K223517 · Wuxi EA Medical Instruments Technologies Limited. · NXC · Jun 13, 2023 · Dental
Device Facts
Record ID
K223517
Device Name
Clear Aligner
Applicant
Wuxi EA Medical Instruments Technologies Limited.
Product Code
NXC · Dental
Decision Date
Jun 13, 2023
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 872.5470
Device Class
Class 2
Attributes
Therapeutic, Real-World Evidence, Pediatric
Real-World Evidence
Submission
Device
Sponsor
RWD Sources
RWE Use Summary
Key Tags
K223517 · Jun 13, 2023
Clear Aligner
Wuxi EA Medical Instruments Technologies Limited.
Retrospective clinical data
Retrospective clinical data was used to demonstrate that successful alignment of teeth and treatment of malocclusion was achieved with the proposed device when worn 20 to 22 hours a day for at least 10 days.
Retrospective study; Follow-up/Duration: 20 to 22 hours of daily wear for at least 10 days
Patients undergoing orthodontic treatment for misalignment and malocclusion
Not applicable for this study
Successful alignment of teeth
Indications for Use
The Proposed Device is indicated for use in the alignment or mixed dentition teeth through orthodontic treatment of misalignment and malocclusion, with 20 to 22 hours of daily wear for at least 10 days. The Proposed Device allows the option of forward repositioning of the mandible to a more optimal bite profile.
Device Story
Device consists of sequential thermoformed plastic aligners; worn 20-22 hours/day; removed for cleaning. Input: digital dental scans from dental professionals. Process: 3D models of sequential stages designed/approved by physician; 3D printed models thermoformed with dental thermoplastic sheets. Features: 'A6' mandibular repositioning for retrognathic Class II malocclusions. Output: physical aligners. Used in clinical settings; monitored by orthodontists/dentists. Benefit: incremental tooth movement via continuous gentle force; bite profile optimization.
Clinical Evidence
Retrospective clinical data provided. Study demonstrated successful tooth alignment and malocclusion correction with 20-22 hours daily wear for at least 10 days per stage.
Technological Characteristics
Materials: Multilayer thermoformed copolyester/polyurethane composite and thermoplastic polyurethane. Sensing/Actuation: Passive mechanical force via thermoformed plastic. Form factor: Sequential aligners. Connectivity: None. Sterilization: Non-sterile. Biocompatibility: ISO 10993-1 compliant.
Indications for Use
Indicated for orthodontic treatment of malocclusion and tooth alignment in patients with permanent or mixed dentition, including those requiring mandibular repositioning for retrognathic Class II malocclusions.
Regulatory Classification
Identification
An orthodontic plastic bracket is a plastic device intended to be bonded to a tooth to apply pressure to a tooth from a flexible orthodontic wire to alter its position.
{1}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
June 13, 2023
Wuxi EA Medical Instruments Technologies Limited. % Breanne Butler Regulatory Affairs Consultant Prime Path Medtech 1321 Upland Dr. Suite 6792 Houston, Texas 77043
Re: K223517
Trade/Device Name: Clear Aligner Regulation Number: 21 CFR 872.5470 Regulation Name: Orthodontic Plastic Bracket Regulatory Class: Class II Product Code: NXC Dated: May 12, 2023 Received: May 15, 2023
Dear Breanne Butler:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{2}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
# Michael E. Adjodha -
## S
Michael E. Adjodha, M.ChE. Assistant Director DHT1B: Division of Dental and ENT Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory. ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
#### Indications for Use
510(k) Number (if known)
K223517
Device Name Clear Aligner
Indications for Use (Describe)
Indicated for use in alignment of teeth through orthodontic treatment of malocclusion.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|--|
| | |
X Prescription Use (Part 21 CFR 801 Subpart D)
| | Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
### 510(k) Summary
#### 510(k) SUMMARY - K223517
A summary of 510(k) safety and effectiveness information for this special 510(k) in accordance with the requirements of 21 CFR 807.92.
| Submitter: | Wuxi EA Medical Instruments Technologies Limited<br>No.1619 Huishan Avenue, Huishan Economic Development Zone<br>Wuxi, Jiangsu Province<br>China |
|-----------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------|
| Company Contact Person:<br>Phone:<br>Email: | Jessica Luo, Regulatory Affairs, Angel Align<br>+86.0510.83591717(181)<br>luoyuqing@angelalign.com |
| Submission Correspondent:<br>Address:<br>Phone:<br>Email: | Breanne Butler, Regulatory Affairs Consultant<br>1321 Upland Dr. Suite 6792 Houston, TX 77043<br>860-810-5594<br>bbutler@primepathmedtech.com |
| Date Prepared: | November 15, 2022 |
| Proprietary Name: | Clear Aligner |
| Common Name: | Orthodontic plastic bracket. |
| Product Code: | NXC – Orthodontic plastic bracket. |
| Device Classification: | Class II, 21 CFR 872.5470 |
| Predicate Device: | Invisalign MAF Aligners (K181739) |
| Reference Predicate: | Clear Aligner (K203688) |
#### Device Description:
This premarket notification is submitted to notify the FDA of Wuxi EA Medical Technologies' intent of changes to the Clear Aligner (hereafter referred to as "Proposed Device") to the currently marketed reference device's, Clear Aligner (K203688, cleared October 8, 2021) (hereafter referred to as "Reference Device") indications for use to include mixed dentition, a wear time of at least 10 days, and mandibular repositioning features.
The Proposed Device consists of multiple stages of thermoformed plastic aligners designed to be worn in sequence to facilitate the movement of a patient's teeth to the final desired treatment position. The sequential stages of aligners introduce incremental move teeth by way of gentle
{5}------------------------------------------------
continuous force and reposition the mandible to achieve a more optimal bite profile. The aligners are to be worn 20 to 22 hours a day and are to be removed for cleaning. Patients can be prescribed with more than one aligner of different materials within a single stage.
The Proposed Device is designed from digital scans of a patient's dentition submitted by a dental health professional (e.g. dentist or orthodontist). Using the scan, sequential stages of dental models are designed and approved by a physician prior to physical manufacturing. The Proposed Device includes mandibular repositioning features known as "A6" on each upper and lower aligner. A6 is intended for use in patients with growing mandibles presenting with retrognathic Class II malocclusions in permanent dentition or stable late mixed dentition.
Once the treatment plan is reviewed and approved by a dental health professional, the 3D model of each stage from the treatment plan is manufactured by 3D printed models are then thermoformed over with a suitable dental thermoplastic sheet. The final manufactured aligner stages are then delivered to the patient. The patients' dental health professional then monitors their treatment from the placement of delivered initial aligner stage to the final aligner stage.
#### Intended Use:
The Proposed Device is indicated for use in the alignment or mixed dentition teeth through orthodontic treatment of misalignment and malocclusion, with 20 to 22 hours of daily wear for at least 10 days. The Proposed Device allows the option of forward repositioning of the mandible to a more optimal bite profile.
#### Comparison to Predicate and Reference Devices:
The Proposed Device is functionally equivalent to the following predicate device, Invisalign System with Mandibular Advancement Features (Invisalign MAF) (K181739, cleared October 26, 2018) (hereafter referred to as "Predicate Device"), and possesses minor differences to the previously cleared Reference Device. The following table demonstrates the functional specifications of the Proposed Device is substantially equivalent to the Predicate Device, minorly different to Reference Device, and raises no questions regarding safety and effectiveness of the device.
| | Subject Device: | Predicate Device: | Reference Device: | Comparison | |
|--------------------------------------|------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------|
| Specification | Clear Aligner | Invisalign MAF | Clear Aligner | Result | |
| | (K223517) | (K181739) | (K203688) | | |
| Regulation<br>Number | 21 CFR 872.5470 | 21 CFR 872.5470 | 21 CFR 872.5470 | Same | |
| Classification<br>Name | Orthodontic Plastic<br>Bracket | Orthodontic Plastic<br>Bracket | Orthodontic Plastic<br>Bracket | Same | |
| Product Code | NXC | NXC | NXC | Same | |
| Classification | Class II | Class II | Class II | Same | |
| OTC or Rx | Rx | Rx | Rx | Same | |
| Material | (1) Multilayer<br>thermoformed<br>copolyester and<br>polyurethan<br>composite and<br>(2) Thermoplastic<br>Polyurethane | Thermoplastic<br>polyurethane-<br>polyester composite<br>resin | (1) Multilayer<br>thermoformed<br>copolyester and<br>polyurethan<br>composite or<br>(2) Thermoplastic<br>Polyurethane | Similar to<br>Predicate (1),<br>Same as<br>Reference (1<br>& 2) | |
| | Specification | Subject Device:<br>Clear Aligner<br>(K223517) | Predicate Device:<br>Invisalign MAF<br>(K181739) | Reference Device:<br>Clear Aligner<br>(K203688) | Comparison<br>Result |
| | Material<br>Properties | Acceptable material<br>properties<br>established for use<br>as an aligner. | Acceptable<br>materials properties<br>established for use<br>as an aligner. | Acceptable<br>materials<br>properties<br>established for use<br>as an aligner. | Same |
| | Material<br>Testing | Tested for<br>performance in<br>accordance with<br>internal design<br>specification and<br>with the applicable<br>performance<br>standards. | Tested for<br>performance in<br>accordance with<br>internal design<br>specification and<br>with the applicable<br>performance<br>standards. | Tested for<br>performance in<br>accordance with<br>internal design<br>specification and<br>with the applicable<br>performance<br>standards. | Same as<br>Reference<br>Device |
| | Biocompatible | Biocompatible<br>according to ISO<br>10993-1 (Biological<br>evaluation of<br>medical devices –<br>Part 1: Evaluation<br>and testing within a<br>risk management<br>system). | Biocompatible<br>according to ISO<br>10993-1 (Biological<br>evaluation of<br>medical devices –<br>Part 1: Evaluation<br>and testing within a<br>risk management<br>system). | Biocompatible<br>according to ISO<br>10993-1 (Biological<br>evaluation of<br>medical devices –<br>Part 1: Evaluation<br>and testing within<br>a risk management<br>system). | Same |
| | Sterile | Non-sterile | Non-sterile | Non-sterile | Same |
| | Device<br>Description | Sequential<br>thermoformed<br>plastic aligners | Sequential<br>thermoformed<br>plastic aligners | Sequential<br>thermoformed<br>plastic aligners | Same |
| | Patient<br>Removable? | Yes | Yes | Yes | Same |
| | Indication for<br>Use | Indicated for use in<br>alignment of teeth<br>through orthodontic<br>treatment of<br>malocclusion. | Indicated for the<br>orthodontic<br>treatment of<br>malocclusion. | Indicated for the<br>alignment of teeth<br>during orthodontic<br>treatment of<br>malocclusion in<br>patients with<br>permanent<br>dentition (i.e. all<br>second molars). | Similar |
| | Intended Use | Orthodontic tooth<br>movement | Orthodontic tooth<br>movement | Orthodontic tooth<br>movement | Same |
| | Mode of<br>Action | Continuous gentle<br>force applied to<br>teeth following the<br>prescribed and<br>approved treatment | Continuous gentle<br>force applied to<br>teeth following the<br>prescribed and<br>approved treatment | Continuous gentle<br>force applied to<br>teeth following the<br>prescribed and<br>approved | Same |
| Specification | Subject Device:<br>Clear Aligner<br>(K223517) | Predicate Device:<br>Invisalign MAF<br>(K181739) | Reference Device:<br>Clear Aligner<br>(K203688) | Comparison<br>Result | |
| | plan to achieve<br>orthodontic<br>movement. | plan to achieve<br>orthodontic<br>movement | treatment plan to<br>achieve<br>orthodontic<br>movement | | |
| Lower Jaw<br>Adjustment<br>Mechanism | Mandibular<br>repositioning<br>features known as<br>"A6" | Mandibular<br>Advancement<br>Features also known<br>as "Precision Wings" | N/A | Same as<br>predicate<br>device | |
#### Table 5.1: Device Comparison
{6}------------------------------------------------
Wuxi EA Medical Instruments Technologies Limited
{7}------------------------------------------------
Wuxi EA Medical Instruments Technologies Limited
#### Comparison of Indications for Use to Predicate Devices and Previously Cleared Devices:
The Proposed Device in this submission is the same as the Predicate Device as they are both indicated for use in the alignment of teeth and mandibular repositioning through orthodontic treatment of malocclusion. It can be used for both permanent and mixed dentition with a wear time 20 to 22 hours a day for at least 10 days. The differences between the indications for use of the Proposed Device, Predicate Device, and the previously cleared Reference Device do not raise questions of substantial equivalence. The Proposed Device guides patients' teeth and mandible to their desired treated position by way of continuous gentle forces in sequential aligner stages. Thus, the Proposed Device can be considered substantially equivalent to the Predicate Device and Reference Device.
The difference between the Proposed Device in this submission and the Reference Device is the inclusion of mandibular repositioning and of teeth for treatment of malocclusion. The Proposed Device additionally differs from the Predicate Device in allowing the use of multiple aligner materials. The Thermoplastic Polyurethane material of the Proposed Device is the same as the Reference Device and the Copolyester and Polyurethane Multilayer material is similar to the Predicate Device and raises no questions in the safety or effectiveness of the Proposed Device, demonstratial equivalence to the Predicate Device.
#### Comparison of Technological Characteristics to Predicate Devices:
Based on the above comparison, the design, construction, and performance characteristics of the Proposed Device are similar to the Predicate Device. Thus, Proposed Device can be considered substantially equivalent to the Predicate Device.
#### Non-clinical performance testing:
The use of thermoplastic materials for sequential aligners intended to treat malocclusions has been well documented in scientific literature regarding incremental tooth moving forces. An internal manufacturing validation was performed to test the manufacturing process for Proposed Device. The robustness of the process was demonstrated from 3D printing through thermoforming.
The thermoplastic materials used for the Proposed Device has passed the required testing for material characterization. Material testing was conducted on the aligner materials according to internal design specification and with the applicable performance standards. Biocompatibility testing for the aligner materials, the only patient contacting aspect, was conducted in accordance with International Standard ISO 10993-1, "Biological evaluation of medical devices – Part 1: Evaluation and testing within a risk management process."
#### Clinical performance testing:
{8}------------------------------------------------
Retrospective clinical data has been provided the Proposed Device. The data collected for the study showed that successful alignment of teeth through orthodontic treatment of misalignment and malocclusion with at 20 to 22 hours of daily wear for at least 10 days was achieved when using the Proposed Device.
#### Conclusion:
Based on similarities in indications for use, technological characteristics, non-clinical performance data, the Proposed Device is substantially equivalent to the Predicate Device, Invisalign MAF (K181739) and the previously cleared Reference Device, Angel Align System (K203688).
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.