K223522 · Xiamen Intretech, Inc. · FLL · Jun 20, 2023 · General Hospital
Device Facts
Record ID
K223522
Device Name
NC° Thermometer (Gen 3), Model Number: NCTG3
Applicant
Xiamen Intretech, Inc.
Product Code
FLL · General Hospital
Decision Date
Jun 20, 2023
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.2910
Device Class
Class 2
Attributes
Pediatric
Indications for Use
The NC° Thermometer (Gen3) is used orally for the intermittent measurement and monitoring of human body temperature. The device can be used by adults and children over the age of 5 years old.
Device Story
NC° Thermometer (Gen3) is a battery-powered, predictive digital thermometer for oral temperature measurement. It utilizes an NTC thermistor in the probe tip to sense resistance changes, which are processed via a predictive algorithm to estimate body temperature. The device features a local screen for display, three user buttons (power, setting, history), and an audio buzzer. It supports BLE and NFC connectivity to transmit temperature data and timestamps to the Natural Cycles smartphone application. Designed for home use by adults and children (5+), it provides an alternative to manual monitoring. Healthcare providers or users view the output on the thermometer screen or the app to track temperature trends, aiding in clinical or personal health decision-making.
Clinical Evidence
Clinical validation study per ISO 80601-2-56 included 105 participants (32 febrile; 64 female, 41 male; all >5 years old). Results confirmed the device meets clinical accuracy requirements for an adjusted thermometer.
Technological Characteristics
Predictive digital thermometer; NTC thermistor sensor; 3V CR2032 battery; materials: SUS316L, ABS, PMMA, SMMA; BLE and NFC connectivity; oral measurement; accuracy ±0.1°C to ±0.2°C depending on range; meets ISO 80601-2-56, ASTM E1112, AAMI/ANSI ES 60601-1, IEC 60601-1-2, and IEC 60601-1-11.
Indications for Use
Indicated for oral intermittent measurement and monitoring of human body temperature in adults and children over 5 years old.
Regulatory Classification
Identification
A clinical electronic thermometer is a device used to measure the body temperature of a patient by means of a transducer coupled with an electronic signal amplification, conditioning, and display unit. The transducer may be in a detachable probe with or without a disposable cover.
Special Controls
(1) Device is not a clinical thermometer with telethermographic functions;
(2) Device is not a clinical thermometer with continuous temperature measurement functions; and
(3) Appropriate analysis and testing (such as that outlined in the currently FDA-recognized editions, as appropriate, of ISO 80601-2-56, “Medical electrical equipment—Part 2-56: Particular requirements for basic safety and essential performance of clinical thermometers for body temperature measurement,” or ASTM E1965, “Standard Specification for Infrared Thermometers for Intermittent Determination of Patient Temperature,” or ASTM E1112, “Standard Specification for Electronic Thermometer for Intermittent Determination of Patient Temperature,” or ASTM E1104, “Standard Specification for Clinical Thermometer Probe Covers and Sheaths”) must validate specifications and performance of the device.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo for the U.S. Food and Drug Administration (FDA). The logo includes the FDA acronym in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text. To the left of the FDA logo is the Department of Health & Human Services logo.
June 20, 2023
Xiamen Intretech Inc. % Megan Callanan Regulatory and Compliance Manager Natural Cycles 510 5th Avenue 3rd Floor New York, New York 10036
Re: K223522
Trade/Device Name: NCº Thermometer (Gen 3), Model Number: NCTG3 Regulation Number: 21 CFR 880.2910 Regulation Name: Clinical Electronic Thermometer Regulatory Class: Class II Product Code: FLL Dated: May 19, 2023 Received: May 19, 2023
Dear Megan Callanan:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE(@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
David Walloschek
David Wolloscheck, Ph.D. Assistant Director DHT3C: Division of Drug Delivery and General Hospital Devices, and Human Factors OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K223522
Device Name NCº Thermometer (Gen3), Model Number: NCTG3
#### Indications for Use (Describe)
The NC Thermometer (Gen3) is used orally for the intermittent measurement and monitoring of human body temperature. The device can be used by adults and children over the age of 5 years old.
| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|
Prescription Use (Part 21 CFR 801 Subpart D)
X Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# K223522
# 510(k) Summary: NC° Thermometer (Gen3)
| Applicant: | Xiamen Intretech Inc.<br>No.100 Dongfu West Road<br>Haicang District, Xiamen<br>Fujian, China Zip:361027 |
|-------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Applicant Contact: | Name: Sunny Sun<br>Phone: 86-15750706715<br>Email: xmshj@intretech.com |
| Correspondent: | Name: Megan Callanan<br>Phone: (216)744-4524<br>Email: Megan.callanan@naturalcycles.com |
| Date Prepared: | June 17, 2023 |
| Trade Name: | NC° Thermometer (Gen3),<br>Model Number: NCTG3 |
| Common Name: | Clinical electronic thermometer |
| Proposed Class: | Class II |
| Classification Name: | Clinical electronic thermometer |
| Regulation Number: | 21 CFR 880.2910 |
| Product Code: | FLL |
| Predicate Device: | K173730 KSA-110 Clinical electronic thermometer from KINSA,INC ,<br>marketed as the Kinsa QuickCare Thermometer. |
| Device Description: | The purpose of this traditional 510(k) premarket notification is to introduce<br>NC° Thermometer (Gen3). NC° Thermometer (Gen3) is a battery powered,<br>thermistor- based Bluetooth Low Energy (BLE) and Near Field<br>Communication (NFC) enabled adjusted mode digital thermometer that<br>uses a predictive algorithm for the measurement and monitoring of human<br>body temperature. Body temperature is measured orally with the NC°<br>Thermometer. Temperature measurements are displayed on the<br>thermometer and can also (optionally) be transmitted to the Natural Cycles<br>application on the user's smartphone by women over the age of 18. The<br>compatible application displays the thermometer temperature value. The |
| | thermometer is reusable for home use on adults and children ages 5 and<br>above. The thermometer is made of biocompatible metals and resins. The<br>thermometer consists of a Negative Temperature Coefficient (NTC)<br>thermistor located in the probe tip to sense human body temperature,<br>three input user facing buttons, an output buzzer for audio indications, a<br>screen for user display, and back lights for illuminating the buttons. |
| Indications for<br>Use: | The NC° Thermometer (Gen3) is used orally for the intermittent<br>measurement and monitoring of human body temperature. The device can<br>be used by adults and children over the age of 5 years old. |
{4}------------------------------------------------
# Summary of Technical Characteristics:
The technology of the NC° Thermometer (Gen3) is identical to the predicate device in terms of its principle of operation, thermistor, use of a prediction algorithm, and probe tip design. The industrial design is similar to the predicate except for the outer shell form.
| Element of<br>comparison | Subject Device<br>NC° Thermometer (Gen3)<br>K223522 | Predicate Device<br>Kinsa QuickCare<br>Thermometer<br>K173730 | Verdict |
|-----------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------|
| Thermometer<br>type | Predictive digital | Predictive digital | Same |
| Indications for<br>Use | The NC° Thermometer<br>(Gen3) is used orally for the<br>intermittent measurement<br>and monitoring of human<br>body temperature. The<br>device can be used by<br>adults and children over the<br>age of 5 years old. | The Kinsa QuickCare<br>Thermometer is used for the<br>intermittent measurement<br>and monitoring of human<br>body temperature, orally,<br>rectally and under the arm.<br>The device is for the adult<br>and pediatric population. | Similar1 |
| Fundamental<br>technology &<br>Operating<br>principle | Utilizes an NTC thermistor<br>located in the probe tip to<br>sense human body<br>temperature when placed in<br>the desired measuring site<br>(mouth). The resulting<br>change in resistance is<br>sensed and monitored in<br>order to through use of | Utilizes an NTC thermistor<br>located in the probe tip to<br>sense human body<br>temperature when placed in<br>the desired measuring site<br>(mouth, underarm, and<br>rectum). The resulting change<br>in resistance is sensed and<br>monitored in order to through | Similar2 |
| | | | |
| | predictive algorithms<br>estimate human body<br>temperature. | use of predictive algorithms<br>estimate human body<br>temperature. | |
| Sensor | Thermistor based | Thermistor based | Same |
| Signal<br>processing<br>and display | Internal firmware and local<br>screen display<br>- Also able to transfer<br>transmit data to an auxiliary<br>device for secondary display | Internal firmware and local<br>LCD display<br>- Also able to transfer<br>transmit data to an auxiliary<br>device for secondary display | Same |
| Wireless<br>Interface | Bluetooth Low Energy (BLE)<br>and Near Field<br>Communication (NFC) | Bluetooth Low Energy (BLE) | Different1 |
| Compatible<br>Application<br>Name | Natural Cycles | Kinsa | Different2 |
| Compatible<br>Application<br>Display of<br>Thermometer<br>Related Data | Temperature, timestamp | Temperature, timestamp, tips<br>from our in-house clinician,<br>share your illness timeline<br>with another caregiver or<br>your doctor | Different3 |
| Power<br>requirements | Battery powered CR2032<br>(3V) | Battery powered (CR2032<br>(3V) | Same |
| Patient<br>Contacting<br>Materials | SUS316L, ABS, PMMA,<br>ABS+SMMA | SU304 Stainless Steel, ABS,<br>PMMA | Similar3 |
| Scale | °F / °C | °F / °C | Same |
| Measurement<br>locations | Oral | Oral, Axillary and Rectal | Different4 |
| Measurement<br>Range | 32 to 42 °C (89.6 to 107.6<br>°F) | 32 to 42.8 °C (89.6 to 109.2<br>°F) | Different5 |
| Number of<br>measurements<br>that can be<br>saved and<br>viewed on<br>thermometer | 10 | N/A- No history button on<br>thermometer | Different6 |
| Thermometer<br>Buttons | Power button, setting<br>button, history button | Power button | Different7 |
| Operating<br>Environment | 15 to 40°C<br>< 95% relative humidity non- | 15 to 35 °C<br>15 to 85% Relative humidity | Different8 |
| Storage<br>Environment | -25 to 50 °C<br>≤ 95% relative humidity<br>70kPa to 106kPa | -25 to 70 °C<br>Up to 90% relative humidity,<br>non-condensing, up to 50 hPa | Different9 |
| Accuracy | ±0.2°C/0.3°F (Temperature less<br>than 37°C/98.0°F)<br>±0.1°C/0.2°F (Temperature<br>37.0°C to 39.0°C/98.0°F to<br>102.0°F)<br>±0.2°C/0.3°F (Temperature<br>greater than 39.0°C/102.0°F) | ± 0.2°C within measurement<br>range of 32 to 42.8°C (89.6<br>to 109.2°F) | Different10 |
| Response<br>Time | 40 seconds nominally | 8 seconds nominally | Different11 |
| Resolution of<br>Display | 0.01 °C / 0.01°F | 0.1 °C / 0.1°F | Different12 |
| Performance | Meets ISO 80601-2-56:2017 | Meets ISO 80601-2-56 | Same |
| Biocompatibility | Meets ISO 10993-1:2018,<br>ISO 10993-5:2009, ISO<br>10993-10:2010 | Meets ISO 10993-1:2009,<br>ISO 10993-5:2009, ISO<br>10993-10:2010 | Same |
| Electrical<br>Safety | Meets AAMI/ANSI ES<br>60601-1:2005+A2:2020 | Meets AAMI/ANSI ES 60601-<br>1:2005+A1:2012 | Same |
| Electromagnetic<br>Disturbances | Meets IEC 60601-1-2:2014 | Meets IEC 60601-1-2:2014 | Same |
{5}------------------------------------------------
Traditional 510(k)
{6}------------------------------------------------
Traditional 510(k)
# Discussion of Similar Elements of Comparison:
- 1. Similar indications for use: The subject device can only be used for adults and children over the age of 5. The predicate device indications for use do not specify a pediatric age limit. As the subject device is intended for oral use only, the intended user was limited to children over the age of 5 years old and adults. Performance testing demonstrates that the subject device meets ISO 80601-2-56 and the different pediatric age limit does not raise any new safety and effectiveness questions.
- 2. Similar fundamental technology and operating principle: Similar hardware is used in both devices. Performance testing demonstrates that the subject device meets ISO 80601-2-56 and ASTM E1112. The differences in fundamental technology and operating principle do not raise any new safety and effectiveness questions.
- 3. Similar patient contacting materials: The thermometer materials are similar except for the probe tip and buttons. The subject device probe tip uses an equivalent biocompatible stainless steel and the subject device buttons include an additional thermoplastic, SMMA. All skin contacting materials in the subject device have been tested successfully for biocompatibility: cytotoxicity in accordance with AAMI/ANSI/ISO 10993-5:2009/(R)2014, Biological Evaluation of
{7}------------------------------------------------
Medical Devices Part 5: Tests for In Vitro Cytotoxicity, as well as irritation and sensitization in accordance with AAMI/ANSI/ISO 10993-10:2010/(R)2014. Biological Evaluation of Medical Devices Part 10: Tests for Irritation and Skin Sensitization. The materials used in the subject device do not raise any new questions of safety and effectiveness.
### Discussion of Different Elements of Comparison:
- 1. Different wireless interface: The subject device has NFC interface in addition to BLE. while the predicate device has BLE. NFC is a widely used technology and the subject device has passed electrical safety and electromagnetic compatibility testing in accordance with the FDA recognized consensus standards: AAMI/ANSI ES 60601-1:2005/(R)2012 + A1:2012, C1:2009/(R)2012 + A2:2010/(R)2012, IEC 60601-1-2:2014, IEC60601-1-11 Edition 2.1 2020-07, FCC Part 15 Subpart C & 15.247, and FCC Part 15 Subpart C & 15.225. Additionally, performance data consisting of software verification, cybersecurity analysis, wireless coexistence testing, usability testing demonstrate the safe and effective use of this feature. The essential performance of the subject device is not compromised as the primary display of the temperature readings on the thermometer's local screen is independent of the wireless transmission. The addition of the NFC wireless interface does not raise different questions of safety and effectiveness.
- 2. Different compatible application name: Subject and predicate thermometers are compatible with different applications. Both applications are downloaded onto smartphones and allow users to see temperature and timestamp. The essential performance of the subject device is not compromised as the primary display of the temperature readings on the thermometer's local screen is independent of the wireless transmission. The subject device has passed electrical safety and electromaqnetic compatibility testing in accordance with the FDA recognized consensus standards: AAMI/ANSI ES 60601-1:2005/(R)2012 + A1:2012, C1:2009/(R)2012 + A2:2010/(R)2012, IEC 60601-1-2:2014, IEC60601-1-11 Edition 2.1 2020-07, FCC Part 15 Subpart C § 15.247, and FCC Part 15 Subpart C § 15.225. Additionally, performance data consisting of software verification, cybersecurity analysis, wireless coexistence testing, usability testing demonstrate the safe and effective use of the subject device and compatible application.
- 3. Different compatible application display of temperature related data: The predicate thermometer's compatible app provides more information regarding illness but both apps function equivalently when receiving temperature and timestamp data and displaying it on the phone. The predicate thermometer's compatible application has additional features related to illness such as tips from clinician and illness timeline but these features do not impact the safe and effective use of the thermometer. The subject thermometer compatible app is a medical device while the predicate is not but this does not impact the app with respect to the thermometer. The subject device has passed electrical safety and electromagnetic compatibility testing in accordance with the FDA recognized
{8}------------------------------------------------
consensus standards: AAMI/ANSI ES 60601-1:2005/(R)2012 + A1:2012, C1:2009/(R)2012 + A2:2010/(R)2012, IEC 60601-1-2:2014, IEC60601-1-11 Edition 2.1 2020-07, FCC Part 15 Subpart C § 15.247, and FCC Part 15 Subpart C & 15.225. Additionally, performance data consisting of software verification, cybersecurity analysis, wireless coexistence testing, usability testing demonstrate the safe and effective use of the subject device and compatible application.
- 4. Different measurement locations: The subject device is used for oral measurement only. The route of temperature measurement for the subject device is a subset of the routes available using the predicate device. Both the subject and predicate devices are used for oral measurements. The subject device meets the requirements of ISO 80601-2-56 and ASTM E1112 for oral use and the limitation on measuring routes does not raise any different questions of safety and effectiveness.
- 5. Different measurement range: The subject device has a lower upper limit of measurement range than the predicate device. The ISO 80601-2-56 performance testing of the subject device demonstrated that the difference does not raise any new questions of safety and effectiveness.
- 6. Different number of measurements that can be saved and viewed on thermometer: There is no history button on the predicate device and thus no measurements can be saved and viewed on the thermometer. The subject device has a history button, but this does not impact the essential performance of the subject device. Software verification, design verification, and usability testing demonstrates that the history button and display of previous measurements on the thermometer does not raise any new questions of safety and effectiveness.
- 7. Different thermometer buttons: Compared to the predicate device, there are two more buttons (history button and settings button) on the subject device. The essential performance of the subject device is not compromised as temperature measurement and display of temperature readings is independent of History and Settings buttons. Performance data consisting of software verification, design verification, and usability testing demonstrates the safe and effective use of all buttons. It does not raise new questions of safety and effectiveness.
- 8. Different operating environment: The subject device has larger operating range and higher humidity requirements than the predicate device. The operating environment requirements on the subject device and the predicate device both meet the IEC 60601-1-11 and ISO80601-2-56 standards and performance testing conducted according to these standards demonstrates that the difference in operating environment does not raise any new questions of safety and effectiveness.
- 9. Different storage environment: The subject device has a smaller storage range and higher humidity limitation. The storage environment requirements on the subject device and the predicate device both meet the IEC 60601-1-11 and ISO80601-2-56 standards and performance testing conducted according to these standards demonstrates that the difference in storage environment does not raise any new questions of safety and effectiveness.
{9}------------------------------------------------
- 10. Different accuracy: The subject device meets ISO80601-2-56 and ASTM E1112 requirements, while the predicate device meets ISO80601-2-56. The subject device has higher accuracy than the predicate device from 37 to 39°C because ASTM E1112 has stricter accuracy requirements. Performance testing was conducted on the subject device according to ASTM E1112 and ISO80601-2-56. The difference does not raise any new performance or safety concerns.
- 11. Different response time: The subject device has longer measurement time than the predicate device, but usability testing showed the measuring time is well within user expectations and far below the 3 minutes that would be required without a predictive algorithm. This specification has been validated by the testing per ISO 80601-2-56 and ASTM E1112. The difference does not raise any new questions of safety and effectiveness.
- 12. Different resolution of display: The subject device has a higher resolution of display than the predicate device. Performance testing demonstrated that the subject device is in compliance with both ISO 80601-2-56 and ASTM E1112 requirements. The difference in resolution of display does not raise any new questions of safety and effectiveness.
## Testing to Demonstrate Substantial Equivalence:
### Non-clinical Testing
Non-clinical performance reports were provided to document verification and validation activities intended to demonstrate substantial equivalence of the subject device to the predicate device:
- 1. Design Verification results confirmed the device meets the product requirements set by and the performance standard requirements of ISO 80601-2-56:2017, Medical Electrical Equipment – Part 2-56: Particular Requirements For Basic Safety And Essential Performance Of Clinical Thermometers For Body Temperature Measurement and ASTM E 1112-00 (2018) Standard Specification for Electronic Thermometer for Intermittent Determination of Patient Temperature.
- 2. Software Verification and Validation results confirmed the firmware and software units meet the software requirements specifications and the system performs as intended. Software documentation is provided in accordance with the FDA "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices" (May 11, 2005).
- 3. Cybersecurity hazards and risks associated with the thermometer have been evaluated. Cybersecurity information is provided in accordance with the FDA Guidance "Content of Premarket Submission for Management of Cybersecurity in Medical Devices"
- 4. All skin contacting materials have been tested successfully for biocompatibility: cytotoxicity in accordance with AAMI/ANSI/ISO 10993-5:2009 /(R)2014, Biological Evaluation of Medical Devices Part 5: Tests for In Vitro Cytotoxicity, as well as irritation and sensitization in accordance with AAMI/ANSI/ISO 10993-
{10}------------------------------------------------
10:2010/(R)2014, Biological Evaluation of Medical Devices Part 10: Tests for Irritation and Skin Sensitization.
- 5. Electrical and Mechanical Safety as well as essential performance was confirmed through compliance testing to AAMI/ANSI ES 60601-1:2005/(R)2012, Medical Electrical Equipment - Part 1: General Requirements for Basic Safety and Essential Performance and IEC 60601-1-11:2015, Medical Electrical Equipment Part 1-11: Collateral Standard: Requirements for Medical Electrical Equipment and Medical Electrical Systems Used in the Home Healthcare Environment.
- 6. Electromagnetic Compatibility was confirmed through compliance testing to IEC 60601-1-2:2014, Medical Electrical Equipment - Part 1-2: Electromagnetic Disturbances – Requirements and Tests.
- 7. FCC compliance was confirmed through testing to FCC Part 15 Subpart C §15.247/ RSS 247.
- 8. Interoperability testing was completed to ensure that the thermometer is able to communicate with the Natural Cycles application as intended.
## Clinical Testing
To support the performance of the NC° Thermometer (Gen3), a clinical validation study was performed. The study evaluated the clinical performance of the NC° Thermometer (Gen3) to a reference clinical thermometer in accordance with ISO 80601-2-56. Clinical validation study reports were provided with this submission. The pivotal clinical study included 105 participants, 32 (30%) were febrile. All participants were over the age of 5 years old. 64 participants (61%) were female and 41 participants (39%) were male. Based on the clinical performance as documented in the pivotal clinical study, the NC° Thermometers (Gen3) was found to meet the ISO 80601-2-56 requirements of clinical accuracy for an adjusted thermometer.
## Conclusion:
Based on the design features, the use of established well known materials, feature comparisons, and indications for use, the NC° Thermometer (Gen3) has demonstrated substantial equivalence to the identified predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.