LTX Interbody System<br>Intervertebral body fusion device<br>Intervertebral fusion device with bone graft, lumbar<br>Intervertebral fusion device with integrated fixation. lumbar<br>Intervertebral fusion device with bone graft, thoracic |
| | Regulation Number:<br>Regulatory Class:<br>Product Code: | 21 CFR 888.3080<br>Class II<br>MAX, OVD, PHM |
#### III. LEGALLY MARKETED PREDICATE DEVICES
| 510(k) | Product Code | Trade Name | Manufacturer |
|------------------------------|------------------|---------------------------------------------|----------------|
| Primary Predicate Device | | | |
| K222028 | MAX, OVD,<br>PHM | IdentiTi and Transcend Interbody<br>Systems | Alphatec Spine |
| Additional Predicate Devices | | | |
| K222455 | MAX, OVD,<br>PHM | Calibrate LTX Interbody System | Alphatec Spine |
#### IV. DEVICE DESCRIPTION
The subject Calibrate LTX Interbody System is a lordotic expandable thoracolumbar intervertebral body fusion system designed to be inserted through a lateral or anterolateral surgical approach. The subject interbody spacers are manufactured from titanium alloy (Ti-6Al-4V ELI) per ASTM F136. The Calibrate LTX System consists of a variety of shapes and sizes of interbody spacers, inserters, trials, and general instruments to create lordotic
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for a company called "atec". The "a" is green, and the "tec" is dark blue. There is a trademark symbol next to the "c". There is a black line underneath the logo.
expansion, restore sagittal alignment, and provide indirect decompression. Implants are offered with anti-migration teeth and grit-blast treatment on the bone-contacting endplate surfaces. Certain Calibrate LTX Interbody System offerings also accept fixation bone screws manufactured from titanium alloy per ASTM F136.
This purpose of this 510(k) is to receive clearance for an anti-migration plate ("AMP-X") which may be used with certain LTX interbody offerings. AMP-X includes fixation plates. center locking screws, and bone screws manufactured from titanium alloy per ASTM F136. Additionally, the purpose of this 510(k) is to receive clearance for expanded indications for the treatment of degenerative spondylolisthesis, multilevel degenerative scoliosis, spinal stenosis, and sagittal deformity, and use of allogenic bone consisting of cortical bone, as well as allogenic bone consisting of demineralized allograft bone with bone marrow aspirate.
#### V. INDICATIONS FOR USE
The Calibrate LTX Interbody System is indicated for spinal fusion procedures from T1 to S1 in skeletally mature patients for the treatment of a symptomatic degenerative disc disease (DDD), degenerative spondylolisthesis, spinal stenosis, and/or thoracic disc herniation (myelopathy and/or radiculopathy with or without axial pain) at one or two adjacent levels. DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies.
Additionally, the Calibrate LTX Interbody System can be used as an adjunct to fusion in patients diagnosed with multilevel degenerative scoliosis and sagittal deformity.
The Calibrate LTX Interbody System is intended for use on patients who have had at least six months of non-operative treatment. It is intended to be used with autograft and/or allogenic bone graft comprised of cortical, cancellous, and/or corticocancellous bone, and/or demineralized allograft bone with bone marrow aspirate and supplemental fixation systems that are cleared by FDA for use in the thoracic and lumbar spine.
Calibrate LTX spacers expanded greater than 20° must be used with the provided bone screws in addition to supplemental fixation. Calibrate LTX spacers without integrated fixation features may be used with AMP-X System as integrated fixation in addition to supplemental fixation.
#### VI. TECHNOLOGICAL COMPARISON TO PREDICATES
The technological design features of the subject implants were compared to the predicate IdentiTi and Transcend Interbody Systems (K222028) and additional predicate Calibrate LTX Interbody System (K222455) in intended use, indications for use, design, function, and technology, and it was demonstrated that they are substantially equivalent. Calibrate LTX interbody implants are substantial equivalent in design, material, and function as predicate Calibrate LTX interbody implants (K222455). The new modular AMP-X plate that must be used with Calibrate LTX is substantially equivalent in design, materials, and
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the logo for a company called "atec". The "a" is green, while the "tec" is dark blue. There is a trademark symbol to the right of the "c". There is a black line underneath the logo.
function to the integrated fixation features of the predicate Calibrate LTX (K222455) as well as LIF AMP that must be used with ATEC LIF interbody implants (K222028).
#### VII. PERFORMANCE DATA
Nonclinical testing performed on the Calibrate LTX Interbody System supports substantial equivalence to other predicate devices. The following testing was performed:
- · Dynamic Axial Compression (per ASTM F2077)
- · Dynamic Compression-Shear (per ASTM F2077)
- · Static Screw Push-Out Testing
Since the technological characteristics of the subject Calibrate LTX Interbody System is substantially equivalent to the predicate systems, no further clinical or non-clinical testing is required to support the expanded indications for use of the subject system.
## Clinical Information
Not applicable; determination of substantial equivalence is not based on an assessment of clinical performance data.
#### CONCLUSION VIII.
Based upon the information provided in this 510(k) submission, it has been determined that the subject devices are substantially equivalent to legally marketed devices in regard to indications for use, intended use, design, technology, and performance,
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.